Sterilization of Schweickhardt Instruments

Similar documents
The stainless steels used in the manufacture of surgical instruments can be divided into two main classes.

Surgical Instruments - Instructions for use

CAUTION: FEDERAL LAW (USA) RESTRICTS THESE DEVICES TO SALE BY OR ON THE ORDER OF A PHYSICIAN.

Water for Instrument Processing

Performance, Beauty & Long Life STAINLESS STEEL CARE & MAINTENANCE. Stainless Steel Performance. Stain less Steel

Technical Data Monograph. Stainless Steel Instrument Maintenance Using Prolystica Restore Descaler & Neutralizing Detergent

Stainless Steel Maintenance Guide

Instructions for the Use Surgical Instruments

Made to perform. Extraction Instruments .,. ISFV. Quality Through Innovation

HURST CORPORATION STAINLESS STEEL

Instruction Guide Re-Usable Surgical Instruments. Instruction Guide Re-usable surgical instruments (GA 30.1)

Guide to Optimal Steam Generation

What Are Metal Standards and Why Are They Important?

Recommendation

1. Scaling. H.O.: H-5/21, KRISHNA NAGAR, DELHI Tel.: , Fax:

Galvanic Corrosion Prevention Guide for Water Cooling Systems

PROCESS CHECK REPORT

Selection criteria for Biomaterials

Stainless Steels for the Food Processing Industries

Corrosion Control and Cathodic Protection Data Sheet

ATI 332 ATI 332. Technical Data Sheet. Stainless Steel: Austenitic GENERAL PROPERTIES TYPICAL ANALYSIS PHYSICAL PROPERTIES

Secondary Drinking Water Standards: Guidance for Nuisance Chemicals On this page:

Evaluation of. Wonder Gel Stainless Steel Pickling Gel and Derustit SS-3 Stainless Steel Cleaner. Prepared for

CLEANING AND PASSIVATING OF STAINLESS STEEL EQUIPMENT

Connexion Developments Ltd + Solenoid Valves (UK) Ltd Unit 3, Rainbow Court, Armstrong Way, Great Western Business Park, Yate, Bristol, BS37 5NG,

THE EXPERTS IN INDUSTRIAL CLEANING

ATI 825 ATI 825. Technical Data Sheet. Nickel-base Alloy INTRODUCTION PRODUCT FORMS SPECIFICATIONS & CERTIFICATES (UNS N08825)

Environmental concerns include the presence of oxygen, moisture (water), contact with dissimilar metals, and chemicals.

Laboratory Balances of the Cubis Balance Range Meet the Requirements of the Pharmaceutical Industry for Cleanability. White Paper

CURRENT METHODOLOGIES & CHEMISTRIES UTILIZED IN EFFECTIVE PASSIVATION PROCEDURES

Water supplied by Marafiq does not meet the process requirements.

SAFETY & SANITARY DESIGN STANDARDS. S afety & Sanitary. Design Standards. Revision: B Date: 10/2013

Loaner Program Guidance

Glossary of Steel Terms

CARE & MAINTENANCE INITIAL CLEANING AND MAINTENANCE

Metal Finishing Products and Service META-MATE ZINCATE 40 "A CONCENTRATED LIQUID ZINCATE FORMULATION FOR THE PRETREATMENT OF ALUMINUM AND ITS ALLOYS"

RR-17 AN IDEAL & SAFE RUST REMOVER

Water The Universal Solvent

MEGA CHECK Mega check (Cleaner) Coloring Penetrant Detection Agent for Crack Detection

1. CORROSION OF REINFORCEMENT

OUTLINE ATMOSPHERIC CORROSION INTEGRITY MANAGMENT

Pipe Protection Tape System

Section # 2. Assembly Fasteners. ICCA Assembly Fasteners - Section # 2 #

SECTION 9. MAGNESIUM AND MAGNESIUM ALLOYS

Materials of Engineering ENGR 151 CORROSION ELECTRICAL PROPERTIES

Cleaning of Stainless Steel

Loaner Programme Guidance

Loaner Programme Guidance

Rusting is an example of corrosion, which is a spontaneous redox reaction of materials with substances in their environment.

Thermal Diffusion Galvanizing

CHEMICAL METAL PRE-TREATMENT

2.3 Chemical Changes corrosion Kinds of Corrosion

Use the arrow keys to easily navigate through the course.

Cleaning and Sanitizing

Corrosion and batteries

Ultrex CAA Product Code: Revised Date: 10/09/2006. Ultrex CAA Heavy Duty Alkaline Product Derusting - Descaling - Activation

Welding Job Knowledge

Understanding Your Test Results

CONCRETE STEPS, HANDRAILS, AND SAFETY RAIL

The material is also available in a High Strength condition with a guaranteed minimum of 140 ksi 0.2% yield strength.

TECHNICAL BULLETIN. Dissimilar Metal Corrosion in Roofing Installations

family of stainless steels can be divided into five (5) categories:

AL 29-4C AL 29-4C. Technical Data Sheet. Stainless Steel: Superferritic INTRODUCTION (UNS S44735)

ATTENTION: READ THESE INSTRUCTIONS VERY CAREFULLY BEFORE YOU PREPARE AND USE THE PRODUCT FOR THE FIRST TIME.

Zinc Silicate (Inorganic)

Technical Data Monograph. The Instrument Protection Properties of Prolystica 2X Concentrate Neutral Detergent Versus Other Neutral Detergent Products

ATI 321 /ATI 347 /ATI 348

Corrosion Issues & Test Methods. Jerry Byers

The Basics of Alkaline In-Process Cleaning. for Metal Substrates. John Sparks Oakite Products, Inc. Berkeley Heights, New Jersey

Team Metal Finishing Inc.

Medical Device Reprocessing Assessment - General (Cleaning, Disinfection, and/or Sterilization)

Pre-Lab Exercises Lab 5: Oxidation and Reduction

Metallurgical Failure Analysis of Power Transmission Components

Workshop on Medical Device Cleanliness: How Clean is Clean Enough? Sponsored by ASTM Committee F04 on Medical and Surgical Materials and Devices

SEASTROM Manufacturing Co., Inc Seastrom Street Twin Falls, Idaho Fax (208)

MINIMIZING POLLUTION FROM CHEMICAL PRETREATMENT PROCESSES

Hydroblasting Standards

Dr. Ali Kadhim. Bricks

Guideline for Prevention of Damage in Water Heating Installations

Stainless Steel Innosoft - Innoclean - Innoprotect

Demineralisation. PUROTAP expert. Installation Function Operation Service. Demineralised heating water.

Anticorrosive Coatings

Specification Product Form ASTM ASME AMS Federal or Military Plate, Sheet and Strip B 127 SB QQ-N-281 B 163 B165 B164 B564

CHAPTER INTRODUCTION

Drinking Water Corrosion Control and POU/POE Treatment

Distribution Review. Corrosion Control. Corrosion Control Vocabulary. American Water College 1. Corrosion Control Training Objectives

Professional instrument reprocessing in CSSD

Daraclean Selection. Finding the right formulation MAGNAFLUX SOURCE

Label Tape FAQs and Technical Information Pertaining to Brother Label Printers PT-9700PC and PT-9800PCN

Stainless Steel (17/4PH&630) Bar

Heat Treatment of Tubes for Condenser, Feedwater Heater, and Shell & Tube Heat Exchangers

Residential Doors. Limited Warranties Care & Maintenance Painting Instructions. / / Date Installed

These elements are in carbon steels in minimal amounts, usually less than 1%.

MONITORING PASSIVATION ON STAINLESS STEEL WITH OPEN CIRCUIT POTENTIAL TECHNOLOGY

Identification. Type Analysis

Energy Transfer in a Hydraulic System. Lecture (1) Hydraulic and Pneumatic Circuits 4th year Mech. power Engineering Dept.

Metalworking/Automotive

Stainless Steel Bar

OVERVIEW OF CORROSION PROTECTION ALLOY MP-ACQ FASTENER COATING SYSTEM FOR USE IN ACQ-PRESSURE TREATED LUMBER

Reinforcing Wire in Solutions with Chloride and Cement

Transcription:

Sterilization of Schweickhardt Instruments

INTRODUCTION l Most dental instruments are made of corrosion-resistant highgrade steels. The requirements for grades of steel are defined in national and international standards. Properties are adapted to meet specific functions and conditions of use. Good elasticity, high tensile strength, and good corrosion resistance are required for non-cutting instruments such as ligature holders or tweezers. l Cutting instruments such as cutters and scissors mainly require a high hardness and resistance to wear and corrosion.

l Users often think the terms stainless steel or corrosionresistant high-grade steel mean the steels are physically indestructible and have high resistance to chemical corrosion. l Users are often surprised to learn that stainless steels can be susceptible to various types of mechanical, thermal, or chemical attack.

l l l l There are some metals that spontaneously form a thin layer on its surface that protects it from corrosion. Both aluminum and titanium form surface oxides to protect the interior. The formation of these protective layers is called passivation. For stainless steel, its corrosion-resistance is related to its internal structure; passivated surface layers protect the steel from corrosion. Chromium in the alloy, combined with oxygen, builds the passivation layer. However, external influences can destroy these protective layers and cause severe damage to the instruments. Therefore, it is extremely important to preserve the passivated layers on the instruments by appropriate sterilization methods.

Storage of Used Instruments in the Operatory l Staff Protection l Lay down instruments carefully, don`t throw them!

Cleaning and Disinfection There are two methods to clean and disinfect instruments: 1. Cleaning: immerse in a bath, with or without ultrasonic cleaning. 2. Clean and disinfect using an automatic washer designed for this process.

Manual Cleaning/Disinfection Choose suitable cleaning and disinfectant agents. Follow manufacturer instructions for: l correct concentration of solution l correct exposure time l correct temperature

Manual Cleaning/Disinfection l Prepare new solutions daily l Completely dissolve powder products l Open articulated instruments l Immerse instruments completely in the solution l Eliminate air bubbles in hollow instruments l Rinse efficiently with water l Dry instruments immediately

Never Use Metal Brushes l Avoid using metal brushes due to transfer of abrasives that remain on the brush to the pliers. l When two different metals are in contact in a solution, galvanic corrosion can result that can lead to corrosion.

Ultrasonic Cleaning l Fill cleaning solution to the mark indicated on the unit l Use suitable cleaning agents l Do not exceed a temperature of 40ºC l Prepare a new solution daily l Adhere to the specified frequency (35-60 khz) l When loading, cutting edges must not touch metal parts l Rinse well with water l Dry instruments immediately

Automatic Cleaning/Disinfection Units Standardized cleaning and disinfection is best achieved by automatic procedures in machines using heat to disinfect. Pliers should be stored under dry conditions before processing in automatic washers.

Automatic Cleaning/Disinfection l l l l l l Clean/disinfect pliers immediately after use Caution: danger of foaming Clean/disinfect instruments only at temperatures recommended by the manufacturer for heating, approximately 105ºC Arrange instruments correctly Do not overload trays Open pliers at joints or hinges

Check Maintenance and Care Recommendations l Check maintenance, care, and packaging recommendations before sterilization instruments should be macroscopically clean. l Check cleanliness at critical areas -grips, joints and hinges. l Important -residues remaining on pliers must be entirely removed before sterilization.

Water Stains and Discoloration Water stains and discoloration may be due to: l Errors in procedure, e.g., new instruments were not cleaned before the first time they were sterilized. l Cleaning agent, disinfectants, and maintenance products are not compatible. l Insufficient final rinsing. l Non-adherence to recommended concentrations for cleaning agents, disinfectants, and maintenance products.

Water Stains and Discoloration l Drug residues. l Poor water quality, e.g., due to instruments held for excessively long cycles in the ion exchangers. l Water used for final rinsing has not been fully desalinated. l Contaminated steam during sterilization. l Inadequate removal of sterilization indicators.

Discoloration of Pliers- Insufficient Rinsing

Lubricating Joints and Hinges Joints and hinges of pliers must be lubricated with a paraffin based maintenance product before a functional check is performed. The lubricant must be permeable to steam, suitable for sterilization, and thermally stable.

Corrosion Due to Lack of Maintenance and/or No Lubricant Used

Cool Down Functional Check Before the functional check is performed, pliers must be cool and lubricated to avoid metal wear and abrasion. Metal abrasion can result in corrosion.

Storage and Correct Packaging l Be aware of special storage of sensitive parts. l Remove defective instruments to prevent contamination of good instruments. l Use correct packaging.

Sterilization Autoclave Recommended for Schweickhardt Pliers Dry heat sterilization Cold sterilization Not Recommended for Schweickhardt Pliers

Autoclave l The sterilizer and sterilization methods must comply with current standards, e.g., German Standard DIN EN 13060 Type B (for packaged / wrapped, solid, hollow and porous products) or other country standards. l General sterilization cycles are 127ºC for 15 minutes or 132ºC for 5 minutes. l Adhere to the specified routine checks and maintenance regulations.

Autoclave l Use only fully demineralized or distilled water. l Dirty Water = Dirty Steam l Only autoclave specific items. l After sterilization, open sterilizer without vacuum drying.

Dry Heat Sterilization l When using dry heat sterilizers, the manufacturer's instructions should be strictly followed. l It is particularly important to load the instruments correctly. l Closely follow the specifications for sterilization time.

Dry Heat Sterilization l Maintain, but do not exceed 180ºC. Paraffin oil will form a resin and lose its lubricating function if temperatures exceed 180ºC. Do not use silicone oil to avoid the danger of staining. l Even if the temperature briefly exceeds180ºc, there is a risk of loss in hardness and impaired function.

Dry Heat Sterilization l There is also a risk of surface changes if the temperature exceeds180ºc. l Closely adhere to the recommended sterilization time. l Never open the unit during the sterilization cycle.

l Dry Storage of Pliers That Have Not Been Sterilized l Protect from dust l Separate from chemicals l Secure sensitive parts l Do not store in the open

Storage of Sterile Pliers l Caution: unpackaged pliers are not sterile. l For dust free and protected storage follow your country s standard. l Strictly separate sterile and non-sterile pliers.

New Instruments l Remove the transportation/shipping packaging. l Clean new instruments before sterilizing. l New Instruments must be sterilized before their first use.

Problems with Water Affect Tarnish and Corrosion of Pliers Drinking water High chloride content Stains and pitting corrosion Discoloration Foreign corrosion

Drinking Water l Drinking water always contains dissolved salts and minerals. l Excessively high concentrations of salts and minerals in drinking water can lead to staining or damage to the pliers during sterilization. l When water dries on pliers, the dissolved ingredients are deposited as salt incrustations or stains.

Stained Pliers Cause: Water with high salt content

High Chloride Content l Different concentrations of chloride salts are always found in drinking water. They cannot be eliminated by a water softening process. l In most cases, the correlation between chloride content in the water and pitting corrosion is not obvious. l When water dries, chlorides concentrate causing pitting corrosion at some sites.

Pitting Corrosion

Stains and Pitting Corrosion l Since water softening will not reduce the total content of dissolved salts or chloride, the use of fully desalinated water is recommended. l Softened water is adequate for operations such as pre-rinsing, cleaning, and intermediate rinsing.

Discoloration is Not Corrosion l Other ingredients in the water can cause brown, blue, gray, black or rainbow-colored discolorations even at low concentrations. l This discoloration can be caused by silicates; iron, copper and manganese compounds. l This discoloration can be removed using a suitable acidic cleaning agent.

Foreign Corrosion l Sometimes rust is found in drinking water, in addition to the substances found naturally in water. l Rust usually comes from corroded pipes. l Rust is deposited on pliers during sterilization causing stains and subsequent corrosion.

Foreign Corrosion

General Problems Discoloration Deposits Corrosion

Discoloration l Pliers become discolored with time. l These surface changes are always due to the sterilization procedure. l Find the cause eliminate the cause remove discoloration. l Removing discoloration from pliers without eliminating the cause, is not a permanent solution to the problem.

Discoloration Discoloration can be caused by: l Water stains when water containing dissolved salts/minerals dries on pliers. l Irregular stains on new pliers, often with a clearly visible margin. l Water stains can only be reliably avoided by using fully desalinated water for the final rinsing cycle.

Discoloration Discoloration occurs when: l Cleaning solution or disinfectant solution have not been sufficiently rinsed. l The mineral content of these solutions dries on the surface. l Is avoided by: q Improved, additional rinsing. q Reduced load on the trays during automated cleaning.

Discoloration Rainbow discoloration can be avoided by: l Using fully desalinated water in the final rinse l Using fully demineralized water for steam generation. l Using fully desalinated water for steam generation.

Deposits l Rust is avoided by: - Using fully desalinated water for generating pure steam. l Organic residues can be avoided by: - Sufficient cleaning - Changing baths regularly Do not sterilize disposable items!

Corrosion Pitting corrosion Tensile crack corrosion Fissure corrosion Frictional corrosion Contact corrosion Surface corrosion

Pitting Corrosion Pitting corrosion is only observed on metal materials. Stainless steels are not resistant to this type of corrosion. Pitting corrosion is avoided by: - Cleaning instruments immediately after use. - Using low chloride water and only fully desalinated water for the final rinsing cycle.

Pitting Corrosion

Tensile Crack Corrosion Tensile crack corrosion is usually observed on stainless steel pliers and my have negative effects on the service life of pliers. Tensile crack corrosion is avoided by: - Cleaning pliers that remain open. - Closing pliers to the first ratchet position during sterilization. (e.g., Mathieu)

Tensile Crack Corrosion

Fissure Corrosion Fissure corrosion is observed in narrow joints or hinges. Fissure corrosion can be avoided by: - Adequately drying narrow gaps in joints and hinges.

Fissure Corrosion

Corrosion Due to Friction Joints and hinges of pliers must be treated with maintenance products based on medical grade white oil / paraffin oil to keep the joints from getting blocked. Otherwise, metal wear and abrasion will form, leading to corrosion. Prevention: Regular treatment of joints and hinges with medical grade white oil.

Corrosion Due to Friction We recommended the medicine With oil Nr. 3297

Contact Corrosion When pliers are machine-cleaned, contact corrosion may occasionally be observed. Contact corrosion is avoided by: - Using desalinated water for cleaning and rinsing. - Separating stainless steel from non-stainless steel pliers.

Contact Corrosion

Surface Corrosion Surface corrosion is particularly severe if stainless steel pliers contact non-stainless steel pliers. Surface corrosion is avoided by: - Removing non-stainless steel pliers having a damaged (non-passivated) protective layer. - Not using alkaline cleaning agents in the automated cleaning of aluminum and aluminum oxide products.

Surface Corrosion

Help on the Internet www.a-k-i.org For problem solving and recommendations for maintenance of instruments, consult this website.