Danalyzer 370XA Gas Chromatograph
Then 1995
Then and now Complexity C9+, H2S, methanol, wet gas Fiscal / Custody transfer Simpler installation /operation 1995 2014
Danalyzer 370XA developed to address Gas Measurement Sector s Major Issues Skills Shortage in the Field Less Time For Construction, Start-up, and Maintenance Reduced CAPEX and OPEX
Danalyzer 370XA Gas Chromatograph Smaller, Simpler, and Smarter C6+ Analysis Unique Maintainable Module Full Function Local Operator Interface Trusted Danalyzer C6+ 4-minute Analysis
Danalyzer 370XA developed to address Gas Measurement Sector s Major Issues Skills Shortage in the Field Extensive use of software assistants to guide the user Module replacement requires no GC specific skills Self checking algorithms Less Time For Construction, Start-up, and Maintenance Simplified installation Quick-start guide to help get units up and online in hours, not days Field Module replacement for quick maintenance turnaround Reduced CAPEX and OPEX Smaller footprint Lower Carrier Gas usage Cal gas saver Maintainable Module to minimize cost of repairs
Local Operator Interface Simplifies Operation Perform Common Functions Locally without a Computer Analysis Report Calibration Report Check/Ack. Alarms Start/Stop Analysis Cycle Run Calibration Run Validation Configure I/O
Easy to Use Software Assistants Simplify Operation Change Calibration Cylinder Step by step instructions and selfchecking algorithms to ensure correct change of standard, and accurate operation Auto Valve Timing (AVT) Automatic adjustment of valve timing and integration events to optimize performance previously requiring a very high skill level Module Validation Auto start-up of a new Maintainable Module that purges, configures, and calibrates a new module
Calibration Gas Saver Halves or more the Use of Your Most Expensive Gas Sample Stream Analysis Calibration Analysis Run 1 Calibration Analysis Run 2 Calibration Analysis Run 3 Inje ection Calibration Gas Purging Inje ection Calibration Gas Purging Inje ection Calibration Gas Purging Inje ection Sample Gas Purging 300 cc of Calibration Gas is used for the standard 3-run Calibration Gas Cycle (@ 25 cc/min) Injection Calibration Gas Purging Injection Calibration Gas Saver No Flow Calibration Gas Purging Injection Calibration Gas Saver No Flow Calibration Gas Purging Injection Sample Gas Purging Calibration Gas Saver uses only 135 cc of Calibration Gas less than half the standard cycle! Saves gas and associated cylinder changing costs
Maintainable Module Absolutely Unique Heat sink, airless analytical oven Columns, valves, detector Detector (TCD) Oven configuration based on 500 Danalyzer Columns 6-port valves
Unique Maintainable Module Technology Valve Actuation Solenoids Electronic Pressure Control Valve Intelligent Module Board Stream Selection Solenoids Carrier Pressure Sensor
Maintainable Module Designed to be replaced by whoever is available, not a GC expert Complex start-up, calibration, and tuning performed by software, not a GC expert Removed module can be repaired costeffectively for re-use
4 Maintainable Module Swap Procedure 1 3 2 1.Turn off power and isolate gases 2.Remove module 4 screws, 3 connectors 3.Install new Module, 4 screws, 3 connectors 4. Turn on gases 5. Turn on power 6. Module Validation process starts automatically
Module Start-up 1. 370XA recognizes new Module 2. Downloads Module timed events and calibration information 3. Heats up oven and cycles the valves to purge oven 4. Runs calibration gas and validates measurement 5. Runs Auto- Valve-Timing if required 6. Calibrates 7. Starts Auto- Analysis Sequence Entire Process Runs Without the User Touching a Button! Typically takes less than 3 hours
Uses XA-Series MON2020 Software Same look and feel of MON2000 Common across all XA series GC s Multiple, resizable windows Faster communication
Danalyzer 370XA developed to address Gas Measurement Sector s Major Issues Skills Shortage in the Field Extensive use of software assistants to guide the user Module replacement requires no GC specific skills Self checking algorithms Less Time For Construction, Start-up, and Maintenance Simplified installation Quick-start guide to help get units up and online in hours, not days Field Module replacement for quick maintenance turnaround Reduced CAPEX and OPEX Smaller footprint Lower Carrier Gas usage Maintainable Module to minimize cost of repairs
Thank you any questions?