Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds. Overview. Data Analysis Tutorial

Similar documents
User Bulletin. ABI PRISM 7000 Sequence Detection System DRAFT. SUBJECT: Software Upgrade from Version 1.0 to 1.1. Version 1.

Use the AccuSEQ Software v2.0 Mycoplasma SEQ module

PrimerArray Analysis Tool for Embryonic Stem Cells

PrimerArray Analysis Tool Ver. 2.2

Quantitative Real time PCR. Only for teaching purposes - not for reproduction or sale

HID Real-Time PCR Analysis Software v1.1

AccuSEQ Real-Time PCR Detection Software

QuantStudio 6 qpcr instrument (Applied Biosystems) quick start guide

Validation Using SDS Software Version on the Applied Biosystems 7500 Real-Time PCR System and the ABI PRISM 7000 Sequence Detection System

Applied Biosystems 7500 Fast Real-Time PCR System

ARM-D Kits Data Acquisition and Analysis Guide

TaqMan Residual DNA Detection Kit Chinese Hamster Ovary v2 Protocol

Module1TheBasicsofRealTimePCR Monday, March 19, 2007

Setting Up a Multiplex (two probe) Real Time 5 Nuclease Reaction on the ABI PRISM 7700 Sequence Detection System

ARM-D Kits Data Acquisition and Analysis Guide. Real-Time PCR Platform: Applied Biosystems QuantStudio 7 Flex Real-Time PCR System

Technical Review. Real time PCR

PowerSeq Quant MS System

Real-Time PCR Principles and Applications

Guidelines for Developing Robust and Reliable PCR Assays

Q: Using at least 3 biological replicates in an experiment is recommended to do. What do you suggest: At which step of calculation of the relative

ARM-D Kits Data Acquisition and Analysis Guide

Quantitative Real time PCR. Only for teaching purposes - not for reproduction or sale

High Resolution Melting Protocol

ReadMe Software v2.0.1 For Applied Biosystems 7500 and 7500 Fast Real-Time PCR Systems

User Bulletin. Veriti 96-Well Thermal Cycler AmpFlSTR Kit Validation. Overview

Georgia Bureau of Investigation - Division of Forensic Sciences DNA Procedures Manual DNA QUANTIFICATION USING QUANTIFILER HUMAN / Y HUMAN MALE KIT

Real-Time PCR Workshop Gene Expression. Applications Absolute and Relative Quantitation

Troubleshooting Chemistry and Assay Runs

Real-Time PCR ABI Prism 7500 (PSC 563, NSC 338) Real-Time PCR StepOne (PSC 563)

Setting up a Run of Real-time PCR cycler Rotor Gene 3000

Designing TaqMan MGB Probe and Primer Sets for Gene Expression Using Primer Express Software Version 2.0

For in vitro Veterinary Diagnostics only. Real-Time PCR Detection Kit for detection of Histomonas meleagridis.

HID Real-Time PCR Analysis Software v1.3

Principals of Real-Time PCR. Amira A. T. AL-Hosary Lecturer of Infectious Diseases, Faculty of Veterinary Medicine, Assiut University, Egypt

Applied Biosystems 7900HT Fast Real-Time PCR System. Performing Fast Gene Quantitation with 384-Well Plates DRAFT

Using AccuCaL-D and the RealCount software for robust quantitation of gene copy number

Troubleshooting of Real Time PCR Ameer Effat M. Elfarash

For in vitro Veterinary Diagnostics only. Kylt Brachyspira spp. Real-Time PCR Detection.

TaqMan Cytokine Gene Expression Plate I Protocol

Real Time Quantitative PCR Assay Validation, Optimization and Troubleshooting

Applied Biosystems TaqMan Array Microfluidic Cards

For in vitro Veterinary Diagnostics only. Detection Reagents specific for. Chlamydia psittaci Real-Time PCR.

Quantifiler Human DNA Quantification Kit Quantifiler Y Human Male DNA Quantification Kit

Clostridium difficile Toxin Genes A and B. Real-Time PCR Detection Kit for detection and differentiation of.

Roche Applied Science Technical Note No. LC 10/update 2003

Applied Biosystems Real-Time PCR Rapid Assay Development Guidelines

AccuSEQ v2.1 Real-Time PCR Detection and Analysis Software

For in vitro Veterinary Diagnostics only. Kylt E. coli F18, F41, Stx2e. Real-Time PCR Detection.

Perform an HRM Methylation Study

For in vitro Veterinary Diagnostics only. Kylt MHRS Triplex. Real-Time PCR Detection.

For feed and food applications only. Kylt Staphylococcus aureus. Real-Time PCR Detection Kit for detection of Staphylococcus aureus.

A Recommended Procedure for Real-Time Quantitative TaqMan PCR for Roundup Ready Canola RT73 Monsanto Biotechnology Regulatory Sciences

Fully Automated Library Quantification for Illumina Sequencing on the NGS STAR

For in vitro Veterinary Diagnostics only. Kylt Avian Nephritis Virus. Real-Time RT-PCR Detection.

For in vitro Veterinary Diagnostics only. Kylt Chicken Astrovirus. Real-Time RT-PCR Detection.

QUANTITATIVE RT-PCR PROTOCOL (SYBR Green I) (Last Revised: April, 2007)

Only for teaching purposes - not for reproduction or sale

For in vitro Veterinary Diagnostics only. Kylt MHRS. Real-Time PCR Detection kit for Mycoplasma hyorhinis and hyosynoviae.

For in vitro Veterinary Diagnostics only. Kylt Chlamydiaceae Screening. Real-Time PCR Detection.

For in vitro Veterinary Diagnostics only. Kylt Avian Hepatitis E. Real-Time RT-PCR Detection.

Quantitative Real Time PCR USING SYBR GREEN

PathoSEEK Analysis Quick Reference Tables:... 2 FLOWER... 2 NON FLOWER Matrices... 2 GUMMY... 3

Data normalization and standardization

Event-specific Method for the Quantification of Soybean SYHT0H2 by Real-time PCR. Validated Method

DNA Extraction and Real-Time PCR Detection Kit for Mycoplasma gallisepticum live vaccine strain ts-11.

User Bulletin #2. ABI PRISM 7700 Sequence Detection System. Introduction. Contents. December 11, 1997

For in vitro Veterinary Diagnostics only. Kylt Influenza A - H9. Real-Time RT-PCR Detection.

PathoSEEK Analysis Quick Reference Tables:... 2 FLOWER... 2 NON FLOWER Matrices... 2 GUMMY... 3

Software. Version Code TP700/TP760. Quick Reference TAKARA BIO INC.

Quantitation of mrna Using Real-Time Reverse Transcription PCR (RT-PCR)

Forensic News. FAS Corner April The Maximizing Data Quality Series Part 2: DNA Quantitation

TaqMan SNP Genotyping Protocol

Real-Time PCR Detection Reagents for Detection of Marek Disease Virus and differentiation between field strains and Rispens vaccine CVI988 strain

validation report for TaqMan Roundup Ready Quantification Kit

Executive Summary. clinical supply services

Event-specific Method for the Quantification of Cotton GHB119 Using Real-time PCR. Protocol

RT 2 Profiler PCR Array: Web-based data analysis tutorial

Event-specific Method for the Quantification of Cotton Line GHB614 Using Real-time PCR. Protocol

Validation Report. Elena Nardini Marco Mazzara Joachim Kreysa. Report EUR EN

Applied Biosystems 7500 Fast, 7500 and 7300 Real-Time PCR Systems

The use of these guideline is dedicated for Argene kits that mentioned the instrument ABI PRISM 7900 in their package insert.

Event-specific Method for the Quantification of Soybean DAS by Real-time PCR. Validated Method

Event-specific Method for the Quantification of Maize Line MON Using Real-time PCR. Validation Report

We measured telomere length in DNA isolated from leukocytes in peripheral blood, a measurement

Applications and Uses. (adapted from Roche RealTime PCR Application Manual)

Event COT102 Cotton. Real-time, Event-specific Polymerase Chain Reaction Method

Event-specific Method for the Quantification of Cotton Line LLCotton25 Using Real-time PCR. Protocol

Introducing a new generation of affordable choices from the leader in real-time PCR. Applied Biosystems 7500 Fast Real-Time PCR System

KAPA Library Quantification Kits For Applied Biosystems SOLiD platform

Evaluation of the PowerQuant System on the QuantStudio 5 Real-Time PCR System

Touchscreen (standalone capabilities, PIN-protected user accounts, and dye calibration/rnasep functionality)

TaqMan Protein Assays

Real-Time PCR. Real-Time PCR meets Automation qtower 3 auto

Event-specific Method for the Quantification of Soybean Line A Using Real-time PCR. Validation Report

A revolution in PCR analysis!

resdnaseq Quantitative DNA Kits

Joint Research Centre

PCR validation. August 2014 Willem van Leeuwen

Roche Molecular Biochemicals Technical Note No. LC 10/2000

Event-specific Method for the Quantification of Soybean CV127 Using Real-time PCR. Validation Report

Transcription:

Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds Overview In order for accuracy and precision to be optimal, the assay must be properly evaluated and a few adjustments need to be made. There are three important parameters to be assessed: Baseline Threshold Ct value Data Analysis Tutorial To accurately reflect the quantity of a particular target within a reaction, i.e. the amount of PCR product, it is critical that the point of measurement be accurately determined. Real-time analysis on the ABI PRISM 7700 Sequence Detection System involves three principle determinants for more accurate, reproducible data. Baseline Value During PCR, changing reaction conditions and environment can influence fluorescence. In general, the level of fluorescence in any one well corresponds to the amount of target present. Fluorescence levels may fluctuate due to changes in the reaction medium creating a background signal. The background signal is most evident during the initial cycles of PCR prior to significant accumulation of the target amplicon. During these early PCR cycles, the background signal in all wells is used to determine the baseline fluorescence across the entire reaction plate. The goal of data analysis is to determine when target amplification is sufficiently above the background signal, facilitating more accurate measurement of fluorescence. Threshold The threshold is the numerical value assigned for each run, which reflects a statistically significant point above the calculated baseline. Ct Value The Threshold Cycle (Ct) reflects the cycle number at which the fluorescence generated within a reaction crosses the threshold. The Ct value assigned to a particular well thus reflects the point during the reaction at which a sufficient number of amplicons have accumulated, in that well, to be at a statistically significant point above the baseline. 1

Baseline Setting The first step of data analysis is to accurately set the baseline. By default, the software sets the baseline from cycles 3-15. An example of this initial analysis is shown here: Figure 1: Default Baseline Default Threshold Baseline default is cycles 3 to 15 Threshold default is 10 standard deviations above mean fluorescence generated during baseline cycles To establish if the baseline needs to be adjusted, it is important to determine which reaction in the assay emerges earliest above the baseline. In order to determine this, double-click on the y-axis and change the plot to Linear View. Figure 2: Baseline Adjustment Initial Amplification Change plot to linear view for easiest viewing Verify that the reactions are flat at zero Determine which reaction emerges earliest above baseline If the reaction emerges after cycle 15, then no adjustments need to be made 2

If the first reaction emerges after cycle 15, then no adjustments to the baseline are necessary. Adjusting the Baseline A highly abundant target (i.e., 18s rrna) will often amplify very early in the PCR cycles. In this type of assay, it is necessary to adjust the stop value of the baseline setting. As shown in the example below, the amplification plot displays the characteristics of a highly abundant target, and reflects the need to adjust the baseline. Figure 3: Log View of Early Amplification To determine the appropriate stop value, double-click on the y-axis and change the view to linear. Figure 4: Linear View 3

For a quick check, adjust the Baseline stop value to six and hit the Update button. Next, determine the first cycle at which amplification is detected. Adjust the stop value one to two cycles before the earliest amplification. In the example below the best stop value was determined to be at cycle 8. Figure 5: Good Baseline Setting Assess earliest CT. Adjust stop value to one to two cycles before amplification. In this case the stop value was determined to be 8. Double-click on the y-axis to return to log view. Incorrect Baseline Settings The following examples depict what an amplification plot will look like if the baseline stop value is set too high or too low. 4

Baseline Stop Value is Too Low The settings in the example below were changed to cycles 3-8. Note that the plot begins to dip below zero around cycle 14. To fix a plot like this, extend the baseline stop back out to 15 and click on update calculations. Figure 6: Baseline Setting Too Low Initial baseline set too low Baseline Stop Value is Too High The settings in this plot were changed from 3-28. The fluorescence is not flat at zero; therefore the baseline stop is set to high. To correct it, lower the baseline stop point to a value of 15. View the plot in linear view to assist in determining the best stop point. Figure 7: Baseline Setting Too High 5

Threshold Value Once the baseline has been set correctly, the software will automatically set the threshold at 10 standard deviations above the mean baseline fluorescence. At this point, the threshold can be adjusted manually by placing the cursor on the threshold line and dragging the line to the point on the amplification plot that follows these parameters: 1. Linear phase of exponential amplification. The threshold should be placed in the region of exponential amplification across all of the amplification plots. This region is depicted in the log view of the amplification plots as the portion of the plot, which is linear. The threshold line should neither be placed in the plateau phase nor in the initial linear phase of amplification, which may be too low, and into the background fluorescence. 2. Precision maximization. The threshold should be set at a point that maximizes the precision of replicates. It is common to find that the precision amongst the replicates increases as the amplification progresses further into the exponential phase of the reaction. Therefore, to assure maximal precision, the threshold value must be set within the exponential range above any background noise within the assay. 3. Sensitivity maximization. The threshold should also be set to maximize the sensitivity of the assay. It is important that the threshold be placed at the point which best reflects all orders of magnitude in the assay(s) across the plate. The example below depicts the ideal location for setting the threshold. Note: There will usually be a "window" or range of values within which a threshold setting will fit the aforementioned criteria and yield optimum results. 6

Figure 8: Good Threshold Setting Plateau Range Exponential Phase Threshold Value Background Fluorescence The Exponential phase of amplification is the most accurate place to take a measurement of the amplicon. The Exponential phase of amplification appears as the middle third of the reaction in the logarithmic view. The default numerical value assigned for the threshold depicts a value equal to a certain multiplier (default of 10) above the standard deviation of the mean baseline fluorescence. The default Threshold value may not be the best setting for a particular assay. There are two ways of determining the appropriate threshold setting: the standard curve method and the replicate method. Standard Curve Method The figures below depict the default setting for the threshold; it is too low and close to the background noise. The default threshold is also in a place where the replicates are not as tight. This may result in impaired precision for quantitative values 7

Figure 9: Default Threshold Setting Threshold Bar Threshold Value Figure 10: Corresponding Standard Curve Values Discrimination between unknown replicates is poor Slope value is poor 8

Figure 11: Adjusted Threshold Threshold Bar Threshold adjusted higher in exponential phase and below plateau Threshold Value Moving the threshold higher in the exponential phase of amplification, as shown above, gave rise to more precise results. Figure 12: Adjusted Threshold Discrimination between replicates is good Slope value is improved Slope value is improved Correlation Coefficient has improved By adjusting the threshold, the Slope and Correlation Coefficient values have improved. 9

Replicate Method: When a standard curve has not been run, the replicate method can be used to properly set the threshold. In the figure below replicates of sample sets are being compared. Once again the default threshold is too low; The Ct s of the replicates are scattered a bit and the background is coming above the threshold. Figure 13: Thresholds for Replicates Background Default Threshold is near background Replicates are scattered and not very tight. The difference in Ct values are greater than 0.5 When the X-axis is expanded, it is easier to see just how spread apart the replicates are. The threshold can be easily set when viewed this way. Figure 14: Expanded X-axis Replicates are spread apart at threshold Expanded X-axis 10

The threshold can be adjusted to improve the precision of the replicates. After adjustment, the Ct values of the replicates are much tighter. Figure 15: Adjusted Threshold Threshold is set properly Replicates are tight, within 0.5 Cts Data Analysis Summary Step 1: Determine the correct Baseline Setting Step 2: Set the proper Threshold In most cases for a given assay (i.e. same primer and probe set) use the same threshold and baseline settings. Using the same settings from assay to assay will improve the precision and make the data more comparable from plate to plate and instrument to instrument. Setting appropriate Baselines and Thresholds will help to assess the quality of an assay If threshold and baseline settings are not helping with precision and accuracy, the assay may not have been properly optimized Note: The threshold and baseline settings will not be saved. It is best to record the baseline and the numerical threshold and manually set them the next time the run is opened or another assay is performed. Data Capture After a run is complete, the data can be manipulated for presentation purposes. After the appropriate settings have been determined, the Experimental Report, Amplification Plot, Standard Curve and other views can be printed out and filed. 11

The data can also be exported to Microsoft Excel or another spreadsheet program. When the Sequence Detection file is opened and all settings have been adjusted, go to File > Export > Experimental Report or Results. This will create a text file, which can be opened in programs like Microsoft Excel and manipulated. There are several tutorials on-line to help with some manipulations of data in an Excel like program. Summary: The SDS software and instrumentation collects fluorescent data as amplicon accumulates. The data collected is displayed in an Amplification Plot that can be viewed in two ways: Linear and Logarithmic. The Threshold Cycle (Ct) is determined when the level of fluorescence from accumulating amplicon crosses the Threshold. Ct vs. Well position is a good way of determining outliers in your reaction. Standard curves can be used in the reaction to determine unknown quantities. The software displays the standard curve as well as the unknowns in the Standard Curve view. The quantities of the unknowns are determined automatically and displayed in the Experimental Report. Evaluating the Raw Data and Multicomponent View are ways to monitor the fluorescence in the reaction as the reaction progresses. Setting Baselines and Thresholds properly will increase the precision of a given assay. 2002. Applied Biosystems. All rights reserved. For Research Use Only. Not for use in diagnostic procedures. Information in this document is subject to change without notice. Applied Biosystems assumes no responsibility for any errors that may appear in this document. This document is believed to be complete and accurate at the time of publication. In no event shall Applied Biosystems be liable for incidental, special, multiple, or consequential damages in connection with or arising from the use of this document. Applied Biosystems and ABI PRISM are registered trademarks of Applera Corporation or its subsidiaries in the US and certain other countries. Macintosh is a registered trademark of Apple Computer, Inc. All other trademarks are the sole property of their respective owners. 117GU05-01 Part Number 4370923 Revision A 12