Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Similar documents
Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water. Pranathi R Perati Thermo Fisher Scientific, Sunnyvale, CA, USA

Analysis of Fracking Flowback Water From the Marcellus Shale Using In-line Conductivity, Automated Dilution, and Ion Chromatography

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

Jonathan R. Beck and Charles T. Yang; Thermo Fisher Scientific, San Jose, CA

The Benefits of SOLAμ Technology in Sample Preparation. Jon Bardsley and Ken Meadows Thermo Fisher Scientific, Runcorn, UK

Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry

High-speed, High-resolution Oligonucleotide Separations Using Small Particle Anion-Exchangers

A Novel ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge-Variant Analysis

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants

Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

High-Resolution Separation of Intact Monoclonal Antibody Isoforms

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

Nitrogen Determination in Soils and Plants by Flash Combustion using Argon as Carrier Gas

Quantification of Drug Metabolites in Early-Stage Drug Discovery Testing

Extraction of Phthalates from Solid and Liquid Matrices

In-Cell and External Hydrolysis of Biomass for Carbohydrate Determinations Using Automated Solvent Extraction

Alex J. Zhang, Pranathi Perati, and Monika Verma; Thermo Fisher Scientific, Sunnyvale, CA, USA

Introduction. Yan Liu, Victor Barreto, Rosanne Slingsby, Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA

Analysis of Monoclonal Antibody and Related Substances using a New Hydrophobic Interaction Chromatography (HIC) Column

for water and beverage analysis

Thermo Scientific Dionex ICS-900 Ion Chromatography System

Method Optimisation in Bottom-Up Analysis of Proteins

High-Resolution Charge Variant Analysis for Top-Selling Monoclonal Antibody Therapeutics Using a Linear ph Gradient Separation Platform

Determination of 1,4-Dioxane in Drinking Water by Gas Chromatography/Mass Spectrometry (GC/MS) with Selected Ion Monitoring (SIM)

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

High-Throughput Oligonucleotide Analysis on a Small Particle Pellicular Anion-Exchanger

Thermo Scientific EASY-Spray Technology. Plug-and-spray with. State-of-the-art performance

Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass

Highlighted Capillary IC Applications

A Novel Screening Method for Anthropogenic Sewage Pollutants in Waste Water, Ground Water and Drinking Water Samples by LC HRAM Analysis

How to Create Your Own Affinity Titer Column. Kelly J. Flook*, Yury Agroskin and Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA

Chromatography Column Performance and Data Analysis Success Guide. Hints and Tips for Better Purifications

Thermo Scientific EASY-nLC 1200 System. Leading in simplicity. and performance

Implementation of a Walk-Up High- Pressure Capillary Ion Chromatograph for the Fast Separation of Pharmaceutical Relevant Inorganic Anions and Cations

Determination of Manganese in Brine ( 30% NaCl)

Nano LC at 20 nl/min Made Easy: A Splitless Pump Combined with Fingertight UHPLC Nano Column to Boost LC-MS Sensitivity in Proteomics

Claudio De Nardi, 1 Marta Kozak, 2 David Kasper 3 1. Thermo Fisher Scientific, Dreieich, Germany; 2 Thermo Fisher Scientific, San Jose, CA (USA); 3

Using HRAM Survey Analysis Combined with Rapid MS2 Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition

Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography

Parallel LC with Capillary PS-DVB Monolithic Columns for High-Throughput Proteomics

Passion. Power. Productivity. Corporate Headquarters

Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines

Minimizing Carbon Dioxide Contamination in Anion Reagent-Free Ion Chromatography

Agilent 1290 Infinity Binary LC System with ISET Emulation of a Waters Alliance 2695 LC Applying Concave, Convex, and Linear Gradients

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance

electrochemical detection

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

NISTmAb characterization with a high-performance RP chromatography column

Analysis of Monoclonal Antibodies and Their Fragments by Size Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

New Developments in Capillary Ion Chromatography Systems with Electrochemical Detection and Their Applications

Thermo Scientific EASY-nLC 1000 System. Effortless, split-free nanoflow UHPLC Top performance in LC/MS

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

ALLIANCE SYSTEM FOR CARBAMATE ANALYSIS

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column

Determination of Inorganic Anions in Acid Rain Using a Dedicated High-Pressure Capillary Ion Chromatography System

Determination of Cations in Hydraulic Fracturing Flowback Water from the Marcellus Shale

Determination of Trace Concentrations of Oxyhalides and Bromide in Municipal and Bottled Waters Using a Compact Ion Chromatography System

Direct Carbohydrate Analysis in Beverages and Foods Using Pulsed Amperometric Detection or Charged Aerosol Detection

columnsi o n P a c C S 1 6 C a t i o n - E x c h a n g e C o l u m n

TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet

A Fast and Effective Approach for Running EPA Method 539: Determination of Hormones in Drinking Water using SPE and LC/MS/MS

The Quantitative Analysis of Curcuminoids in Food and Food Additives Using Rapid HPLC With Electrochemical, UV, or Fluorescence Detection

Performance characteristics of the 1260 Infinity Quaternary LC system

The system has an optimized porosity distribution and acts as a depth filter, retaining the precipitated protein without well blockage.

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

Application Note. Author. Abstract. Pharmaceuticals. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Direct Carbohydrate Analysis in Beverages and Foods Using Pulsed Amperometric Detection or Charged Aerosol Detection

Quality Control Measures for Routine, High-Throughput Targeted Protein Quantitation Using Tandem Capillary Column Separation

Fast Ion Determinations and Efficient Separations of Complex Samples Using 4 µm Particle-Size Ion Exchange Columns

Thermo Scientific Solutions for Intact-Protein Analysis. Better, Faster Decisions for. Biotherapeutic Development

Optimize Your Process Operations by Improving Process Monitoring

Practical guidance for switching between anion and cation analysis systems

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

columns PepSwift and ProSwift Capillary Monolithic Reversed-Phase Columns

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

Emulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture

Method Transfer from an Agilent 1200 Series LC to an Agilent 1260 Infinity II LC

CX-1 ph Gradient Buffer

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Hillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION

Technical Overview. Author. Abstract. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Analysis of Atrazine in Drinking Water at the ppb Level Using New Agilent Reversed Phase LC Columns. Application. Author. Abstract.

Fast Method Development Using the Agilent 1290 Infinity Quaternary LC System with Column Selection Valve

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental

FRACTIONLYNX APPLICATION MANAGER

Intelligent LC: Advanced control

ph gradient analysis of IgG1 therapeutic monoclonal antibodies using a 5 µm WCX column

ApplicationNOTE Introduction

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

ISET. Agilent 1290 Infinity Binary LC with ISET, emulation of the Waters Alliance 2695 LC system analyzing aromatic acids. Application Note.

Analytically Monitoring the Effect of Fracturing Activity

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo

Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS

Simple, fast and reliable analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

Transcription:

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography Carl Fisher, Monika Verma, and Pranathi Perati; Thermo Fisher Scientific, Sunnyvale, CA, USA

Overview A method was developed to automate the capture of human hormones from drinking water onto solvent extraction cartridges and then eluting them using a solid phase extraction system. The resultant was subsequently quantified with HPLC. Introduction The presence of hormones in drinking water is a human health concern, although safe exposure limits have yet to be established due to the need for further studies to determine the impact of long-term and synergistic exposure. 1 Estriol, estrone, estradiol, ethynylestradiol, equilin, androstenedione, and testosterone (Figure 1) are routinely monitored as part of the U.S. Environmental Protection Agency (EPA) Unregulated Contaminant Monitoring program. 2 FIGURE 1. Sex hormones routinely monitored in drinking water. The naturally occurring human estrogens and androgens are highlighted. Estriol, estrone, and estradiol are estrogens that control female traits, while testosterone is an androgen controlling the development of male characteristics. These hormones are naturally present in humans and are derived from androstenedione during steroidogenesis. Various forms of estrogen are prescribed as a hormonal contraceptive device and to counter hormone deficiencies. Ethynylestradiol is an orally bioactive synthetic estrogen used as an oral contraceptive, while equilin is an equine-derived estrogen used for hormone replacement therapy. Testosterone is prescribed for hormone replacement therapy in males, while androstenedione is a performance enhancing anabolic steroid. 2 Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Methods Sample Preparation Drinking water was fortified with 1 mg/l hormones and mixed by vortexing. Instrumentation Thermo Scientific Dionex AutoTrace 280 Solid-Phase Extraction instrument Thermo Scientific Dionex UltiMate 3000 RSLC system: DGP-3600M Dual-Gradient Micro Pump WPS-3000TRS Rapid Separation Wellplate Sampler, Thermostatted TCC-3000RS Rapid Separation Thermostatted Column Compartment DAD-3000RS Rapid Separation Diode Array Detector (Without Flow Cell) Thermo Scientific Dionex AS-AP Autosampler Data Analysis Thermo Scientific Dionex Chromeleon Chromatography Data System (CDS) software. Sample Concentration and Elution Prior to loading, Dionex SolEx SPE HRPHS cartridges were conditioned by rinsing with methanol, then water, and then drying with nitrogen gas. Twenty ml of sample was loaded onto cartridges, followed by a rinse with water and elution with five ml acetonitrile. This results in a four fold concentration. Two µl from each eluent sample were then separated by HPLC and the hormones quantified using standard curves. FIGURE 2. Dionex AutoTrace 280 SPE instrument; a Dionex SolEx cartridge and the fluid connections are shown. Thermo Scientific Poster Note PN71117_HPLC_2014_E_05/14S 3

Results Chromatographic Separation An absorbance of 214 nm was chosen for run analysis due to the approximately equivalent signal response obtained for all hormones at this wavelength (Figure 3). Only testosterone and androstenedione, structurally very similar molecules, showed substantial absorbance at 240 and 254 nm. All of the peaks had close to baseline or better resolutions (R s (EP)), ranging from 1.4 to 2.9. FIGURE 3. Hormone standards at various wavelengths. 4 150 1 3 7 mau 2 4 5 6 A -5 B C 6 7 8 9 Minutes 10 11 Column: Thermo Scientific Acclaim RSLC 120 C18 column, 2.1 150 mm Mobile phase: Water/Acetonitrile (CH 3 CN) Gradient : CH 3 CN, 10 54% (0 4 min) 54% (4 12 min) 100% (12 16 min) 10% (16 20 min) Flow Rate: 0.20 ml/min Inj. Volume: 2 µl Column Temp.: 20 C Detection: UV, A: 214 nm, B: 240 nm, C:254 nm Sample: Standard mix, 50 mg/l each Peaks: R s (EP) 1. Estriol -- 2. Estradiol 2.9 3. Testosterone 1.7 4. Ethynylestradiol 1.4 5. Equilin 1.7 6. Estrone 1.4 7. Androstenedione -- Chromatographic Limits of Detection (LOD), Limits of Quantification (LOQ), and Linearity As can be seen in Table 1, the LOD and LOQ ranged from 0.025 to 0.4 mg/l and 0.083 to 1.3 mg/l, respectively. TABLE 1. LOD and LOQ for hormones at 214 nm. 4 To determine linearity, triplicate injections of standards ranging from 0.05 to 10 mg/l were run. The coefficient of determination (r 2 ) was >0.999, indicating good fit of the standard curve to the data. 4 Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Recovery To determine the accuracy of the method, three aliquots of drinking water fortified with 1 mg/l of seven hormones were extracted in parallel and analyzed using HPLC. Figure 4 shows a comparison of the chromatograms for the fortified water prior to cartridge loading, a cartridge elution, and an injection of acetonitrile (used as the eluent for SPE) to show the background signal. As can be seen in Table 2, recovery of hormones from drinking water was approximately 100%, ranging from 90 to 124% with an RSD from 1.2 to 4.4. FIGURE. 4. Chromatograms of hormones eluted from a Dionex SolEx SPE HRPHS cartridge (A), fortified drinking water prior to loading (B), and a blank injection (C). 4 15 1 2 4 5 6 mau 3 7 A B C 0 6 7 8 9 10 11 Minutes Column: Dionex Acclaim RSLC 120 C18 column, 2.1 150 mm Mobile phase: Water/Acetonitrile (CH 3 CN) Gradient : CH 3 CN, 10 54% (0 4 min) 54% (4 12 min) 100% (12 16 min) 10% (16 20 min) Flow Rate: 0.20 ml/min Inj. Volume: 2 µl Column Temp.: 20 C Detection: UV, 214 nm Sample: A. Cartridge elution B. Drinking water + 1 mg/l hormones C. Acetonitrile (eluent blank) Peaks: 1. Estriol 2. Estradiol 3. Testosterone 4. Ethynylestradiol 5. Equilin 6. Estrone 7. Androstenedione TABLE 2. Percent recovery of hormones added to a drinking water sample from three Dionex SolEx SPE HRPHS cartridges. 4 Thermo Scientific Poster Note PN71117_HPLC_2014_E_05/14S 5

Conclusion The Dionex AutoTrace 280 SPE instrument greatly facilitates the analysis of aqueous samples that contain low-analyte concentrations by automating the process of concentrating them onto cartridges or discs containing absorbents with specific retention properties. Baseline resolution of all of the hormones analyzed was demonstrated using the Dionex SolEx SPE HRPHS cartridges for sample extraction, followed by analysis on the Dionex UltiMate 3000 RSLC system using the Acclaim 120 C18 analytical column. Good recoveries from drinking water, ranging from 90% for androstenedione to 124% for ethynylestradiol, were shown. The recoveries achieved verify the suitability of these cartridges as a means to concentrate human hormones for subsequent quantification as per EPA Method 539. References 1. World Health Organization (WHO), United Nations Environment Programme, Inter- Organization Programme for the Sound Management of Chemicals. State of the Science of Endocrine Disrupting Chemicals 2012: Summary for Decision-Makers. Bergman, A.; Heindel, J.J.; Jobling, S.; Kidd, K.A.; Zoeller, R.T. WHO Document Number WHO/HSE/PHE/IHE/2013.1. 2. U.S. Environmental Protection Agency (EPA). Basic Information about the Unregulated Contaminant Monitoring Rule 3 (UCMR3). http://water.epa.gov/lawsregs/rulesregs/sdwa/ucmr/ucmr3/basicinformation.cfm#four (accessed April 15, 2014). 3. Smith, G.A.; Zaffiro, A.D.; Zimmerman, M.L.; Munch, D.J. Method 539: Determination of Hormones in Drinking Water by Solid Phase Extraction (SPE) and Liquid Chromatography Electrospray Ionization Tandem Mass Spectrometry (LC-ESI- MS/MS). Document No. 815-B-10-001. U.S. Environmental Protection Agency, Cincinnati, OH, 2010. 4. Thermo Fisher Scientific Technical Note 148: Automated Extraction and Determination of Human Hormones in Drinking Water Using Solid-Phase Extraction and HPLC with UV Detection. Thermo Scientific P/N TN70898_E, Sunnyvale, CA, 2013. [Online] http://www.thermoscientific.com/content/dam/tfs/atg/cmd/cmd%20documents/tn- 148-SP-HPLC-Hormones-Water-TN70898-E.pdf (accessed April 15, 2014). 6 Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

www.thermoscientific.com 2014 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa +43 1 333 50 34 0 Australia +61 3 9757 4300 Austria +43 810 282 206 Belgium +32 53 73 42 41 Brazil +55 11 3731 5140 Canada +1 800 530 8447 China 800 810 5118 (free call domestic) 400 650 5118 Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland +358 9 3291 0200 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9494 Italy +39 02 950 591 Japan +81 6 6885 1213 Korea +82 2 3420 8600 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Norway +46 8 556 468 00 Thermo Fisher Scientific, Sunnyvale, CA USA is ISO 9001:2008 Certified. Russia/CIS +43 1 333 50 34 0 Singapore +65 6289 1190 Sweden +46 8 556 468 00 Switzerland +41 61 716 77 00 Taiwan +886 2 8751 6655 UK/Ireland +44 1442 233555 USA +1 800 532 4752 PN71117_E 05/14S