HYDROCARBONS, BP C 1500

Similar documents
Formulas: See Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1. METHOD: 2555, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003

ANILINE, o-toluidine, AND NITROBENZENE 2017

MERCAPTANS, METHYL-, ETHYL-, and n-butyl- 2542

THIOPHANATE-METHYL in air 5606

POLYCHLOROBENZENES 5517

METHOD: 2016, Issue 2 EVALUATION: FULL Issue 1: 15 January 1998 Issue 2: 15 March 2003

CORESTA Recommended Method No. 57

2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial waste waters.

Corn Syrup Analysis E-1a- 1. ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

ACETALDEHYDE. The procedure is applicable to corn syrups including those containing fructose, and to most other starch hydrolyzates (Note 1).

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D*

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

STANDARD OPERATING PROCEDURES

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

METHOD 310A - DETERMINATION OF RESIDUAL HEXANE THROUGH GAS CHROMATOGRAPHY

NWTPH-Gx. Volatile Petroleum Products Method for Soil and Water

PARTICULATES NOT OTHERWISE REGULATED, RESPIRABLE 0600

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016

Ethylene Oxide. Fully validated method. This method has been subjected to the established evaluation procedures of the Methods Development Team.

EXTRACTION OF BIOLOGICAL TISSUES FOR AROMATIC AND CHLORINATED HYDROCARBONS

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE.

Standard Operation Procedure (SOP) for the Extraction and Instrument Analysis of Siloxanes from Cartridges

Validation of Sulfur Dioxide Using. SKC UME x 200 Passive Sampler Cat. No Page 1 of 14 Publication 1781 Rev 1607

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D7504 on the Agilent 8890 GC System

New Developments in Thermal Desorption (TD) Tube & Canister Technologies for Collection & Analysis of Soil Gas

Hydrocarbon Exposures During Tank Gauging and Sampling Operations

Certified Reference Material - Certificate of Analysis

Field Tests of Chloroform Collection/Analysis Methods

Analytical Method of Limit Test for Hexachlorobenzene. in Picloram TC and Method Validation Data*

C146-E096 TD-20. Shimadzu Thermal Desorption System

Course Outline. in Soil/Sediment&WasteSediment&Waste by. Overview of VOCs Analysis Sample Collection VOCs Analysis

JAI RESEARCH FOUNDATION A Global Contract Research Organisation

Large Volume Injection of Organochlorine Pesticide Extracts Using a Programmable Temperature Vaporizer in the Solvent Split Mode

CHROMATOGRAPHY/MASS SPECTROMETRY

Principles, Practices, and Equipment for Performance Evaluation of Personal Monitoring Badges. Maria D.R. Peralta, PhD and C.R.

Evaluation by Competent Authorities

Liners and consumables for OPTIC Inlet. Liner Selection Guide

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

CORESTA Recommended Method No. 63

Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP1 and DB-CLP2

Volatile Organic Carbon Contaminated Site Assessment

METHOD 3610B ALUMINA CLEANUP

POLYCHLOROBIPHENYLS in serum 8004

Certified Reference Material - Certificate of Analysis

Volatile Organic Compounds in Solids PBM

Certified Reference Material - Certificate of Analysis

Industrial Hygiene Report Control of Glycol Ether Solvent Fumes Using Activated Carbon Filter Media: Ethylene Glycol Monobutyl Ether.

Hydrocarbon Processing Industry Gas Chromatographs. Standard GC Analyzer Solutions

4002 Synthesis of benzil from benzoin

Joette L. Steger, Joan T. Bursey, and David Epperson

Project Summary. Field Evaluation of EPA Method 0040 (Volatiles Using Bags)

Partly O-(2-hydroxyethylated) cellulose. It may contain suitable ph-stabilisers such as phosphates.

Application Note AN1501

Certified Reference Material - Certificate of Analysis

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

Method of determination of phthalates by gas chromatography / mass spectrometry in wines

3005 Synthesis of 7,7-dichlorobicyclo[4.1.0]heptane (7,7-dichlornorcarane) from cyclohexene

STANDARD OPERATING PROCEDURES SOP: 1502 PAGE: 1 of 11 REV: 0.0 DATE: 10/24/94

CORESTA RECOMMENDED METHOD N 61

Residual Solvents in medical cannabis analysis using the SRI 8610C FID GC June 2014

Passive Monitoring: A Guide to Sorbent Tube Sampling for EPA Method 325. Nicola Watson July 2015

Gas Chromatography SSI-GC-003

EURL-SRM - Analytical Observations Report

TECHNICAL MEMORANDUM. Pamela Moss, EA Team (EA Engineering, Science, and Technology, Inc., PBC/ CH2M HILL)

Certified Reference Material - Certificate of Analysis

ELEMENTS by ICP 7301 (Aqua Regia Ashing)

Certified Reference Material - Certificate of Analysis

Chromatography 2015, 36, Technical Note

Certified Reference Material - Certificate of Analysis

ELEMENTS by ICP 7301 (Aqua Regia Ashing)

3015 Synthesis of iodocyclohexane from cyclohexane and iodoform

Standard Test Methods for Aromatics in Mineral Spirits by Gas Chromatography 1

METHOD 3560 SUPERCRITICAL FLUID EXTRACTION OF TOTAL RECOVERABLE PETROLEUM HYDROCARBONS

HYDROCARBONS, HALGENATED 1003

Analysis of PCBs in Sewage sludge for the characterization study of ERM CC392

ELEMENTS by ICP 7300 (Nitric/Perchloric Acid Ashing)

USEPA Methods and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors

SILICA, CRYSTALLINE by IR 7602

Method 0040 for Volatile Total Organics as Unspeciated Mass in Stack Gas Emissions

Experiment 11. DNA Composition by HPLC 1,2

USEPA Methods and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors

PuriTox Aflatoxin Product Code: P25

Analysis of styrene leached from polystyrene cups using GCMS coupled with Headspace (HS) sampler

When optimization is discussed throughout this talk, it s in terms of performing more samples, in less time, while minimizing investment in capital

TurboMatrix Headspace and Headspace Trap Samplers

Capillary GC Troubleshooting: a Practical Approach

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

NCASI METHOD CI/SG/PULP CHILLED IMPINGER TEST METHOD FOR USE ON PULP MILL SOURCES TO QUANTIFY METHANOL EMISSIONS

Direct Injection of Fish Oil for the GC-ECD Analysis of PCBs: Results Using a Deans Switch With Backflushing. Application. Author.

EDB July ETHYLENE DlBROMlDE. F. Analytical Quantitation... I. Determinative Method. Contents. A. Introduction I3. Equipment...

Validation Data for the Analysis of Chlorotalonil in Leek Using Mini-Luke Method Followed by GC-QqQ-MS/MS

2023 Reduction of D-(+)-camphor with lithium aluminium hydride to an isomeric mixture of (+)-borneol and ( )-isoborneol

2008 Acid catalyzed esterification of propionic acid with 1-butanol to propionic acid 1-butyl ester

Total Petroleum Hydrocarbons By Gas Chromatography in Colombian Waters and Soils

Certified Reference Material - Certificate of Analysis

SpeedExtractor E-916 Determination of Total Petroleum Hydrocarbons in Soil, Sediment, and Waste Samples using the SpeedExtractor E /2009

Separation and Quantitation of a Mixture of Alcohols by Gas Chromatography

SOP SMPS BaltimorePM supersite Version 2 November 20, 2000 Page 1 of 1. SEAS Chemistry

Transcription:

HYDROCARBONS, BP 36 216 C 1500 FORMULA: Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1 METHOD: 1500, Issue 3 EVALUATION: PARTIAL Issue 1: 15 August 1990 Issue 3: 15 March 2003 OSHA : Table 2 NIOSH: Table 2 ACGIH: Table 2 PROPERTIES: Table 1 COMPOUNDS: (Synonyms in Table 1) cyclohexane cyclohexene ndecane ndodecane nheptane nhexane methylcyclohexane nnonane noctane npentane nundecane SAMPLING SAMPLER: SOLID SORBENT TUBE [1] (coconut shell charcoal, 100 mg/50 mg) FLOW RATE: Table 3 VOLMIN: Table 3 MAX: Table 3 SHIPMENT: Routine SAMPLE STABILITY: 30 days @ 5 C BLANKS: 10% of samples ACCURACY RANGE STUDIED: Table 3 BIAS: Table 3 OVERALL PRECISION ( rt ): Table 3 ACCURACY: Table 3 MEASUREMENT TECHNIQUE: GAS CHROMATOGRAPHY, FID [1] ANALYTE: DESORPTION: INJECTION VOLUME: 1 µl Hydrocarbons listed above 1 ml CS 2 ; stand 30 min TEMPERATURES INJECTION: 250 C DETECTOR: 300 C COLUMN: 35 C (8 min) 230 C (1 min) ramp (7.5 C /min) CARRIER GAS: COLUMN: Helium, 1 ml/min Capillary, fused silica, 30 m x 0.32mm ID; 3.00µm film 100% dimethyl polysiloxane CALIBRATION: Solutions of analytes in CS 2 RANGE: Table ESTIMATED LOD: Table PRECISION ( r ): Table APPLICABILITY: This method may be used for simultaneous measurements; however, interactions between analytes may reduce breakthrough volumes and alter analyte recovery. INTERFERENCES: At high humidity, the breakthrough volumes may be reduced. Other volatile organic solvents such as alcohols, ketones, ethers, and halogenated hydrocarbons are potential interferences. OTHER METHODS: This method is an update for NMAM 1500 issued on August 15, 199 [2] which was based on methods from the 2 nd edition of the NIOSH Manual of Analytical Methods: S28, cyclohexane [3]; S82, cyclohexene [3]; S89, heptane [3]; S90, hexane [3]; S9, methylcyclohexane [3]; S378, octane []; and S379, pentane [].

HYDROCARBONS, BP 36216 C: METHOD 1500, Issue 3, dated 15 March 2003 Page 2 of 8 REAGENTS: 1. Eluent: Carbon disulfide *, low benzene, chromatographic quality. 2. Analytes, reagent grade.* 3. Helium, prepurified and filtered.. Hydrogen, prepurified and filtered. 5. Air, prepurified and filtered. * See SPECIAL PRECAUTIONS EQUIPMENT: 1. Sampler: glass tube, 7 cm long, 6mm OD, mm ID, flamesealed ends, containing two sections of activated coconut shell charcoal (front = 100 mg, back = 50 mg) separated by a 2mm urethane foam plug. A silylated glass wool plug precedes the front section, and a 3mm urethane foam plug follows the back section. Tubes are commercially available. 2. Personal sampling pumps (0.01 to 1.0 L/min, Table 3) with flexible tubing. 3. Gas chromatograph, FID, integrator, and a Rtx1 or equivalent capillary column (page 15001).. Glass autosampler vials (2m L) with PTFElined caps. 5. Pipettes (1mL) and pipette bulb. 6. Syringes (10, 25, 100, and 250 µl). 7. Volumetric flasks (10mL). SPECIAL PRECAUTIONS: Carbon disulfide is toxic and extremely flammable (flash point = 30 C). Prepare samples and standards in a wellventilated hood. SAMPLING: 1. Calibrate each personal sampling pump with a representative sampler in line. 2. Break the ends of the sam pler immediately before sampling. Attach the sampler to a personal sampling pump with flexible tubing. 3. Sample at an accurately known flow rate between 0.01 and 0.2 L/min (0.01 to 0.05 L/m in for npentane) for a total sample size as shown in Table 3.. Cap the sam plers with plastic (not rubber) caps and pack securely for shipm ent. SAMPLE PREPARATION: 5. Place the front and back sorbent sections of the sampler tube in separate vials. Include the glass wool plug in the vial with the front sorbent section. Discard the foam plugs. 6. Add 1.0 m L carbon disulfide to each vial. Attach crimp cap to each vial immediately. 7. Allow to stand at least 30 min with occasional agitation. CALIBRATION AND QUALITY CONTROL: 8. Calibrate daily with at least six working standards from below the LOD to 10 times the LOQ. Additional standards may be added to extend the calibration curve if necessary. a. Add known amounts of analytes to carbon disulfide in 10mL volumetric flasks and dilute to the mark. Prepare additional standards by serial dilution in 10mL volumetric flasks. b. Analyze with samples and blanks (steps 11 and 12). c. Prepare a calibration graph (peak area of analyte vs. µg of analyte per sample). 9. Determine the desorption efficiency (DE) at least once for each batch of charcoal used for sampling in the calibration range (step 8). a. Prepare three tubes at each of five levels plus three media blanks. b. Remove and discard the back sorbent section of a media blank sampler.

HYDROCARBONS, BP 36216 C: METHOD 1500, Issue 3, dated 15 March 2003 Page 3 of 8 c. Inject a known amount of stock solution (5 to 25 µl) directly onto the front sorbent section with a microliter syringe. d. Allow the tubes to air equilibrate for several minutes, then cap the tubes and allow to stand overnight. e. Desorb (steps 5 through 7) and analyze with standards and blanks (steps 11and 12). f. Prepare a graph of DE vs. µg analyte recovered. 10. Analyze at least three quality control blind spikes and three analyst spikes to insure that the calibration graph and DE graph are in control. MEASUREMENT: 11. Set the gas chromatograph according to the manufacturer's recommendations and to the conditions given on page 15001. Inject a 1µL aliquot manually using a solvent flushing technique or with an autosampler. NOTE: If the peak area is above the linear range of the working standards, dilute with solvent, reanalyze and apply the appropriate dilution factor in the calculations. Analyte Approximate Retention Time (min) npentane 7.5 solvent (CS 2 ) 9.6 nhexane 13.0 cyclohexane 16.1 cyclohexene 16.8 nheptane 17.7 methylcyclohexane 18.9 noctane 21.6 nnonane 2.9 ndecane 27.8 nundecane 30.5 ndodecane 32.9 NOTE: Retention times may vary slightly due to column manufacturer and age of column, and be influenced by other GC instrumental parameters. 12. Measure the peak area. CALCULATIONS: 13. Determine the mass, µg (corrected for DE), of analyte found in the sample front (W f ) and back (W b ) sorbent sections, and in the average media blank front (B f ) and back (B b ) sorbent sections. NOTE: If W b > W f / 10, report breakthrough and possible sample loss. 1. Calculate the concentration, C, of analyte in the air volume, V(L), sampled: NOTE: g/l = mg/m 3

HYDROCARBONS, BP 36216 C: METHOD 1500, Issue 3, dated 15 March 2003 Page of 8 EVALUATION OF METHOD: Issues 1 and 2: Precisions and biases (Table 3) were determined by analyzing generated atmospheres containing onehalf, one, and two times the OSHA standard. Table 3 does not contain data for ndecane, ndodecane and n heptane since they were not evaluated previously. Generated concentrations were independently verified. Breakthrough capacities were determined in dry air. Storage stability was assessed at 7, 1, and 30 days. Measurement precisions (Table ) were determined by spiking sampling media with amounts corresponding to onehalf, one, and two times the OSHA standard for nominal air volumes. Desorption efficiencies for spiked sam plers containing only one compound exceeded 75% [2,3,,8]. Issue 3: The desorption efficiency, at levels ranging from 10 times the LOQ to 0.1 times the REL, was determined by spiking known amounts of analytes (in CS 2 ) on coconut shell charcoal tubes. All analytes exhibited acceptable desorption efficiency recovery results at six levels evaluated. Each analyte was evaluated for its storage stability. Sorbent tubes were spiked at approximately 100 µg and stored in a drawer for 7 days, then transferred to a refrigerator at 5 C. Samples were analyzed after 7, 1, and 30 days. All analytes had acceptable recoveries (>90%), except cyclohexene, which had a 30 day recovery of 85% [1]. REFERENCES: [1] Pendergrass SM and May L, Backup Data Report, ACS/CEMB/DART/NIOSH (1999). [2] NIOSH[199]. Hydrocarbons, BP 36136 C. In: Eller PM, ed. NIOSH Manual of Analytical Methods, th rev. ed. Cincinnati, OH: U.S. Department of Health and Human Services, Public Health Service, Centers for Disease Control, National Institute for Occupational Safety and Health, DHHS (NIOSH) Publication No. 9113. [3] NIOSH Manual of Analytical Methods, 2nd. ed., V. 2, S28, S82, S89, S90, S9, U.S. Department of Health, Education, and Welfare, Publ. (NIOSH) 77157B (1977). [] NIOSH Manual of Analytical Methods, 2nd. ed., V. 3., S378, S379, U.S. Department of Health, Education, and Welfare, Publ. (NIOSH) 77157C (1977). [5] Code of Federal Regulations; Title 29 (Labor), Parts 1900 to 1910; U.S. Government Printing Office, Washington, (1989); 29 CFR 1910.1000. [6] NIOSH Recommendations for Occupational Safety and Health. U.S. Department of Health and Human Services, DHHS (NIOSH) Publication No. 92100 (1992). [7] 1993 Threshold Limit Values for Chemical Substances and Physical Agents and Biological Exposure Indices. ACGIH, Cincinnati, OH (1993). [8] Documentation of the NIOSH Validation Tests, S28, S82, S89, S90, S9, S378, S379, U.S. Department of Health, Education, and W elfare, Publ. (NIOSH) 77185 (1977). AUTHORS: Issues 1 and 2: R. Alan Lunsford, Ph.D., NIOSH/DPSE Issue 3: Stephanie M. Pendergrass, NIOSH/DART and Leroy R. May, NIOSH/PRL

HYDROCARBONS, BP 36216 C: METHOD 1500, Issue 3, dated 15 March 2003 Page 5 of 8 TABLE 1. SYNONYMS, FORMULA, MOLECULAR WEIGHT, PROPERTIES Name/ Synonyms Empirical Formula Molecular Weight Boiling Point ( C) Vapor Pressure @ 25 C (mm Hg) (kpa) Density @ 25 C (g/ml) cyclohexane hexahydrobenzene CAS # 110827 RTECS GU6300000 cyclohexene tetrahydrobenzene CAS # 110838 RTECS GW 2500000 ndecane CAS # 12185 RTECS HD6550000 ndodecane CAS # 11203 RTECS JR2125000 nheptane CAS # 12825 RTECS MI7700000 nhexane CAS # 11053 RTECS MN9275000 methylcyclohexane CAS # 108872 RTECS GV6125000 nnonane CAS # 11182 RTECS RA6115000 noctane CAS # 111659 RTECS RG800000 npentane CAS # 109660 RTECS RZ950000 nundecane hendecane CAS # 112021 RTECS YQ1525000 C 6 H 12 8.16 80.7 97.6 13.0 0.779 C 6 H 10 82.15 83.0 88.8 11.8 0.811 C 10 H 22 12.28 17 NA NA 0.730 C 12 H 26 170.3 216.2 NA NA 0.750 C 7 H 16 100.21 98. 5.8 6.1 0.68 C 6 H 1 86.18 68.7 151.3 20.2 0.659 C 7 H 1 98.19 100.9 6.3 6.2 0.769 C 9 H 20 128.26 151 NA NA 0.718 C 8 H 18 11.23 125.7 1.0 1.9 0.703 C 5 H 12 72.15 36.1 512.5 68.3 0.626 C 11 H 2 156.31 196 NA NA 0.70

HYDROCARBONS, BP 36216 C: METHOD 1500, Issue 3, dated 15 March 2003 Page 6 of 8 TABLE 2. EXPOSURE LIMITS, PPM [57] OSHA NIOSH ACGIH mg/m 3 Substance TWA PEAK TWA C TLV STEL per ppm cyclohexane 300 300 300 3. cyclohexene 300 300 300 3.36 ndecane none none none 5.82 ndodecane none none none 6.97 nheptane 500 85 0 00 500.10 nhexane a 500 50 50 3.52 methylcyclohexane 500 00 00.01 nnonane none 200 200 5.25 noctane 500 75 385 300 375.67 npentane 1000 120 610 600 750 2.95 nundecane none none none 6.39 a The ACGIH recommendation for other hexane isomers is: TLV 500, STEL 1000.

HYDROCARBONS, BP 36216 C: METHOD 1500, Issue 3, dated 15 March 2003 Page 7 of 8 TABLE 3. SAMPLING FLOWRATE a, VOLUME, CAPACITY, RANGE, OVERALL BIAS AND PRECISION [3,, 8] Sampling Range of Overall Flowrate Volume (L) Breakthrough Volume Generated Samples Bias Precision Accuracy Vol Concentration Substance (L/min) MIN MAX b (L) (mg/m 3 ) (mg/m 3 ) (%) ( rt ) (%) cyclohexane cyclohexene nhexane methylcyclohexane nnonane noctane npentane nundecane 0.01 0.2 0.01 0.2 0.01 0.2 0.01 0.2 0.01 0.2 0.01 0.2 0.010.2 0.010.05 2.5 5 not 2 5 7 studied 2 7.6 10. 6.1 5.9 6.1 6.5 3.1 1650 2002 060 3679 391 612 560 5102010 5102030 968060 8773679 90391 105003 1766190 1.1 10.6 6.5 1.8 6.1 2.0 8. 0.060 c 0.073 0.056 0.062 0.052 0.060 0.055 ±11.5 ±20.7 ±15.0 ±12.5 ±15.2 ±12.1 ±16.6 a Minimum recommended flow is 0.01 L/min. b Approximately twothirds the breakthrough volume. c Corrected value calculated from data in Ref. 3

HYDROCARBONS, BP 36216 C: METHOD 1500, Issue 3, dated 15 March 2003 Page 8 of 8 TABLE. MEASUREMENT RANGE AND PRECISION [1, 3,, 8] Substance LOD (µg/sample) Range (µg) Measurement Precision ( r ) cyclohexane a 0.1 5300 0.012 cyclohexene a 0.08 3 9700 0.01 ndecane 0.06 2 58 0.020 ndodecane 0.05 2 600 0.027 nheptane 0.06 2 16300 0.01 nhexane 0. 10 1500 0.011 methylcyclohexane 0.1 16100 0.013 nnonane 0.0 1 57 0.018 noctane 0.3 11 18900 0.022 npentane 0.6 19 11800 0.012 nundecane 0.05 2 592 0.02 a Corrected value, calculated from the data in [1,8].