Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy. Jean-Pierre Bissonnette, PhD, MCCPM

Similar documents
AAPM IGRT Recommendations

TG-104: In-room kv Computed Tomography for Image-guidance

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Imaging/Imagine Needs for Proton Therapy: Treatment Planning. Lei Dong, Ph.D. Scripps Proton Therapy Center San Diego, CA

Respiratory-phase phase correlated imaging, treatment planning and delivery Clinical Implementation. Disclosures. Session Objectives.

The Role of In-Room kv X-Ray Imaging for Patient Setup and Target Localization (TG104)

Integrated on-board CBCT-US imaging system for soft tissue IGRT and real-time intra-fraction monitoring

Disclosures 7/31/2017. Clinical Impact and Applications of 4D Imaging (in RT)

RT Workspace from Standard Imaging

Disclosures I have provided Physics Consultation Services to Varian and Vision RT

IMAGE GUIDANCE DOSES IN RADIOTHERAPY. Scott Crowe

TG142 Quality Assurance of Medical Accelerators

Stereotactic Body Radiation Therapy: Planning and Delivery

Imaging for Proton Treatment Planning and Verification

Practical Workflow and the Cost of Adaptive Therapy. Rojano Kashani, Ph.D., DABR Washington University School of Medicine March 7, 2015

} } 8/3/2016. Accounting for kv Imaging Dose. kilovoltage imaging devices/techniques. Current imaging dose determination methods

Controlli di qualità in SBRT

Status of ART in the Clinic

Adaptive Re-planning for Lung RT with Multi-targets

Title. CitationPhysica Medica, 30(5): Issue Date Doc URL. Type. File Information.

Evaluation of Dynamic Conformal Arc Therapy for Treatment of Lung and Liver

Quality Assurance for particle beam therapy

UNIVERSITY OF WISCONSIN-LA CROSSE Graduate Studies A WRITTEN DIRECTIVE SURROGATE FOR PHYSICIAN REAL-TIME IGRT APPROVAL

Christie Medical Physics and Engineering (CMPE), The Christie NHS Foundation Trust, Wilmslow Road, Manchester M20 4BX, UK

03/08/2017. Advances and Challenges in Contour QA for Adaptive RT. Objective. Disclosures

Optimizing Adaptive Workflows Using RapidPlan at the Beatson West of Scotland Cancer Centre

Clinical Evaluation of real-time, patient-centric QA using the Delta 4 Discover

Image based Brachytherapy- HDR applications in Partial Breast Irradiation

E2E SBRT Thorax Phantom

MR Guided Radiation Therapy: ViewRay System Sasa Mutic, Ph.D.

MRI-Guided On-line Adaptive Radiotherapy The UCLA Physics Experience. Disclosures 8/3/2016

Image guided Radiation Therapy (IGRT) a perspective

Transitioning to PBS Proton Therapy From IMRT or Passive Scattered Based Proton Therapy

This is probably the kind of radiotherapy that you are used to delivering in your country.

N.MAFFEI, G.GUIDI, C.VECCHI, G.BALDAZZI Physics Department, University of Bologna, via Irnerio Bologna, Italy

ViewRay Experience. Sasa Mutic, Ph.D. Washington University School of Medicine

Strategies for Adaptive RT

SPECIFICATION CT SIMULATOR FOR BLACK LION HOSPITAL, ADDIS ABEBA, ETHIOPIA

Integrating the MR-Linac into Radiation Therapy Practice

RapidArc: Clinical Implementation

EVOLUTION IN 4D TREATMENT

Precise imaging. Personalized treatment. Varian Exact IGRT couch

MV Cone Beam CT Imaging for daily localiza6on: (part II)

I M A G E G U I D E D R A D I A T I O N T H E R A P Y. Elekta XVI. Inspiring clinical confidence

Going Clinical with the Accuray Radixact System. The Montefiore Experience

8/17/2011. Implementation and Quality Assurance Considerations for Volumetric Modulated Arc Therapy (VMAT)*

Initial experience and clinical comparison of two image guidance methods for SBRT treatment: 4DCT versus respiratory-triggered imaging

SBRT LUNG CANCER CLINICAL PATHWAY

Approved [insert date]*

UNIVERSITY OF CALGARY. An X-ray Source Model and Characterization Method. for Computing kv Radiation Dose. Yannick Poirier A THESIS

Vendor Tools and Uses for TG-142

Clinical feasibility of dual tube floor based kilo voltage X-ray localization system for image guided radiation therapy

Towards a SMART proton PBS clinic

Nathan Childress, Ph.D., DABR

Proton Therapy: Application and Quality Assurance

Characterization of interplay errors in step-and-shoot IMRT of the lung

Benchmark & FAQ for NRG BR001: A Phase 1 Study of SBRT for the Treatment of Multiple Metastases

Clinical Experience with Philips Auto-Planning

Clinical trials with hypo-fractionation for breast conserving therapy (BCT): Shorter overall treatment time Effective radiation treatment

An updates on ART for prostate, pancreas and breast

Computational Verification in Interventional Radiation Oncology (Brachytherapy)

Elekta VMAT. Freedom to Treat. Enhanced dose conformance for patient-specific treatment

Real-Time Monitoring of Treatment Delivery. Sonja Dieterich, Ph.D., FAAPM Professor

Proton Therapy Workflow Martijn Engelsman HollandPTC and TU Delft

THE PRECISE ANSWER TO STEREOTACTIC RADIATION THERAPY

Technical White Paper: An Educational Report On Helical IMRT EXECUTIVE SUMMARY. CRCPD Publication #E-11-3

Initial Certification

IMRT Planning: Concepts and Recommendations of the ICRU report n. 83

At that time, do not reference or utilize the initial plan in any manner or study data may be compromised

Thanks to Jeff Williamson, Mike Steinberg, James Purdy

Impact of patient rotational errors on target and critical structure dose in IMRT: A 3D simulation study

BodyFIX Maximize positional accuracy. Support delivery with precise patient positioning and immobilization

DESIGN OF AN ANTHROPOMORPHIC INTENSITY MODULATED RADIATION THERAPY QUALITY ASSURANCE PHANTOM

Acceptance testing, commissioning and routine QA of the Varian TrueBeam Richard Popple

Quality assurance: Fundamental reproducibility tests for 3D treatment-planning systems

Proposal on Guideline for Quality Assurance of Radiation Treatment Planning System

Part 3 Oral Exam Content Guide

Target point correction optimized based on the dose distribution of each fraction in daily IGRT

AAPM TG-275: Live Demonstration of Chart Checks

Analisi della letteratura

AAPM Medical Physics Practice Guideline 8.a.: Linear accelerator performance tests

Introduction to Radiation Therapy : Basics, and Recent Technologies

Assessing dose variance from immobilization devices in VMAT head and neck treatment planning: A retrospective case study analysis

Quality assurance for clinical implementation of an Optical Surface monitoring system

8.0 Technical Guidelines. 8.1 Statement of treatment aim

ArcCHECK & 3DVH. The Ultimate 4D Patient QA Solution. Your Most Valuable QA and Dosimetry Tools

8/1/2018. Optimization Strategies for Pediatric CT Imaging. Educational Objectives. Brain CT

Research Article On-Line Adaptive Radiation Therapy: Feasibility and Clinical Study

Tra n s m is s ion dete c to r fo r a t-t re a t m e n t Q A

VMAT monthly QA using two techniques: 2D ion chamber array with an isocentric gantry mount and an in vivo dosimetric device attached to gantry

Clinical experience of the importance of daily portal imaging for head and neck IMRT treatments

Multiple Metastasis Insert for the Lucy 3D QA Phantom

Process Mapping. Acknowledgements 7/21/2014. Disclosures

Autoplanning in the IGRT era. Ben Heijmen

AQUA. Data-backed Assurance

RADIATION SAFETY ISSUES LINKED TO THE OMNIPRESENCE OF COMPUTERS

Knowledge-based treatment planning: fundamentally different, or more of the same?

The application of error reduction QA philosophy in HDR brachytherapy

GENESIS Edition. Transforming CT

Guidance for the Physics Aspects of Clinical Trials

Transcription:

Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy Jean-Pierre Bissonnette, PhD, MCCPM

Disclosure Work supported, in part, by Elekta Oncology Systems Commercial Interest in Penta-Guide Phantom, Modus Medical Inc. Chair, AAPM TG-179

Learning Objectives Brief review of the reasoning behind IGRT How does IGRT improve quality? Identify key components of CBCT QA A glimpse of AAPM TG-179 report IGRT as a QA tool: case studies on lung RT

Image-Guided Radiation Therapy Frequent imaging during a course of treatment as used to direct radiation therapy It is distinct from the use of imaging to enhance target and organ delineation in the planning of radiation therapy.

Justification for IGRT Accuracy: verify target location (QA) + Precision: tailor PTV margins (patient-specific) Adaptation to on-treatment changes Correct & moderate setup errors Assess anatomical changes Re-planning ( naïve or explicit) + +

Issues around IGRT Reduction of positional errors Margin definition 3D vs 2D Deformable anatomy Patient throughput Staff training and workload Dose Applicability

Rationale against IGRT Increased complexity New information New issues Patient dose Redefining workload (more?) Therapy, Physics, Oncology Time Resources/Infrastructure

IGRT reduces geometric errors Random (σ) Systematic (Σ) Margin: = + van Herk approach: 2.5Σ + 0.7σ 95% confidence interval: One size does not fit all!

IGRT reduces geometric errors If Systematic Errors are minimized, margins can be reduced

IGRT reduces geometric errors Minimizing Systematic and Random Errors can allow even smaller margins Treated volume r 3 Requires on-line image guidance

ICRU 62 Margin Construction Adapt Setup margin reduced to spare normal tissues Thoughtful Combination CTV GTV Subclinical Involvement Internal Margin Setup Margin Conservative Addition CTV

QA: CBCT systems Geometric performance Image quality Lung IGRT

Structure of the presentation Geometric accuracy = distance to shift is physically correct Image quality = highest quality images are available for IGRT Lung IGRT: improving quality of lung radiation therapy

QA: Devices and Processes Geometric Performance Image quality Lung IMRT

Geometric accuracy Three worlds Cone beam (kv) space Megavoltage (MV) space Patient (ME) Space Objective is to ensure that the geometry of each is the same

MV Geometry Treatment and imaging beams are coincident

MV/kV Coincidence Treatment and imaging beams are orthogonal

kv/mv Calibration Concept BB (reconstruction Isocentre) MV mechanical isocentre kv y MV radiation isocentre z x Calibrated isocentre

1. MV Localization (0 o ) of BB; collimator at 0 and 90 o. 2. Repeat MV Localization of BB for gantry angles of 90 o, 180 o, and 270 o. +1mm 3. Analyze images and adjust BB to Treatment Isocentre (± 0.3 mm) θ gantry -1mm -180 u v θ gantry +180 θ gantry 4. Measure BB Location in kv radiographic coordinates (u,v) vs. θ gantry. 5. Analysis of Flex Map and Storage for Future Use. Reconstruction 6. Employment of Flex Map During Routine Clinical Imaging.

Geometric Calibration Analogous to the Winston-Lutz test used for brain stereotactic QA Lutz, Winston, & Maleki, IJROBP 14, pp. 373-81 (1988)

MV/kV Calibration Procedure v u

Flexmap A plot of the apparent travel of a point as a function of gantry angle. Removes the effect of component flexes and torques prior to reconstructions. Ties the 3D image matrix to the radiation isocentre of the accelerator.

Results for Six Units Absolute U displacement (mm) 5 4.5 4 3.5 3 2.5 2 1.5 1 0.5 0-0.5-1 -1.5-2 -2.5-3 -3.5-4 -4.5-5 -180-135 -90-45 0 45 90 135 180 Gantry angle (degrees) Unit 7 Unit 8 Unit 9 Unit 10 Unit 14 Unit 17

Effect of Incorrect Calibration

Daily Geometry QA Align phantom with lasers Acquire portal images (AP & Lat) & assess central axis Acquire CBCT Difference between predicted couch displacements (MV & kv) should be < 2 mm

Daily Geometry QA Align phantom with lasers Acquire portal images (AP & Lat) & assess central axis Acquire CBCT Difference between predicted couch displacements (MV & kv) should be < 2 mm

Compare Portal Image & DRR

MV vs kv Shift Data Deviation from isocentre (mm) 2.5 2.0 1.5 1.0 0.5 0.0-0.5-1.0-1.5-2.0 X Y Z -2.5 A B C D G Unit

QA: Devices and Processes Geometric Performance Image quality Lung IMRT

Summary of image analysis tests Scale CT number linearity Spatial resolution

Image quality AAPM CT Performance Phantom CatPhan 500 phantom

Scale Geometric calibration to tie isocentre to centre of volumetric reconstruction 5 cm Scale to relate all pixels to isocentre

Scale Geometric calibration to tie isocentre to centre of volumetric reconstruction Scale to relate all pixels to isocentre

CT Numbers - calibration

CT numbers: 7 units 2000 1800 1600 Measured Hounsfield unit 1400 1200 1000 800 600 Unit 7 Unit 8 Unit 9 Unit 10 Unit 12 Unit 16 Unit 16 with annulus Unit 17 400 200 0 0 200 400 600 800 1000 1200 1400 1600 1800 2000 Theoretical Housfield unit

Spatial Resolution 1.2 1.0 Mean Elekta Mean Varian 0.8 MTF 0.6 0.4 0.2 0.0 0 2 4 6 8 10 12 Spatial frequency (cm -1 ) Droege, Radiology 146, pp. 244-246 (1983)

Effect of Scatter on MTF Acceptance Narrow x-ray field Window & level identical

Spatial Resolution 1.2 1.0 Mean Elekta Mean Varian 0.8 1.4 mm MTF 0.6 0.4 1.2 mm Resolving Power 0.2 0.0 0 2 4 6 8 10 12 Spatial frequency (cm -1 )

Spatial Resolution 1.2 1.0 Mean Elekta Mean Varian Unit E small field 0.8 0.8 mm MTF 0.6 0.4 1.4 mm Resolving Power 0.2 0.0 0 2 4 6 8 10 12 Spatial frequency (cm -1 )

Quality Control: AAPM TG-179 Daily tests Safety Quality Metric System operation and accuracy Quality Check Interlocks: interrupts or prevents irradiation Warning lights Laser/image/treatment isocentre coincidence OR Phantom localization and repositioning with couch shift Tolerance Functional Functional ± 2 mm ± 2 mm

Quality Control: AAPM TG-179 Monthly tests, or after major upgrade Quality Metric Geometry Image quality If used for dose calculations Quality Check Geometric calibration maps* OR kv/mv/laser alignment Couch shifts: accuracy of motions Scale, distance, and orientation accuracy* Uniformity, noise* High contrast spatial resolution* Low contrast detectability* CT number accuracy and stability* Tolerance Replace/refresh ± 1 mm ± 1 mm Baseline Baseline 2 mm (or 5 lp/cm) Baseline Baseline

Quality Control: AAPM TG-179 Annual tests Dose Quality Metric Imaging system performance Geometric System operation Quality Check Imaging dose X-ray generator performance (kv systems only): tube potential, ma, ms accuracy and linearity AP, ML, and CC orientations Planning of resources Tolerance Baseline Baseline Accurate Support clinical use

QA: Devices and Processes Geometric Performance Image quality Lung IMRT

IGRT: Improving Quality of Lung RT Experience with CBCT for lung is positive Excellent bone visualization Improved soft tissue visualization Lower doses Fast, less ambiguous image registration Questions were raised: How often should I image? What is a more stable setup? What about patient dose? Oh no, the patient has moved I think!

IGRT: Improving Quality of Lung RT Published articles: Acta Oncol. 2006;45(7):915-22 Int J Radiat Oncol Biol Phys. 2007 May 1;68(1):243-52 Clin Oncol (R Coll Radiol). 2007 Apr;19(3 Suppl):S10 J Thorac Oncol. 2008 Nov;3(11):1332-41 Int J Radiat Oncol Biol Phys. 2009 March; 73 (3): 927-934

A Case Study: How Often to Image? CBCT offers advantages over portal imaging Excellent bone visualization Improved soft tissue visualization Lower doses Fast, less ambiguous image registration 2D MV 3D kv-cbct

A Case Study: How Often to Image? However, some concerns are raised (risk analysis): Imaging dose to a large volume from daily CBCT? Extra time spent in treatment suite due to image guidance? Can we get away with less frequent imaging?

A Case Study: How Often to Image? Retrospective study (REB approval) Identified 100 stage III lung cancer pts Radical RT Daily CBCT acquired (4237 scans) Localization CBCT Verification CBCT when couch shift is performed Using this data, simulate 4 image guidance approaches Higgins et al, IJROBP, in press

Methods: Simulated IG approaches No IG 1 st 5 Days Represents Skin Tattoo Setup; no couch adjustments made Average Setup Error for 1st 5-Days; apply average shift to remaining fractions (6+) Weekly IGRT performed weekly Alternate Day IGRT performed every other day

Systematic Setup Error ( ) No IG 1st 5-Day IG Weekly IG RL SI AP Alternate IG Daily IG 0 0.5 1 1.5 2 2.5 3 Setup Error (mm)

Random Setup Error (σ) No IG 1st 5-Day IG Weekly IG RL SI AP Alternate IG Daily IG 0 0.5 1 1.5 2 2.5 3 3.5 4 Setup Error (mm)

Uniform Setup Error 15.0 10.0 Residual Error (mm) RL 5.0 0.0-5.0-10.0-15.0 1 3 5 7 9 11 13 15 17 19 21 23 25 Fractions No IG 1st 5 Fractions IG Daily IG

Extreme Setup Error 15.0 10.0 Residual Error (mm) RL 5.0 0.0-5.0-10.0-15.0 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 Fractions No IG 1st 5 IG Daily IG

Population-Based Setup Margins No IG 1st 5-Day IG RL Weekly IG Alternate IG SI AP Daily IG 0 1 2 3 4 5 6 7 8 9 Setup Margin (mm) van Herk Margin Recipe used to approximate PTV margins (2.5 + 0.7σ) van Herk M et al. IJROBP, 2000

Findings No IG & 1 st 5-Day IG No reduction Random Error (σ) Potential introduction further Systematic Error ( ) For many Lung patients, setup on a given day is not predictive of the next day Weekly IG & Alternate Day IG Some reduction Systematic Error ( ) but little reduction Random Error (σ) Consideration if adequate PTV margins used (5-7 mm)

Findings Daily IG Substantial reductions of residual setup error ( and σ) Anisotropic setup margins 3-4mm Potential to treatment toxicity w/out compromising target coverage Daily IG Standard Practice at PMH

A Case Study: What Immobilization is More Stable for Lung SBRT? SBRT lung patients spend a long time on the treatment couch. Between 9-13 treatment beams Long beam-on times (12-20 Gy per fx) Couch kicks (2 sagittals) Is patient assessment done at beginning of treatment representative of the whole treatment?

A Case Study: SBRT Lung Retrospective study (REB approval) Identified 108 stage I-II lung cancer pts SBRT Image guidance Localization CBCT Verification CBCT when couch shift is performed Mid-treatment CBCT, before couch kicks End treatment CBCT Immobilization Evacuated cushions Evacuated cushions + abdominal compression Chest board Li et al, IJROBP, in press

SBRT Lung: IGRT Workflow Patient immobilized Setup using skin marks Acquire Localization CBCT Image Registration Target to ITV, 3 mm tolerance Acquire Verification CBCT Treat all co-planar beams Acquire Mid-treatment CBCT Assess Target to PTV Treat all non co-planar beams Acquire Post-treatment CBCT

Overall Positional Accuracy 120% Frequency of fractions within tolerance 100% 80% 60% 40% 20% 0% 99% 88% 94% 68% 39% 15% End treat 0 1 2 3 4 5 6 7 8 9 10 Tolerance (mm) Localization Verification Midway 133 patients, 2047 CBCT scans analysed

Accuracy vs Immobilization Device Localization (solid) + Verification (dotted) CBCT 100% Frequency of fractions within tolerance 90% 80% 70% 60% 50% 40% 30% 20% 10% 0% 0 1 2 3 4 5 6 7 8 9 10 Tolerance (mm) Vakloc Vakloc + compression Chestboard Overall Vaclok Vakloc + compression Chestboard Overall

Accuracy vs Immobilization Device Midway (solid) + Verification (dotted) CBCT 100% Frequency of fractions within tolerance 90% 80% 70% 60% 50% 40% 30% 20% 10% 0% 0 1 2 3 4 5 6 7 8 9 10 Tolerance (mm) Vakloc Vakloc + compression Chestboard Overall Vaclok Vakloc + compression Chestboard Overall

Accuracy vs Immobilization Device End treat (solid) + Verification (dotted) CBCT 100% Frequency of fractions within tolerance 90% 80% 70% 60% 50% 40% 30% 20% 10% 0% 0 1 2 3 4 5 6 7 8 9 10 Tolerance (mm) Vakloc Vakloc + compression Chestboard Overall Vaclok Vakloc + compression Chestboard Overall

Findings Initially, accuracy is excellent and independent from immobilization device Evacuated cushion + abdominal compression is most consistent Chestboard pts tend to drift out of position faster Can correct for this drift with additional couch shifts Can limit drift by reducing time spent on couch (i.e. VMAT/RapidArc)

Conclusions Need to establish a program for QA of CBCT systems QA program aligned with your objectives Primary focus on geometric accuracy and precision, secondary focus on image quality Quality principles can be used to improve the quality (i.e., accuracy) of clinical processes.