EPA Method Proper Collection and Analysis of Soil Samples for Volatile Organics....Green Technology through Innovation

Similar documents
Required Use of Method SW

Best Practices for Collecting Soil Samples for VOC Analysis

Best Practices for Collecting Soil Samples for VOC Analysis. David Kaminski - QED Environmental Systems Inc. Dave Turriff - En Novative Technologies

Best Practices for Collecting Soil Samples for VOC Analysis

Best Practices for Collecting Soil Samples for VOC Analysis

Best Practices for Collecting Soil Samples for VOC Analysis. David Kaminski - QED Environmental Systems Inc. Dave Turriff - En Novative Technologies

Required Use of Method Sw

Course Outline. in Soil/Sediment&WasteSediment&Waste by. Overview of VOCs Analysis Sample Collection VOCs Analysis

Best/Worst Practices for Soil VOC Sampling Webinar VOLATILE ORGANIC COMPOUND ANALYSIS

Volatile Organic Compounds in Solids PBM

OPERATIONAL MEMO GEN-14 Revision 3. November, Volatile Organic Compound (VOC) Sediment Sampling Method 5035 Field Sampling Procedure

NWTPH-Gx. Volatile Petroleum Products Method for Soil and Water

WM 03 Conference, February 23-27, 2003, Tucson, AZ IN-SITU CHEMICAL OXIDATION OF CHLORINATED HYDROCARBONS IN THE PRESENCE OF RADIONUCLIDES

ANALYTICAL METHODOLOGY FOR GROUNDWATER AND SOIL ASSESSMENT GUIDELINES

Field Analytical Method Development for C 6 -C 9 Range Hydrocarbon Measurement

SW 846 METHOD 5035 FIELD SAMPLING GUIDE

Method 0040 for Volatile Total Organics as Unspeciated Mass in Stack Gas Emissions

SOP-C-130 Determination of Total Solids and Percent Solids

Investigation Requirements for Ethanol-Blended Fuel Releases Petroleum Remediation Program

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D*

Volatile Organic Compounds in Soil PBM

Volatile Organic Compounds in Solids PBM

Ground Water & UST Samples: Containers, Preservation and Hold Times Table

NEW JERSEY DEPARTMENT OF ENVIRONMENTAL PROTECTION

HACH TU 5 SERIES TURBIDIMETERS SOP: COMPARISON LAB AND PROCESS. The next standard in the evolution of turbidity

14 May 2003 Project # John E. Dustman Summit Envirosolutions, Inc Bandana Boulevard North St. Paul, Minnesota 55108

STANDARD OPERATING PROCEDURE FOR SOIL, SEDIMENT, AND SOLID WASTE SAMPLING

SOIL SAMPLING FOR VOLATILE COMPOUNDS

SUGGESTIONS FOR REDUCTION OF ANALYTICAL COSTS BY ELIMINATION OF UNNECESSARY QUALITY CONTROL (QC) SAMPLES

2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial waste waters.

The CDS 7000 E Purge & Trap Concentrator

CHROMATOGRAPHY/MASS SPECTROMETRY

Topical Report. By Susan S. Sorini John F. Schabron Joseph F. Rovani, Jr. September 2003

Soil Sampling and NAPLANAL

Back TO Basics: Sample Collection & Handling

CORESTA Recommended Method No. 57

Driving ISM Laboratory Processing Selection with Project Goals

Residual Solvents in medical cannabis analysis using the SRI 8610C FID GC June 2014

CT DEP Reasonable Confidence Protocols. Glen Breland, VP Member CT DEP QA/QC Workgroup

Parsons. APPENDIX E Laboratory Data Validation Reports Calscience Environmental Laboratories, Inc.

1 Laboratory Processing. Learning Objectives

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE.

Improved Accuracy of Dissolved Hydrocarbon Gas Analysis in Water

COPY. Region4 U.S. Environmental Protection Agency Science and Ecosystem Support Division Athens, Georgia

Application Note AN1501

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

Available Laboratory Methods for Soil Gas Analysis

OPTIMIZATION OF HIGH-STRENGTH HYDROCARBON BIODEGRADATION USING RESPIROMETRY

New Developments in Thermal Desorption (TD) Tube & Canister Technologies for Collection & Analysis of Soil Gas

MIP System Field Test for Mid-Low Level TCE

MEMO. July 1, To: Julie Moore, P.E. From: Kim Watson, RQAP GLP Amy Macrellis

Soil Vapor Reproducibility: An Analytical and Sampling Perspective

ASTM D19.06 SUBCOMMITTEE

EXTRACTION OF BIOLOGICAL TISSUES FOR AROMATIC AND CHLORINATED HYDROCARBONS

State of Alaska DEPARTMENT OF ENVIRONMENTAL CONSERVATION DIVISION OF SPILL PREVENTION AND RESPONSE CONTAMINATED SITES PROGRAM

Ensuring Data Integrity

DoD Environmental Monitoring & Data Quality Workshop

Baseline Characterization. The baseline characterization consists of the following tasks:

DETERMINATION OF METHOD UNCERTAINTY A PRACTICAL EXAMPLE FOR THE ANALYSIS OF TRIHALOMETHANES IN DRINKING WATER USING PURGE&TRAP GC-MS

D.9 Particulate Nitrogen and Particulate Carbon

Dissolved Methane Sampling and Analysis Techniques

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

METHOD 24A - DETERMINATION OF VOLATILE MATTER CONTENT AND DENSITY OF PUBLICATION ROTOGRAVURE INKS AND RELATED PUBLICATION ROTOGRAVURE COATINGS

Result. Laboratory Sample Analysis Record

2. Crystallization. A. Background

Surface Water Samples: Containers, Preservation and Hold Times Table

ANALYTICAL REPORT. Job Number: Job Description: CETCO - DK033009

When optimization is discussed throughout this talk, it s in terms of performing more samples, in less time, while minimizing investment in capital

SOP-C-107 Determination of Non Filterable Residue

2. Crystallization. A. Background

METHOD 3500A ORGANIC EXTRACTION AND SAMPLE PREPARATION

Volatile organic compounds Water No CAS Number assigned

DoD Training Environmental Laboratory Accreditation Program Requirements for 8330B

CLEANUP PROCEDURES. P.O. Box Dallas, Texas Ph: (972) Fax: (469) Web:

Guide. Moisture Determination in Test Laboratories. Moisture Determination in Test Laboratories

APPENDIX K. Information sources that may be useful for identification of available water sources that could be selected as sampling locations include:

Oklahoma Checklist (OK)

Advancing Mobility Testing and Aqueous-Phase Sampling in NAPL Zones

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

Investigating the Extraction Efficiency of Dioxins in Several Types of Aqueous Matrices by Automated Solid Phase Extraction Utilizing Disks

Experiment 1 MOLAR MASS DETERMINATION BY FREEZING POINT DEPRESSION

Best Practices for the Interpretation and Use of the Weathered and Biologically Degraded Samples in the Ignitable Liquids Database

Soil Vapor Reproducibility

METHOD 5A - DETERMINATION OF PARTICULATE MATTER EMISSIONS FROM THE ASPHALT PROCESSING AND ASPHALT ROOFING INDUSTRY

Jeff, Thank you for your timely report. Sincerely, Allen Gilliam ADEQ State Pretreatment Coordinator E/NPDES/NPDES/Pretreatment/Reports

Introduction. Think of this Guide as a collection of cheat sheets - a quick reference of FAQs and issues that we get calls on all the time:

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016

MICROWAVE SOLVENT EXTRACTION TECHNIQUES FOR VARIOUS REGULATORY COMPOUNDS

PRACTICAL ASPECTS OFCONDUCTING SOIL VAPOR INVESTIGATIONS in general accordance with the 2015 DTSC Advisory

STANDARD OPERATING PROCEDURES

USEPA Methods and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors

Analysis of Semivolatiles by GC/FID using the Ultra Inert Inlet Liners with Wool

USEPA Methods and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors

METHOD 7550 OSMIUM (ATOMIC ABSORPTION, DIRECT ASPIRATION)

MERCAPTANS, METHYL-, ETHYL-, and n-butyl- 2542

Montana Bureau of Mines and Geology. BASELINE GROUNDWATER MONITORING SAMPLING AND ANALYSIS PLAN Elizabeth Meredith May

Development of standard laboratory based test to measure compost stability Annex A

Volatile Organic Carbon Contaminated Site Assessment

Total Petroleum Hydrocarbons By Gas Chromatography in Colombian Waters and Soils

olyols: Cost-Effective Alternative to Conventional Polyether and Polyester Polyols

Transcription:

EPA Method 5035 Proper Collection and Analysis of Soil Samples for Volatile Organics

History and Justification of Method 5035 Application of Method 5035 Requirements Frequently Asked Questions

History and Justification of Method 5035 Prior to 1997 and publication of SW846 Update III, Volatile Organics for all matrices were collected and analyzed using Method 5030. Update III of SW846 (pub Dec. 1996), soil matrices were removed from Method 5030 and a new Method (5035) was published In 2002, USEPA updated to Method 5035A, removing bulk collection as an option for samples with high concentrations of VOCs

Effect of Volatilization on Bulk Soils VOC Concentrations In soils and solid waste materials, VOCs coexist in gaseous, liquid, and solid phases. Initial concern is in retention of the gaseous phase As the gaseous phase dissipates from the sample, the liquid and solid phases volatilize to maintain equilibrium Loss of TCE was investigated in a narrow mouth container completely filled with sample and held at ambient temperature for 1 hour in the field

Effect of Volatilization on Bulk Soils TCE Concentrations Alan D. Hewitt, USACE, May 1999

Effect of Volatilization on Bulk Soils VOC Concentrations Nearly 100% of the TCE was lost to evaporation within 1 hour of collection Change in loss rate indicative of phase change from liquid or solid (adsorbed) states after initial loss of gaseous TCE in maintenance of equilibrium

Effects of Aerobic Microbial Degradation on VOC Concentrations Even when stored and transported at <6 C, microbial degradation can impact VOC concentrations in soils and solid waste material Slower loss rate than from volatilization Loss rate is compound selective

Effects of Aerobic Microbial Degradation on Select Volatile Organic Analytes Alan D. Hewitt, USACE, May 1999

Effects of Aerobic Microbial Degradation on VOC Concentrations Aromatic compounds selectively lost over halogenated analytes Loss of > 98% of Benzene originally present in the sample occurred within 5 days of collection Historical holding time for VOCs collected in bulk was 14 days to analysis only

History and Justification of Method 5035 To address these potential issues, the EPA published method 5035 On 13 June 1997 Collection of VOCs into a sealed vial in the field to prevent loss to evaporation Require preservation technique be employed within 48 hours to prevent microbial activity

Application of Method 5035 Requirements Collect samples into a tared VOA vial containing either water or methanol and a Teflon coated stirring bar Preserve immediately in the field by acidification with NaH2SO4 or methanol or freeze (<-7 C) within 48 hours of collection.

TCEQ Remediation Guidance Document Collect soils via a closed-system using an open-bore syringe and 40-milliliter (ml) volatile organic analysis (VOA) vials without preservatives. NOTE: Samples collected in VOA vials can only be analyzed one time. Therefore, collect at least three separate vials at each sample location

TCEQ Remediation Guidance Document one vial for low-level concentration analysis, one vial for low-level concentration reanalysis, if needed; and one vial in the event of high levels of VOCs requiring methanol extraction. Collect an additional container (e.g., 40-mL vial or 4-ounce wide-mouth jar) for VOC screening and moisture determination.

TCEQ Remediation Guidance Document Wipe the threads of the vial clean prior to sealing the vial, then quickly seal and chilling the vial and hold and ship the vials at <6 C to the laboratory.

TCEQ Remediation Guidance Document Section 3.0: This recommended collection procedure does not require the use of chemical preservatives in the field. The Remediation Division recommends collecting the samples with no chemical preservation, because soils in Texas can have significant levels of calcareous material that will react with acid preservatives to cause effervescence in the sample

Laboratory Practices Samples properly submitted to the laboratory are analyzed within 48 hours or properly preserved by freezing to extend the holding time to 14 days Sample vials collected in DI Water are never opened in the laboratory prior to analysis Dilutions can only be performed by analyzing an aliquot not to exceed 200 ul / 5 ml of the methanol solution collected in the field (apparent 25 fold dilution)

Frequently Asked Questions What happens if the sample needs a 5 fold dilution for analysis? The only available dilutions for soil samples are undiluted or 25 fold or greater.

Frequently Asked Questions How does the laboratory manage an MS/MSD on low level samples for VOC analysis An MS/MSD can only be performed for VOC analysis when a sample is designated for this purpose and collected in triplicate in the field. If no samples in the analytical batch are submitted in triplicate, MS/MSD results can not be provided for the batch. LCS/LCSD data is used to provide precision and accuracy representations.

Frequently Asked Questions Since the triplicate collected vials are not truly split samples, how does that affect the MS/MSD analysis While not a true split sample, the aliquot historically used for soil VOC MS/MSD was also not a true split sample. Only a 1-5 gram aliquot of the submitted samples was used for analysis. Considering the non-homogeneity of soil samples, this was in effect the same process as defined in Method 5035

Frequently Asked Questions Does adherence to Method 5035 requirements affect the cost of the analysis? Laboratories typically pass on the additional cost of supplying sampling devices to the client. This is accomplished by the addition of a separate device charge or by increasing the cost of the VOC analysis in soil.

Frequently Asked Questions When collecting samples for VOCs, why do we need both VOAs preserved with methanol and unpreserved VOAs? The purpose of the methanol preserved VOA vial is to provide a means of dilution for the analysis when analyte concentrations exceed the upper limit of the calibration curve, typically 200 ug/kg wet weight.

Frequently Asked Questions What are the stir bars for in the NaH2SO4 Or DI Water VOC VOAs? The purpose of the magnetic stirring bars is to provide agitation of the soil during the purge phase of the analysis, primarily to allow sufficient movement of helium through the matrix to continually evolve the target compounds into the headspace

Frequently Asked Questions What is percent moisture used for? The moisture content of the soil is used for two determinations. First, it allows all samples to be normalized to only the concentration present in the solid sample (dried sample) and second allows for the correction of the water miscible solvent volumes when analyzing dilutions from the methanol preserved container

Frequently Asked Questions How do we manage field screening of cores before deciding which samples to prepare for analysis? Collect all samples and preserve or seal in VOA Vials with the proper preservatives. Additionally collect a bulk sample for moisture determination that can be used for screening in the field to determine which samples to submit to the laboratory.

Additional Questions or Comments? Johnny Mitchell ESC Lab Sciences www.esclabsciences.com