Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging

Similar documents
Patient doses from CT examinations in region of Prishtina, Kosovo

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

Leading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children

Study of Image Quality, Radiation Dose and Low Contrast Resolution from MSCT Head by Using Low Tube Voltage

Dose comparison between Ziehm Vision RFD 3D, Medtronic O-arm O2 and Samsung NeuroLogica BodyTom

2017 ACR Computed Tomography Quality Control Manual FAQS

Optimization of Parameters in 16-slice CT-scan Protocols for Reduction of the Absorbed Dose

Size-based protocol optimization using automatic tube current modulation and automatic kv selection in computed tomography

Jessop M*(1), Thompson JD(2), Coward J(2), Sanderud A(3), Jorge J(4), Groot M de(5), Lança L(6), Hogg P(2).

CQIE PET PROCEDURES. American College of Radiology Clinical Research Center. Centers for Quantitative Imaging Excellence LEARNING MODULE

40TH ANNUAL MEETING. CTA Dose Reduction: Special Considerations in Children. Jeffrey C. Hellinger, MD FACC. October 13 16, 2012 Pasadena, CA

Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals. Koos Geleijns, medical physicist

SPECIFICATION CT SIMULATOR FOR BLACK LION HOSPITAL, ADDIS ABEBA, ETHIOPIA

MEASURING RADIATION DOSE IN COMPUTED TOMOGRAPHY USING ELLIPTIC PHANTOM AND FREE-IN-AIR, AND EVALUATING ITERATIVE METAL ARTIFACT REDUCTION ALGORITHM

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

AAPM ACTIVITIES WITH RESPECT TO CT IMAGING Cynthia McCollough, PhD, DABR, FAAPM, FACR, FAIMBE

The Latest in Radiation Dose Reduction Techniques in CT

Medical Radiation Reporting: A Manufacturer s Perspective

Image Gently Development of Pediatric CT Protocols 2014

Clarity CT Technology

Patient-centered CT imaging: New methods for patient-specific optimization 1 of image quality and radiation dose

Low-dose and High-resolution Cardiovascular Imaging with Revolution* CT

CALIFORNIA LAW AND RESPONSE + FIVE QUICK LESSONS ON CT DOSE OPTIMIZATION

High performance comes easily

PERFORM Operating Document. Radiation Dose Estimates, Calculations and Tracking Methodology

Not for publication in the USA Erlangen, November 26, 2017

IAEA-TECDOC Dose Reduction in CT while Maintaining Diagnostic Confidence: A Feasibility/Demonstration Study

Handzettel 1. CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014

NEMA XR Standard Attributes on CT Equipment Related to Dose Optimization and Management

GE Healthcare. Optima NM/CT 640

QIBA Profile: Small Lung Nodule Volume Assessment and Monitoring in Low Dose CT Screening

8/1/2018. Optimization Strategies for Pediatric CT Imaging. Educational Objectives. Brain CT

CT QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring ACCREDITATION CT IMAGE QUALITY QA CTDI CT CHARACTERIZATION CT PERFUSION QA

Toward Large-Scale Process Control to Enable Consistent CT Radiation Dose Optimization

Instructions and Checklist for the Preparation of Data for the Pediatric Administered Activity Survey

Quality control of CT systems by automated monitoring of key performance indicators: a two-year study

New Material for Shields in X-Ray Diagnostic Procedures

From dose to risk: an uncertain predicament or an ethical mandate? Ehsan Samei Department of Radiology Duke University Medical Center

Computed tomography. High. performance. no trade-offs. Philips Ingenuity CT family

Accuracy of Patient-Specific Organ Dose Estimates Obtained Using an Automated Image Segmentation Algorithm

QIBA Profile: Small Lung Nodule Volume Assessment and Monitoring in Low Dose CT Screening

Optimizing Radiation Dose for Samsung GC85A

3D-Printing Patient-Specific Phantoms For Imaging and Radiation Dosimetry: Recent Progress, Challenges, and Future Directions

Trusted Performance. Smart Investment. 80 detector row Ultra Helical CT

Equipment Quality Control for Fluoroscopy February 22, Imaging Physics CancerCare Manitoba

Device Improvements for Pediatric X-Ray Imaging; Public Meeting; Request for Comments

RADIATION DOSE MONITOR. The DACS for Building a Low-Dose Culture

Part 3 Oral Exam Content Guide

RADIATION DOSE MONITOR

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels

7/28/2017. Michael Speidel. University of Wisconsin - Madison

INVALUABLE VERSATILITY. GO DIGITAL. YOUR WAY. DRX-Evolution

Impact of miscentering on patient dose and image noise in x-ray CT imaging: Phantom and clinical studies

Jonathon A. Nye, Ph.D Emory University School of Medicine Atlanta, GA

Optimisation in CT. A case for shared approach. Stephen J. Golding University of Oxford

FLORIDA HOSPITAL DIAGNOSTIC RADIOLOGY RESIDENCY PROGRAM PHYSICS GOALS AND OBJECTIVES

TENDER # SUPPLY OF MOBILE C-ARM RADIOGRAPHY/FLUOROCOPY SYSTEM FOR WESTERN HEALTH

Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences

AAPM IGRT Recommendations

Radiation Protection Dosimetry (2008), Vol. 129, No. 4, pp Advance Access publication 30 October 2007

Discovery NM630 Innovation to expand your care.

UNIVERSITY OF GHANA COLLEGE OF BASIC AND APPLIED SCIENCES

Physicists Quality Control for MR Equipment

By completing this questionnaire, you are indicating your consent to participate in this research.

CT QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring

Assessment of volumetric absorbed dose for mobile fluoroscopic 3D image acquisition

Precision Measurement of Small Gap within Closed Components by Industrial Computed Tomography

Easy that excels Discovery RF180

} } 8/3/2016. Accounting for kv Imaging Dose. kilovoltage imaging devices/techniques. Current imaging dose determination methods

Table of Contents. Adaptive Diagnostics...7 Integrated Dose Reduction Streamlined Workflow Clinical Images... 26

On the Importance of Computation in Clinical Radiology

How Tracking Radiologic Procedures and Dose Helps: Experience From Finland

Fluoroscopy units come in many different physical configurations but all consist of the main components shown in this slide.

Imaging for Proton Treatment Planning and Verification

Physical structure of matter. Monochromatization of molybdenum X-rays X-ray Physics. What you need:

New PET/CT from Siemens helps more patients benefit from premium technologies

VOLUME 17 ISSUE ISSN = H.E. Henriksen. T. Laurence. D. Caramella et al. m.p. culp, J.s. Harford, s.g. Jordan

Donna Newman B.A. RT ( R) CNMT Director of Professional Practice ISRRT

TG-104: In-room kv Computed Tomography for Image-guidance

THE NEW 640 SLICE CT SCANNER

R/F. Experiences Using SONIALVISION safire and the Utility of Tomosynthesis. 1. Introduction. 2. Basics of Tomosynthesis.

SPECIFICATION. SPECT/CT Gamma Camera System

Survey of Radiation Exposure for adult and pediatric patients undergoing CT procedures in Latin America

The FUTURE has ARRIVED NeuViz 128

Radiation Dose Terms Made Simple v1

Philips PET/CT Gemini GXL. Total performance. Total confidence.

The Utility of Patient-Specific CT Dose Estimation Maps

Optimizing the Use of Pulsed Fluoroscopy to Reduce Radiation Exposure to Children

Joint Commission Medical Imaging Requirements. Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Department of Medical Imaging SPR Annual Meeting

Patient radiation dose and image quality optimization in CT. Optimiser la dose patient et la qualité image en scanner

Quality Control Concerns for New Generations of CT, MRI, & PET

Disclaimer and Notices.

LASER ULTRASONICS INSPECTIONS OF AERONAUTICAL COMPONENTS VALIDATED BY COMPUTED TOMOGRAPHY

More flexibility More performance Less dose

IMAGE GUIDANCE DOSES IN RADIOTHERAPY. Scott Crowe

A Methodology for Optimizing the Acquisition Time of a Clinical PET Scan Using GATE

IMRT PHANTOMS CIRS IMRT HOMOGENEOUS PHANTOM CIRS IMRT PHANTOMS C-1

IV. 1. Skin Dose Measurement for Patients Using Imaging Plates in Interventional Radiology Procedures

GENESIS Edition Transforming CT

The use of CT guidance for spine steroid injections has increased

Transcription:

Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging Hanna Piwowarska-Bilska 1, Leszek J. Hahn 2, Bozena Birkenfeld 1, Katarzyna Cichon-Bankowska 1, Maria H. Listewnik 1, and Piotr Zorga 1 1 Department of Nuclear Medicine, Pomeranian Medical University, Szczecin, Poland; and 2 Department of Diagnostic Imaging, Foothills Medical Centre, Calgary, Alberta, Canada This study was performed to find the optimal low-dose CT protocol for children being imaged on SPECT/CT scanners not equipped with automatic dose control. For SPECT/CT systems with manually adjustable x-ray tube voltage (kv) and anode current (ma), an optimized protocol makes it possible to minimize the dose to patients. Methods: Using the 4-slice low-dose CT component of a commercially available SPECT/ CT scanner, we compared the signals reaching the CT detector after radiation passes through objects of different sizes. First, the exit dose rates were measured for combinations of available voltages and currents. Next, imaging parameters were selected on the basis of acceptable levels of exit dose rates, cylindric phantoms of different diameters approximating children of different sizes were scanned using these parameters, and the quality of the CT images was evaluated. Finally, weighted CT dose indexes for abdomen and head CT dose phantoms simulating, respectively, adult and pediatric patients were measured using exactly the same techniques to estimate and compare doses to these 2 groups of patients. Results: For children with torsos smaller than 150 mm, imaging can be performed using the lowest available voltage and current (120 kv and 1 ma, respectively). For children with torsos less than 250 mm, 140 kv and 1.5 ma can be used. For patients with torsos greater than 250 and less than 300 mm, 140 kv and 2 ma can be used. Regarding the signal-to-noise ratio, all these parameters give an excellent signal and fully acceptable noise levels. Conclusion: For the SPECT/CT system studied, even the lowest available voltage and current used for scanning pediatric patients did not cause signal-to-noise degradation, and the use of these settings substantially lowered the dose to the patients. Key Words: low-dose CT; SPECT/CT; pediatric radiation exposure; CT dose optimization J Nucl Med Technol 2010; 38:181 185 DOI: 10.2967/jnmt.109.073486 Diagnostic SPECT/CT systems allow CT scans to be registered with nuclear medicine scans so that radioisotope Received Dec. 30, 2009; revision accepted Jul. 2, 2010. For correspondence or reprints contact: Hanna Piwowarska-Bilska, Department of Nuclear Medicine, Pomeranian Medical University, ul. Unii Lubelskiej 1, 72-252 Szczecin, Poland. E-mail: hannap@pam.edu.pl COPYRIGHT ª 2010 by the Society of Nuclear Medicine, Inc. uptake can be localized. CT scans are additionally used for attenuation correction of radioisotope images. However, compared with SPECT-only examinations, SPECT/CT examinations increase the radiation exposure of patients. SPECT/CT systems with adjustable x-ray tube voltage (kv) and anode current (ma) make it possible to minimize the dose to patients. Reasonable reduction of CT-related dose in individual patients is an important issue, particularly in children (1,2). The aim of this study was to find the optimal pediatric CT protocol that reduces radiation risk. MATERIALS AND METHODS All investigations used the 4-slice low-dose CT component of the Infinia Hawkeye 4 SPECT/CT system (GE Healthcare). In this system, both the voltage and the current are adjustable. To compare the signals reaching the CT detector after radiation passes through objects of different sizes, we measured the exit dose for all combinations of available voltages and currents. To simulate patients of different sizes, we used phantoms containing various volumes of water. During all measurements, the x-ray tube was immobilized at the 12-o clock position in the scanner service mode. To validate this approach and to find the relationship between exit dose rates (detector signal) and CT image quality, we scanned 4 cylindric phantoms filled with water. For this purpose, the system was switched from stationary to tomographic mode, and noise levels representing signalto-noise ratios in CT images of these phantoms were measured. The diameters of these phantoms corresponded to children of different weights and sizes. To estimate the doses to large and small patients imaged at variable voltages and currents, we scanned the abdomen and head phantoms and measured the weighted CT dose index (CTDI w ). Measurement of Exit Dose Radiation exiting the water phantom and reaching the CT detector was measured using a Dosemeter/Ratemeter 2026C (Radcal Corp.) ionization chamber, 20X6-3. For this purpose, 8 combinations of available voltage (120 or 140 kv) and current (1, 1.5, 2, or 2.5 ma) were used at a constant time of 23 s, which corresponds to the time of full gantry rotation in the CT mode of scanning. To estimate the strength of the signal reaching the CT detectors, we PROTOCOL OPTIMIZATION OF LOW-DOSE CT Piwowarska-Bilska et al. 181

calculated dose rates by dividing the measured cumulative dose by that time. An acrylic water-filled cylinder was used as a phantom simulating attenuation by the human body. The cylinder was 330 mm in height and 210 mm in diameter. The cylinder was placed on the patient table, vertically (Fig. 1), with its center coinciding with the geometric center of the x-ray fanbeam. The ionization chamber was positioned directly under the cylinder at its center. The water level in the cylinder was increased in steps of 50 mm, from 0 mm to a maximum of 300 mm. All these measurements were performed twice to check repeatability. Overall, 168 measurements of exit dose were performed under the unchanged geometry. Noise Level Evaluation Noise levels for calculation of signal-to-noise ratios (3) were evaluated by imaging sets of water-filled cylinders 270, 200, 160, and 100 mm in diameter in SPECT/CT mode using a bone SPECT protocol. Voltage and current were identical to those used for the exit dose measurements. The reference to which the data obtained from all scanned cylinders were compared corresponded to an adult patient, as represented by a Data Spectrum torso anthropomorphic phantom. The water-filled part of this phantom was scanned at 140 kv and 2.5 ma, and the time of rotation was 23 s. It was assumed that these parameters were tested in the factory and then selected as optimal for adult patient imaging. The slice thickness and collimation are not adjustable for this SPECT/CT system and for all scans were 5 and 20 mm, respectively. For data analysis, circular regions of interest were used. One large region of interest covering about 80% FIGURE 1. geometry. Schematic illustration of exit dose measurement of the image of a scanned cylinder was used to calculate the average SD in the image. This SD represented the noise level for a particular set of parameters and cylinder scan. To evaluate the range of SDs in different parts of the CT image, the set of smaller regions of interest with a diameter equal to about 40 mm was placed circumferentially (12-, 3-, 6-, and 9-o clock positions) and along the x- and y-axes of CT images of cylinders and torso phantom. SDs in selected regions of interest for a particular slice were calculated using the system software and recorded from a region-ofinterest statistics database. Measurements of CTDI w The most commonly used dose descriptor is CT dose index. The CTDI w, measured at the center and periphery of CT dose index phantoms, provides a convenient estimate of patient dose for a procedure (4,5). CTDI w is defined as follows: CTDI w 5 ð1=3þðctdi 100 Þ center 1ð2=3ÞðCTDI 100 Þ periphery : Eq. 1 CTDI 100 is the integral of the dose profile along a line perpendicular to the tomographic plane divided by the product of the nominal tomographic section thickness and the number of tomograms produced in a single axial scan. Because the ionization chamber measures an integrated exposure along its 100-mm length, CTDI 100 can be calculated according to the following formula: CTDI 100 5 ðf EÞðL=TÞ: Eq. 2 f is the conversion factor from exposure (mr) to a dose in air (mgy), E is the measured value of exposure, L is the active length of the pencil ionization chamber, and T is the nominal width of each acquired section. For CT dose index measurements, a CT dose probe was used: a 10X5-3CT pencil-type ionization chamber with a sensitive length of 100 mm (Radcal). Two phantoms were used: solid polymethylmethacrylate disks measuring 160 mm in diameter (head phantom simulating a child s body) and 320 mm in diameter (abdomen phantom simulating an adult s body), both 140 mm thick. There are 5 holes in each of the phantoms, 4 circumferential (4 mm from the edge) and 1 central, with acrylic rods to plug the holes when not in use. The holes are 13.1 mm in diameter to accommodate the size of a standard CT probe. Because the main aim of this work was to evaluate the dose reduction when small patients (children) are scanned using the lowest available current and voltage, only 2 sets of scanning parameters were tested. First, 140 kv and 2.5 ma (maximal available values) were used for scanning the abdomen and head phantoms. These parameters would correspond to the use of a nonadjustable scanner, independent of a patient s size. Then, the lowest available voltage (120 kv) and current (1 ma) were used for scanning the head phantom only, which would correspond to a scanner that is adjustable according to the weight or size of smaller patients. 182 JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY Vol. 38 No. 4 December 2010

TABLE 1 Average Exit Dose Rates Exit dose rate (10 22 mgy/s)* Voltage (kv) Current (ma) 0 mm 50 mm 100 mm 150 mm 200 mm 250 mm 300 mm 140 2.5 2,138 1,100 522 235 110 48 24 140 2.0 1,705 887 426 193 87 39 18 140 1.5 1,295 675 323 148 67 29 15 140 1.0 883 460 218 105 44 20 9 120 2.5 1,396 713 332 147 65 29 13 120 2.0 1,126 574 264 117 53 23 10 120 1.5 853 417 201 87 40 17 7 120 1.0 580 280 140 61 28 12 5 *Data are for the stated different depths of water in a cylindric phantom simulating attenuation in a patient. Rates are calculated by dividing cumulative exit doses by time of measurement. RESULTS Table 1 shows the exit dose rates. These calculations used the arithmetic average of 3 cumulative exit doses measured with the 20X6-3 ionization chamber. Because the dispersion of measured cumulative exit doses was small (coefficient of variation, 1%), it is not shown in Table 1. Figure 2 is the graphic representation of the results in Table 1. The shaded area represents parameters giving a detector signal equal to or higher than that produced using the maximum voltage and current available. Table 2 shows noise levels in CT images of cylindric phantoms of different diameters. They represent children of varying sizes scanned using different parameters. These measurements helped to justify the validity of using the exit dose values as a basis for selecting parameters adequate for a child s size. CTDI w results for both CT dose phantoms are shown in Table 3. Table 4 contains CT acquisition protocols proposed for technologists performing patient studies on the Infinia Hawkeye 4. The information on weight, age, and corresponding patient dimensions is based on the available literature (6). In this study, the arithmetic mean of the chest and abdomen dimensions was used as the average patient size because all measurements of CT dose index and noise were FIGURE 2. Semilogarithmic graph of exit dose rates for various voltages and currents vs. increasing depth of water in phantom simulating attenuation in patient. PROTOCOL OPTIMIZATION OF LOW-DOSE CT Piwowarska-Bilska et al. 183

TABLE 2 Signal-to-Noise Results for CT Images of Water Phantoms Phantom diameter (mm) Voltage (kv) Current (ma) Reference phantom* 270 200 160 100 140 2.5 42 (35 52) 27 (23 34) 18 (15 21) 11 (9 14) 140 2.0 30 (27 35) 21 (22 25) 12 (10 15) 140 1.5 35 (26 45) 25 (23 32) 14 (12 16) 140 1.0 43 (36 54) 29 (28 33) 16 (14 19) 120 2.5 33 (27 43) 22 (21 26) 13 (11 16) 120 2.0 37 (25 49) 25 (24 31) 15 (13 16) 120 1.5 43 (37 52) 29 (27 34) 17 (15 19) 120 1.0 53 (44 69) 36 (33 42) 20 (17 23) 11 (10 12) *Torso phantom (260 380 mm). Data are SDs in Hounsfield units, followed by ranges in parentheses. performed on cylindric phantoms. The relationship between body size and weight for adolescents was extrapolated from the data on children and adult patients because of a lack of relevant data in the literature. The results from Table 1 and information about the relationship between children s ages, weights, and sizes were the basis for the proposed low-dose CT acquisition protocols. DISCUSSION The results of this study indicate the possibility of substantially decreasing doses to patients without losing the detector signal. Thus, there is a chance that the risk in children imaged with SPECT/CT systems can be reduced. Table 1 shows that for water layers thinner than 200 mm, the exit dose measured at the minimum available voltage and current (120 kv, 1 ma, respectively) is stronger than that detected when a 300-mm-thick layer of water is penetrated by radiation generated using maximum voltage and current (140 kv and 2.5 ma, respectively). Drawing a line parallel to the water level in cylinder axis to intersect the object exit dose axis at a point corresponding to the detected exit dose when using 140 kv and 2.5 ma for objects 300 mm in diameter (which corresponds to penetration of a 300-mm-thick water layer), one can find all the parameters for all object diameters that give a signal equal to or greater than this first one (dashed area in Fig. 2). On this basis, parameters optimizing doses to patients of different sizes can be selected. Thus, for infants 150 mm in diameter or smaller, the technique utilizing the lowest available voltage and current can be used because it still gives an excellent detector signal. For patients greater than 150 mm and up to 200 mm in diameter, 120 kv and 2 ma are sufficient. Patients greater than 200 mm and up to 250 mm can be scanned at 140 kv and 1.5 ma. For patients from 250 to 300 mm (adolescents), scanning at 140 kv and 2 ma is suitable. The fact that these parameters translate into a substantial reduction in doses to children and smaller patients is an important message to users of SPECT/CT systems with adjustable settings for voltage and current. The signal-to-noise results for scanning various sizes of water-filled phantoms with various parameters (Table 2) supports the exit-dose-level criterion as a basis for selecting the imaging parameters. The noise level in the reference image (torso phantom) averaged 42 and ranged from 35 (minimum, the best) to 52 (maximum, the worst). This level is comparable to results obtained for the phantom with a diameter of 270 mm (the largest) scanned at 140 kv and 1 ma: the noise level in that case averaged 43 and ranged from 36 to 54. The noise levels for all remaining phantoms scanned, even at 120 kv and 1 ma, were better than the reference level. Additionally, comparison with the data from Table 1 and Figure 2 shows that using the exit dose rate as a basis for selection of scanning parameters is actually more restrictive than would be selection based on analysis of noise levels only. The dose reduction can clearly be seen from the CTDI w results for the various scanning parameters and for the 2 phantoms (Table 3). When one switches from 140 kv and 2.5 ma to 120 kv and 1 ma for children 150 mm in diameter, the dose decreases from 7.23 mgy to 2.01 mgy. For TABLE 3 CTDI w for Various Scanning Parameters and Phantoms CT parameters CTDI w for abdomen phantom (mgy) CTDI w for head phantom (mgy) 120 kv; 1 ma; axial scanning; collimation, 20 mm; slice thickness, 5 mm 140 kv; 2.5 ma; axial scanning; collimation, 20 mm; slice thickness, 5 mm 2.01 5.42 7.23 184 JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY Vol. 38 No. 4 December 2010

TABLE 4 Proposed CT Protocol for SPECT/CT in Children Relationship between patient average diameter* and weight (mm/kg) Patient weight (kg) Voltage (kv) Current (ma) 110/4,10 120 1 150/11 10 19 120 2 200/20 20 29 140 1.5 250/35 30 40 140 2 300/70.40 140 2.5 *Arithmetic mean of diameters of thorax and abdomen. For adolescents, diameter/weight relation was extrapolated because of lack of data in handbook (6). infants smaller than 150 mm, the dose reduction would be even greater. The results presented in this article should be instructive for producers of cheaper SPECT/CT systems in which the CT component is not equipped with automatic dose control. There is definitely some room for improvement. For imaging of small children, the mas as a combination of changes in kv, ma, and times of data acquisition can be decreased further. The CTDI w is 3.6-fold higher when the parameters are not adjustable than when the parameters can be properly adjusted to the weight and height of the patient. Table 4 presents our proposed protocol for the CT part of the acquisition on an Infinia Hawkeye 4. For clinical acquisitions, technologists can set the optimal current and voltage based on patient size and weight. CONCLUSION The radiation dose to children imaged with low-dose CT as part of SPECT/CT studies can be reduced. Proper reduction of the voltage and current of the Infinia Hawkeye 4 for patients of smaller size and weight does not decrease the signal strength reaching the CT detectors and does not affect image quality. REFERENCES 1. Committee to Assess Health Risks from Exposure to Low Levels of Ionizing Radiation, National Research Council of the National Academies. Health Risks from Exposure to Low Levels of Ionizing Radiation: BEIR VII, Phase 2. Washington, DC: National Academies Press; 2006. 2. Brenner DJ, Hall E. Computed tomography: an increasing source of radiation exposure. N Engl J Med. 2007;357:2277 2284. 3. Sprawls P. CT image detail and noise. Radiographics. 1992;12:1041 1046. 4. Bongartz G, Golding SJ, Jurik AG, et al. European Guidelines for Quality Criteria for Computed Tomography: EUR 16262. Available at: http://www.drs.dk/ guidelines/ct/quality/. Accessed August 30, 2010. 5. Shope TB, Gagne RM, Johnson GC. A method for describing the doses delivered by transmission x-ray tomography. Med Phys. 1981;8:488 495. 6. Kereiakes GJ, Rosenstein M. Handbook of Radiation Doses in Nuclear Medicine and Diagnostic X-Ray. Boca Raton, FL: CRC Press; 1980:187, 214. PROTOCOL OPTIMIZATION OF LOW-DOSE CT Piwowarska-Bilska et al. 185