PHOSPHORUS, REACTIVE (0.00 to 5.00 mg/l PO 4

Similar documents
Phosphorus, Reactive (Orthophosphate)

Phosphorus, Reactive (Orthophosphate)

1,5-Diphenylcarbohydrazide Method* (Powder Pillows or AccuVac Ampuls) USEPA accepted for wastewater analyses** Using Powder Pillows

TPTZ Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows

3. Add 0.4 ml of. 7. Use a TenSette

3. Add 0.4 ml of. 7. Use a TenSette

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

FerroVer Method (Powder Pillows or AccuVac Ampuls) USEPA approved for reporting wastewater analysis (digestion is required; see Section 2 * )

PaSEC Colorimeter Instructions DR850

Sulfate. USEPA 1 SulfaVer 4 Method 2 Method to 70 mg/l SO 4. Powder Pillows. Test preparation. Before starting.

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

1-(2-Pyridylazo)-2-Naphthol PAN Method 1 Method to mg/l Mn (LR) Powder Pillows

Phosphorus, Reactive (Orthophosphate)

Sulfate DOC SulfaVer 4 Method 1. Method (0 70, and mg/l) Powder Pillows. Test procedure

Method 8017 (0.7 to 80.0 µg/l) Powder Pillows. Scope and Application: For water and wastewater; digestion is required to determine total cadmium.

USEPA 1 Periodate Oxidation Method 2 Method to 20.0 mg/l Mn (HR) Powder Pillows

USEPA 1,2 Bicinchoninate Method 3 Method 8506 (CuVer 1) and Method 8026 (CuVer 2) 0.04 to 5.00 mg/l Cu Powder Pillows or AccuVac Ampuls

DR/4000 PROCEDURE. IRON, Ferrous. Using Powder Pillows

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement, 0 to 40 mg/l COD

1,10 Phenanthroline Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows

Method to 500 µg/l CH 2 O Powder Pillows

Color, True and Apparent, LR

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Scope and application: For water, wastewater, seawater, brine solutions, produced waters and hydraulic fracturing waters.

Oxygen Demand, Chemical

Scope and application: For water and wastewater. Turbid samples are filtered during the analysis.

Oxygen Demand, Chemical

Dimethylphenol Method Method LR (0.23 to mg/l NO 3 N or 1.00 to mg/l NO 3 ) TNTplus 835

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Method (0.1 to 8.0 mg/l Cu) TNTplus 860

Table 163 Instrument-specific information. Instrument Pour-thru Kit Cell orientation Adapter. DR 6000 LQV Arrow faces right

STP (off-line) Water Technologies & Solutions analytical procedure. colorimetric method DCA reagent 1 (5/5/1) 2 to 15 mg/l as STP

Nitrogen, Simplified TKN (s-tkn )

Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell

Oxygen Demand, Chemical

Nitrogen, Simplified TKN (s-tkn ) DOC

Instrument Adapters Light shield DR 6000, DR 5000 DR 3900 LZV849 DR 3800, DR 2800 LZV646 DR or (A)

Oxygen Demand, Chemical

NACA Laboratory Equipment and Supplies

Direct ISE Method Method to 10.0 mg/l F Fluoride ISE

Amperometric Titrator assembly. Digital Titrator. Beaker, low-form, 250-mL. Stir bar, octagonal, Teflon-coated, 50.8 x 7.9 mm

Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric)

Direct ISE Method Method to 4.00 mg/l NO 3 N Nitrate ISE

Oxygen Demand, Biochemical

Professional Boiler Water Test Kit

Ground Water & UST Samples: Containers, Preservation and Hold Times Table

DR/4000 PROCEDURE OXYGEN SCAVENGERS. 4. Fill a second sample cell (the blank) with 25 ml of deionized water.

Why Hach Reagents? Alphabetical Listing of Reagents

BOD/CBOD, TSS ph OPERATOR BASICS. Presented by: Marcy Bolek - Alloway

Heterotrophic Bacteria

Surface Water Samples: Containers, Preservation and Hold Times Table

INSTRUCTION MANUAL. ba76126e01 03/2013. photoflex Series TEST INSTRUCTIONS AND SPECIFICATIONS, PROG. V 2.02 Y

Hardness, Calcium/Magnesium

Application Notes for COD Analysis DETERMINATION OF CHEMICAL OXYGEN DEMAND (COD) IN WATER AND WASTE WATER.

DR/800. Series Colorimeters. High-Value, High-Capability Instruments for Water Quality Analysis

C 99 & C 200 Series Multiparameter Bench Photometers

Coliforms, Total and E. coli

ANALYTICAL PROCEDURE. Bacteria: Hydrogen Sulfide Producing

:00 am 6:00 pm Mon-Thu (MT) 6:00 am 5:00 pm Fri (MT)

POCKET COLORIMETER II ANALYSIS SYSTEMS

USEPA Membrane Filtration Method Method m-tec. Scope and Application: For potable water, nonpotable water, recreation water and wastewater.

SECTION D.3 CHLOROPHYLL AND PHEOPHYTIN

Method 108C - Determination of Arsenic Content in Ore Samples From Nonferrous Smelters (Molybdenum Blue Photometric Procedure)

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.

METHOD DETERMINATION OF TURBIDITY BY NEPHELOMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.

Let s take a look several devices and methods for obtaining measurements of water quality in the field.

Zinc Testing and On Site Calibration for Water Testing Detection Range: 1 15ppm

Uranium Testing and Site Calibration for Water Testing Detection Range: 2 60ppb

Cadmium Testing and On Site Calibration for Water Testing Detection Range: ppm

Chemical Oxygen Demand

Human Myostatin, ELISA Kit (MSTN)

Cu (s) Cu 2+ (aq) Cu(OH) 2 (s) CuO (s) Cu 2+ (aq) Cu (s)

Technical Process Bulletin

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.

Suspended Metals Digestion Procedure

Engineered Systems & Designs TURBIDITY METER

STATE OF OHIO DEPARTMENT OF TRANSPORTATION SUPPLEMENT 1122 DETERMINING SULFATE CONTENT IN SOILS. July 17, 2015

CEE 311 Environmental Engineering I Fall 2015 Dr. Jagadish Torlapati BIOCHEMICAL OXYGEN DEMAND (BOD)

Chemical Test Kits. Table of Contents Introduction Reference Table Single Parameter Test Kits Multiparameter Test Kits

Technical Manual No Version

Rat gamma-aminobutyric acid (GABA) ELISA Kit

Human anti-hepatitis A virus (HAV) antibody (IgM) ELISA Kit

CTRP15/MYONECTIN (HUMAN) ELISA KIT

Mouse hepatitis E virus (HEV) antibody (IgG) ELISA Kit

Human rotavirus (RV) antibody (IgG) ELISA kit

Standard Methods for the Examination of Water and Wastewater

COLORIMETERS HACH POCKET COLORIMETER II. Fluoride 0.1 to 2.0 mg/l H

Edward F. Askew PhD Askew Scientific Consulting

Instruction Manual GB

Transcription:

PHOSPHORUS, REACTIVE (0.00 to 5.00 mg/l PO 4 ) PhosVer 3 Method, Test N Tube Procedure USEPA accepted for reporting wastewater analysis * Method 8048 For water, wastewater, and seawater 1. Enter the stored program number for reactive phosphorus (PO 4 ), Test N Tube. Press: PRGM The display will show: PRGM? Note: For most accurate results, perform a Reagent Blank Correction using deionized water (see Section 1). 2. Press: 82 ENTER The display will show mg/l, PO4 and the ZERO icon. Note: For alternate forms (P, P 2 O 5 ), press the CONC key. 3. Insert the COD/TNT Adapter into the cell holder by rotating the adapter until it drops into place. Then push down to fully insert it. Note: A diffuser band covers the light path holes on the adapter to give increased performance.the band should NOT be removed. 4. Use a TenSette Pipet to add 5.0 ml of sample to a Reactive Phosphorus Test 'N Tube Dilution Vial. Cap and mix. Note: For samples with extreme ph, see the Interference section. * Procedure is equivalent to USEPA Method 365.2 and Standard Method 4500-P E for wastewater. 481

5. Clean the outside of the vial with a towel. Note: Wiping with a damp towel, followed by a dry one, will remove fingerprints or other marks. 6. Place the sample vial into the adapter. Push straight down on the top of the vial until it seats solidly into the adapter. Note: Do not move the vial from side to side as this can cause errors. 7. Tightly cover the sample vial with the instrument cap. 8. Press: ZERO The cursor will move to the right, then the display will show: 0.00 mg/l PO4 Note: For multiple samples, zero only on the first sample. Read the remaining samples after adding the PhosVer 3 Reagent. 9. Using a funnel, add the contents of one PhosVer 3 Phosphate Powder Pillow to the vial. 10. Cap the vial tightly and shake for 10-15 seconds. Note: The powder will not completely dissolve. 11. Press: TIMER ENTER A 2-minute reaction time will begin. Note: Read samples between 2 and 8 minutes after the addition of the PhosVer 3 reagent. Note: A blue color will develop if phosphate is present. 12. Immediately after the timer beeps, place the sample vial in the adapter. Push straight down on the top of the vial until it seats solidly into the adapter. Note: Do not move the vial from side to side as this can cause errors. 482

13. Tightly cover the vial with the instrument cap. 14. Press: READ The cursor will move to the right, then the result in mg/l phosphate (PO 4 ) will be displayed. Note: Standard Adjust may be performed using a prepared standard (see Section 1). Sampling and Storage Collect samples in plastic or glass bottles that have been acid cleaned with 1:1 Hydrochloric Acid Solution and rinsed with deionized water. Do not use commercial detergents containing phosphate for cleaning glassware used in this test. Analyze samples immediately after collection for best results. If prompt analysis is impossible, preserve samples for up to 48 hours by filtering immediately and storing at 4 C. Warm to room temperature before analyzing the sample. Accuracy Check Note: Clean glassware with 1:1 hydrochloric acid solution. Rinse again with deionized water. Do not use detergents containing phosphates to clean glassware. Standard Additions Method a) Fill three 25-mL graduated mixing cylinders with 25 ml of sample. b) Snap the neck off a Phosphate PourRite Ampule Standard, 50 mg/l as PO 4. c) Use the TenSette Pipet to add 0.1 ml, 0.2 ml and 0.3 ml, respectively, to the three 25-mL aliquots of sample prepared in step a. Mix well. 483

d) Analyze each sample as described in the procedure; use 5.0 ml of the prepared standard additions for each test. The concentration should increase as follows: 0.2 mg/l, 0.4 mg/l, 0.6 mg/l PO 4, respectively. e) If these increases do not occur, see Standard Additions in Section 1 for more information. Method Performance Standard Solution Method To check accuracy, use a 1.0 mg/l Phosphate Standard Solution listed under Optional Reagents. Or, prepare a 1.0-mg/L PO 4 standard by pipetting 2 ml of solution from a Phosphate Voluette Ampule Standard for Phosphate, 50 mg/l as PO 4, into an acidwashed, Class A 100-mL volumetric flask. Dilute to the mark with deionized water. Substitute this standard for the sample and perform the procedure as described. Precision In a single laboratory, using a standard solution of 5.00 mg/l PO 4 and two lots of reagent with the instrument, a single operator obtained a standard deviation of ±0.08 mg/l PO 4. Estimated Detection Limit (EDL) The EDL for program 82 is 0.07 mg/l PO 4. For more information on derivation and use of Hach s estimated detection limit, see Section 1. 484

Interferences The following may interfere when present in concentrations exceeding these listed below: Aluminum Arsenate Chromium Copper Iron Nickel Silica Silicate Sulfide Substance Turbidity (large amounts) Zinc Highly buffered samples or extreme sample ph 200 mg/l Interference Level and Treatment Interferes at any level 100 mg/l 10 mg/l 100 mg/l 300 mg/l 50 mg/l 10 mg/l 6 mg/l. Sulfide interference may be removed by oxidation with Bromine Water as follows: 1. Measure 25 ml of sample into a 50-mL beaker. 2. Swirling constantly, add Bromine Water dropwise until a permanent yellow color develops. 3. Swirling constantly, add Phenol Solution dropwise until the yellow color just disappears. Proceed with step 1. May cause inconsistent results because the acid present in the powder pillows may dissolve some of the suspended particles and because of variable desorption of orthophosphate from the particles. 80 mg/l May exceed the buffering capacity of the reagents and require sample pretreatment; see ph Interferences (Section 1). The PhosVer 3 Phosphate Reagent Powder Pillows should be stored in a cool, dry environment. Sample Disposal Information Final samples will contain molybdenum. In addition, final samples will have a ph less than 2 and are considered corrosive (D002) by the Federal RCRA. 485

Summary of Method Orthophosphate reacts with molybdate in an acid medium to produce a phosphomolybdate complex. Ascorbic acid then reduces the complex, giving an intense molybdenum blue color. REQUIRED REAGENTS Reactive Phosphorus Test N Tube Reagent Set...50 tests... 27425-45 Includes: (1) 21060-46, (50) Orthophosphate Dilution Vials* Quantity Required Description Per Test Unit Cat. No. PhosVer 3 Phosphate Reagent Powder Pillows...1...50/pkg... 21060-46 50 Orthophosphate Test N Tube Dilution Vials...1...50/pkg... * REQUIRED APPARATUS COD/TNT Adapter...1...each... 48464-00 Funnel, micro...1...each... 258435 Pipet, TenSette, 1 to 10 ml...1...each... 19700-10 Pipet Tips, for 19700-10 TenSette Pipet...1...50/pkg... 21997-96 Test Tube Rack... 1-3...each... 18641-00 OPTIONAL REAGENTS Bromine Water, 30 g/l... 29 ml... 2211-20 Drinking Water Standard, Inorganic, F, NO 3 N, PO 4 3, SO 4 2......500mL... 28330-49 Hydrochloric Acid Standard Solution, 6.0 N (1:1)... 500 ml... 884-49 Phenol Solution, 30 g/l... 29 ml... 2112-20 Phosphate Standard Solution, 1 mg/l as PO 4... 500 ml... 2569-49 Phosphate Standard Solution, Voluette ampule, 50 mg/l as PO 4, 10 ml...16/pkg... 171-10 Phosphate Standard Solution, PourRite ampule, 50 mg/l as PO 4, 2 ml...20/pkg... 171-20H Wastewater Effluent Standard, Inorganic (NH 3 N, NO 3 N, PO 4, COD, SO 4, TOC)...500 ml... 28332-49 Water, deionized... 4 L... 272-56 * These items are not sold separately. 486

OPTIONAL APPARATUS Ampule Breaker, Pour Rite (2-mL ampule)... each...24846-00 Ampule Breaker Kit... each...21968-00 Aspirator, vacuum... each...2131-00 Cylinder, graduated, mixing, 25 ml (3 required)... each...20886-40 Dispenser, Repipet Jr., 2 ml... each...22307-01 Filter Holder, 47 mm, 300 ml, graduated... each...13529-00 Filter, membrane, 47 mm, 0.45 microns... 100/pkg...13530-00 Flask, filtering, 500 ml... each...546-49 Flask, volumetric, Class A, 100 ml... each...14574-42 ph Indicator Paper, 1 to 11 ph units... 5 rolls/pkg...391-33 ph Meter, sension 1, portable with electrode... each...51700-10 Pipet, TenSette, 0.1 to 1.0 ml... each...19700-01 Pipet Tips, for 19700-01... 50 pkg...21856-96 Pipet Tips, for 19700-01...... 1000/pkg...21856-28 Pipet Filler, Safety Bulb... each...14651-00 Pipet, volumetric, Class A, 5.00 ml... each...14515-37 Pipet, volumetric, Class A, 2.00 ml... each...14515-36 For Technical Assistance, Price and Ordering In the U.S.A. call 800-227-4224 Outside the U.S.A. Contact the Hach office or distributor serving you. 487