Ultrahydrogel Columns

Similar documents
Protein-Pak Hi Res HIC Column and HIC Protein Standard

Agilent MetaCarb 87H Organic Acids Column H + Form User Manual Part No. A5214

ION EXCHANGE KIT FOR MAB SEPARATIONS

[ Care and Use Manual ]

MassPREP On-Line Desalting Cartridge

HPLC Columns. Protein LW-803 MANUAL. Shodex HPLC Columns Europe, Middle East, Africa, Russia

[ Care and Use Manual ] Waters AccellPlus QMA and CM Bulk Media. I. calculating bulk media needs. II. packed column use. IIi.

HPLC Columns. HILICpak VN-50 2D, 4D MANUAL. Shodex HPLC Columns Europe, Middle East, Africa, Russia

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

HPLC Columns HIC PH-814 MANUAL. Shodex HPLC Columns Europe, Middle East, Africa, Russia. For technical support please use contact details shown below:

Agilent MetaCarb Ca PLUS Carbohydrate Column Ca ++ Form User Manual Part No. A5205

Rev. CO Packed Column for Ion Exchange Chromatography. TSKgel STAT Series INSTRUCTION MANUAL TOSOH CORPORATION

( ) IONPAC ATC-HC ANION TRAP COLUMN

3. Close the bottom end of the column and apply the packing device on top. Pump water through the upper adaptor to remove air.

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime

the power of discovery USER GUIDE FOR ORGANIC ACIDS ANALYSIS COLUMNS

G Sep Size Exclusion Columns

Sartobind TM Membrane Adsorbers Discs and Cassettes

G-Sep Ion Exchange Agarose Fast Flow

G-Sep Ion Exchange Agarose Fast Flow

AFFINITY HIS-TAG PURIFICATION

Gelatin Sepharose. instructions

ACQUITY UPC 2 BEH, CSH, and HSS Columns


Red Sepharose CL-6B INSTRUCTIONS

Table of contents. p. 2

Econo-Pac High Capacity Ion Exchange Cartridges, 5 ml Instruction Manual. Catalog Numbers , ,

Operating Instructions

Arginine Sepharose 4B

5 AMP Sepharose 4B. instructions

Userguide: SepFast Gel Filtration columns

Nuvia S and Q High-Capacity Ion Exchange Media Instruction Manual

Superdex prep grade. GE Healthcare Life Sciences

Agilent Multiple Affinity Removal Spin Cartridges for the Depletion of High-Abundant Proteins from Human Proteomic Samples.

for Acclaim Carbamate Column

Size Exclusion Chromatography

CNBr-activated Sepharose 4B

TSK-GEL BioAssist Series Ion Exchange Columns

Superdex 200 HR 10/30 INSTRUCTIONS

order:

GORE. Protein Capture Devices. Operating Instructions Contents. Product Description Device Characteristics... 2 General Handling...

Module 16: Gel filtration: Principle, Methodology & applications. Dr. Savita Yadav Professor Department of Biophysics AIIMS, New Delhi

HiTrap Desalting, 5 ml

Sartobind IEX nano 3 ml

Blue Sepharose CL-6B. instructions

Glutathione Agarose Resin User s Guide

AFFINITY HIS-TAG PURIFICATION

Operating Instructions

Preparative, Pilot & Process Columns

Phenyl Sepharose CL-4B

[ Care and Use Manual ]

UNOsphere Q & S Ion Exchange Media. Instruction Manual ...

2 5 ADP Sepharose 4B. GE Healthcare Life Sciences

Caring for Turbo V Ion Source Electrodes

Product Guide for LudgerSep TM N2 High Resolution Amide HPLC Columns for Glycan Analysis

ACQUITY UPLC Protein BEH SEC Columns and Standards

Introduction Ron Majors is a Senior Scientist at Agilent. Bill Champion is a chemist in Agilent s HPLC Columns tech support group.

1 ml gel corresponds to ml of 75% (v/v) Glutathione Agarose suspension.

Superose 6 HR 10/30 Superose 12 HR 10/30

G Trap Ion Exchange Selection Kit

WorkBeads Protein A INSTRUCTIONS IN

G-Sep IDA Agarose Fast Flow

ProteoSpin CBED (Concentration, Buffer Exchange and Desalting)

Hamilton PRP Polymeric Columns

Isolation of Protein

CLEAN SCREEN FASt FILTER AND SHOOT UCT FORENSICS. CLEAN SCREEN FASt SPE COLUMNS

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

MBP Excellose handbook - Purification of MBP fusion proteins -

G-Sep Ni-NTA Agarose Fast Flow

Experiment 11. DNA Composition by HPLC 1,2

The Impact of Spare Parts. Performance

AFFINITY HIS-TAG PURIFICATION

Cellufine MAX Q-r, Cellufine MAX Q-h

Improving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract

Instructions AD Hydrophobic interaction

Dionex IonPac AS14 Columns

Principles of Gel Filtration Chromatography

AFFINITY GST PURIFICATION

Erstellt mit einer Testversion von PDF Annotator - FLUIDIC & OPTICAL PRODUCTS AND INFORMATION

Product Manual for the Dionex IonPac Ultra Trace Anion Concentrator (UTAC)

Operating Instructions

Data File. HiLoad Superdex 30 prep grade HiLoad Superdex 75 prep grade HiLoad Superdex 200 prep grade. Pre-packed gel filtration columns

Continuously Regenerated Trap (CR-TC 500, Capillary)

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

Operating Instructions

High Throughput Sub-4 Minute Separation of Antibodies using Size Exclusion Chromatography

Bioneer Corporation 8-11,Munpyeongseo-ro, Daedeok-gu, Daejeon 34302, Republic of Korea Tel: Fax:

Bio-Scale Mini Affi-Prep Protein A Cartridges, 1 and 5 ml. Instruction Manual. Catalog # ,

Q SepFast HighRes. DEAE SepFast TM HighRes S SepFast TM HighRes CM SepFast TM HighRes

Techniques for Improving the Efficiency of Large Volume Loading in Preparative Liquid Chromatography

AFFINITY GST PURIFICATION

Technical data and operating instructions. Vivacon 500. For in vitro use only

Principles of Gel Filtration Chromatography

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Inertsil ODS-3 2 µm. Base Silica Physical Properties and Chemical Modification. Base Silica: High Purity Silica Gel % Surface Area: 450 m 2 /g

Phenyl Sepharose High Performance Butyl Sepharose High Performance

AFFINITY HIS-TAG PURIFICATION

KOBRA CELL Art. No. K01

Sean M. McCarthy and Martin Gilar Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL RESULTS AND DISCUSSION

SERVA IMAC Ni-IDA Test Kit Agarose for Affinity Purification of His-Tag Fusion Proteins

Transcription:

CONTENTS I. INTRODUCTION II. INSTALLATION a. Solvents b. Column c. Organic solvents d. Aqueous salt and buffer solution e. Flow rate f. Solvent and sample preparation g. Guard columns h. Column installation III. OPERATING TIPS a. Column efficiency b. System efficiency IV. COLUMN STORAGE a. Short term (several days) b. Long term (more than three days) I. INTRODUCTION The Waters Ultrahydrogel Column you have purchased is a sophisticated aqueous-compatible gel column for analytical and preparative separations of water-soluble polymers. The gel is a cross-linked hydroxylated polymer and contains some residual carboxyl functionality. Ultrahydrogel DP contains some residual amine functionality. Compounds normally characterized include oligomers and biological substances such as polysaccharides, nucleic acids, proteins, and peptides. This Ultrahydrogel Column provides: High resolution Low adsorption A wide range of molecular weight separations Stability in a ph range of 2 to 12 A temperature range of 10 C to 80 C Please take a few moments to read this manual carefully. Use the procedures it contains to ensure that you obtain quality results and take full advantage of the features your Waters column offers. V. WARRANTY VI. ORDERING INFORMATION

II. INSTALLATION a. Solvents are packed in distilled water for shipping. Table 1 shows the maximum solvent replacement flow rate for each Ultrahydrogel Column. Perform solvent changeover at a lower flow rate than that shown in Table 2 for analysis. Caution: are flow (pressure) sensitive and extreme care must be taken not to exceed maximum flow rate on start-up. b. Column Table 1. Flow rates for replacing solvents Column Column size Max. flow rate during solvent change Ultrahydrogel (all pore sizes) 7.8 mm I.D. x 30 cm 0.3 ml/min Ultrahydrogel DP 7.8 mm I.D. x 30 cm 0.3 ml/min Ultrahydrogel DNA 7.8 mm I.D.. x 30 cm 0.3 ml/min Ultrahydrogel Linear 7.8 mm I.D. x 30 cm 0.3 ml/min Note: Change solvents as seldom as possible. Frequent solvent exchanges accelerate the degradation of column efficiency. c. Organic solvents are physically and chemically stable in ordinary water-soluble organic solvent mixtures. Aqueous solutions of methanol, ethanol, acetonitrile, formic acid, and dimethylsulfoxide also provide dependable separations. Organic solvent concentration should not exceed 20%. It is possible however, to use organic solvents up to 50% concentration by gradual introduction via gradient method. d. Aqueous salt and buffer solution Although some non-ionic samples can be analyzed in distilled water, most water soluble polymers require aqueous salt or buffer solutions. Some typical aqueous salt solutions: Sodium sulfate Sodium acetate Sodium dihydrogenphosphate Compatible buffer solutions: Phosphate Tris phosphate Tris hydrochloric acid Citrate Tris acetate Acetate e. Flow rate For best resolution and maximum column life the flow rate should not exceed the maximum flow rate listed below. Use Table 2 below to determine recommended flow rates and maximum pressures. Table 2. Recommended flow rates Column Typical operating flow rate (ml/min) Maximum flow rate (ml/min) Ultrahydrogel 120 0.5 0.8 1.0 Ultrahydrogel 250 0.5 0.8 1.0 Ultrahydrogel 500 0.3 0.6 1.0 Ultrahydrogel 1000 0.3 0.6 1.0 Ultrahydrogel 2000 0.3 0.6 1.0 Ultrahydrogel Linear 0.3 0.6 1.0 Ultrahydrogel DP 0.5 0.8 1.0 f. Solvent and sample preparation Use LC grade solvents, filtered through a GHP (or other suitable) membrane to remove microparticulate matter above 0.45 μm. Use of the Waters Solvent Clarification Kit (p/n WAT085113) is recommended. Particulate-free solvents reduce the problem of plugged filters and column beds and preserve column life. Vacuum filtration or sonication of solvents may be used to remove dissolved gases which could affect the solvent delivery system. Always filter prepared sample solutions to prevent excessive pressure buildup due to particulate matter (Acrodisc Filters are available for this purpose). Use of a Waters In-Line Precolumn Filter (p/n WAT084560) is recommended to obtain maximum column life. Ammonium acetate Ammonium formate 2

g. Guard Columns While guard columns are not generally used for organic GPC, they will remove and/or filter many impurities found in aqueous soluable samples such as saccharides, peptides and hydrolyzates that might adsorb irreversibly on the Ultrahydrogel Column. Guard columns are also used to remove particulates and strongly retained sample components which, ultimately decrease column life from the sample and solvent. Guard columns will aid in obtaining consistently reliable and reproducible results with your Ultrahydrogel Column. Ultrahydrogel Guard Columns are available in both Ultrahydrogel (p/n WAT011565) and Ultrahydrogel DP (p/n WAT011570) configurations. Guard columns have a finite capacity for impurities. Either the number of sample injections or an observed change in backpressure will indicate when a change of guard column is required. If the capacity of the guard column is exceeded, undesirable contaminants will elute onto the analytical column. or guard column. Follow the next four steps of this procedure to cut tubing to connect a new steel column, or to improve the end connections on your existing fittings. Note: When changing columns, a new ferrule must be carefully reseated to ensure that no dead volume is present. 1. Using a three-cornered file with a cutting edge scribe the circumference of the tubing at the desired break. 2. Grasp the tubing on both sides of the scribe mark with clothcovered pliers (to prevent marring the tube surface) and gently work the tube back and forth until it separates. 3. Slide the compression fitting, followed by the ferrule (large end of the taper first) over the tube. Be certain to bottom the tube in the fitting seat to assure a leak-free connection. Note: Attach a union in place of the column and flush the lines free of previous solvents before attaching the column. While guard columns will help lengthen column life, they perform a specific function for a short period of time and must be changed or regenerated often. It is important to use good laboratory practices to increase column life. Ultrahydrogel Guard Columns can be regenerated by flushing with 6 10 column volumes of one or more of the following solvents: Tube Compression screw or nut Critical distance to be determined by each application (union, column fitting, etc.) Ferrule End must be straight and smooth to achieve maximum column efficiency Salt buffer (0.5M 1.OM) High ph buffer (ph 9 12) Urea, SDS or other surfactants Low ph buffer (ph 2 3) Methanol or acetonitrile h. Column Installation Remove the end cap fittings from your steel column and save them to recap the column when it is removed from the system. The column outlet is indicated by an arrow on the label showing the direction that the solvent should flow. If solvent is not visible when the end cap is removed, connect the outlet side of the column to the system and slowly pump solvent through the column to expel the air. Rapid pressurization or solvent flow may cause degradation of the column. When solvent flow is visible at the inlet side of the column, connect the column in the direction of normal flow and attach the inlet side of the column to the injector Figure 1. Ferrule and compression screw assembly If you wish to use in series: 1. Connect the columns in order of decreasing pore size (e.g. Ultra 500 before Ultra 250). 2. To minimize dead volume insert interconnecting tubing into the compression fittings before tightening the fittings. 3. Connect the outlet end of the last column to the detector. 3

II. OPERATING TIPS Use the column only in the direction indicated by the arrow. Flow in the reverse direction may cause degradation in column performance. Be careful not to introduce air into the column during installation or removal. Air bubbles will cause voids in the column packing. Slowly increase flow rates and change solvents. Continue to pump solvent through the column after separation is complete. Air may be sucked into the column by contraction of the solvent if the pump is turned off while the column is still hot. Maintain a slow flow rate (below 0.2 ml/min) through the column to prevent precipitation of buffer salts during daily operation. If halides are being used, replace the buffer with distilled water for overnight storage. maybe operated at a temperature range of 10 C to 80 C. High temperature analysis reduces viscosity, increases theoretical plate count and resolution and reduces adsorptivity. a. Column efficiency Liquid chromatography columns have a finite life that is directly related to the care and use they receive. Column life is influenced by the number of injections, sample and solvent cleanliness, frequency of solvent changeover, and handling and storage procedures. To perform a plate count to check for column efficiency, proceed as follows: Flow rate: 0.8 ml/min Detector: Refractive Index Eluent: Water Temperature: Ambient Injection vol.: 20 μl Marker: Ethylene glycol Test standard conc.: 0.0025 ml/ml Calculate the plates (MN) at the half height using the calculations shown in Figure 2. INJECT V R h W½ ½h N = Column efficiency (plates) V R = Volume to peak apex (µl) W½= Volume of peak at ½ peak height (µl) N = 5.54 x V R W½ 2 Figure 2. Half height method test calculations 4

b. System efficiency Column degeneration may not always be the cause of unacceptable sample resolution values. Troubleshooting the entire system may be necessary to determine the exact cause of low performance. In order to isolate the cause of poor system efficiency, always measure bandspreading before installing a column. If loss of resolution is observed after column use, repeat the test for system bandspreading and compare with the initial bandspreading results. If bandspreading has increased, focus troubleshooting efforts on the system hardware (such as the injector or an in-line filter). Isolating or eliminating the column as the source of reduced efficiency in an LC system allows the operator to perform maintenance procedures to restore system efficiency with minimal downtime. V. WARRANTY Waters Corporation warrants its high performance liquid chromatography columns in accordance with the following terms and conditions: Waters will repack or replace (at our discretion) without cost any column that fails to perform satisfactorily if notified within 90 days from your receipt of a steel column. Any column that is returned must have a return authorization number granted by the Waters Customer Service Department. Approval is subject to the following exclusions: Physical damage to the column because of misuse or abuse. Chemical damage to the packing material because of use with incompatible solvents or buffers, or at an incorrect ph. Physical damage to the packing material because of operation IV. COLUMN STORAGE a. Short term (several days) If the solvent does not contain buffers or other harmful salts, (e.g. halogens), remove the column from the system and seal both ends with the end cap fittings provided. at incorrect temperatures or pressures. Particulate buildup or precipitation in the column or end fittings causing high internal pressure which has occurred because of improper solvent or sample filtration practices. If the solvent does contain buffers, etc., rinse the column with distilled water, remove the column from the system and seal with the end cap fittings provided. Store at room temperature. Avoid direct sunlight. Keep away from corrosive gases. b. Long term (more than three days) Replace the solvent with distilled water. Remove the column from the system and seal both ends with the end cap fittings provided. Adding sodium azide (0.05%) to the water is recommended for storage periods of 72 hours or more. Store at room temperature. Avoid direct sunlight. Keep away from corrosive gases. 5

VI. ORDERING INFORMATION Item Ultrahydrogel 120 Ultrahydrogel 250 Ultrahydrogel 500 Ultrahydrogel 1000 Ultrahydrogel 2000 Ultrahydrogel Linear Ultrahydrogel DP Guard Columns Ultrahydrogel (for Ultrahydrogel 120 2000 and Linear) Ultrahydrogel DP (for Ultrahydrogel DP) Part number WAT011520 WAT011525 WAT011530 WAT011535 WAT011540 WAT011545 WAT011550 WAT011565 WAT011570 Waters, The Science of What s Possible, and Ultrahydrogel are registered trademarks of Waters Corporation. All other trademarks are the property of their respective owners. 2015 Waters Corporation. Produced in the U.S.A. March 2015 WAT011556 Rev F AW-PDF Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com