Bioinstrumentation Light Sources Lasers or LEDs?

Similar documents
LASERS, LEDS AND OTHER LIGHTING SOURCES FOR LIFE SCIENCE APPLICATIONS. Wallace Latimer Coherent

Considerations Dynamic Range Detection Linearity Components Software Comparing qpcr Instrumentation Sensitivity Specificity of Detection sigma.

BASICS OF FLOW CYTOMETRY

Compensation: Fundamental Principles

Design for Manufacturability (DFM) in the Life Sciences

Each question may have MULTIPLE correct answers. Select all that are correct.

Sapphire. Biomolecular Imager THE NEXT GENERATION OF LASER-BASED IMAGING

The Measure of Performance Key efficiency metrics in assessing LED lighting technology. July 30 th, 2012

Proceedings of Meetings on Acoustics

UV-LED Curing for Industrial Printing


Evaluating Microplate Detection Instruments

Simultaneous multi-color, multiphoton fluorophore excitation using dual-color fiber lasers

Bringing Raman Spectroscopy to the Field

Short Length High Gain ASE Fiber Laser at 1.54µm by High Co-doped Erbium and Ytterbium Phosphate Laser Glasses

MoFlo Astrios High-Speed Cell Sorter

Sapphire SF: Low-Noise Ultra-Narrow Bandwidth Performance Delivers Superior Raman Spectroscopy Results

Laser longevity. by Glenn Binder, founder and president, Laspar LLC. Getting the most out of a laser investment requires knowing all the variables

Confocal Microscopy of Electronic Devices. James Saczuk. Consumer Optical Electronics EE594 02/22/2000

Absorption of an electromagnetic wave

Cavity Filters. KIGRE, INC., 100 Marshland Road, Hilton Head, SC 29926, USA PH: FAX: Web:

ANAT 3231 Cell Biology Lab12 Stem Cell Analysis

Biophotonics. Light Matter Interactions & Lasers. NPTEL Biophotonics 1

Quantum Dot applications in Fluorescence Imaging for Calibration and Molecular Imaging

MF-ChemiBIS. Today s most comprehensive solution for your bio-imaging needs and applications. Documenting Nature

Lasers service a diverse range of biomedical

792 : January Steven R. Bowman Brandon Shaw Barry J. Feldman John A. Moon Barry B. Harbison Ishwar D. Aggarwal Joseph Ganem NOTICE

Cell analysis and bioimaging technology illustrated

Lesson 1 Good Diffraction Data

Sapphire. Biomolecular Imager THE NEXT GENERATION OF LASER-BASED IMAGING

A Survey of Laser Types. Gas Lasers

Principles of flow cytometry: overview of flow cytometry and its uses for cell analysis and sorting. Shoreline Community College BIOL 288

FLUORESCENCE. Matyas Molnar and Dirk Pacholsky

Confocal Microscopes. Evolution of Imaging

Measurement data at orders of magnitude lower cost than other techniques BENEFITS Simple UVC source ideal for compact sensors

PALM/STORM, BALM, STED

Measurement data at orders of magnitude lower cost than other techniques BENEFITS Simple UVC source ideal for compact sensors

Biophotonics?? Biophotonics. technology in biomedical engineering. Advantages of the lightwave

EDUCATION EXPERIMENT. Fluorescence of an Unknown Fluorophore. Introduction.

Increasing Distribution Center Title Productivity Using Automated Put Walls

Special Techniques 1. Mark Scott FILM Facility

A Thin Layer Imaging with the Total Internal Reflection Fluorescence Microscopy

by Ken Tsukii *, Akihiro Murakami *, Toru Takahashi * and Jie Xu *

DETECTION OF LASER ULTRASONIC SURFACE DISPLACEMENT BY WIDE APERTURE FIBER OPTIC AMPLIFIER M.L. Rizzi and F. Corbani CESI, Milano, Italy

Azure cseries. A new way to see the light. c600 c500 c400 c300

NONTRADITIONAL MANUFACTURING PROCESSES

In situ semi-quantitative assessment of single cell viability by resonance

Hard Coated Silica/Silica (Low OH) Radius

SpectraMax Multi-Mode Microplate Readers

DNA Microarray Technology

Basic Principles in Flow Cytometry. Flow Cytometry

Newport Corporation. 10 th Annual Needham & Company Growth Conference January 8, 2008

Miniature fibre optic probe for minimally invasive photoacoustic sensing

SURFACE ENHANCED RAMAN SCATTERING NANOPARTICLES AS AN ALTERNATIVE TO FLUORESCENT PROBES AN EVALUATION

Azure cseries. A new way to see the light. c600 c500 c400 c300 c200

Fluorescence & UV- Visible Workshop

Theremino DNA Meter. Fluorometer for DNA Measurements. Ana Rodriguez (MUSE) - Lodovico Lappetito (Theremino)

OCTOPLUS QPLEX FLUORESCENCE IMAGER. for fast & powerful fast 2D Gel image acquisition

Innovations in Molecular Lab Operation. Joseph M. Campos, PhD, D(ABMM), F(AAM)

Spectroscopy with UVC LEDs in Environmental Monitoring & Life Sciences

Introduction. (b) (a)

OPTAN BALL LENS. Data Sheet Updated December, 2018 BENEFITS

LED GUIDE. The LED store that saves you more

Concept review: Fluorescence

Fundamentals of X-ray diffraction and scattering

Automation In Life Sciences and Diagnostics Applications

Photonic integrated circuits in biochemical food analysis

LOW TEMPERATURE PHOTONIC SINTERING FOR PRINTED ELECTRONICS. Dr. Saad Ahmed XENON Corporation November 19, 2015

Thermo Scientific GENESYS 10S Bio spectrophotometer

Color-Rich Fluoro-Max Dyed Microparticles March 2008

Physical structure of matter. Monochromatization of molybdenum X-rays X-ray Physics. What you need:

Intracellular Delivery Using Nanosecond-Laser Excitation of Large-Area Plasmonic Substrates

Supporting Information for. Electrical control of Förster energy transfer.

Bridgelux Gen. 7 V6 HD LED Array. Product Data Sheet DS401

Introduction to Energy Efficient Lighting ECEN 2060

Experiment title: Picosecond Crystallography I-III

1. INTRODUCTION 2. EXPERIMENTAL 3. REFERENCES

SBDUV-APD Solar-Blind Deep UV APDs

Attune TM Acoustic Focusing Cytometer Training. Manik Punj Attune Training

Plastic Coated Silica/Silica (Low OH) FIBER CROSS SECTION Polyimide and Acrylate Coated. Nylon and Tefzel Coated

For outstanding performance in real-time PCR

Skills and excellence formation on basis of Laboratory of Plasma Physics & Atomic Spectroscopy Institute of Spectroscopy (ISAN) of Russian Academy of

Real-Time PCR Principles and Applications

Laser Produced Plasma for Production EUV Lithography

Pacific Biosciences Technology Backgrounder 1 11/24/2008

Wavelength LDH - P / D - _ / C / F / FA / TA - N - XXX - _ / B / M / L / XL. Narrow linewidth (on request) Tappered amplified

SICCE LED Lighting Systems

Analysis of strongly absorbing chromophores by UV-visible ATR spectroscopy

Fiber lasers. 1 Introduction. 2 Optical bers. Rui Miguel Cordeiro (F141104) April 13, 2015

Diode Pumped Miniature Eye-Safe Laser Q-Switched by Passive Materials

Features. Benefits. Min Typical Max Min Max. OPTAN-250H-BL 245 nm 250 nm 255 nm 0.5 mw 1.0 mw. OPTAN-255H-BL 250 nm 255 nm 260 nm 0.5 mw 1.

Characterization of laser-material interaction during laser cladding process P.-A. Vetter,* J. Fontaine,* T. Engel," L. Lagrange,& T.

Interferometric optical biosensor. Xingwei Wang

SLS-process monitoring and temperature control

AMREP FLOW CYTOMETRY CORE FACILITY. Flow Cytometry Data Analysis Workshop Friday 24 th October

Introduction to Flow Cytometry:

NEWTON 7.0 BIOLUMINESCENCE & FLUORESCENCE IMAGING IN VIVO - IN VITRO IMAGING

Scintillating Optical Fibers

Technical Proposal. A UV Laser Setup for Thinning Radiator Diamonds for GlueX

Bridgelux Vesta Series Dim-To-Warm 13mm and 15mm Arrays. Product Data Sheet DS151

Transcription:

Bioinstrumentation Light Sources Lasers or LEDs? A comprehensive analysis of all the factors involved in designing and building life sciences instrumentation reveals that lasers provide superior performance compared to LEDs, and an overall cost advantage. Summary: Higher Performance and Lower Cost Biotech instrumentation manufacturers are currently developing products that feature increased miniaturization and lower costs per use to support worldwide trends in healthcare. For instruments based on fluorescence, this has opened a debate as to whether the laser or LED represents the best choice in light source. While the superior performance of lasers is well-known, it may not be obvious that lasers nearly always also represent a lower cost option. True Cost Many instruments that analyze biochemical samples and reactions utilize fluorescent probes. Examples include cytometry, genetic sequencing, hematology, polymerase chain reaction (PCR), high-throughput drug screening, and microarray scanners. The key to successful operation in all these instruments is the ability to maximize signal-to-noise ratio. This is accomplished by delivering the necessary amount of useable excitation light (light that actually excites the fluorescent probe) into the required illumination sample region, while minimizing the amount of light that ends up focused or scattered outside of this region. For current commercial instruments, these focused spot sizes can range from tens of microns in diameter, to several millimeters in size. Application Gene Sequencing/ Medical Diagnostics Focused Spot Size 1 to 10 x 1 to 10mm Medical Diagnostics 100 s x 100 s µm Cytometry 30 x 30 µm In general, the optical characteristics of the laser make it a much more efficient source for accomplishing this task than the LED, especially at smaller focused spot sizes. Although the LED source itself might be much less costly than a laser, all the ancillary measures that must be employed to correct for its inherent output deficiencies can end up making the total system cost greater than for a laser-based instrument. The following sections detail why this is the case. Output Power vs. Rated Power One important difference between lasers and LEDs is not in how they operate, but in how their output is traditionally specified. Lasers, originally developed by the photonics community, have always been specified by their output power. Simply stated, a 1 Watt laser delivers 1 Watt of light output. In contrast, LEDs are specified by their power consumption. Since even the best LEDs rarely surpass 10% efficiency, a 1 Watt LED usually outputs 100 milliwatts or less of light. Moreover, the maximum current and voltage specifications are often specified for pulsed operation. Thus, a 1 Watt LED might only be continuously operable at 750 milliwatts or less, which translates into an optical output of only 75 milliwatts! Light Throughput: Advantages of a True Point Source The most significant inherent difference between lasers and LEDs is in the area of light collection and focusing efficiency. The output of a laser is a well-behaved, narrow, low-divergence beam. Such a beam behaves like an ideal, true point source, meaning that virtually all the light can be collected and focused into the precise spot size needed by the application. Similarly, a laser beam can be very efficiently coupled into a single-mode fiber (whose output can be focused to a small, well behaved spot). On the other hand, a LED is a conventional, extended light source, like a light bulb. The ability to focus an extended source is defined by the size of the emitting surface area and the solid angle into which it radiates in a property called etendue. The important thing about etendue is that, in the best case scenario, it is www.coherent.com I LAS.sales@coherent.com I (800) 343-4912 I (408) 764-4042 1

preserved across an optical system (and in a real world optical system, it always gets a bit worse). For a perfect, Lambertian source: Etendue = π x emitting area x (NA)² Where the numerical aperture (NA) is the smaller of either the source s emission angle (i.e., source NA) or acceptance angle of the collecting optics (i.e., lens NA). Since real LEDs are never perfect Lambertian sources, this equation represents a theoretical best case. To see the impact of etendue, consider the case of a 100 milliwatt LED that has an active area of 1 mm² and which emits into a hemisphere (a solid angle of 2π steradians corresponding to a NA=1). If we wish to focus this output into a spot diameter of 100 µm. The etendue equation tells us: Collecting etendue = Focusing etendue π 1 mm² Collecting NA² = π (.1)² mm² Focusing NA² Collecting NA =.1 Focusing NA The highest theoretically possible NA on the focusing side is a value of 1. Therefore, the theoretical maximum for this scenario is a collecting NA of 0.1. This means that about 6% (6 milliwatts) of the LED output is focused into the desired 100 µm spot (using NA n sin θ). Still, the NA=1 value for the focusing optics is not practically realizable. Even a value of NA=0.3 would represent a fairly sophisticated and costly optical system. In this case, less than about 2% (2 milliwatts) of the LED output is contained in the 100 µm focused spot. Due to this phenomenon, it s usually not practical to concentrate even 1% of the output of a typical LED into the focused spot in an instrument. The etendue tradeoff for extended sources, such as LEDs, means that collecting more light means a bigger focused spot, or substantially more complex and costly focusing optics, or both. This problem can t be avoided by combining multiple LEDs; that increases source size even further, making the problem even worse. What would help is brighter LEDs which produce greater light output without an increase in emitter size. At present, typical LED brightness is increasing by only 10% per year in response to the marketplace. In addition, since most LEDs are used in illumination or display applications that are orders of magnitude greater than the bioinstrumentation market, biotech has no leverage whatsoever to change this situation. Figure 1 Etendue dictates that if a source is imaged at a magnification of 0.1X, then the collection NA must be 0.1 times the focusing NA. Since the highest possible focusing NA is 1.0, this severely limits the ability to collect all the output from the source LED. www.coherent.com I LAS.sales@coherent.com I (800) 343-4912 I (408) 764-4042 2

Monochromatic Laser vs. Broadband LED Another major difference between lasers and LEDs is that the laser is a monochromatic source (it only emits at one wavelength), whereas a LED emits over a much wider spectral range. Why does this matter? Whether it s a simple dye or a genetically modified protein, a fluorophore used in life sciences has a broad absorption (i.e., excitation) spectrum. When irradiated with light anywhere in this spectrum, part of the light will be absorbed and a very small part of this absorbed light is re-emitted as fluorescence at a longer wavelength. With a laser, all that is required is that the laser wavelength is chosen to be somewhere near the excitation maximum. As indicated in the figure, a longpass filter between the interaction zone and the photodetector effectively blocks all scattered laser light and ensures that the photodetector only sees the fluorescence, despite the fact that it is far dimmer than the laser excitation. Thus, with a laser, it is very easy to get a high signal-to-noise ratio in the detection system. With a LED, the output has a broad spectrum and the shape and center of this spectrum can shift with changes in operating conditions, e.g., operating temperature, operating current, pulsed operation versus continuous wave (CW) operation. This causes several problems. First, the long wavelength part of the LED emission overlaps with the fluorescence emission spectrum. And, the intensity of this scattered light can exceed the intensity of fluorescence, which can easily be as weak as 1/1000 of the absorbed light or less. To minimize this problem, a shortpass filter must also be included between the LED and the interaction zone to block the long wavelength tail of the LED output its bad photons. The end result is an increase in system cost, due to the extra optic, and a reduction in signal-tonoise ratio because some of the LED s output is blocked by this optic. The situation is even worse when multiple fluorophores are being interrogated, as in most flow cytometers. With lasers, the instrument only needs to incorporate laser wavelengths that are close to the excitation peaks of the flourophores. The signals from the different fluorophores can be readily distinguished by ratioing the signal from multiple detectors, each with a different longpass filter and, if necessary, by pulsing of the lasers and correlation of the signal timing. With LEDs the situation is more complex and messy. Extra frontend filters have to be carefully matched to both the excitation/emission spectra of the fluorophores and the output spectrum of the LED(s), lowering optical efficiency still further. In addition to this trade-off between signal and cross-talk, any variations in the LED emission spectrum generally requires changes in filter specifications. These variations are beyond the control of the instrument manufacturer and even their LED vendor. Figure 2 Every fluorophore is characterized by an excitation spectrum and a long wavelength emission spectrum. With a single wavelength laser, it is easy to separate scattered excitation light from the emission spectrum using a long-wave cutoff filter. But with a typical LED, the long wavelength tail of its output overlaps the fluorescence emission and must be somehow eliminated. Plus the size and shape of this tail can vary significantly between different batches of LEDs. www.coherent.com I LAS.sales@coherent.com I (800) 343-4912 I (408) 764-4042 3

Impact on Detector Costs Using a LED can also increase the cost and complexity of the photodetection system used to sense the fluorescence and/or scatter, particularly in applications with a small illuminated interaction zone. Specifically, the low brightness of LEDs means that the signal will be correspondingly lower, compared to the use of a laser. Achieving target speed and/or signal-to-noise performance may therefore dictate using a photodetector with higher gain and/or lower noise. For example, a particular instrument may only require a zero-gain photodiode or photodiode array when based on laser excitation, but may need a photomultiplier tube or a CCD array when used with a LED source. For instances where a laser-based instrument uses a room temperature CCD array, a LED version of the same instrument might need an actively cooled CCD array. Unit-to-Unit Variability: Supply Chain Realities Today s lasers are characterized by incredible unit-tounit consistency. All solid-state laser construction and the implementation of automated, robotic assembly and testing methods have largely eliminated individual variations. And, just as important, laser manufacturers like Coherent recognize bio-instrumentation as an important market for our products. For example, lasers are optimized for specific biotech applications and market segments (e.g., research, clinical, point of care). Additionally, specific laser models have a long (multiyear) life cycle with a minimum seven-year life cycle on spare parts and replacement units. This continuity is critical for OEM instrument manufacturers; it allows them to confidently design new instruments and upgrades with no fear of component obsolescence and enables them to service instruments in the field for many years after initial purchase. In contrast, LEDs are produced by the billions to support high volume application such as displays and lighting fixtures. They are designed specifically to service these high volume applications, which typically have very different performance requirements than a precision biotech instrument. Indeed, the entire bioinstrumentation market is an irrelevantly small size in the eyes of these LED manufacturers. They do not even sell their products at these small (thousands) volumes. Instead, instrument builders have to purchase any LEDs through third party re-sellers, who have no control over product development and product continuity. Their only control over product quality is by occasionally offering testing and pre-selection at a price premium, of course. Thus, while two LEDs might have the same nominal center wavelength, this can vary by several nanometers, and the shape of the entire output spectrum can be very different. Switching between two LEDs that differ in this way often necessitates installing one or more new filters and complete re-calibration. LEDs are manufactured in large batches called bins and are assigned a bin number. The only way to know that two LEDs are going to be the same is to buy them from the same bin. If a LED-based instrument design is chosen and ready to go into production, the instrument maker needs to advance purchase a lifetime supply of thousands of identical LEDs, i.e., an entire reel of LEDs from the same bin. Lasers: Scalable Performance In several applications (e.g., cytometry), instrument makers need to supply instruments to quite a number of different market sectors, such as advanced research, pre-clinical uses, clinical labs, and even point of care tools. These different sectors can have diverse needs from their new instrument, covering factors such as sensitivity, throughput and cost per measurement. This puts different demands on the characteristics and acceptable cost of the light source. With over 50 years supporting life sciences, laser manufacturers like Coherent understand this challenge and provide a very wide range of laser technologies and products to optimally match the needs of all the different instrument types, from the highest performance research instrument to the lowest cost point of care analyzer. Contrarily, with LEDs there is virtually no range of optical performance to choose from, other than size (power) and center wavelength. www.coherent.com I LAS.sales@coherent.com I (800) 343-4912 I (408) 764-4042 4

BioRay In many bioinstrumentation applications, there is a shift from the research laboratory to the clinical laboratory, and, subsequently, to the point of care. This trend manifests itself in greater streamlining, miniaturization and automation. To support this trend, Coherent has recently introduced the BioRay series of lasers that emphasize economy, compact packaging, and simple integration over bleeding-edge performance. These lasers deliver up to 50 milliwatts, and are available at several visible wavelengths (e.g., 405 nm, 450 nm, 488 nm, 520 nm, and 640 nm) that match the optimum excitation of common fluorescent probes and genetically encoded markers. These products are based on laser diode technology, as that is the simplest and lowest cost method of generating CW laser output at these wavelengths and in this power range. Laser diodes also offer the highest efficiency, lowering the required power budget for the final instrument. Since edge emitting laser diodes emit a highly divergent and asymmetric (elliptical) beam, optics are used in the laser head to produce a collimated, elliptical beam. To reduce the complexity of downstream beam delivery optics, each head has an adjustable output lens in order to enable smooth adjustment of the beam waist location. Conclusion There s certainly no question that it s cheaper to buy a LED than a laser. However, when the total cost of design, the complete bill of materials (BOM), the service strategy, and the true cost of ownership are considered, lasers are often a more cost effective option than LEDs. This is even true for next generation, point of care instrumentation, where component costs are paramount. Figure 3 A new generation of economical CW visible lasers called the BioRay series has been developed with output parameters tailored specifically to meet current and future trends in costsensitive biotech instrumentation. www.coherent.com I LAS.sales@coherent.com I (800) 343-4912 I (408) 764-4042 5