Application Note. Authors. Abstract

Similar documents
Quantification of Host Cell Protein Impurities Using the Agilent 1290 Infinity II LC Coupled with the 6495B Triple Quadrupole LC/MS System

Applying High Sensitivity LC/MS and Automated Liquid Handling Technology to Peptide Quant

Jet Stream Proteomics for Sensitive and Robust Standard Flow LC/MS

Thermo Scientific Mass Spectrometric Immunoassay (MSIA) Pipette Tips. Next generation immunoaffinity. Robust quantitative platform

Strategies for Quantitative Proteomics. Atelier "Protéomique Quantitative" La Grande Motte, France - June 26, 2007

The Agilent Metabolomics Dynamic MRM Database and Method

Simultaneous Quantitation of a Monoclonal Antibody and Two Proteins in Human Plasma by High Resolution and Accurate Mass Measurements

Simplifying ImmunoAffinity Capture Workflow

Agilent AssayMAP Platform BIOLOGICS SAMPLE PREPARATION INTELLIGENTLY AUTOMATED

Ultra-Sensitive Quantification of Trastuzumab Emtansine in Mouse Plasma using Trap-Elute MicroLC MS Method

Bravo with AssayMAP. Automated Sample Preparation

Application Note. Authors. Abstract. Automated Protein Sample Preparation for LC/MS

Universal Solution for Monoclonal Antibody Quantification in Biological Fluids Using Trap-Elute MicroLC-MS Method

A High Throughput Microchromatography Platform

Agilent Technologies. Technological Innovation and Growth of Liquid Chromatography Mass Spectrometry for Research in the Clinical Laboratory

LC/MS Based Quantitation of Intact Proteins for Bioanalytical Applications

Appendix. Table of contents

Generating Quantitative Assays for Biomarker Development. Jeff Whiteaker, Ph.D. Paulovich Laboratory, Fred Hutchinson Cancer Research Center

Automation for LC/MS Sample Preparation: High Throughput In-Solution Digestion and Peptide Cleanup Enabled by the Agilent AssayMAP Bravo Platform

BioBA High Capacity Enrichment Sample Preparation Kit:

Accelerating Throughput for Targeted Quantitation of Proteins/Peptides in Biological Samples

Agilent AssayMAP Bravo Platform AUTOMATED PROTEIN AND PEPTIDE SAMPLE PREPARATION FOR MASS SPEC ANALYSIS

A Rapid and Reproducible Immuno-MS Platform from Sample Collection to Quantitation of IgG

Technical tips for ELISA and multiplex

Agilent RapidFire 365 High-throughput Mass Spectrometry System ACCELERATING PHARMACEUTICAL DRUG DISCOVERY

Measurement of Hepcidin-25 in Serum Using the Agilent 6490 Triple Quadrupole LC/MS with ifunnel Technology

A Highly Accurate Mass Profiling Approach to Protein Biomarker Discovery Using HPLC-Chip/ MS-Enabled ESI-TOF MS

Improving Sensitivity for an Immunocapture LC-MS Assay of Infliximab in Rat Plasma Using Trap-and-Elute MicroLC-MS

The Agilent StreamSelect LC/MS Solution: Increasing the Throughput of a Triple Quadrupole Mass Spectrometer

DISCOVERY AND VALIDATION OF TARGETS AND BIOMARKERS BY MASS SPECTROMETRY-BASED PROTEOMICS. September, 2011

Human Cytokine Assay Products from Meso Scale Discovery

Pharmaceutical LC/MS Solutions from Agilent Technologies

Robustness of the Agilent Ultivo Triple Quadrupole LC/MS for Routine Analysis in Food Safety

SUPPLEMENTAL MATERIALS AND METHODS

Introduction to Proteomics

Development and evaluation of Nano-ESI coupled to a triple quadrupole mass spectrometer for quantitative proteomics research

Rapid quantitation of therapeutic antibodies in animal plasma

Immunoassays and Mass Spectrometry: Transitioning Technology

Benefits of 2D-LC/MS/MS in Pharmaceutical Bioanalytics

Sample Preparation Automation and QQQ Workflows for Peptide Quantitation in DMPK

reverse transcription! RT 1! RT 2! RT 3!

A Quality Control Procedure for the Agilent Bravo Platform

Mouse Sclerostin ELISA

ProteinPilot Report for ProteinPilot Software

Gary A. Schultz, Advion BioServices, Inc. Ithaca, NY 14850

Rapid Extraction of Therapeutic Oligonucleotides from Primary Tissues for LC/ MS Analysis Using Clarity OTX, an Oligonucleotide Extraction Cartridge

Understanding life WITH NEXT GENERATION PROTEOMICS SOLUTIONS

Online affinity and digestion: A flexible and robust tool for the characterization and quantification of proteins

Faster, easier, flexible proteomics solutions

AssayMAP. Mitu Chaudhary Product Marketing Manager Bay area product specialist Robert (Bob) Boice

CLSI C60: Assay Validation & Post-Validation Monitoring

Bio-Plex suspension array system

Application Note. Authors. Abstract. Ravindra Gudihal Agilent Technologies India Pvt. Ltd. Bangalore, India

Increasing MS Sensitivity for Proteomics. ifunnel Technology for the QQQ and Q-TOF

LC/MS/MS Determination of PFOS and PFOA in Water and Soil Matrices

Bio-Plex suspension array system

Agilent 6490 Triple Quadrupole LC/MS System REACH ULTIMATE SENSITIVITY WITH BREAKTHROUGH IFUNNEL TECHNOLOGY

Agilent Automation Solutions Portfolio FLEXIBLE AUTOMATION MADE EASY

Accelerate mab Characterization Using Automated Sample Prep

Peptide Mapping: A Quality by Design (QbD) Approach

Fast, accurate sequencing starts here. Accelerate your NGS success with automated sample prep

Application Note. Kwasi Antwi, Amanda Ribar, Urban A. Kiernan, and Eric E. Niederkofler Thermo Fisher Scientific, Tempe, Arizona

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

IMPROVE SPEED AND ACCURACY OF MONOCLONAL ANTIBODY BIOANALYSIS USING NANOTECHNOLOGY AND LCMS

Providing Traceability for Protein Biomarkers

Thermo Scientific Peptide Mapping Workflows. Upgrade Your Maps. Fast, confident and more reliable peptide mapping.

Agilent 6430 Triple Quadrupole LC/MS System

Peptide enrichment and fractionation

Eliminating Bottlenecks with the Agilent Encore Multispan Liquid Handler

U-PLEX Biomarker Group 2 (Human, Mouse, and NHP)

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Cytokine Assays: Base Kit

Automated Protocol for ANTI-FLAG High Sensitivity, M2 coated 96-well plate Using the Sciclone ALH 3000 Workstation (Caliper Life Sciences)

Comparison of Luminex 200 to MAGPIX using the Poultry Serology Assay Technical Notes

New Approaches to Quantitative Proteomics Analysis

LECTURE 8. Biologics and Future Technologies for the Clinical Laboratory

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI

MagSi Beads. Magnetic Silica Beads for Life Science and Biotechnology study

MultiQuant Software 2.0 with the SignalFinder Algorithm

ichemistry SOLUTIONS Integrated chemistries to boost mass spectrometry workflows

Analysis of Nucleosides Using an Agilent Infinity II High Speed UHPLC with the 6130 Single Quadrupole Mass Selective Detector

MULTI-ARRAY. 1-Plate Kit 5-Plate Kit 20-Plate Kit

Host Cell Protein Analysis Using Agilent AssayMAP Bravo and 6545XT AdvanceBio LC/Q-TOF

HyperCap, an automatable workflow on the Agilent Bravo B

Human Serum Albumin Assay

Jet Steam Proteomics with Automated Sample Prep for Superior Peptide Quantitation

For Research Use Only Ver

Total Mouse IgG Immunoglobulin Assay

Human Cytokine Assays: Tissue Culture Kit (96-well)

Application Note. Abstract. Authors. Environmental

Automated Plasma Proteomic Sample Preparation

Skyline & Panorama: Key Tools for Establishing a Targeted LC/MS Workflow

Separation of Native Monoclonal Antibodies and Identification of Charge Variants:

Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals.

A Sub-picogram Quantification Method for Desmopressin in Plasma using the AB SCIEX Triple Quad 6500 System

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Mouse Cytokine Assays: Tissue Culture Kit

Bioanalysis of Intact Therapeutic Monoclonal Antibodies from Non-Human Primates Using MSIA D.A.R.T. S Technology

Supporting Information. for. Development and Application of a Quantitative Multiplexed Small GTPase Activity. Assay Using Targeted Proteomics

Proinflammatory (hu) Demonstration Kit

Automated Plasma Proteomic Sample Preparation

KLH IgM (Mouse) ELISA Kit

Transcription:

Automated, High Precision Tryptic Digestion and SISCAPA-MS Quantification of Human Plasma Proteins Using the Agilent Bravo Automated Liquid Handling Platform Application Note Authors Morteza Razavi, N. Leigh Anderson, Matthew E. Pope, Richard Yip, and Selena S. Larkin SISCAPA Assay Technologies, Inc., Box 5339, Washington DC 29, USA Christine A. Miller Agilent Technologies, Inc., 531 Stevens Creek Boulevard, Santa Clara, CA 9551, USA Abstract An addition only tryptic protein-digestion protocol, followed by multiplexed SISCAPA peptide enrichment, was developed and automated in 96-well format on the Agilent Bravo Automated Liquid Handling Platform. These combined protocols, taking as little as 4 hours total, were coupled with high throughput mass spectrometry for quantitation of plasma proteins over a wide range of abundance. The complete workflow was demonstrated for suitability of routine quantitation of plasma protein biomarkers based upon sensitivity, multiplexing capability, scalability, reproducibility, and throughput. Terry W. Pearson Department of Biochemistry and Microbiology, University of Victoria, Victoria, B.C., V8W 3P6, Canada

Introduction Quantitation of proteotypic surrogate peptides in relation to internal standards allows quantitative, multiplexed measurement of proteins using methods derived from classical isotope-dilution mass spectrometry (MS) of small molecules 1. Triple quadrupole mass spectrometer (TQMS) instruments have measured light:heavy peptide ratios with CVs of less than 2 % 2. Since the precision of MS measurements depends on the variability of interfering ion signals (matrix and chromatography effects) as well as the intensity of peptide ion signals, achieving low CVs for low abundance peptides is not trivial and adds to the need for reproducible digestion. Quantitative reproducibility of tryptic digestion is essential to support the evolution of mass spectrometric assays for identification and quantitation of peptides, and ultimate measurement of proteins as clinically relevant biomarkers. This Application Note presents the development of a highly reproducible, automated addition only method for tryptic digestion of plasma, followed by the previously-described bead-based SISCAPA enrichment 3 of target peptides and quantitation using TQMS. This method offers precision, simplicity of multiplexing, generality, and is used to improve sensitivity and increase the dynamic range and throughput of peptide MS assays. The combined, automated workflow, together with an optimized liquid chromatography procedure, facilitates multiplexed peptide quantitation with a 3 minute LC/MS sample cycle time, significantly increasing the throughput of protein quantitation. Materials and Methods Hardware An addition-only method was developed to enable facile, efficient, and highly reproducible tryptic digestion of plasma samples followed by SISCAPA enrichment of target peptides, using an Agilent Bravo Automated Liquid Handling Platform (Figure 1). Addition-only tryptic digestion, no SPE SISCAPA capture, wash, elute LC-MRM detection SISCAPA kit Alkylating agent Trypsin Quench 3 3 1 6 Wash Elute Inj Denaturant plate Figure 1. Protocol Overview: Trypsin digestion, SISCAPA enrichment, LC/MS quantitation. 2

All steps in the workflow protocol were carried out using the Bravo Automated Liquid Handling Platform with 96-channel LT disposable tip head and one VarioMag Teleshake plate shaker (Figure 2). The steps were programed for execution using VWorks software. The workflow requires use of three distinct Bravo Automated Liquid Handling Platform Layouts as shown in Figure 3. Trypsin digestion (Figure 3A) A 1 µl solution of pooled human plasma was added to each well of a 96- well plate containing a dried denaturation mix, followed by a 3-minute incubation at room temperature with vigorous mixing. The alkylation and diluting agents were then added to the sample plate, followed by the addition of trypsin at an enzyme:protein ratio of 1:1. The digest plate was sealed and manually moved to an incubator at 37 C for tryptic digestion of 1 to 5 hours. Agilent Bravo Automated Liquid Handling Platform deck configuration 1 2 3 Alignment station Alignment station Orbital shaking station 4 5 6 Alignment station Alignment station Alignment station 7 8 9 Alignment station Alignment station Alignment station Figure 2. Agilent Bravo Automated Liquid Handling Platform deck configuration for all methods. 1 2 3 Tips out Reagent tips Digest plate Contents: Dried denaturation mixture Plate type: 96 LT 2 µl tips Plate type: 96 LT 2 µl tips Plate type: 96-well DeepWell plate 4 5 6 Plasma tips Plasma sample plate Reagent plate Plate type: 96 LT 2 µl tips Contents: Plasma samples Plate type: 96-well PCR plate Contents: Digestion reagents Plate type: 8-well LP reagent plate 7 8 9 Diluent Tips Diluent plate Empty Plate type: 96 LT 2 µl tips Contents: Diluent Plate type: 8-well LP reagent plate Figure 3A. Agilent Bravo Automated Liquid Handling Platform deck layout: trypsin digestion. 3

Stopping digestion, SIS addition, SISCAPA target peptide enrichment (Figure 3B) During tryptic-digest incubation, the deck was rearranged as shown in Figure 3B. The tryptic cleavage reaction was stopped by adding a stopping agent. Each sample was spiked with 5 fmol each of SIS peptides and.5 1. μg each of corresponding antipeptide antibodies conjugated to magnetic beads. The sample plate containing the antibody bead complex was vigorously shaken for 1 hour at 1, RPM to capture the peptides. Bead wash and elution (Figure 3C) The Bravo Automated Liquid Handling Platform deck was set up a third time, (Figure 3C). Three washes were performed to remove unbound matrix, with a custom magnet array designed by SISCAPA Assay Technologies (Figure 4) used to pull the magnetic beads to the sides of the wells for the unbound peptide digest to be removed. The enriched peptides were released from the beads using acid elution. Liquid chromatography tandem mass spectrometry (LC/MS/MS) Analysis was carried out using an Agilent 129 Infinity ultra high performance liquid chromatograph and an Agilent 649 TQMS (Agilent Technologies, CA). Agilent MassHunter Workstation Software was used for both data acquisition and analysis: MassHunter LC/MS Data Acquisition for Agilent 64 Series Triple Quadrupole LC/MS v. B.6. was used for data acquisition while MassHunter Quantitative Analysis v. B.5.2 and MassHunter Qualitative Analysis v. B.6. were used for data analysis. 1 2 3 Tips out Reagent tips Digest plate Plate type: 96 LT 2 µl tips Plate type: 96 LT 2 µl tips Plate type: 96 LT 2 µl tips Contents: SIS peptide Plate type: 96-well PCR plate Contents: Digested plasma Plate type: 96-well DeepWell plate 4 5 6 SIS tips SIS reagent plate SISCAPA reagent plate Contents: SISCAPA reagents Plate type: 96-well PCR plate 7 8 9 Empty Empty Empty Figure 3B. Agilent Bravo Automated Liquid Handling Platform deck layout: stopping digestion, SIS addition, target peptide enrichment. 1 2 3 Tips 1 Elution plate Digest plate Plate type: 96 LT 2 µl tips Plate type: 96-well DeepWell plate Plate type: 96-well PCR plate Contents: Wash/elution buffers Plate type: 384-well plate Contents: Digested plasma Plate type: 96-well DeepWell plate 4 5 6 Clean plate Wash/elution stock plate Digest recovery plate Plate type: 96-well DeepWell plate 7 8 9 Tips 2 Tips 3 SISCAPA magnet plate Plate type: 96 LT 2 µl tips Plate type: 96 LT 2 µl tips Contents: Magnet plate Plate type: Magnet array plate Figure 3C. Agilent Bravo Automated Liquid Handling Platform deck layout: bead wash and elution. 4

Results and Discussion Sensitivity The Limit of Detection (LOD) was calculated as the amount of the SIS peptide below which the CV was consistently greater than 5 % and Limit of Quantitation (LOQ) as the amount of the SIS peptide below which the CV was consistently greater than 2 %. The performance of four example peptide assays is summarized in Table 1. Comparison to a reference method A 5-level calibrator set for Soluble Transferrin Receptor (stfr) was analyzed using the automated protocol and the results were compared to the results of the reference method provided with the calibrator set. As shown in Figure 5, a strong correlation (R 2 =.998) was observed between the SISCAPA measurement of stfr and its measurement by this reference method. The linear regression model had a slope of.1963 ±.3 and a y-intercept of.439 ±.82 at a 95 % confidence interval. Figure 4. SISCAPA custom magnet array used for washing of magnetic beads. Table 1. Agilent Bravo Automated Liquid Handling Platform SISCAPA assay performance data. stfr (mg/l) as measured by SISCAPA Mesothelin Soluble transferrin receptor Protein C inhibitor LPS binding protein Measured in 1 µl plasma (fmol) 9.98 (±.7) 97.8 (± 3.7) 755 (± 15) 392 (± 36) LOD (fmol).99 (±.1).72 (±.7) 1.5 (±.3).89 (±.15) LOQ (fmol) 5.11(± 1.6).72 (±.7) 1.5 (±.3).89 (±.15) 1 8 6 4 y =.1963x +.439 R 2 =.9979 2 1 2 3 4 5 stfr (mg/l) as measured by Roche/Hitachi analyzers Figure 5. Comparison of stfr measurements using robotic SISCAPA and a Roche/Hitachi analyzer. 5

Scalability of digestion and linearity of peptide responses Linearity of peptide recovery is shown as a function of plasma volume for four targets (Figure 6) measured by mixing equal volumes of pooled human plasma (known to contain the four proteins under study) with chicken plasma (deficient in the target peptides). In all cases, linear responses (R 2 >.999) were observed across four technical replicates. Assay performance is shown using forward and reverse curves (bottom panel). Forward curves were generated by titrating synthetic light peptide from 1, fmol to fmol while the corresponding SIS peptide was kept constant at 5 fmol. For SIS-varying reverse curves the SIS peptide was titrated from 1, fmol to 56 amol while maintaining constant light peptide by spiking 5 fmol in all 12 samples. The curves were independently generated and processed using 1 µl/point of pooled human plasma in triplicate. Endogenous levels of the analytes (plateaued part of the forward curve) can be measured,along with limits of detection (LOD) and limits of quantitation (LOQ). Measured analyte (fmol) 1 8 6 4 2 Meso R 2 =.973 5 1 15 1,25 1, 75 5 25 stfr R 2 =.999 5 1 15 1, 8, 6, 4, 2, PCI R 2 =.961 5 1 15 LPSBP R 2 =.992 5 1 15 6 Meso 72 stfr 7,2 PCI 3, LPSBP 5 6 6, 2,5 4 48 4,8 2, 3 36 3,6 1,5 2 R 2 =.999 24 R 2 = 1 2,4 R 2 =.999 1, R 2 = 1 1 12 1,2 5 Chicken 1:1Mix Human Chicken 1:1Mix Human Chicken 1:1Mix Human Chicken 1:1Mix Human 1, Meso 1, stfr 1, PCI 1, LPSBP 1, 1, 1, 1, 1, 1, 1, 1, 1 1 1 1 1 Fwd 1 Fwd 1 Fwd 1 Fwd Rev Rev Rev Rev 1 1 1 1.1 1 1 1,.1 1 1 1,.1 1 1 1,.1 1 1 1, Varying spiked peptide (fmol) Varying spiked peptide (fmol) Varying spiked peptide (fmol) Varying spiked peptide (fmol) Measured analyte (fmol) Figure 6. Scalability of digestion and linearity of peptide responses. Measured analyte (fmol) Measured analyte (fmol) 4, 3,2 2,4 1,6 8 6

Reproducibility of trypsin digestion and SISCAPA enrichment To evaluate the precision of the automated SISCAPA protocol, including the digestion and peptide enrichment steps, three sets of 12 samples (N = 36) were digested and processed independently on 3 days. Coefficients of variation for the total workflow (digestion + SISCAPA peptide enrichment + mass spectrometric analysis) were consistently less than 5 %, and as low as 2.8 % for proteotypic peptides (Table 2). The coefficients of variation attributable to the tryptic digestion step were estimated from % to 2.2 % for given targets, with the limiting factor in SISCAPA assay precision being MS performance rather than sample preparation. Table 2. Precision of Agilent Bravo Automated Liquid Handling Platform digestion and SISCAPA workflow. Addition-only protocol on Agilent Bravo Automated Liquid Handling Platform 1 µl plasma samples in 96-well plates Mesothelin %CV (n = 12) stransferrin receptor MS Day 1 5.1 4. 2. MS Day 2 5.4 2.8 2. MS Day 3 6.7 1.8 2.6 SISCAPA + MS Day 1 7. 3.1 2.1 SISCAPA + MS Day 2 5.1 2.2 2.6 SISCAPA + MS Day 3 5.9 2.8 2.5 Digest + SISCAPA + MS Day 1 6.1 2.6 4.7 Digest + SISCAPA + MS Day 2 4.6 3.3 5.1 Digest + SISCAPA + MS Day 3 6.3 3.4 4.7 LPS binding protein Throughput This workflow, implemented on the Bravo Automated Liquid Handling Platform, facilitates digestion and enrichment of 96 samples in parallel in as little as 4 hours. The near-purity of analyte peptides enriched by SISCAPA antibody capture means the subsequent LC/MS quantitation is reduced to a total of 3 minutes per sample. Indeed, it has been demonstrated elsewhere that SISCAPA enriched peptides can be quantitated using RapidFire MS for a cycle time of approximately 1 seconds per multiplexed sample. The combined throughput of this automated process is appropriate for industrial applications including biomarker verification and quantitative analysis. 7

References 1. Rittenberg, D., Foster, G.L., A new procedure for quantitative analysis by isotope dilution with application to the determination of amino acids and fatty acids. J. Biol. Chem., 194, 133, 737-774. 2. Anderson, et al. Mass spectrometric quantitation of peptides and proteins using Stable Isotope Standards and Capture by Anti-Peptide Antibodies (SISCAPA). J. Proteome Res., 24, 3, 235-244. 3. Anderson, et al. High-Throughput SISCAPA-Based Peptide Quantitation Using an Agilent RapidFire High Throughput Mass Spectrometry System. Agilent Application Note 5991-38EN www.agilent.com/chem For Research Use Only. Not for use in diagnostic procedures. This information is subject to change without notice. Agilent Technologies, Inc., 214 Published in the USA, March 4, 214 5991-412EN