High Resolution Analysis of Heparin and Heparin-like Impurities on Glycomix TM an Anion Exchange Column

Similar documents
SEC Column Selection Guide for the Separation of Water-soluble Polymers

Update on Heparin Contamination: Assessment of Optimized Capillary Electrophoresis, Proton NMR and Ion-Exchange HPLC Methodologies

Determination of Oversulfated Chondroitin Sulfate and Dermatan Sulfate in Heparin Sodium Using Anion-Exchange Chromatography with UV Detection

Intact Glycoprotein Isoforms Separation on Proteomix Strong Anion Exchange (SAX) Chromatography Using Salt and ph Gradients

ION EXCHANGE KIT FOR MAB SEPARATIONS

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

A Faster Solution with Increased Resolution for Determining Chromatographic Identity and Absence of OSCS in Heparin Sodium

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

Sepax Technologies, Inc.

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Heparin Sodium - USP method

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

Faster Separations Using Agilent Weak Cation Exchange Columns

Capillary electrophoresis of heparin and related impurities using highly concentrated buffers in a 25 µm bubble cell capillary

Protein Separation with ph Gradients Using Composite Buffer Systems Calculated by the Agilent Buffer Advisor Software

Application Note. Authors. Abstract. Biopharmaceuticals

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Chromatography column for therapeutic protein analysis

ph gradient analysis of IgG1 therapeutic monoclonal antibodies using a 5 µm WCX column

Size Exclusion BioHPLC Columns

HEPARIN CALCIUM. Heparinum calcicum

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Thermo Scientific. MAbPac HIC-10. Product Manual. P/N: June Part of Thermo Fisher Scientific

Columns for Biomolecules BioLC Column Lines

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

The Agilent 1260 Infinity BioInert Quaternary Pump. Scope of a low-pressure mixing UHPLC pump with Bio-Inert Capabilities

Commentary Pharmacopeial Forum 35(5) Sept.-Oct Interim Revision Announcements to USP 33-NF 28 Reissue

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Size Exclusion BioHPLC columns Ion Exchange BioHPLC columns

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

2012 Waters Corporation 1

9/2/2014. USP Monograph Modernization. Todays topics. USP basic. Todays topics. - USP basic. - USP publications. - USP monograph modernization

Automated QbD-Based Method Development and Validation of Oxidative Degraded Atorvastatin

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

Bivalirudin Purification:

mab and ADC Analysis Shanhua Lin, Ph.D. The world leader in serving science

TSK-GEL BioAssist Series Ion Exchange Columns

Fast protein separations with ion exchange columns

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

Thermo Scientific MAbPac HIC Columns. Novel Hydrophobic Interaction HPLC Columns. Designed for Monoclonal Antibody Analysis

Analysis of Monoclonal Antibody and Related Substances using a New Hydrophobic Interaction Chromatography (HIC) Column

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

Fast, Low Pressure Analysis of Food and Beverage Additives Using a Superficially Porous Agilent Poroshell 120 EC-C18 Column

High Throughput Sub-4 Minute Separation of Antibodies using Size Exclusion Chromatography

Scale-up with the Agilent SD-1 Purification System analytical and preparative runs on a single system

Method Transfer from an Agilent 1200 Series LC to an Agilent 1260 Infinity II LC

Agilent PL-SAX Anion-Exchange Media for Nucleotide and Oligonucleotide Analysis

Mobile Phase Optimization in SEC Method Development

Optimizing Protein Separations with Agilent Weak Cation-Exchange Columns

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Guidance for Industry

Analytical Instrument Qualification According to USP <1058>: Requirements and Examples for the Agilent 1290 Infinity LC System

Fast Agilent HPLC for Large Biomolecules

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Note #1. Heparin manufacturing process (Supplementary Fig. 1) Note #4. Molecular weight determinations (Supplementary Table 1)

Monitoring of Mammalian Cell Culture Media with HILIC LC/MS

CX-1 ph Gradient Buffer

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Certificate of Analysis

4/4/2013. BioHPLC columns. Paul Dinsmoor Biocolumn Technical Specialist. April 23-25, Size Exclusion BioHPLC Columns

High Throughput Methods for the Identity Tests of Low Molecular Mass Heparins and Heparin Sodium

High-Throughput LC/MS Purification of Pharmaceutical Impurities

Performance characteristics of the 1260 Infinity Quaternary LC system

A Simple Rapid and Sensitive Method Development for Quantification of Quetiapine Fumarate in Bulk and Dosage Forms Using RP-HPLC

DNAPac PA200 and PA200 RS Columns Solutions for Nucleic Acid Analysis

Emulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture

A universal chromatography method for aggregate analysis of monoclonal antibodies

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

Separation of Monoclonal Antibodies Using TSKgel HPLC Columns

columns P r o P a c H I C C o l u m n S o l u t i o n s f o r P r o t e i n A n a l y s i s

using the method described in the European Pharmacopoeia JASCO INTERNATIONAL CO., LTD

Automated Scouting of Stationary and Mobile Phases Using the Agilent 1290 Infinity II Method Development Solution

Encompassing Method Development, Small Scale Purification, and Post-Purification Analysis with the Investigator SFC System

Analysis of biomolecules by SEC and Ion-Exchange UPLC

Fast Analysis of Environmental Phenols with Agilent Poroshell 120 EC-C18 Columns

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

BioHPLC columns. Tim Rice Biocolumn Technical Specialist

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

BioPro SmartSep. Benefits of using BioPro SmartSep. Product information

mabs and ADCs analysis by RP

Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column

High-Resolution Analysis and Purification of Oligonucleotides with the DNAPac PA-100 Column

MAbPac SCX-10 Columns

Choosing the Right Calibration for the Agilent Bio SEC-3 Column

Improved Sensitivity with the Agilent 1290 Infinity Evaporative Light Scattering Detector (ELSD)

Online LC Sample Preparation with BioTrap 500 C18

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Agilent Anion-Exchange Media for Proteins

Ion Exchange Chromatography. Learning Objectives:

Transcription:

High Resolution Analysis of and -like Impurities on Glycomix TM an Anion Exchange Column Authors Xiaomi Xu Yuna Li Aiyan Cui Haiying Chen Sepax Technologies, Inc. 5 Innovation Way Newark, DE 19711 USA Glycan Separation Abstract Analysis of heparin is challenging due to its heterogeneity, both in its size distribution and charge variance. Moreover, its impurities usually have properties that resemble heparin which makes it difficult to distinguish heparin and its impurities using analytical methods. To effectively separate heparin from its impurities, including naturally occurring impurities during manufacturing process (e.g., dermatan sulfate ) and non-native contaant from adulteration (e.g., oversulfated chondroitin sulfate ), U.S. Pharmacopeia (USP) published a chromatographic method to identify heparin and heparin-like impurities using ion-exchange liquid chromatography. However, current commercially available ion-exchange column barely meets USP separation resolution criteria; the situation calls for a new stationary phase which can improve the resolution. Sepax has recently developed an ion-exchange column Glycomix TM SAX. It grafts charged and a certain level of hydrophilic functional groups to polymer-based resin. Here we present the high efficiency separation of highly charged glycans.

Introduction has been widely used as an anticoagulant or anti-thrombotic agent. It is a complex mixture of sulfated glycosaoglycans (GAGs), which are highly negatively charged. During its extraction process from mammalian tissue (such as pig intestine), other polyanionic GAGs can co-purify with heparin since they share heparin-like properties [1]. One representative example of such impurities is dermatan sulfate (chondroitin sulfate B). Oversulfated chondroitin sulfate () was identified as a non-native contaant, which can induce severe side effects even death. The most recent heparin sodium monograph published by U.S. Pharmacopeia [2] details the analytical method that heparin manufacturers must follow to ensure the quality of their heparin products. This application note reports the separation of heparin, and using Glycomix TM SAX with USP recommended method. The performance of Glycomix SAX is also discussed in comparison to competitor D s column separation efficiency. Experimental HPLC system Agilent 12 HPLC with binary pump Ion-exchange columns Glycomix TM -SAX (4.6 x 25 mm, P/N 91665-4625) Glycomix TM -SAX guard column (4.6 x 5 mm, P/N 91665-465) Chemicals and Reagents sodium salt and Dermatan sulfate ( or chondroitin sulfate B) were purchased from Sigma-Aldrich. Oversulfated chondroitin sulfate (part no. 113358) was purchased from U.S. Pharmacopeia. Monobasic sodium phosphate and sodium perchlorate were purchased from EMD. All solutions were made with Milli-Q water. LC Method Mobile phase: A:.4% NaH 2 PO 4, ph3.; B:.4% NaH 2 PO 4 + 14% NaClO 4, ph3. Gradient: 2% B to 9% B in 6 Flow rate:.22 ml/ Column temperature: 25 C Detection: UV 22 nm Injection volume: 1 μl Pressure: 1 bar Results Following the conditions listed on USP heparin sodium monograph, Figure. 1 demonstrates the separation of heparin from its impurities, dermatan sulfate () and oversulfated chondroitin sulfate (). The three peaks, according to elution order, are identified as, heparin sodium salt and, the concentration of which are.2 mg/ml, 2 mg/ml and.2 mg/ml, respectively. The resolutions are 3.8 between and heparin and 5.8 between heparin and ; both are well beyond the USP resolution requirements, which are 1. and 1.5, respectively. Glycomix TM column shows the separation resolution much higher than that of Competitor D column, which is 1.1 and 1.8 [3]. It was also stated in competitor D s literature [3] that dermatan sulfate may be unresolved from heparin at >2 mg/ml concentrations as the heparin peak starts eluting from the column within 2 after the dermatan peak. loading is not a concern on Glycomix TM column. The high resolution between heparin and dermatan sulfate also enables identification of any unknown impurities that elutes between heparin and dermatan sulfate. As shown in Figure. 2, there is an impurity peak eluted after dermatan sulfate, which is marked by an arrow on the graph. Comparing it with each single standard, it is found that this unknown impurity peak, which may be chondroitin sulfate, is also present on heparin sodium salt (Sigma-Aldrich) sample. Figure 3 is another example of a real heparin injection product analysis, which shows the presence of. Glycomix can be used in the quantitative analysis of heparin and its impurities. Figure 4 shows the calibration curve with different heparin and impurity loadings over concentration range from.5 mg/ml to 1 mg/ml for heparin and from.25 mg/ml to 4 mg/ml for. The R 2 values exceed.99 for both analytes (R 2 heparin=.9997; R 2 =.9972). 2

mau 25 2 15 1 5 5 1 15 2 25 3 35 4 Figure 1. Elution profile of standard solution, containing.2 mg/ml dermatan sulfate (), 2 mg/ml heparin sodium and.2 mg/ml oversulfated chondroitin sulfate (). mau 4 35 3 25 2 15 1 5 5 1 15 2 25 3 35 4 Figure 2. Comparison of standard solution (1 mg/ml, 2 mg/ml heparin and 1 mg/ml ) and one heparin injection product provided by an undisclosed drug company. injection Standard solution 5 1 15 2 25 3 35 4 Figure 3. Comparison of standard solution (.2 mg/ml, 2 mg/ml heparin and.2 mg/ml ) and another heparin injection product provided by an undisclosed drug company, a different source than that on Figure 2. 3

Figure 4. calibration curve over the concentration range from.5 mg/ml to 1 mg/ml, from.25 mg/ml to 4 mg/ml. Conclusion It has been demonstrated that with careful control of surface chemistry, high resolution separation of heparin and its impurities is successfully achieved on Sepax Glycomix TM column. Glycomix TM column is well suitable for quality control for heparin products, as well as quantitative analysis due to its high sensitivity. Reference 1. Advances in the separation, sensitive detection, and characterization of heparin and heparin sulfate. Anal Bioanal Chem (29) 393: 155-169 2. Sodium, Pharmacopeia Forum 29, 35 (5), 1-4. 3. Deteration of oversulfated chondroitin sulfate and dermatan sulfate in heparin sodium using anion-exchange chromatography with UV detection. Dionex Application Note 235. Ordering Information Product Description Part number Glycomix SAX 4.6 x 25 mm 91665-4625 Glycomix SAX guard column 4.6 x 5 mm 91665-465 Glycomix SAX Kit Column + Guard 91665-KIT Sodium 5 mg HP-5 Chondroitin Sulfate B 5 mg CS-5 Oversulfated Chondroitin Sulfate 5 mg -5 Glycomix SAX Custom size Inquire For more information on Glycomix TM products, please visit Sepax website: http://www.sepaxtech.com or contact us at 1-877-SEPAX-US 4

Analysis on SEC SF11 Column: SRT SEC- 3 (7.8x3mm), SRT SEC- 5 (7.8x3mm) and SRT SEC- 3 guard (4.6x5mm), Column temperature: 3 C, Flow rate:.6 ml/, Detection: RI, Mobile phase:.1 M ammonium acetate,.2% sodium azide in water, Injection Volume: 2 µl of 5 mg/ml nriu 5 4 3 2 1-1 1 2 3 4 5 www.sepax-tech.com Better Surface Chemistry for Better Separation Sepax Technologies, Inc.