Choosing the Right Calibration for the Agilent Bio SEC-3 Column

Similar documents
Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Multiple Detector Approaches to Protein Aggregation by SEC

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Size exclusion chromatography for Biomolecule analysis: A "How to" Guide

ProSEC 300S. Protein Characterization columns

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Selecting the Correct Agilent PLgel Column for Polyolefin Analysis with GPC/SEC

Aggregate/Fragment Analysis

Faster Separations Using Agilent Weak Cation Exchange Columns

Fast Agilent HPLC for Large Biomolecules

Mobile Phase Optimization in SEC Method Development

Size Exclusion BioHPLC Columns

SEC Column Selection Guide for the Separation of Water-soluble Polymers

Understanding the Effects of Common Mobile Phase Additives on the Performance of Size Exclusion Chromatography

BioHPLC columns. Tim Rice Biocolumn Technical Specialist

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

Characterization of IgG monomers & their aggregates

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

Optimizing Protein Separations with Agilent Weak Cation-Exchange Columns

4/4/2013. BioHPLC columns. Paul Dinsmoor Biocolumn Technical Specialist. April 23-25, Size Exclusion BioHPLC Columns

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

Size Exclusion Chromatography

2012 Waters Corporation 1

Validation of a Dual Wavelength Size Exclusion HPLC Method with Improved Sensitivity to Detect Aggregates of a Monoclonal Antibody Biotherapeutic

Size Exclusion BioHPLC columns Ion Exchange BioHPLC columns

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Molecular Characterization of Biotherapeutics The Agilent 1260 Infi nity Multi-Detector Bio-SEC Solution with Advanced Light Scattering Detection

Application Note. Authors. Abstract. Biopharmaceuticals

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

TSK-GEL BioAssist Series Ion Exchange Columns

Reproducible LC/MS Peptide Separation Using Agilent AdvanceBio Peptide Plus Columns

Agilent PL-SAX Anion-Exchange Media for Nucleotide and Oligonucleotide Analysis

Higher Order mab Aggregate Analysis using New Innovative SEC Technology

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS

BioBasic SEC Columns TG Silica-based size exclusion columns for superior resolution and efficiency

Optimization of the Effective Separations for Peptides and Proteins Using High Durable Packing Materials for HPLC. YMC CO., LTD.

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

Protein Separation with ph Gradients Using Composite Buffer Systems Calculated by the Agilent Buffer Advisor Software

High Throughput Sub-4 Minute Separation of Antibodies using Size Exclusion Chromatography

Agilent OpenLAB CDS MatchCompare for the Comparison of Peptide Mapping Samples

Analysis of biomolecules by SEC and Ion-Exchange UPLC

CHARACTERIZATION & PURIFICATION

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

The New Standard in High Resolution Size Exclusion

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

Agilent Anion-Exchange Media for Proteins

EAG.COM MATERIALS SCIENCES APPLICATION NOTE. By J. Tompkins 1, T. Spurgeon 1, R. Tobias 1, J. Anders 1, E. Butler-Roberts 2, and M.

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

TSKgel STAT Columns for High Performance Ion Exchange Chromatography

A universal chromatography method for aggregate analysis of monoclonal antibodies

Bio-Monolith Protein G Column - More Options for mab Titer Determination

Superdex 200 Increase columns

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

Superdex 200 Increase columns

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

TSKgel G2000SWXL Columns for the Reproducible Analysis of Bovine Serum Albumin

Application Note. Authors. Abstract. Introduction. Pharmaceutical

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Lifetime stability of size exclusion chromatography columns for protein aggregate analysis

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

NISTmAb characterization with a high-performance RP chromatography column

Data File. HiLoad Superdex 30 prep grade HiLoad Superdex 75 prep grade HiLoad Superdex 200 prep grade. Pre-packed gel filtration columns

Protein Quantitation using various Modes of High Performance Liquid Chromatography

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs)

EPA 8330A Analysis of Explosives Using Agilent Poroshell 120 EC-CN and EC-C18 Columns

Thermo Scientific MAbPac HIC Columns. Novel Hydrophobic Interaction HPLC Columns. Designed for Monoclonal Antibody Analysis

Reversed-Phase HPLC Columns

Columns for Biomolecules BioLC Column Lines

TOYOPEARL GigaCap Series

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

Superdex 200 HR 10/30 INSTRUCTIONS

Chromatography column for therapeutic protein analysis

Zwitterion Chromatography ZIC

Novel Wide-Pore Superficially Porous Particles for Biomacromolecular Separations

Size Exclusion. Tools For Polymer Characterization. Gel Permeation / Size Exclusion / Gel Filtration - GPC / SEC / GFC SIZE EXCLUSION CHROMATOGRAPHY

Agilent 1290 Infinity II 2D-LC Solution Biopharmaceutical Polymer Analysis. WCBP Jan 2017 Washington, DC

Biomolecule Purification. Purification columns and media for peptides, oligonucleotides, and proteins

Superose 6 HR 10/30 Superose 12 HR 10/30

CX-1 ph Gradient Buffer

PRINCIPLES OF CHROMATOGRAPHY SEC COLUMNS

Discovery BIO Wide Pore

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis

Sepax Technologies, Inc.

Agilent Technologies April 20,

Performance characteristics of the 1260 Infinity Quaternary LC system

Sean M. McCarthy and Martin Gilar Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL RESULTS AND DISCUSSION

Characterize mab Charge Variants by Cation-Exchange Chromatography

Agilent Prep LC Columns for Small Molecules and Biomolecules MAINTAIN RAPID, RELIABLE SEPARATIONS AS YOU SCALE-UP

'Tips and Tricks' for Biopharmaceutical Characterization using SEC

Userguide: SepFast Gel Filtration columns

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Transcription:

Choosing the Right Calibration for the Agilent Bio SEC-3 Column Application Note Biotherapeutics & Biosimilars Author Keeley Mapp Agilent Technologies, Inc Introduction Size exclusion chromatography (SEC) is a technique for separating molecules by size using aqueous eluents. This typically includes molecules such as proteins, oligonucleotides, and other complex biopolymers. For molecules of discreet molecular weight, such as proteins, SEC can be used to detect and quantitate monomers, dimers, aggregates, and fragments. The Agilent Bio SEC-3 300Å column is ideally suited to the analysis of globular proteins. Calibration choice One of the factors that influences column choice in SEC is the column s resolving range. This can be identified by running a calibration, using standards appropriate to the analyte type, molecular weight range, and size in solution. Thus, for biopolymers such as polysaccharides, we used pullulan polysaccharide calibrants, for proteins we used protein standards, and for synthetic polydispersed polymers we used polyethylene glycol and polyethylene oxide standards.

Conditions Column: Agilent Bio SEC-3, 300Å, 4.6 300 mm, 3 µm (p/n 5190-2513) Samples: Agilent pullulan polysaccharide standards kit SAC-10 (p/n PL2090-0) Agilent polyethylene glycol standards kit PEG-10 (p/n PL2070-0) Agilent polyethylene oxide standards kit PEO-10 (p/n PL2080-0101) Protein standards (Sigma-Aldrich Corp.) Eluent: 150 mm Sodium phosphate buffer, ph 7 Flow rate: 0.35 ml/min Detector: RI for pullulan polysaccharides, PEGs and PEOs UV, 220 nm for proteins System: Agilent 1260 Infinity LC Results and Discussion Pullulan polysaccharide calibration Figure 1 shows a typical calibration curve using pullulan polysaccharide standards on a Bio SEC-3, 3 µm column. The pullulan polysaccharide calibrant kit contains mono-, di-, tri-, and tetra-saccharides to define the bottom end of the calibration. The pullulan polysaccharides are well characterized with a narrow polydispersity. These are linear macromolecules of maltotriose units. The molecular weight range of the kit is 180, glucose to 788,000. The suggested resolving range, the shallowest part of the calibration curve for this type of biopolymer, is to 120,000 molecular weight. Table1 shows the pullulan polysaccharide calibration data. Log MW,000,000 4 5 6 7 8 9 10 11 12 Figure 1. Pullulan polysaccharide calibration curve using an Agilent Bio SEC-3, 300Å column. Table1. Pullulan polysaccharide calibration data for an Agilent Bio SEC-3, 300Å column. Pullulan or polysaccharide Nominal Mp (g/mol) 788 K 788,000 5.55 380 K 380,000 5.66 K,000 6.67 48 K 48,000 7.77 23.7 K 23,700 8.97 5.8 K 5,800 10.42 Maltotriose 504 11.46 Maltose 360 11.51 Glucose 180 11.61 Retention time (min) Protein calibration The molecular weight range used for protein calibration was 931 to 670,000. Figure 2 shows the resulting plot achieved using the Bio SEC-3, 300Å column. In solution, low molecular weight proteins can deviate from linear elution behavior. This is due to the characteristics of proteins in solution and the type of structure they form, for example, how they are folded. Therefore, the elution is influenced by their actual size in solution and not just the molecular weight. 2

The suggested protein molecular weight resolving range for the column is approximately 5,000 to 1,250,000 molecular weight. Table 2 shows the protein calibration data. Log MW,000,000 4 6 8 10 12 14 Figure 2. Protein calibration curve using an Agilent Bio SEC-3, 300Å column. Table 2. Protein calibration data using an Agilent Bio SEC-3, 300Å column. Protein or peptide Nominal Mp (g/mol) Thyroglobulin 670,000 6.66 g-globulin 150,000 8.31 BSA 66,000 9.03 Ovalbumin 44,300 9.50 B-lactoglobulin 36,800 9.77 Myoglobin 17,600 10.15 Cytochrome C 12,327 10.33 Insulin B chain 3,496 11.43 Neurotensin 1,672 12.09 Vitamin B12 1,355 12.22 Angiotensin II 1,040 12.28 Retention time (min) Polyethylene glycol and polyethylene oxide calibration Figure 3 shows the Bio SEC-3, 300Å column calibration curve obtained when calibrated with PEG/PEO calibration standards. The glycols cover the lower molecular weight range and the oxides the higher MW range. The two types are chemically similar, differing only in the end group. The PEGs have a dihydroxyl end group and the PEOs have a single hydroxyl end group. If they are used together, they can span a very wide molecular weight range. In this instance, the resolving range is 1,500 to 1,250,000 molecular weight. Table 3 shows the polyethylene glycol and polyethylene oxide calibration data. Log MW,000,000,000 10 1 4 5 6 7 8 9 10 11 12 Figure 3. Polyethylene glycol and polyethylene oxide calibration curve using an Agilent Bio SEC-3, 300Å column. Table 3. Polyethylene glycol and polyethylene oxide calibration data using an Agilent Bio SEC-3, 300Å column. PEG or PEO Nominal Mp (g/mol) Retention time (min) > 5M 5,000,000 5.42 1,522,000 1,522,000 5.40 905,000 905,000 5.44 692,000 692,000 5.51 454,000 454,000 5.58 305,500 305,500 5.64 135,800 135,800 5.94 77,350 77,350 6.50 46,470 46,470 7.24 21,300 21,300 8.56 12,140 12,140 9.42 8,730 8,730 9.89 3,870 3,870 10.68 1,480 1,480 11.46 420 420 11.48 106 106 11.47 With proteins, it is important to recognize that the SEC mechanism works by separating solutes dependant on their size in solution and not their molecular weight. This is evident when comparing the calibration plot of the proteins/peptides with the pullulan/polysaccharide and PEG/PEO curves, as shown in Figure 4. The pullulan/polysaccharide and PEG/PEO calibrants provided quite similar calibration curves but the protein/peptide curve was shifted and a different shape. Proteins are composed of complex polypeptide chains that form three-dimensional structures. These structures are affected by the environment to which they are exposed, for example ph or ionic strength. The chains will form the shape that is most suited to them and so the structure and size may vary. 3

Log MW,000,000,000 Peptides and proteins PEGs and PEOs Pullulan and polysaccharides 10 4 6 8 10 12 Figure 4. Comparison of calibration plots generated for all three types of calibrant. To demonstrate that elution time is due to size rather than molecular weight, consider the retention times for calibrants with a molecular weight of approximately 50,000, in which there is a significant difference (Figure 5). The PEG elutes just after 7 minutes, the polysaccharide elutes at just over 7.5 minutes, but the protein elutes at approximately 9.5 minutes. This clearly demonstrates that the SEC separation mechanism is based on the actual size and not molecular weight. Therefore, when using calibration curves it is important to specify what calibrants have been used. For example, it can be stated that the sample of interest has a pullulan/polysaccharide equivalent molecular weight of 50,000. mau 4,000 PEG 46,702 (nominal MV) Polysaccharide 48,000 (nominal MV) Protein, ovalbumin 44,300 (nominal MV) 3,000 2,000 0 2 4 6 8 10 min Figure 5. Overlay of chromatograms obtained for calibrants of similar molecular weight. 4

Conclusions Calibrating Agilent Bio SEC columns to identify their resolving range is straightforward, provided the calibrant is matched to the analyte. The plots shown in this application note demonstrate that both pullulan/polysaccharide and PEG/PEO standards can be used with the Agilent Bio SEC-3 300Å columns, and that calibration plots are similar and simple to interpret. Therefore, for maximum resolution in SEC, it is important to work within the resolving range of the column and where the calibration curve is less steep. The linear section of the plots identifies the molecular size range for which optimum resolution can be achieved by using this Bio SEC-3 300Å column. Agilent manufactures a range of polymer standards that are ideal reference materials for generating accurate and reliable SEC column calibrations. The calibration curves presented here relate to the Agilent Bio SEC-3 300Å columns but the same calibrants can be used to calibrate the full range of Agilent Bio SEC columns, which includes Å and 150Å pore sizes, and 5 µm particle size columns in 4.6 300 mm and 7.8 300 mm dimensions. For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. 5

www.agilent.com/chem For Research Use Only. Not for use in diagnostic procedures. Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2013, 2017 Printed in the USA November 2, 2017 5991-2463EN