Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides

Similar documents
A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

Low-level Detection of Ethanol, 1,4-Dioxane, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator

Analysis of Organotin Compounds in Biological and Environmental Samples by Gas Chromatography and Pulsed Flame Photometric Detection (GC-PFPD)

Application Note. Abstract. Authors. Environmental

Pulsed Flame Photometric Detector (PFPD) for CP-3800

Large Volume Injection of Organochlorine Pesticide Extracts Using a Programmable Temperature Vaporizer in the Solvent Split Mode

Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D7504 on the Agilent 8890 GC System

Improving Productivity in Your Environmental and Food Testing Labs with Agilent 7890A/5975C GC/MS System

Olga Shimelis, Katherine Stenerson, Michael Ye, Carmen T. Santasania, and Ed Mauney. Supelco, Div. of Sigma-Aldrich, Bellefonte, PA USA

Improved Gas Chromatograph Method for the Analysis of Trace Hydrocarbon Impurities in 1, 3-Butadiene

Ghost Peaks in Gas Chromatography Part 2: The Injection Port

USEPA Methods and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors

USEPA Methods and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

Analysis of Semivolatiles by GC/FID using the Ultra Inert Inlet Liners with Wool

Agilent Consumables Improve Performance with Shimadzu GC/MS

Gas Chromatography Assignment Chem 543/443

Technical Article. Innovative Chromatography Solutions. The Rtx -CLPesticides Column Set Lets You Make the Choice

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run

Gas Chromatography SSI-GC-003

ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER

Direct Injection of Fish Oil for the GC-ECD Analysis of PCBs: Results Using a Deans Switch With Backflushing. Application. Author.

Petroleum Fuels. ogen process analyzers 6000 SERIES

Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES

Improving GC-MS Method Robustness and Cycle Times Using Capillary Flow Technology and Backflushing

JAI RESEARCH FOUNDATION A Global Contract Research Organisation

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

Improved Performance and Productivity of Purge-and-Trap Analyses

Contract Laboratory Program (CLP) Pesticide Analysis with 0.18 mm ID High-Efficiency GC Columns Utilizing Helium Carrier Gas. Application.

Prefractionator for Reliable Analysis of the Light Ends of Crude Oil and other Petroleum Fractions

System Overview. Valve oven. Restrictor. Valve. Sample in and out

Multiresidue Pesticide Analysis with the Agilent Intuvo 9000 GC and Agilent 7000 Series Mass Spectrometer

DETECTORS Dry Electrolytic Conductivity Detector - DELCD. Overview

Pesticide Analysis in a Complex Matrix using a 2D GC/MS System

Drinking Water Semivolatiles Analysis using the 6890N/5975B inert GC/MSD. Application. Author. Abstract. Instrument Operating Parameters.

Fluorescence Detectors for High Performance Liquid Chromatograph. RF-20A RF-20Axs C193-E026C

An Examination of the Presence, Formation, and Transformation of Volatile Halogenated Organic Species in Wastewater Extracts Using GC-ICP-MS

TRACE GCxGC. Comprehensive Two Dimensional GC. Unmatched separation capability for complex samples

Methods for Increased Sample Throughput with Dirty Samples

CryoTrap. For Agilent Technologies 7890/6890/6850 GC Systems. CryoTrap

Application Note: Multi-Residue Pesticides in Fruit and Vegetables (9/16/08)

Capillary GC Troubleshooting: a Practical Approach

Analysis of Nucleosides Using an Agilent Infinity II High Speed UHPLC with the 6130 Single Quadrupole Mass Selective Detector

Nitrosamines Analysis in Drinking Water Using GC/MS/MS Meeting Equivalence to EPA Method 521

Electron Capture Detector for GC-2010 Plus

Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS

Scale-up with the Agilent SD-1 Purification System analytical and preparative runs on a single system

Benefits of running organic matrices using the Agilent 5100 ICP-OES fast, robust, high performance analysis

STANDARD OPERATING PROCEDURES

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC

Evaluating Inert Flow Path Components and Entire Flow Path for GC/MS/MS Pesticide Analysis

Olefin Analyzer Solutions for Polymer Grade Ethylene and Propylene

Corn Syrup Analysis E-1a- 1. ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

A New Horizon of Performance Agilent G1888 Network Headspace Sampler

Validated Multi-Residue Method for Extraction and Analysis of Trace-Level Pesticides in Surface Water Application

TurboMatrix Headspace and Headspace Trap Samplers

TDLAS. Application Note. Tunable Diode Laser Absorption Spectroscopy (TDLAS) for Carbon Dioxide Measurements In Natural Gas Processing

Gas Chromatograph. GC-8A Series C184-E016B

Validation Data for the Analysis of Chlorotalonil in Leek Using Mini-Luke Method Followed by GC-QqQ-MS/MS

FOOD PESTICIDE RESIDUES LABORATORY REFERENCE NO INSPECTOR DATE

Technical Overview. Author. Abstract. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS

Historical Review: Fast & micro Gas Chromatography

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

Application Note: Overview. Introduction

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations

for water and beverage analysis

A Single Analyzer Solution for Refinery Flare Emissions Monitoring (40 CFR 60 Subpart Ja) Using Mass Spectrometry

Agilent 1290 Infinity Binary LC System with ISET Emulation of a Waters Alliance 2695 LC Applying Concave, Convex, and Linear Gradients

Separation and Quantitation of a Mixture of Alcohols by Gas Chromatography

Confirmation of THC in Oral Fluids Using High-Resolution 2-D GC/MS Application Note

DANI A SCENT OF FUTURE HIGH PRODUCTIVITY ACCURACY AND PRECISION FLEXIBILITY AND UPGRADEABILITY USER FRIENDLY INTERFACE FAST GAS C HROMATOGRAPH

Lab 7 Measurement of Ozone

Certified Reference Material - Certificate of Analysis

Analysis of steel and its alloys using the GB/T standard and an Agilent 5100 ICP-OES in dual view mode

Certified Reference Material - Certificate of Analysis

Automated Purification Software for Agilent OpenLAB CDS

Fast(er) GC: How to Decrease Analysis Time using Existing Instrumentation? Part I: Impact of column dimensions.

A Presentation by OI Analytical. Oxygenate Analysis on the Eclipse Purge-and-Trap Sample Concentrator

CX-1 ph Gradient Buffer

Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP1 and DB-CLP2

A Direct Column-Performance Comparison for Rapid Contract Laboratory Program (CLP) Pesticide Analysis. Application. Authors. Abstract.

Certified Reference Material - Certificate of Analysis

Lab 6 Measurement of Ozone

7890A An Essential Partner for the Trace

MARKET FOCUSED - environmental ORGANOCHLORINE PESTICIDES

Certified Reference Material - Certificate of Analysis

GDMS. High Resolution GDMS.

When optimization is discussed throughout this talk, it s in terms of performing more samples, in less time, while minimizing investment in capital

Energy Saving Capillary Gas Chromatograph

Turnkey configurations for gas chromatographic analysis, including instrumentation, methods, supplies and service. Analysis of Natural Gas

The Use Of Automated Backflush on the 7890A/5975A GC-MS System. Application. Setup Example Using Biomarkers. Authors. Abstract.

Meet Ja, H2S, and BTU flare regulations while reducing your total cost of ownership

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Automated alternating column regeneration on the Agilent 1290 Infinity LC

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

Introducing the Eclipse: A New Generation of Purge-and-Trap Sample Concentrators

Composition Analysis of Animal Feed by HR ICP-OES

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

Transcription:

Application Note 25390206 Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides Keywords Model 5380 Organophosphorus Pesticide PFPD PFPDView Phosphorus Pulsed Flame Photometric Detector Sulfur USEPA Method 1618 USEPA Method 8141B Introduction Organophosphorus (OP) pesticides are among the most widely applied commercial pesticides in the world. Their toxic pesticide residues are present in many agricultural products, often at concentrations in the low parts-per-billion (ppb) range. Food, water, plants, and soils are some of the familiar matrices analyzed for low-level OP pesticide residues. Two of the most common methods for analyzing OP pesticides include USEPA Method 8141B and Method 1618. Both methods specify using a flame photometric detector (FPD) operated in the phosphorus mode for selective detection of OP pesticides. The OI Analytical Model 5380 Pulsed Flame Photometric Detector (PFPD) shown in Figure 1 has emerged as the phosphorus-selective detector of choice for trace-level OP pesticide analysis. The PFPD uses a pulsed flame, instead of the continuous or static flame of the conventional FPD. The pulsed flame introduces a time-dependent variable to the analysis. As a result, the PFPD outperforms the conventional FPD in detector sensitivity, selectivity for phosphorus with respect to hydrocarbon, lower gas consumption, ease of operation, long-term stability, and dual-element detection capability. This application note discusses considerations for configuring the PFPD for low-level analysis of OP pesticide residues. Modifying gate settings to extend the calibration range, and using dual gate subtraction to eliminate any residual sulfur interference are discussed. PFPDView software is used as a post-acquisition data processing tool. Presented at the 2006 Pittsburgh Conference on Analytical Chemistry and Applied Spectroscopy, Orlando, FL March 12 17, 2006 Figure 1. OI Analytical Model 5380 Pulsed Flame Photometric Detector (PFPD)

Experimental The OP pesticides listed in Methods 8141B and 1618 contain phosphorus as the heteroatom, or in many cases, both phosphorus and sulfur. The PFPD can be easily configured for high selectivity of either element with a few simple configuration changes. But when configured for the phosphorus-only mode, the PFPD s sensitivity is approximately ten times lower than in the sulfur mode. The detector can also be configured to detect both phosphorus and sulfur simultaneously using dual-gate subtraction, which can produce mutually selective phosphorus and sulfur chromatograms. For these experiments, a standard test mixture was selected that included nine pesticides determined by one or both of the USEPA methods. Each OP pesticide contained at least one sulfur and one phosphorus atom so calibration curves in both configurations could be compared. All analyses were conducted using an Agilent 6890N Gas Chromatograph (GC) configured with a split/splitless injector and an OI Analytical Model 5380 PFPD. Agilent ChemStation was used to acquire and process the data. Table 1 shows instrument operating conditions. Table 1. Instrument operating conditions Parameter Specification GC Agilent 6890N Inlet Split/splitless inlet, splitless mode, 4-mm single gooseneck splitless liner, 220 C, 1-µL injections Column Rtx -5, 30 m x 0.32-mm I.D. x 0.5-µm film Carrier gas Helium, 1.1 ml/minute, constant flow mode Oven program 50 C (2 minutes), 8 C/minute to 300 C (5 minutes) Detector Model 5380 PFPD Board parameters PMT voltage = 550 Ignitor current = 2.8 Trigger level = 500 Range = 10 Phosphorus mode 3-mm combustor Yellow RG-495 optical filter H 2 /air combustion gas in approximately equal proportions Phosphorus gate at 4 10 msec (see discussion for other gates) Sulfur mode 3-mm combustor (using a 2-mm combustor improves sulfur sensitivity, but sacrifices phosphorus sensitivity) Blue BG-12 optical filter H 2 /air combustion gas, slightly H 2 -rich Sulfur gate at 6 25 msec, square root ON P+S dual element mode 3-mm combustor Clear UV-34 optical filter H 2 /air combustion gas same as phosphorus mode Dual-gate subtraction (see discussion) 2

Results and Discussion Phosphorus Mode Calibration standards were prepared containing all nine compounds at concentrations ranging from 1 to 2,000 ppb. The standards were analyzed using the PFPD configured in the phosphorus mode and a calibration curve was plotted. The chromatogram from analysis of the 100-ppb standard is shown in Figure 2. Table 2 lists the compound names with their chemical formulas, molecular weights, and weight percents of phosphorus and sulfur. 1 2 3 4 6 7 8 5 9 Figure 2. PFPD chromatogram of the 100-ppb calibration standard containing nine of the OP pesticides found in USEPA Methods 8141B and 1618. The PFPD was configured for phosphorus-selective detection with a 3-mm combustor, RG-495 optical filter, and 4 10 msec gate. See Table 2 for compound identification and percent phosphorus content. No. Name CAS No. Formula Table 2. OP pesticide compounds Mol. Weight Wt. % P Wt. % S 8141B Analyte 1618 Analyte 1 o,o,o- Trimethylphosphorothioate 152-20-5 C 3 H 9 O 3 PS 156.1 19.8 20.5 2 Thionazin 297-97-2 C 8 H 13 N 2 O 3 PS 248.2 12.5 12.9 3 Sulfotepp 3689-24-5 C 8 H 20 O 5 P 2 S 2 322.3 19.2 19.9 4 Phorate 298-02-2 C 7 H 17 O 2 PS 3 260.4 11.9 36.9 5 Dimethoate 60-51-5 C 5 H 12 NO 3 PS 2 229.3 13.5 27.9 6 Disulfoton 298-04-4 C 8 H 19 O 2 PS 3 274.4 11.3 35.1 7 Methyl parathion 298-00-0 C 8 H 10 NO 5 PS 263.2 11.8 12.2 8 Parathion 56-38-2 C 10 H 14 NO 5 PS 291.3 10.6 11.0 9 Famphur 52-85-7 C 10 H 16 NO 5 PS 2 325.3 9.5 19.7 The PFPD is a highly selective detector, and when configured for phosphorus detection, responds only to the phosphorus in the compound. It also has the added advantage of an equimolar response for phosphorus, so it produces the same total response for equivalent amounts of phosphorus regardless of the parent compound. This effect can be observed by comparing peak areas from the chromatogram in Figure 2 to the weight percent phosphorus for each compound. The calculated response factors for the phosphorus in each compound are shown in Table 3. The low 6.1% RSD over the nine response factors is a confirmation of the equimolarity of the phosphorus response. This conclusion is further confirmed by the chromatogram of the 1-ppb standard (Figure 3), which only shows peaks for the two compounds with the highest percent phosphorus content, o,o,o-trimethylphosphorothioate and sulfotepp. The phosphorus content for these two peaks was approximately 0.2 ppb (0.2 pg of phosphorus on the column). Note that peak heights cannot be used for this test as the peak width varies for 3

individual pesticides across the chromatogram. Table 3. PFPD area counts for the nine OP pesticides shown in Figure 2, with calculated response factors illustrating equimolar response of the PFPD. Each compound was present in the mixture at 100 ppb. No. Name Wt. % P Area Counts Response Factor 1 o,o,o-trimethylphosphorothioate 19.8 46,573 2,352 2 Thionazin 12.5 25,967 2,077 3 Sulfotepp 19.2 43,507 2,266 4 Phorate 11.9 25,798 2,168 5 Dimethoate 13.5 31,832 2,358 6 Disulfoton 11.3 24,336 2,154 7 Methyl parathion 11.8 25,318 2,146 8 Parathion 10.6 23,227 2,191 9 Famphur 9.5 18,385 1,935 Average Response Factor 2,183 RF % RSD 6.1 1 (0.2 pg P) 3 (0.2 pg P) Figure 3. PFPD chromatogram (phosphorus mode) of the 1-ppb calibration standard showing peaks for the two compounds with the highest percent phosphorus content. These two peaks each represent approximately 0.2 pg of phosphorus on the column. When plotting the PFPD responses for sulfotepp as a function of concentration over the entire calibration range (1 to 2,000 ppb), the calibration curve clearly shows that the linear range of the detector has been exceeded using these conditions (Figure 4). A close examination of the individual phosphorus emissions using PFPDView software makes the reason clear. Figure 5 is a screen from PFPDView showing the overlaid phosphorus emissions from sulfotepp at 50 and 2,000 ppb, using a gate setting of 4 10 msec. As long as the phosphorus emission within the defined gate remains within scale and does not saturate the window, the signal remains linear. In this example, when the gate was set at 4 10 msec, the emission from the 50-ppb sulfotepp peak remained within scale and within the linear range. Once the emission reached the top of the scale, the curve was no longer linear. For sulfotepp, the phosphorus emission begins to saturate the 4 10 msec gate at 100 ppb (approximately 19.2 ppb of phosphorus), producing a linear calibration curve up to 100 ppb with an R 2 value of 0.9971 (Figure 6). 4

300000 PFPD Response (Area Cnts) 250000 200000 150000 100000 50000 R 2 = 0.9809 0 0 500 1000 1500 2000 2500 Concentration (ppb) Figure 4. Calibration curve for sulfotepp over the full calibration range (1 to 2,000 ppb) using a 4 10 msec phosphorus gate 4 10 msec P gate Gate saturation at 2,000 ppb Sulfotepp 2,000 ppb P emission Sulfotep 50 ppb P emission Figure 5. PFPDView screen illustrating overlaid phosphorus emissions from 50 ppb and 2,000 ppb of sulfotepp with a gate setting of 4 10 msec. Note that the 2,000-ppb phosphorus emission is saturated within the 4 10 msec gate, leading to a nonlinear response. 5

50000 PFPD Response (Area Cnts) 45000 40000 35000 30000 25000 20000 15000 10000 R 2 = 0.9971 5000 0 0 20 40 60 80 100 120 Concentration (ppb) Figure 6. Linear calibration curve for sulfotepp from 1 to 100 ppb using a 4 10 msec phosphorus gate When necessary, the linear range of the PFPD can be easily extended by simply modifying the gate setting. For sulfotepp, when the gate was changed to 6 10 msec, the phosphorus emission no longer saturated the gate, even at 2,000 ppb. Using the shorter delayed gate setting, the calibration range for sulfotepp was extended to 2,000 ppb, and resulted in R 2 values of 0.9936 for a linear fit or 0.9997 when using a quadratic fit (Figure 7 and Figure 8). 6 10 msec P gate No gate saturation even at higher concentrations Sulfotepp 2,000 ppb P emission Sulfotep 50 ppb P emission Figure 7. PFPDView screen illustrating overlaid phosphorus emissions from 50 ppb and 2,000 ppb of sulfotepp with a gate setting of 6 10 msec. Note that the 6 10 msec gate is not saturated, even at 2,000 ppb, leading to a linear calibration up to 2,000 ppb. 6

180000 PFPD Response (Area Cnts) 160000 140000 120000 100000 80000 60000 40000 20000 R 2 = 0.9997 R 2 = 0.9936 0 0 500 1000 1500 2000 2500 Concentration (ppb) Figure 8. Calibration curve for sulfotepp over the full calibration range (1 to 2,000 ppb) using a 6 10 msec phosphorus gate. Either a linear (0.9936) or quadratic (0.9997) curve fit can be used for this calibration. The one drawback to using the delayed 6 10 msec gate can be a slight loss of sensitivity at the low end of the calibration curve. One technique for mitigating this loss is to configure two separate gates for phosphorus. Since the PFPD controller uses a digital processor, it is possible to output two simultaneous signals and two simultaneous chromatograms by simply defining two different phosphorus gates. For sulfotep, the first signal would be configured with a 4 10 msec gate for the low-level concentrations (up to about 100 ppb), and a second signal would be configured with the delayed 6 10 msec gate for quantitation at the high end of the concentration range. Using this technique, the calibration range can be extended by one full order of magnitude without sacrificing sensitivity at the low end. Sulfur Mode The PFPD was reconfigured for selective detection of sulfur by switching to the BG-12 optical filter, changing the gate to 6 25 msec to capture the sulfur emissions, and turning on the square root function to linearize the naturally quadratic sulfur signal. Results from sulfur selective analyses were similar to the phosphorus results; the primary difference was that the PFPD was approximately 10 times less sensitive in the sulfur mode. Figures 9, 10, and 11 illustrate the sulfur results. In PFPDView (Figure 10), the sulfur emission from phorate at 50 ppb is not saturated and is within the linear range of the 6 25 msec gate. The second emission, phorate at 500 ppb, is just below the saturation point and represents the top of the linear range. The sulfur emission of phorate at 2,000 ppb is completely saturated throughout most of the 6 25 msec gate and is beyond the linear range for the gate shown. In this example, the linear calibration range for phorate was 10 to 500 ppb (Figure 11). The 500-ppb calibration standard represented approximately 185 pg of sulfur on the column and was the saturation point for the sulfur emission. If desired, the sulfur calibration range can also be extended by defining a second sulfur gate for a portion of the sulfur emission that does not saturate at the higher concentrations. In Figure 10, the second gate at 3 5 msec does not saturate until approximately 1,500 pg sulfur on column, and would extend the linear calibration range by almost one order of magnitude. This is exactly analogous to the phosphorus example previously discussed. 7

1 (2 pg S) 3 (2 pg S) 4 (3.7 pg S) 6 (2.7 pg S) Figure 9. PFPD chromatogram (sulfur mode) of the 10-ppb calibration standard, showing peaks for four of the compounds with the highest sulfur content. The peaks shown here represent approximately 2 to 4 pg of sulfur on the column. See Table 2 for compound identification and sulfur content. Unsaturated emission Optional 3 5 msec S gate 6 25 msec S gate Phorate 2,000 ppb S emission fully saturated Phorate 500 ppb S emission just below saturation point Phorate 50 ppb S emission within linear range Figure 10. PFPDView screen illustrating overlaid sulfur emissions for phorate at 50 ppb, 500 ppb, and 2,000 ppb, and a gate setting of 6 25 msec. The optional second gate at 3 5 msec can be used to extend the linear calibration range by one full order of magnitude. 7000 PFPD Response (Area Cnts) 6000 5000 4000 3000 2000 1000 R 2 = 0.9996 0 0 100 200 300 400 500 600 Concentration (ppb) Figure 11. Linear calibration curve for phorate from 10 to 500 ppb using a 6 25 msec sulfur gate 8

PFPDView Software and Dual Gate Subtraction of Sulfur Interference Some sample types, particularly agricultural products like onions, garlic, and peppers, may contain very high concentrations of naturally occurring sulfur compounds in the sample extract that have the potential to interfere when the PFPD is configured for selective phosphorus detection. These sulfur compounds have weak spectral emission lines that fall within the transmission curve of the RG-495 optical filter used in phosphorus analyses, and are normally not observed under standard conditions. However, when the sulfur compounds are present at extremely high concentrations, as in the case of some food products, false positive peaks in the phosphorus chromatogram may result. PFPDView can be used to optimize PFPD parameters, effectively remove the sulfur interference from the chromatogram, and result in phosphorus-only detection. The individual emissions (pulses) from a chromatographic run are saved as a data file (*.dat) using WinPulse. These data files are then imported into PFPDView, which allows the operator to observe the individual emission profiles from any peak in the chromatogram, and identify the heteroatom as either sulfur or phosphorus. The operator can also use PFPDView to modify the PFPD gate parameters and reprocess the original data to determine the effect of new gate settings on the chromatogram without having to rerun the analysis. This is a powerful method development tool that predicts the outcome of PFPD acquisition variables on the chromatogram in a matter of minutes, instead of the hours it would take to run multiple combinations of variables. To illustrate PFPDView and dual-gate subtraction, the PFPD was configured in the phosphorus selective mode with an RG-495 filter, 3-mm combustor, and 4 10 msec phosphorus gate. A garlic extract with high sulfur content was spiked with four OP pesticides at varying concentrations and analyzed on the PFPD. The result of the original PFPD analysis is shown in the screen from PFPDView (Figure 12). The vertical black cursor in Figure 12A is positioned on the chromatographic peak at approximately pulse 201; the corresponding emission in the top half of the screen clearly identifies this as a sulfur-containing compound, a false positive in this phosphorus chromatogram. Moving the cursor to the peak at pulse 520 (Figure 12B) displays a waveform now dominated by a phosphorus emission and identifies this peak as one of the OP pesticides spiked into the matrix. Since the extract has been spiked with only four phosphorus containing compounds, most of the peaks in the chromatogram are clearly the result of sulfur interference. 9

A Sulfur emission Cursor B Phosphorus emission Cursor Figure 12. PFPDView screen illustrating the original PFPD phosphorus-mode chromatogram obtained from analysis of a garlic extract spiked with four OP pesticides. The peak at pulse 201 (Figure 12A) is identified as coming from a sulfur compound by examining the corresponding sulfur emission profile in the top half of the screen. The peak at pulse 520 (Figure 12B) is identified as coming from one of the spiked OP pesticides by the distinctly characteristic shape of the corresponding phosphorus emission. 10

For this example, the interfering sulfur response across the chromatogram was eliminated using the dual gate subtraction technique. Briefly, dual gate subtraction eliminates interference between two species by assigning reciprocal gate assignments (one for phosphorus and one for sulfur), and using a scaling factor, called the Alpha value, to subtract the overlapping portion of the sulfur emission from the phosphorus gate. (For a complete discussion of the dual gate subtraction technique, see OI Analytical Application Note 1127.) Using PFPDView, reciprocal phosphorus and sulfur gates were defined, and small adjustments were made to the Alpha value until the resulting chromatogram revealed no sulfur interference. Figure 13 shows the reciprocal phosphorus and sulfur gates and Alpha values used in this example. An Alpha value of 1.3 was found to eliminate the observed sulfur interference and result in a clean phosphorus chromatogram (Figure 14). This parameter is then entered into the Gate Parameters window in WinPulse and saved to memory for future analyses. Figure 13. Gate Parameters screen from PFPDView illustrating five different settings for the Alpha value. For this example, the Alpha value of 1.3 produced a phosphorus-only chromatogram with no sulfur interference. 11

Phosphorus emission from the peak at ~pulse 520 Cursor OP pesticides All sulfur responses have been eliminated Solvent front Figure 14. PFPDView screen showing the phosphorus-only chromatogram generated using dual gate subtraction with reciprocal phosphorus and sulfur gates and an Alpha value of 1.3. The insert is an expanded view of the baseline illustrating that all of the sulfur interference has been eliminated. 12

Phosphorus + Sulfur Dual Element Mode Since many of the OP pesticides also contain sulfur, the PFPD can be configured with a clear, UV-34 optical filter to collect simultaneous sulfur and phosphorus chromatograms. In this mode, pesticides that contain both sulfur and phosphorus present an emission profile similar to the one shown in Figure 15. 4 10 msec P gate Phosphorus emission Phosphorus and sulfur emissions overlap Sulfur emission Figure 15. PFPDView screen illustrating the emission profile from a mid-concentration pesticide that contains both phosphorus and sulfur. The PFPD was configured with a clear filter for simultaneous detection of both elements. Note the significant overlap between the sulfur and phosphorus emissions, and the lack of a pure phosphorus emission within the 4 10 msec phosphorus gate. As can be seen in Figure 15, a considerable overlap of the phosphorus and sulfur emissions appears between 6 and 12 msec. A sulfur-only chromatogram can be generated using a gate from 12 to 25 msec, where there is no interfering phosphorus waveform. Configuration of the proper phosphorus gate is slightly more complex, due to the presence of a portion of the sulfur emission within the 4 10 msec phosphorus gate. Dual-gate subtraction must be used to remove this sulfur contribution from the total emission and result in a pure phosphorus chromatogram. Since the phosphorus response is naturally linear and the sulfur response is naturally quadratic, the sulfur emission makes up a disproportionately larger portion of the emission at higher concentrations. But since the overall sulfur emission remains proportional at all concentrations, a single Alpha value can be defined that works throughout the entire phosphorus linear range. The iterative process of defining an Alpha value for simultaneous phosphorus and sulfur detection is exactly as described previously. When used during analysis of standards or samples, this Alpha value produces mutually selective phosphorus and sulfur chromatograms. In some cases, independent and simultaneous sulfur- and phosphorus-selective chromatograms are desired without employing dual-gate subtraction, such as in chemical warfare agent (CWA) monitoring or when only limited sample is available. The dual-headed PFPD (Figure 16) can be used in these situations. With the dual-headed PFPD, a single PFPD body houses a single combustor and a single combustion process. Two optical filter/pmt combinations are mounted on either side of the body, each configured for selective detection of a single element. Using this configuration, a single GC detector port can be used to collect simultaneous and mutually selective phosphorus and sulfur chromatograms from a single analysis without using dual-gate subtraction. 13

Figure 16. Dual-headed PFPD used for generating simultaneous and mutually selective phosphorus and sulfur chromatograms from a single analysis. This configuration is particularly useful for CWA monitoring or in cases where the sample size may be limited. Conclusions The OI Analytical Model 5380 PFPD can be configured to selectively detect and quantify organophosphorus pesticides at single-digit ppb concentrations. The linear calibration range can be extended to 2,000 ppb by using two phosphorus gates, and sulfur interference from complex food matrices can be eliminated using PFPDView software and dual gate subtraction. Simultaneous, mutually selective phosphorus and sulfur chromatograms can be acquired in a single analysis using the dual-headed PFPD. References 1. OI Analytical Application Note 1127, Using Dual Gate Subtraction to Enhance the Selectivity of a Pulsed Flame Photometric Detector (PFPD). 2. OI Analytical Application Note 1166, Model 5380 PFPD Photomultiplier Tube and Optical Filter Configurations. Dr. Aviv Amirav, Professor of Chemistry at the University of Tel Aviv, Israel, developed and patented the PFPD, and licensed it to O.I. Corporation. Agilent and ChemStation are registered trademarks of Agilent Technologies, Inc. Rtx is a registered trademark of Restek Corporation. P.O. Box 9010 College Station, TX 77842-9010 Tel: (979) 690-1711 FAX: (979) 690-0440 www.oico.com