Yes, there is going to be some math (but not much) STATISTICAL APPROACH TO MEDICAL DEVICE VERIFICATION AND VALIDATION
|
|
- Lambert Shepherd
- 5 years ago
- Views:
Transcription
1 Yes, there is going to be some math (but not much) STATISTICAL APPROACH TO MEDICAL DEVICE VERIFICATION AND VALIDATION
2 Medical Device Verification and Validation In the medical device world, verification and validation are the backbone of design control This is how you prove you have produced what you said you would and what the customer wants! It s also required
3 What The FDA Says About Design Verification/Validation 21CFR820: QUALITY SYSTEM REGULATION Sec Definitions Verification means confirmation by examination and provision of objective evidence that specified requirements have been fulfilled Design verification. Each manufacturer shall establish and maintain procedures for verifying the device design. Design verification shall confirm that the design output meets the design input requirements. The results of the design verification, including identification of the design, method(s), the date, and the individual(s) performing the verification, shall be documented in the DHF
4 What The FDA Says About Design Verification/Validation 21CFR820: QUALITY SYSTEM REGULATION Sec Definitions Validation means confirmation by examination and provision of objective evidence that the particular requirements for a specific intended use can be consistently fulfilled Design validation means establishing by objective evidence that device specifications conform with user needs and intended use(s)
5 What The FDA Says About Design Verification/Validation 21CFR820: QUALITY SYSTEM REGULATION Sec Definitions Design validation. Each manufacturer shall establish and maintain procedures for validating the device design. Design validation shall be performed under defined operating conditions on initial production units, lots, or batches, or their equivalents. Design validation shall ensure that devices conform to defined user needs and intended uses and shall include testing of production units under actual or simulated use conditions. Design validation shall include software validation and risk analysis, where appropriate. The results of the design validation, including identification of the design, method(s), the date, and the individual(s) performing the validation, shall be documented in the DHF This is NOT the same as Test Method or Process Validation
6 Where Does V&V Fit Into the Design Control Picture?
7 A Few Quick Points Design V&V studies are NOT experiments! Y0u should know the outcome before you start. Require predefined acceptance criteria Which is why they are not experiments Before you start V&V, you should have enough characterization data to understand your system (product)
8 Ok Enough Introduction But before we get into the nerdy stuff a few hints to make it easier: Ensure you have clearly defined and measurable product requirements Use standards when available (more in a minute) Hire a good statistician!!
9 Standards and Guidelines There are several good guidelines and standards published to help CLSI publishes a variety of guidelines for statistical analysis useable for verification studies Normally protocols based on CLSI guidance is accepted by the FDA Common Technical Specifications (CTS) for in vitro diagnostics medical devices These CTS requirements are associated with the IVD Directive 2009/108/EC of the European Parliament and the Council of 3rd February 2009 on in vitro diagnostic medical devices
10 Standards and Guidelines A few words on CLSI EP Protocols Provide estimates for individual parameters (limit of detection, precision, bias ) CLSI does not have a standard to combine parameters CLSI standards in many cases proscribe sample sizes and outliers routines
11 Two Requirements for CLSI EP Protocols Doable Don t need a Nobel prize to figure out how to do it Right As in right enough. If CLSI method deviates from a more complicated gold standard method after the 7th decimal place, that s ok An EP rule The least burdensome of several alternatives will always be used if the minimum sample size is 40, 40 will be the only sample size used
12 Speaking of Sample Size Three ways to get sample size From a standard This is the easiest Math Good enough Most common questions I get (how many???)
13 Number of Replicates Math for Replicate Number: nn zz 11 αα/22 + zz 11 ββ/22 SSSS/dddddddddd 22 SD is the standard deviation estimate based on characterization delta is the allowable error limit n is the number of samples α is the significance level, usually β is the power of the test z 1-α/2 and z 1-β/2 are the z values corresponding to the probability of 1-α/2 and 1-β/2 respectively on a normal distribution table (z table)
14 Number of Replicates Good Enough for Replicate Number: You will find that unless you have unusual circumstances, the calculation on the last slide will normally give you a small number of replicates Three (3) is the minimum number you need to see a possible outlier and get a good mean Power calculations will tend to give you an answer which is more than you have time or money for Unless y0ur system has a lot of rep to rep variability, a smaller number of reps will work Variability can be determined through good characterization and used as justification for number of reps.
15 Speaking of Sample Size The same type of logic applies for how many samples are needed: From a standard Again, this is the easiest Math Good enough
16 Speaking of Sample Size Math: Some definitions: Margin of Error (Confidence Interval) No sample will be perfect, so you need to decide how much error to allow. The confidence interval determines how much higher or lower than the population mean you are willing to let your sample mean fall Confidence Level How confident do you want to be that the actual mean falls within your confidence interval? The most common confidence intervals are 90% confident, 95% confident, and 99% confident
17 Speaking of Sample Size Math: Confidence Level Your confidence level corresponds to a Z-score. This is a constant value needed for this equation. Here are the z- scores for the most common confidence levels 90% Z Score = % Z Score = % Z Score = nn = zz22 SSSS (11 SSSS) MMMMMMMMMMMM oooo EEEEEEEEEE 22 You will find this will give you pretty large sample numbers You also have to know the SD
18 Speaking of Sample Size Good Enough Large vs. Small Sample Numbers Larger sample numbers are more likely to approach a closer model to the total population than smaller samples numbers Larger sample numbers are more likely to approach a normal distribution than small sample numbers A rule of thumb is to have 16 degrees of freedom (D.F.) for any average standard deviation used in building a model "Understanding Industrial Designed Experiments" (Schmidt & Launsby) This would be 17 samples
19 Speaking of Sample Size "The closeness of the approximate probability 1- alpha to the exact probability depends on both the underlying distribution and the sample size...as the underlying distribution becomes 'less normal' (i.e., badly skewed or discrete), a larger sample size might be required to keep a reasonably accurate approximation. But, in all cases, an n of at least 30 is usually quite adequate. Probability And Statistical Inference 2nd Edition" (Hogg & Tanis)
20 Speaking of Sample Size As a rule of thumb, you can consider n=30 observations or more to constitute a large' sample So If n=30 is a large sample, and n-17 is a reasonable number to get enough degrees of freedom Therefore, by using n=20 there is a compromise in dropping the theoretical value (n=30) to a value that is more practical to use in industrial settings
21 Speaking of Sample Size A few final notes on sample size If SD is known, then use z distribution. If SD is not known: If n is large, then use z distribution. If n is small, then use t-distribution
22 Design Verification Examples There are so many different types of product requirements and potential ways to verify You really don t want to be here all night, do you? Here some suggestions of product requirement verifications and what elements you should have in your protocols and reports
23 Precision and Accuracy These are sometimes confused You need to be sure you are measuring the correct parameter And that parameter meets you needs and requirements
24 Evaluation of Imprecision and Inaccuracy Imprecision Refers to random analytic error Lack of repeatability and/or reproducibility Inaccuracy Refers to systematic analytic error Lack of trueness Constant Proportional Total Error Combined error for a single result
25 Precision and Accuracy
26 How Much Error is Ok?? Performance is acceptable when Observed error < allowable error Performance is not acceptable when Observed error > allowable error How much error is acceptable? Based on medical and or clinical need User needs traced to product requirements which are what you verify/validate
27 Protocols Should contain (at a minimum) Purpose including design input, design output being verified Rationale for the verification design Measurable acceptance criteria. Materials required Verification Method/Procedure. Include procedure or reference for evaluation of unexpected test results (e.g. no-tests, invalids) Boundary values, must be assessed and included Data Analysis including validity criteria, if applicable Statistical methods with references to published methods, if applicable Include method for handling exceptions and outliers.
28 Reports Should contain (at a minimum) Materials used (including part number, lot number, expiration date, storage condition, as applicable) Equipment used and calibration status Revisions of any reference procedures Date and location of testing Individual(s) performing the verification Results including or referencing raw data Summary of procedural errors and investigation of failures or unexpected results, if any Summary of excluded data, if any Disposition, i.e., whether or not the acceptance criteria was met. If acceptance criteria are not met, a discrepancy report should be initiated
29 Design Validation I did not spend a lot of time (or any) on design validation In many cases, the number of sites, samples and statistics required are dictated by Regulatory agencies Cost of external studies Availability of clinical samples Time (mainly cost and management driven) You still have to justify what you do and have the data analyzed statistically
30 In Summary You always need to have some sort of statistical justification for your studies You need to be able to defend your justification There are many guidelines and standards available to reference HIRE A GOOD STATISTICIAN
31 Almost Last Slide
32 That s All Folks! My Contact Information: Alan Golden Manager QA, Operations Support Abbott Molecular Alan.golden@abbott.com
Secondary Math Margin of Error
Secondary Math 3 1-4 Margin of Error What you will learn: How to use data from a sample survey to estimate a population mean or proportion. How to develop a margin of error through the use of simulation
More informationTest Method Validation of In-Vitro Diagnostic Products. Tim Carr QA Manager, Beckman Coulter
Test Method Validation of In-Vitro Diagnostic Products Tim Carr QA Manager, Beckman Coulter IVD Overview What is an In-Vitro Diagnostic Test IVDs are medical devices and accessories Used to perform tests
More informationMeasurement Uncertainty Guide. ISO Accreditation Program
Measurement Uncertainty Guide ISO 15189 Accreditation Program Background Why This is Necessary The ISO 15189:2012 standard contains enhanced expectations regarding measurement uncertainty (MU) in clause
More informationVerification of Method XXXXXX
Reviewed By: Date: Technical Supervisor Reviewed By: Date: General Supervisor Reviewed By: Date: Quality Assurance Coordinator Approved By: Date: Director Page 1 of 11 Table of Contents Purpose and Principle:...
More informationMOLECULAR TESTING: VERIFYING/VALIDATING INSTRUMENTS, REAGENTS AND ASSAYS. Richard L. Hodinka, Ph.D.
MOLECULAR TESTING: VERIFYING/VALIDATING INSTRUMENTS, REAGENTS AND ASSAYS Richard L. Hodinka, Ph.D. University of South Carolina School of Medicine Greenville Greenville Health System, Greenville, SC hodinka@greenvillemed.sc.edu
More informationV&V Best Practices. CASSS, CMC Strategy Forum Steven W. Badelt, PhD Managing Partner Suttons Creek, Inc. SUTTONSCREEK.COM
V&V Best Practices CASSS, CMC Strategy Forum Steven W. Badelt, PhD Managing Partner Suttons Creek, Inc. Solving the problem of Complexity. 21CFR820.30 & FDA Guidance INCOSE System Engineering Handbook
More informationThe Complete Guide to FDA Design Controls
The Complete Guide to FDA Design Controls Jon D. Speer Founder & VP QA/RA of greenlight.guru ABOUT THE PRESENTER Jon D. Speer is the founder and VP of QA/RA of greenlight.guru 20+ years in medical device
More informationChapter 9 Assignment (due Wednesday, August 9)
Math 146, Summer 2017 Instructor Linda C. Stephenson (due Wednesday, August 9) The purpose of the assignment is to find confidence intervals to predict the proportion of a population. The population in
More informationTotal Analytic Error From Concept to Application
Página 1 de 5 Clinical Laboratory News Subscribe CLN Daily CLN Stat CLN Articles Total Analytic Error From Concept to Application Author: James O. Westgard, PhD, and Sten A. Westgard, MS // Date: SEP.1.2013
More informationSelecting a Sample Size
Selecting a Sample Size A Model for Calculating Sample Size to Allow Development Decisions An Adept Packaging White Paper Selecting a Sample Size A Model for Calculating Sample Size to Allow Development
More informationChapter 4: Foundations for inference. OpenIntro Statistics, 2nd Edition
Chapter 4: Foundations for inference OpenIntro Statistics, 2nd Edition Variability in estimates 1 Variability in estimates Application exercise Sampling distributions - via CLT 2 Confidence intervals 3
More informationGuide to Fulfillment of Validation and Verification of Examination Requirements
DIAGNOSTIC ACCREDITATION PROGRAM College of Physicians and Surgeons of British Columbia 300 669 Howe Street Telephone: 604-733-7758 ext. 2635 Vancouver BC V6C 0B4 Toll Free: 1-800-461-3008 (in BC) www.cpsbc.ca
More informationEstablishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation
Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation For informational purposes only. WHO REQUIRES IT AND WHAT THEY REQUIRE Regulatory agencies require the
More information9. Verification, Validation, Testing
9. Verification, Validation, Testing (a) Basic Notions (b) Dynamic testing. (c) Static analysis. (d) Modelling. (e) Environmental Simulation. (f) Test Strategies. (g) Tool support. (h) Independent Verification
More informationTHE NORMAL CURVE AND SAMPLES:
-69- &KDSWHU THE NORMAL CURVE AND SAMPLES: SAMPLING DISTRIBUTIONS A picture of an ideal normal distribution is shown below. The horizontal axis is calibrated in z-scores in terms of standard deviation
More informationBOOTSTRAPPING AND CONFIDENCE INTERVALS
BOOTSTRAPPING AND CONFIDENCE INTERVALS Fill out your reading report on Learning Suite MPA 630: Data Science for Public Management November 8, 2018 PLAN FOR TODAY Why are we even doing this? Confidence
More informationTechnical Guidance on Development of In Vitro Companion Diagnostics and Corresponding Therapeutic Products
Administrative Notice December 26, 2013 To: Division of Pharmaceutical Affairs, Prefectural Health Department (Bureau) From: Evaluation and Licensing Division, Pharmaceutical and Food Safety Bureau Ministry
More informationMETHOD VALIDATION: AN ESSENTIAL COMPONENT OF THE MEASUREMENT PROCESS. Presented by : Jerry Parr, The NELAC Institute
METHOD VALIDATION: AN ESSENTIAL COMPONENT OF THE MEASUREMENT PROCESS Presented by : Jerry Parr, The NELAC Institute Source Material The NELAC Institute, Volume 1, Module 2, Quality Systems General Requirements,
More informationCrowe Critical Appraisal Tool (CCAT) User Guide
Crowe Critical Appraisal Tool (CCAT) User Guide Version 1.4 (19 November 2013) Use with the CCAT Form version 1.4 only Michael Crowe, PhD michael.crowe@my.jcu.edu.au This work is licensed under the Creative
More informationValidating, Verifying, and Evaluating Your Test Methods: It s NOT a Regulatory Exercise!
Validating, Verifying, and Evaluating Your Test Methods: It s NOT a Regulatory Exercise! Pat Garrett, Ph.D., DABCC Renee Howell, Ph.D., MT(ASCP) SeraCare Life Sciences, Inc. AACC Annual Meeting July 29,
More informationLECTURE 10: CONFIDENCE INTERVALS II
David Youngberg BSAD 210 Montgomery College LECTURE 10: CONFIDENCE INTERVALS II I. Calculating the Margin of Error with Unknown σ a. We often don t know σ. This requires us to rely on the sample s standard
More informationAvailable online at ScienceDirect. Procedia Engineering 132 (2015 )
Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 132 (2015 ) 811 815 The Manufacturing Engineering Society International Conference, MESIC 2015 Metrological Regulations for
More informationEstoril Education Day
Estoril Education Day -Experimental design in Proteomics October 23rd, 2010 Peter James Note Taking All the Powerpoint slides from the Talks are available for download from: http://www.immun.lth.se/education/
More informationMolecular Diagnosis Challenges & Solutions. Using Molecular Kits or Laboratory Developed Tests (Home Brew), Emphasis on Validation
Using Molecular Kits or Laboratory Developed Tests (Home Brew), Emphasis on Validation Molecular Diagnosis Challenges & Solutions Behzad Poopak, DCLS PhD Tehran Medical Branch- Islamic Azad University
More informationSupplemental Verification Methodology
Supplemental Verification Methodology To: ALL FOREST PROJECT VERIFIERS Date: July 03, 2012 Re: Updated Guidance for Verification of sampled pools for Forest Projects Version 3.3 of the Forest Project Protocol
More informationChapter 12. Sample Surveys. Copyright 2010 Pearson Education, Inc.
Chapter 12 Sample Surveys Copyright 2010 Pearson Education, Inc. Background We have learned ways to display, describe, and summarize data, but have been limited to examining the particular batch of data
More informationi) Wild animals: Those animals that do not live under human supervision or control and do not have their phenotype selected by humans.
The OIE Validation Recommendations provide detailed information and examples in support of the OIE Validation Standard that is published as Chapter 1.1.6 of the Terrestrial Manual, or Chapter 1.1.2 of
More informationQ: Using at least 3 biological replicates in an experiment is recommended to do. What do you suggest: At which step of calculation of the relative
The questions below have been asked by attendees of the qpcr webinar series, Part 2: Analyzing Your Data. All the questions, including the questions that could not be answered during the webinar have been
More informationLifecycle of NIR Analytical Methods: Regulatory Perspective
Lifecycle of NIR Analytical Methods: Regulatory Perspective Bogdan Kurtyka, Ph.D. FDA/OPQ/OPF IFPAC Annual Meeting January 27, 2015 Outline Method lifecycle Impact of selected elements of lifecycle on
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 18113-1 First edition 2009-12-15 In vitro diagnostic medical devices Information supplied by the manufacturer (labelling) Part 1: Terms, definitions and general requirements
More informationSawtooth Software. Sample Size Issues for Conjoint Analysis Studies RESEARCH PAPER SERIES. Bryan Orme, Sawtooth Software, Inc.
Sawtooth Software RESEARCH PAPER SERIES Sample Size Issues for Conjoint Analysis Studies Bryan Orme, Sawtooth Software, Inc. 1998 Copyright 1998-2001, Sawtooth Software, Inc. 530 W. Fir St. Sequim, WA
More informationThe road to a perfect EQA-programme
The road to a perfect EQA-programme The Dutch SKML experience Christa Cobbaert DEPARTMENT OF CLINICAL CHEMISTRY AND LABORATORY MEDICINE, LEIDEN UNIVERSITY MEDICAL CENTRE, LEIDEN, THE NETHERLANDS Content
More informationImproving the validity and quality of our research Daniël Lakens Eindhoven University of
Improving the validity and quality of our research Daniël Lakens Eindhoven University of Technology @Lakens / Human-Technology Interaction 1-2-2016 PAGE 1 Sample Size Planning / Human-Technology Interaction
More informationLifecycle Management of Process Analytical Technology Procedures
Lifecycle Management of Process Analytical Technology Procedures IFPAC 2015 Marta Lichtig Senior Scientist in New Testing Technologies, ACS Member Contents General Comparison : PV guide to NIR model development
More informationSection 7-3 Estimating a Population Mean: σ Known
Section 7-3 Estimating a Population Mean: σ Known Created by Erin Hodgess, Houston, Texas Revised to accompany 10 th Edition, Tom Wegleitner, Centreville, VA Slide 1 Key Concept Use sample data to find
More informationAir Monitoring Directive Chapter 5: Quality System
Air Monitoring Directive Chapter 5: Quality System Version Dec 16, 2016 Amends the original Air Monitoring Directive published June, 1989 Title: Air Monitoring Directive Chapter 5: Quality System Number:
More informationAAMI Quality Systems White Paper
AAMI s White Paper Comparison of 21 CFR Part 820 to ISO 13485:2016 February 2017, Updated February 2018 AUTHORS Seb Clerkin, GMP Advisory Services Nicola Martin, Owner, Nicola Martin Consulting Jack Ward,
More informationSystems-Based Inspections for Cleaning Validation
Systems-Based Inspections for Cleaning Validation FDA DG 230 July 21, 2014 Rockville, MD Destin A. LeBlanc Cleaning Validation Technologies www.cleaningvalidation.com 1 Cleaning Definition: The process
More informationEFFICACY OF ROBUST REGRESSION APPLIED TO FRACTIONAL FACTORIAL TREATMENT STRUCTURES MICHAEL MCCANTS
EFFICACY OF ROBUST REGRESSION APPLIED TO FRACTIONAL FACTORIAL TREATMENT STRUCTURES by MICHAEL MCCANTS B.A., WINONA STATE UNIVERSITY, 2007 B.S., WINONA STATE UNIVERSITY, 2008 A THESIS submitted in partial
More informationUnderstanding the Characteristics and Establishing Acceptance Criteria for Analytical Methods Validation
Understanding the Characteristics and Establishing Acceptance Criteria for Analytical Methods Validation Ying Verdi IVT LAB WEEK EUROPE June 2017 Partners in Health Since 1919 Regulatory View of Method
More informationVerification or validation, that is the question
Perspective Page 1 of 7 Verification or validation, that is the question Giorgia Antonelli 1, Andrea Padoan 1, Ada Aita 1, Laura Sciacovelli 2, Mario Plebani 1,2 1 U.O.C. of Laboratory Medicine, Department
More information10 Things You Should Do Before You Validate Your Next Package
sponsored by 10 Things You Should Do Before You Validate Your Next Package Before you set out to validate a medical device package design or process, you ve got to do your homework. You need to understand
More informationValidation. An Easy to Understand Guide Validating Purified Water Systems
Validation Overview Validation Documentation Validation Activities Validation Mistakes Validation Protocols Test Selection Common Verifications When to test Test Layout and Content Validation Protocol
More informationGUIDE TO GOOD MANUFACTURING PRACTICE FOR MEDICINAL PRODUCTS
PHARMACEUTICAL INSPECTION CONVENTION PHARMACEUTICAL INSPECTION CO-OPERATION SCHEME GUIDE TO GOOD MANUFACTURING PRACTICE FOR MEDICINAL PRODUCTS PH 1/97 (Rev. 3), 15 January 2002 ANNEX 15 QUALIFICATION AND
More informationLifecycle Management Concepts to analytical Procedures: A compendial perspective. Horacio Pappa, Ph.D. Director - General Chapters U.S.
Lifecycle Management Concepts to analytical Procedures: A compendial perspective Horacio Pappa, Ph.D. Director - General Chapters U.S. Pharmacopeia USP Definitions Validation of Compendial Procedures
More informationGuidance for Industry and FDA Staff. Replacement Reagent and Instrument Family Policy
Guidance for Industry and FDA Staff Replacement Reagent and Instrument Family Policy Document issued on: December 11, 2003 This document supersedes Data for Commercialization of Original Equipment Manufacturers
More informationMarch 19, Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane Room 1061 Rockville, MD 20852
March 19, 2018 Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane Room 1061 Rockville, MD 20852 RE: Docket No. FDA-2017-D-6765; Draft Guidance for Industry and Food
More informationWhile the recognition
Designing the Perfect Change Control System Change control systems today are expected to be designed in a way that provides a system to not only document and approve changes, but also to anticipate change
More informationIndustry Perspective on Manufacturing in Early Development
Industry Perspective on Manufacturing in Early Development IQ Workshop, Feb 4-5, 2014, Washington, D.C. Eric Schmitt AbbVie IQ Drug Product Manufacturing Working Group August 2012 issue of Pharmaceutical
More informationINVESTIGATE OOT AND OOS IN STABILITY STUDIES
INVESTIGATE OOT AND OOS IN STABILITY STUDIES Kim Huynh-Ba Executive Director Pharmalytik LLC www.pharmalytik.com BIOGRAPHY Kim Huynh-Ba has over 25 years of experience in analytical development, project
More informationPROCESS VALIDATION. A Systematic Approach 2015 WHITE PAPER WHITE PAPER PRODUCED BY MAETRICS
WHITE PAPER PROCESS VALIDATION A Systematic Approach 2015 WHITE PAPER PRODUCED BY MAETRICS For more information, please contact: USA Office: +1 317 706 1493 UK Office: +44 115 921 6200 globalsales@maetrics.com
More informationBasic Good Laboratory Practice. Christopher Jerome, B.Vet.Med., Ph.D
Basic Good Laboratory Practice Christopher Jerome, B.Vet.Med., Ph.D Goals Outline the concept of Good Laboratory Practice (GLP) Provide some specific guidelines applicable to day to day work in the laboratory
More informationSAMPLE. Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures
2nd Edition EP21 Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures This guideline provides manufacturers and end users with an understanding of concepts related
More informationOOS: Back to Basics. Compliant, Effective, Efficient PATH
OOS: Back to Basics Even with the number of trainings, seminars, online webinars and consultant guided investigations, companies are still seeing FDA 483 observations around how they are handling and investigating
More informationDesign Quality. Indu Lakshman
Design Quality Indu Lakshman Overview New product development (NPD) covers the complete process of bringing a new product to market. In commercial terms, new product development is described in the literature
More informationTransfer of Methods Supporting Biologics and Vaccines
Transfer of Methods Supporting Biologics and Vaccines Timothy Schofield Arlenda Inc. tim.schofield@arlenda.com Presented at the BPD event on Case Studies in Tech Transfer esearch Triangle Park, NC, December
More informationKEY CONSIDERATIONS AND CHALLENGES OF EDC IN THE IMPLEMENTATION AND STATISTICS OF CLINICAL TRIALS
KEY CONSIDERATIONS AND CHALLENGES OF EDC IN THE IMPLEMENTATION AND STATISTICS OF CLINICAL TRIALS Liora Bosch and EDC expert, Omrix Biopharmaceuticals, Johnson & Johnson EDC 1 A Brief Survey: Which of the
More informationApplication of Statistical Sampling to Audit and Control
INTERNATIONAL JOURNAL OF BUSINESS FROM BHARATIYA VIDYA BHAVAN'S M. P. BIRLA INSTITUTE OF MANAGEMENT, BENGALURU Vol.12, #1 (2018) pp 14-19 ISSN 0974-0082 Application of Statistical Sampling to Audit and
More informationSpeaker: Neville Taylor
Paper: The importance of Stability Testing in Proficiency Testing Author: Vijay Padayachee Vijay Consulting cc PO Box 2707, Johannesburg, 2000, South Africa E-mail: vpad@telkomsa.net Phone: +27 825795242
More informationSurvey Question Analysis (Draft )
The purpose of this tutorial is to analyze two types of questions commonly found on surveys: proportion (yes/no) questions and Likert scale (preferences) questions. (This tutorial doesn t tell you how
More informationChapter 8 Analytical Procedures
Slide 8.1 Principles of Auditing: An Introduction to International Standards on Auditing Chapter 8 Analytical Procedures Rick Hayes, Hans Gortemaker and Philip Wallage Slide 8.2 Analytical procedures Analytical
More information21CFR Ventilation, air filtration, air heating and cooling. 21CFR211 Details, Details. 21CFR Equipment Cleaning and Maintenance
21CFR211 Details, Details Kirsten L. Vadheim, Ph.D., RAC 7710 196th Avenue N.E. Redmond WA 98053 Tel: 651.260.6560 Fax: 425.868.4302 klvadheim@hotmail.com 21CFR211.46 Ventilation, air filtration, air heating
More informationCopyright K. Gwet in Statistics with Excel Problems & Detailed Solutions. Kilem L. Gwet, Ph.D.
Confidence Copyright 2011 - K. Gwet (info@advancedanalyticsllc.com) Intervals in Statistics with Excel 2010 75 Problems & Detailed Solutions An Ideal Statistics Supplement for Students & Instructors Kilem
More informationSolving Statistical Mysteries What Does the FDA Want?
3 7 25Temperature(degC) 40 3.4 6.6 2.6 3.6 75 Humidity % 1.6 5.3 50 No Base Present Yes Solving Statistical Mysteries What Does the FDA Want? Ronald D. Snee, PhD IVT Statistics in Validation Conference
More informationDetailing key considerations and challenges of EDC to enable users to reap the cost, time and quality benefits
Detailing key considerations and challenges of EDC to enable users to reap the cost, time and quality benefits Liora Bosch and EDC expert, Omrix Biopharmaceuticals, Johnson & Johnson EDC 1 A Brief Survey:
More informationINTERNATIONAL RESEARCH JOURNAL OF PHARMACY ISSN Review Article
INTERNATIONAL RESEARCH JOURNAL OF PHARMACY www.irjponline.com ISSN 2230 8407 Review Article INTRODUCTION AND GENERAL OVERVIEW OF PHARMACEUTICAL PROCESS VALIDATION: A REVIEW Pandita Rachna* 1, Rana A C
More informationAP Stats ~ Lesson 8A: Confidence Intervals OBJECTIVES:
AP Stats ~ Lesson 8A: Confidence Intervals OBJECTIVES: DETERMINE the point estimate and margin of error from a confidence interval. INTERPRET a confidence interval in context. INTERPRET a confidence level
More informationCS 147: Computer Systems Performance Analysis
CS 147: Computer Systems Performance Analysis Approaching Performance Projects CS 147: Computer Systems Performance Analysis Approaching Performance Projects 1 / 35 Overview Overview Overview Planning
More informationValidation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals.
Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals. 1 Analytical Method Validation: 1..1 Philosophy: Method validation is
More informationLCA in decision making
LCA in decision making 1 (13) LCA in decision making An idea document CHAINET LCA in decision making 2 (13) Content 1 INTRODUCTION 2 EXAMPLE OF AN INDUSTRIAL DEVELOPMENT PROCESS 2.1 General about the industrial
More informationRecent Trends in the Evaluation of Analytical Biosimilarity
Recent Trends in the Evaluation of Analytical Biosimilarity WCBP 2016, Washington D.C. Thomas Stangler, Senior Scientist, Process Development Strategy Sandoz Biopharmaceuticals 2016 Sandoz. All rights
More informationChapter 19. Confidence Intervals for Proportions. Copyright 2012, 2008, 2005 Pearson Education, Inc.
Chapter 19 Confidence Intervals for Proportions Copyright 2012, 2008, 2005 Pearson Education, Inc. Standard Error Both of the sampling distributions we ve looked at are Normal. For proportions For means
More informationProcess development and basic GMP
Process development and basic GMP Aulton Chapter 45, handouts Specification, stability, inprocess controls and validation Product development Process development Critical Product Qualities issues Critical
More informationUnderstanding and Managing Uncertainty in Schedules
Understanding and Managing Uncertainty in Schedules Realistic Plans for Project Success Presented by: John Owen MPUG Project Integration Month July 20 th, 2016 John Owen 2 1981-1985 Worley Engineering
More informationGuide to Fulfillment of Laboratory Results Comparability Requirements
DIAGNOSTIC ACCREDITATION PROGRAM College of Physicians and Surgeons of British Columbia 300 669 Howe Street Telephone: 604-733-7758 ext. 2635 Vancouver BC V6C 0B4 Toll Free: 1-800-461-3008 (in BC) www.cpsbc.ca
More informationGuidance for Industry
Guidance for Industry Q8, Q9, and Q10 Questions and Answers U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research (CDER) Center for Biologics
More informationDetermining the Effectiveness of Specialized Bank Tellers
Proceedings of the 2009 Industrial Engineering Research Conference I. Dhillon, D. Hamilton, and B. Rumao, eds. Determining the Effectiveness of Specialized Bank Tellers Inder S. Dhillon, David C. Hamilton,
More informationNotes from Round Table 5. January 26 and 27, 2016
Notes from Round Table 5 January 26 and 27, 2016 Topic: Facilitator: Scribe: Potency Assays: Cell Based vs. Non Cell Based Formats Sally Seaver, Seaver Associates LLC Cheryl Blaise, Bristol-Myers Squibb
More information48 Market Rate Analysis
783 48 Market Rate Analysis Key concepts and terms Arithmetic mean or average Capsule job description Inter-quartile range Lower quartile Market rate survey Median Benchmark jobs Derived market rate Job
More informationMETHOD VALIDATION TECHNIQUES PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER 2012
METHOD VALIDATION PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE TECHNIQUES OCTOBER/NOVEMBER 2012 Prepared by for ENAO Assessor Calibration B SCOPE Introduction House Rules Central Tendency Statistics Population
More informationISO 13485:2016 Medical Devices-Quality Management Systems- Requirements for Regulatory Purposes. Theresa McCarthy
ISO 13485:2016 Medical Devices-Quality Management Systems- Requirements for Regulatory Purposes Theresa McCarthy Welcome and Introductions Please interview and collect: Name Primary Quality Management
More informationGLP/SC/01 Basic statistical tools for analytical chemistry (2 days)
GLP Consulting http://consultglp.com Courses on offer (See outlines below) GLP/SC/01 Basic statistical tools for analytical chemistry (2 days) GLP/SC/02 The GUM bottom-up evaluation technique of measurement
More informationCOMPARING THE STANDARDS
COMPARING THE STANDARDS Codex/NACMCF Preventive Rule BRC Comparison 2.0 FUNDAMENTAL Preparatory stages 1 - Assemble the team The team: ideally should be multidisciplinary can use outside expert help should
More informationVQA Proficiency Testing Scoring Document for Quantitative HIV-1 RNA
VQA Proficiency Testing Scoring Document for Quantitative HIV-1 RNA The VQA Program utilizes a real-time testing program in which each participating laboratory tests a panel of five coded samples six times
More informationSuppose we wanted to use a sample of 25 students to estimate the average GPA of all students. Now suppose we choose our sample by random sampling and
Margin of Error When a sample is used to draw inferences about a larger population, there is always a possibility that the sample is non-representative, i.e. does not reflect the true attributes of the
More informationACC: Review of the Weekly Compensation Duration Model
ACC: Review of the Weekly Compensation Duration Model Prepared By: Knoware Creation Date July 2013 Version: 1.0 Author: M. Kelly Mara Knoware +64 21 979 799 kelly.mara@knoware.co.nz Document Control Change
More informationExercise Confidence Intervals
Exercise Confidence Intervals (Fall 2015) Sources (adapted with permission)- T. P. Cronan, Jeff Mullins, Ron Freeze, and David E. Douglas Course and Classroom Notes Enterprise Systems, Sam M. Walton College
More informationWINDOWS, MINITAB, AND INFERENCE
DEPARTMENT OF POLITICAL SCIENCE AND INTERNATIONAL RELATIONS Posc/Uapp 816 WINDOWS, MINITAB, AND INFERENCE I. AGENDA: A. An example with a simple (but) real data set to illustrate 1. Windows 2. The importance
More informationThe Dummy s Guide to Data Analysis Using SPSS
The Dummy s Guide to Data Analysis Using SPSS Univariate Statistics Scripps College Amy Gamble April, 2001 Amy Gamble 4/30/01 All Rights Rerserved Table of Contents PAGE Creating a Data File...3 1. Creating
More informationRevision of 30 April 2013 draft, 4 November 2013
GUIDANCE DOCUMENT FOR SINGLE LABORATORY VALIDATION OF QUANTITATIVE ANALYTICAL METHODS USED IN SUPPORT OF PRE- AND POST-REGISTRATION DATA REQUIREMENTS FOR PLANT PROTECTION AND BIOCIDAL PRODUCTS INTRODUCTION
More informationClinical qualification of specifications - a Regulator s view
Clinical qualification of specifications - a Regulator s view Mats Welin Medical Products Agency, Uppsala, Sweden Disclaimer: The opinions expressed are my own and do not necessarily represent those of
More informationCEC SDG Glossary of Terms Notes
CEC SDG Glossary of Terms Notes - Terms are listed alphabetically - Acronyms are either identified by underlined words (e.g. Co-ordinating European Council), or refer to a separate full-text entry - Each
More informationOMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT
OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (13) 113 2R Evaluation and Reporting of Results Core document Full document title and reference Evaluation and Reporting of Results
More informationEstablishment of an Accredited Reference Measurement Laboratory
Establishment of an Accredited Reference Measurement Laboratory Francesco Dati, PhD IVD-Consulting Marburg / Germany E-mail: f.dati@t-online.de Quality of Analytical Systems Reference Measurement Systems
More informationWHITEPAPER: PROCESS VALIDATION A SYSTEMATIC APPROACH
WHITEPAPER: PROCESS VALIDATION A SYSTEMATIC APPROACH White paper produced by Maetrics For more information, please contact global sales +1 610 458 9312 +1 877 623 8742 globalsales@maetrics.com With offices
More informationKeyMath Revised: A Diagnostic Inventory of Essential Mathematics (Connolly, 1998) is
KeyMath Revised Normative Update KeyMath Revised: A Diagnostic Inventory of Essential Mathematics (Connolly, 1998) is an individually administered, norm-referenced test. The 1998 edition is a normative
More informationAP Statistics Scope & Sequence
AP Statistics Scope & Sequence Grading Period Unit Title Learning Targets Throughout the School Year First Grading Period *Apply mathematics to problems in everyday life *Use a problem-solving model that
More information