Simple, fast and reliable analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

Similar documents
Multi-element determination in pharmaceutical preparations using the Thermo Scientific icap Q ICP-MS

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

Analytically Monitoring the Effect of Fracturing Activity

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance

EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo

Analysis of elemental impurities in drug products using the Thermo Scientific icap 7600 ICP-OES Duo

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo

Determination of 1,4-Dioxane in Drinking Water by Gas Chromatography/Mass Spectrometry (GC/MS) with Selected Ion Monitoring (SIM)

Nitrogen Determination in Soils and Plants by Flash Combustion using Argon as Carrier Gas

Jonathan R. Beck and Charles T. Yang; Thermo Fisher Scientific, San Jose, CA

TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

Functionality. Normal Operation

for water and beverage analysis

The Benefits of SOLAμ Technology in Sample Preparation. Jon Bardsley and Ken Meadows Thermo Fisher Scientific, Runcorn, UK

Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry

EDS Phase Mapping of a Contact Metamorphosed Calc-Silicate Rock

A Novel Screening Method for Anthropogenic Sewage Pollutants in Waste Water, Ground Water and Drinking Water Samples by LC HRAM Analysis

Collaborative Efforts for Creating a Robust Trace Metals Analytical Procedure for Flue Gas Desulfurization Wastewaters by ICP-MS

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

Simplifying Complex Multi-Residue Pesticide Methodology in GC-MS/MS

High matrix tolerance of the Thermo Scientific icap 7000 Plus Series ICP-OES with the radial Ceramic D-Torch and the sheath gas adaptor

Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass

Analysis of Mineral and Heavy Metal Content in Beverages Using the Teledyne Leeman Labs Prodigy Plus ICP-OES

Thermo Scientific Qtegra Intelligent Scientific Data Solution. Delivering quality. Driving productivity

Automating EPA 6020 Compliant Analysis with the Agilent 7900 ICP-MS and ESI prepfast Autodilution System

Thermo Scientific Solutions for Intact-Protein Analysis. Better, Faster Decisions for. Biotherapeutic Development

ESI SC-FAST Sample Introduction Solutions for PerkinElmer ICP-OES and ICP-MS Systems

Claudio De Nardi, 1 Marta Kozak, 2 David Kasper 3 1. Thermo Fisher Scientific, Dreieich, Germany; 2 Thermo Fisher Scientific, San Jose, CA (USA); 3

California State University, Sacramento: Lead Study

Using HRAM Survey Analysis Combined with Rapid MS2 Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition

US EPA Method using the XSERIES 2 ICP-MS

DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, & USING THE AGILENT 5100 ICP-OES

Isotope Ratio MS. Continuous Flow. Consumables and Spare Parts for Thermo Scientific Elemental Analyzers

Analysis of steel and its alloys using the GB/T standard and an Agilent 5100 ICP-OES in dual view mode

prepfast Autodilution Autocalibration Auto QC Dilution Elemental Scientific

Selenium in Natural Waters by Graphite Furnace Atomic Absorption using EPA Method 200.9

Increased Sample Throughput for ICP-OES Applied to U.S. EPA Method 200.7

EU water analysis using the Thermo Scientific icap 7400 ICP-OES Duo

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Application Note. Abstract. Authors. Environmental

Selenium in Natural Waters by Graphite Furnace Atomic Absorption using EPA Method 200.9

Ultra-Trace Analysis of Beryllium in Water and Industrial Hygiene Samples by ICP-MS Application

Thermo Scientific EASY-nLC 1200 System. Leading in simplicity. and performance

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

Analysis of inorganic impurities in lubricating oils by ICP-MS

Direct Analysis of Photoresist by ICP-MS. Featuring the Agilent Technologies 7500s ICP-MS

Analysis of Heavy Metals in Iron-based Fertilizers by HR ICP-OES

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water. Pranathi R Perati Thermo Fisher Scientific, Sunnyvale, CA, USA

Thermo Scientific icap Q ICP-MS

Thermo Scientific icap 6000 Series ICP Emission Spectrometer

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance

Analysis of elemental impurities in drug products using the Thermo Scientific icap 7400 ICP-OES Duo

Analysis of Samples Related to Nuclear Submarine Operation

Discover the Importance of Third Party Quality Control

Amplifying ICPMS Productivity Using Discrete Sampling Technology

Thermo Scientific icap RQ ICP-MS. Simplicity, productivity and robustness for routine labs

Thermo Scientific TSQ Quantiva Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative performance. with unprecedented ease

Trace Elements in Petrochemical Samples: Achieving Excellent Analytical Results

Analysis of Trace Metals in Surface and Bottled Water with the Optima 7300 DV ICP-OES

icap 7000 Plus Series ICP-OES Streamlined performance and ultra-low ICP-OES detection limits for routine laboratories

Ultra-fast ICP-OES determinations of base metals in geochemical samples using next generation sample introduction technology

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

Quantitative Analysis of Wheat Flour Using FT-NIR

Composition Analysis of Animal Feed by HR ICP-OES

Introduction. Yan Liu, Victor Barreto, Rosanne Slingsby, Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA

Multi-Element Analysis of Trace Metals in Animal Feed using ICP-OES

Lebedev A, 1 Damoc E, 2 Makarov A, 2 Samguina T 1 1. Moscow State University, Moscow, Russia; 2 ThermoFisher Scientific, Bremen, Germany

The Agilent Atomic Spectroscopy portfolio for Environmental applications AA, MP-AES, ICP-OES, ICP-MS & ICP-QQQ

Benefits of running organic matrices using the Agilent 5100 ICP-OES fast, robust, high performance analysis

Analysis of Environmental Samples with the Agilent 730-ES Following US EPA Guidelines

Table I: MCL and MRL Concentrations for Contaminants Monitored Under the Safe Drinking Water Act National Primary Drinking Water Regulations

Analysis of Fracking Flowback Water From the Marcellus Shale Using In-line Conductivity, Automated Dilution, and Ion Chromatography

Analysis of Toxic Trace Elements in Coffee Bean Products by HR ICP-OES

Atomic Spectroscopy Solutions for Environmental Applications

Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation

Analysis of Triazine Herbicides in Drinking Water Using LC-MS/MS and TraceFinder Software

Analysis of Triazine Herbicides in Drinking Water Using LC-MS/MS and TraceFinder Software

Analysis of wear metals and contaminants in engine oils using the 4100 MP-AES

Thermo Scientific KIEL IV Carbonate Device

High performance radio frequency generator technology for the Thermo Scientific icap 7000 Plus Series ICP-OES

Increase productivity for environmental sample analysis using the SVS 2+ Switching Valve System for Agilent 5100 SVDV ICP-OES

Dr. Jianfeng Cui, Thermo Fisher Scientific, Bremen, Germany APPLICATION NOTE AN43384

Implementation of a Walk-Up High- Pressure Capillary Ion Chromatograph for the Fast Separation of Pharmaceutical Relevant Inorganic Anions and Cations

Correlation coefficient

Highlighted Capillary IC Applications

EXPAND YOUR RANGE EXTEND YOUR RESOURCES. Avio 500 ICP Optical Emission Spectrometer

Application Note AN1501

Application Note PlasmaQuant PQ 9000 Elite. Analysis of Rare Earth Elements in Granite and Sandstone by HR ICP-OES. Challenge.

Determination of Elemental Impurities in Graphite-based Anodes using the Agilent 5110 ICP-OES

Elemental Scientific. prepfast. The Next Leap in Automation. Autodilution Autocalibration Intelligent QC Dilution

The Determination of Toxic Metals in Waters and Wastes by Furnace Atomic Absorption

single-use fermentors

The image part with relationship ID rid4 was not found in the file. Welcome

Transcription:

Simple, fast and reliable analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS Kent W. Warnken, Thermo Fisher Scientific, USA Patrick Messina, Quest Diagnostics, USA Application Note 43137 Key Words Blood, Clinical Toxicology, ICP-MS, Lead Goal To develop a fully quantitative, high throughput method for the determination of Pb levels in blood by ICP-MS. Introduction Man has used the unique properties of Lead (Pb) in a wide variety of ways for hundreds of years. For example the Romans used it for plumbing (lead is plumbum in Latin). More recently, lead was used in gasoline and paints to improve performance before it was phased out in the 1970s when the dangers associated with its use were first recognized. Lead is a neurotoxic metal that affects areas of the brain that regulate behavior and nerve cell development 1. A series of exposure pathways allow lead to enter the body, including inhalation of particles (e.g. dust from soils exposed to gasoline and paint lead sources) as well as drinking water and (particularly for children) contaminated toys. On entering the body, lead sequesters itself into soft tissue and bone, where it can be released back to the blood. Young children and fetuses are particularly vulnerable since the blood-brain barrier is not yet fully developed and absorption rates are higher than in adults. Lead, even at low concentrations, has been shown to impair cognitive development. Currently, there is no agreed safe concentration of lead in blood; however the United States Environmental Protection Agency (EPA) has defined standards for dangerous levels of lead in dust, paint and soil 2 to safeguard the population. The United States Centers for Disease Control and Prevention (CDC) states that Blood Lead Levels (BLL) >70 µg/dl (700 ng/ml) can cause serious health effects up to and including coma and death. BLL as low as 10 µg/dl (100 ng/ml) are associated with cognitive development, growth, and behavioral issues in children between the ages of 1-5 years. BLL at 10-25 µg/ dl do not show clinical symptoms of lead poisoning so it is necessary to screen children. The CDC now designates a reference value for BLL of >5 ug/dl as elevated in children ages 1-5. The Occupational Safety & Health Administration (OSHA) also has guidelines for adult BLLs as it relates to work place exposure. All companies that expose their employees to lead must test employees at least once a year. OSHA guidelines include medical removal from the job if levels are > 60 µg/dl (600 ng/ml), and require follow up testing before returning to work. Because of these requirements, the United States performs millions of Blood Lead Level tests per year. All 50 States have BLL programs and require results to be reported back for statistical purposes.

2 Historically, atomic absorption (AA) has been employed for the determination of Pb in whole blood. However, with analysis times often in excess of three minutes, this technique can be quite slow, especially if complicated ashing and atomization programs are used, leading to reduced sample throughput. In addition, AA is not a multi-element approach and often utilizes matrix-modifiers, which can complicate the sample preparation and add cost to the analysis. More recently, ICP-MS has emerged as a rapid, multi-element alternative, with improved detection limits (allowing for reduced sample amounts) and sample throughput over AA. Here, we report a simple, rapid method for BLL by ICP-MS and evaluate between run consistency and sample carry-over. Sample Preparation Because of the high instrumental sensitivity of the Thermo Scientific icap Q ICP-MS, samples and reference standards could be diluted 100 times in 0.5% HNO 3 (prior to an additional 2x dilution from on-line addition of an internal standard) while still maintaining the required detection limits. An additional advantage of the relatively high dilution factor used is the improved nebulization of whole blood samples as well as reduced maintenance of the sample introduction system, both leading to longer analysis periods. Calibration was carried out using external aqueous standards in 0.5% HNO 3 with concentrations of 0.001, 0.010, 0.100, 1.00, 2.00, 5.00 and 10.0 ng/ml. The method was validated by measuring the Lyphochek Whole Blood Level Controls (Levels 1, 2 and 3 (561, 562 and 563)) (Bio-Rad Laboratories Inc.). Instrument Settings All analyses were made using an icap Qc ICP-MS. While the icap Qc is equipped with collision cell technology for the suppression of spectral interferences, it was not used in this study and the icap Qc ICP-MS was therefore operated in STD mode (Table 1). The standard icap Qc sample introduction system consisted of a PFA-ST nebulizer, quartz glass peltier-cooled cyclonic spraychamber and a 2.5 mm ID quartz injector. The 1:100 diluted samples were presented for analysis using a SC-2DX autosampler with integrated 7-port FAST valve from Elemental Scientific Inc (Omaha, NE, USA). A 1 ng/ml Bi (in 0.5% HNO 3 ) internal standard was added on-line (a further 1:1 dilution) using the 7-port FAST valve, giving a total uptake rate of 400 µl/min (an effective 200x sample dilution). Through the combination of the icap Q and the FAST sample introduction system, sample to sample analysis times (including uptake, data acquisition and washout) of < 24 seconds per sample were achieved. Table 1. icap Qc instrument parameters. Parameter Peristaltic pump speed Pump tubing Nebulizer Interface cones RF Power Cool gas flow Auxiliary gas flow Nebulizer gas flow Injector Type; ID Value 32 rpm orange/green tubing for both carrier (sample) and internal standard PFA-ST Nickel 1550 W 14 L/min 0.8 L/min 1.06 L/min Quartz; 2.5 mm Results The analytical figures of merit for five consecutive runs are listed in Table 2, and an example calibration curve is shown in Figure 1. In order to keep analysis times at a minimum, particularly important when sample sizes are limited as is often the case in clinical analyses, a dwell time of 10 ms was used with only 10 sweeps. The instrument detection limit (IDL) obtained with aqueous calibration standards prepared in 0.5% HNO 3, was ~1 ng/l (0.0001 µg/dl) and the blank equivalent concentration (BEC) was ~2 ng/l. Sensitivity is expressed as 206 Pb+ 207 Pb+ 208 Pb and is half of the normal STD mode level, due to the 1:1 on-line IS addition. The percentage recovery of the internal standard, Bi, over the course of the analysis (Run #3) is shown in Figure 2 and ranged from 94% to 103%. This indicates the analytical robustness of the method and freedom from matrix suppression.

Run # Analyte Dwell (sec) Mode Sweeps Sensitivity cps/ng/l R value BEC (ng/l) LOD (ng/l) 1 Pb 0.010 STD 10 125 0.999996 1.7 1.0 2 Pb 0.010 STD 10 140 0.999990 1.9 0.9 3 Pb 0.010 STD 10 142 0.999993 0.8 1.1 4 Pb 0.010 STD 10 121 0.999950 1.3 0.2 5 Pb 0.010 STD 10 137 0.999991 1.6 0.8 Table 2. Analytical figures of merit for five consecutive analytical runs. 3 Figure 1. Example calibration curve for Pb from 0 to 10 ng/ml. The insert shows the same curve with concentrations between 0 and 0.01 ng/ml. Figure 2. Screenshot from the Thermo Scientific Qtegra software showing the percentage recovery of the internal standard, Bi, during Run #3.

4 To determine the ability of the icap Q to accurately achieve low level BLL measurements, an additional 10x dilution was performed on the low level control (Table 3). The excellent agreement (97-105 %) with the reference values indicates that the method could easily support dilution factors in excess of the 200x used in this work. Table 3. Serial dilution results for the low level Bio-Rad control (561). Dilution Factor Pb Meas. (µg/l) D.F. Corr. (µg/dl) Certified Value (ug/dl) % Recovery 1000 0.085 8.51 8.14 105 100 0.810 8.10 8.14 100 1000 0.079 7.89 8.14 97 100 0.854 8.54 8.14 105 1000 0.083 8.28 8.14 102 100 0.794 7.94 8.14 98 Due to the clinical implications of this test, it is also important to assess the washout characteristics of the sample introduction system, and the likelihood that a false positive could be generated by inadequate washout from a previously analyzed sample. To assess this, a blank was repeatedly analyzed after five measurements of the high control standard (Figure 3). Using a FAST washout of two seconds, the lead background was reduced by >2 orders of magnitude. This equates to a BLL of ~0.001 µg/dl and is, therefore, a more accurate measure of the effective detection limit of the method described. Figure 3. Washout results from five analyses of the high (Level 3) Bio-Rad control followed by five blank measurements. Finally, an assessment of the accuracy of the proposed method was carried out by analyzing the control standards during each of the five analytical runs. The results for the BioRad Level 1-3 controls are given in Table 4.

Table 4. BLL (µg/dl) measured as the sum of the stable Pb isotopes 206, 207 and 208 for the Bio-Rad Control Levels 1-3. The % RSD for each measurement is also shown. 5 Run # Analyte Level 1 % RSD Level 2 % RSD Level 3 % RSD Range (AA) Pb 6.28-9.42 20.4-30.6 34.2-51.3 Value (AA) Pb 7.85 25.5 42.8 Value (ID-ICP-MS) Pb 8.14 25.2 43.2 Run 1 1a Pb 7.98 4.8 25.2 1.9 41.9 1.9 Run 2 Pb 2a Pb 8.22 1.3 25.7 2.0 42.2 1.6 2b Pb 8.16 2.6 25.5 1.2 42.1 0.8 2c Pb 8.09 2.7 25.7 1.7 41.8 2.1 Mean ± SD 8.16 ± 0.065 25.6 ± 0.12 42.0 ± 0.21 % RSD 0.80 0.45 0.50 Run 3 3a Pb 8.10 1.4 25.7 0.4 43.5 0.1 3b Pb 8.32 1.2 25.5 1.6 43.0 2.0 3c Pb 8.16 0.7 25.1 1.1 42.4 1.3 Mean ± STD 8.19 ± 0.11 25.4 ± 0.31 43.0 ± 0.55 % RSD 1.4 1.2 1.3 Run 4 4a Pb 8.54 3.8 25.8 1.3 40.6 3.1 4b Pb 8.40 0.6 26.5 2.9 41.2 1.9 4c Pb 8.43 2.9 26.0 0.8 41.3 2.5 4d Pb 8.29 3.3 26.1 3.0 41.5 0.9 Mean ± STD 8.42 ± 0.10 26.1 ± 0.29 41.2 ± 0.39 % RSD 1.2 1.1 0.94 Run 5 5a Pb 7.94 0.6 25.0 1.1 42.5 1.3 5b Pb 8.15 2.8 25.9 1.3 42.4 2.6 5c Pb 8.19 1.3 25.1 1.0 41.6 2.3 5d Pb 8.25 2.2 25.0 1.5 43.0 2.4 5e Pb 8.07 1.2 24.5 2.2 42.5 1.4 Mean ± STD 8.12 ± 0.12 25.1 ± 0.50 42.4 ± 0.50 % RSD 1.5 2.0 1.2 Run 1 5 Mean ± STD 8.21 ± 0.16 25.5 ± 0.51 42.1 ± 0.75 % RSD 2.0 2.0 1.8

Conclusions A fully quantitative method for the determination of blood Pb by ICP-MS has been described. Calibration using aqueous calibration standards is possible and the accuracy was assessed by the measurement of control samples. The proposed method, with an instrumental detection limit of 0.0001 µg/dl (and effective method detection limit of 0.001 µg/dl due to sample carry-over) can be used for the analysis of real samples with dilution factors of 1000 as demonstrated by recoveries of the low level control. Due to the high sensitivity of the Thermo Scientific icap Q ICP-MS, this increased dilution factor would further decrease system maintenance, enabling high volume clinical laboratories to process samples at a rate of 2.5 per minute. Application Note 43137 References 1. NRC (National Research Council). 1993. Measuring lead exposure in infants, children, and other sensitive populations. Washington, DC: National Academies Press. 2. United States Federal Register, Part III, Environmental Protection Agency, 40 CFR Part 745; Lead: Identification of Dangerous Levels of Lead; Final Rule. Disclaimer The method presented is intended for research purposes only. www.thermoscientific.com 2013 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. Lyphochek and Bio-Rad are trademarks of Bio-Rad Laboratories, Inc. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific, (Bremen) GmbH Management System Registered is ISO 9001:2008 Certified. Africa-Other +27 11 570 1840 Australia +61 3 9757 4300 Austria +43 1 333 50 34 0 Belgium +32 53 73 42 41 Canada +1 800 530 8447 China +86 10 8419 3588 Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland/Norway/Sweden +46 8 556 468 00 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9434 Italy +39 02 950 591 Japan +81 45 453 9100 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Russia/CIS +43 1 333 50 34 0 South Africa +27 11 570 1840 Spain +34 914 845 965 Switzerland +41 61 716 77 00 UK +44 1442 233555 USA +1 800 532 4752 AN43137_E 03/13C