Quantification of Drug Metabolites in Early-Stage Drug Discovery Testing

Similar documents
Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Nano LC at 20 nl/min Made Easy: A Splitless Pump Combined with Fingertight UHPLC Nano Column to Boost LC-MS Sensitivity in Proteomics

High-speed, High-resolution Oligonucleotide Separations Using Small Particle Anion-Exchangers

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

BIOANALYTICAL STRATEGY FOR IN VITRO METABOLITE SCREENING WITH EXACT MASS USING THE Q-Tof micro. Jose M. Castro-Perez 1, Carina Leandersson 2

Implementation of a Walk-Up High- Pressure Capillary Ion Chromatograph for the Fast Separation of Pharmaceutical Relevant Inorganic Anions and Cations

Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry

High-Resolution Separation of Intact Monoclonal Antibody Isoforms

Thermo Scientific EASY-Spray Technology. Plug-and-spray with. State-of-the-art performance

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

electrochemical detection

Passion. Power. Productivity. Corporate Headquarters

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants

A Novel ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge-Variant Analysis

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

Metabolite ID. Introduction

The Benefits of SOLAμ Technology in Sample Preparation. Jon Bardsley and Ken Meadows Thermo Fisher Scientific, Runcorn, UK

Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation

Method Optimisation in Bottom-Up Analysis of Proteins

Minimizing Carbon Dioxide Contamination in Anion Reagent-Free Ion Chromatography

Introduction. Yan Liu, Victor Barreto, Rosanne Slingsby, Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA

High-Resolution Charge Variant Analysis for Top-Selling Monoclonal Antibody Therapeutics Using a Linear ph Gradient Separation Platform

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

The Quantitative Analysis of Curcuminoids in Food and Food Additives Using Rapid HPLC With Electrochemical, UV, or Fluorescence Detection

Optimize Your Process Operations by Improving Process Monitoring

Hillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION

Thermo Scientific EASY-nLC 1000 System. Effortless, split-free nanoflow UHPLC Top performance in LC/MS

Analysis of Monoclonal Antibody and Related Substances using a New Hydrophobic Interaction Chromatography (HIC) Column

TechnicalNOTE METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES INTRODUCTION

Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography

Identification of RNA Linkage Isomers by Anion-Exchange Purification with ESI-MS of Automatically-Desalted Phosphodiesterase-II Digests

NISTmAb characterization with a high-performance RP chromatography column

Measurement and Control of Copper Additives in Electroplating Baths by High-Performance Liquid Chromatography

Benefits of 2D-LC/MS/MS in Pharmaceutical Bioanalytics

for water and beverage analysis

Emulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture

Using HRAM Survey Analysis Combined with Rapid MS2 Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition

[ VION IMS QTOF ] BEYOND RESOLUTION

Analysis of Fracking Flowback Water From the Marcellus Shale Using In-line Conductivity, Automated Dilution, and Ion Chromatography

Thermo Scientific Dionex ICS-900 Ion Chromatography System

New Developments in Capillary Ion Chromatography Systems with Electrochemical Detection and Their Applications

of peptide mapping chemistries, 130Å or 300Å with tunable UV, photodiode array, or mass spectrometry

Separation Science Built for Biopharma

A Simple Rapid and Sensitive Method Development for Quantification of Quetiapine Fumarate in Bulk and Dosage Forms Using RP-HPLC

FRACTIONLYNX APPLICATION MANAGER

Direct Carbohydrate Analysis in Beverages and Foods Using Pulsed Amperometric Detection or Charged Aerosol Detection

Key Words Q Exactive Focus, SIEVE Software, Biomarker, Discovery, Metabolomics

Direct Carbohydrate Analysis in Beverages and Foods Using Pulsed Amperometric Detection or Charged Aerosol Detection

Determination of Manganese in Brine ( 30% NaCl)

Jonathan R. Beck and Charles T. Yang; Thermo Fisher Scientific, San Jose, CA

Carbohydrate Analysis in Beverages and Foods Using Pulsed Amperometric Detection or Charged Aerosol Detection

Technical Overview. Author. Abstract. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

A Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS

ph gradient analysis of IgG1 therapeutic monoclonal antibodies using a 5 µm WCX column

Chromatography Column Performance and Data Analysis Success Guide. Hints and Tips for Better Purifications

Simultaneous in vivo Quantification and Metabolite Identification of Plasma Samples Using High Resolution QTof and Routine MS E Data Analysis

Performance characteristics of the 1260 Infinity Quaternary LC system

How discovery activities can influence metabolic profiling in the regulatory space? C. DELATOUR EBF, 25 th September 2015

Impurity Control in the European Pharmacopoeia

Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution

Parallel LC with Capillary PS-DVB Monolithic Columns for High-Throughput Proteomics

Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column

Lifetime stability of size exclusion chromatography columns for protein aggregate analysis

Nitrogen Determination in Soils and Plants by Flash Combustion using Argon as Carrier Gas

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

The system has an optimized porosity distribution and acts as a depth filter, retaining the precipitated protein without well blockage.

BIOANALYSIS CONSUMABLE SOLUTIONS. Sample Preparation and Liquid Chromatography Solutions for Quantitative BIOANALYSIS

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

ENHANCED MS WITH THE TURN OF A KEY

Thermo Scientific MAbPac HIC Columns. Novel Hydrophobic Interaction HPLC Columns. Designed for Monoclonal Antibody Analysis

Analysis of Nucleosides Using an Agilent Infinity II High Speed UHPLC with the 6130 Single Quadrupole Mass Selective Detector

Introduction. Three different cases for different dosage forms and pharmacopeias, respectively, must be distinguished (Table 1).

Developing Purification Strategies for the Agilent 1260 Infinity II Preparative LC/MSD System

Techniques for Improving the Efficiency of Large Volume Loading in Preparative Liquid Chromatography

Thermo Scientific EASY-nLC 1200 System. Leading in simplicity. and performance

Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS

Application Note. Author. Abstract. Pharmaceuticals. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Method Transfer from an Agilent 1200 Series LC to an Agilent 1260 Infinity II LC

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

Laboratory Reagents. for life science research.

High-Throughput LC/MS Purification of Pharmaceutical Impurities

A universal chromatography method for aggregate analysis of monoclonal antibodies

P. Wadhwani College of Pharmacy, Yavatmal , India. *Corres.author: Cell No

Highlighted Capillary IC Applications

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

Analysis of Monoclonal Antibodies and Their Fragments by Size Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Simplifying Methods Transfer: Novel Tools for Replicating Your Established Methods on an ACQUITY Arc System

VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY *)

Informatics and High Resolution QTof MS. How can we ask better questions and get better answers in DMPK? Mark D. Wrona

Effect of Temperature on IC Signal Noise

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water. Pranathi R Perati Thermo Fisher Scientific, Sunnyvale, CA, USA

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns

Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins

In-Cell and External Hydrolysis of Biomass for Carbohydrate Determinations Using Automated Solvent Extraction

columns PepSwift and ProSwift Capillary Monolithic Reversed-Phase Columns

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

Transcription:

Quantification of Drug Metabolites in Early-Stage Drug Discovery Testing Christopher Crafts, Bruce Bailey, John Waraska, Ian Acworth Thermo Fisher Scientific, Chelmsford, MA, USA

Overview Purpose: The use of a high efficiency UHPLC column and charged aerosol detection were evaluated for the quantification of drug metabolites. Methods: A 2.1 250 mm 2.2 µm UHPLC C18 column with a simple water/acetonitrile gradient was used to compare the performance of charged aerosol detection vs. traditional UV detection techniques. Results: A more universal response for the majority of the drugs and their metabolites was achieved by charged aerosol detection with an inverse gradient when compared to traditional UV techniques. The reproducibility and sensitivity of the two detectors were similar. Charged aerosol detection is an important complementary technique to those already being used during the in vitro drug development stage of testing. Introduction Due to new regulations, the identification and quantification of drug metabolites in early-stage drug development testing has become an essential part of the drug development process. Analytical systems typically use reversed phase HPLC or UHPLC with diode array and high resolution mass spectrometry to detect drug metabolites. For many samples however, differences in UV absorbance or inability to ionize can lead to incorrect estimation of metabolite levels. Unfortunately, this over- or under-estimation is typically only revealed at later stage testing when radiochemical approaches are used. In late-stage testing, differences in metabolite quantity can be detrimental to the acceptance process and extremely costly to rectify. Presented here is an orthogonal technique that uses an inverse gradient with the Thermo Scientific Dionex Corona ultra RS charged aerosol detector (CAD ) incorporated into the UV/MS system to overcome most quantification issues. Charged aerosol detection is a highly sensitive universal technique that offers near uniform response for non-volatile analytes. 1 Like with all nebulization-based techniques, analyte response changes during gradient elution due to changes in nebulization efficiency. This can be rectified by using solvent compensation or an inverse gradient after the analytical column but before the detector. 2 Another challenge with the analysis of drug metabolites is the need for a fast and robust method capable of resolving the API and metabolites for an array of compounds. In early-stage development it is often not possible to optimize HPLC conditions or detector settings from sample to sample. For this reason a long UHPLC column with an acidic water/acetonitrile gradient was chosen for the analysis. The 2.2 µm 250 mm column has sufficient i resolving capacity even with injections volumes of 30 µl. Previous work highlighted the use of charged aerosol detection for the analysis of buspirone and erythromycin and their metabolites without inverse gradient compensation. 3 Although both the Corona ultra RS charged aerosol detector and low wavelength UV detector showed similar results, late eluting analytes typically y showed an increase in response when measured by the CAD. In later studies, this response issue was overcome using an inverse gradient. 4 For the work presented in this poster, an inverse gradient was used for all analyses, thereby keeping solvent composition to the CAD constant and analyte response uniform. In the first part of the study, several parent drugs and metabolite standards were prepared at equal mass quantities and were analyzed using a CAD and an UV detector. The sensitivity and consistency of response were compared for several compounds including acetaminophen and its metabolites. In the second part of the study, drugs were incubated with liver microsomes and then evaluated for mass balance (pre vs. post incubation). The reproducibility and sensitivity of the Corona ultra RS is discussed for the analysis of standards d and in vitro samples. 2 Quantification of Drug Metabolites in Early-Stage Drug Discovery Testing

Methods Sample Preparation Samples were provided already prepared by a major pharmaceutical company for evaluation of the UHPLC/CAD platform. This included standards for active pharmaceutical ingredients (APIs), acetaminophen, clozapine, efavirenz and ranitidine, and metabolites which were prepared at 10 µg/ml in acetonitrile/water solution. Ten micromolar clozapine pre- and post-incubation in rat liver microsomes were also included. The samples were extracted with acetonitrile prior to analysis. Liquid Chromatography System: Column: Thermo Scientific Dionex UltiMate 3000 RSLC system equipped with: DGP-3600RS, WPS-3000TRS, TCC-3000RS, DAD-3000RS, Corona ultra RS Thermo Scientific Acclaim RSLC 120 C18, 2.1 250 mm, 2.2 µm 0.1% Formic acid in DI water 0.1% Formic acid in acetonitrile 0.75 ml/min pumps left and right. Mobile Phase A: Mobile Phase B: Flow Rate Gradient Conditions Analytical: 5% B hold 0.1 min, 95% B in 10 min hold for 2, return and equilibrate for 4. Gradient Conditions Inverse: 95% B hold 1.1 min 5% B in 10 min hold for 2, return. Injection Volume: 10 µl for standards, 30 µl for microsome samples. CAD settings: Filter: 3 Nebulizer Temp On at 25 C Power Function Value 1.0 Flow Diverted 0.6 1.1 minutes Mass Spectrometry Mass Spectrometry was not used for this evaluation, see discussion of microsome results. Data Analysis Thermo Scientific Dionex Chromeleon 7.1 SR.1 Chromatography Data System (CDS) was used for all data collection and processing. Results Analysis of 10 µg/ml Standards The major metabolites of interest vary depending on the API and its metabolism in the body. In addition to the in vitro incubation with microsomes or hepatocytes some compounds undergo further testing for bioactivation and idiosyncratic toxicity using a glutathione conjugation assay. 5 The adduct formation often significantly increases molecular weight of the compound from the API and can have varying impact on the UV absorbance depending on the conjugate and its UV chromophore. To evaluate this problem acetaminophen-glutathione conjugates were prepared and analyzed at concentrations of 10 µg/ml on the UHPLC system with UV (at 210 and 254 nm) detection and charged aerosol detection. Figure 1 shows the overlaid chromatogram of the API and three metabolites by charged aerosol detection. FIGURE 1: Overlay of acetaminophen and three metabolites at 10 µg/ml (100 ng on-column) measured by charged aerosol detection. Response [pa] 3 Quantification of Drug Metabolites in Early-Stage Drug Discovery Testing

A comparison of the peak areas obtained by the CAD (Figure 1) to those from the two UV wavelengths is presented in Figure 2. Charged aerosol detection is mass sensitive and therefore no data manipulation is required as equal masses of each metabolite were analyzed. UV detection, on the other hand is concentration dependent and will show changes in molar concentration due to differences in MW between the different metabolites. In Figure 3, UV area values are presented corrected for concentration. The results show the consistency of response with charged aerosol detection with a RSD across the four areas of only 5.1%. This is in contrast to the 11.5 and 58% RSD values obtained for the UV at 254 and 210 nm, respectively. FIGURE 2: Response area for charged aerosol detection and the two UV channels for 10 µg/ml samples of acetaminophen and three metabolites. (UV areas are raw area/mw * 10 to correct for MW and to scale properly). CAD UV @ 254nm Corrected UV @210nm corrected 1 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 acetaminophen acetaminophen Glucuronide acetaminophen Sulfate acetaminophen Glutathione The results using the specific UV wavelength of 254 nm for acetaminophen showed acceptable deviation levels in this case. However, low wavelength UV which typically offers a more universal or consistent response and is the preferred wavelength for these experiments had a response deviation of 58%. Even in cases were the response deviation is more in line, the sensitivity of low wavelength UV is often insufficient, as was the case for the analysis of ranitidine. Figure 3 illustrates the analysis of 8-hydroxyefavirenz using the same detection parameters. In this case, two peaks were observed by all three detection techniques. The charged aerosol detector trace on top and the UV at 210 nm trace on the bottom are in close agreement to the relative abundance, with peak 1 being 54% by charged aerosol detector and 58 % by UV at 210 nm. However er the middle trace using UV at 254 nm has peak 1 with a relative e abundance of 92%. Such discrepancies would be very difficult to resolve by UV alone without the use of authentic standards for each peak. Data agreement of the CAD with one of the two other UV results provides sufficient confirmation and speeds up the analysis without need for further investigation or the use of authentic standards. FIGURE 3: Overlay of chromatograms for analysis of 8-hydroxyefavirenz by charged aerosol detection and UV detection. CAD UV @ 254 nm UV @ 210 nm 4 Quantification of Drug Metabolites in Early-Stage Drug Discovery Testing

Microsome Samples The goal in early-stage drug development is to obtain in vitro results that are similar to the in vivo results in later stage testing. It is often not the expectation of a DMPK chemist to obtain 100% recovery or accuracy of the results in this early-stage testing. General practice requires accuracy levels of ~ 20% of the true value to allow confidence in the technique. While this value is fairly large, no single technique is able to deliver these results for every analyte with the sensitivity and turnaround time needed. This is the main reason why a more complex system with multiple detection techniques is required. As presented in our previous report, 3 this approach would normally require a high h resolution MS to identify the metabolites. The MS is used in parallel with the UV/charged aerosol detection using a split of 90% post UV (150 µl flow to the MS). However, as the metabolites were already identified and provided for the current study, the MS is not required. To illustrate the power of this approach, data for the analysis of the antipsychotic medication clozapine (10 µm) are presented in Figure 4. FIGURE 4: Metabolism of clozapine by rat liver microsomes. Data from charged aerosol detection: Top pre-incubation (T=0); Bottom post-incubation (T=60). CAD T=0 nse [pa] Response [pa] CAD T=60 Respon Table 1: Comparison of Area results pre- and post-incubation for clozapine sample. Incubation Point Total Raw Area Value (Minutes) CAD UV @ 254 UV @ 210 T= 0 0.4539 4.677 6.68496849 T= 60 0.3574 2.9771 4.3853 % Recovery 79% 64% 66% Following metabolism by rat liver microsomes, the clozapine peak was significantly decreased, whereas the two major metabolites, n-oxide clozapine (peak 4) and n-desmethyl clozapine (peak 9) significantly increased. The peaks not present in the matrix blank or in the non-incubated sample were used to calculate an area recovery for all three detection techniques. The results using charged aerosol detection were close to the expected results as shown in Table 1, with a 79% area recovery. Sensitivity and reproducibility are important factors when weighing the effectiveness of quantification techniques for drug metabolites. Figure 5 shows an overlay of three injections of the clozapine sample at T=0. The relative standard deviation for the clozapine peak in the matrix for 3 interlaced injections was 1.4% by the CAD and 1.3% by UV @ 210 nm. The results for the residual clozapine in the T=60 minutes for three interlaced injections (data not shown) was 2.4% with an average signal-to-noise (S/N) of 14/1. The results for the same injections by UV at 210 nm was 4.2% with a S/N of ~ 20/1. The estimated mass on-column for the residual clozapine in the T=60 minute sample is 7 ng on-column. Charged aerosol detection results were still in line with those for the UV technique even at these low levels (near the limit-of-detection) with the added benefit of the near universal response for the non-volatile API and metabolites. 5 Quantification of Drug Metabolites in Early-Stage Drug Discovery Testing

FIGURE 7. Overlay of three interlaced injections of clozapine pre-incubation samples obtained with charged aerosol detection. 16.0 Clozapine 10.0 5.0 0.0-2.0 0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 Time [min] Conclusion The inclusion of the Corona ultra RS detector as part of a UHPLC/UV/MS system provides another piece of valuable information to help confirm or quantify drug metabolites in early-stage testing. The addition of an inverse gradient post-column prior to the CAD and MS detectors is not required but improves the inter-analyte response of these nebulizer-based detectors. The sensitivity and reproducibility of the technique was sufficient to analyze the major and minor metabolites formed when clozapine (10 µm) was incubated with rat liver microsomes. Charged aerosol detection is an important complementary technique to those already being used during the in vitro drug development stage of testing. When analyzing compounds at specific wavelengths, charged aerosol detection helps confirm UV absorbance characteristics of APIs and their metabolites. Charged aerosol detection, while not a stand-alone solution, provides additional data to make accurate interpretations of in vitro data without excessive cost or time requirements. References 1. Górecki, T.; Lynen, F.; Szucs, R.; Sandra, P. Universal Response in Liquid Chromatography Using Charged Aerosol Detection. Anal. Chem. 2006, 78, 3186 3192. 3192 2. DeLand P.; Waraska J.; Crafts C.; Acworth I.; Steiner F.; Fehrenbach T. Improving the Universal Response of Nebulization-Based UHPLC Detection. LC/GC. 2011, April Supplement 45-49. 3. Cai, H.; Crafts, C.; Ramanathan, R.; Humphreys, W.; Bailey, B.; Josephs J.; A Novel Detection Technology: Charged Aerosol Detection (CAD) Coupled with HPLC,UV, and LTQ-Orbitrap MS for Semi-Quantitation of Metabolites in Drug Discovery Metabolism Studies, ASMS oral presentation, Salt Lake City, May, 2010. 4. Crafts, C.; Plante, M.; Bailey, B.; Malek, G.; Neely, M. Use of Charged Aerosol Detection as an Orthogonal Quantification Technique for Drug Metabolites in Safety Testing (MIST). Eastern Analytical Symposium poster, 2011. 5. Nedderman, A., Metabolites in Safety Testing: Metabolite Identification Strategies in Discovery and Development. Biopharm. Drug Dispos. 2009, 30, 153-162. 6 Quantification of Drug Metabolites in Early-Stage Drug Discovery Testing

www.thermofisher.com/dionex 2016Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. U.S./Canada (847) 295 7500 Brazil (55) 11 3731 5140 Austria (43) 1 616 51 25 Benelux (31) 20 683 9768 (32) 3 353 42 94 Denmark (45) 36 36 90 90 France (33) 1 39 30 01 10 Germany (49) 6126 991 0 Ireland (353) 1 644 0064 Italy (39) 02 51 62 1267 Sweden (46) 8 473 3380 Switzerland (41) 62 205 9966 United Kingdom (44) 1276 691722 Australia (61) 2 9420 5233 China (852) 2428 3282 India (91) 22 2764 2735 Japan (81) 6 6885 1213 Korea (82) 2 2653 2580 Singapore (65) 6289 1190 Taiwan (886) 2 8751 6655 PN70036_E 09/16S