System Overview. Valve oven. Restrictor. Valve. Sample in and out

Similar documents
DETECTORS Dry Electrolytic Conductivity Detector - DELCD. Overview

Residual Solvents in medical cannabis analysis using the SRI 8610C FID GC June 2014

Operating the SRI RGD ( Reduction Gas Detector )

Improved Gas Chromatograph Method for the Analysis of Trace Hydrocarbon Impurities in 1, 3-Butadiene

Large Valve Oven Analyzers for Refinery Gas Analysis. ASTS 2015 Houston Shannon Coleman Application Specialist Agilent Technologies

Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D7504 on the Agilent 8890 GC System

Two Headspace Solutions for the Analysis of Dissolved Gases in Water by Method RSKSOP-175

2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial waste waters.

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations

Mach-RGA. The Fast Refinery Gas Analyzer. Analyzer

Sulfur in Natural Gas /LPG GC configuration Quick Start March 2013

Expanding the Versatility of Micro GC in the Shale Gas Play; The Agilent 490 Micro GC offers unprecedented reliability and flexibility

TVOC Heated FID VE 7

of Methane, Ethylene, Acetylene and Ethane in Water by Headspace-Gas Chromatography (HS-GC) with

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC

MicroSAM / Sitrans CV

STANDARD OPERATING PROCEDURES

ADVANCES IN MICRO GAS CHROMATOGRAPHY

Turnkey configurations for gas chromatographic analysis, including instrumentation, methods, supplies and service. Analysis of Natural Gas

Hydrocarbon Processing Industry Gas Chromatographs. Standard GC Analyzer Solutions

The Use Of Automated Backflush on the 7890A/5975A GC-MS System. Application. Setup Example Using Biomarkers. Authors. Abstract.

ADVANCES IN MICRO GAS CHROMATOGRAPHY

Gas Chromatography SSI-GC-003

Heated FID 109A. Rack Mount/Table Top NMGOC Analyzer

ADVANCES IN MICRO GAS CHROMATOGRAPHY

Agilent Gas Phase Solutions

Energy Saving Capillary Gas Chromatograph

Simplicity. Accuracy. Durability. Lotix. Automated High Temperature Combustion TOC Analyzer

Meet the airgc. Ambient air carrier. Top performance. catalytics accelerated

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE.

SCION GC Series. The Gas Chromatographers Choice for Separations

Prefractionator for Reliable Analysis of the Light Ends of Crude Oil and other Petroleum Fractions

Olefin Analyzer Solutions for Polymer Grade Ethylene and Propylene

Direct Injection of Fish Oil for the GC-ECD Analysis of PCBs: Results Using a Deans Switch With Backflushing. Application. Author.

METHOD TRANSLATION OF HJ FOR INTUVO

Agilent 490 Micro GC Biogas Analyzers

Analysis of Permanent Gases and Extended Hydrocarbons in Natural Gas and Natural Gas Liquids by Capillary Gas Chromatography

Using a Micro GC to measure, monitor, and optimize your process. ASTS, Houston June 2015 Shannon Coleman Agilent Technologies

Capillary GC Troubleshooting: a Practical Approach

Fast On-Site Mine Safety Analysis by the Agilent 490 Micro GC

Heated FID OVF Portable THC/TVOC Analyzer

Fast Refinery Gas Analysis Using the Agilent 490 Micro GC QUAD

METHOD 25 - DETERMINATION OF TOTAL GASEOUS NONMETHANE ORGANIC EMISSIONS AS CARBON

ENVIRONMENTAL MONITORING OF EMISSIONS FROM PAINT MANUFACTURING PROCESS

FINAL REPORT. Participants in the reported tasks 1. Ravi Subramanian 2. Manoranjan Misra 3. Kent Hoekman

8 stream capability with dual mode (H 2 S & total Sulphur)

Pesticide Analysis in a Complex Matrix using a 2D GC/MS System

Improving GC-MS Method Robustness and Cycle Times Using Capillary Flow Technology and Backflushing

Simplicity. Accuracy. Durability. Lotix. Automated High Temperature Combustion TOC Analyzer

Agilent 490 Micro GC. The Measure Anywhere Micro GC February 2011

ABB MEASUREMENT & ANALYTICS DATA SHEET. PGC5007 Total sulfur analyzer

Solutions Guide. MX Series II Modular Automation for Nano and Analytical Scale HPLC And Low Pressure Fluid Switching Applications

GC8000 Process Gas Chromatograph. Bulletin 11B08A01-01E

Improving Productivity in Your Environmental and Food Testing Labs with Agilent 7890A/5975C GC/MS System

METHOD 310A - DETERMINATION OF RESIDUAL HEXANE THROUGH GAS CHROMATOGRAPHY

Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides

AUTOMATIC PEROXYACETYL NITRATE (PAN) ANALYZER

Markes International Thermal Desorption G8125A, G8126A, G8127A, G8128A, G8131A, G8132A, G8136A, G8133A, MKI-MicroC Site Preparation Checklist

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application

Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730

SRI Gas Chromatographs for Cannabis Testing September 2016

Inlet-Residue-NGL Solutions for Natural Gas Processing Applications

Fast Analysis of Extended Natural Gas using the New Transportable INFICON Micro GC Fusion

Hichrom Limited 1 The Markham Centre, Station Road, Theale, Reading, Berkshire, RG7 4PE, UK Tel: +44 (0) Fax: +44 (0)

Chemical Analysis. DHA Analyzer Family. Optimized Solutions for Detailed Hydrocarbon Analysis. Gas Chromatography. think forward

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Historical Review: Fast & micro Gas Chromatography

GC Troubleshooting. Frank Michel, Katherine K. Stenerson T HCE

SIMPLIFY AND ACCELERATE GAS ANALYSIS. Micro GC Fusion GAS ANALYZER

Meet Ja, H2S, and BTU flare regulations while reducing your total cost of ownership

When optimization is discussed throughout this talk, it s in terms of performing more samples, in less time, while minimizing investment in capital

Complexity. C9+, H2S, methanol, wet gas. Fiscal / Custody transfer. Simpler installation /operation

Tar-Analyser TA Norman Poboss

Confirmation of THC in Oral Fluids Using High-Resolution 2-D GC/MS Application Note

Gas Chromatography Assignment Chem 543/443

1. Isotope Ratio Mass spectrometer

Standard Test Methods for Aromatics in Mineral Spirits by Gas Chromatography 1

Dionex ICS-6000 Consumables Tracking Quick Start Guide

Lab 6 Measurement of Ozone

TRACE ANALYTICAL. ta7000 Series UHP Gas Purity Monitors At the critical link in your manufacturing process. Features /Benefits

Customers need and want the answers Not just the hassles often associated with analytical chemistry!

Passive Monitoring: A Guide to Sorbent Tube Sampling for EPA Method 325. Nicola Watson July 2015

Preparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions

Petroleum Fuels. ogen process analyzers 6000 SERIES

Agilent Automated Card Extraction Dried Blood Spot LC/MS System

Greenhouse Gas Measurement Technology

TDLAS. Application Note. Tunable Diode Laser Absorption Spectroscopy (TDLAS) for Carbon Dioxide Measurements In Natural Gas Processing

GAS CHROMATOGRAPH FOR MULTIPLE IMPURITIES

Experiment 11. DNA Composition by HPLC 1,2

Simple Run Table Entry Cuts Sample onto Column 2. Method Developers Calculator to Tools correctly set flows and restrictor size

MARKET FOCUSED - environmental ORGANOCHLORINE PESTICIDES

Fast Gas Chromatography using Heated Headspace Gas Auto Sampling Techniques: Polyethylene Pellets, Product

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

Please enclose a item wise compliance report along with the technical bid.

Stack Sampling. requires airflow through the sampling probe to be at the same rate as that flowing in the waste gas stream isokinetic

Envantage DAKServe LTD LLC. SuperGC TM High Pressure Hydrocarbon Liquid/Gas Analyzer

A New Horizon of Performance Agilent G1888 Network Headspace Sampler

Separation and Quantitation of a Mixture of Alcohols by Gas Chromatography

Large Volume Injection of Organochlorine Pesticide Extracts Using a Programmable Temperature Vaporizer in the Solvent Split Mode

Transcription:

System Overview The GC system is designed to provide a continous reading of total hydrocarbons in a gas stream, while periodically performing a chromatographic separation of the sample to determine the composition of the sample gas stream. This is accomplished with a flow restrictor for the monitoring FID (#2), and 10-port gas sampling valve & sample loop for the speciation FID (#1). In the temperature-programmable column oven, the 3-meter Hayesep-D packed column efficiently separates the constituent hydrocarbons peaks (C 1 ). Valve oven Restrictor Valve Column oven Hayesep-D packed column Restrictor Sample in and out Dual FID detectors Sample in Valve 30µL sample loop Sample out In the heated valve oven, the sample inlet is split between the 10-port gas sampling valve and the restrictor. The restrictor controls the sample flow rate for the hydrocarbon monitoring FID (#2). The gas sampling valve periodically injects the contents of the sample loop onto the column for speciation by FID #1. In the temperature-programmable (ambient to 400 o C) column oven, the 3-meter Hayesep-D packed column efficiently separates the hydrocarbons up to C 6. Valve injections are automated by the built-in PeakSimple data system, and the timing is determined by the user. FID #1 FID #2 3-meter Hayesep-D packed column Designated as FID #1, the speciation FID signal occupies channel 1. FID #2, the total hydrocarbon monitor, is displayed on channel 2. Using Data logger mode, PeakSimple will display a scaled and calibrated result in large numbers for at-a-glance visibility. PeakSimple has an alarm function which you can set to alert you when the system receives out-of-range readings. A built-in, whisper quiet air compressor supplies the combustion air for both FID detectors.

Theory of Operation The sample gas stream is connected to the bulkhead fitting on the front of the valve oven. Inside the valve oven, the sample flow is split through a T. On one side of the T, the flow is directed through a restrictor, through the sampling loop on the 10-port valve, and on to the column and FID #1 for separation and detection. The other side of the T directs the sample gas stream through a much longer restrictor, and on to FID #2. FID #2 FID #1 10-port gas sampling valve with 30uL sample loop 7ft. restrictor The sample gas stream is connected here. 3in. restrictor Tee Heated valve oven Sample In Sample Out The 10-port gas sampling valve is plumbed to backflush the column to the detector. At an automatically repeating time interval controlled through the built-in PeakSimple data system, the gas sampling valve injects the contents of its 30µL sample loop into the Hayesep-D column. There, it is separated into the constituent hydrocarbon peaks, which are detected by FID #1. In the LOAD position, the carrier gas flows into the column while sample gas flows through the sample loop and out to vent. In the INJECT position, the carrier gas flows first through the sample loop, then sweeps the sample into the Hayesep-D column.

1. Set the valve oven temperature to 90 o C. General Operating Procedure 2. For FID #1, set the amplifier gain switch to MEDIUM. For FID #2, set the gain switch to LOW. 3. Set the FID hydrogen flow to 25mL/minute, and the FID air supply flow to 250mL/minute. The approximate pressures required to achieve these flow rates are printed on the right-hand side of your GC. 4. Ignite each FID by holding up the ignitor switch for a couple of seconds until you hear a small POP. The ignitor switches are located on the front panel of the GC. Verify that the flame is lit by holding the shiny side of chromed wrench directly in front of the collector outlet/fid exhaust vent. If condensation becomes visible on the wrench surface, the flame is lit. To prevent flameout, set the ignitor voltage to -750 by adjusting the trimpot on the FLAME IGNITE zone with the supplied screwdriver. 5. Connect zero gas to the sample inlet at 10psi. (Zero gas has no hydrocarbons). Zero the FID #2 signal by clicking on the Auto Zero button on the left side of the channel 2 chromatogram window. Auto Zero button 6. Disconnect the zero gas, and connect calibration gas standard (typically 100% methane) to the sample inlet at 10psi. The FID signal should increase approximately 300 millivolts while running 100% methane.

General Operating Procedure continued 7. In PeakSimple, open the Channel details dialog box for channel 2. The Datalogger mode area is in the lower right corner of the Channel details dialog box. Click in the checkbox to activate Datalogger mode, then enter the gain factor which will multiply the 300 millivolt signal to produce the desired concentration unit. For example: 300 x.33 = 100 if the desired unit is percent; or 300 x 3333 = 1,000,000 if the desired unit is parts per million (ppm). The FID #2 signal times the gain factor will display in the channel 2 chromatogram window. Event table 8. Enter an isothermal temperature program in the Temperature Program dialog box. For example: Initial Temp Hold Ramp Final Temp 180 o C 15.00 0.00 180 o C 9. Type in an Event table. Example: Time Event 0.00 Zero signal 0.050 G ON 4.00 G OFF 10. Start the analysis by pressing the spacebar on your computer keyboard, or by pressing the RUN button on the front of the GC. 11. In PeakSimple, input the retention windows ( Add component ) to identify the individual hydrocarbon peaks (methane, ethane, propane, butane, etc.) Calibrate the individual hydrocarbon peaks. NOTE: The GC system is plumbed for backflush. This gives you the option to set the valve program to backflush the heavier hydrocarbons after the desired peaks have been separated. For example, if your application required separation of hydrocarbons up to C 5, you could set the valve to backflush after the elution of the C 5 component(s), and all the heavier hydrocarbons would together produce one large peak. See the Expected Performance: Column Backflush to FID #1 page.

Expected Performance Total Hydrocarbon Monitoring with Simultaneous Chromatographic Separation The top chromatogram window is channel 1, displaying the FID #1 signal response to a valve injection of natural gas standard. The bottom chromatogram window, in Data logger mode, shows the concentration of hydrocarbons in the desired unit. Designate the unit in the Channel details dialog box, in the Data logger mode section. FID #1 gain = MED FID #1 temp = 150 o C FID #1 ignitor = -700 Valve temp = 90 o C FID #2 gain = LOW FID #2 temp = 150 o C FID #2 ignitor = -700 Events: Time Event 0.000 Zero 0.050 G ON 4.000 G OFF Temperature program: Initial Hold Ramp Final 180 o C 100.00 0.00 180 o C RESULTS: Component Retention Area Methane 0.616 7175.4520 Ethane 0.900 1836.1900 Propane 1.500 1864.8260 Iso-butane 2.566 1186.8560 N-butane 2.916 1258.8330 C 5 + 7.533 989.2620

Expected Performance Column Backflush to FID #1 The GC system is plumbed for backflush. This gives you the option to set the valve program to backflush the heavier hydrocarbons after the desired peaks have been separated. In the example below, the valve was set to backflush after the elution of ethane, and the C 3 produce one large peak. Sample: 1mL 1000ppm C 1 standard FID gain = HIGH FID temp = 300 o C FID ignitor = -600 Valve temp = 90 o C Events: Time Event 0.000 G ON 3.000 G OFF C 3 Temperature program: Initial Hold Ramp Final 70 o C 1.00 10.00 220 o C 220 o C 5.00 0.00 220 o C RESULTS: Component Retention Area Methane 0.733 3530.1805 Ethane 2.200 13632.4805 C 3 4.583 111209.5330