Photoacoustic assessment of oxygen saturation: effect of red blood cell aggregation

Similar documents
Photoacoustic Imaging in Biomedicine Critical Review by Saurabh Vyas Group 9: Interventional Photoacoustic Ultrasound CIS II: 600.

Photoacoustic imaging of vascular networks in transgenic mice

Basic principles of quantification using optical techniques

In vivo label-free photoacoustic flow cytography and on-the-spot laser killing of single circulating. melanoma cells

PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

PREPARED FOR: U.S. Army Medical Research and Materiel Command Fort Detrick, Maryland

Miniature fibre optic probe for minimally invasive photoacoustic sensing

IN VIVO BLOOD OXYGENATION LEVEL MEASUREMENTS USING PHOTOACOUSTIC MICROSCOPY. A Thesis MATHANGI SIVARAMAKRISHNAN

Realtime Photoacoustic Microscopy of Murine Cardiovascular Dynamics

Optoacoustic System for 3D Functional and Molecular Imaging in Nude Mice

1st Faculty of Medicine, Charles University in Prague Center for Advanced Preclinical Imaging (CAPI)

Imaging acute thermal burns by photoacoustic microscopy

Classification of biological cells using a sound wave based flow cytometer

Biophotonics?? Biophotonics. technology in biomedical engineering. Advantages of the lightwave

REVOLUTIONIZING EARLY CANCER DETECTION

Experts in Femtosecond Laser Technology. DermaInspect. Non-invasive multiphoton tomography of human skin

Enhanced Sensitivity Carbon Nanotubes as Targeted Photoacoustic Molecular Imaging Agents

Biomedical Applications of Molecular Spectroscopy

SUPPLEMENTAL INFORMATION. Virus-mimicking nano-constructs as a contrast agent for near infrared photoacoustic imaging

FUNCTIONAL PHOTOACOUSTIC MICROSCOPY. A Dissertation HAO ZHANG

SIMULTANEOUS IMAGING OF A lacz-marked TUMOR AND MICROVASCULATURE MORPHOLOGY IN VIVO BY DUAL-WAVELENGTH PHOTOACOUSTIC MICROSCOPY

Toward High-speed Transcranial Photoacoustic Imaging using Compact Near-infrared Pulsed LED Illumination System

Three-dimensional combined photoacoustic and optical coherence microscopy for in vivo microcirculation studies

TITLE PAGE. Citation Format: Copyright notice: DOI abstract link:

Absorption of an electromagnetic wave

In situ semi-quantitative assessment of single cell viability by resonance

Measurement of cell aggregation characteristics by analysis of laser-backscattering in a microfluidic rheometry

Time-resolved optical spectroscopy and imaging of breast

7/16/2009, LV WANG. 1 cm

Laser treatment of gravure-printed ITO films on PET

DETECTION OF LASER ULTRASONIC SURFACE DISPLACEMENT BY WIDE APERTURE FIBER OPTIC AMPLIFIER M.L. Rizzi and F. Corbani CESI, Milano, Italy

Light dosimetry for Low-Level Laser therapy: Accounting for differences in tissue and depth

Translational Photoacoustic Tomography

Effects of Laser Peening Parameters. on Plastic Deformation in Stainless Steel

Supporting Information


In-vitro study on the hemorheological characteristics of chicken blood in microcirculation

University of Michigan

Deliverable 2.1: Definition of paradigms representing exemplary breast lesions cases

Transient and selective suppression of neural activity with infrared light

PATTERNING OF OXIDE THIN FILMS BY UV-LASER ABLATION

Nayar Prize I Quarterly Progress Report (Quarters 2&3) August, 2016

CENTER FOR BRAIN EXPERIMENT

Principle and technique of NIRS-Imaging for human brain FORCE: fast-oxygen response in capillary event

Multiplexed 3D FRET imaging in deep tissue of live embryos Ming Zhao, Xiaoyang Wan, Yu Li, Weibin Zhou and Leilei Peng

Noninvasive label-free imaging of microhemodynamics by optical-resolution photoacoustic microscopy

In-situ laser-induced contamination monitoring using long-distance microscopy

EFFECT OF LASER RADIATION ON ELECTRICAL CONDUCTIVITY OF HUMAN BLOOD

Spectroscopy and Imaging IV

Fs- Using Ultrafast Lasers to Add New Functionality to Glass

Femtosecond micromachining in polymers

Supplementary Information

Qswitched lasers are gaining more interest because of their ability for various applications in remote sensing, environmental monitoring, micro

ABSTRACT 1. INTRODUCTION

Evaluation of Ultrasonic Attenuation and Estimation of Ultrasonic Grain Noise in Copper

SUPPLEMENTARY INFORMATION

Investigation of the influence of laser irradiation on the relative concentration of oxyhemoglobin in the blood

Biomedical Optical Imaging Martin Frenz Biomedical Photonics Department Institute of Applied Physics, University of Bern

Skin Imaging NTU

Deeply penetrating in vivo photoacoustic imaging using a clinical ultrasound array system

HHS Public Access Author manuscript Annu Rev Biomed Eng. Author manuscript; available in PMC 2014 August 11.

Translational Multimodality Optical Imaging

Laser- and Light-Induced Autofluorescence Spectroscopy of Human Skin in Dependence on Excitation Wavelengths

THE USE OF A DIODE LASER IN THE TREATMENT OF HEMORRHOIDAL DISEASE

Submicron-resolution Photoacoustic Microscopy of Endogenous Light-absorbing Biomolecules

micromachines ISSN X

Compact Plasmonic Blackbody for Cancer Theranosis in Near-Infrared II Window

Structural changes of austenitic steel obtained by 532 nm and 1064 nm Nd:YAG laser radiation

Enhanced Light Trapping in Periodic Aluminum Nanorod Arrays as Cavity Resonator

Contact Details. Dr Alexander Galkin. Office: MBC Room 186. Tel: (028) Frequency and wavelength.

Citation for published version (APA): Bremmer, R. H. (2011). Non-contact spectroscopic age determination of bloodstains

HHS Public Access Author manuscript Phys Rev Lett. Author manuscript; available in PMC 2014 March 18.

STRUCTURAL DIAGNOSTICS OF GEOMATERIALS BY LASER ULTRASONIC SPECTROSCOPY

F* techniques: FRAP, FLIP, FRET, FLIM,

SLS-process monitoring and temperature control

This document is downloaded from DR-NTU, Nanyang Technological University Library, Singapore.

Ultrasound-Aided High-Resolution Biophotonic Tomography

Introduction. (b) (a)

Study of SOFC Operational Behavior by Applying Diagnostic Methods

Examples of the application of light-tissue interaction to biomedical engineering

Widefield Microscopy Bleed-Through

HHS Public Access Author manuscript Nat Biomed Eng. Author manuscript; available in PMC 2018 January 12.

Dielectric II-VI and IV-VI Metal Chalcogenide Thin Films in Hollow Glass Waveguides (HGWs) for Infrared Spectroscopy and Laser Delivery

Principles of translational medicine: imaging, biomarker imaging, theranostics

Genetically targeted all-optical electrophysiology with a transgenic Credependent

BME101 Introduction to Biomedical Engineering Medical Imaging Özlem BİRGÜL Ankara University Department of Biomedical Engineering

Interventional multi-spectral photoacoustic imaging in laparoscopic surgery

Michael T. Anderson, Stephen E. Cumblidge, Steven R. Doctor Pacific Northwest National Laboratory, Richland, WA, USA

Radio-opaque isotropic fibre optic probes for in vivo photodynamic therapy dosimetry

HELICA BIOSYSTEMS, INC. HIGH SENSITIVITY HUMAN C-REACTIVE PROTEIN FOR RESEARCH USE ONLY (Not for in vitro diagnostic use)

Fabrication of Micro and Nano Structures in Glass using Ultrafast Lasers

Assessment of Blood Trauma in Circulatory-assist Devices. Sangria Project. Amanda Daly. Advisor: Dr. Marina Kameneva

Optimization of Water based Optical Filter for Concentrated Crystalline Si PV/T System - A Theoretical Approach

Supporting Information: Gold nanorod plasmonic upconversion microlaser

Electronic Supplementary Information

The quenching effect in PRESAGE dosimetry of proton beams: Is an empirical correction feasible?

LOW TEMPERATURE GROWTH OF SMOOTH INDIUM TIN OXIDE FILMS BY ULTRAVIOLET ASSISTED PULSED LASER DEPOSITION

Formation of Droplets on Thin Film Surface in Pulsed Laser Deposition Using Metal Targets*

Photothermal Optical Coherence Tomography of Nanoparticle Contrast Agents

Transcription:

Photoacoustic assessment of oxygen saturation: effect of red blood cell aggregation Eno Hysi a, Ratan K Saha b and Michael C Kolios a a Department of Physics, Ryerson University, Toronto, Canada b Applied Material Science Division, Saha Institute of Nuclear Physics, Kolkata, India; mkolios@ryersonca ABSTRACT The simultaneous photoacoustic assessment of oxygen saturation and red blood cell aggregation is presented Aggregation was induced on porcine red blood cells using Dextran-70 at multiple hematocrit levels Samples were exposed to 750 nm and 1064 nm for each hematocrit and aggregate size in order to compute the oxygen saturation As the size of the aggregate increased, the photoacoustic signal amplitude increased monotonically The same trend was observed for increasing hematocrit at each aggregation level The oxygen saturation of aggregated samples was 30% higher than non-aggregated samples at each hematocrit level This suggests that the presence of red blood cell aggregates impairs the release of oxygen to the surrounding environment Such a result has important implications for detecting red blood cell aggregation non-invasively and making clinical decisions based on the simulatenous assessment of oxygen saturation Keywords: Photoacoustics, red blood cell aggregation, oxygen saturation 1 INTRODUCTION Estimating a patient s need for oxygen is an important assessment of the patient s vital signs as life cannot thrive in the absence of oxygen Monitoring the oxygen saturation (SO 2 ) of blood is a routine method for assessing clinical conditions where oxygenation levels fluctuate The technique is useful in many settings where the patient s oxygenation is unstable (ie in intensive care, operating rooms, and emergency or other hospital wards) for determining the effectiveness or need for supplemental oxygen 1 Typically derived from the hemoglobin concentration, SO 2 is crucially important in mapping brain hemodynamics in various applications such as during responses to sensory simulations 2 In addition, monitoring of SO 2 has been used to evaluate the treatment of tumors with chemotherapy and radiation therapy 3, monitor the healing of wounds 4 and even study gene expression 5 Non-invasive monitoring of SO 2 is commonly achieved through the near-infrared spectroscopy (NIRS) method 6 NIRS relies on 2-wavelength emissions, typically through translucent body parts (ie finger tip or earlobes) At the wavelengths typically used (660 nm and 910 nm), the absorption of oxyhemoglobin (OHb) and deoxyhemoglobin (DHb) differs significantly Therefore, the OHb/DHb ratio can be calculated from the ratio of the absorbance at each respective wavelength However, a major limitation of NIRS is the small penetration depth due to significant light scattering in tissue In addition, NIRS is unable to distinguish between venous and arterial blood due to poor spatial resolution 7 Furthermore, this technique does not provide a complete measure of circulatory sufficiency It can provide false SO 2 readings based on the blood that arrives at the instant the measurement is taken despite the fact tissue can be hypoxic or hemoglobin insufficient (ie anemic) 8 Photoacoustic (PA) imaging can potentially overcome some of these limitations of NIRS in monitoring SO 2 The primary advantage of PA imaging lies in the fact that the scattering of sound waves detected by passive ultrasonic transducers is 2-3 orders of magnitude smaller than the scattering of light waves thus enabling the probing of deeper structures due to lower attenuation 9 By using multiple wavelengths of illumination, PA imaging has been able to not Photons Plus Ultrasound: Imaging and Sensing 2013, edited by Alexander A Oraevsky, Lihong V Wang, Proc of SPIE Vol 8581, 85813T 2013 SPIE CCC code: 1605-7422/13/$18 doi: 101117/122004960 Proc of SPIE Vol 8581 85813T-1 Downloaded From: http://spiedigitallibraryorg/ on 03/10/2013 Terms of Use: http://spiedlorg/terms

only measuree SO 2 for in-vivo samples but also simultaneously form high resolution images using the absorption of the oxygen-carrying cells in circulation, red blood cells (RBCs) 10 12 Our group has recently shown that the aggregation of RBCs is a phenomenon which affects the PA signals 13,14 The aggregation of RBCs results in the formation of clusters of RBCs of different sizess which can increase the viscosity of blood and impair the circulation It is observed in a variety of disorders such as myocardial infarctions, bacterial infections, type 2 diabetes and sickle cell disease 15 Theoretical simulations showed that as the size of the aggregates increased, the dominant frequency peak of the PA signal power spectrum decreased, approaching the diagnostic frequency range (<30 MHz) This could be attributed to the fact that an aggregatee of RBCs behaves as a single entity rather than individual RBCs 13 Experimentally, we observed that the strength of the PA signal and spectral features of the power spectrum were dependent on the size of the RBC aggregatess and that the signal changes correlated well with independent measurements of aggregation 14 It has been shown that the formation of RBC aggregates contributess to a thickening of the plasma layer around blood aggregates leading to a significant reduction in the diffusion of oxygen in surrounding tissues 16,17 Since PA techniques are capable of measuring the SO 2 of a blood sample in addition to detecting the presence of aggregation, we propose using PA methods for assessing the effect of RBC aggregation on SO 2 This simultaneous detection could potentially provide a non-invasive method for assessing RBC aggregation and thee effect that it has on tissue oxygenation 2 METHODS 21 Sample preparation The guidelines on handling blood weree adapted from the recommendation of the International society for Clinical Hemorheology and the European Society for Clinical Hemorheology y and Microcirculation 18 The source of blood was the femoral vein of Yorkshire pigs (Comparative Research, Toronto,, Canada) Thee blood was drawn into spray-coated EDTA vacutaners (Becton, Dickinson and Company, Franklin Lakes, New Jersey) in order to avoid coagulation In order to separate the RBCs from the plasma, the blood was centrifuged (1500 g for 6 minutes at room temperature) and then washed twice with phosphate-buffered saline (PBS) The collected, packed RBCs were then suspended in PBS in order to achieve three hematocritt levels (10%, 20% and 40%) ) Aggregation was induced using Dextran-70 by changing the concentration of (Sigma-Aldrich, St Louis, Missouri) dissolvedd in PBS Varying degrees of aggregationn were achieved Dextran in PBS ([Dex-PBS]) In this study, the [Dex-PBS] was 1%,, 3% and 8% for each hematocrit level The blood was kept in full contact with air prior to experimentation and all experiments weree completed within 4 hours of blood collection The presence of aggregation was assessed through light microscopy (Olympus Canada Inc, Markham, Canada) Representative images of non-aggregated and aggregated RBCs are shown in figure 1 :: ; :;' T' : : Figure 1: (a) Non-aggregated RBC suspended in PBS and (b) aggregated RBCs suspended in 3% [Dex- PBS] The scale bar denotes 10 µm Proc of SPIE Vol 8581 85813T-2 Downloaded From: http://spiedigitallibraryorg/ on 03/10/2013 Terms of Use: http://spiedlorg/terms

22 PA measurements The PA measurements were conducted using the Imagio Small Animal PA Imaging Device (Seno Medical Instruments Inc, San Antonio, Texas) shown in figure 2 Laser So urce Trans& ucer San iple CCD Camera Wa'ter Figure 2: The Imagio Small Animal PA Imaging Device Thee blood samplee was placed in a cylindrical tube that was placed between the laser (left) and ultrasound transducer (right) The system consisted c of a Q-switched, pulsed Nd:YAG laser deliveredd through an articulated arm coaxial to a transducer 60 mm apartt allowing raster scanning of the sample The laser beam diameter was 9 mm, the pulse width was 6 ns, the pulse repetition rate was 10 Hz and the maximum fluence was 25 mj/cm 2 /pulse Each sample was irradiated to 750 nm and 1064 nmm lasers The PA signals were collected through a 4 element annular array transducer with 5 MHz center frequency and 60% -6 db bandwidth The blood samples were loaded into cylindrical tubes (diameter 05 mm) at the focus of the transducer (295 mm) A vertical raster scan was performed for each sample During each scan, the laser emitted 4 pulses at each location and the 4 PA signals recorded weree averaged to produce 1 PA signal used for analysis A total of 80 laser pulses were fired for each sample thus corresponding to 20 PA signals at different locationss inside each tube (05 mm apart over a length of 10 cm) 23 Signal analysis and estimation of SO 2 The PA signal amplitude (SA) of each recorded signal was computedd as the integration of the Hilbert Transform of the PA signal The average SA and standard deviation was computed by averaging thee SA from all 20 signals recorded for each hematocrit and [Dex-PBS] The normality of the data was confirmed using a Shapiro-Wilk test (normality criterion W > 005) An unpaired t-test was used to compare the SA from non-aggregated and aggregated RBCs Statistical significance was w established for p-values of 005 or less The SO 2 for each exposed sample was computed by taking into account the fact that the SA is directly proportional to the concentrationn of OHb and DHb 12 The SO 2 can be computed by exposing each sample at 2 optical wavelengths ( λ 1 = 750nm and λ 2 = 1064nm ): Proc of SPIE Vol 8581 85813T-3 Downloaded From: http://spiedigitallibraryorg/ on 03/10/2013 Terms of Use: http://spiedlorg/terms

SO 2 [ OHb] SA( λ2) ε( DHb, λ1) SA( λ1) ε( DHb, λ2) = = [ OHb] + [ DHb] SA( λ ) Δε( λ ) SA( λ ) Δε( λ ) 1 2 2 1 (1) Here, Δ ε ( λ) = ε( OHb, λ) ε( DHb, λ) is the difference in extinction coefficient ε for each wavelength of illumination λ 3 RESULTS AND DISCUSSION Figure 3 contains the summary of the PA SA of each RBC sample exposed to both illumination wavelengths The SA is plotted as a function of both hematocrit and aggregation level (determined by the [Dex-PBS]) For both exposure wavelengths, the SA increased monotonically with increasing hematocrit (~13x per doubling hematocrit level) This is in accordance to previous reports which suggest that the increase is due to the increase in the concentration of optical absorbers (the RBCs) 13,14,19 21 The relationship between the PA SA and hematocrit suggests that it might be possible to use PA techniques as a non-invasive means of estimating the hematocrit, an improvement over other non-invasive methods such as ultrasound backscattering where the relationship between the ultrasound backscatter and hematocrit is relatively complex 22 Increasing aggregation levels resulted in changes in the PA SA as shown in figure 3 For a [Dex-PBS] of 3%, the SA was highest for all hematocrit levels The 1% and 8% [Dex-PBS] were nearly identical for all hematocrit levels (p = 02) The SA for the highest aggregation level was ~16x higher than the non-aggregated sample (0% [Dex-PBS]) for both wavelengths of illumination (p = 0001) This can be attributed to the fact that the 3% [Dex-PBS] forms the largest aggregate while the 1% and 8% concentrations yield smaller aggregates as previously reported through independent measures of aggregation 23,24 The SA for the 1064 nm exposure was ~13x smaller than the 750 nm exposure for all samples recorded This could be attributed to the oxygen-depended optical absorption of the RBCs 9 3 (a) SA (au) 2 1 0 3 0 1 3 8 (b) H=10% H=20% H=40% SA (au) 2 1 0 0 1 3 8 [Dex-PBS] (%) Figure 3: PA SA for (a) 750 nm and (b) 1064 nm exposures of RBC samples at 10%, 20% and 40% hematocrit and 1%, 3% and 8% [Dex-PBS] levels The 0% [Dex-PBS] corresponds to the non-aggregated sample Figure 4 shows the SO 2 computed using equation 1 for the RBC samples examined A linear increase in the SO 2 with increasing hematocrit was observed with an average increase of ~7% per doubling of the hematocrit level The largest Proc of SPIE Vol 8581 85813T-4 Downloaded From: http://spiedigitallibraryorg/ on 03/10/2013 Terms of Use: http://spiedlorg/terms

SO 2 was recorded for the 3% [Dex-PBS] samples which was ~30% higher than the non-aggregated case (p = 000001) The 1% and 8% samples were nearly identical and ~20% higher than then non-aggregated samples (p = 000003) These results suggest that the presence of aggregation increases the SO 2 level As this oxygen is bound within the aggregate, the ability of the clustered RBCs to release oxygen to the surrounding environment is diminished This has important implications for adequate oxygen transport to surrounding tissues and further investigations are warranted in order to understand the role of RBC aggregation in vascular pathologies such as atherosclerosis Similar increases in SO 2 with aggregation have been reported using a microscope system coupled with a spectrophotometer 17 The impaired oxygen diffusion due to aggregation is attributed to an increased layer of aggregant/plasma engulfing the RBC aggregates The fact that PA imaging is capable of measuring changes in oxygenation and RBC aggregation simultaneously suggests that it could potentially be used for making assessments of RBC aggregation in diseased tissues while providing information about the level of oxygen and the capacity of the RBCs to deliver oxygen for the tissues of interest 85 80 75 H=10% H=20% H=40% 70 65 SO 2 (%) 60 55 50 45 40 35 30 0 1 3 8 [Dex-PBS] (%) Figure 4: The SO 2 calculated using equation 1 for all blood samples for 3 hematocrit and 3 aggregation levels The 0% [Dex-PBS] corresponds to the non-aggregated sample 4 CONCLUSIONS This paper presents the potential for using PA techniques for assessing oxygenation of blood samples in the presence of RBC aggregates The results of this study suggest that in the presence of RBC aggregation, the release of oxygen to the surrounding environment is impaired It might be possible to use PA imaging for assessing the presence of RBC aggregation in circulatory disorders while measuring the oxygenation of the same area and thereby monitoring the need for oxygen to the surrounding tissues ACKNOWLEDGEMENTS This work was made possible due to the financial support of the following granting agencies awarded to M C Kolios: Natural Sciences and Engineering Research Council of Canada, Canadian Institutes of Health Research, Canadian Foundation for Innovation, Canada Research Chairs Program and the Terry Fox Foundation E Hysi was supported Proc of SPIE Vol 8581 85813T-5 Downloaded From: http://spiedigitallibraryorg/ on 03/10/2013 Terms of Use: http://spiedlorg/terms

through the Alexander Graham Bell Graduate Scholarship A Worthington is gratefully acknowledged for the technical support provided along with J Barry for providing the porcine blood The authors are also very thankful to M Rui and E Berndl for their assistance with the blood specimens REFERENCES [1] CD Hanning and J M Alexander-Williams, Pulse oximetry: a practical review, Brit Med J 311, 367-370 (1995) [2] H F Zhang et al, Imaging of hemoglobin saturation variation in single vessels in vivo using photoacoustic microscopy, Appl Opt 90(5), 053901 (2007) [3] M Henke et al, Blood hemoglobin level may affect radiosensitivitypreliminary results on acutely reacting normal tissues, Int J Radiat Oncol Biol Phys 48(2), 339 345 (2000) [4] B Venkatesh et al, Monitoring tissue oxygenation during resuscitation of major burns, J Trauma 50(3), 485 494 (2001) [5] S S Foo et al, Functional imaging of intratumoral hypoxia, Mol Imaging Biol 6(5), 291 305 (2004) [6] M E George et al, Noninvasive tissue oxygen saturation measurements identify supply dependency, J Surg Res 160, 40-46 (2010) [7] S Hu and L V Wang, Photoacoustic imaging and characterization of microvasculature, J Biomed Opt 15(1), 011101 (2010) [8] K Maslov et al, Effects of wavelength-dependent fluence attenuation of the noninvasive photoacoustic imaging of hemoglobin oxygen saturation of subcutaneous microvasculature in vivo, Inverse Probl 23, S113-S122 (2007) [9] L V Wang, Prospects of photoacoustic tomography, Med Phys 35(12), 5758 2767 (2008) [10] C Li and L V Wang, Photoacoustic tomography and sensing in biomedicine, Phys Med Biol 54(19), R59 R97 (2009) [11] Y Wang et al, In vivo intergrated photoacoustic and confocal microscopy of hemoglobin oxygen saturation and oxygen partial pressure, Opt Let 36(7), 1029-1031 (2011) [12] XWang et al, Noninvasive imaging of hemoglobin concentration and oxygenation in rat brain using highresolution photoacoustic tomography, J Biomed Opt 11(2), 024015 (2006) [13] R K Saha and M C Kolios, A simulation study on photoacoustic signals from red blood cells, J Acoust Soc Am 129(5), 2935-2943, (2011) [14] E Hysi, R K Saha, and M C Kolios, On the use of photoacoustics to detect red blood cell aggregation, Biomed Opt Express 3(9), 2326 2338 (2012) [15] O K Baskurt, B Neu, and H J Meiselman, Red Blood Cell Aggregation, pp 1 304, CRC Press, Boca Raton, FL (2011) [16] N Tateishi et al, Reduced oxygen release from erythrocytes by the acceleration-induced flow shift, observed in an oxygen-permable narrow tube, J Biomech 35(9), 1241 1251 (2002) [17] N Tateishi, N Maeda, and T Shiga, A method for measuring the rate of oxygen release from single microvessels, Circ Res 70(4), 812 819 (1992) Proc of SPIE Vol 8581 85813T-6 Downloaded From: http://spiedigitallibraryorg/ on 03/10/2013 Terms of Use: http://spiedlorg/terms

[18] O K Baskurt et al, New guidelines for hemorheological laboratory techniques, Clin Hemorheol Micro 42(2), 75 97 (2009) [19] R K Saha and M C Kolios, Effects of erythrocyte oxygenation on optoacoustic signals, J Biomed Opt 16(11), 115003 (2011) [20] E Hysi et al, Photoacoustic ultrasound spectroscopy for assessing red blood cell aggregation and oxygenation, J Biomed Opt 17(12), 125006 (2012) [21] A B Karpiouk et al, Combined ultrasound and photoacoustic imaging to detect and stage deep vein thrombosis: phantom and ex vivo studies, J Biomed Opt 13(5), 055406 (2008) [22] R S C Cobbold, Foundations of Biomedical Ultrasound, pp 100 123, Oxford University Press, New York, NY (2007) [23] O K Baskurt, R A Farley, and H J Meisleman, Erythrocyte aggregation tendency and cellular properties in horse, human, and rat: a comparative study, Am J Physiol Heart Circ Physiol 273, H2604 H2612 (1997) [24] O K Baskurt et al, Handbook of Hemorheology and Hemodynamics, pp 1 455, IOS Press, Amsterdam (2007) Proc of SPIE Vol 8581 85813T-7 Downloaded From: http://spiedigitallibraryorg/ on 03/10/2013 Terms of Use: http://spiedlorg/terms