Available online at ScienceDirect. Procedia Engineering 132 (2015 )

Similar documents
OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Pharmaceutical Reference Standards: Overview and Role in Global Harmonization

European Pharmacopoeia Reference Standards A Lodi, Head of the Laboratory Department, EDQM, Council of Europe

Good practices in quality control in pharmaceutical industry - Overview of regulatory guidelines

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

Standard Practices for Measurement Traceability in Forensic Toxicology

SCC Requirements and Guidance for Calibration and Measurement Traceability in Testing Laboratories

GMP GUIDELINES. GMP Guides from Industry Organisations. FDA cgmp WHO GMP. cgmp Guide Drugs 21 CFR 210 GAMP. ISPE Technical Guides

á1225ñ VALIDATION OF COMPENDIAL PROCEDURES

The procurement, qualification and calibration of lab instruments: An Overview

A016 Traceability of measurement results compared to national and international measurement standards

Process development and basic GMP

Calibration in the Laboratory

Reference Standard Characterization. Steve Lane. General Manager, NSF Reference Standards.

INTRODUCTION TO THE QUALITY SYSTEMS APPROACH TO CGMP COMPLIANCE

Registration and renewal of the credential certificate for registration

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

At Your ServIce. Microbiology Services. The life science business of Merck operates as MilliporeSigma in the U.S. and Canada.

General Accreditation Criteria Equipment assurance, in-house calibration and equipment verification

FDA Quality Metrics, Data Integrity and Application of Statistics Throughout Process Validation in a Global Economy

22 January Division of Dockets Management (HFA 305) Food and Drug Administration 5630 Fishers Lane, rm Rockville, MD 20852

A2LA. R231 Specific Requirements: Threat Agent Testing Laboratory Accreditation Program. December 6, 2017

Pharmacopoeial Reference Standards

Understanding the Characteristics and Establishing Acceptance Criteria for Analytical Methods Validation

ISO INTERNATIONAL STANDARD

Regulatory Perspective on Assuring Ingredient Quality

DECEMBER 2008 GUIDANCE NOTES ON PRODUCT QUALITY REVIEW

Drug Quality Assurance: Systems at ChemCon Author: Dr. Peter Gockel

Analytical Procedures and Methods Validation for Drugs and Biologics

Documenta tion and Records

Phase Appropriate GMPs for IMPs. Presented by: Karen S. Ginsbury For: IFF, October 31, Nov 02, 2017

APPENDIX H TRACEABILITY OF MEASUREMENT

Guide to Fulfillment of Validation and Verification of Examination Requirements

EUROPEAN COMMISSION ENTERPRISE AND INDUSTRY DIRECTORATE-GENERAL. EudraLex The Rules Governing Medicinal Products in the European Union

EUROPEAN INDUSTRIAL PHARMACISTS GROUP. Guidance on CPD for QUALIFIED PERSONS

CRITERIA FOR VALIDATION OF METHODS USED BY CHEMICAL LABORATORIES AND RELATED INDUSTRIES

Pharmaceutical Compounding Sterile and Nonsterile Preparations. Jeanne Sun, PharmD

Excipients Facing Increased Scrutiny How to Use Secondary Reference Standards to Help Maintain Regulatory Compliance

Validation Needs for Sterilization by Aseptic Filtration

Validation/Verification of Test Methods An FDA Perspective. Laure H. Kairawicz, Ph.D. Senior Scientist Expert Witness

Analytical Method Transfer Program. Richard B. Nguyen. Associate Scientific Liaison, Chemical Medicines, USP

MAURITAS R3:2017 Issue No. 7:Revision 1 Date: November 2017 MAURITAS. Traceability of measurement. Mauritius Accreditation Service. Copyright MAURITAS

The Council of Europe. EDQM Milestones (1) EDQM Milestones (2) European Directorate for the Quality of

ICH Q2(R1) A Primer. cgmp. GxP PIC/S SOP ISO QA/QC LOD/LOQ. Validation of Analytical Methods API EP FDA OECD GCP

TECHNICAL REQUIREMENTS (CLAUSE 5)

Subpart B Organization and Personnel, 21 CFR Responsibilities of the quality control unit.

Good Manufacturing Practices Purpose and Principles of GMP. Tony Gould

International Pharmaceutical Excipients Council Of The Americas

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

GUIDANCE NOTE 37 MEDICAL GASES DATA INTEGRITY

Perspectives on Method Validation: Importance of Adequate Method Validation

Journal of Chemical and Pharmaceutical Research

Trinity College Dublin QP Forum 2017 Tuesday 25 th April

References Concept. Principle. EU Annex 11 US FDA , (g), (i), 11 Orlando Lopez 2/15/11. Old Annex 11.

EU and FDA GMP Regulations: Overview and Comparison

ISO INTERNATIONAL STANDARD

Validation & Transfer of Methods for

American Association for Laboratory Accreditation

European Regulations for Medicines. Place and Role of the European Pharmacopoeia in Europe Ph. Eur. concept

Synopsis: FDA Process Validation Guidance

Guidance for Industry Quality Systems Approach to Pharmaceutical Current Good Manufacturing Practice Regulations

Q & A on USP General Chapter <823> Revision Proposal Webinar held on 21 February 2011

Analytical Procedures and Methods Validation for Drugs and Biologics

Overview of Regulatory Requirements for API and Formulations

Thank You to United States Pharmacopeia for Their Sponsorship and Assistance

Latino America Consultores Innovation & Technology to make your Business Compliant

1 Analytical Validation within the Pharmaceutical Lifecycle

RADIOPHARMACY GMP QC HPLC VALIDATION

Contents. Contents (13) 1 Production (23)

GMP, GLP OR ISO 17025: HOW DO THESE APPLY TO OUTSOURCED ANALYTICAL TESTING?

APLAC GUIDANCE ON REFERENCE MATERIAL USE AND PRODUCTION

Your partner in the pharmaceutical industry

Measurement Uncertainty Guide. ISO Accreditation Program

PROCESS VALIDATION ROCHAPON WACHAROTAYANKUN, PH.D.

According to USP <1058>

CASE STUDIES OF DUST AMS CALIBRATION AT EXTREME CONDITIONS

Implementation of the ICH Q3D guideline in the Ph. Eur.

International Journal of Pharma and Bio Sciences DEVELOPMENT OF ACCELERATED STABILITY PROTOCOL FOR SILDENAFIL TABLETS A EUROPEAN PERSPECTIVE REVIEW

Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals.

CONCEPT PAPER FOR COMMENT

QUALITY AGREEMENT. The following Agreement has been concluded between

Sam Chappell, CIML Honorary Member, USA

Lessons from Pharmaceutical Laboratory related FDA Warning Letters

Impacto de las Nuevas Tendencias Regulatorias en la Producción de Parenterales. Innovacion en Packaging Primario

COMPLIANCE BY DESIGN FOR PHARMACEUTICAL QUALITY CONTROL LABORATORIES INSIGHT FROM FDA WARNING LETTERS

Validation, Verification and Transfer of Analytical Methods (Understanding and implementing guidelines from FDA/EMA, USP and ICH)

Guidance for Industry

Method Development and Validation for Online UV-Dissolution Methods Using Fiber-Optic Technology

Industry Perspective on Manufacturing in Early Development

ISPE-FDA 3 rd Annual CGMP Conference 2 4 June 2014 Baltimore, MD. Detecting GMP Data Integrity Issues

Industry Perspective on PET Manufacturing Comparison of EU and US

INTERNATIONAL RESEARCH JOURNAL OF PHARMACY ISSN Review Article

TECHNICAL AND PRODUCTION REQUIREMENTS of ISO Guide 34:2009

Regulatory Expectations, Standards & Guidelines

DESIGN AND OPERATION OF AN INTERLABORATORY COMPARISON SCHEME

EU Annex 11 US FDA , (g), (i), 11 Orlando López 09-APR

ISPE NORDIC COP CLEAN UTILITIES SEPTEMBER TUUSULA FINLAND. Timo Kuosmanen STERIS Finn-Aqua

DEC STD ISO Quality Systems - Model for Quality Assurance in Design /Development, Production, Installation and Servicing

Testo Expert Knowledge. The GxP Dictionary 1st edition. Definitions relating to GxP and Quality Assurance

Transcription:

Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 132 (2015 ) 811 815 The Manufacturing Engineering Society International Conference, MESIC 2015 Metrological Regulations for Quality Control Equipment Calibration in Pharmaceutical Industry F. Losada-Urzáiz a, *, C. González-Gaya a, M.Á. Sebastián-Pérez a a Department of Construction and Manufacturing Engineering. ETSII UNED C/ Juan del Rosal 12, 28040 Madrid, Spain Abstract Drug manufacturing is an extremely regulated industry, where regulatory agencies established actions to assure high quality of pharmaceuticals. Testing final products in laboratory requires adequate facilities and equipment. Error occurrence in measurement of drug physicochemical properties introduces risk in quality of products, which could be harmful for patients. Errors are frequently produced by use of inadequately calibrated instruments. This derived in a need of establishing methods and frequency of calibration. Due to the ambiguity of regulations and continuous technological advances, regulations are not well defined. European and USA Pharmacopoeias are main sources of information for analytical testing, thus, should describe requirements for calibration. Several EP and USP methods require calibration or verification of instruments, but in most cases these operations are not well defined. FDA states that equipment should be calibrated following specific procedures, with defined limits and intervals according to written programs. Because regulations provide inadequate guidance for laboratories, independent organizations release guidelines to promote best practices. Integration of guidelines and standards in regulatory documents would improve calibration procedures and diminish risk of errors. 2015 2016 Published The Authors. by Elsevier Published Ltd. by This Elsevier is an open Ltd. access article under the CC BY-NC-ND license Peer-review (http://creativecommons.org/licenses/by-nc-nd/4.0/). under responsibility of the Scientific Committee of MESIC 2015. Peer-review under responsibility of the Scientific Committee of MESIC 2015 Keywords: Equipment calibration; Pharmaceutical industry; Laboratory. 1. Introduction The process of drug manufacturing is an extremely regulated industry, where several regulatory agencies around the globe have established several actions to control, maintain and assure higher degree of quality of * Corresponding author. Tel.: +34-679-96 2-720 E-mail address: flosada14@alumno.uned.es 1877-7058 2015 Published by Elsevier Ltd. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/). Peer-review under responsibility of the Scientific Committee of MESIC 2015 doi:10.1016/j.proeng.2015.12.564

812 F. Losada-Urzáiz et al. / Procedia Engineering 132 ( 2015 ) 811 815 pharmaceuticals. Testing samples of final products in the Quality Control Laboratory requires adequate facilities and qualified and adequately calibrated or verified equipment and instruments. Error occurrence in the correct measurement of physicochemical properties of drug substances introduces a risk in the quality of these products, which can be harmful for the patients. Measurement errors are frequently produced by the use of inadequately calibrated instruments of equipment. This derived in a need of establishing valid methods and frequency of equipment calibration. But due to the ambiguity of current regulations and the continuous technological advances, regulations are incomplete and/or not well defined. Because regulations provide inadequate guidance for modern laboratories, independent organizations release standards and guidelines to improve industry operations and promote best practices. So, several calibration regulations appeared from ISO, ASTM, AOAC and other institutions that help industry to improve their standard operating procedures. The aim of this study is to show the needs to establish, describe and unify defined methods for calibration and or verification (metrological confirmation) of main instruments and equipment used in pharmaceutical laboratories to check quality of drugs in the different Pharmacopoeias and regulation agencies. 2. Methodology and Discussion According to ISO 9000, on fundamentals and vocabulary on quality Management systems, Quality control is the part of quality management focused on fulfilling quality requirements. Analytical quality control, commonly shortened to AQC, refers to all those processes and procedures designed to ensure that the quality requirements of the products being analysed are of the highest level of confidence, which means that results of laboratory analysis are consistent, comparable, accurate, and within specified limits of precision. Quality control Samples submitted to the analytical laboratory must be accurately described to avoid faulty interpretations, approximations or incorrect results. The qualitative and quantitative data generated from the laboratory can then be used for decision making. In a chemical sense, quantitative analysis refers to the measurement of the amount or concentration of an element or chemical compound that differs from other elements or compounds. Several fields such as industry, medicine or law enforcement can make use of AQC. In a few words, quality control means the measure of precision, or how well the measurement system reproduces the same result over time and under varying operating conditions. Quality control begins with sample collection and ends with the reporting of data. AQC is achieved through laboratory control of analytical performance. Initial control of the complete system can be achieved through specification of laboratory services, instrumentation, glassware, reagents, solvents and gases. However, evaluation of daily performance must be documented to ensure continual production of valid data. A check should first be done to ensure that the data should be seen is precise and accurate. This, applied to equipment, usually constitutes the initial calibration and/or adjustment of the measuring instrument. Next, systematic daily checks such as analysing blanks, standards, quality control check samples, and references must be performed to establish the reproducibility of the data. These checks help to certify that the methodology is measuring what is in the sample. These checks are also verifications, and, with initial and/or periodic calibration and adjustment, constitute the metrological confirmation, that is well described in ISO 10012 standard. But measuring instruments tend to change in its response in output or set point value over long periods of time due to such factors as temperature, voltage, and time. This effect is called drift, and is the main reason of the needs of periodic recalibration and adjustment of instruments. At present, European and United States Pharmacopoeias (EP [1] / USP [2]) are the main source of information for analytical testing in a Pharmaceutical Quality Control laboratory, thus, should describe main requirements for calibration of equipment. In fact, there are several monographs in both pharmacopoeias that stablish periodic calibration or, at least, a verification of equipment used for the described analysis, but in most cases, there s not a description on how to develop these activities, or limits for accuracy and precision are not well described. Several EP and USP methods require calibration and/or verification of instruments, but in most cases there s not a clear difference between these two concepts, methods are not described and limits for these operations are not well defined. There are as many definitions of calibration as there are methods. According to the International Vocabulary of Metrology (VIM) [3], calibration is defined as an operation that, under specified conditions, in a first step, establishes a relation between the quantity values with measurement uncertainties provided by measurement

F. Losada-Urzáiz et al. / Procedia Engineering 132 ( 2015 ) 811 815 813 standards and corresponding indications with associated measurement uncertainties and, in a second step, uses this information to establish a relation for obtaining a measurement result from an indication. According to The Automation, Systems, and Instrumentation Dictionary (ISA), the word calibration is defined as a test during which known values of measurand are applied to the transducer and corresponding output readings are recorded under specified conditions. The definition includes the capability to adjust the instrument to zero and to set the desired span. An interpretation of the definition would say that a calibration is a comparison of measuring equipment against a standard instrument of higher accuracy to detect, correlate, adjust, rectify and document the accuracy of the instrument being compared. Typically, calibration of an instrument gives its metrological characteristics, is checked at several points throughout the calibration range of the instrument. The calibration range is defined as the region between the limits within which a quantity is measured, received or transmitted, expressed by stating the lower and upper range values. The limits are defined by the zero and span values. The zero value is the lower end of the range. Span is defined as the algebraic difference between the upper and lower range values. The calibration range may differ from the instrument range, which refers to the capability of the instrument. Calibration finishes with the emission of a calibration certificate. The concept of verification is different than a calibration: VIM and ISA define verification as a confirmation by examination and provision of objective evidence that a given item fulfils specified requirements. Verification includes several activities of confirmation, for example that performance properties or legal requirements of a measuring system are achieved, or that a given reference material is homogeneous for a quantity value and a measurement procedure concerned, or that a target measurement uncertainty can be met. In some cases, where the specified requirements are adequate for an intended use, the verification is called validation. While calibration indicates the error of the instrument and compensates for any lack of trueness by applying a correction, verification indicates that the measurement error is smaller than a so called maximum permissible error. Users define the maximum permissible error as the largest error that they are prepared to accept. These two concepts are included in a unique concept, called Metrological confirmation, which comprises a set of operations required to ensure that measuring equipment conforms to the requirements for its intended use. Metrological confirmation generally includes calibration and verification, any necessary adjustment or repair, and subsequent recalibration, comparison with the metrological requirements for the intended use of the equipment, as well as any required sealing and labelling. In all these process, it is very important to know the acceptance limits or requirements. Because pharmaceutical industries are governed by the Good Manufacturing Practices (GMPs), instrumentation must remain in a validated state, and the site calibration program is a critical part of this effort. GMPs in Europe, and current GMPs (cgmps) in the United States, are the practices required in order to conform to guidelines recommended by agencies that control authorization and licensing for manufacture and sale of food, drug products and active pharmaceutical products. These guidelines provide minimum requirements that a pharmaceutical or a food product manufacturer must meet to assure that the products are of the highest quality and do not suppose any risk to the consumer or public. cgmps are enforced by the U.S. FDA in a Code of Federal Regulations, 21CFR Parts 210 and 211 [4], where they establish several requirements for equipment and measuring instruments. Specifically, 21 CFR 211 states that automatic, mechanical, or electronic equipment or other types of equipment, including computers, or related systems that will perform a function satisfactorily, may be used in the manufacture, processing, packing, and holding of a drug product. Concretely, FDA, in Subpart D of 21 CFR 211, Code of Federal Regulations, in its section 211.68 states that equipment used to manufacture drugs should be routinely calibrated, inspected or checked. Section 211.160 of Subpart I emphasizes laboratory controls and requires calibration of all instruments at suitable intervals in accordance with an established written program, and Calibration procedures shall include specific directions and limits for accuracy and precision, and provisions for remedial action in the event accuracy and/or precision limits are not met. Instruments, apparatus, gauges, and recording devices not meeting established specifications shall not be used. It is left up to individual laboratories to define and defend their choices for calibration, limits and corrective actions. In the same way, in the hole world by the International Council of Harmonization (ICH) on Good Manufacturing Practice Guide for Active Pharmaceutical Ingredients (through the compliance policy guide Q7A [5]) and in Europe by the European Commission (through the Annex 15 to the EU Guide to Good Manufacturing Practice [6-7]) state

814 F. Losada-Urzáiz et al. / Procedia Engineering 132 ( 2015 ) 811 815 that Control, weighing, measuring, monitoring and test equipment that is critical for assuring the quality of intermediates or APIs should be calibrated according to written procedures and an established schedule, but neither particular procedures of calibration, verification or metrological confirmation nor acceptance criteria are defined. The European Directorate for the Quality of Medicines and Healthcare (EDQM) is responsible for the Secretariat of the European Pharmacopoeia Commission and for preparing the General Chapters and Monographs of the European Pharmacopoeia with the groups of experts. The Official Medicines Control Laboratories (OMCL) is a network of laboratories belonging to the Council of Europe that develop and upgrade several guidelines of calibration and/or qualification of laboratory measuring instruments and equipment. These guidelines are not mandatory, but can be used by quality control laboratories to establish calibration methods and frequencies. Several calibration/qualification methods are developed, for example, for calibration of balances [8], pipettes [9]. In the United States of America, another association called the American Society for Testing Materials (ASTM) developed also similar calibration guidelines for instruments and reference materials. All these regulations are not only compatible but also completed with standards from International Organization for Standardization (ISO), an organization that developed general standards for managing measurement systems, including processes and measuring systems, such as ISO 10012, managing quality systems such as ISO 9001, general requirements for the competence of testing and calibration laboratories such as ISO 17025, and also specific standards such as UNE EN ISO 45501 on Metrological aspects of non-automatic weighing instruments or ISO 4787 for calibration of volumetric instruments of laboratory glassware. Table 1. Calibration Regulations and Guidelines for an analytical balance Test USP FEU ISO OMCL Cal Freq. ND ND ND ND Linearity Yes No Yes Yes Sensitivity Yes No Yes No Accuracy* Yes No Yes Yes Precision Yes No Yes Yes Eccentricity Yes No Yes Yes Drift No No No Yes Minimum weight Yes No No Yes Measurement uncertainty No No Yes Yes * Note that in the International Vocabulary for Metrology and in documents of the ISO, the term Accuracy is referred to as trueness. An example of some of these regulations and guidelines applied to a laboratory weighing balance is shown in Table 1. 3. Conclusions As it is described in the present article, calibration of measuring instruments and laboratory equipment is essential for controlling and maintaining the highest quality of pharmaceuticals, because laboratory results are critical for the safety of patients. The article shows the needs of establishing adequate calibration procedures for these instruments and equipment in order to demonstrate the quality and trueness of measurements given by the laboratory. Many pharmaceutical companies are not sure about how to calibrate their laboratory equipment, which tests apply in each case or which the best frequency of calibration is. In this study, several international documents about pharmaceutical industry and its regulation about Metrology have been analysed, showing that in most cases, Pharmacopoeias don t define well how to calibrate, while regulatory agencies such as FDA or European agencies are requiring in their regulations all these aspects.

F. Losada-Urzáiz et al. / Procedia Engineering 132 ( 2015 ) 811 815 815 The detailed explanation in this paper gives a clear idea of the needs of establishing common calibration procedures for basic laboratory equipment, showing some differences between mandatory documents and guidelines. The Integration of guidelines and standards of private and public organizations in unique regulatory documents such as Pharmacopoeias would lead to improve and unify equipment calibration procedures, facilitating inspections and diminishing risk of measuring errors, which ultimately, leads to improvement in quality of pharmaceuticals. Acknowledgements This work was supported by the School of Industrial Engineering (UNED) under Grants 2014-IFC01 and 2014- IEN16. References [1] European Pharmacopoeia, Ph. Eur. 8th Edition, 2014 (8.4). [2] United States Pharmacopoeia, USP37-NF32, 2014. [3] International Vocabulary of Metrology Basic and General Concepts and Associated Terms (VIM 3rd edition). JCGM 200:2012. [4] US FDA. Code of Federal Regulations Title 21, Part 211. Current Good Manufacturing Practice For Finished Pharmaceuticals. US Department of Health and Human Services, US FDA, Center for Drug Evaluation and Research, MD, USA. [5] Guidance for Industry, Q7A Good Manufacturing Practice Guidance for Active Pharmaceutical Ingredients. Center for Drug Evaluation and Research; Center for Biologics Evaluation and Research; Food and Drug Administration; 5600 Fishers Lane; Rockville, MD 20857. [6] EudraLex - Volume 4 Good Manufacturing Practice (GMP) Guidelines, Annex 15 Qualification and Validation, 2001 [7] EudraLex - Volume 4 Good Manufacturing Practice (GMP) Guidelines, Annex 15 Qualification and Validation, 2014. Draft. [8] Qualification of Balances. Annex 8 to the OMCL Network Guideline Qualification of Equipment PA/PH/OMCL (12) 77 7R. [9] Qualification of Equipment. Annex 6: Qualification of piston pipettes. PA/PH/OMCL (09) 64 2R.