Optimizing the Purification of a Standard Chiral Compound Utilizing a Benchtop, Multi-Purpose, Semi-Preparative to Preparative HPLC System

Similar documents
A Practical User Guide for the Determination of Optimal Purification Gradients for the Gilson PLC 2020

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

Encompassing Method Development, Small Scale Purification, and Post-Purification Analysis with the Investigator SFC System

Preparative HPLC is still the

A Simple Rapid and Sensitive Method Development for Quantification of Quetiapine Fumarate in Bulk and Dosage Forms Using RP-HPLC

What differentiates Dynamic Extractions?

Simple Techniques for Improving the Isolation of Synthetic Peptides Jo-Ann Jablonski Principal Scientist Waters Corporation

Streamlined Purification of Assay-Ready Oligonucleotides by Automated HPLC

A Comparison of Compendia Normal Phase LC Methods Run Under Supercritical Fluid Chromatography Conditions

High-Throughput LC/MS Purification of Pharmaceutical Impurities

Improving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract

Fast Preparative Column Liquid Chromatography (PCLC)

Optimizing Purified Sample Recovery on Reverse-Phase HPLC Systems

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

Effective Scaling of Preparative Separations Using Axial Compression Columns

Overview of preparative HPLC. Analytical Technologies Limited

Peptide Mapping: A Quality by Design (QbD) Approach

Bivalirudin Purification:

Preparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions

HPLC/FLASH Purification ADVANCE THE PACE OF YOUR DISCOVERY TO ACHIEVE THE RESULTS OF TOMORROW

Agilent 218 Purification System. Purify your way

Quick Manual to the Waters UPLC System

Preparative Purification of Corticosteroids by HPLC; Scalability and Loadability Using Agilent Prep C18 HPLC Columns Application

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Injection by Extraction in Preparative SFC; Isolation of Impurities and Natural Products

Practical Aspects of Preparative HPLC in Pharmaceutical Development and Production

TSK-GEL BioAssist Series Ion Exchange Columns

CHAPTERS 1, 2 and 3 CHAPTER-4 CHAPTER-5,

Agilent SD-1 Purification System. Purify your way SD-1

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Introduction Ron Majors is a Senior Scientist at Agilent. Bill Champion is a chemist in Agilent s HPLC Columns tech support group.

Flexibility with. Preparative HPLC

Emulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

PLC Purification Systems

Scale-up with the Agilent SD-1 Purification System analytical and preparative runs on a single system

Method Transfer from an Agilent 1260 Infinity LC to an Agilent 1260 Infinity II LC

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis

Automated QbD-Based Method Development and Validation of Oxidative Degraded Atorvastatin

Performance characteristics of the 1260 Infinity Quaternary LC system

Automated fraction re-analysis does it really make sense? Application. Udo Huber. Abstract

Column for High Performance,High-Binding Capacity Ion Exchange Chromatography:TSKgel SuperQ-5PW and Its Applications

Developing Purification Strategies for the Agilent 1260 Infinity II Preparative LC/MSD System

Application Note No. 300/2017. Loading capacity of silica flash cartridges

Method Transfer from an Agilent 1200 Series LC to an Agilent 1260 Infinity II LC

BÜCHI Labortechnik AG

Fast and Comprehensive 2-Dimensional Liquid Chromatography of Block Copolymers through Temperature Gradient Driven Trapping

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

Chromatography Column Performance and Data Analysis Success Guide. Hints and Tips for Better Purifications

P. Wadhwani College of Pharmacy, Yavatmal , India. *Corres.author: Cell No

Benefits of 2D-LC/MS/MS in Pharmaceutical Bioanalytics

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

Gradient Elution. Slide 2

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D7504 on the Agilent 8890 GC System

Analysis of Atrazine in Drinking Water at the ppb Level Using New Agilent Reversed Phase LC Columns. Application. Author. Abstract.

Fundamentals and Techniques of Preparative HPLC. Parto Zist Behboud Tel: 42108

Is there a Preparative Chromatography Green, Economic and Productive?

Analysis of amoxicillin and five impurities on the Agilent 1220 Infinity LC System

Determination of additives and byproducts in an acid copper plating bath by liquid chromatography

Genotoxicity is the property of a compound

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

High Throughput Sub-4 Minute Separation of Antibodies using Size Exclusion Chromatography

A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts

Application Note. Authors. Abstract. Biopharmaceuticals

Techniques for Improving the Efficiency of Large Volume Loading in Preparative Liquid Chromatography

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

Protein-Pak Hi Res HIC Column and HIC Protein Standard

9/2/2014. USP Monograph Modernization. Todays topics. USP basic. Todays topics. - USP basic. - USP publications. - USP monograph modernization

Fast Analysis of Environmental Phenols with Agilent Poroshell 120 EC-C18 Columns

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Qualification of High-Performance Liquid Chromatography Systems

Agilent 1290 Infinity LC with ISET under Waters Empower control Emulation of Agilent 1100 Series LC

GenElute FFPE DNA Purification Kit. Catalog number DNB400 Storage temperature -20 ºC TECHNICAL BULLETIN

INTERNATIONAL RESEARCH JOURNAL OF PHARMACY

DIDANOSINE ORAL POWDER Final text for addition to The International Pharmacopoeia

Chromatography column for therapeutic protein analysis

Lakshmana Rao et.al Indian Journal of Research in Pharmacy and Biotechnology ISSN: (Print) ISSN: (Online)

Analytical Methods Development and Validation

Agilent 1290 Infinity Binary LC System with ISET Emulation of a Waters Alliance 2695 LC Applying Concave, Convex, and Linear Gradients

Product Guide for LudgerSep TM N2 High Resolution Amide HPLC Columns for Glycan Analysis

As reported before, the employed dose was 10 mg/ml is in according with commonly used dose.

Comparing gradient transfer of isocratic hold and delay volume addition using the Agilent 1290 Infinity LC with ISET

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Techniques for Reducing the Effects of Sample Solvent on UHPLC Analyses

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

Promix TM. Enter a New Era in Biomolecule Analysis with. Columns. Unsurpassed Selectivity and Peak Capacity for Peptides and Proteins

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

Technical Overview. Author. Abstract. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

FUTURE-PROOF SOLUTIONS FOR REGULATED LABORATORIES IN THE FACE OF CHANGING USP <621> GUIDELINES

VARIAN, INC. Flash Chromatography

Rapid Extraction of Therapeutic Oligonucleotides from Primary Tissues for LC/ MS Analysis Using Clarity OTX, an Oligonucleotide Extraction Cartridge

Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

Protein Separation with ph Gradients Using Composite Buffer Systems Calculated by the Agilent Buffer Advisor Software

columns P r o P a c H I C C o l u m n S o l u t i o n s f o r P r o t e i n A n a l y s i s

Transcription:

Optimizing the Purification of a Chiral Compound Utilizing a Benchtop, Multi-Purpose, Semi-Preparative to Preparative HPLC System Application Note PHA0111 Keywords Chiral, Enantiomer, HPLC, Normal Phase Chromatography, PLC 2020 Personal Chromatography System, Preparative Chromatography, Purification, Resolution, trans-stilbene Oxide Introduction This study was presented at Pittcon 2011 held in Atlanta Georgia in March 2011. There are often significant differences in the effectiveness and toxicity of drug enantiomers in biological systems (Cox, G. and Ellis, D., 2010; Li, B. and Haynie, D., 2006). This has led to the need for the separation of chiral drug candidates into their respective enantiomers for further testing. Semi-preparative and preparative chiral chromatography has become a common tool for the separation and purification of chiral compounds into their enantiopure form (Franco, P. and Zhang, T., 2010; Li, B. and Haynie, D., 2006). The use of several chiral phases and solvents of different polarities are often employed at this step. This may require the use of either normal phase or reverse phase chromatography during the optimization process. Using, learning and maintaining two different instruments or platforms to accomplish these separations can be both costly and time consuming. The purpose of this application was to optimize the chiral separation of a standard compound using a stand alone, benchtop, preparative chromatography system (Figure 1) that is capable of performing separation by either normal phase or reverse phase chromatography with flow rates up to 100 ml/min. Percent recovery, loading capacity, fraction collection parameters, precision, and its percent impurity are reported for trans-stilbene oxide using normal phase chromatography. In addition, a matrix was developed to assist a chemist with the determination of optimal sample loading parameters as well as methodology for performing simple fraction recovery for each enantiomer. P a g e 1

Figure 1. Gilson PLC 2020 Personal Purification System (Part No. 21130000) Materials & Methods Materials All solvents were distilled in glass suitable for GC, HPLC, LC/MS and spectrophotometry. All reagents were ACS grade quality or better. Hexane and Isopropyl Alcohol were purchased from Burdick & Jackson (part nos. AH212-4 and AH323-4). Trans-Stilbene oxide (98 purity) was obtained from Sigma Aldrich (Figure 2). The standard was prepared in HPLC mobile phase. Figure 2. Chemical Structure of trans-stilbene oxide P a g e 2

HPLC Conditions HPLC purification was carried out using a Gilson PLC 2020 Personal Purification System. Separation was achieved using a 5 micron 10mm x 250 mm Phenomenex Lux Cellulose-2 column at a flow rate of 5 ml/min. The isocratic mobile phase was 1:1 Hexane: Isopropyl Alcohol. Peaks were monitored with UV detection at 220 and 254 nm. See Figure 3 and Table 1 for complete details, including fraction collection parameters. Figure 3. PLC 2020 Software Method (above) Used for the Chiral Separation of trans-stilbene Oxide Enantiomers Table 1. PLC 2020 Chiral Normal Phase Method Normal Phase Mobile Phase Solvent A = Hexane B = Isopropyl Alcohol Mobile Phase Gradient A 0-12 min = 90 Fraction Collection Conditions Front Slope = 65 Back Slope = 65 Maximum Collection Volume per Tube = 20 ml Run Time Flow Rate (ml/min) UV Detection (nm) 12.0 5.0 220 and 254 Method Design Percent recovery, loading capacity, fraction collection parameters, precision and impurity were calculated for trans-stilbene oxide from multiple injections (n=3) using a 100 µl total loop injection. The loading capacity was increased until a resolution of approximately 1 was achieved. This mg load was used for evaluating a single set of fraction collection parameters to determine fraction recovery using reinjection of each enantiomer (Peak 1 and Peak 2 seen in Figure 4). P a g e 3

Results Chiral purification can be optimized by first performing a resolution study to determine the amount of compound that can be loaded onto the specific chiral column. The resulting matrix can assist the researcher with determining the best resolution for purification based on peak width and peak shape (see Table 2 and Figure 4 below). A resolution of 1.15 for the 1.8 mg trans-stilbene oxide load was determined to be optimal and was used for this application. Table 2. Resulting Resolution Matrix for trans-stilbene Oxide Peak 1 Peak 2 Column Loading Study (mg in 90:10 Hexane:IPA) Retention Time Peak Width Retention Time Peak Width Resolution 0.5 4.60 0.67 7.06 0.83 3.29 0.6 4.55 0.73 7.05 0.84 3.19 0.7 4.40 0.75 6.70 0.91 2.77 0.8 4.66 0.77 7.27 0.98 1.88 1.5 4.64 0.82 7.17 1.15 1.39 1.8 4.65 0.91 7.16 1.22 1.15 Figure 4. Chromatogram of trans-stilbene Oxide at resolution = 1 P a g e 4

The PLC 2020 Personal Purification System software allows the user to set both front and back fraction collection parameters (Table 1). This allows for accurate and optimal sample recovery when peak tailing or peak fronting is observed. It is effective for the collection of both large and small peaks. Sharp peaks with smaller peak widths will result in lower fraction collection fraction collection volumes that require less dry-down time. In this application, fraction volumes for trans-stilbene oxide peaks 1 and peak 2 were 3.79 ml and 5.24 ml, respectively (Table 3). Fraction volumes were kept low as a result of narrow peak widths in combination with a 5 ml/min flow rate. Triplicate injections were performed at the 1.8 mg level. Peak 1 and peak 2 were collected separately for each injection. Each of the triplicate fractions from peaks 1 and 2 were mixed manually prior to re-injection of the fractions. The chromatograms of the re-injected fractions are shown in Figures 5 and 6. Average recovery values were calculated at 103.17 and 99.70 for peak 1 and 2 respectively. Consistency in fraction collection was optimal at a load of 1.8 mg, with values at < 1.5. Table 3. Comparison of trans-stilbene Oxide Peak 1 and Peak 2 Collected Fraction Volume and Re-injected Collected Fraction Recoveries Collected Fraction Volume Peak 1 Re-injected Recovery Peak 1 Column Load Resolution Volume (ml) (n=3) Average Re-injected Recovery (n=3) 1.8 1.15 3.79 0.04.01 103.17 0.01 1.17 Collected Fraction Volume Peak 2 Re-injected Recovery Peak 2 Column Load Resolution Volume (ml) (n=3) Average Re-injected Recovery (n=3) 1.8 1.15 5.24 0.05.01 99.70 0.00 0.30 P a g e 5

Figure 5. Re-injection Chromatogram for trans-stilbene Oxide Peak 1 Chiral standards can often have impurities. A consistent impurity peak was visible at ~ 3 minutes in both injection and re-injection chromatograms. At a 1.8 mg trans- Stilbene oxide load, the average area impurity value was calculated to be 2.94 (Table 4). The area ratio of peak 1 and peak 2 in these same 1.8 mg load injections was calculated at 97.9. Area averages of peak 1 and peak 2 and the impurity peak resulted in a final recovery value of 100.9. This was consistent with the listed purity of the standard. Table 4. trans-stilbene Oxide Ratio Area Average Comparison to Impurity Peak Area Coverage Peak 1/Peak 2 Area Ratio Average (n=3) Impurity Peak Area Average (n=3) Peak 1/Peak 2 Ratio Area Average + Impurity Peak Area Average (n=3) 97.9 2.94 100.9 P a g e 6

Summary Optimizing HPLC purification of enantiomers using a benchtop solution that allows for front and back slope peak collection can greatly increase purification efficiency for a variety of separation conditions and applications and insures collection of the entire sample or peak of interest. In this application, a simple resolution matrix study enabled the maximum load that maintained adequate resolution. Fractions were re-injected onto the same column using the minimum volume possible and the same conditions as the original purification method for recovery verification. Consistent recovery and values confirm the optimization of fraction collection volume and recovery rates. References Cox, G.B. and Ellis, D.W. (2010). Method Development for Analysis and Isolation of Chiral Compounds Using Immobilized Stationary-Phase Technology Pharmaceutical Technology Cphl/ICSE 2010, s14-s16. Franco, P. and Zhang, T. (2010). Finding the Best Separation for Enantiomeric Mixtures. LC GC North America 28 (9): 818 822. Li, B. and Haynie, D.T. (2006). Chiral Drug Separation. In Encyclopedia of Chemical Processing DOI: 10.1081/E-ECHP-120039232. Taylor & Francis, p 449-458. LUX is a registered trademark of Phenomenex Corporation P a g e 7