Urinalysis: Comparison between Microscopic Analysis and a New Automated Microscopy Image-Based Urine Sediment Instrument

Similar documents
Cornelia Ottiger and Andreas R. Huber * Analytical Techniques. Clinical Chemistry 49: (2003) Automation and

True Walkaway Automated Urinalysis, Chemistry to Microscopy The CLINITEK AUWi System Answers for life.

Comparison between IRIS iq 200ELITE and microscopy for urinalysis and evaluation of performance in predicting outcome of urine cultures

Meets Standards AFTER REVIEW ( 2.5 pts) Fail to meet the standard safety precautions, counseled one time

Verification or validation, that is the question

Pushing limits in routine laboratory haematology with the XT-4000i

Automated Hematology Analyzer. Small Footprint. Big Difference.

CONCURRENT 10: THE COMPLETE BLOOD COUNT AND BEYOND: QUALITY ISSUES AND REFERENCE METHODS

Automated Hematology System. Your Complete Choice

Automated Hematology Analyzer. Differentiate with the XT-4000i

BEYOND A BETTER BOX DI-60 INTEGRATED SLIDE PROCESSING SYSTEM. Seamless Integration

SAMPLE. Urinalysis; Approved Guideline Third Edition

Evaluation of the Automated Immature Granulocyte Count (IG) on Sysmex XE-2100 Automated Haematology Analyser vs. Visual Microscopy (NCCLS H20-A)

BEYOND A BETTER BOX XN-V SERIES TM MULTISPECIES HEMATOLOGY ANALYZERS. Accelerate Your Multispecies Research Laboratory XN-V IS FOR ANIMAL USE ONLY

Immature Granulocyte Enumeration Our Journey from Manual to Automated Reporting

SAMPLE. Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard Second Edition

MOLECULAR TESTING: VERIFYING/VALIDATING INSTRUMENTS, REAGENTS AND ASSAYS. Richard L. Hodinka, Ph.D.

Guide to Fulfillment of Validation and Verification of Examination Requirements

ICSH guidelines for the verification and performance of automated cell counters for body fluids

Application Note for Gentian Cystatin C Immunoassay on. ABX Pentra 400 1, Horiba Medical 1 v03en-june 2011

MANAGEMENT BRIEF. How Automation is Revolutionizing White Cell Differential Analysis.

AUTOMATED HEMATOLOGY IN THE CLINICAL LAB

M-TP PRINCIPLE REF ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

Clinical Chemistry Approach to Evaluation of Commutability

S1725 Method Validation and Verification: Case Studies and Laboratory Challenges

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer

SAMPLE. Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline Third Edition

Performance Evaluation of the Coulter LH 750 Hematology Analyzer

MORE SAMPLES, LESS MICROSCOPE

Greater Possibilities By Design

SAMPLE. Risk Management Techniques to Identify and Control Laboratory Error Sources; Approved Guideline Second Edition

Ready, Set, Test! AACC Conference Mass Spectrometry in the Clinical Lab: Best Practice and Current Applications September 17-18, 2013 St.

WAIT! Ready, Set, Test! Financial Disclosure. Research/Educational grants/consulting/salary support

Optimizing an internal quality control program of automated ANA IIF

9/7/17. Implementing the ARK Immunoassay for Therapeutic Drug Monitoring of High-Dose Methotrexate. Disclosure. Learning Objectives

5 Tips for Better Cell Counting

Date. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver and/or kidneys.

BEYOND A BETTER BOX XN-SERIES TM AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Compact Automation

CLSI C60: Assay Validation & Post-Validation Monitoring

Upon completion of the Clinical Hematology rotation, the MLS student will be able to:

XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS

AAG PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

Quality Control in Flow. Dr David Westerman Head of Haematopathology Peter MacCallum Cancer Centre

XN-9100 AUTOMATED HEMATOLOGY SYSTEMS

TELE-HEMATOLOGY: CRITICAL DETERMINANTS FOR SUCCESSFUL IMPLEMENTATION

COMPARISON OF PLATELET COUNTS BY SYSMEX XE 2100 AND LH 750 WITH THE INTERNATIONAL FLOW REFERENCE METHOD IN THROMBOCYTOPENIC PATIENTS

SF Cube. Exclusive Technology, Inclusive Approach. BC-6800 Auto Hematology Analyzer. Cell Analysis Technology

Examples of body fluid analysis with the XE-5000

APB PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

Mouse Collagen Type III ELISA

Screening of urine samples by flow cytometry reduces the need for culture

Mouse OVA IgG1 ELISA. Catalog Number M046072

Molecular Diagnosis Challenges & Solutions. Using Molecular Kits or Laboratory Developed Tests (Home Brew), Emphasis on Validation

Urine Protein Electrophoresis and Immunoelectrophoresis Using Unconcentrated or Minimally Concentrated Urine Samples

BC Auto Hematology Analyzer. A CUTE 5-part

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

The Future of Hematological Diagnosis

Prepared by Date Adopted Supersedes Procedure # Adapted from HPTN Policy. Review Date Revision Date Signature

BC Auto Hematology Analyzer. A CUTE 5-part

Reshaping compact automation

METHOD VALIDATION TECHNIQUES PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER 2012

Reshaping compact automation

Detection of Significant Bacteriuria. by Use of the iq200 Automated Urine Microscope

SAMPLE. Evaluation of Total Analytical Error for Quantitative Medical Laboratory Measurement Procedures

Novel POC analysis for determination of total and 5-part differential WBC count among a US population, in comparison to Beckman Coulter LH750

Cellometer. for Cell Counting & Analysis. Brewing Yeast Wine Yeast Platelets and Other Small Cells. Image Cytometer

BIOEQUIVALENCE TRIAL INFORMATION FORM (Medicines and Allied Substances Act [No. 3] of 2013 Part V Section 39)

ABX Pentra XL 80. Hematology analyzer. Auto-sampling 5-part Differential Customized Dilution Ratio Integrated Validation Station

Designing QC Activities to More Precisely Manage Analytical Accuracy and Patient Risk

Improving the efficiency and efficacy of pre-analytical and analytical work-flow of urine cultures with urinary flow cytometry

Test Method Validation of In-Vitro Diagnostic Products. Tim Carr QA Manager, Beckman Coulter

School of Allied Medical Sciences. Course Description Guide Associate in Medical Laboratory Sciences. Page 1 of 15

ABX Pentra DX120. Hematology analyzer. 45 parameters Integrated slide-maker Expert validation system

Image Analysis. Tools for basic and translational research. visualize analyze interpret report archive customize

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer

Starrsed RS. Gold Standard ESR

Highly Accurate, Reliable and Real-time Enumeration of Individual Bacteria for the Determination of Raw milk Hygienic Quality

GOOD LABORATORY PRACTICE Agilent Cary 8454 UV-Visible Spectroscopy System

ACCURACY THE FIRST TIME. Bring Your Esoteric Assays In-House Routinely SCIEX DIAGNOSTICS MASS SPECTROMETERS

Carl Zeiss MicroImaging AIS. Digital Pathology The Future. Andrew Lesniak Director, Product Management

XN Series. What s in it for me? Krista Curcio, MT(ASCP), MBA Product Manager, Hematology Systems Sysmex America, Inc. All rights reserved.

LabGuide 13 How to Verify Performance Specifications

Date. Alkaline phosphatase measurements are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.

PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

An Automated Method for the Determination of Intestinal Disaccharidase and Glucoamylase Activities

Measure-iT Thiol Assay Kit (M30550)

Total Analytic Error From Concept to Application

Cellometer Auto 2000 Cell Viability Counter Optimized Analysis of Primary Cells Cellometer Features of the Cellometer Auto 2000

CR-E PRINCIPLE REF A ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

BEYOND A BETTER BOX XN-9000 AUTOMATED HEMATOLOGY SYSTEMS. Reshaping Scalable Automation

Successful performance in proficiency testing (PT) is required

The utility of qnmr to improve accuracy and precision of LC-MS bioanalysis

HST-330 TM and Total Laboratory Automation

YOU Can Make a Difference!

Automation In Life Sciences and Diagnostics Applications

WADA Technical Document TD2019BAR. Blood Analytical Requirements for the Athlete Biological Passport

Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals.

The Journey Towards Standardization in Validation of Fluorescent Cell-Based Assays

Critical appraisal checklist for papers on biological variation

Fibrinogen Universal Kit

Transcription:

Clin. Lab. 2014;60:693-697 Copyright SHORT COMMUNICATION Urinalysis: Comparison between Microscopic Analysis and a New Automated Microscopy Image-Based Urine Sediment Instrument PAULA V. BOTTINI, MAYARA H. M. MARTINEZ, CÉLIA R. GARLIPP Division of Clinical Pathology, University of Campinas/UNICAMP, Campinas, SP, Brazil SUMMARY Background: Urinalysis is a high demand procedure, with a large amount of manual labor and poorly standardized. Recently a new walk-away automated urine analyzer has been introduced. The aim of this study was to evaluate the performance of UriSed as an alternative to the microscopic analysis of urine samples. Methods: Four hundred mid-stream urine samples from patients with several clinical conditions were analyzed by bright field microscopy and by UriSed. The validation protocol included studies of precision, carryover, and comparison with the gold standard microscopy. Results: Our data showed that UriSed is a precise method. Repeatability and reproducibility ranged from 8 to 15%. Carryover was negligible. All the elements showed good agreement between both methods, with an extremely high correlation between the erythrocyte and leukocyte counts (r > 0.95) Conclusions: UriSed is a precise and accurate alternative to microscopy that allows a better workflow and may significantly improve turnaround time. (Clin. Lab. 2014;60:693-697. DOI: 10.7754/Clin.Lab.2013.130725) KEY WORDS urinalysis, microscopic examination, automated sediment analysis INTRODUCTION Urinalysis is a high demand procedure which includes physical, chemical, and microscopic examinations. Although the analysis of urine sediment provides essential information on the functional state of the kidneys, it is a labor intensive and time consuming procedure. Besides that, it is poorly standardized and features a wide interobserver variability as several factors may contribute to the inaccuracy of this method (e.g., centrifugation, differences in interpretation of cellular elements, cylinders, etc.) [1,2]. Automation seems to be the answer to this issue [3]. In an attempt to automate the microscopic analysis of urine, some urine analysis systems have been developed beginning in the late 80s. The great advantage of automated analysis over microscopic procedure is that the former reduces the inter-observer variability and increases the accuracy and the productivity. While microscopic analysis usually takes several minutes to be performed automation can release a result in about one minute [4]. A few years ago, a new walk-away automated urine sediment analyzer based on the KOVA method with on-screen review of the images was introduced. This analyzer is basically the automation of the traditional manual microscopy [5]. The aim of this study was to evaluate the performance of this image based automated sediment analyzer (UriSed, also called sedimax in some countries) as an alternative to the microscopic analysis of urine samples. MATERIALS AND METHODS We analyzed 400 mid-stream urine samples from patients (262 women and 138 men) with several clinical conditions. Their ages varied from 2 to 91 years (median = 41 years). All samples were analyzed by bright field microscopy Short Communication accepted July 24, 2013 Clin. Lab. 4/2014 693

P. V. BOTTINI et al. Table 1. Precision of UriSed. Intra-assay precision Inter-assay precision LEVEL 1 LEVEL 2 LEVEL 1 LEVEL 2 element/hpf mean ± SD CV mean ± SD CV mean ± SD CV mean ± SD CV RBC 25.7 ± 2.0 8% 83.6 ± 7.2 9% 25.5 ± 2.0 8% 95.8 ± 7.7 8% WBC 11.7 ± 1.2 10% 64.7 ± 6.9 11% 16.1 ± 2.4 15% 79. 9 ± 10.6 13% RBC - red blood cell, WBC - white blood cell hpf - high power field, SD - standard deviation, CV - coefficient of variation. Table 2. Diagnostic performance of UriSed. RBC WBC Hyaline Casts Path Casts Crystals Bacteria Sensitivity (%) 75 72. 52 54 100.0 69.4 Specificity (%) 99 98 69 82 100.0 85.0 Positive Predictive value (%) 94 94 17 27 100.0 68.8 Negative Predictive value (%) 96 92 92 94 100.0 86.2 Agreement (%) 95.0 91 67 79 100.0 80.1 2 269.812 242.585 7.195 26.686 397.000 120.247 p < 0.001 < 0.001 < 0.05 < 0.001 < 0.001 < 0.001 RBC - red blood cell, WBC - white blood cell, Path - pathological. using the KOVA method with the results expressed in terms of quantitative elements (erythrocytes-rbc and leukocytes-wbc; arithmetic mean of 10 fields of 400x) and qualitative elements (casts, crystals, and bacteria, considering all fields), and by a fully automated sediment analyzer (UriSed - 77 Elektronika Kft, Budapest, Hungary). UriSed is based on microscopic examination of urine samples (KOVA method). The analyzer homogenizes, aspires, and transfers the sample to a special disposable cuvette. The cuvette is centrifuged and placed in microscopy position and whole-field images are obtained and evaluated by an auto image evaluation module (AIEM). Results were expressed as particle concentration/hpf (high power field). These high definition images are displayed on-screen and can be stored for further review and for educational purposes [5]. All 15 images of each sample were reviewed by an experienced analyst and manually edited, if necessary. The validation protocol was based on local regulation, on CLSI documents EP05-A2 [6] and EP15-A2 [7], and on ICSH guideline [8] and included studies of precision (repeatability/reproducibility) and carryover. Repeatability (intra assay precision) was obtained from the analysis of 20 replicates of two urine samples with different particle numbers. The inter assay precision (reproducibility) was performed by analyzing two levels of control samples (Liquichek Urinalysis Control - Biorad) on 20 consecutive days. Carryover effect was assessed by testing a high-level urine sample consecutively in triplicate (H1, H2 and H3) followed by a low-level sample (L1, L2 and L3). Carryover was calculated using the following formula: carryover (%) = [(L1 - L3)/(H1 - L3)] x 100. Accuracy of the automated method was evaluated by comparison with microscopic analysis. Simple linear regression (least square method) was used to evaluate the correlation between quantitative analysis (RBC and WBC) by both methods. The Chi-squared statistical method was used to determine the agreement between microscopic analysis and UriSed. RBC and WBC counts > 5/hpf were classified as abnormal. Qualitative elements were compared according to the detection of them by each method. RESULTS Intra assay precision (repeatability) and inter assay precision (reproducibility) ranged from 8 to 15%, depending on the considered parameter (Table 1). As expected, there was no carryover since UriSed uses single-use cuvette for each sample. The results of high to low carryover returned negative values: -3.2% (RBC), -1.8% (WBC) and -1.2% (bacteria). There was a good correlation between the erythrocyte 694 Clin. Lab. 4/2014

URINALYSIS: MICROSCOPIC AND AUTOMATED ANALYSIS Figure 1. Distribution of RBC and WBC counts. and leukocyte counts by both methods (r > 0.95). Figure 1 shows the distribution of RBC and WBC counts by manual and automated analysis. All the elements showed good agreement between microscopic analysis and UriSed, as seen in Table 2. The analyzer easily detected the most common crystals seen in urine samples. The major causes of misclassification are the presence of squamous epithelial cells, mucus or amorphous crystals, emphasizing the importance of a recently collected mid-stream urine sample. DISCUSSION There is no doubt that microscopy is an important part of urinalysis and may provide useful information. As it is a labor intensive and time consuming procedure many efforts have been made to develop a better, cost-effective strategy without loss in the diagnostic yield. The first attempt in reducing the number of urine microscopy analyses was based on urine dipsticks. This procedure led to conflicting results as studies showed that a significant percentage of urine samples with normal chemistry results on the dipsticks had abnormal findings on microscopic analysis of sediment [9-11]. Studies have demonstrated significant inter-observer variation in the identification and interpretation of urine structures [1,12,13]. Automated urine microscopy is able to classify, quantify, and report urine particles and it seems to be an efficient alternative to reduce the manual workload and improve inter-observer reliability. Microscopic examination of urine sediment involves classification and quantification of particles present in this fluid and automated devices were developed to facilitate and standardize this procedure. This goal can be achieved through three systems with different methodologies. One of them (Sysmex UF-100/UF-1000i) is based on the use of fluorescence flow cytometry to classify the elements present in urine. Results are displayed as scattergrams and numerical values are reported [2]. Samples whose results have been flagged need to be reviewed by microscopy. Another analytical system is based on flow cell digital imaging with automatic particle recognition software (Iris iq 200) to identify and classify isolated images based on the texture, contrast, size, and shape of each element. The images are displayed to the analyst, allowing edition and reclassification, if necessary [14]. These images are considerably different from those observed in microscopy requiring substantial training before using the equipment [15]. The third analyzer (UriSed) is based on microscopic urine sediment analysis with digital images and automatic recognition of particles. UriSed corresponds to the automation of manual microscopy. Due to its characteristics, UriSed eliminates the need of microscopic Clin. Lab. 4/2014 695

P. V. BOTTINI et al. review of most of the samples. Images are very similar to those observed by microscopy and can be easily reviewed by the analyst. It is noteworthy that automated urinalysis, enables better workflow and may significantly improve turnaround time, since its processing time is much faster than the manual method (80-100 samples/hour versus around 10 samples hour/technologist, respectively). Our data showed that UriSed is a precise method. Interand intra-assay precision ranged from 8% to 15%. This rate is much lower than that observed in manual microscopy where several studies showed a variation of more than 50% depending on the particle counts [14-16]. Carryover was negligible, as described by Zaman et al. [17]. All the elements showed good agreement between microscopic and automated analysis. Our data were in accordance with those obtained by Akin et al. [14] and Zaman et al. [17]. Other authors, using automated urine analyzers based on flow cytometry (UF-100/UF-1000i) or digital image with particle recognition (iq 200) also observed good correlation with the gold standard microscopy [16,18,19]. Akin et al. compared the analytical performance of UriSed with another automated analyzer (iq 200) and observed a significant correlation between them [14]. Appropriate preanalytical procedures are the basic condition for obtaining an adequate sample that allows reliable results and a proper clinical interpretation. As stated before, one of the major causes of misclassification was the presence of contaminating elements such as squamous epithelial cells and mucus. Collecting a midstream sample is the most relevant step to prevent contamination. A recent multicenter study conducted by Manoni et al. [20] clearly confirmed mid-stream urine as the most suitable sample for routine urinalysis. In summary, UriSed is a precise and accurate alternative to the gold standard microscopy. The implementation of this automated routine allows a better workflow and may significantly improve turnaround time. Besides that, it has the potential to eliminate the need of a microscopic review of most of the samples as UriSed permits an on-screen review of the images. It is noteworthy that the use of a system that photographs and stores high-definition images provides an invaluable training tool for technologists and medical students. Acknowledgement: We thank José Ricardo Lauand, Luciane Franco-Fernandes, and Solange Gomes Lara Cioffi for helping us during the validation process. Declaration of Interest: The authors declare that there are no conflicts of interest. References: 1. Winkel P, Statland BE, Jorgenson J. Urine microscopy: an ill-defined method examined by a multifactorial technique. Clin Chem 1974;20:436-9. 2. Delanghe JR, Kouri TT, Huber AR, et al. The role of automated urine particle flow cytometry in clinical practice Clinica Chimica Acta 2000;301:1-18. 3. Langlois MR, Delanghe JR, Steyaert SR, Everaert KC, De Buyzere ML. Automated flow cytometry compared with an automated dipstick reader for urinalysis. Clinical Chemistry 1999;45: 1 118-22. 4. Block DR, Lieske JC. Automated urinalysis in the clinical lab. Medical Laboratory Observer 2012;44(10):8-10. 5. Barta Z, Kránicz Tünde, Bayer G. UriSed technology - A standardised automatic method of urine sediment analysis. European Infectious Disease 2011;5(2):139-42. 6. Clinical and Laboratory Standards Institute. Evaluation of precision performance of quantitative measurement methods. CLSI document EP05-A2. Wayne, PA, USA: CLSI; 2004. 7. Clinical and Laboratory Standards Institute. User verification of performance for precision and trueness; approved guideline - second edition. CLSI document EP15-A2. Wayne, PA, USA: CLSI; 2005. 8. International Council for Standardization in Haematology. Guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting and cell marker applications. Clin Lab Haemat 1994;16:157-74. 9. Szwed JJ, Schaust C. The Importance of microscopic examination of the urinary sediment. American Journal of Medical Technology 1982;48(2):141-3. 10. Shaw ST Jr, Poon SY, Wong ET. Routine urinalysis : is the dipstick enough? JAMA 1985;253:1596-600. 11. Resnick M. Comment on simplifying urinalysis. Clin Chem 1985;31:450-1. 12. Wald R, Bell CM, Nisenbaum Ret al. Interobserver reliability of urine sediment interpretation. Clin J Am Soc Nephrol 2009;4: 567-71. 13. Tsai JJ, Yeun JY, Kumar VA, Don BR. Comparison and interpretation of urinalysis performed by a nephrologist versus a hospitalbased clinical laboratory. American Journal of Kidney Diseases 2005;46(5):820-9. 14. Akin OK, Serdat MA, Cizmeci Z, et al. Comparison of LabUMatwith-UriSed and iq 200 fully automatic urine sediment analysers with manual urine analysis. Biotechnol Appl Biochem 2009:53:139-44. 15. Block DR, Lieske JC. Automated urinalysis in the clinical lab. MLO Med Lab Obs 2012;44(10):8-10,12;quiz 14. 16. Mayo S, Acevedo D, Quiñones-Torrelo C, Canós I, Sancho M. Clinical laboratory automated urinalysis: comparison among automated microscopy, flow cytometry, two test strips analyzers, and manual microscopic examination of the urine sediments. J Clin Lab Anal 2008;22(4):262-70. 17. Zaman Z, Fogazzi GB; Garigali G et al. Urine sediment analysis: Analytical and diagnostic performance of sedimax - A new automated microscopy image-based urine sediment analyser. Clinica Chimica Acta 2010;411:147-54. 696 Clin. Lab. 4/2014

URINALYSIS: MICROSCOPIC AND AUTOMATED ANALYSIS 18. Jiang T, Chen P, Ouyang J, Zhang S, Cai D. Urine particles analysis: performance evaluation of Sysmex UF-1000i and comparison among urine flow cytometer, dipstick, and visual microscopic examination. Scand J Clin Lab Invest 2011;71(1):30-7. 19. Manoni F, Tinello A, Fornasiero L, et al. Urine particle evaluation: a comparison between the UF-1000i and quantitative microscopy. Clin Chem Lab Med 2010;48(8):1107-11. 20. Manoni F, Gessoni G, Alessio MG, et al. Mid-stream vs. firstvoided urine collection by using automated analyzers for particle examination in healthy subjects: an Italian multicenter study. Clin Chem Lab Med 2011;50(4):679-84. Correspondence: Paula V. Bottini MD, PhD Divisão de Patologia Clínica/HC - UNICAMP Rua Vital Brasil 251- Cidade Universitária Zeferino Vaz CEP: 13083-888, Campinas, S.P., Brasil Tel.: + 55 19 3521-7539 Fax: + 55 19 3521-7510 Email: paula@hc.unicamp Clin. Lab. 4/2014 697