Method Suitability Report Membrane Filtration Sterility Test with QTMicro Apparatus

Similar documents
3.2 Test for sterility

Compounding Pharmacies and the USP <71> Sterility Tests

á61ñ MICROBIOLOGICAL EXAMINATION OF NONSTERILE PRODUCTS: MICROBIAL ENUMERATION TESTS

IN THIS SECTION MICROBIOLOGY TESTING EXPERT SOLUTIONS FOR PRODUCT DEVELOPMENT. Bacterial Endotoxin (LAL) Testing

3.3.1 Microbial enumeration tests

Growth Promotion Test Guide for Media Used in Sterility Tests

Final text for addition to The International Pharmacopoeia

ICH Topic Q4B Annex 8 Sterility Test General Chapter. Step 3

Tests to Support Sterility Claim. Imtiaz Ahmed

á62ñ MICROBIOLOGICAL EXAMINATION OF NONSTERILE PRODUCTS: TESTS FOR SPECIFIED MICROORGANISMS

2.6. BIOLOGICAL TESTS

2.6. BIOLOGICAL TESTS

Standard Operating Procedure Title: Stock Suspensions of Micro-Organisms

The Microbiological Requirements of a Stability Study. Ngoc Anh-Thu Phan 19 th June 2012

Final text for addition to The International Pharmacopoeia

MICROBIOLOGICAL EXAMINATION OF NON-STERILE PRODUCTS: TEST FOR SPECIFIED MICRO-ORGANISMS Test for specified micro-organisms

ANALYTICAL REPORT: Comparison of the Microbial Recovery Efficacy of QI Medical EnviroTest Paddles versus a Conventional Contact Plate

Guidance. Media Fills for Validation of Aseptic Preparations for Positron Emission Tomography (PET) Drugs DRAFT GUIDANCE

HardyVal TM CSP RANDOM TEST KIT

European Pharmacopoeia

BIO & PHARMA ANALYTICAL TECHNIQUES

Today s Topics. General Quality Control Best Practices. Practices Antimicrobial Effectiveness Testing(AET) Best Practices Environmental Isolates

Specifications, Methods and precedence of source with particular reference to microbiological requirements in TGOs.

Microbiological Quality Control as Described in the Compendia. Scott Sutton, Ph.D.

Erin Patton, MS Senior Product Specialist Charles River Labs, Microbial Solutions

PRESERVATIVE EFFICACY TEST FOR COSMETIC PRODUCT

Validation of the BacT/Alert as an Alternative Sterility Test for Dendreon s Autologous Cell Therapy Product Timothy Wood QC Scientist

Cosmetics Microbiology Detection of Candida albicans

Industry Case Study: A Microbial Investigation of Contamination by Burkholderia multivorans. Jim Klein and Geert Verdonk Merck & Co., Inc.

Protocol Reference: Verification Protocol

STERILITY TESTING PHARMACEUTICAL MICROBIOLOGY SEMESTER V. Dr. Jigar Shah

2.1 Tryptone Soya Broth containing 4% Tween 80 (TSB + T), or another appropriate deactivating broth.

Validation Guide Sterisart NF

Validation Guide Sterisart NF

Published Standards ISO General instructions for microbiological examination ISO Enumeration and detection of aerobic mesophilic bacteri

This document is a preview generated by EVS

COMPARE USP <71> TEST METHODS TO ALTERNATIVE METHODS

KILL TIME STUDIES Antimicrobial Activity of Advanced Cellular Silver (ACS) 200 Using Methicillin resistant Staphylococcus aureus (MRSA)

Submitted May 16, May 23, 2014 BY:

DUAL SPECIES BIOLOGICAL INDICATOR SPORE STRIPS For Monitoring Ethylene Oxide (EO), Steam and Chemiclave Processes

Overview of a sterility assurance program for PET drugs

ASSESSMENT OF THE MICROBICIDAL ACTIVITY OF AN ACCELERATED HYDROGEN PEROXIDE- BASED FORMULATION (AHP-5) AGAINST VRE AND MRSA

EC-type approval protocol No. 2 Microbiological safety of toys containing aqueous media REV 2

Introduction. Michael J. Miller, Ph.D. RMM»

Study Title Antimicrobial Activity and Efficacy of Seal Shield's Electroclave. Test Method Custom Device Study. Study Identification Number NG7233

EC-type approval protocol No. 2 Microbiological safety of toys containing aqueous media REV 3

Microbiology for Oral and Topical Products - The basics Scott Colbourne Business Manager NSW ALS Food & Pharmaceutical

Submitted May 16, May 23, 2014 BY:

MICROBIOLOGICAL EXAMINATION OF NON-STERILE PRODUCTS: TEST FOR SPECIFIED MICRO-ORGANISMS (2)

KILL-TIME STUDIES Antimicrobial Activity of EO H 2 O Using Clostridium difficile spores Test Solutions: Acidic H 2 O Submitted December 12, 2011

ENVIRONMENTAL PARAMETERS OF GROWTH

TECHNICAL LEAFLET. MEDI-MEDIA-FILL KIT SUPPLY PACKAGE (Code: MR-25/S)

KILL-TIME STUDIES Antimicrobial Activity of Experimental Solutions Using Legionella pneumophila Test Solution: ACS 200 Submitted June 6, 2012

New York State Department of Health - Wadsworth Center Laboratory of Environmental Biology NYS ELAP Laboratory ID 10765

New York State Department of Health - Wadsworth Center Laboratory of Environmental Biology NYS ELAP Laboratory ID 10765

TSA + LTHT CSG Contact Plate

HardyVal TM CSP MEDIUM-RISK LEVEL MEDIA-FILL CHALLENGE KITS

USP Chapter <797> & <800> Testing Supplies

CLENDO INDUSTRIAL LABORATORY. CUSTOMER SAMPLE ANALYSIS REPORT Page1 of 8. FDA No CLENDO CONTROL NO. 9081, 9095

Test Method for the Continuous Reduction of Bacterial Contamination on Copper Alloy Surfaces

New York State Department of Health - Wadsworth Center Laboratory of Environmental Biology NYS ELAP Laboratory ID 10765

ENVIRONMENTAL PARAMETERS OF GROWTH

<51> Uji Batas Mikroba Farmakope Indonesia ed.4 (<61>Microbial Limit Test) and Identification method. Marlia Singgih Wibowo School of Pharmacy ITB

Cosmetics Industry. Culture media for microbiological analysis of cosmetic products

Antimicrobial effectiveness evaluation of Isocide powder coating versus stainless steel plate

Study Summary. Results: All tested media-filled vials were negative for growth of any microorganisms.

Microbiological Analysis of Pharmaceutical Preparations

ISO INTERNATIONAL STANDARD

Cleaning Efficiency of The MicroOne Wipe and Mop Cleaning System

Environmental Monitoring of Aseptic Processing Areas - 2

Reagents for Pharma Industry. Chapter 3

Antimicrobial Lab Test Report

TRYPTIC SOY AGAR (TSA) WITH LECITHIN AND TWEEN 80

Application of the BacT/ALERT R 3D system for sterility testing of injectable products

Biocidal Surface Test - Clinell Wipes ~ Project Report Prepared for GAMA Healthcare Ltd ~ Huddersfield Microbiology Services Oct 06

Test Method for Efficacy of Copper Alloy Surfaces as a Sanitizer

Media Fill Test Kits USP <797> COMPLIANT. HARDY DIAGNOSTICS. Low Risk. Medium Risk. High Risk

Disinfectant Qualification A Multifaceted Study

This study report is published with the express written consent of Antimicrobial Test Laboratories. Business Mobility Now!

Laboratory Testing Service Guide

Why Do I Test, What Do I Test & When Do I Test It? Ross Caputo, PhD Chief Technical Officer Eagle Analytical Services

2/25/2013. Psychrotrophs Grow between 0 C and C Cause food spoilage Food Preservation Temperatures

Microbiology Research Associates

The Effect of Air Pockets on the Efficiency of Disinfection of Respiratory Equipment by Pasteurization

Disinfection Qualification Testing

Key words: Paracetamol, antibacterial activity, chemical preservative, zone of inhibition.

DISINFECTION QUALIFI CATION TESTING CONSIDERATIONS FOR THE ASEPTIC AND CLEANROOM MANUFACTURING ENVIRONMENT

Sterisart BD Media and Fluids One-Stop Shopping for Sterility Testing

Microbiological Quality of Drug Products after Penetration of the Container System for Dose Preparation Prior to Patient Administration

Evaluation methods of disinfectants. Kwandong Univ. Coll. Med. Hyukmin Lee

Alpha HydroMAID Cleaning Effectiveness of the Alpha HydroMAID Cleaning System versus Conventional Mopping

Dentistry Test methods for dental unit waterline biofilm treatment

REUSABLE MEDICAL DEVICE VALIDATION

New York State Department of Health - Wadsworth Center Laboratory of Environmental Biology NYS ELAP Laboratory ID 10765

Releasing capacity of pre-sterile cotton swabs for discharging sampled microorganisms

Validation Guide for and

WANG, DER-YUAN Ph.D. ISCT GRP, Paris April 23 rd 2014 衛生福利部食品藥物管理署

Solutions for All Your Aseptic Parenteral Drug Formulation/Filling Needs. BUBBLE-FREE FILLING : A New Option in Prefilled Syringe Filling

CHAPTER 4 DISCUSSION. Many types of suitable media can be used to support the fungal growth and there is no

Working document QAS/ June 2011 RESTRICTED DRAFT FOR COMMENT

Transcription:

Method Suitability Report Membrane Filtration Sterility Test with QTMicro Apparatus Bevacizumab (Avastin) 1.25 mg/0.05ml This report provides details on the specifics of a Membrane Filtration Sterility Test using the QTMicro apparatus. The method suitability study was performed as part of protocol MNI 2013 001, where this sample had the sample ID MNI 2013 001 009 and the title Bevacizumab (Avastin) 1.25 mg/0.05ml. The qualified formulation 1 (normalized to 100 ml) is: Avastin 2.5 g α,α trehalose dihydrate 6 g Sodium phosphate monobasic 0.58 g Sodium phosphate dibasic 0.12 g Polysorbate 20 0.04 g Water for Injection This formulation was qualified for a USP <71> Sterility Test at a maximum volume 2 of 0.35 ml with the QT Micro filtration apparatus. This method suitability report applies only to this specific formulation at a maximum volume per membrane as listed above using the QTMicro. The basic method suitability procedure and the qualified Sterility Test are described below. 1 This study is valid only for the CSP formulation submitted. Changing preservatives (or preservative concentration), active(s) concentration or suspending solution components renders this study invalid. Also note that this formulation description is not sufficient for compounding instructions and is not presented as such but only to list the ingredients and their concentrations. 2 The volume of material used in a USP <71> Sterility Test is determined by a combination of the unit fill volume and the number of units made during a single run (a batch ) as described in Tables 2 and 3 of this chapter. A specific situation may require a larger volume of CSP for a sterility test than was qualified in this study. In this case, it may be necessary to use more than one QTMicro per recovery medium, or to use a different sterility test method consistent with USP <71>. Page 1 of 5

Method Suitability Procedure The purpose of the Method Suitability Test is to demonstrate that the CSP does not inherently (by composition) inhibit detection of low levels of viable bacteria (Method Suitability) after appropriate rinsing in the QTMicro. It is possible that antimicrobial residues of the CSP could adhere to the membrane in the QTMicro, inhibiting the growth of contaminating microorganisms. This test employs the maximum volume of CSP qualified, assuming that lesser volumes should leave less antimicrobial residue on the filter to be neutralized. The basic outline of the Method Suitability Study is as follows: 1. Filter the volume of CSP determined by USP <71> through each of 7 QTMicro Devices. The volume of CSP filtered is determined by the category of CSP (antibiotic, ophthalmic or other) and by details provided in USP <71>: a) Table 2 Minimum quantity per each container based on unit fill volume b) Table 3 Minimum number of containers based on batch size 2. Leave the QT Micro filter moist 3. Filter a 20 ml aliquot of Diluting Fluid D through each QTMicro leave the filter moist Diluting Fluid D (as per USP <71>) 4. Filter a second 20 ml aliquot of Diluting Fluid D through each QTMicro leave the filter moist 5. The third 20 ml aliquot of Diluting Fluid D is then inoculated with <100 CFU of each challenge organism individually (see Table 1). Two QTMicro will not be inoculated and serve as the Microbial Purity Controls 6. The inoculated rinse is filtered through each QTMicro leave the filter moist 7. Depending on the challenge organism, fill the QTMicro with either Soybean Casein Digest Broth (a.k.a. SCDB, Trypticase Soy Broth, or TSB) or Fluid Thioglycollate Medium (a.k.a. FTM) (see Table 1). One of the Microbial Purity Control QTMicro samples will be filled with SCDB and the other with FTM. 8. Test tubes containing at least 10 ml the different growth media (SCDM or FTM) (a.k.a. Media Blanks ) will be inoculated with <100 CFU of the challenge organisms and serve as growth promotion controls (Positive Controls). 9. Two additional QTMicro devices will be filled with either SCDM or FTM and serve as Media Sterility Controls (Negative Controls) 10. The test samples and controls will be incubated as described in Table 1 11. A passing Method Suitability Test will meet all of the acceptance criteria in Table 2. Page 2 of 5

Table 1. Challenge Organisms and Growth Conditions Name Staphylococcus aureus (S. aureus) Pseudomonas aeruginosa (P. aeru) Gram/type Gram Positive Coccus (round) Gram Negative Bacillus (rod) ATCC Number 3 Media Incubation Temperature Incubation Time 6538 FTM 30 35 C 3 days 9027 FTM 30 35 C 3 days Bacillus subtilis (B. subtilis) Gram Positive Bacillus, Spore former 6633 SCD 20 25 C 5 days Clostridium sporogenes (C. sporo) Gram Positive Bacillus, spore former, anaerobic 19404 or 11437 FTM 30 35 C 3 days Candida albicans (C. albicans) Yeast 10231 SCD 20 25 C 5 days Aspergillus brasiliensis (A. brasiliensis) Mold 16404 SCD 20 25 C 5 days 3 American Type Culture Collection identification number Page 3 of 5

Table 2. Method Suitability Acceptance Criteria Media Organism Result Method Suitability Challenge Samples CSP Microbial Purity Control Positive Control Negative Control Method Suitability Challenge Samples CSP Microbial Purity Control C. sporo, P. aeru, S. aureus C. sporo, P. aeru, S. aureus B. subtilis, A. brasiliensis, C. albicans Growth expected in this Test Sample in not more than 3 days Growth not expected in this Microbial Purity Control (product negative control) after 14 days Growth expected in this Positive Control in not more than 3 days Growth not expected in this Negative Control after a minimum of 5 days Growth expected in this Test Sample in not more than 3 days (bacteria) or 5 days (yeast and mold) Growth not expected in this Microbial Purity Control in 5 days Positive Control B. subtilis, A. brasiliensis, C. albicans Growth expected in this Positive Control in 5 days Negative Control Growth not expected in this Negative Control after a minimum of 5 days Page 4 of 5

Sterility Test Method Qualified The USP <71> Sterility Test is designed to check a specific number and volume of samples for sterility in both an aerobic recovery medium and an anaerobic recovery medium. It is necessary that the correct number of samples and volumes (according to Tables 2 and 3 in USP <71> are used in this test. The Method Suitability test establishes the ability of the recovery method to neutralize antimicrobial residues from this amount of CSP. Therefore, the compliant Sterility Test Method Suitability Test must have used at least the volume of CSP to be tested. The exact conditions of the test (volume of diluting rinses, number of diluting rinses and formulation of diluent) must be established in the Method Suitability Study. Conditions of a compliant USP Sterility test for the specific formulation and total volume of CSP described above are escribed below: 1. Determine the amount of product needed for a compliant Sterility Test using Tables 2 and 3 of USP <71> a) If this volume of material to be tested is less than or equal to the volume qualified, proceed b) If this volume of material to be tested exceeds the volume qualified: i. Perform an additional method suitability test to evaluate the larger volume ii. Adjust the fill volume and number of units for this fill to bring the volume to be testing in line with the qualified parameters iii Use different equipment. NOTE: All compliant Sterility Tests must have a relevant method suitability test. 2. The formulation cited above is passed through a QTMicro apparatus for TSB, and a second QTMicro for FTM recovery media 3. Three 20 ml volumes of Diluting Fluid D (see USP <71>) are passed through each QTMicro apparatus to rinse the filter. Note: The diluting fluid volumes should be completely filtered (without drying the membrane) before the next volume is filtered. 4. One QTMicro is filled with TSB, the other with FTM. a) The TSB filled container is incubated at 22.5 o C (± 2.5 o C) for 14 days b) The FTM filled container is incubated at 32.5 o C (± 2.5 o C) for 14 days 5. The incubated, media filled QTMicro are checked regularly through the 14 days and at the end of the incubation period (if necessary) for growth. The Sterility Test fails if either microbial media shows evidence of growth. Scott Sutton, Ph.D. 12/8/14 Date Page 5 of 5