Best Practices for Reprocessing Robotic Instruments

Similar documents
Multi-Dimensional Study of Cleaning Efficacy in a Hospital CSSD

Getinge da Vinci Solution

New Requirements of ISO 17664/AAMI ST 81

Monitoring Flexible Endoscope Channels to Assure Cleaning

Validation of cleaning and sterilization of reusable medical devices

applying a unique combination of disciplines consulting, preclinical, toxicology,

Lotus. Reprocessing instructions for reusable instrumentation

Lotus. Reprocessing instructions for reusable instrumentation. French/Swiss protocol

GENERAL SURGERY DEVICES

How is Sterilization Validation Performed by the Medical Device Manufacturer?

DISCOVER INNOVATION DRIVING INCREASED EFFICIENCY

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces

One such problem is blood residual inside the channels of a robotic arm.

BIOTECH-GERMANDE ENDOSCOPE REPROCESSING WITH BRUSHLESS CLEANING PROCESS CONDITIONS. Lionel PINEAU PhD

Flexible endoscopy in a robotic world Challenges? HONG KONG HOSPITAL AUTHORITY - ADVANCED COURSE FOR INFECTION CONTROL NOVEMBER 2017

Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy!

The Significance of Validation to US CSSD Managers

Manner of Assessment Code (check all that apply) & Surveyor Notes. Yes N/A. Yes. Yes N/A N/A. Yes N/A. N/A Yes N/A

Guide for Cleaning, Sterilization and Storage of Reusable Introducers

Loaner Program Guidance

Endoscope Washer Disinfector Cleaning Process Challenge Devices Research & Validation

The Reusable Surgical Instrument Cycle

pump pressure conveying quantity chamber geometry flow conditions Washing arms trays water quantity dosing temperature foam performance

Loaner Programme Guidance

Loaner Programme Guidance

Surgical Instruments - Instructions for use

2016. All rights reserved.

Overview: High Level Disinfection

Instruction Guide Re-Usable Surgical Instruments. Instruction Guide Re-usable surgical instruments (GA 30.1)

The Power of Prolystica Cleaning Chemistries SO MUCH MORE THAN. One Integrated Approach to Healthcare.

CARE SERVICE USER GUIDE. Sonicision System

Competency Guide: Care and Handling of Rigid Endoscopes

In use test of a Washer-disinfector reprocessing of flexible endoscopes

Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants

VALIDATION OF WASHER-DISINFECTOR

CLEANING AND HANDLING OF MICROPORT INSTRUMENTS The following languages are included in this packet:

Facility: Audit Date Auditor Name/s

Instructions for use. Reusable surgical instruments. Hersteller:

Professional instrument reprocessing in CSSD

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation

Ethicon, Inc All rights reserved. EVOTECH and the EVOTECH logo are registered trademarks owned by Advanced Sterilization Products.

ISO/TC 198. Secretariat: ANSI

CDWA3 / CDWA4 CDWAH. Page 4. BASKET ZONE: water pressure, spray system performance, detergent chemistry and concentration, temperature.

Validation Strategy for an Automated Endoscope Reprocessor

MDRAO Importance of Bedside & Manual Precleaning CONMED.CA

What Does Validation Involve?

ANSI/AAMI ST91: 2015

The processing and cleaning of instruments provided by the company Weber Instrumente GmbH & Co. KG

Technical Data Monograph. The Instrument Protection Properties of Prolystica 2X Concentrate Neutral Detergent Versus Other Neutral Detergent Products

Monitoring your decontamination and reprocessing cycle

Improving the cleaning of surgical instruments the importance of pretreatment. qualification testing. Professor Andrew Smith

SUBJECT: Detection of Blood residue inside the biopsy channel of a scope

6/12/2016. Objectives. Manufacturers Instructions for Use: Impact on Patient Safety. Background

Amalgatome SD Sterilization Tray User Guide

Ideal Sterilization Method. Sterilization. Sterilization Practices in Healthcare Facilities

SUBJECT: Detection of Blood (hemoglobin) residue inside the biopsy channel of an endoendoscope

Low Temperature Processing. An Update on Guidelines, Standards and Requirements

TOTAL SOLUTION FOR ENDOSCOPE REPROCESSING

WASHING / DISINFECTION. Using the AMSCO WASHER / DISINFECTOR. Easy to use. Reliable. Compact LEAN STERISHEALTHCARE

OLYMPUS AUG -1 A9 :02

Clean Collaboration: Putting the pieces of the puzzle together on difficult to clean medical devices a Collaboration approach. Jahan Azizi BS, CBET

AS/NZS 4187 RELEVANT STANDARDS. Robyn Williams NUM CSSD LBHHS & Standards Australia Chair HE-023

Medical Device Reprocessing Assessment - General (Cleaning, Disinfection, and/or Sterilization)

Is Your Sterile Processing Department Ready for Survey?

RIGID AUTOCLAVABLE ENDOSCOPES

The efficacy of cleaning surgical instruments

Water for Instrument Processing

CAUTION: FEDERAL LAW (USA) RESTRICTS THESE DEVICES TO SALE BY OR ON THE ORDER OF A PHYSICIAN.

Getinge Automation System for washer-disinfectors

Instructions for the Use Surgical Instruments

ENDOSCOPE DECONTAMINATION- EDUCATION AND TRAINING COMPETENCY FRAMEWORK

Entellus Medical Sinuscope INSTRUCTIONS FOR USE

CSA Z314 Series Standards Understanding the Standards

Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Webinar FAQ

Important information

Safety, Efficacy & Microbiological Consideration

Instructions for: Cleaning, Sterilization, Inspection and Maintenance of Reusable Medical Devices. Instructions for Stryker Orthopaedics

Instructions for: Cleaning, Sterilization, Inspection and Maintenance of Reusable Medical Devices. Instructions for Stryker Orthopaedics

Testing the Endoscope Journey vs the Decontamination Process. Wayne Spencer

Endoscopy Reprocessing Trends. Wayne Spencer United Kingdom

Workshop on Medical Device Cleanliness: How Clean is Clean Enough? Sponsored by ASTM Committee F04 on Medical and Surgical Materials and Devices

Steam Sterilization Validation for Healthcare Institutions

Sterilization Service Revolution Service Enhancement in HA Hospitals

GETINGE 8666/8668 WASHER-DISINFECTOR. Always with you

T E S T R E P O R T. RehaWash Systems GmbH Berliner Strasse Weißwasser SN - No.: SN 12205

Christina Bradley Endoscope Decontamination A Webber Training Teleclass. Christina Bradley

Processing of Medical Devices: Risk Management and Plausibility Control

Cleaning and disinfection of surfaces and instruments

Understanding ANSI/AAMI ST 91. Flexible and Semi-Rigid Endoscope Reprocessing in Healthcare Facilities

If All Else Fails Read the Directions. Disclosure. Manufacturer s Instructions. Presenter: Sue Lafferty CHICA Canada Conference, May 2010

OBJECTIVES. Most Relative to SPD 10/22/ Top Ten Health Technology Hazards. Keeping Your Patients Safe And Your Hospital Out of The Headlines

Instructions for: Cleaning, Sterilization, Inspection and Maintenance of Reusable Medical Devices. Instructions for Stryker Orthopaedics

Endoscope Decontamination. A Tube in a Hole??

Critical aspects of endoscope decontamination-an overview of endoscope decontamination training for both engineers and users.

Reprocessing Complex Medical Devices The Way of the Future

Quality Control EQUIPMENT CONTROL. Monitor critical performance characteristics of sterilization, decontamination or cleaning equipment

ANNEX 2. to Guideline. for Testing, Validation and Monitoring of Automated Cleaning and Disinfection Processes for Medical Devices

Reusable 5x Holmium Laser Fibers Instructions for Use Model: LF0201X200, LF0201X272, LF0201X365, LF0201X550, LF Holmium and Nd:YAG Fiber

Safety, Efficacy & Microbiological Consideration

Transcription:

Best Practices for Reprocessing Robotic Instruments Nupur Jain Manager, Validation Engineering Services Intuitive Surgical, US

Agenda 1. Background on robotic instruments 2. Reprocessing procedure for robotic instruments 3. Demonstration of effective reprocessing

Robotic instruments Background

Background - Instruments

Perceived to be difficult to clean Intricate components Look different

Designed for reprocessing

Designed for reprocessing

Designed for reprocessing

Designed for reprocessing

Agenda 1. Background on robotic instruments 2. Reprocessing procedure for robotic instruments 3. Demonstration of effective reprocessing

Reprocessing Flow Storage Use Prep at point of use Sterilize Pre-clean Dry and Package Disinfect Manual / Automated Process

Cleaning Steps Pre-Clean 1. Soak in enzymatic detergent 2. Flush, Spray, Brush 3. Rinse and Inspect Manual Ultrasonic bath Automated Washer-Disinfector

Cleaning Steps: Pre-Clean

Cleaning Steps: Pre-Clean

Cleaning Steps: Pre-Clean

Cleaning Steps: Pre-Clean

Cleaning Steps: Pre-Clean

Cleaning Steps

Washer-Disinfectors Type Testing according to ISO 15883-1 Belimed WD 290 Getinge T88 Medisafe SI PCF Medisafe SI PCF Niagara Miele 8528 Steelco DS1000 Steelco DS610 Specialized load carriers and cycles Refer to Appendices of user manuals

Cleaning Steps

Cleaning Steps: Manual Process

Cleaning Steps: Manual Process

Advantages of Reprocessing procedure One process for all robotic instruments Straightforward and models best practices from standards Multiple washer-disinfectors are validated

Agenda 1. Background on robotic instruments 2. Reprocessing procedure for robotic instruments 3. Demonstration of effective reprocessing

Cleaning validation Intuitive Surgical follows international standards: ISO 15883-1 ISO 15883-5 AAMI TIR 12 AAMI TIR 30 ISO 17664 If instructions are not followed, Central Processing Department is responsible for validating cleaning steps that are performed.

Cleaning Validation Perform testing for reprocessing under worst case conditions: Soiling Cleaning Visual Inspection Extraction Protein Analysis Use of positive controls and negative control Acceptance criteria applied to visual inspection and protein analysis to determine if process renders the product clean

Soiling Worst case clinical use determines: Type of Test Soil Location of Test Soil Amount of Test Soil As a worst case, tips are completely immersed in Test Soil As a worst case, a safety factor is applied to the inner shaft volume. Product Family Amount of Test Soil in Shaft (µl) 8mm 600 Single Site 270 Stapler 50

Soiling

Soiling

Soiling Cleaning Cleaning is performed under worst case parameters: Test Soil is allowed to dry Time, pressure, concentration of detergent Visual inspection is not performed during cleaning but as an acceptance criterion at the end of the process During Washer-Disinfector validation, thermal disinfection is not performed

Soiling Cleaning Visual Inspection Extraction Extraction method is validated by performing a recovery efficiency calculation. Product Family Tips Recovery Efficiency Inner Shaft Recovery Efficiency 8mm 94% 82% Single Site 99% 87% Stapler 95% 85%

Soiling Cleaning Visual Inspection Extraction Protein Analysis Quantitative protein analysis is performed using the modified OPA method using a spectrophotometer Modified OPA method was validated method: Correlation coefficient > 0.999

Soiling Cleaning Visual Inspection Extraction Protein Analysis Acceptance criteria Total residual protein < 200µg Local guidelines: e.g. KRINKO: < 100µg Total residual protein corrected for surface area < 6.4 µg/cm 2 Product Family Surface Area (cm 2 ) 8mm 280 Single Site 237 Stapler 320 Visual Inspection no visible soil (magnification: 4x)

Results

Results

Results from Clinical Use All hospitals had total protein residuals of below 200ug. Hospital A-05: cleaning process was not followed and after corrections to the process validation, results are below 100ug. Data from hospital is consistent with validation data.

Results Visual Inspection Post disassembly visual examination 40 8mm da Vinci instruments from 11 hospitals

Results Visual Inspection

Summary Reprocessing method for robotic instruments: Straightforward Effective One process for all instruments Multiple WDs validated Robotic instruments can be reproducibly and effectively reprocessed Supported by laboratory validation data Clinical data consistent with laboratory validation data

THANK YOU