Investigating the Extraction Efficiency of Dioxins in Several Types of Aqueous Matrices by Automated Solid Phase Extraction Utilizing Disks

Size: px
Start display at page:

Download "Investigating the Extraction Efficiency of Dioxins in Several Types of Aqueous Matrices by Automated Solid Phase Extraction Utilizing Disks"

Transcription

1 Investigating the Extraction Efficiency of Dioxins in Several Types of Aqueous Matrices by Automated Solid Phase Extraction Utilizing Disks James Fenster and Michael Ebitson, Horizon Technology, Inc., Salem, NH USA Key Words Dioxins, SPE, Solid Phase Extraction, POPs Introduction Dioxins are of great environmental concern due to their teratogenic, mutagenic, and carcinogenic impact. These lipophilic compounds bioaccumulate in humans and wildlife and can have half lives of up to 132 years, making them persistent organic pollutants (POPs). Traditional SPE requires a filtration step to remove particulates when using SPE cartridges, because of clogging issues. The filtered particulates then require a separate extraction step, such as soxhlet extraction. Liquid-liquid Extraction (LLE) often has issues with emulsions that can adversely affect the proper extraction of the particulate matter. Using the automated 47-mm SPE-disk approach allows for the extraction to take place without the problems such as clogging, filtration or emulsions. The SPE disk allows for the dioxin compounds in water to be retained on divinylbenzene (DVB) disk packing while the water passes through. It also allows the particulate matter to sit on the top of the disk. This way, when the elution solvents are delivered to extract the disk, they also extract any dioxin compounds that may have attached to suspended particulates or sediment, allowing for extraction of the whole sample. Disks combine fast flow rates without compound breakthrough, allowing efficient extraction of particulate laden samples larger than 1 liter with excellent recoveries that result in lower detection limits. This application will focus on SPE extraction techniques, results and recoveries of dioxins in deionized water, particulate-laden river water samples and wastewater. Experimental SPE-DEX 4790 Automated Extraction System Extractions of samples were accomplished using the automated SPE-DEX 4790 with the Envision platform controller (Horizon Technology, Inc.). SPE extraction disks were Atlantic DVB disks (47 mm), and pre-filters. Extraction solvents were methylene chloride, acetone, methanol, reagent water, hexane, and nonane. Samples were dried using sodium sulfate and then cleaned up using a coupling column, containing silica gel, alumina and Florisil. Samples were concentrated on a mivac Evaporator (Genevac), and analytical analysis was carried out by high resolution gas chromatography (HRGC)/high resolution mass spectroscopy (HRMS). The samples and characterization information are summarized in Table 1.

2 Table 1. Summary of Samples Prepared Sample Type Number of Samples Sample Volume ml Disk Type ph Prefilter Large Particulates Fine Particulates Sediment DI ( LCS) DVB 2 No No No No River Water DVB 2 Yes No Yes No River Water DVB 2 Yes Yes Yes Yes Wastewater DVB 2 Yes Yes Yes No Results and Discussion Due to the persistent nature of these compounds great care needs to be taken to ensure all glassware and extraction system components are clean in order to assure no cross contamination between samples or interferences in method blanks or low level samples. A cleaning procedure was developed using organic solvents and mild detergent that assures the extractor s liquid flow path is free of contaminants between samples. Twelve samples as described in Table 1, were prepared, extracted, and analyzed in order to demonstrate the effectiveness of the SPE method and to compare matrix effects. The samples represent a variety of matrices, including distilled water, very fine particulate laden river samples, river samples containing fine particles and sediment, and finally two wastewater influent samples. All sample sizes were reduced compared to the usual 1 L of sample taken due to the use of the 47 mm extraction disks in the smaller holder. If larger samples were required for detection limit criteria, larger particulated samples could be run using the EZ Flow disk holder which allow the layering of filters above the SPE disk so the particles are retained on the surface of the filter for rinsing with the elution solvent, but flow through the SPE disk is not compromised. Dioxins and furans are known to bind to sediment or suspended particles so these types of samples were ideal as reference matrix spikes. Each of the 12 samples were spiked with 25 µl of a 20 pg/µl solution containing CDDs and CDFs and a 40 pg/µl for OCDD and OCDF. As specified in USEPA Method 1613 the samples were allowed to sit for 3 hours to equilibrate and become more representative of a native sample. The method in Table 2 was programmed into the SPE-DEX 4790 Envision controller for a 47mm disk and was used for all of the extraction runs in this study. P a ge 2 Table 2. SPE-DEX 4790 Extraction Method Steps Solvent Soak Air-Dry Pre-Wet 1 DCM 1 Min 1 Min Pre-Wet 2 Acetone 1 Min 1 Min Pre-Wet 3 MeoH 1 Min 4 sec Pre-Wet 4 Water 5 sec 2 sec Pre-Wet 5 Water 0 sec 0 sec Sample Introduction Air-Dry Time 5 min Rinse 1 Acetone 1 min 30 sec 30 sec Rinse 2 DCM 1 min 30 sec 30 sec Rinse 3 DCM 1 min 30 sec 30 sec Rinse 4 DCM 1 min 30 sec 30 sec Rinse 5 DCM 1 min 30 sec 1 min

3 The clean LCS samples were processed very quickly and passed through the disk at speeds up to 150 ml/min. A sample pre-filter was used with the particulate-laden river samples and the sediment-laden samples. The flow rates through these samples were slower than the LCS samples. Additionally, it was determined that increasing the soak times with the multiple rinse steps in the particulate-laden samples will optimize the extraction efficiency of the automated extractor. One of the unique features in the design of this extraction system is that the original sample bottle is automatically rinsed. Twenty-five milliliters of extract consisting of methylene chloride and acetone 80:20 was collected, dried with sodium sulfate and concentrated to 100 µl on the evaporation system. Solvent extracts were exchanged with 3.0 ml of hexane, twice to ensure complete solvent exchange takes place. The sample extracts were then brought up to 1 ml volume and cleanup was carried out with a coupling column with silica gel, alumina and florisil. Final concentration was carried out on the evaporation system to a volume of 100 µl, 25 µl of nonane is added and placed back on the concentrator until a final volume of 25 µl is reached. Results for these samples are shown in tables 4, 5, 6, and 7. Samples were analyzed by HRGC/HRMS under the following conditions. Instrument Condition HRGC/MS Column DB5 (60 m) Flow 1mL/min Column Temperature 180 o C to 190 o C at 2 o C/min then 190 o C to 240 o at 5 o C/min Sample Injection Volume 1.8 µl Run Time 52 minutes In order to know how much recovery would be lost to the drying and concentration steps, 25 ml methylene chloride was spiked with 25 µl of a 20 pg/µl solution containing CDDs and CDFs and a 40 pg/µl for OCDD and OCDF. These spikes were dried, treated with florisil, silica gel, and alumina and concentrated. The following table reports losses in recovery associated with these steps. Results for the concentrated methylene chloride spike is shown in Table 3. Losses were fairly small for these final cleanup and concentration steps with the exception of the hepta and octa chlorinated furans. The results show that recovery losses may occur in the clean-up and evaporation steps, even if the extraction step provides a high recovery. Care must be taken in each step to ensure the best overall recovery of analyte. Several different types of water samples were examined to see if significant differences in recovery were observed. Samples with suspended particles yielded higher recoveries (Table 5) than the LCS (Table 4) by a small amount. Table 6 shows results for samples with both particulate and sediment, providing an increasingly challenging matrix. The suspended particles and sediment in the bottle did lower the recoveries due to the compounds natural tendency to adhere to soil particles. Wastewater influent samples, a challenging matrix, were analyzed and the results shown in (Table 6). Even in this matrix most of the compounds showed a reasonable recovery and good precision. Page 3

4 Table 3. Methylene Chloride Spike (Dried, Treated and Concentrated) Compounds %Recovery % Loss 2,3,7,8-TCDF ,2,3,7,8-PeCDF ,3,4,7,8-PeCDF ,2,3,4,7,8-HxCDF ,2,3,6,7,8-HxCDF ,3,4,6,7,8-HxCDF ,2,3,7,8,9-HxCDF ,2,3,4,6,7,8-HpCDF ,2,3,4,7,8,9-HpCDF OCDF ,3,7,8-TCDD ,2,3,7,8-PeCDD ,2,3,4,7,8-HxCDD ,2,3,6,7,8-HxCDD ,2,3,4,6,7,8-HpCDD OCDD Table 4. Results -DI Water LCS Compounds Sample 1 %Recovery Sample 2 %Recovery SD 2,3,7,8-TCDF ,2,3,7,8-PeCDF ,3,4,7,8-PeCDF ,2,3,4,7,8-HxCDF ,2,3,6,7,8-HxCDF ,3,4,6,7,8-HxCDF ,2,3,7,8,9-HxCDF ,2,3,4,6,7,8-HpCDF ,2,3,4,7,8,9-HpCDF OCDF ,3,7,8-TCDD ,2,3,7,8-PeCDD ,2,3,4,7,8-HxCDD ,2,3,6,7,8-HxCDD ,2,3,4,6,7,8-HpCDD OCDD Page 4

5 Table 5. Results - River Water with Fine Particulates Compounds 1 %Recovery 2 %Recovery 3 %Recovery 4 %Recovery 5 %Recovery 6 %Recovery Average SD 2,3,7,8-TCDF ,2,3,7,8-PeCDF ,3,4,7,8-PeCDF ,2,3,4,7,8-HxCDF ,2,3,6,7,8-HxCDF ,3,4,6,7,8-HxCDF ,2,3,7,8,9-HxCDF ,2,3,4,6,7,8-HpCDF ,2,3,4,7,8,9-HpCDF OCDF ,3,7,8-TCDD ,2,3,7,8-PeCDD ,2,3,4,7,8-HxCDD ,2,3,6,7,8-HxCDD ,2,3,4,6,7,8-HpCDD Table 6. Results - Fine Particulate and Sediment-Laden River Water Compounds Sample 1 %Recovery Sample 2 %Recovery SD 2,3,7,8-TCDF ,2,3,7,8-PeCDF ,3,4,7,8-PeCDF ,2,3,4,7,8-HxCDF ,2,3,6,7,8-HxCDF ,3,4,6,7,8-HxCDF ,2,3,7,8,9-HxCDF ,2,3,4,6,7,8-HpCDF ,2,3,4,7,8,9-HpCDF OCDF ,3,7,8-TCDD ,2,3,7,8-PeCDD ,2,3,4,7,8-HxCDD ,2,3,6,7,8-HxCDD ,2,3,4,6,7,8-HpCDD OCDD Page 5

6 Table 7. Results - Wastewater Influent Samples Compounds Sample 1 %Recovery Sample 2 %Recovery SD 2,3,7,8-TCDF ,2,3,7,8-PeCDF ,3,4,7,8-PeCDF ,2,3,4,7,8-HxCDF ,2,3,6,7,8-HxCDF ,3,4,6,7,8-HxCDF ,2,3,7,8,9-HxCDF ,2,3,4,6,7,8-HpCDF ,2,3,4,7,8,9-HpCDF OCDF ,3,7,8-TCDD ,2,3,7,8-PeCDD ,2,3,4,7,8-HxCDD ,2,3,6,7,8-HxCDD ,2,3,4,6,7,8-HpCDD OCDD Conclusion We were able to demonstrate an efficient SPE disk technology for the extraction of several types of known water matrices for the detection of tetra through octa-chlorinated dioxins and furans with analytical sensitivities to ppt (ng/l) levels. We were able to achieve acceptable recoveries with particulate-laden water samples without the means of another extraction apparatus such as SDS (Soxhlet/Dean-Stark extractor). Automated SPE extraction, using the SPE-DEX 4790, is very efficient, and provides a fast method of extracting these difficult compounds from water matrices. Future work on these compounds will involve the extraction of larger volumes of samples (1 L or greater) using the EZ Flow Disk Holder. These studies will enable the analysis of samples with heavy particulate matter to be processed in a timely and efficient manor while lowering detection limits to the range of ppq (pg/l). Acknowledgements: Horizon Technology, Inc. gratefully acknowledges the CARSO Group France. References: 1. EPA Method 1613, Tetra- through Octa-Chlorinated Dioxins and Furans by Isotope Dilution HRGC/HRMS, Revision B, October ISO Method 18073:2004 Determination of Tetra- through Octa-Chlorinated Dioxins and Furans by Isotope Dilution HRGC/HRMS, reviewed and confirmed in AN _01 16 Northwestern Drive, Salem, NH USA Tel: (603) Support-Service@horizontechinc.com

Automated Solid Phase Extraction of PAH Compounds Utilizing the SPE-DEX 5000

Automated Solid Phase Extraction of PAH Compounds Utilizing the SPE-DEX 5000 Automated Solid Phase Extraction of PAH Compounds Utilizing the SPE-DEX 5000 Melissa Lever, Horizon Technology, Inc., Salem, NH USA Key Words PAH, polycyclic aromatic hydrocarbons, EPA 870, water extraction,

More information

Ruud Addink and Tom Hall Toxic Report, LLC Watertown MA

Ruud Addink and Tom Hall Toxic Report, LLC Watertown MA Ruud Addink and Tom Hall Toxic Report, LLC Watertown MA POPs (PCDD/Fs, PCBs) continue to attract interest around the world due to strict regulations enforced in many countries Rapid and quality sample

More information

Determining Diesel and Oil Range Organics in Water using Automated Solid Phase Extraction

Determining Diesel and Oil Range Organics in Water using Automated Solid Phase Extraction Determining Diesel and Oil Range Organics in Water using Automated Solid Phase Extraction David Gallagher, Horizon Technology, Inc. Key Words Diesel, Oil, Organics in Water, Seawater, Solid Phase Extraction,

More information

Occupational Exposure of Dioxins in Taiwan

Occupational Exposure of Dioxins in Taiwan Occupational Exposure of Dioxins in Taiwan T. S. Shih 1, C. C. Lee 2 1 Institute of Occupational Safety and Health, Council of Labor Affairs, Taiwan 2 Research Center of Environmental Trace Toxic Substances,

More information

Determination of Monocrotophos, Diazinon, Malathion, EPN, and Methamidaphos from Aqueous Samples Using Atlantic HLB SPE Disks

Determination of Monocrotophos, Diazinon, Malathion, EPN, and Methamidaphos from Aqueous Samples Using Atlantic HLB SPE Disks Determination of Monocrotophos, Diazinon, Malathion, EPN, and Methamidaphos from Aqueous Samples Using Jim Fenster, Julie McGettrick, Horizon Technology, Inc., Salem, NH Introduction Monocrotophos, Diazinon,

More information

Determination of Endocrine Disrupting Hormones in Drinking and Wastewater

Determination of Endocrine Disrupting Hormones in Drinking and Wastewater Determination of Endocrine Disrupting Hormones in Drinking and Wastewater Michael Ebitson, Horizon Technology, Inc. Key Words Endocrine Disrupting Hormones, drinking water, SPE, LC/MS/MS Introduction Hormones

More information

Evaluation of PCBs and Dioxin/Furans (DxFs) Concentrations in Sediment from the Delaware Estuary

Evaluation of PCBs and Dioxin/Furans (DxFs) Concentrations in Sediment from the Delaware Estuary Evaluation of PCBs and Dioxin/Furans (DxFs) Concentrations in Sediment from the Delaware Estuary Delaware Estuary Science and Environmental Summit February 1, 2011 Gregory J. Cavallo, P.G. Thomas J. Fikslin,

More information

PAH Analysis in Drinking Water using Disk SPE Automated with the SPE-DEX 5000

PAH Analysis in Drinking Water using Disk SPE Automated with the SPE-DEX 5000 PAH Analysis in Drinking Water using Disk SPE Automated with the SPE-DEX 5000 Alicia Cannon, Horizon Technology, Salem, NH USA Key Words PAH, Polycyclic aromatic hydrocarbons, USEPA 550.1 Introduction

More information

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D*

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D* Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D* UCT Products: ECUNIPAH (ENVIRO-CLEAN Universal PAH/DRO 2000 mg/83 ml cartridge) Method Summary Method 8015D may be used

More information

Analysis of Drinking Waters for a Suite of Pesticides using GC with dual ECD (US EPA Method 508.1) after SPE Concentration

Analysis of Drinking Waters for a Suite of Pesticides using GC with dual ECD (US EPA Method 508.1) after SPE Concentration Analysis of Drinking Waters for a Suite of Pesticides using GC with dual ECD (US EPA Method 508.1) after SPE Concentration Eric Scott and Keith Ewing, Kentucky Department for Environmental Protection,

More information

PowerPrep TM. Automated Multi Column Sample Cleanup for Trace Analysis

PowerPrep TM. Automated Multi Column Sample Cleanup for Trace Analysis PowerPrep TM PowerPrep TM Automated Multi Column Sample Cleanup for Trace Analysis PowerPrep Multicolumn Sample Cleanup Automated Sample Cleanup for Trace Analysis The Powerprep Multi Column Sample Cleanup

More information

Extraction of POPs from Environmental Samples using Accelerated Solvent Extraction (ASE )

Extraction of POPs from Environmental Samples using Accelerated Solvent Extraction (ASE ) Extraction of POPs from Environmental Samples using Accelerated Solvent Extraction (ASE ) Frank Höfler* and Khalil Divan** *Dionex (Europe) Management AG Solothurner Strasse 259 4600 Olten Switzerland

More information

High sample throughput in a modern dioxin laboratory using dual acquisition

High sample throughput in a modern dioxin laboratory using dual acquisition High sample throughput in a modern dioxin laboratory using dual acquisition Wim Traag *1 and Jeroen Markesteijn *2 *1 DSP-Systems, Erichemseweg 7, 4117GG Erichem The Netherlands *2 Nofalab, Jan van Galenstraat

More information

Multi-Media, Multi-Concentration Chlorinated Biphenyl Congeners Analytical Service for Superfund (CBC01.0)

Multi-Media, Multi-Concentration Chlorinated Biphenyl Congeners Analytical Service for Superfund (CBC01.0) United States Office of OSWER Document 9240.1-48FS Environmental Protection Solid Waste and EPA Publication 540-F-05-002 Agency Emergency Response Draft Final - September 2005 Multi-Media, Multi-Concentration

More information

Dual Data Acquisition

Dual Data Acquisition 38th International Symposium on Halogenated Persistent Organic Pollutants (POPs) Thermo Lunch Seminar: DFS Magnetic Sector GC-HRMS Krakow, August 26 31, 2018 Dual Data Acquisition Experiences and Application

More information

CHROMATOGRAPHY/MASS SPECTROMETRY

CHROMATOGRAPHY/MASS SPECTROMETRY DETERMINATION OF 1,4-DIOXANE IN DRINKING WATER BY SOLID PHASE EXTRACTION (SPE) AND GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) WITH SELECTED ION MONITORING (SIM)* EPA Method 522 Part Number: EU52112M6

More information

EXTRACTION OF BIOLOGICAL TISSUES FOR AROMATIC AND CHLORINATED HYDROCARBONS

EXTRACTION OF BIOLOGICAL TISSUES FOR AROMATIC AND CHLORINATED HYDROCARBONS EXTRACTION OF BIOLOGICAL TISSUES FOR AROMATIC AND CHLORINATED HYDROCARBONS Donell. S. Frank, Bo Wang, Juan A. Ramirez, Susanne J. McDonald, Rebecca Price, and James M. Brooks TDI-Brooks International/B&B

More information

CO-DISTILLATION OF AGENT ORANGE AND OTHER PERSISTENT ORGANIC POLLUTANTS IN EVAPORATIVE WATER DISTILLATION

CO-DISTILLATION OF AGENT ORANGE AND OTHER PERSISTENT ORGANIC POLLUTANTS IN EVAPORATIVE WATER DISTILLATION CO-DISTILLATION OF AGENT ORANGE AND OTHER PERSISTENT ORGANIC POLLUTANTS IN EVAPORATIVE WATER DISTILLATION Jochen F. Müller 1, Caroline Gaus 1, Katie Bundred 1, Vincent Alberts 2, Michael R. Moore 1,2 and

More information

Experimental. Instrumentation

Experimental. Instrumentation Efficient Identification of Deoxynivalenol in the Global Wheat Food Chain Testing Unprocessed, Processed, and Consumer Wheat Products using Automated Solid Phase Extraction Cleanup and Quick High Performance

More information

Puget Sound Sediment Reference Material: Requesting, Analyzing, Validating and Reporting Data

Puget Sound Sediment Reference Material: Requesting, Analyzing, Validating and Reporting Data Puget Sound Sediment Reference Material: Requesting, Analyzing, Validating and Reporting Data Introduction The Puget Sound Sediment Reference Material (SRM) has been developed to help assess/evaluate measurement

More information

Urban vs. Wildland Wildfires: Implications for Assessing Dioxin & Furan Impacts. Heather Lord, Terry Obal

Urban vs. Wildland Wildfires: Implications for Assessing Dioxin & Furan Impacts. Heather Lord, Terry Obal Urban vs. Wildland Wildfires: Implications for Assessing Dioxin & Furan Impacts Heather Lord, Terry Obal Overview of Presentation 1. Introduction to Dioxins and Furans 2. Dioxin and Furan Levels from Wildland

More information

EVOLUTE ABN SPE Columns for Solid Phase Extraction of Environmental Samples

EVOLUTE ABN SPE Columns for Solid Phase Extraction of Environmental Samples TN136 EVOLUTE ABN SPE Columns for Solid Phase Extraction of Environmental Samples This Chemistry Data Sheet provides guidelines for the simultaneous extraction of acidic, basic and neutral compounds from

More information

Justine Barton, Environmental Review and Sediment Management Unit, EPA Region 10. Keith Strout, CB&I Federal Services, LLC, EPA QATS Program

Justine Barton, Environmental Review and Sediment Management Unit, EPA Region 10. Keith Strout, CB&I Federal Services, LLC, EPA QATS Program CB&I 2700 Chandler Avenue, Building C Las Vegas, NV 89120 Tel: +1 702 795 0515 Fax: +1 702 795 8210 www.cbi.com Technical Memorandum To: From: Through: Justine Barton, Environmental Review and Sediment

More information

SpeedExtractor E-916 Determination of Total Petroleum Hydrocarbons in Soil, Sediment, and Waste Samples using the SpeedExtractor E /2009

SpeedExtractor E-916 Determination of Total Petroleum Hydrocarbons in Soil, Sediment, and Waste Samples using the SpeedExtractor E /2009 SpeedExtractor E-916 Determination of Total Petroleum Hydrocarbons in Soil, Sediment, and Waste Samples using the SpeedExtractor E-916 019/2009 019/2009 SpeedExtractor E-916 SHORT NOTE Determination of

More information

Validation of Horizon Technology Disk Extraction Technology for US EPA Wastewater Method 625.1

Validation of Horizon Technology Disk Extraction Technology for US EPA Wastewater Method 625.1 Validation of Horizon Technology Disk Extraction Technology for US EPA Wastewater Method 625.1 Zoe Grosser, Alicia Cannon, Michael Ebitson, Melissa Lever, Horizon Technology, Salem, NH USA Nic Rasnake,

More information

Atmospheric Pressure Gas Chromatography: Background & Applications. Jody Dunstan MS Systems Evaluation, Manchester 2011 Waters Corporation 1

Atmospheric Pressure Gas Chromatography: Background & Applications. Jody Dunstan MS Systems Evaluation, Manchester 2011 Waters Corporation 1 Atmospheric Pressure Gas Chromatography: Background & Applications Jody Dunstan MS Systems Evaluation, Manchester 2011 Waters Corporation 1 What is APGC? 2011 Waters Corporation 2 What is APGC Waters APGC

More information

Investigation of Possible PCB Impact during the Akron Airdock May 18, 2006 Fire

Investigation of Possible PCB Impact during the Akron Airdock May 18, 2006 Fire Investigation of Possible PCB Impact during the Akron Airdock May 18, 2006 Fire Submitted to EPA Region 5 on June 23, 2006 Lockheed Martin 1210 Massillon Road Akron, Ohio 44315 SUMMARY Following the Akron

More information

Extraction of a Full Suite of Semivolatile Compounds from Drinking Water using Automated Solid Phase Extraction

Extraction of a Full Suite of Semivolatile Compounds from Drinking Water using Automated Solid Phase Extraction Extraction of a Full Suite of Semivolatile Compounds from Drinking Water using Automated Solid Phase Extraction Melissa Lever, Horizon Technology, Inc., Salem, NH, USA Key Words US EPA 525.2, SPE, solid

More information

PCDDs/PCDFs in Soils at a Former Tannery Site-- Profiles as Evidence of PCP Contamination

PCDDs/PCDFs in Soils at a Former Tannery Site-- Profiles as Evidence of PCP Contamination Proceedings of the Annual International Conference on Soils, Sediments, Water and Energy Volume 17 Article 13 2012 PCDDs/PCDFs in Soils at a Former Tannery Site-- Profiles as Evidence of PCP Contamination

More information

Nature of Dioxin Contamination - Regulatory History and Authorities. Dow Chemical Midland Facility. March 15, 2010

Nature of Dioxin Contamination - Regulatory History and Authorities. Dow Chemical Midland Facility. March 15, 2010 Nature of Dioxin Contamination - Regulatory History and Authorities Dow Chemical Midland Facility March 15, 2010 Al Taylor DNRE Dow Overview 1,900 acre facility in Midland, Michigan. Operated from 1890

More information

SmartPrep. Automated Cartridge Extractor. Horizon Technology, Inc. Salem, NH USA

SmartPrep. Automated Cartridge Extractor. Horizon Technology, Inc. Salem, NH USA Automated Cartridge Extractor Horizon Technology, Inc. Salem, NH USA Trace Organics Contaminants Sample Matrices Drinking Waters Municipal Water Bottled Water Reservoirs Waste Waters Storm Water Waste

More information

PCDD, PCDF, and Dioxin-like PCB in Herring Gull Eggs from the North Sea and Baltic Sea: Levels, Patterns, and Temporal Trends

PCDD, PCDF, and Dioxin-like PCB in Herring Gull Eggs from the North Sea and Baltic Sea: Levels, Patterns, and Temporal Trends PCDD, PCDF, and Dioxin-like PCB in Herring Gull Eggs from the North Sea and Baltic Sea: Levels, Patterns, and Temporal Trends Christa Schröter-Kermani 1, Erik Schmolz 1, Thomas Herrmann 2, and Olaf Päpke

More information

CHAR CHARACTERIZATION OF THERMOCHEMICAL TREATED KENTUCKY BLUEGRASS SEED SCREENINGS

CHAR CHARACTERIZATION OF THERMOCHEMICAL TREATED KENTUCKY BLUEGRASS SEED SCREENINGS CHAR CHARACTERIZATION OF THERMOCHEMICAL TREATED KENTUCKY BLUEGRASS SEED SCREENINGS S.M. Griffith and G.M. Banowetz The on-farm production of bioenergy from straw and seed screenings and the recycling of

More information

Standard Operation Procedure (SOP) for the Extraction and Instrument Analysis of Siloxanes from Cartridges

Standard Operation Procedure (SOP) for the Extraction and Instrument Analysis of Siloxanes from Cartridges Standard Operation Procedure (SOP) for the Extraction and Instrument Analysis of Siloxanes from Cartridges Created March 17, 2016 Updated April 14, 2016 Dr. Rachel Marek (Asst. Research Scientist) 4105

More information

Improved Analysis of 1,4-Dioxane in Drinking Water

Improved Analysis of 1,4-Dioxane in Drinking Water Improved Analysis of 1,4-Dioxane in Drinking Water Melissa Lever a, Michael Ebitson a and Philip Bassignani b a Horizon Technology, Salem NH USA b Alpha Analytical, Inc., Mansfield, MA USA Page 1 1, 4

More information

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water Application Note Environmental Authors Min Cai and Yun Zou Agilent Technologies Co. Ltd, 412 Ying Lun Road Waigaoqiao

More information

METHOD 3610B ALUMINA CLEANUP

METHOD 3610B ALUMINA CLEANUP METHOD 3610B ALUMINA CLEANUP 1.0 SCOPE AND APPLICATION 1.1 Alumina is a highly porous and granular form of aluminum oxide. It is available in three ph ranges (basic, neutral, and acidic) for use in chromatographic

More information

EPA S RISK ASSESSMENT PROCESS AND APPLICATION TO DIOXIN

EPA S RISK ASSESSMENT PROCESS AND APPLICATION TO DIOXIN EPA S RISK ASSESSMENT PROCESS AND APPLICATION TO DIOXIN Tittabawassee-Saginaw Rivers Contamination Community Advisory Group April 16, 2012 Helen Dawson, PhD U.S. Environmental Protection Agency dawson.helen@epa.gov

More information

Development of an industrial standard method for Total Petroleum Hydrocarbon (TPH) analysis

Development of an industrial standard method for Total Petroleum Hydrocarbon (TPH) analysis Development of an industrial standard method for Total Petroleum Hydrocarbon (TPH) analysis Vasilios G. Samaras email: vsamaras@tellab.ie Topics of presentation Introduction Aim of the Project Project

More information

METHOD 3500A ORGANIC EXTRACTION AND SAMPLE PREPARATION

METHOD 3500A ORGANIC EXTRACTION AND SAMPLE PREPARATION METHOD 3500A ORGANIC EXTRACTION AND SAMPLE PREPARATION 1.0 SCOPE AND APPLICATION 1.1 The 3500 Methods are procedures for quantitatively extracting nonvolatile and semivolatile organic compounds from various

More information

Nonylphenols and Ethoxylates in Water by LC/MS/MS

Nonylphenols and Ethoxylates in Water by LC/MS/MS Nonylphenols and Ethoxylates in Water by LC/MS/MS Organics Revision Date: Sept 15, 2017 Parameter MDLs and EMS Analyte Codes Analytical method Introduction Method summary Nonylphenol and Nonylphenol Ethoxylates

More information

Simultaneous Extraction of PAHs and PCBs from Tissue and Soil Samples Using a New Accelerated Solvent Extraction (ASE) Procedure

Simultaneous Extraction of PAHs and PCBs from Tissue and Soil Samples Using a New Accelerated Solvent Extraction (ASE) Procedure Simultaneous Extraction of PAHs and PCBs from Tissue and Soil Samples Using a New Accelerated Solvent Extraction (ASE) Procedure Richard Jack, Ph.D., Brett Murphy, Selvan Lingam, David Knowles, Bruce Richter

More information

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental Samples Michael S.Young, Jeremy C. Shia, Kim vantran, Kevin M. Jenkins and Masayo Yabo Waters Corporation 34 Maple Street,

More information

Extraction of Phthalates from Solid and Liquid Matrices

Extraction of Phthalates from Solid and Liquid Matrices Extraction of s from Solid and Liquid Matrices David E. Knowles, Brian C. Dorich, Richard E. Carlson, Brett J. Murphy, Eric S. Francis, Jennifer Paterson, and Bruce E. Richter; Thermo Fisher Scientific,

More information

Agilent Environmental Analysis

Agilent Environmental Analysis Agilent Environmental Analysis Providing innovation for your laboratory so together we can in Protect our future. Dale R Walker GCMS Application Scientist Huntsville Texas A Complex Relationship Environment,

More information

FULLY AUTOMATED SPE-DEX Horizon Technology. understands the. challenges today s. laboratories face. The automated. extraction systems

FULLY AUTOMATED SPE-DEX Horizon Technology. understands the. challenges today s. laboratories face. The automated. extraction systems FULLY AUTOMATED SPE-DEX 4790 The Horizon Technology SPE-DEX 4790 Automated Extraction System is a programmable, multipurpose, automated SPE system, capable of processing aqueous samples directly from their

More information

Measuring PCBs in Lake Ontario Surface Water Using Full-Scan Congener Methods

Measuring PCBs in Lake Ontario Surface Water Using Full-Scan Congener Methods Measuring PCBs in Lake Ontario Surface Water Using Full-Scan Congener Methods Implications for designing environmental assessment programs Frederick Luckey U.S. Environmental Protection Agency, Region

More information

Soil Dioxin and PAH Study Urban and Rural Study Area Results

Soil Dioxin and PAH Study Urban and Rural Study Area Results Soil Dioxin and PAH Study Urban and Rural Study Area Results By Roger McGinnis, Ph.D. Hart Crowser and Charles San Juan, LHG, Ecology Toxics Cleanup Program Study Objectives Quantify ambient soil dioxin,

More information

Kopeopeo Remediation Project

Kopeopeo Remediation Project Baseline Ambi ent Air Monitoring Summar y Kopeopeo Remediation Project Bay of Plenty Regional Council Kope Canal Road Ambient Air Monitoring Report IZ006800-0008-NN-RPT-002 Final 29 March 2016 Kope Canal

More information

FOOD PESTICIDE RESIDUES LABORATORY REFERENCE NO INSPECTOR DATE

FOOD PESTICIDE RESIDUES LABORATORY REFERENCE NO INSPECTOR DATE FOOD PESTICIDE RESIDUES LABORATORY REFERENCE NO INSPECTOR DATE TYPES OF PESTICIDE ANALYSES: OrganoChlorine Pesticides OrganoPhosphorus Pesticides Nitrogen Containing Pesticides Carbamates Herbicides Standard

More information

Analysis of Polychlorinated Biphenyls (PCBs) in Water Samples Collected February 28 March 1, 2000 from the Bayou Creek System. Wesley J.

Analysis of Polychlorinated Biphenyls (PCBs) in Water Samples Collected February 28 March 1, 2000 from the Bayou Creek System. Wesley J. Analysis of Polychlorinated Biphenyls (PCBs) in Water Samples Collected February 28 March 1, 2000 from the Bayou Creek System Wesley J. Birge David J. Price DRAFT REPORT February 13, 2001 Submitted to

More information

EVOLUTE ABN FOR EXTRACTION OF DRUGS FROM BIOLOGICAL FLUIDS

EVOLUTE ABN FOR EXTRACTION OF DRUGS FROM BIOLOGICAL FLUIDS Technical ote 3 EVLUTE AB FR EXTRACTI F DRUGS FRM BILGICAL FLUIDS EVLUTE Sample Preparation Products are a new generation of advanced polymeric solid phase extraction sorbents for the high throughput extraction

More information

Analytical Method for aldrin, endrin and dieldrin (Targeted to Agricultural, Animal and Fishery Products)

Analytical Method for aldrin, endrin and dieldrin (Targeted to Agricultural, Animal and Fishery Products) Analytical Method for aldrin, endrin and dieldrin (Targeted to Agricultural, Animal and Fishery Products) The target compound to be determined is aldrin, endrin and dieldrin. 1. Instrument A gas chromatograph

More information

Comparison Between Lipid Removing Methods Using SPMDs and GPC for the Analysis of PCDD/Fs in Pine Needles

Comparison Between Lipid Removing Methods Using SPMDs and GPC for the Analysis of PCDD/Fs in Pine Needles w y wz 8«( 1y) 13~17, 2005 J. of the Korean Society for Environmental Analysis ù PCDD/Fs SPMDs GPC w ½k Á w w y œw Comparison Between Lipid Removing Methods Using SPMDs and GPC for the Analysis of PCDD/Fs

More information

Method Implementation Document (MID 1948)

Method Implementation Document (MID 1948) Method Implementation Document (MID 1948) BS EN 1948 Parts 1-3: 2006 Stationary source emissions - Determination of the mass concentration of PCDDs/PCDFs and dioxin-like PCBs Part 1: Sampling of PCDDs/PCDFs

More information

Analysis of Atrazine in Drinking Water at the ppb Level Using New Agilent Reversed Phase LC Columns. Application. Author. Abstract.

Analysis of Atrazine in Drinking Water at the ppb Level Using New Agilent Reversed Phase LC Columns. Application. Author. Abstract. Analysis of Atrazine in Drinking Water at the ppb Level Using ew Agilent Reversed Phase LC Columns Application Environmental Author Rongjie Fu Agilent Technologies, Inc. 412 Ying Lun Road Pu Dong, Shanghai

More information

DIOXIN EMISSION FROM TWO OIL SHALE FIRED POWER PLANTS IN ESTONIA

DIOXIN EMISSION FROM TWO OIL SHALE FIRED POWER PLANTS IN ESTONIA DIOXIN EMISSION FROM TWO OIL SHALE FIRED POWER PLANTS IN ESTONIA Ole Schleicher 1, Allan Astrup Jensen 1, Thomas Herrmann 2, Ott Roots 3, Arvo Tordik 4 1 FORCE Technology, Gladsaxe Møllevej 15, DK-2860

More information

Preparation. Sample. Extraction Glassware Liquid/Liquid Extraction SUPELCO

Preparation. Sample. Extraction Glassware Liquid/Liquid Extraction SUPELCO 186 Liquid/Liquid Extraction US EPA methods specify solvent extraction procedures for semivolatile, pesticide, PCB, and dioxin contaminants in water ardous and haz waste samples. Although analysts encounter

More information

EFFECTIVE SPE STRATEGIES FOR LC/MS DETERMINATION OF SUDAN DYES IN CHILI PRODUCTS

EFFECTIVE SPE STRATEGIES FOR LC/MS DETERMINATION OF SUDAN DYES IN CHILI PRODUCTS EFFECTIVE SPE STRATEGIES FOR LC/MS DETERMINATION OF SUDAN DYES IN CHILI PRODUCTS KimVan Tran, Michael S. Young a Uwe Neue Waters Corporation, Milford, MA, USA. INTRODUCTION Recently there have been numerous

More information

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water. Pranathi R Perati Thermo Fisher Scientific, Sunnyvale, CA, USA

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water. Pranathi R Perati Thermo Fisher Scientific, Sunnyvale, CA, USA Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water Pranathi R Perati Thermo Fisher Scientific, Sunnyvale, CA, USA Overview A new method was developed to automate the capture

More information

Dioxin emission measurements at Fortum Klaipeda, Lithuania. 25 March Report number: r01. Relation number: 96 Edition: 01

Dioxin emission measurements at Fortum Klaipeda, Lithuania. 25 March Report number: r01. Relation number: 96 Edition: 01 Dioxin emission measurements at Fortum Klaipeda, Lithuania Report number: 8406514r01 Relation number: 96 Edition: 01 Project manager: A. van Meekeren Author: A. van Meekeren Edited: AvM Checked: Initials

More information

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water Johannes Ritter 1 and Pranathi R Perati 2 1 Thermo Fisher Scienti ic, Vienna, Austria, 2 Thermo Fisher Scienti ic, Sunnyvale,

More information

SPE-DEX 4790 Automated Extraction System. A Fully Automated Sample Preparation System. votre distributeur

SPE-DEX 4790 Automated Extraction System. A Fully Automated Sample Preparation System. votre distributeur SPE-DEX 4790 Automated Extraction System A Fully Automated Sample Preparation System votre distributeur FULLY AUTOMATED SPE-DEX 4790 The Horizon Technology SPE-DEX 4790 Automated Extraction System is a

More information

DioxinMonitoringSystem. General device details and features

DioxinMonitoringSystem. General device details and features DioxinMonitoringSystem General device details and features Page 1 DioxinMonitoringSystem Main features Automatic isokinetic sampling using EN 1948-1/TS 1948-5: dilution method Industrial design - application

More information

SPE in Disk Format. Experimental. Table 1 Comparison of Typical SPE Cartridge and Empore Disk

SPE in Disk Format. Experimental. Table 1 Comparison of Typical SPE Cartridge and Empore Disk SPE in Disk Format A technology has been developed by the 3M Company in which small (8 µm) chromatographic particles are enmeshed in a network of PTFE fibrils to form a strong, porous sheet or particle-loaded

More information

CLEAN SCREEN FASt FILTER AND SHOOT UCT FORENSICS. CLEAN SCREEN FASt SPE COLUMNS

CLEAN SCREEN FASt FILTER AND SHOOT UCT FORENSICS. CLEAN SCREEN FASt SPE COLUMNS FORENSICS UCT CLEAN SCREEN FASt FILTER AND SHOOT CLEAN SCREEN FASt SPE COLUMNS CLEAN SCREEN FASt SPE COLUMNS Since the introduction and implementation of LC/MS as a staple analytical tool in forensic laboratories,

More information

Extraction of Vitamin B7 (Biotin) from Serum Using EVOLUTE

Extraction of Vitamin B7 (Biotin) from Serum Using EVOLUTE Application Note AN880 Extraction of Vitamin B7 (Biotin) from Serum Using EVOLUTE EXPRESS ABN Prior to LC-MS/MS Analysis Page 1 Extraction of Vitamin B7 (Biotin) from Serum Using EVOLUTE EXPRESS ABN Prior

More information

Extraction Products Separator Paper 188 Solid Phase Extraction (SPE) 189 Extraction Thimbles 192. Extraction Products

Extraction Products Separator Paper 188 Solid Phase Extraction (SPE) 189 Extraction Thimbles 192. Extraction Products Separator Paper 188 Solid Phase Extraction (SPE) 189 192 : Used to determine the oil and fat content in food, an indispensable tool for food manufacturers and health practitioners. Also used to test for

More information

Application Note. Abstract. Authors. Introduction. Environmental

Application Note. Abstract. Authors. Introduction. Environmental EPA Method 538: Determination of Selected Organic Contaminants in Drinking Water by Direct Aqueous Injection with the Agilent 646 Triple Quadrupole LC/MS System Application Note Environmental Authors Imma

More information

EPA Drinking Water Method 544

EPA Drinking Water Method 544 OAWWA-2016 EPA Drinking Water Method 544 Jody A. Shoemaker, Daniel R. Tettenhorst, Armah de la Cruz Disclaimers: Mention of trade names or commercial products does not constitute endorsement or recommendation

More information

LC/MS/MS Determination of PFOS and PFOA in Water and Soil Matrices

LC/MS/MS Determination of PFOS and PFOA in Water and Soil Matrices Application Note Environmental LC/MS/MS Determination of PFOS and PFOA in Water and Soil Matrices Using an Agilent 19 Infinity II LC with Ultivo Tandem Mass Spectrometry Authors Wenlong Yang, Jing Guo,

More information

3.0 DETECTION LIMIT 3.1. See Procedure ALQAP Procedure for the Determination of Uncertainty

3.0 DETECTION LIMIT 3.1. See Procedure ALQAP Procedure for the Determination of Uncertainty 1.0 IDENTIFICATION 1.1. SM 2540D Total Suspended Solids dried at 103-105 C. 2.0 APPLICABLE MATRIX 2.1. Drinking Water 3.0 DETECTION LIMIT 3.1. See Procedure ALQAP 011 - Procedure for the Determination

More information

Alex J. Zhang, Pranathi Perati, and Monika Verma; Thermo Fisher Scientific, Sunnyvale, CA, USA

Alex J. Zhang, Pranathi Perati, and Monika Verma; Thermo Fisher Scientific, Sunnyvale, CA, USA Pesticides in My Beverage Screening (and Subsequent Quantification) of Pesticides in Beverages Originating from Leaves, Grapes, Grasses, and the Hydrologic Cycle (Tea, Wine, Milk, and Water) Using Automated

More information

Sensitive Determination of Hexavalent. Brian De Borba, Lipika Basumallick, Jeffrey Rohrer

Sensitive Determination of Hexavalent. Brian De Borba, Lipika Basumallick, Jeffrey Rohrer Sensitive Determination of Hexavalent Chromium in Drinking Water Brian De Borba, Lipika Basumallick, Jeffrey Rohrer Outline Why do we need a sensitive method for hexavalent chromium analysis? U.S. EPA

More information

Interlaboratory Validation Study for the Analysis of Nitrosamines in Drinking Water using GC-MS/MS

Interlaboratory Validation Study for the Analysis of Nitrosamines in Drinking Water using GC-MS/MS Interlaboratory Validation Study for the Analysis of Nitrosamines in Drinking Water using GC-MS/MS Diana Wong, PhD National Environmental Monitoring Conference Washington, D.C. August 7, 2017 1 Outline

More information

TSERAWG Technical Guidance

TSERAWG Technical Guidance Page 1 of 7 TSERAWG Technical Guidance This paper has been technically reviewed and endorsed by the Tri-Service Environmental Risk Assessment Working Group for use by Army, Navy, and Air Force Personnel

More information

Guideline for the Implementation of Canada-wide Standards for Emissions of Mercury and of Dioxins and Furans. and

Guideline for the Implementation of Canada-wide Standards for Emissions of Mercury and of Dioxins and Furans. and GUIDELINE A-8 Guideline for the Implementation of Canada-wide Standards for Emissions of Mercury and of Dioxins and Furans and Monitoring and Reporting Requirements for Municipal Waste Incinerators Biomedical

More information

MICROWAVE SOLVENT EXTRACTION TECHNIQUES FOR VARIOUS REGULATORY COMPOUNDS

MICROWAVE SOLVENT EXTRACTION TECHNIQUES FOR VARIOUS REGULATORY COMPOUNDS MICROWAVE SOLVENT EXTRACTION TECHNIQUES FOR VARIOUS REGULATORY COMPOUNDS J.R. Fish, FAS Analytical Services, Monroe, NC M.D. Miller, CEM Corp., Matthews, NC Presented at the 1996 Pittsburgh Conference

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 30 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 2.1 Water Samples 2.2 Soil Samples 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 3.1 Sample Storage 3.2 Holding Times 4.0

More information

METHOD 9013A CYANIDE EXTRACTION PROCEDURE FOR SOLIDS AND OILS

METHOD 9013A CYANIDE EXTRACTION PROCEDURE FOR SOLIDS AND OILS METHOD 9013A CYANIDE EXTRACTION PROCEDURE FOR SOLIDS AND OILS SW-846 is not an analytical training manual. Therefore, method directions assume that they will be followed by individuals formally trained

More information

Phosphorus, Reactive (Orthophosphate)

Phosphorus, Reactive (Orthophosphate) Method 8048 DR/2400 * Adapted from Standard Methods for the Examination of Water and Wastewater ** Procedure is equivalent to USEPA method 365.2 and Standard Method 4500-P-E for wastewater. Phosphorus,

More information

Determination of N-nitrosodimethylamine in drinking water by UPLC-MS/MS

Determination of N-nitrosodimethylamine in drinking water by UPLC-MS/MS Journal of Environmental Sciences 2010, 22(10) 1508 1512 Determination of N-nitrosodimethylamine in drinking water by UPLC-MS/MS Wanfeng Wang 1, Jianying Hu 2, Jianwei Yu 1, Min Yang 1, 1. State Key Lab

More information

PROCEDURE FOR THE EXTRACTION AND PURIFICATION OF ORGANOCHLORINE PESTICIDES (OCP) IN FISH SAMPLES

PROCEDURE FOR THE EXTRACTION AND PURIFICATION OF ORGANOCHLORINE PESTICIDES (OCP) IN FISH SAMPLES Page 1 of 6 PROCEDURE FOR THE PURIFICATION OF 1. Introduction 2. Objective 3. Scope 4. Implementation 4.1. Extraction 4.2. Purification 4.3. Transfer to vial 4.4. Addition of the internal or syringe standard

More information

Performance of the CAPE Technologies DF1 Dioxin/Furan Immunoassay Kit for Soil and Sediment Samples

Performance of the CAPE Technologies DF1 Dioxin/Furan Immunoassay Kit for Soil and Sediment Samples United States Office of Research and EPA/540/R-08/002 Environmental Protection Development February 2008 Agency Washington, DC 20460 Performance of the CAPE Technologies DF1 Dioxin/Furan Immunoassay Kit

More information

Energized Dispersive Guided Extraction

Energized Dispersive Guided Extraction Energized Dispersive Guided Extraction EDGE Worldwide Patents Pending The new leading EDGE of sample preparation. The EDGE has revolutionized the extraction process for sample preparation. Extract up to

More information

Sources of dioxins in Baltic Sea herring

Sources of dioxins in Baltic Sea herring Sources of dioxins in Baltic Sea herring A modelling study for identification of dioxin sources and quantification of their temporal and spatial impacts Institutionen för vatten och miljö, SLU, Rapport

More information

Crystallization & Filtration

Crystallization & Filtration Crystallization & Filtration Purification of Salicyclic Acid and Sodium Chloride Separation processes Liquid-liquid extraction Adsorption Filtration Solid-liquid extraction (leaching) Elution chromatography

More information

EDGE. Automated Extraction System

EDGE. Automated Extraction System EDGE Automated Extraction System EDGE Worldwide Patents Pending Automates the technologies of PFE and dspe. Faster than Soxhlet, more automated than QuEChERS, and simpler than other automated solvent extraction

More information

for Acclaim Carbamate Column

for Acclaim Carbamate Column for Acclaim Carbamate Column Product Manual for Acclaim Carbamate Page 1 of 13 Product Manual for Acclaim Carbamate Analytical Columns 5 µm, 4.6 x 250 mm, P/N 072924 3 µm, 4.6 x 150 mm, P/N 072925 3 µm,

More information

Prepacked, Disposable Büchner Funnels for Medicinal Chemistry

Prepacked, Disposable Büchner Funnels for Medicinal Chemistry Prepacked, Disposable Büchner Funnels for Medicinal Chemistry An Trinh, Todd Williams, Dave Bell, and Michael Ye Supelco, 595 North Harrison Road, Bellefonte, PA 16823 T404044 GUR Abstract/Introduction

More information

Institute of Paper Science and Technology Atlanta, Georgia IPST TECHNICAL PAPER SERIES

Institute of Paper Science and Technology Atlanta, Georgia IPST TECHNICAL PAPER SERIES Institute of Paper Science and Technology Atlanta, Georgia IPST TECHNICAL PAPER SERIES NUMBER 476 PRE- AND POST-TREATMENTS OF PULPS TO ELIMINATE DIOXINS D.R. DIMMEL, L.B. SONNENBERG, K.B. RIGGS, G. PITTS,

More information

ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER

ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER ENVIRONMENTAL ANALYSIS ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER Solutions for Your Analytical Business

More information

EPA Method 540: Selected Organic Contaminants Using Agilent Plexa Cartridges and the Agilent 6460 Triple Quadrupole LC/MS

EPA Method 540: Selected Organic Contaminants Using Agilent Plexa Cartridges and the Agilent 6460 Triple Quadrupole LC/MS EPA Method 540: Selected Organic Contaminants Using Agilent Plexa Cartridges and the Agilent 6460 Triple Quadrupole LC/MS Application Note Environmental Authors Don Noot and Ralph Hindle Vogon Laboratory

More information

Multiresidue Screening of Agricultural Chemicals in a Food Matrix using Varian SPE and Triple Quadrupole LC-MS/MS

Multiresidue Screening of Agricultural Chemicals in a Food Matrix using Varian SPE and Triple Quadrupole LC-MS/MS Multiresidue Screening of Agricultural Chemicals in a Food Matrix using Varian SPE and Triple Quadrupole LC-MS/MS Eugene Chang, Anita DerMartirosian, Guang Li, Nigel Simpson, and Ritu Arora Varian, Inc.

More information

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters

More information

STEPS TO DMR-QA SUCCESS. BY: Marcy Bolek

STEPS TO DMR-QA SUCCESS. BY: Marcy Bolek STEPS TO DMR-QA SUCCESS BY: Marcy Bolek marcy@alloway.com Presentation Overview Purpose of DMR-QA Preparation DMR-QA Ethics Analysis Ammonia BOD Residual Chlorine ph TSS Quality Control Reporting What

More information

Comparison of ASTM D7979 PFAS compounds by the ASTM method versus Solid Phase Extraction

Comparison of ASTM D7979 PFAS compounds by the ASTM method versus Solid Phase Extraction Comparison of ASTM D7979 PFAS compounds by the ASTM method versus Solid Phase Extraction William Lipps, Eurofins Eaton Analytical Brahm Prakash, Shimadzu Scientific Instruments Monrovia South Bend www.eatonanalytical.com

More information

Comparison of Three Methods for the Analysis of PFAS

Comparison of Three Methods for the Analysis of PFAS Comparison of Three Methods for the Analysis of PFAS William Lipps, Eurofins Eaton Analytical Monrovia South Bend www.eatonanalytical.com Full Service Laboratories Eurofins Lancaster Laboratories Env.

More information

3 Empore. High Performance Extraction Disk Cartridges October Answers to Frequently Asked Questions. Product Description

3 Empore. High Performance Extraction Disk Cartridges October Answers to Frequently Asked Questions. Product Description 3 Empore High Performance Extraction Disk Cartridges October 1998 Answers to Frequently Asked Questions Product Description What is the configuration within the Empore Extraction Disk Cartridge? An Empore

More information

UCT ENVIRO-CLEAN HL DVB

UCT ENVIRO-CLEAN HL DVB E N V I R O UCT ENVIRO-CLEAN HL DVB ENVIRO-CLEAN HL DVB ENVIRO-CLEAN HL DVB UCT Part# ENVIRO-CLEAN HL DVB extraction columns are manufactured from an extremely clean, and highly cross-linked divinylbenzene

More information