Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Similar documents
Dimethylphenol Method Method LR (0.23 to mg/l NO 3 N or 1.00 to mg/l NO 3 ) TNTplus 835

Phosphorus, Reactive (Orthophosphate)

Method (0.1 to 8.0 mg/l Cu) TNTplus 860

Nitrogen, Simplified TKN (s-tkn ) DOC

Nitrogen, Simplified TKN (s-tkn )

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

USEPA 1 Periodate Oxidation Method 2 Method to 20.0 mg/l Mn (HR) Powder Pillows

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

1-(2-Pyridylazo)-2-Naphthol PAN Method 1 Method to mg/l Mn (LR) Powder Pillows

Oxygen Demand, Chemical

Oxygen Demand, Chemical

USEPA 1,2 Bicinchoninate Method 3 Method 8506 (CuVer 1) and Method 8026 (CuVer 2) 0.04 to 5.00 mg/l Cu Powder Pillows or AccuVac Ampuls

Method 8017 (0.7 to 80.0 µg/l) Powder Pillows. Scope and Application: For water and wastewater; digestion is required to determine total cadmium.

Scope and application: For water, wastewater, seawater, brine solutions, produced waters and hydraulic fracturing waters.

Hardness, Calcium/Magnesium

Oxygen Demand, Chemical

Instrument Adapters Light shield DR 6000, DR 5000 DR 3900 LZV849 DR 3800, DR 2800 LZV646 DR or (A)

Scope and application: For water and wastewater. Turbid samples are filtered during the analysis.

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Method to 500 µg/l CH 2 O Powder Pillows

Table 163 Instrument-specific information. Instrument Pour-thru Kit Cell orientation Adapter. DR 6000 LQV Arrow faces right

Phosphorus, Reactive (Orthophosphate)

Color, True and Apparent, LR

Sulfate DOC SulfaVer 4 Method 1. Method (0 70, and mg/l) Powder Pillows. Test procedure

Phosphorus, Reactive (Orthophosphate)

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement

DR/4000 PROCEDURE. IRON, Ferrous. Using Powder Pillows

Sulfate. USEPA 1 SulfaVer 4 Method 2 Method to 70 mg/l SO 4. Powder Pillows. Test preparation. Before starting.

Oxygen Demand, Chemical

Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement, 0 to 40 mg/l COD

FerroVer Method (Powder Pillows or AccuVac Ampuls) USEPA approved for reporting wastewater analysis (digestion is required; see Section 2 * )

TPTZ Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

Direct ISE Method Method to 4.00 mg/l NO 3 N Nitrate ISE

PHOSPHORUS, REACTIVE (0.00 to 5.00 mg/l PO 4

1,5-Diphenylcarbohydrazide Method* (Powder Pillows or AccuVac Ampuls) USEPA accepted for wastewater analyses** Using Powder Pillows

1,10 Phenanthroline Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows

STP (off-line) Water Technologies & Solutions analytical procedure. colorimetric method DCA reagent 1 (5/5/1) 2 to 15 mg/l as STP

Amperometric Titrator assembly. Digital Titrator. Beaker, low-form, 250-mL. Stir bar, octagonal, Teflon-coated, 50.8 x 7.9 mm

3. Add 0.4 ml of. 7. Use a TenSette

Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric)

3. Add 0.4 ml of. 7. Use a TenSette

PaSEC Colorimeter Instructions DR850

USEPA Membrane Filtration Method Method m-tec. Scope and Application: For potable water, nonpotable water, recreation water and wastewater.

Direct ISE Method Method to 10.0 mg/l F Fluoride ISE

Oxygen Demand, Biochemical

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

Coliforms, Total and E. coli

Background Statement for SEMI Draft Document 4795 (Yellow Ballot) NEW STANDARD: SPECIFICATIONS FOR NITRIC ACID, USED IN PHOTOVOLTAIC APPLICATIONS

Heterotrophic Bacteria

Cuvette Test LCK 555 BOD 5 /BOD [n]

DR/4000 PROCEDURE OXYGEN SCAVENGERS. 4. Fill a second sample cell (the blank) with 25 ml of deionized water.

Surface Water Samples: Containers, Preservation and Hold Times Table

Cuvette Test LCK 555 BOD 5 /BOD [n]

Cuvette Test LCK 555 BOD 5 /BOD [n]

Analysis of dissolved nitrite (NO 2- )

METHOD #: Approved for NPDES (Issued 1971) Oxygen, Dissolved (Modified Winkler, Full-Bottle Technique) ANALYTE: CAS # O Oxygen

Ground Water & UST Samples: Containers, Preservation and Hold Times Table

Why Hach Reagents? Alphabetical Listing of Reagents

INSTRUCTION MANUAL. ba76126e01 03/2013. photoflex Series TEST INSTRUCTIONS AND SPECIFICATIONS, PROG. V 2.02 Y

Test sticks and test papers for semi-quantitative determinations

Professional Boiler Water Test Kit

SOP-C-102 Determination of Chemical Oxygen Demand

Uranium Testing and Site Calibration for Water Testing Detection Range: 2 60ppb

Cadmium Testing and On Site Calibration for Water Testing Detection Range: ppm

OXIDATION-REDUCTION TITRATIONS-Permanganometry

BONDERITE NT-1 Phosphate-free, Regulated Heavy Metal-Free Pre-treatment

Background The document will be revised for its five year review and updated per the committee agreed upon changes.

Zinc Testing and On Site Calibration for Water Testing Detection Range: 1 15ppm

National Food Safety Standard

E.Z.N.A. Water DNA Kit. D preps D preps D preps

HUMAN PLASMA (SOLUBLE) GELSOLIN ELISA KIT

Suspended Solids Quality Test for Bleach Using Vacuum Filtration

The image part with relationship ID rid4 was not found in the file. Welcome

CTRP15/MYONECTIN (HUMAN) ELISA KIT

Standard Methods for the Examination of Water and Wastewater

MOUSE OPG ELISA KIT PAGE

CEL212 Environmental Engineering

HUMAN CONNECTIVE TISSUE GROWTH FACTOR (CTGF) ELISA KIT

Standard Operating Procedure for GLNPO Turbidity: Nephelometric Method

HUMAN HIGH SENSITIVE C- REACTIVE PROTEIN (CRP) ELISA KIT

Human breast cancer susceptibility protein 2, BRCA-2 ELISA Kit

Uranium Testing and On Site Calibration for Water Testing Detection Range: 2 60ppb

Application Notes for COD Analysis DETERMINATION OF CHEMICAL OXYGEN DEMAND (COD) IN WATER AND WASTE WATER.

SEMI C SPECIFICATIONS AND GUIDELINES FOR AMMONIUM HYDROXIDE

Human Myostatin, ELISA Kit (MSTN)

MOUSE SOLUBLE FLT-1/VEGF-R1 ELISA KIT

Mouse Clusterin ELISA

Rev Experiment 10

IRON (Colorimetric) 2. Muffle Furnace: Equipped with pyrometer and capable of operating at controlled temperatures up to 600 C

SUMMARY OF STANDARDS FOR WATER ANALYSIS ON COURSE WAT-E2120 SPRING 2017

RAT PROPROTEIN CONVERTASE SUBTILISIN /KEXIN TYPE 9 (PCSK9) ELISA KIT

HUMAN SOLUBLE HEAT SHOCK PROTEIN 70 (HSP70) ELISA KIT

NACA Laboratory Equipment and Supplies

HUMAN NEUROTROPHIN 3 (NT3) ELISA KIT

RECOMMENDED SCOPE OF ACCREDITATION (For Testing Laboratories) * Test Method / Standard against which tests are performed

HUMAN ENDOTHELIAL- CELL SPECIFIC MOLECULE- 1 (ESM-1) ELISA KIT

HUMAN TOTAL MATRIX METALLOPROTEINASE 1 (MMP-1) ELISA KIT

EXPERIMENT 15C. Qualitative Analysis Scheme of Main Group and Transition Metal Cations without Hazardous Waste

Transcription:

, TNTplus856, 10220 DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and Application: For Water and Wastewater Test preparation Before starting the test: DR 2800: Install the light shield in Cell Compartment #2 before performing this test. Please read Safety Advice and Expiration Date on package. Recommended sample ph is 3 10. Recommended sample and reagent temperature is 15 25 C (59 77 F). concentrations greater than the measuring range may cause precipitation in the vial. In such cases, the water sample must first be diluted with deionized water. Undissolved nickel or nickel contained in complexes can only be determined after digestion with Metals Prep Set TNT 890. TNTplus methods are activated directly from the Main Menu when the sample vial is inserted into the sample cell holder. Collect the following items: Description Quantity Light Shield (DR 2800 only) 1 TNT856 Reagent Set 1 Pipettor, variable, 100 1000 µl 1 Pipettor tips, for 100 1000 µl pipettor 1 Pipettor, variable, 1 5 ml 1 Pipettor, tips for 1 5 ml pipettor 1 See Consumables and replacement items for reorder information. Page 1 of 6

Dimethylglyoxime method, TNTplus 1. Pipet 2.0 ml of sample into the vial. 2. Cap the vial and invert 2 3 times until the freeze-dried contents in the vial are completely dissolved. 3. Pipet 0.2 ml (200 µl) of Solution A into the vial. 4. Cap and invert the vial 2 3 times. 5. Wait three minutes. Install the Light Shield in Cell Compartment #2. 6. Thoroughly clean the outside of the vial. Insert the prepared vial into the cell holder. The instrument reads the barcode, then selects and performs the correct test. Results are in mg/l Ni No instrument zero is required. Reagent blanks A reagent blank can be measured and the value subtracted from the results of each test performed using the same reagent lot number. Use deionized water in place of sample and run the procedure as described. To subtract the value of the blank from a series of measurements, measure the blank per step 6. Press OPTIONS>MORE>REAGENT BLANK. Select ON. The measured value of the blank should be displayed in the highlighted box. Press OK to accept this value. The reagent blank value will now be subtracted from all results until the function is turned off, or a different method is selected. Alternately, the blank can be recorded and entered at any later time by pressing the highlighted box and using the keypad to enter the value. Page 2 of 6

Sample blanks Interferences Colored or turbid samples can cause high results. To compensate for color or turbidity, the procedure is repeated and the color forming reagent that is present in Solution A is not added. To determine the sample blank, run the procedure as given, but substitute 0.2 ml of deionized water in place of the 0.2 ml of Solution A in step 3. Use the red stopper to cap the sample vial. The value obtained in step 6 is then subtracted from the value obtained on the original sample to give the corrected sample concentration. Alternatively, samples that contain only turbidity may be first filtered through a membrane filter and then analyzed. Report the results as soluble nickel. Samples without color or turbidity do not require sample blanks. The ions listed in Interfering substances have been individually checked up to the given concentrations and do not cause interference. The cumulative effects and influence of other ions has not been determined. Undissolved nickel or nickel contained in complexes can only be determined after digestion with Metals Prep Set TNT 890. Table 1 Interfering substances Interfering substance K +, Na +, Cl, SO 2 4 NH + 4, NO 3, Ca 2+, PO 3 4, CO 2 3 Cr 6+, Zn 2+, F, NO 2 Al 3+, Cr 3+, Cd 2+, Co 2+, Sn 2+, Pb 2+ Fe 2+, Fe 3+, Mn 2+, Cu 2+, Mg 2+, Hg 2+ Ag + Interference level 1000 mg/l 500 mg/l 50 mg/l 10 mg/l 5 mg/l 1 mg/l Sample collection, preservation and storage Accuracy check Collect samples in acid-washed glasses or plastic containers. Adjust the ph to 2 or less with nitric acid (about 2 ml per liter). Store preserved samples up to six months at room temperature. Adjust the ph between 3 10 with 5.0 N sodium hydroxide before analysis. Correct the test result for volume additions. Standard solution method Note: Refer to the instrument user manual for specific software navigation instructions. Required for accuracy check: Standard Solution, 1000 mg/l Deionized water 250-mL volumetric flask 1-mL Volumetric pipet Page 3 of 6

Pipet filler 1. Prepare a 4.0 mg/l nickel standard solution as follows: a. Pipet 1.0 ml of Standard, 1000-mg/L, into a 250-mL volumetric flask. b. Dilute to the mark with deionized water. Mix well. Prepare this solution daily. 2. Use 2.0 ml of this solution in place of the sample. Follow the Dimethylglyoxime method, TNTplus test procedure. 3. To adjust the calibration curve using the reading obtained with the standard solution, navigate to Standard Adjust in the software. Instrument DR 5000 DR 2800 DR 2700 DR/2500 DR/2400 Navigate to: OPTIONS>MORE>STANDARD ADJUST OPTIONS>MORE>STANDARD ADJUST OPTIONS>MORE>STANDARD ADJUST OPTIONS>STANDARD ADJUST OPTIONS>STANDARD ADJUST Summary of method 4. Turn on the Standard Adjust feature and accept the displayed concentration. If an alternate concentration is used, enter the concentration and adjust the curve to that value. In the presence of an oxidizing agent, nickel ions react with dimethylglyoxime in an alkaline solution to form an orange-brown-colored complex. Test results are measured at 463 nm. Consumables and replacement items Required reagents Description Quantity/Test Unit Catalog number TNT 856 Reagent Set 1 25/pkg TNT856 Required apparatus Description Quantity/Test Unit Catalog number Light Shield (DR 2800 only) 1 each LZV646 Pipettor, variable volume, 1 5 ml 1 each 2795100 Pipettor Tips, for 27951-00 pipettor 1 100/pkg 2795200 Pipettor, variable volume, 100 1000 µl 1 each 2794900 Pipettor Tips, for 27949-00 pipettor 1 400/pkg 2795000 Recommended standards Description Unit Catalog number Standard Solution, 1000 mg/l 100 ml 1417642 Nitric Acid, ACS 500 ml 15249 Sodium Hydroxide Standard Solution, 5.0 N 100 ml MDB 245032 Page 4 of 6

Recommended standards Description Unit Catalog number Sodium Hydroxide Standard Solution, 5.0 N 50 ml SCDB 245026 Water, deionized 4 L 27256 Optional reagents and apparatus Description Unit Catalog number Bottle, sampling, low density poly, w/cap, 500 ml 12/pkg 2087079 Filter Holder, glass for vacuum filtration each 234000 Filter, membrane, 47 mm; 0.45 -micron, hydrophilic, polyethersulfone for SUVA each 2894700 Flask, filtering, glass, 1000-mL each 54653 Flask, volumetric 250 ml, Class A each 1457446 Pipet, volumetric 1.0 ml, Class A each 1451535 Metals Prep Set TNT 890 each TNT890 Test Tube Rack for 13-mm vials each 2497900 Tubing, rubber 12 ft 56019 Pipet Filler, Safety bulb each 1465100 Aspirator with check valve each 213100 Page 5 of 6

FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932 Hach Company, 2007. All rights reserved. Printed in the U.S.A. Updated February 2008, Edition 5