Setting Up a Multiplex (two probe) Real Time 5 Nuclease Reaction on the ABI PRISM 7700 Sequence Detection System

Similar documents
Designing TaqMan MGB Probe and Primer Sets for Gene Expression Using Primer Express Software Version 2.0

QIAGEN Supplementary Protocol

Applied Biosystems Real-Time PCR Rapid Assay Development Guidelines

User Bulletin. ABI PRISM 7000 Sequence Detection System DRAFT. SUBJECT: Software Upgrade from Version 1.0 to 1.1. Version 1.

Applied Biosystems 7900HT Fast Real-Time PCR System. Performing Fast Gene Quantitation with 384-Well Plates DRAFT

Applied Biosystems 7500 Fast Real-Time PCR System

Real-Time PCR Workshop Gene Expression. Applications Absolute and Relative Quantitation

ProNex DNA QC Assay Calibration Kit, 7500

TaqMan Residual DNA Detection Kit Chinese Hamster Ovary v2 Protocol

TaqMan SNP Genotyping Protocol

resdnaseq Quantitative DNA Kits

TaqMan Cytokine Gene Expression Plate I Protocol

LightCycler 480 Instrument Quick Guide

Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds. Overview. Data Analysis Tutorial

iq 5 Calibration: Contents.

Absolute Quantification Getting Started Guide

Applied Biosystems TaqMan Array Microfluidic Cards

Real-Time PCR ABI Prism 7500 (PSC 563, NSC 338) Real-Time PCR StepOne (PSC 563)

Introducing a new generation of affordable choices from the leader in real-time PCR. Applied Biosystems 7500 Fast Real-Time PCR System

JAK2 MutaScreen TM Kit & Reference Scale

Applied Biosystems 7500 Fast, 7500 and 7300 Real-Time PCR Systems

Internal RNA extraction control. Instructions for use of RNA real-time PCR internal extraction control kit

Internal DNA extraction control

Relative Quantification Getting Started Guide

Relative Quantification Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System

M. tuberculosis_mpb64/is611. genesig Advanced Kit. 150 tests. Primerdesign Ltd. For general laboratory and research use only

Xerox Configurator Pricing Manager Guide

Genomic DNA quantification assay. Quantification of genomic DNA by real-time PCR

Quantifiler Human DNA Quantification Kit Quantifiler Y Human Male DNA Quantification Kit

Plexor HY System for the Applied Biosystems 7500 and 7500 FAST Real-Time PCR Systems

Escherichia coli_o104: H4

Use of the GenePrint 10 System with the Applied Biosystems 3500 and 3500xL Genetic Analyzers and POP-7 Polymer

Internal RNA extraction control. Instructions for use of RNA real-time PCR internal extraction control kit

Pistacia vera. Introduction to Pistacia vera. 100 tests. Techne qpcr test. For general laboratory and research use only

Plus/Minus Assay. Getting Started Guide. Applied Biosystems 7900HT Fast Real-Time PCR System. Introduction. Designing a. Plus/Minus Assay.

HID Real-Time PCR Analysis Software v1.1

Techne qpcr test. A graveolens. NADPH-dependent mannose 6- phosphate reductase (m6pr) 100 tests

Techne qpcr test. Pisum sativum. Ribosomal protein S12 (rps12) gene. 100 tests

Roadway ExpressWORKS to WorldShip Commodity Transfer

AccuSEQ Real-Time PCR Detection Software

Mycobacterium genus (all species)

Maize CaMV promoter & NOS terminator (GMO)

User Bulletin. ABI PRISM 373 DNA Sequencer. Using the ABI PRISM 373 BigDye Filter Wheel. Contents. Introduction. February 16, 1998 (updated 06/2001)

Module1TheBasicsofRealTimePCR Monday, March 19, 2007

User Bulletin #2. ABI PRISM 7700 Sequence Detection System. Introduction. Contents. December 11, 1997

Dengue Virus subtypes 1, 2, 3 and 4 (Multiplex kit)

Primerdesign Ltd. High risk Human Papillomavirus. Multiplex screening kit. genesig kit. 100 tests. For general laboratory and research use only

M. tuberculosis_mpb64/is61 10 genesig Standard Kit

ReadMe Software v2.0.1 For Applied Biosystems 7500 and 7500 Fast Real-Time PCR Systems

Anacardium occidentale

ARM-D Kits Data Acquisition and Analysis Guide. Real-Time PCR Platform: Applied Biosystems QuantStudio 7 Flex Real-Time PCR System

NOS terminator (GMO) genesig Advanced Kit. NOS terminator. 150 tests. Primerdesign Ltd. For general laboratory and research use only

Relative Quantification Getting Started Guide

Love Bird (Agapornis) Sexing Sex chromosome specific spindlin gene. Advanced Kit. 150 tests. For general laboratory and research use only

Design and Ordering Guide. Custom TaqMan Assays. For New SNP Genotyping and Gene Expression Assays

Arachis hypogaea A.hypogaea. genesig Allergen detection Kit. 100 tests. Primerdesign Ltd. For general laboratory and research use only

PowerSeq Quant MS System

genorm TM Reference Gene Selection Kit

MicroSEQ E. coli O157:H7 Detection Kit

Table of Contents. Welcome to igo Figure...1 About this Guide...1 A Few Important Things to Know...1

Corylus avellana C.avellana. genesig Allergen detection Kit. 100 tests. Primerdesign Ltd. For general laboratory and research use only

Soya Round up Ready (RR)

PowerPlex Matrix Standards, 3100/3130

High Resolution Melting Protocol

CaMV promoter & NOS terminator detection

TECHNICAL MANUAL. ProNex DNA QC Assay. for Use on the Bio-Rad CFX96 Touch Real-Time. PCR Detection System

Human Y-Chromosome detection assay

Prunus dulcis genesig Allergen detection Kit

Sus scrofa Pig genesig Advanced Speciation Kit

Real-Time PCR Validations

OPEN ENROLLMENT. Employee Benefits

UC Any and All Regular Payroll History

myt KRAS qpcr primers for detection of seven human KRAS codon 12/13 mutations:

Sus scrofa Pig genesig Speciation Kit

PathoSEEK Microbial Safety Testing Platform on the BIO-RAD CFX96 Touch Real-Time PCR Detection System Decontamination Step Included Page 1 of 13

Touchscreen (standalone capabilities, PIN-protected user accounts, and dye calibration/rnasep functionality)

Mycobacterium Tuberculosis

TaqPath ProAmp Master Mixes

SUBJECT: System Software Patch Provides Support for Fast SYBR Green Master Mix and RNA-to-C T

Epstein Barr Virus (Human Herpes virus 4)

Lactobacillus 16S Ribosomal RNA. genesig Advanced Kit. 150 tests. Primerdesign Ltd. For general laboratory and research use only

Loss of Native Allele Protocol for KOMP-Regeneron derived ESC lines January 9 th 2009

Real-Time Quantitative PCR (qpcr) Protocol... 5 Glossary and Definitions DICSCLAIMER LIMITED USE LABEL LICENSE... 10

Job Aid. Enter a Rate Based Contract

Human Papillomavirus 16

HR APPROVALS. Version 5.5

Expression Array System

Human Herpes Virus 1 (Herpes simplex type 1)

Investigator Quantiplex Handbook

Relative Quantification (Mono-Color) Unknown samples (purified total RNA, mrna or cdna or genomic DNA)

Staphylococcus epidermidis

Setting up a Run of Real-time PCR CFX 96 Real-Time System

Prunus dulcis P.dulcis. genesig Allergen detection Kit. 100 tests. Primerdesign Ltd. For general laboratory and research use only

BioBank cdna kit. Instructions for the use of BioBank control cdna in real-time PCR

APPENDIX C: LAB SOLUTIONS

Epstein Barr Virus (Human Herpes virus 4)

Apium graveolens Celery/Celeriac. genesig Allergen detection Kit. 100 tests. Primerdesign Ltd. For general laboratory and research use only

MYOB Support Note. Invoices Aging Incorrectly

FMV 35S promoter (GMO) Advanced Kit. FMV 35S promoter. 150 tests. For general laboratory and research use only

MANUAL MY.DHLPARCEL.NL

Transcription:

Setting Up a Multiplex (two probe) Real Time 5 Nuclease Reaction on the ABI PRISM 7700 Sequence Detection System Overview The goal of this tutorial is to demonstrate how to correctly label a plate set-up for a multiplex reaction (two reporter dyes in one tube) on an ABI PRISM 7700 Sequence Detection System (SDS). For this example we will use a TaqMan FAM dye-labeled probe and a TaqMan VIC dye-labeled probe. Throughout this tutorial, words in bold type are actual menu options in the ABI PRISM 7700 SDS software. Please Note: If you are not familiar with setting up the ABI PRISM 7700 SDS software, please refer to the tutorial Setting-up Singleplex (single probe) Real Time 5' Nuclease Reactions on the ABI PRISM 7700 Sequence Detection System before proceeding. The Singleplex tutorial covers certain basic setup and operation steps that will not be covered in this tutorial. Launching ABI PRISM 7700 SDS Software Launch the ABI PRISM 7700 SDS software. For a real-time run, the Plate Type is Single Reporter, Plate Format is Standard Plate, and Run is Real Time. Labeling the Dye Layers The SDS software is set up so that FAM is always the default dye layer. Set up the FAM dye layer first, labeling all of the wells that contain the TaqMan FAM dye-labeled probe. Page 1 of 8

Then change to the VIC dye layer using the Dye Layer drop-down menu, which is just to the right of the Show Analysis button on the plate set-up. Sample Type Setup In order to set up the VIC dye layer, you will have to generate some new Sample Type names. To do this, go to Sample Type Setup... which is found in the drop down menu next to Sample Type, at the top left of the setup window. Click the ADD button on the bottom left corner of this window. Scroll down within the Page 2 of 8

window to the newly created row in which you can assign the Acronym, Name (Sample Type), Color, and Reporter. When choosing an Acronym, do not use any of the existing acronyms displayed in the list in this window. For example, if you want to add an unknown in the VIC dye layer, do not use the acronym UNKN, because this is already used in the FAM dye layer. Instead choose something different like UK or UNK2. Then choose a name for the sample. The Color does not necessarily need to be changed. The default Color in the Sample Type Setup is gray. If you would like to change it, double-click on the gray Color box. This will open up a window in which you can pick a color. Simply click anywhere in the color circle, and the New Color will be shown in the top right corner of the window. Once you have changed the color, click the OK button. The last column is Reporter. Change the None to VIC. Page 3 of 8

Repeat this for as many new sample types as you need to create. The new sample types will now appear in the Sample Type menu of the VIC dye layer and be available for you to finish labeling your plate. Changing the Quencher Setting If you are using at least one TaqMan TAMRA dye-labeled probe in the multiplex reaction, do not change the Quencher setting. If you are using TaqMan MGB probes you will need to change the Quencher setting. Since a TaqMan MGB probe has a nonfluorescent quencher, simply uncheck the Quencher box. Then click OK. Note: It is possible to run TaqMan TAMRA and TaqMan MGB probes in multiplex reactions. If you are running a TaqMan TAMRA and a TaqMan MGB probe in the same wells, leave the Quencher set to TAMRA, and analyze as usual. Page 4 of 8

Labeling the VIC Dye Layer Label the VIC dye layer using the new Sample Types that you have created. Once both layers are set up, click the Show Analysis button so that it now reads Show Setup. Page 5 of 8

Checking Instrument Settings Under the Instrument menu, at the top of your screen, go down to Diagnostics > Advanced Options Make sure that, in the Miscellaneous box, you have checked Use Spectral Compensation for Real Time. Page 6 of 8

Click OK. You will receive the following message: It is not necessary to re-launch the application. Simply click OK and start your run. Data Analysis of Singleplex and Multiplex Samples on the Same Plate When running singleplex and multiplexed samples on the same plate, you need to make some software adjustments for both sample types to be effectively analyzed. The Spectral Compensation option should only be checked when analyzing multiplex samples. The Spectral Compensation settings are unique for the analysis of two reporter dyes in one well. While the multiplex algorithm (Spectral Compensation) enhances the analysis of a multiplex reaction, the advantages are not realized in analyzing a singleplex reaction, and therefore should not be used for this. If both singleplex and multiplex samples are run on the same plate, you will need to take one of these sets out of use, post run, for analysis of the other. For example, one could first take the singleplex samples out of use and analyze the multiplex samples (using Spectral Compensation). After analysis and exporting of the results you would reverse this process and take the multiplex samples out of use and analyze the singleplex samples (without Spectral Compensation). The example below will illustrate this process. Here the gold-labeled wells are singleplex reactions. To take them out of use, highlight the desired wells, go to the Sample Type Menu, and choose Not In Use. Since the rest of the wells are multiplex, they would then be analyzed with Use Spectral Compensation for Real Time checked. You may wish to save this setup as a separate file using the Save As command. To analyze the singleplex samples you would put the singleplex reaction wells back in use, and put all multiplex samples Not in Use. Analyze the singleplex wells (with Use Spectral Compensation for Real Time unchecked). You may want to save this configuration as a separate plate for future analysis. Page 7 of 8

To analyze the results of both singleplex and multiplex samples in a secondary software, such as Microsoft Excel, you would select File > Export > Results. This will export the results (including all dye layers, Ct values, sample names, etc.) in a tab-delimited format for additional downstream analysis. 2004. Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic procedures. The PCR process and 5 nuclease process are covered by patents owned by Roche Molecular Systems, Inc. and F. Hoffmann-La Roche Ltd. Applied Biosystems and ABI PRISM are registered trademarks and AB (Design), Applera, FAM, and VIC are trademarks of Applera Corporation or its subsidiaries in the US and/or certain other countries. TaqMan is a registered trademark of Roche Molecular Systems, Inc. Page 8 of 8