1 SYSTRONIK Elektronik u. Systemtechnik GmbH Description and User Manual for Flue Gas Analyser Read the instructions before using the instrument! Consider security advices! Keep instructions for further questions! Version:

2

3

4 Contents 1. General instructions Saftey instructions Battery and device disposal Application area Product Description Technical Data General Operating Information Measurement Principle Physical Specification Calculation Formulae System Maintenance & Instrument Care Display Symbols Key Pad Functions Program Start Calibration Phase Fuel Selection Flue Gas Measurement Program Print Data End of Measurement Program Time / Date Connection-schema Gas-Processing & Watertrap unit Notices Appendix... 30

5

6 1. General instructions 1 Please read this user s manual carefully and make sure to know how the instrument works before using it. Keep the user s manual to read up on details when required. 2. Saftey instructions 2 1 Avoid danger due to electricity - Do not touch parts under voltage with the instrument or the sensor! Protection of the measuring instrument - Keep the instrument away from paint, solvent and glue and store it in a dry place. The product s safety for maintenance - The instrument can only be used within the specified data. - Treat the instrument according to its purpose. - Do not use force! - Only authorized staff is allowed to repair the instrument, otherwise the manufacturer is not responsible for functioning. This is the same for the validity of licence. 1

7 3. Battery and device disposal U= Empty or damaged batteries are to deliver to authorized collection points. P= After removal from service, please dispose the instrument eco-friendly. 4. Application area This high-quality instrument is only for the using in the following application area definite: The instrument is for professional settings and for control-measurements at all small-firings-facilities (low temperatureand burner-value-boilers and -thermal) for gas and oil applicable. This instrument is for measurement appropriate DIN EN

8 5. Product Description The BLUELYZER is one of the smallest gas analyzer for continuous measurement of flue gases. The instrument measures the values as follows: T.Air Air or ambient temperature in C T.Gas Flue gas temperature in C O 2 Flue oxygen content in Vol.-% Carbon monoxide in ppm Draft Draft or pressure in hpa The values determined are used to calculate and display the following data: Flue gas losses in % 2 Carbon dioxide in Vol.-% u Carbon monoxide, undiluted in ppm Lambda(λ) Excess air value Additional values that can be printed: T.Diff Differential temperature (T.Gas - T.Air) in C Eta Combustion Efficiency in % 3

9 6. Technical Data Supply Battery: 2 x 1.5 V Alkaline Battery (Type: MIGNON, IEC LR6, AA) 2 x 1,2 V NiMH-Akku (Typ: MIGNON, IEC HR6, VH 1101, AA) or 2 x 1,5 V AccuCell-Akku (Typ: MIGNON, IEC LR6, AA) Power Supply: BLUELYZER Power Supply V~/50-60Hz Display: Dual line 7 Segment LCD Printer: External Infrared Thermoprinter (EuroPrinter or HP82240B) Operating Temperature: + 5 C to + 40 C Storage and Transport Temperature: - 20 C to + 50 C Sensor Warm-Up time (Instrument cold): From 0 C to ambient in less than 10 minutes Dimensions / Weight: L190 x W 60 H 36 mm / approx. 400 g 4

10 7. General Operating Information 1 Note! In the event of program disturbances please remove batteries! This causes the instrument to return to its initial settings (Reset Function). Attention! In case of repetitive malfunction, replace batteries or connect the instrumentspecific power supply (refer to Technical Data ). Protection of measuring cell In case of overload (-Values exceeding 9999 ppm) the display will alternate between the actual value measured and a warning message " HIGH". Simultaneously, the pump will be switched off! The gas probe must be removed from the flue gas tube immediately! After an initially slow reduction, the concentration at the sensor will fall continuously. The purging of the instrument pump will not start until the excessive gas concentration has fallen below a value of less than 9999 ppm (this is supported by the self-purging function of the instrument). If the display value reaches < 20 ppm, the instrument should be switched OFF and ON again. This initiates a new calibration phase.. Preparing the analyzer for a measurement Sensor connections: Connect the tubings to and from the sensor by means of quick-changes connectors. Connect the temperature plug to the inlet sockets. The connections are colour-coded to prevent any confusion: Yellow = Gas-Temperatur Blue = Air temperature 1 Note! Keep the probe out of the flue during the 60 second calibration phase! 5

11 8. Measurement Principle Temperature measurement: Draft Measurement (Option): K-type Thermocouple (NiCr-Ni) for flue gas temperature K-type Thermocouple (NiCr-Ni) for combustion air temperature Piezo-resistive semiconductor sensor O 2-Measurement: Electrochemical sensor -Measurement: Electrochemical sensor Duration Measurement: Flue Gas Measurement: Max. 60 minutes are possible, followed by a new calibration phase with ambient air. The flue gas is fed to the sensors by means a gas pump via an external water separator and protective filter. Sensor Calibration: Flue gas measurement: Calibration (60 sec) with ambient air. Draft measurement: Calibration (10 sec) to ambient pressure. Flue Gas Sampling: Standard flue gas probe with condensate trap, filter and retainer cone. 6

12 9. Physical Specification Measurement Ranges: ppm - 0 % ppm O Vol.-% T-Gas C T-Air C max. Vol.-% % Lambda Draft (Option) 0 ± 40 hpa Flue Gas Temperature Measurement Measurement range: 0 C C Max. Deviation: ±1 C + 1 digit (up to 300 C) ±0.5 % of reading value (above 300 C) Resolution: 1 C Response time (T 90): 50 Sek. Combustion Air Temperature Measurement range: -20 C +200 C Max. Deviation: ±3 C + 1 digit (-20.0 to 0,0 C) ±1 C + 1 digit (+0.1 to C) Resolution: 0.1 C Response time (T 90): 70 Sek. 7

13 O 2 Measurement Range: Vol.- % Max.-Deviation: ±0.2 Vol.- % Resolution: 0.1 Vol.- % Response time (T 90): 50 sec. 2 Measurement (calculated from O 2 measurement) Range: max. Vol.- % Max. Deviation: ±0.2 Vol.-% Resolution: 0.1 Vol.-% Response Time (T 90): 60 sec. Measurement (with H 2-Compensation) Range: ppm Accuracy: ±5 ppm (up to 150 ppm) ±5 % of reading value (up to 9999 ppm) Resolution: 1 ppm Response time (T90): 90 sec. 8

14 Draft or Pressure Measurement (Option) Range: 0 ± 40 hpa Accuracy: ±0.02 hpa (up to ± 2.00 hpa) ±1 % of reading value (up to ± 40 hpa) Resolution: 0.01 hpa Response time (T90): 10 sec. 10. Calculation Formulae O 2 Calculation of the 2 Value 2 = 2max. * ( ) in % 20,9 2max. : max. 2 value (fuel-specific) in %. O 2: Measured Oxygen content in %. 20,9: Oxygen content in air in %. A2 Calcuation of the waste gas loss () = (T.Gas - T.Air ) * ( B) in % 21 - O2 T.Gas : Flue gas temperature in C. T.Air : Combustion air temperature in C. A2, B: Fuel-specific factors. 9

15 2max. 20,9 Calculation of the Excess Air Value (Lambda) Lambda = = ,9 O 2 2 calculated 2-Value Calculation of the combustion-related efficiency (Eta) Eta = in % Calculation of 0 % (undiluted) 0 % = * Lambda in ppm 0 %: Carbon monoxide, undiluted : Measured -Value 10

16 11. System Maintenance & Instrument Care Condensate Trap/Filter: see drawing page 28! 1 Important! Condensate trap or container: Empty completely after each operation. Water in the instrument may have strong or even destructive effects on the pumps and sensors; Damaged or improperly inserted filters will lead to a failure of the filtering function and thus, to a failure of expensive pumps and sensors. Infiltecmicrofilter: Check for contamination and replace if required. NOTE! The microfilter acts as a filter for soot, smoke and dust particles and absorbs water vapor from the condensate contained in the waste gas! In case of strong contamination and/or saturation caused by condensate, the filter must be replaced immediately! After drying and in a clean condition, the filter can be re-used repetitively. Membrane filter (filter disc): Replace in case of decreasing pump flow. Threaded parts must be properly inserted (engaged) and tightened moderately. Proper sealing is ensured by O-rings. U= Batteries must not be disposed of in household waste. Please use proper collection points for disposal. Use specified batteries cells only! Do not mix different types of batteries or new and used cells. Do not charge non- rechargeable batteries. Do not short the terminals!. Keep the instrument dry. 11

17 12. Display Symbols Carbon Dioxide Concentration Oxygen Content Flue Gas Losses O 2 2 T L Combustion Air-Temp % Units of measure C U Carbon Monoxide Concentration Carbon Monoxid (undiluted) Excess Air Value (Lambda) Menu Arrow Mark Flue Gas Temperatur 12

18 13. Key Pad Functions 13

19 14. Program Start O 2 2 T Luft Switch on Instrument U O 2 2 T Luft All segments and symbols are tested for approx. 2 sec. C hpa % Min HOLD Max C F sec ppm U 14

20 15. Calibration Phase O 2 2 T Luft Immediately after powering up the instrument the calibration phase starts automatically (60 sec.). The gas pump is switched on. sec U If an error should occur during the calibration phase the measurement value it refers to will be indicated by a flashing arrow mark. O 2 2 T Luft Defective sensors will be excluded from the measurement. The measurement program indicates a defective sensor by means of a flashing arrow and by horizontal dashes in lieu of the expected measurement value. sec U 15

21 16. Fuel Selection O 2 2 T L The Fuel Selection menu is automatically displayed after calibration phase. The gas pump is switched off during the fuel selection. Select the desired Fuel U Confirm The following fuel types can be selected successively via the display: In Brackets: Abbreviations for fuel types as indicated by the 7Segment display of the BLUELYZER. Oil - H1 (OIL H1) Natural Gas (nat - GAS) Oil - H2 (OIL H2) Butane Gas (but - GAS) Oil L (OIL L) Propane Gas (PrO - GAS) 16

22 1 Note! The first position of the fuel selection menue always displays the last selected type of fuel (Memory Function). For the test gas measurement please select the test gas menu (test GAS). Select test gas menu Confirm the desired flue type After the desired fuel type has been confirmed the gas pump is switched ON automatically. At the same time the instrument switches to the Measurement program automatically. The first pair of measured values is pre-adjusted to O 2 and. 17

23 17. Flue Gas Measurement Program O 2 2 T L The Flue Gas Measurement Program allows the display of two values in any combination. % The corresponding arrows indicate the measurement value as selected. ppm Selection the measurement value Upper display line (O 2, 2,, T L) U O 2 2 T L % U 18

24 O 2 2 T L % ppm Selection the measurement value Lower display line (, u, λ, ) U O 2 2 T L 1 Note! In case of disturbances during the calibration phase (eg. Temperature Sensor not connected, O 2-Sensor defective, etc.) the corresponding values in the measurement program are indicated by dash symbols % ---- C U 19

25 Each value in the upper row can be combined with any value of the lower display line. O 2 2 T L The corresponding physical units are shown in the same line to the right of the numerical value. % HOLD C Hold measured values (Hold-Function) U The current values are held and will no longer be changed. O 2 2 T L hpa Option: Draft (or Pressure) Measurement Program Select Measurement value (upper row) O 2, 2,, T L, Draft ppm U 20

26 Functional Description: Draft Measurement during Flue Gas Measurement: O 2 2 T Luft Within the exhaust measurement program the Draft-Measurement can be carried out both continuously and selectively. hpa Continuous (permanent) Draft Measurement Immediately after the calibration phase the pressure sensor is set to zero. If the appropriate presettings have been made, the continuously measured draft and/or pressure value will be indicated in the upper display line. If you wish to hold (lock) the displayed draft value together with all other measurement values press the HOLD key twice. U O 2 2 ppm T Luft Hold-Function hpa 1 Note! The continuous draft measurement is only a measurement that indicates certain tendencies of the pressure pump in operation! Any disturbances, such as pump underpressure, sensor drift, etc., can have a very strong impact on the measuring results. To perform a specified draft measurement the draft must be measured selectively! U ppm 21

27 Selective Draft Measurement O 2 2 T L To precisely determine the draft value a zero-value calibration should be carried out shortly before the draft measurement. Procedure: Preselect Draft Measurement in the upper display line. hpa HOLD Measurement selection (Upper line) Hold measurement values (Pump switches off!) ppm U HOLD-Function Calibrate Draft-Zero-Setting: Remove air tubing before pushing the ENTER-Button! O 2 2 T L hpa Calibration Zero-Setting The pressure sensor will be calibrated.following the zero calibration reconnect the air tube. Wait until the measurement value stabilizes, then HOLD the measured pressure value. HOLD ppm U HOLD-Function 22

28 If a different measurement value is assigned and the HOLD-function is deactivated during a locked Draft-Value in the upper display line, the last measured and locked Draft value will still be maintained! O 2 2 T L C Select Measurement Value (Upper display line) ppm Release measured values (Deactivate HOLD-Function) U After branching back to the draft display, the old draft value re-appears in the upper line of the display. The additional up-arrow indicates that only the pressure value as shown is held. All other measurement values are measured successively and can be shown one by one. O 2 2 T L hpa Repetitive pushing of the HOLD-Button will release the draft-value for new measurements. HOLD ppm Release Draft Measurement Value U 23

29 18. Print Data Print Printing takes place via the internal infrared port by an IR-Thermoprinter type: HP 82240B 1 Note! Printing of both current and locked values (HOLD-Function) can only be done from the appropriate measuring program! 19. End of Measurement Program O 2 2 T L Switch OFF the instrument The instrument switches off within 6 seconds. If is possible to abort the shutdown cycle by pressing the ENTER-key. sec Interrupt U 24

30 20. Time / Date To set the time and/or date of BLUELYZER, push the HOLD key during the calibration phase. After that you can select first the time and then the date. O 2 2 T L Time The first display line shows the time in hours and minutes. Select hours U Select minutes Confirm time entry 25

31 Date O 2 2 T L The first display line shows day and month, the second line shows the year. Select day Select month U Select year Confirm date entry After setting time/date, the calibration phase will start automatically! 26

32 27 Power supply Flue gas tempertaur plug (yellow) Air temperatur sensor (blue) Flue Gas Measurement Hose Gas Filtraton Chamber (refer to separate sheet) Adjustable cone adapter Condensate Trap 21. Connection-schema

33 22. Gas-Processing & Watertrap unit 10 Replacement and spare parts (Part-no. VK-00192) Part-no. 00 Spare parts bag VK Inlet piece O-ring 23x Glass piston with arrow mark Middle part with cylinder Glass piston with logo Infiltec mirco filter Connection pieces Teflon membran 23.5 mm Outlet pieces O-ring 18x Maintenance / Care - Empty condensate trap after use. - Check micro filter for contamination, replace if required. - Exchange teflon membran filter in case of degrading pump flow. If damaged or inserted improdperly, the filtering function will be lost! - Grease all O-rings with vaseline or silicongrease as required. 28

34 23. Notices 29

35 Appendix O 2 2 T L At appropriateness from the "EuroPrinter" you can choose between Slow Print and Fast Print Setting as following: sec U Press Print-Button during the switch-off countdown O 2 2 T L Shift from the data-output-speed to the slow print (SPrn) sec U 30

36 O 2 2 T L Shift from the data-output-speed to the fast print (FPrn) 1 Note! The printing form the protocol in the mode of operation "FPrn" leads to disturbances if you use the Printer HP 82240B. Redress: Choose mode of operation "SPrn". U sec 31

37 SYSTRONIK Elektronik u. Systemtechnik GmbH Gewerbestraße 57 D Illmensee Tel: +49 (0) 7558 / Fax: +49 (0) 7558 / info@systronik.com