Hardness, Calcium/Magnesium

Similar documents
Instrument Adapters Light shield DR 6000, DR 5000 DR 3900 LZV849 DR 3800, DR 2800 LZV646 DR or (A)

Nitrogen, Simplified TKN (s-tkn )

Scope and application: For water, wastewater, seawater, brine solutions, produced waters and hydraulic fracturing waters.

USEPA 1 Periodate Oxidation Method 2 Method to 20.0 mg/l Mn (HR) Powder Pillows

Method (0.1 to 8.0 mg/l Cu) TNTplus 860

Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell

Phosphorus, Reactive (Orthophosphate)

Phosphorus, Reactive (Orthophosphate)

Sulfate. USEPA 1 SulfaVer 4 Method 2 Method to 70 mg/l SO 4. Powder Pillows. Test preparation. Before starting.

1-(2-Pyridylazo)-2-Naphthol PAN Method 1 Method to mg/l Mn (LR) Powder Pillows

Dimethylphenol Method Method LR (0.23 to mg/l NO 3 N or 1.00 to mg/l NO 3 ) TNTplus 835

Scope and application: For water and wastewater. Turbid samples are filtered during the analysis.

USEPA 1,2 Bicinchoninate Method 3 Method 8506 (CuVer 1) and Method 8026 (CuVer 2) 0.04 to 5.00 mg/l Cu Powder Pillows or AccuVac Ampuls

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Nitrogen, Simplified TKN (s-tkn ) DOC

Oxygen Demand, Chemical

Sulfate DOC SulfaVer 4 Method 1. Method (0 70, and mg/l) Powder Pillows. Test procedure

Color, True and Apparent, LR

Oxygen Demand, Chemical

Oxygen Demand, Chemical

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Table 163 Instrument-specific information. Instrument Pour-thru Kit Cell orientation Adapter. DR 6000 LQV Arrow faces right

Oxygen Demand, Chemical

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Direct ISE Method Method to 4.00 mg/l NO 3 N Nitrate ISE

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

DR/4000 PROCEDURE. IRON, Ferrous. Using Powder Pillows

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement

Method 8017 (0.7 to 80.0 µg/l) Powder Pillows. Scope and Application: For water and wastewater; digestion is required to determine total cadmium.

Direct ISE Method Method to 10.0 mg/l F Fluoride ISE

Method to 500 µg/l CH 2 O Powder Pillows

STP (off-line) Water Technologies & Solutions analytical procedure. colorimetric method DCA reagent 1 (5/5/1) 2 to 15 mg/l as STP

Phosphorus, Reactive (Orthophosphate)

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement, 0 to 40 mg/l COD

DR/4000 PROCEDURE OXYGEN SCAVENGERS. 4. Fill a second sample cell (the blank) with 25 ml of deionized water.

Amperometric Titrator assembly. Digital Titrator. Beaker, low-form, 250-mL. Stir bar, octagonal, Teflon-coated, 50.8 x 7.9 mm

1,5-Diphenylcarbohydrazide Method* (Powder Pillows or AccuVac Ampuls) USEPA accepted for wastewater analyses** Using Powder Pillows

FerroVer Method (Powder Pillows or AccuVac Ampuls) USEPA approved for reporting wastewater analysis (digestion is required; see Section 2 * )

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

3. Add 0.4 ml of. 7. Use a TenSette

3. Add 0.4 ml of. 7. Use a TenSette

PHOSPHORUS, REACTIVE (0.00 to 5.00 mg/l PO 4

Coliforms, Total and E. coli

1,10 Phenanthroline Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows

Heterotrophic Bacteria

USEPA Membrane Filtration Method Method m-tec. Scope and Application: For potable water, nonpotable water, recreation water and wastewater.

Oxygen Demand, Biochemical

TPTZ Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows

Professional Boiler Water Test Kit

PaSEC Colorimeter Instructions DR850

BIOSCAN* ATP method for biocide (toxicant) evaluations

SOP-C-102 Determination of Chemical Oxygen Demand

Human CRP Precoated ELISA kit

Cuvette Test LCK 555 BOD 5 /BOD [n]

Cuvette Test LCK 555 BOD 5 /BOD [n]

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.

Zinc Testing and On Site Calibration for Water Testing Detection Range: 1 15ppm

Cadmium Testing and On Site Calibration for Water Testing Detection Range: ppm

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.

Analysis of dissolved nitrite (NO 2- )

APPENDIX N BULK MILK TANKER SCREENING TEST FORM. CHARM ROSA TETRACYCLINE-SL TEST (DILUTION CONFIRMATION) (Raw Commingled Cow Milk) IMS #9-C17

CHEMetrics, Inc. Indigo Ozone Vacu-vials Technical Data Summary

Uranium Testing and Site Calibration for Water Testing Detection Range: 2 60ppb

Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric)

COLORIMETRIC SANDWICH ELISA KIT INSTRUCTION MANUAL

Glyphosate ELISA Kit. Catalog Number KA assays Version: 07. Intended for research use only.

Free, Total and Double SO2 in Wine

BONDERITE NT-1 Phosphate-free, Regulated Heavy Metal-Free Pre-treatment

OPERATING INSTRUCTIONS. Portable Turbidity Meter

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

:00 am 6:00 pm Mon-Thu (MT) 6:00 am 5:00 pm Fri (MT)

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.

MOUSE OPG ELISA KIT PAGE

The Caffeine ELISA Kit kit is for the quantitative and/or qualitative screening of Caffeine in water, urine, and saliva samples.

Atrazine ELISA Kit. Catalog Number KA assays Version: 15. Intended for research use only.

BIOSCAN* ATP method for monitoring of biofilm/sessile microorganisms in cooling systems

Technical Process Bulletin

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.

Iron 10 T mg/l Fe Ferrozine / Thioglycolate

Cuvette Test LCK 555 BOD 5 /BOD [n]

Human IL-1 alpha ELISA Kit

REAGENTS AND MATERIALS SUPPLIED Control 2: 25-OH vitamin D in human serum with <0.1% Streptomycin Sulfate (2)

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.

Why Hach Reagents? Alphabetical Listing of Reagents

This kit is intended for Research Use Only. Not for use in diagnostic procedures.

DR /2400 and DREL /2400 Spectrophotometer + Portable Laboratories. Get moving with the DR/ 2400 spectrophotometer: Made-to-measure in situ analysis

Porcine IgM (Immunoglobulin M) ELISA Kit

Human D-Dimer, D2D ELISA Kit

Human placenta lactogen (HPL) ELISA Kit

QualiTube Kit For Microcystin

Bovine rotavirus (RV) antigen (Ag) ELISA Kit

Advanced Photometer System IDEAL FOR DRINKING WATER, POOLS AND SPAS, ENVIRONMENTAL, & EDUCATIONAL TESTING

DR 1900 PORTABLE SPECTROPHOTOMETER

Human procollagen type III (HPCⅢ) ELISA Kit

HUMAN SOLUBLE RANK ELISA KIT

(Applicable to identification of Salmonella in all foods.)

Uranium Testing and On Site Calibration for Water Testing Detection Range: 2 60ppb

HUMAN FOLLISTATIN ELISA KIT

HUMAN CONNECTIVE TISSUE GROWTH FACTOR (CTGF) ELISA KIT

Transcription:

Hardness, Calcium/Magnesium DOC316.53.01528 Metal Phthalein Colorimetric Method Method 10293 20 to 350 mg/l CaCO 3, 5 to 100 mg/l Ca or 3 to 50 mg/l Mg TNTplus 869 Scope and application: For raw (untreated) water, drinking water, surface water, boiler water and process control. Test preparation Instrument-specific information Table 1 shows all of the instruments that have the program for this test. The table also shows the adapter and light shield requirements for the applicable instruments that can use TNTplus vials. To use the table, select an instrument, then read across to find the applicable information for this test. Table 1 Instrument-specific information for TNTplus vials Instrument Adapters Light shield DR 6000, DR 5000 DR 3900 LZV849 DR 3800, DR 2800 LZV646 DR 1900 9609900 or 9609800 (A) Before starting DR 3900, DR 3800, DR 2800: Install the light shield in Cell Compartment #2 before this test is started. Review the safety information and the expiration date on the package. The sample ph must be 4 9 for accurate results. The temperature of the samples and reagents must be 15 25 C (59 77 F) for accurate results. The recommended temperature for reagent storage is 15 25 C (59 77 F). To determine magnesium in samples with a very low magnesium concentration and a high calcium concentration, use the TNTplus 849 Magnesium procedure. DR 1900: Go to All Programs>LCK or TNTplus Methods>Options to select the TNTplus number for the test. Other instruments automatically select the method from the barcode on the Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective equipment. Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory agencies for further disposal information. Items to collect Description Quantity Water Hardness TNTplus Reagent Set 1 Pipet, adjustable volume, 1.0 5.0 ml 1 Pipet, adjustable volume, 0.2 1.0 ml 1 Pipet tips 3 1

Refer to Consumables and replacement items on page 4 for order information. Sample collection Test procedure Total hardness Collect samples in clean glass or plastic bottles with tight-fitting caps. Completely fill the bottle and immediately tighten the cap. Prevent agitation of the sample and exposure to air. Analyze the samples as soon as possible for best results. If immediate analysis is not possible, keep the samples at or below 6 C (43 F) for a maximum of 24 hours. Let the sample temperature increase to room temperature before analysis. 1. Use a pipet to add 4.0 ml of Solution A to the test 2. Tighten the cap on the vial and invert until completely mixed. 3. Start the reaction time of 2 minutes. 4. When the timer expires, clean the Start 5. DR 1900 only: Select program 869. Refer to Before starting on page 1. 6. Insert the vial into the cell holder. DR 1900 only: Push READ 1. The instrument zero is set. 7. Remove the 8. Use a pipet to add 0.2 ml of sample to the test 9. Tighten the cap on the vial and invert the vial 2 3 times. 10. Start the reaction time of 30 seconds. 11. When the timer expires, clean the 12. Insert the vial into the cell holder. DR 1900 only: Push READ 2. The total hardness result shows in mg/l CaCO 3. 2 Water Hardness, TNTplus Method (350 mg/l as CaCO 3 )

Test procedure Total, calcium and magnesium hardness 1. Use a pipet to add 4.0 ml of Solution A to the test 2. Tighten the cap on the vial and invert until completely mixed. 3. Start the reaction time of 2 minutes. 4. When the timer expires, clean the Start 5. DR 1900 only: Select program 869. Refer to Before starting on page 1. 6. Insert the vial into the cell holder. DR 1900 only: Push READ 1. The instrument zero is set. 7. Remove the 8. Use a pipet to add 0.2 ml of sample to the test 9. Tighten the cap on the vial and invert the vial 2 3 times. 10. Start the reaction time of 30 seconds. 11. When the timer expires, clean the 12. Insert the vial into the cell holder. DR 1900 only: Push READ 2. 13. Remove the 14. Use a pipet to add 0.2 ml of Solution B to the test 15. Tighten the cap on the vial and invert the vial 2 3 times. 16. Start the reaction time of 30 seconds. Water Hardness, TNTplus Method (350 mg/l as CaCO 3 ) 3

17. When the timer expires, clean the 18. Insert the vial into the cell holder. DR 1900 only: Push READ 3. Results show: Calcium: mg/l Ca, Magnesium: mg/l Mg, Total Hardness: mg/l CaCO 3. Interferences Accuracy check There are no known interferences for drinking water and boiler water samples. To validate the test results, dilute the sample with a known volume of deionized water. Use the diluted sample in the test procedure and multiply the result by the dilution factor. As an alternative, spike the sample with a standard solution and compare the expected result to the actual result. Standard solution method Use the standard solution method to validate the test procedure, the reagents and the instrument. Items to collect: 100-mg/L (as Ca) Calcium Standard Solution 1. Use the test procedure to measure the concentration of the standard solution. 2. Compare the expected result to the actual result. Note: 100 mg/l calcium as Ca is equivalent to 250 mg/l hardness as CaCO 3. Summary of Method Calcium and magnesium ions react with metal phthalein to give a violet dye. The measurement wavelength is 572 nm. Consumables and replacement items Required reagents Description Quantity/Test Unit Item no. Water Hardness TNTplus Reagent Set 1 25/pkg TNT869 Required apparatus Description Quantity/test Unit Item no. Pipet, adjustable volume, 1.0 5.0 ml 1 each BBP065 Pipet tips, for 1.0 5.0 ml pipet 1 75/pkg BBP068 Pipet, adjustable volume, 0.2 1.0 ml 1 each BBP078 Pipet tips, for 0.2 1.0 ml pipet 2 100/pkg BBP079 4 Water Hardness, TNTplus Method (350 mg/l as CaCO 3 )

Required apparatus (continued) Description Quantity/test Unit Item no. Light shield, DR 3900 1 each LZV849 Light shield, DR 3800, DR 2800, DR 2700 1 each LZV646 Recommended standards Description Unit Item no. Calcium standard solution, 100 mg/l Ca 500 ml 4457649 Water Hardness, TNTplus Method (350 mg/l as CaCO 3 ) 5

FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932 Hach Company/Hach Lange GmbH, 2018. All rights reserved. 10/2018, Edition 1