Peptide Mapping: A Quality by Design (QbD) Approach

Similar documents
Automated QbD-Based Method Development and Validation of Oxidative Degraded Atorvastatin

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

Application Note. Abstract. Author. Biopharmaceutical

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

Emulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Automated Scouting of Stationary and Mobile Phases Using the Agilent 1290 Infinity II Method Development Solution

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Method Transfer from an Agilent 1200 Series LC to an Agilent 1260 Infinity II LC

Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles

Agilent 1290 Infinity Quaternary LC Support of Columns with 2.1 to 4.6 mm ID to 1200 bar

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

Fast Method Development Using the Agilent 1290 Infinity Quaternary LC System with Column Selection Valve

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

BioHPLC columns. Tim Rice Biocolumn Technical Specialist

Performance characteristics of the 1260 Infinity Quaternary LC system

AdvanceBio Peptide Mapping

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis

Mobile Phase Optimization in SEC Method Development

Characterize mab Charge Variants by Cation-Exchange Chromatography

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Application Note. Author. Abstract. Pharmaceuticals. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Peptide Mapping. Hardware and Column Optimization

Application Note. Authors. Abstract. Biopharmaceuticals

Protein Separation with ph Gradients Using Composite Buffer Systems Calculated by the Agilent Buffer Advisor Software

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

Quantification of Host Cell Protein Impurities Using the Agilent 1290 Infinity II LC Coupled with the 6495B Triple Quadrupole LC/MS System

Intact & Subunit Purity

Agilent OpenLAB CDS MatchCompare for the Comparison of Peptide Mapping Samples

Application Note. Author. Abstract. Small Molecule Pharmaceuticals. Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany

4/4/2013. BioHPLC columns. Paul Dinsmoor Biocolumn Technical Specialist. April 23-25, Size Exclusion BioHPLC Columns

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Determination of Asarinin in Xixin (Asari Radix Et Rhizoma)

Reproducible LC/MS Peptide Separation Using Agilent AdvanceBio Peptide Plus Columns

Agilent 1290 Infinity Binary LC System with ISET - Emulation of the Waters Alliance 2695 LC system analyzing b-blockers

Technical Overview. Author. Abstract. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Agilent 1290 Infinity Binary LC System with ISET Emulation of a Waters Alliance 2695 LC Applying Concave, Convex, and Linear Gradients

Method Transfer from an Agilent 1260 Infinity LC to an Agilent 1260 Infinity II LC

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Agilent 1290 Infinity II 2D-LC Solution Biopharmaceutical Polymer Analysis. WCBP Jan 2017 Washington, DC

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Application Note. Authors. Abstract. Ravindra Gudihal Agilent Technologies India Pvt. Ltd. Bangalore, India

Molecular Characterization of Biotherapeutics The Agilent 1260 Infi nity Multi-Detector Bio-SEC Solution with Advanced Light Scattering Detection

Fusion Analytical Method Validation

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs)

Scale-up with the Agilent SD-1 Purification System analytical and preparative runs on a single system

Agilent 1290 Infinity LC with ISET under Waters Empower control Emulation of Agilent 1100 Series LC

Development of Analysis Methods for Therapeutic Monoclonal Antibodies Using Innovative Superficially Porous Particle Biocolumns

Developing Quantitative UPLC Assays with UV

Fast Analysis of Environmental Phenols with Agilent Poroshell 120 EC-C18 Columns

Application Note. Authors. Abstract. Introduction. Pharmaceutical

Benefits of 2D-LC/MS/MS in Pharmaceutical Bioanalytics

Understanding the Effects of Common Mobile Phase Additives on the Performance of Size Exclusion Chromatography

Application Note. Author. Abstract. Small Molecule Pharmaceuticals & Generics. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Fusion Analytical Method Validation

Agilent 1260 Infinity Bio-inert Quaternary LC System. Infinitely better for bio-molecule analysis

Separation of Native Monoclonal Antibodies and Identification of Charge Variants:

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Comparing gradient transfer of isocratic hold and delay volume addition using the Agilent 1290 Infinity LC with ISET

Method Translation in Liquid Chromatography

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

Improved Peptide Maps Using Core-Shell Media: Explaining Better Biopharmaceutical Applications With Kinetex C18 Columns

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns

Analysis of Monoclonal Antibody Digests with the Agilent 1290 Infinity 2D-LC Solution

The Agilent 1260 Infinity BioInert Quaternary Pump. Scope of a low-pressure mixing UHPLC pump with Bio-Inert Capabilities

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI

Method Development Considerations for Reversed-Phase Protein Separations

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

Application Note. Environmental. Agilent 1290 Infinity Binary LC. with ISET and fine-tuning. Agilent 1290 Infinity Binary LC.

Analysis of Monoclonal Antibody (mab) Using Agilent 1290 Infinity LC System Coupled to Agilent 6530 Accurate-Mass Quadrupole Time-of-Flight (Q-TOF)

Analytical Instrument Qualification According to USP <1058>: Requirements and Examples for the Agilent 1290 Infinity LC System

High-Throughput LC/MS Purification of Pharmaceutical Impurities

As reported before, the employed dose was 10 mg/ml is in according with commonly used dose.

CX-1 ph Gradient Buffer

Automated alternating column regeneration on the Agilent 1290 Infinity LC

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

Profiling Glycosylation of Monoclonal Antibodies at Three Levels Using the Agilent 6545XT AdvanceBio LC/Q TOF

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

ISET. Agilent 1290 Infinity Binary LC with ISET, emulation of the Waters Alliance 2695 LC system analyzing aromatic acids. Application Note.

NISTmAb characterization with a high-performance RP chromatography column

Bravo with AssayMAP. Automated Sample Preparation

Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins

Transcription:

Peptide Mapping: A Quality by Design (QbD) Approach Application Note Bio-Pharmaceutical Authors Sreelakshmy Menon and Suresh babu C.V. Agilent Technologies, Inc. Richard Verseput S-Matrix Corporation Abstract Peptide mapping is a widely used technique for the characterization of proteins. This Application Note demonstrates a robust LC method and design space for peptide mapping using a Quality by Design (QbD) approach. A therapeutic monoclonal antibody (mab) was employed as the model protein in the current peptide mapping study. Fusion QbD software s multivariate design and analysis capabilities, which are aligned with a QbD approach, in combination with its OpenLAB CDS experiment automation capabilities, provided a detailed understanding of the critical method parameter (CMP) effects on the various critical method attributes (CMAs) included in the study. This knowledge was translated directly into a robust final peptide mapping method with maximum number of peaks and good separation in far less method development time.

Introduction Peptide mapping is the most commonly used identity test for proteins 1. Peptide mapping is protein fingerprinting; it involves several processes including digestion, separation of peptides, and data analysis, which should result in comprehensive understanding about the protein being analyzed 1. Reversed-phase liquid chromatography (RPLC) with UV detection is the most commonly used method for peptide mapping in quality control (QC). One of the major challenges in traditional LC method development for peptide mapping is the time consumed in screening the various chromatographic parameters using a one-factor-at-a-time (OFAT) approach. The goal of analytical Quality by Design (QbD) is to achieve quality in measurement, leading to consistent quality of a drug product. Quality principles in the ICH guidelines can be implemented in the design of analytical method development work to meet that goal 2,3. Many of the traditional LC method development limitations can be overcome with the QbD approach. For example, design of experiment (DOE) methods can reduce the large number of runs normally needed to support the OFAT approach. DOE can provide increased understanding of method performance and variability from a limited number of experiments selected to specifically characterize independent and multivariate interactions between method parameters 4. This Application Note describes the application of the analytical QbD approach to peptide mapping method development using an Agilent 12 Infinity Bio-inert LC with Fusion QbD method development software for a tryptic digested monoclonal antibody (mab). The Fusion QbD software delivers a totally automated DOE-based experimental approach that evolves systematic method development from start to end. Materials and Methods LC instrumentation All chromatographic analyses were performed on an Agilent 12 Infinity Bio inert Quaternary LC system consisting of the following modules: Agilent 12 Infinity Bio-inert Quaternary LC Pump (G5611A) Agilent 12 Infinity Bio-inert High Performance Autosampler (G5667A) Agilent 12 Infinity Series Thermostat (G133B) Agilent 12 Infinity Thermostatted Column Compartment (G1316C, option 19) Agilent 12 Infinity DAD VL with a 1-mm bio-inert standard flow cell (G1315D, option 28) Software Agilent OpenLAB CDS ChemStation Edition, Version C.1.5. Fusion QbD Software Platform, version 9.7. Build 494 (S-Matrix Corporation). Reagents and chemicals The mab protein was a proprietary therapeutic molecule. Tris buffer, urea, iodoacetamide (IAA), dithiotreitol (DTT), trifluoroaceticacid (TFA), and formic acid were obtained from Sigma-Aldrich. Trypsin is from Agilent Technologies. Acetonitrile, methanol, and isopropanol (IPA) were obtained from Labscan. Tryptic digestion The mab (2 mg) was treated with DTT and IAA for reduction and alkylation. After the reduction and alkylation steps, the ph of the solution was adjusted to ph 7 8, and trypsin digestion (2:1, protein to protease w/w) was performed overnight, incubating at 37 C. The incubated sample was then quenched with.5 % TFA. The samples were either immediately analyzed or stored at 2 C until use. 2

Results and Discussion QbD Workflow In this study, the QbD workflow consists of two phases: 1) screening and 2) method optimization. Figure 1 presents a schematic representation of the QbD workflow for each phase. In each phase, given a defined set of variable inputs, the Fusion QbD software generates a statistical DOE experiment design. It then exports the design to the Agilent OpenLAB ChemStation Software as ready to-run sequences and methods, which were run on an Agilent 12 Infinity Bio-inert Quaternary LC system. The software then imports and models the data from the experiment chromatograms, and generates an automated report identifying the best conditions, in terms of meeting the defined performance goals for the critical method attributes (CMAs) 5. Table 1 summarizes the CMA (response goals) for the current study. Chemistry screening Design the experiment Use DOE to screen parameters: columns, solvents, temperature, injection volume Export the design to the CDS Export the DOE runs to the CDS as ready-to-run methods and sequence. Analyze the data Import the results. Automatically model the data. Use the Best Overall Answer search engine to identify the best method. Verify the answer Verify the predicted best method conditions by exporting the method to the CDS, and running the method. Results Select the best column, solvent, temperature, and injection volume to use in the optimization phase. Optimization Design the experiment Use DOE to study the joint ranges defined for critical method parameters (CMPs). Carry out the experiment Export the design to the CDS, run the Fusion QbD built methods and sequence, import the results, and automatically model the data. Establish mean performance design space Establish mean performance design space with proven acceptable ranges (PARs) for the CMPs. Establish robust final design space Use Robustness Simulator to calculate the robustness of all methods in the mean performance design space. Verify the answers Verify the best method predictions by exporting the method to the CDS and running the method. Results Robust design space with validated points that meet all performance goals. Figure 1. Schematic representation of the QbD workflow with Fusion QbD. Table 1. Critical method attributes (CMAs) for QbD peptide mapping analysis. Response goals (critical method attributes) Target Relative rank No. of peaks Maximize 1 No. of peaks 1. (tangent resolution) Maximize 1 No. of peaks 2. (tangent resolution) Maximize 1 3

Screening This first phase involved screening column and solvent type combinations using a generic gradient 6. The experiment setup required specifying critical method parameters (CMPs) as either variables or constants. The parameters specified as variables in the screening study included different columns (Column A: AdvanceBio Peptide Mapping column, Column B, and Column C), different strong solvents (acetonitrile, methanol, IPA + acetonitrile + water), injection volume (1 5 µl), and column temperature ( C). Gradient profile, flow rate (1 ml/min), and additive concentration (.8 % TFA) were kept constant. In this phase, the goal was to identify the conditions that provide better peptide separation in terms of the number of separated peaks. Figure 2 shows example chromatograms obtained from various phase 1 DOE experiment runs. Table 2 summarizes the outcome of the screening phase results. mau DAD1 A, Sig=214 2 1 Column A 2 1 Column B 2 1 Column C 5 1 15 2 25 3 35 min Figure 2. Chromatographic profiles of tryptic digested mab under three different column conditions. Column A: Agilent AdvanceBio Peptide Mapping column (4.6 1 mm, 2.7 µm); Column B: Agilent Poroshell 12 SB (4.6 mm, 2.7 µm); Column C: Agilent ZORBAX SB (4.6 1 mm, 5 µm). Table 2. Screening phase output predicted by the Fusion QbD software. Best overall answer Variable Level setting Column AdvanceBio Peptide Mapping column 4.6 1 mm, 2.7 µm Strong solvent Acetonitrile Oven temperature 53 C Injection volume 5 µl 4

The Advance Biopeptide column (Column A) and acetonitrile (Solvent A) mobile phase were found to be the best conditions to obtain the maximum number of peaks with tangent resolution >1.5 and 2. Figure 3 presents an example of the response surface graphs used to identify the optimum injection volume and oven temperature. Method optimization The second phase used the best column, strong solvent, and injection volume identified in the screening study. In this phase, the method was further optimized by studying pump flow rate, gradient slope, and additive concentration parameters expected to strongly influence the method s mean (average) performance and the method s robustness. This phase also included the higher end of the column (oven) temperature range, which was shown to be advantageous in the screening study. Including column temperature enables characterizing the effect of temperature variation on method robustness. Tables 3 and 4 respectively summarize the critical method variables included in the study, and the resulting optimized method. Figure 4 presents a response graph showing the combined (interactive) effects of pump flow rate and strong solvent % on the maximum number of separated peaks one of the CMAs in this study. No. of peaks (*) 9 8 7 3 1 Injection volume (L) 2 55 3 4 Oven temperature ( C) 45 5 1 Oven temperature ( C) 55 2 3 45 4 5 Injection volume (L) Legend 9 85.8 81.8 8 77.7 73.7 7 69. 65. 61. 57. 53. 49. 45.3 41.3 3 37.2 33.2 29.1 Figure 3. Fusion QbD software response graph showing the impact of an interaction between injection volume and oven temperature. The graph shows the combined (interactive) effect of increasing both injection volume and oven temperature on the maximum number of separated peaks. Table 3. The constants and variables tested during optimization phase. Experimental details Constants Gradient Equilibration 1. minutes 2 %B Initial hold.1 minutes, 2 %B Final hold 3 minutes at 95 %B Re-equilibration 1 minute at 2 %B Injection volume 5 µl Wavelength Variables Pump flow rate 214 nm ± 4 nm Final % of strong solvent 1 3 % Intermediate hold.5, 1.25, 2. ml/min,, and minutes Oven temperature, 55, and C Additive concentration (TFA),.4, and.8 % No. of peaks (*) Table 4. Optimization phase output predicted by the Fusion QbD software. Best overall answer Variable Level setting Pump flow rate.5 µl Intermediate hold time minutes Final % strong solvent 18 % Oven temperature C Additive concentration.7 5

Design space Design Space is a primary region, predicted by the Fusion QbD software, that defines the CMAs in terms of the CMPs 7. To establish a robust final design space, it is important to quantify the robustness of all possible methods in the mean performance design space. This is done using the software s Robustness Simulator feature to characterize the independent and combined effects of the method parameters on method variability. Figure 5 shows a Trellis graph illustrating the impact of the CMPs on the CMAs (response goals). These graphs facilitate understanding the combined effects of the study variables on the separation of the peptide mixture. No. of peaks (*) 9 8 7.5 Pump flow rate (ml/min) 3 25 1. 2 1.5 Final % strong solvent (%) 15 2. 3 1.5 Final % strong solvent (%) 25 2 1. 15 1.5 2. 1 Pump flow rate (ml/min) 7 9 8 No. of peaks (*) Legend 11. 97.2 92.8 88. 84. 79. 75.2 7.8 66. 62. 57. 53.2 48.8 44.. Figure 4. Fusion QbD software response graph showing the combined (interactive) effects of pump flow rate and strong solvent % on the maximum number of separated peaks. The graph shows that the best result occurs by combining low pump flow rate with high final % strong solvent. Intermediate hold time Intermediate hold time Intermediate hold time. Pump flow rate =. Pump flow rate = 1.2 Pump flow rate = 2. 55... 55... 55.. Oven temperature Oven temperature Oven temperature Additive concentration =.8 Additive concentration =.4 Additive concentration =. Figure 5. Trellis Graphs show the design space (unshaded region), proven acceptable ranges (the PAR rectangle), and final method settings (center dot in rectangle) for oven temperature and intermediate hold time achievable at low pump flow rate and high additive concentration. 6

In Figure 5, the unshaded (white) region in the bottom left graph of the trellis is the robust design space the region containing methods that meet required mean performance and robustness. In this graph, the rectangle demarcates the joint proven acceptable ranges (PARs) for oven temperature, intermediate hold time, and the final method settings (the center point of the rectangle). The Fusion QbD software automatically selects the center point and the four border points of the PAR rectangle for point prediction verification, and automatically includes the pump flow rate and additive concentration associated with the graph. Figure 6 is an expanded view of the selected 2D contour graph containing the robust design space and PARs for the combined method performance requirements (CMAs) of the greatest number of separated peaks and the highest number being baseline resolved and well resolved (Tangent Resolution 1.5 and 2.). Intermediate hold time (min) 59 58 57 56 55 54 53 52 51 49 48 47 46 45 44 43 42 41. Proven acceptable ranges (PARs) 51.67 53.33 55. 56.67 58.33. Oven temperature ( C) Figure 6. Fusion QbD software design space graph with PAR rectangle showing proven acceptable ranges for oven temperature and intermediate hold time. 7

Method verification The final step involved verifying that the optimized method s predicted performance met all response goals. To do this, the Point Predictions feature in Fusion QbD was used to automatically export the center point and the four border point verification methods from the PAR rectangle to OpenLAB for automated execution. These points are identified as A, B, C, D, and T in the 2D graph presented in Figure 6. The experimentally verified chromatographic runs at these five different conditions are presented in Figure 7, along with the results for the CMAs. DAD1 A, Sig=214 A mau 1 B 1 5 1 1 T 4 2 3 D 1 C 1 1 2 Response variables 3 min Predicted Experimental ±2 Sigma confidence limit A No. of peaks No. of peaks 1.5 (Tangent resolution) 93 75 98 78 8 15 66 84 No. of peaks 2. (Tangent resolution) 69 51 69 B No. of peaks 94 88 No. of peaks 1.5 (Tangent resolution) 76 74 No. of peaks 2. (Tangent resolution) 61 62 T No. of peaks No. of peaks 1.5 (Tangent resolution) No. of peaks 2. (Tangent resolution) D No. of peaks No. of peaks 1.5 (Tangent resolution) No. of peaks 2. (Tangent resolution) C No. of peaks No. of peaks 1.5 (Tangent resolution) No. of peaks 2. (Tangent resolution) 81 17 66 86 51 7 94 86 81 18 78 72 68 87 58 51 69 94 88 8 17 79 7 69 88 57 69 93 86 8 15 78 72 68 87 59 55 68 Figure 7. The chromatographic runs of the center (T) and four (A, B, C, and D) robust operating range points along with the response goal values. *selected peaks for %RSD calculations (Table 5). 8

All the values obtained after the point prediction validation studies were within the predicted range (±2 Sigma confidence limit). The method s reproducibility was also evaluated by replicate injections, and the %RSD calculations presented in Table 5 show the precision of the method. Figure 8 shows the comparison of the chromatogram obtained from the initial screening study, and the chromatogram obtained from the final QbD optimized method. A comparison of these chromatograms shows the dramatic improvements in both the number of separated peptide peaks and the number of well resolved peptide peaks obtained from method optimization using the Fusion QbD software and the QbD-aligned approach. Table 5. Precision of the final LC method. Upper limits of acceptable %RSD: Retention time =.1 %, Area = 2. % and Area% = 5 %. %RSD values for selected peaks Peaks Retention time Area Area% 1.5 1.15. 2.3.63 1.6 3.3..66 4.2.97 4. 5.2 1.8 2.1 mau 2 1 Before optimization DAD1 A, Sig=214 No. of peaks No. of peaks ~1.5 tangent resolution 45 41 38 No. of peaks ~2. tangent resolution mau 2 1 QbD optimization DAD1 A, Sig=214 1 2 3 No. of peaks No. of peaks ~1.5 tangent resolution 94 78 No. of peaks ~2. tangent resolution min 1 2 3 min Figure 8. Comparison of chromatograms obtained before and after QbD optimization, along with the response goal values. 9

Conclusion This Application Note demonstrates systematic LC method development for peptide mapping using QbD principles on an Agilent 12 Infinity Bio-inert Quaternary LC system. Fusion QbD method development software with OpenLAB ChemStation facilitates the automation of LC method development by evaluating multiple parameters such as column, solvent, oven temperature, injection volume, additive concentration, and gradient slope. The Fusion QbD software s multivariate experiment design and analysis capabilities, combined with its automated experiment execution technology, enabled a robust final method for mab peptide mapping to be obtained in less than two weeks total development time, including instrument run time and analyst time for instrument setup and chromatogram processing. Therefore, the automated QbD method development approach using Fusion QbD software has provided a far better performing and more robust method in dramatically less time compared to manual method development. References 1. Recombinant Protein Characterization, Agilent Primer, publication number 599-8561EN, August 8, 211. 2. Vogt, et al. Development of Quality-By- Design Analytical Methods, J. Pharma. Sci. Vol. March 211,, No. 3, p.797-812. 3. G. Reid, et al. Analytical Quality by Design (AQbD) in pharmaceutical Development. American Pharmaceutical Review August 213, 144191. 4. https://smatrix.com/pdf/fusion%2 AE%2%2LC%2Method%2 Development%2Brochure.pdf 5. L. Jones, et al. Rapid Peptide Mapping Method with High Resolution Using a sub 2-μm Column, Agilent Technologies, publication number 599-4712EN. 6. J. Martosella, A. Zhu, N. Tang, Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles, Agilent Technologies, publication number 5991-3585EN. 7. S. Lateef, A. K. Vinayak, Automated QbD based Method Development and Validation of Oxidative Degraded Atorvastatin, Agilent Technologies, publication number 5991-4944EN. 1

11

www.agilent.com/chem For Research Use Only. Not for use in diagnostic procedures. This information is subject to change without notice. Agilent Technologies, Inc., 215, 217 Published in the USA, November 2, 217 5991-6338EN