Qualification of High-Performance Liquid Chromatography Systems

Similar documents
Analytical Instrument Qualification According to USP <1058>: Requirements and Examples for the Agilent 1290 Infinity LC System

Validation Strategies for Equipment from Multiple Vendors

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

Using Empower SystemsQT Qualification Tool

According to USP <1058>

Performance characteristics of the 1260 Infinity Quaternary LC system

Agilent 1290 Infinity Binary LC System with ISET Emulation of a Waters Alliance 2695 LC Applying Concave, Convex, and Linear Gradients

Validation Strategies for Equipment from Multiple Vendors

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

2018 by Agilent Technologies Agilent CrossLab Compliance Services

The procurement, qualification and calibration of lab instruments: An Overview

Application Note. Environmental. Agilent 1290 Infinity Binary LC. with ISET and fine-tuning. Agilent 1290 Infinity Binary LC.

GOOD LABORATORY PRACTICE Agilent Cary 8454 UV-Visible Spectroscopy System

2018 by Agilent Technologies Agilent CrossLab Compliance Services

Technical Overview. Author. Abstract. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Emulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture

ICH Q2(R1) A Primer. cgmp. GxP PIC/S SOP ISO QA/QC LOD/LOQ. Validation of Analytical Methods API EP FDA OECD GCP

Agilent 1290 Infinity Binary LC System with ISET - Emulation of the Waters Alliance 2695 LC system analyzing b-blockers

Application Note. Author. Abstract. Pharmaceuticals. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Integrated IC Systems Installation Qualification Dionex Corporation

Analytical HPLC solutions

COMPLIANCE BY DESIGN FOR PHARMACEUTICAL QUALITY CONTROL LABORATORIES INSIGHT FROM FDA WARNING LETTERS

Strategies for Risk Based Validation of Laboratory Systems

Method Transfer from an Agilent 1200 Series LC to an Agilent 1260 Infinity II LC

Comparing gradient transfer of isocratic hold and delay volume addition using the Agilent 1290 Infinity LC with ISET

Controlling Agilent 1200 Series Rapid Resolution LC systems through Waters Empower chromatography data software. Technical Overview.

Technical Overview. Author. Abstract. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Agilent 1290 Infinity LC with ISET under Waters Empower control Emulation of Agilent 1100 Series LC

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

ISET. Agilent 1290 Infinity Binary LC with ISET, emulation of the Waters Alliance 2695 LC system analyzing aromatic acids. Application Note.

Risk-based Approach to Part 11 and GxP Compliance

CHEMICAL SOLUTIONS, Inc.

Agilent 1290 Infinity Quaternary LC Support of Columns with 2.1 to 4.6 mm ID to 1200 bar

á1058ñ ANALYTICAL INSTRUMENT QUALIFICATION

2018 by Agilent Technologies Agilent CrossLab Compliance Services

Automated Scouting of Stationary and Mobile Phases Using the Agilent 1290 Infinity II Method Development Solution

Agilent Max-Light Cartridge Cell Information for G4212A and G4212B DAD

Application Note. Authors. Abstract. Biopharmaceuticals

Developing Purification Strategies for the Agilent 1260 Infinity II Preparative LC/MSD System

2017 by Agilent Technologies Agilent CrossLab Compliance Services

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals.

Spectroscopy Solutions for Pharmaceuticals:

Analysis of amoxicillin and five impurities on the Agilent 1220 Infinity LC System

Scale-up with the Agilent SD-1 Purification System analytical and preparative runs on a single system

Application Note. Author. Abstract. Small Molecule Pharmaceuticals & Generics. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Regulatory Expectations, Standards & Guidelines

Lessons from Pharmaceutical Laboratory related FDA Warning Letters

Welcome to our E-Seminar: Automation Tools for Efficient Result Review According to GMP

ultra why micro-lc? product note introducing the ExpressLC -Ultra system: UHPLC with all the advantages of micro-lc

Automating HPLC and GC Analytical Method Validation Fusion AE Software Program White Paper

2017 by Agilent Technologies Agilent CrossLab Compliance Services

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS

Using a Diode Array UV-Vis in Regulated Environments

Regulations in Pharmaceutical Laboratories

Method Transfer from an Agilent 1260 Infinity LC to an Agilent 1260 Infinity II LC

The Leading Solution for Method Validation VALIDAT

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

Determination of Asarinin in Xixin (Asari Radix Et Rhizoma)

Automating HPLC and GC Analytical Method Validation Fusion AE Software Program White Paper

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

á1058ñ ANALYTICAL INSTRUMENT QUALIFICATION

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

High-Throughput LC/MS Purification of Pharmaceutical Impurities

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

Automated fraction re-analysis does it really make sense? Application. Udo Huber. Abstract

For analytical laboratories. A primer Good laboratory practice and current good manufacturing practice

UV/Vis Spectrophotometry. Time for New Demands SPECORD PLUS

18 Spectroscopy 21(11) November 2006

Fast Method Development Using the Agilent 1290 Infinity Quaternary LC System with Column Selection Valve

CRITICAL ASPECT ANALYTICAL TEST REVIEW

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES

Protein Separation with ph Gradients Using Composite Buffer Systems Calculated by the Agilent Buffer Advisor Software

Performance characteristics of the Agilent 1100 Series capillary LC system using diode-array UV and MS for detection. Technical Note.

Application Note. Author. Abstract. Small Molecule Pharmaceuticals. Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany

A Simple Rapid and Sensitive Method Development for Quantification of Quetiapine Fumarate in Bulk and Dosage Forms Using RP-HPLC

Certified Reference Material - Certificate of Analysis

Agilent 8453 Spectrophotometer and UV-visible ChemStation for Dissolution Testing. Solution Bundles for automated Dissolution Testing

Impurity Control in the European Pharmacopoeia

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Chemical Analysis. DHA Analyzer Family. Optimized Solutions for Detailed Hydrocarbon Analysis. Gas Chromatography. think forward

GOOD CHROMOTOGRAPHY PRACTICES (February 2019) DRAFT FOR COMMENTS

Optimizing Performance Bench Scale Service Checklist

Agilent OpenLAB Chromatography Data System IT S TIME TO ACHIEVE MORE

Automated Purification Software for Agilent OpenLAB CDS

AUTOMATIC PEROXYACETYL NITRATE (PAN) ANALYZER

QUANLYNX APPLICATION MANAGER

Agilent 1260 Infinity LC & LC/MS Purification Systems. Purify your way

Mass Detection for the Chromatographer Waters ACQUITY QDa Detector

GUIDANCE NOTES ON ANALYTICAL METHOD VALIDATION

How Compliant and Efficient is Your Spectroscopy Workflow?

Good Laboratory Practice Facilities: Building and Equipment Hedwig Beernaert

Tank Scale Service Checklist

Automated alternating column regeneration on the Agilent 1290 Infinity LC

Agilent 6400 Series Triple Quad LC/MS System

easy. HPLC has never been so Agilent 1120 Compact LC Our measure is your success.

Transcription:

Qualification of High-Performance Liquid Chromatography Systems L. Huber Hewlett- Packard GmbH, Waldbronn December 1998 Manuscript of an article published in BioPharm, Vol 11 Number 11, November 1998, pages 41 and 65/66 The Author: Dr. Ludwig Huber Worldwide product marketing manager HPLC Hewlett-Packard Waldbronn P.O. Box 1280 D-76337 Waldbronn Fax: +49-7802-981948 Phone: +49-7243-602 209 E-mail: Ludwig_Huber@hp.com URL: www.labcompliance.com LC_QUAL.doc Page 1 of 9 www.labcompliance.com

Two Validation Reference Books for lab managers, chemists, QA- and IT-personnel Validation of Computerized Analytical Systems Ludwig Huber 267 pages, US$ 189.- + tax Interpharm HAA-IPVCCAS Fax +1-847-459-6644 ISBN 0-935184-75-9 Hewlett-Packard P/N 5959-3879 Strategies, SOPs and Checklists for Fast and Troublefree Implementation validation during software development using the software development lifecycle concept validation responsibilities of users and suppliers qualification of vendors installation and operational qualification validation of macros validation of existing computer systems definition and archiving of raw data strategy for security change control practical examples of how to accomplish the required validation in a cost effective and efficient manner Validation and Qualification in Analytical Laboratories Ludwig Huber Approx. 300 pages Interpharm HEA-IPVQAL Fax +1-847-459-6644 ISBN 1-57491-080-9 Hewlett-Packard P/N 5965-0036 US$ 219.- + tax (Nov. 1998) Strategies, SOPs and Checklists for Fast and Troublefree Implementation equipment qualifications: DQ, IQ, OQ, PQ design, installation, operation, routine use validation of analytical methods: implementation strategy and performance characteristics of routine/non routine, standard/non standard methods evaluation of measurement uncertainty qualification of home made and purchased (certified) reference compounds proficiency testing for external analytical quality control tests for more than 30 selected instruments qualification of people SOPs for method validation and equipment testing OQ test cases: selecting parameters and limits Help to improve quality and to comply with GLP, GCP, cgmp, ISO 9000, EN45001 and ISO Guide 25 Free sample chapters from: www.labcompliance.com LC_QUAL.doc Page 2 of 9 www.labcompliance.com

Qualification of High-Performance Liquid Chromatography Systems Ludwig Huber, Hewlett-Packard, D-76337 Waldbronn, Germany Phone: +49-7243-602209, fax: +49-7243-981948 E-mail: ludwig_huber@hp.com URL: http://www.labcompliance.com Introduction Equipment qualification is a formal process that provides documented evidence that an instrument is fit for its intended use and kept in a state of maintenance and calibration consistent with its use (1). The entire qualification process consists of four parts: Design qualification (DQ) Installation qualification (IQ) Operational qualification (OQ) Performance qualification (PQ) General definitions, the overall objectives, and how to select parameters and criteria for of each phase are described in detail in references (1-4). Individual steps and frequency of qualifications depend on the type of equipment and application. This article describes steps required for the qualification of HPLC systems. Qualification steps of other instruments such as UV/Vis Spectrophotometers, Capillary Electrophoresis instruments and Mass-Spectrometers will be described in future articles of this series. This article will focus on qualification of hardware. Qualification of software and computer systems has been described in reference. 5. To be as practical as possible, a real application scenario from a QA/QC laboratory is used. In case the reader has a different application, the templates can be used as described in this paper, however, the qualification parameters and acceptance criteria should be modified accordingly. Scenario A QA/QC laboratory will purchase 15 automated HPLC systems for operation in a Current Good Manufacturing Practice environment. The HPLC systems will be used for compound and impurity analysis. Most of the samples to be analyzed in the laboratory can be run in isocratic mode, but some require gradient analysis. Most of the samples can be monitored with a single wavelength UV signal, but some require in addition peak purity analysis using a spectral UV detector. All instruments should have the same configurations. This has the advantage that they all can be used for all LC_QUAL.doc Page 3 of 9 www.labcompliance.com

applications and there is only one test procedure and one set of acceptance criteria required. For instrument control, data acquisition and data evaluation a computer with chromatography software is used. Design qualification Design qualification (DQ) describes the user requirements and defines the functional and operational specifications of the instrument. It should ensure that instruments to be purchased have the necessary functions and performance that will enable them be suitable for the intended applications. Later on the DQ document will be used as a basis for tests in the OQ phase. Table 1 lists elements with examples that should be included in the design qualification for the HPLC systems in the selected QA/QC laboratory. Please note that all instruments are required to have gradient pumps, thermostatted column compartments and diode-array detectors to ensure that they all can be used for all applications. LC_QUAL.doc Page 4 of 9 www.labcompliance.com

Intended use Design elements User requirement specification for the HPLC analysis Functional specifications Pump Detector Autosampler Column compartment Computer Operational specifications Examples Analysis of drug compounds and impurities Up to 100 samples / day Automated over-night analysis Limit of quantitation: 0.1% Automated confirmation of peak identity and purity with diode-array detection Automated compound quantitation and printing of report Binary or higher gradient UV/Vis Diode-array, 190-900 nm 100 samples, 0.5 ul to 5 ml sample volume 15 to 60 Deg C, peltier controlled System control, data acquisition for signals and spectra, peak integration and quantitation, spectral evaluation for peak purity and compound confirmation. Electronically save all chromatograms generated by the system. Detector: Baseline noise: <5 x 10-5 AU Sampler: Precision inj. volume: <0.5% RSD, sample carry over: <0.5% Pump: precision of retent.time: <0.5% RSD User instructions Operational manual on paper Computer based tutorial Validation/qualification Vendor must provide procedures and services for IQ and OQ Maintenance Vendor must deliver maintenance procedure and recommend schedule Training Instrument must include early maintenance feedback for timely exchange of most important maintenance parts Maintenance procedures must be supplied on Multimedia CD ROM Vendor must provide familiarization and training Table 1. Design qualification elements. LC_QUAL.doc Page 5 of 9 www.labcompliance.com

Installation qualification Installation qualification establishes that the instrument is received as designed and specified, that it is properly installed in the selected environment, and that this environment is suitable for the operation and use of the instrument. Table 2 lists steps as recommended before and during installation. IQ should include analysis of a test sample. A successful run of such a sample verifies correct installation of all modules and electrical and fluid connections. Before installation Obtain manufacturer s recommendations for installation site requirements. Check the site for the fulfillment of the manufacturer s recommendations (utilities such as electricity, and environmental conditions such as humidity and temperature). Allow sufficient shelf space for the equipment, SOPs, operating manuals and software. During installation Compare equipment, as received, with purchase order (including software, accessories, spare parts) Check documentation for completeness (operating manuals, maintenance instructions, standard operating procedures for testing, safety and validation certificates) Check equipment for any damage Install hardware (computer, equipment, fittings and tubings for fluid connections, columns in HPLC and GC, power cables, data flow and instrument control cables) Switch on the instruments and ensure that all modules power up and perform an electronic self-test Identify and make a list with a description of all hardware, include drawings where appropriate. Run test sample and compare chromatogram print-out with reference chromatogram List equipment manuals and SOPs Prepare an installation report Table 2. IQ steps before and during installation (from reference 2) Operational qualification (OQ) LC_QUAL.doc Page 6 of 9 www.labcompliance.com

Operational qualification (OQ) is the process of demonstrating that an instrument will function according to its operational specification in the selected environment It verifies that the HPLC system complies with key functional and operational requirements as specified in the design qualification. In modular HPLC systems it is recommended to perform system tests (holistic testing), rather than performing tests module by module (modular testing). Individual module tests should be performed as part of the diagnosis if the system fails. Furman and Layloff, two US FDA employees (6) have first promoted this holistic approach computerized HPLC systems. Table 3 includes test parameters, procedures and recommended user limits. At the beginning of the test, a leak test based on the instrument s flow rate accuracy is performed. If this test fails, most other tests are likely to fail. Reference 3 gives further background information on the selection of parameters and acceptance limits. Parameter Procedure (*) User Limit Leak testing Flow test by volume or weight/time ± 5 % Baseline drift ASTM Method E19.09, 20 min <2 x 10-3 AU Baseline noise ASTM Method E19.09, 20 x 1 min <5 x 10-5 AU Precision of injection volume 6 x injection of caffeine standard, 0.3 % RSD RSD of peak areas Precision of flow rate 6 x injection of caffeine standard, 0.5 % RSD RSD of retention times Detector linearity inject 5 standards >1.5 AU, 5% Wavelength accuracy holmium oxide filter ± 1 nm Temperature accuracy comparison with external measuring device ± 1 C Temperature precision Autosampler carry over monitoring temperature over 20 min Injection of blank solvent after large concentration ± 0.25 C < 0.5 % Mobile phase composition accuracy (*) For detailed procedure, see reference 7 Step gradients from 4 to 7 % B, step heights relative to 100%, with acetone tracer ± 1 % Table 2. Test parameters and acceptance criteria for case study. LC_QUAL.doc Page 7 of 9 www.labcompliance.com

Performance qualification Performance Qualification (PQ) is the process of demonstrating that an instrument consistently performs according to a specification appropriate for its routine use. Important here is the word consistently. The test frequency is much higher than for OQ. Another difference is that PQ should always be performed under conditions that are similar to routine sample analysis. For a chromatograph this means using the same column, the same analysis conditions and the same or similar test compounds. PQ should be performed on a daily basis or whenever the instrument is used. The test frequency not only depends on the stability of the equipment but on everything in the system that may contribute to the analysis results. For a liquid chromatograph, this may be the chromatographic column or a detector s lamp. The test criteria and frequency should be determined during the development and validation of the analytical method. In practice, PQ can mean system suitability testing, where critical key system performance characteristics are measured and compared with documented, preset limits. For example, a well-characterized standard can be injected 5 or 6 times and the standard deviation of amounts are then compared with a predefined value. PQ tests are applications specific. If the limits of detection and/or quantitation are critical, the lamp s intensity profile or the baseline noise should be tested. They should use the same column and chemicals as used for the real samples. Test should include precision of the amounts precision of retention times resolution between two peaks peak width at half height or peak tailing baseline noise wavelength accuracy of the UV/Visible wavelength detector, preferably using built-in holmium-oxide filters Documentation On completion of equipment qualification, documentation should be available that consists of: Design qualification document IQ document (includes description of hardware and software) Procedures for OQ testing OQ test reports PQ test procedures and representative results OQ test reports should include test parameters, acceptance criteria and actual results. An example of a typical test report is shown in figure 1. LC_QUAL.doc Page 8 of 9 www.labcompliance.com

HPLC Instrument Ver f cat on Test method: Data File Directory: Original Operator C\HPCHEM\1\VERIF\Check.M C\HPCHEM\1\VERIF\Result.D Dr. Watson Test item User limit Actual Com DAD noise <5x10-5AU 1x10-5AU Pass Baseline drift <2x10-3AU/hr 1.5x10-4AU/hr Pass DAD WL calibration ±1 nm ±1 nm Pass DAD linearity 1.5 AU 2.2 AU Pass Preci. of ret.times <0.3 % RSD RT 0.15 % RSD RT Pass Temp.stability ±0.15 C ±0.15 C Pass Precision of peak area < 0.5% RSD 0.09 % RSD Pass Verification Test Overall Results Pass HP 1100 Series System, Friday, January 16, 1998 Test Engineer Name: Signature: Figure 1. OQ report obtained from the HP ChemStation for the HP1100 Series HPLC (from ref 2) References: 1. P.Bedson and M.Sargent, The development and application of guidance on equipment qualification of analytical instruments, Accreditation and Quality Assurance, 1 (6), 265/274 (1996) 2. L. Huber, Validation and qualification in analytical laboratories, Text book, 300 pages, 1998, Interpharm, Bufallo Grove, IL, USA, Nov. 1998 http://www.t-online.de/home/huberl/validat.htm (book) 3. L. Huber, Equipment Qualification in Practice LC/GC Magazine, Vol 16, February 1998, 148-156 http://www.t-online.de/home/huberl/validat.htm (book) 4. L.Huber and L.Welebob, Selecting Parameters and Limits for Equipment Operational Qualification, Accreditation and Quality Assurance, 2 (7), 316/322 (1997) 5. L. Huber, Validation of computerized analytical instruments, Interpharm, Buffalo Grove, IL, USA, ISBN 0-935184-75-9, 267 pages, May 1995, Hewlett-Packard Part Number: 5959-3879 http://www.t-online.de/home/huberl/validat.htm (book) 6. W.B. Furman, T.P. Layloff and R.F. Tetzlaff, Validation of computerized liquid chromatographic systems, J.AOAC Intern, 77 (5), 1314-1318 (1994) 7. Operational Qualification/Performance verification for complete HP 1100 Series HPLC Systems, Hewlett-Packard Waldbronn, 1996, Part Number G2170-90102 LC_QUAL.doc Page 9 of 9 www.labcompliance.com