Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

Similar documents
EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo

High matrix tolerance of the Thermo Scientific icap 7000 Plus Series ICP-OES with the radial Ceramic D-Torch and the sheath gas adaptor

Analysis of wear metals and contaminants in engine oils using the 4100 MP-AES

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Trace Elements in Petrochemical Samples: Achieving Excellent Analytical Results

EU water analysis using the Thermo Scientific icap 7400 ICP-OES Duo

Analytically Monitoring the Effect of Fracturing Activity

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

Table I: MCL and MRL Concentrations for Contaminants Monitored Under the Safe Drinking Water Act National Primary Drinking Water Regulations

Analysis of steel and its alloys using the GB/T standard and an Agilent 5100 ICP-OES in dual view mode

Functionality. Normal Operation

Multi-Element Analysis of Trace Metals in Animal Feed using ICP-OES

Simple, fast and reliable analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

Ultra-fast ICP-OES determinations of base metals in geochemical samples using next generation sample introduction technology

Multi-element determination in pharmaceutical preparations using the Thermo Scientific icap Q ICP-MS

Analysis of Mineral and Heavy Metal Content in Beverages Using the Teledyne Leeman Labs Prodigy Plus ICP-OES

icap 7000 Plus Series ICP-OES Streamlined performance and ultra-low ICP-OES detection limits for routine laboratories

Increased Sample Throughput for ICP-OES Applied to U.S. EPA Method 200.7

Analysis of elemental impurities in drug products using the Thermo Scientific icap 7600 ICP-OES Duo

Determination of Elemental Impurities in Graphite-based Anodes using the Agilent 5110 ICP-OES

Increase productivity for environmental sample analysis using the SVS 2+ Switching Valve System for Agilent 5100 SVDV ICP-OES

The Determination of Alloying Elements in High-Carbon Steel using Radially Viewed Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES)

Automating EPA 6020 Compliant Analysis with the Agilent 7900 ICP-MS and ESI prepfast Autodilution System

Analysis of inorganic impurities in lubricating oils by ICP-MS

DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, & USING THE AGILENT 5100 ICP-OES

Correlation coefficient

Benefits of running organic matrices using the Agilent 5100 ICP-OES fast, robust, high performance analysis

Analysis of elemental impurities in drug products using the Thermo Scientific icap 7400 ICP-OES Duo

ESI SC-FAST Sample Introduction Solutions for PerkinElmer ICP-OES and ICP-MS Systems

Analysis of Heavy Metals in Iron-based Fertilizers by HR ICP-OES

Thermo Scientific icap 7000 Series ICP-OES Accessories Guide

Composition Analysis of Animal Feed by HR ICP-OES

Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

Analysis of Environmental Samples with the Agilent 730-ES Following US EPA Guidelines

The Agilent Atomic Spectroscopy portfolio for Environmental applications AA, MP-AES, ICP-OES, ICP-MS & ICP-QQQ

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance

Radial, Axial or Dual View ICP: Which Do You Choose? Manny Almeida Teledyne Leeman Labs, Inc. Hudson, NH

Atomic Spectroscopy Solutions for Environmental Applications

The Determination of Trace Elements in Waters using the US EPA Method and the Thermo Scientific icap 6500 ICP-OES (Duo)

Thermo Scientific icap 6000 Series ICP Emission Spectrometer

Application Note. Abstract. Authors. Environmental

Determination of 22 Elements Following US EPA Guidelines with a New Megapixel CCD ICP-OES

Analysis of Trace Metals in Surface and Bottled Water with the Optima 7300 DV ICP-OES

prepfast Autodilution Autocalibration Auto QC Dilution Elemental Scientific

Analysis of Toxic Trace Elements in Coffee Bean Products by HR ICP-OES

Amplifying ICPMS Productivity Using Discrete Sampling Technology

Thermo Scientific Qtegra Intelligent Scientific Data Solution. Delivering quality. Driving productivity

Achieving Optimum Throughput in ICP-MS Analysis of Environmental Samples with the Agilent 7500ce ICP-MS Application

Direct Analysis of Photoresist by ICP-MS. Featuring the Agilent Technologies 7500s ICP-MS

Analysis of Samples Related to Nuclear Submarine Operation

Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

High performance radio frequency generator technology for the Thermo Scientific icap 7000 Plus Series ICP-OES

Water Analysis Using ICP-OES with an Ultrasonic Nebulizer

Tender Document. National Centre of Excellence in Analytical. Chemistry, University of Sindh, Jamshoro. Inductive Coupled Plasma (ICP -OES) ISSUED BY:

Thermo Scientific EASY-nLC 1200 System. Leading in simplicity. and performance

for water and beverage analysis

Nitrogen Determination in Soils and Plants by Flash Combustion using Argon as Carrier Gas

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

Jonathan R. Beck and Charles T. Yang; Thermo Fisher Scientific, San Jose, CA

Common Challenges facing Laboratories doing Elemental Analysis Today

Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry

Avio 200 ICP-OES Flat Plate

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

A Novel Screening Method for Anthropogenic Sewage Pollutants in Waste Water, Ground Water and Drinking Water Samples by LC HRAM Analysis

Determination of 1,4-Dioxane in Drinking Water by Gas Chromatography/Mass Spectrometry (GC/MS) with Selected Ion Monitoring (SIM)

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

Accuracy and Precision. Dilution Error in Error in Factor Accuracy Precision 200 <2% <2%

Agilent Atomic Spectroscopy Solutions for Energy & Chemicals AA, MP-AES, ICP-OES, ICP-MS.

Agilent 5100 ICP-OES DUAL VIEW ICP-OES MINUS THE WAIT

Analysis of 14 Trace Elements in Stainless Steel

PROCUREMENT PROCEDURE OF CPRI (NON WORKS) Dt of Revision : Issue Dt. :

The Benefits of SOLAμ Technology in Sample Preparation. Jon Bardsley and Ken Meadows Thermo Fisher Scientific, Runcorn, UK

Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass

Guidelines for Trouble Shooting and Maintenance of ICP-OES Systems. Eric Vanclay Product Manager Spectroscopy Consumables May 2, 2012

Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Application Note PlasmaQuant PQ 9000 Elite. Analysis of Rare Earth Elements in Granite and Sandstone by HR ICP-OES. Challenge.

Inductively coupled plasma (ICP) spectroscopy for online measurements of trace metal emissions at the 250 kw PACT furnace

Accurate. Robust. Reliable. AGILENT 710 SERIES ICP-OES

Accurate. Robust. Reliable. AGILENT 710 SERIES ICP-OES

Product Descriptions and Offerings

ICP Expert software. Technical Overview. Introduction

Thermo Scientific Solutions for Intact-Protein Analysis. Better, Faster Decisions for. Biotherapeutic Development

METHOD 3040A DISSOLUTION PROCEDURE FOR OILS, GREASES, OR WAXES

ICP-OES. Varian Vista-MPX CCD Simultaneous ICP-OES

The Determination of Trace Elements in Stainless Steel by Forked Platform GFAAS

EXPAND YOUR RANGE EXTEND YOUR RESOURCES. Avio 500 ICP Optical Emission Spectrometer

ICP-AES Determination of 15 Kind of impurity elements in the Vanadium-aluminum alloy

VHG Labs. Standards and Supplies for Spectrochemical Analysis QUALITY. TRACEABILITY. DEPENDABILITY. WORLDWIDE

Dr. Jianfeng Cui, Thermo Fisher Scientific, Bremen, Germany APPLICATION NOTE AN43384

Analysis of Fracking Flowback Water From the Marcellus Shale Using In-line Conductivity, Automated Dilution, and Ion Chromatography

Rubber and plastics gloves for food services Limits for extractable substances

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water. Pranathi R Perati Thermo Fisher Scientific, Sunnyvale, CA, USA

Elemental Scientific. prepfast. The Next Leap in Automation. Autodilution Autocalibration Intelligent QC Dilution

Transcription:

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES Marine Beauvir, Application Specialist, Thermo Fisher Scientific, Cambridge, UK Application Note 43161 Key Words High throughput, Lubricating oil, Sprint valve, Used oil, Wear metals Goal This application note describes how the Thermo Scientific icap 7600 ICP-OES Radial and its integrated sampling valve offer high throughput capabilities to laboratories analysing lubricating oil samples. The instrument combines fast analysis time with excellent analytical performance for superior productivity and reduced cost of analysis. Introduction The analysis of lubricating oils is a powerful tool in preventative maintenance of engines and machinery. Regular oil sampling and trend analysis will give precious information about the state of a motor, gear transmission or mechanical part, and signify the need for maintenance before critical failures. In particular, elemental analysis by Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES) is used to determine the concentration of wear metals, contaminants and additives present in used oils that can be sampled from car to train fleets, or even large construction or mining machines. Wear metals are elements such as Fe, Cu and Ni, and their presence may indicate wear of metallic parts. Other elements may give evidence of contamination from foreign matter, for example Si and dirt entering the engine via a faulty filter. Additives which are typically Ca, P or Zn based compounds are added artificially to the oil to improve lubricating properties. Monitoring the depletion of these elements may therefore help in identifying optimum conditions and maintenance scheduling. Preventative maintenance means less downtime on the equipment thereby reducing the need for expensive repairs. The result is that specialist laboratories performing this type of analysis face the challenges of both a high volume of samples and fast turnaround requirements. Lubricating oil analysis by ICP-OES is an established method with a typical analysis time for the determination of more than twenty elements of around one to two minutes per sample. This work describes how coupling an innovative sample introduction system to powerful instrumentation enables analysis time to be reduced significantly while retaining the analytical requirement of the industry. In turn, this allows rapid decisions to be taken and imminent mechanical failure to be identified.

2 Instrumentation The Thermo Scientific icap 7600 ICP-OES Radial was chosen for the analysis. The instrument configuration combines the enhanced matrix tolerance of the dedicated radial plasma view with high throughput capabilities of the integrated Sprint valve. The sampling valve helps to reduce flush and rinse time during analysis, thereby decreasing total analysis time significantly. It uses a vacuum pump, the traditional ICP-OES peristaltic pump and a 6-port valve fitted with a sample loop. When the valve is in the load position, the vacuum pump fills the sample loop in a few seconds while the rinse solution is pumped through the nebulizer by the peristaltic pump. When the position switches to inject, the sample is pushed into the nebulizer for analysis via the carrier/rinse solution by the peristaltic pump. At the same time, the autosampler probe is moved to the rinse station and flushed with rinse solution using the vacuum pump. This system enables a reduction of sample uptake times to mere seconds. The icap 7600 ICP-OES is compatible with the CETAC ASX-1400 stirring autosampler which ensures homogeneity of the solutions analyzed. The use of the CETAC APS-1650 Automated Prep Station is also recommended for this type of analysis as it reduces sample preparation time, typically the limiting step of high throughput analysis. In this case, dilutions are made on a volume to volume basis. Method Reagents The following standards were used in this work: Element Blank Oil 75, Conostan (SCP SCIENCE, Baie-D Urfé, Canada); 21-Element Oil Standard 500 mg/kg containing Ag, Al, B, Ba, Ca, Cd, Cr, Cu, Fe, Mg, Mn, Mo, Na, Ni, P, Pb, Si, Sn, Ti, V and Zn, SPEX CertiPrep (SPEX CertiPrep Group, Metuchen, US); a higher-concentration standard 5000 mg/kg for the additive elements (Ba, Ca, Mg, P, Zn), AM-Special custom blend, Conostan; oil-based standard 5000 mg/kg S and oil-based standard 5000 mg/kg Y, Conostan. General purpose white spirit was used as a solvent. Sample and standard preparation For all elements except S, standards at 50, 100, 250 and 500 mg/kg were prepared, plus standards at 1000, 2500 and 5000 mg/kg for Ba, Ca, Mg, P and Zn. Separate standards were also prepared at 2500 and 5000 mg/kg for S. Y was used as an internal standard and added to the white spirit to obtain a 20 mg/l Y solution used for all further dilutions. All samples and standards were therefore diluted 1:10 (w/v) with the later. ICP-OES was used for this analysis in conjunction with the Sprint analysis mode to enable extremely fast analysis times. The standard organic sample introduction kit was also fitted to the instrument. It comprises a V-groove nebulizer, a baffled spray chamber and a 1 mm centre tube. All parameters used for the analysis including settings of the Sprint valve are listed in Table 1 and Figure 1. When using the sampling valve, flush and rinse behaviors are driven by the valve settings. There is no need therefore for faster flush pump speed or wash time to be set as in traditional methods. The flush pump rate is identical to the analysis pump speed (no pump stabilization time required) and the wash time in the method is set to zero. Table 1. Parameters used for analysis Instrument parameters Inlet pump tubing Outlet pump tubing Loop size Pump speed Uptake time Nebulizer Spray chamber Centre tube Torch Plasma parameters RF power Nebulizer gas flow Auxiliary gas flow Coolant gas flow Radial viewing height Analysis parameters Analysis mode Exposure time Solvent Flex (1.016 mm) Solvent Flex (1.524 mm) 1.02 ml 40 rpm 12 sec V-groove Baffled cyclonic 1 mm EMT 1350 W 0.4 L/min 1.5 L/min 14 L/min 12 mm Sprint 1 sec Repeats 2 By ensuring that the final solution always contains 10% oil (weight or volume), differences in viscosity are minimized; therefore, when required, base oil is added to the standard or sample prior to the addition of the diluent. Furthermore, oil analysis is usually expressed directly in ppm or mg/kg in the oil sample, the solvent dilution being negligible. All concentrations in this work are therefore expressed in sample terms (including calibration standards). Instrument and method parameters The Sprint valve of the Thermo Scientific icap 7600 Figure 1. Sprint valve settings

Method development Using the intuitive wavelength selection tool of the Thermo Scientific Qtegra Intelligent Scientific Data Solution (ISDS), wavelengths were selected that were most likely to be free from interferences in this matrix (Figure 2). Table 2. Wavelengths, calibration standards and detection limits (DL) of the optimised method Wavelength Calibration standards (mg/kg) DL (mg/kg)* Ag 328.068 nm 0, 100, 250 0.5 Al 309.271 nm 0, 100, 250, 500 2 Comments No right background point (V) 3 B 208.959 nm 0, 100, 250, 500 2 Ba 233.527 nm 0, 100, 250, 500 1 Ba 234.758 nm 0, 250, 500, 2500, 5000 5 Ca 220.861 nm 0, 250, 500, 2500, 5000 50 Estimated linearity > 15000 ppm Cd 214.438 nm 0, 100, 250 0.2 Cr 205.560 nm 0, 100, 250, 500 0.5 Cu 224.700 nm 0, 100, 250, 500 1 Fe 261.187 nm 0, 100, 250, 500 1 Mg 277.669 nm 0, 250, 500, 2500, 5000 25 Figure 2. Intuitive wavelength selection of Qtegra ISDS Typically, concentrations ranging from single ppm level for wear metals, up to thousands of ppm for additive elements are expected in lubricating oils. Wavelengths were therefore also chosen following these criteria to achieve the required detection limits for wear metals and linearity for additives. Selected wavelengths are listed in Table 2. Ba and Mg are considered as additive elements, but they are not used as widely as Ca, P and Zn based compounds, hence their concentration can vary significantly in oils. A combination of two wavelengths for these elements was kept with switching from Mg 279.079 nm to Mg 277.669 nm and Ba 233.527 nm to Ba 234.758 nm above 500 mg/kg to extend linearity of the method if required. A similar approach could be taken for Ca and Zn. In this work, linear calibrations (see details in Table 2) were obtained with coefficients of correlation better than 0.9995 for all elements. Y 224.306 nm and 242.220 nm were chosen for internal standard correction. Mg 279.079 nm 0, 100, 250, 500 5 Mn 293.306 nm 0, 100, 250 0.5 Mo 281.615 nm 0, 100, 250, 500 0.5 Na 589.592 nm 0, 100, 250 0.5 Ni 230.300 nm 0, 100, 250, 500 0.5 P 185.942 nm 0, 250, 500, 2500, 5000 5 Pb 220.353 nm 0, 100, 250, 500 2 S 182.624 nm 0, 2500, 5000 50 Si 251.611 nm 0, 100, 250, 500 1 Sn 283.999 nm 0, 100, 250, 500 5 Ti 339.458 nm 0, 100, 250, 500 1 V 268.796 nm 0, 100, 250, 500 1 Zn 330.259 nm 0, 250, 500, 2500, 5000 10 *DL estimated as three times the standard deviation (SD) calculated over 10 blanks Analysis and results Two typical oil samples (Oil A and Oil B) were analyzed following the Sprint method described previously. The results are shown in Table 3. They were compared to the concentrations obtained for the same oils analyzed with a traditional method. The Speed analysis was performed using the instrument peristaltic pump in a conventional way (no Sprint valve). Speed analysis mode was selected with five seconds integration time and two replicates. Analytical wavelengths were optimized for the method and may be different than defined for the Sprint analysis. Estimated linearity > 15000 ppm No right background point (Zn)

4 Table 3. Results (mg/kg) obtained for Oil A and Oil B with Sprint and Speed methods Element Oil A Oil B Sprint Speed Sprint Speed Ag < 0.5 < 0.1 < 0.5 < 0.1 Al 2.1 1.9 8.9 9.6 B < 2 0.8 < 2 1.3 Ba < 1 0.2 3.0 3.2 Ca 2740 2830 17330 17750 Cd < 0.2 < 0.05 < 0.2 < 0.05 Cr 0.9 0.8 0.8 0.7 Cu 1.3 1.6 1.7 1.6 Fe 20.6 20.6 30.6 29.3 Mg 248 246 41 41 Mn 0.6 0.6 5.0 5.3 Mo 4.3 4.4 1.1 0.8 Na 7.0 7.0 87 84 Ni < 0.5 0.3 67 70 P 1040 1060 360 369 Pb < 2 1.0 < 2 < 0.5 S 8180 7670 15010 14030 Si 4.2 4.3 20.0 18.9 Sn < 5 < 1 < 5 < 1 Ti < 1 < 0.2 1.5 0.6 V < 1 < 0.2 62 66 Zn 1220 1240 419 419 Comparison within the two methods was found to be good with relative standard deviation (RSD) below 5% for elements present at significant concentrations in Oil A and Oil B. Although detection limits for this Speed method are estimated to be five times lower then for the Sprint method, this is below the industry requirement which is typically around the single ppm level. The Sprint analysis was noticeably faster with 27 seconds analysis time per sample compared to approximately three times longer for samples analyzed with the Speed method. Accuracy of the Sprint method is also shown in Table 4 with recoveries obtained for a check standard prepared at 50 mg/kg (all elements except additives and S). Recoveries were good with bias below 5% for all elements except B (within 10%). Table 4. Results obtained for a 50 mg/kg check standard Element Measured conc. (mg/kg) Recovery (%) Ag 49.9 99.8 Al 50.3 100.6 B 46.5 93.1 Ba 52.4 104.8 Cd 49.4 98.8 Cr 49.5 99.0 Cu 49.2 98.3 Fe 50.5 101.0 Mg 49.2 98.5 Mn 49.6 99.2 Mo 50.3 100.6 Na 50.1 100.1 Ni 49.2 98.5 Pb 50.0 100.0 Si 50.1 100.2 Sn 52.3 104.5 Ti 50.7 101.5 V 50.2 100.5

Excellent stability of the Sprint analysis was demonstrated by measuring check standards every half an hour over a four hour period (100 mg/kg for wear metals and 1000 mg/kg for additives). Recoveries are shown in Figure 3. Recoveries for all elements over the four hour analysis were well within 10% of the expected concentrations with most elements within 5%. Overall RSD on the measurements were also found to be good and below 2% for all wavelengths. No re-calibration was required although this could be automatically set-up within Qtegra ISDS, as a comprehensive choice of actions related to QC or check standards is available for selection from the software. Application Note 43161 Figure 3. Stability over a four hour period for wear metals and additives Conclusion The Thermo Scientific icap 7600 ICP-OES Radial was successfully used for high throughput lubricating oil analysis. The intelligent design of the sample introduction system with an integrated sampling valve allows analysis time per sample below 30 seconds. Analytical performance was also demonstrated giving accuracy, precision and stability for hundreds of samples, reducing the number of QC, re-calibrations and samples to be re-analyzed, but also reducing costs of analysis per sample. The icap 7600 ICP-OES instrument offers superior capacity of analysis for laboratories seeking optimum productivity. thermofisher.com 2016 Thermo Fisher Scientific Inc. All rights reserved. CONOSTAN : A Division of SCP SCIENCE. SPEX CertiPrep from the SPEX CertiPrep Group LLC. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa-Other +27 11 570 1840 Australia +61 3 9757 4300 Austria +43 1 333 50 34 0 Belgium +32 53 73 42 41 Canada +1 800 530 8447 China +86 10 8419 3588 Denmark +45 70 23 62 60 AN43161-EN 0416C Europe-Other +43 1 333 50 34 0 Finland/Norway/Sweden +46 8 556 468 00 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9434 Italy +39 02 950 591 Japan +81 45 453 9100 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Russia/CIS +43 1 333 50 34 0 South Africa +27 11 570 1840 Spain +34 914 845 965 Switzerland +41 61 716 77 00 UK +44 1442 233555 USA +1 800 532 4752