YSI 5905, 5010, 200-BOD, and Pro-BOD Probe Instruction Manual

Similar documents
Dissolved Oxygen Probe (Order Code: DO-BTA)

Thermo Scientific Orion Dissolved Oxygen Probes

Oyster ph - Conductivity - TDS - Temperature Meters

User Guide. Dissolved Oxygen Probes

Ensure Accurate and Successful BOD Measurements with Electrochemical or Optical Sensors

VAN LONDON COMPANY COMBINATION ph ELECTRODE INSTRUCTION MANUAL

QAM-I-113 Operation and Calibration of the Dissolved Oxygen Meter

Model 435A ph Monitor/Recorder Manual

F U C H S L U B R I C A N T S C O.

For products: ConeFET CupFET LanceFET LanceFET+H MicroFET MiniFET SI line ph PROBE MANUAL

Manual No Revision MODEL GLI HAND-HELD/PORTABLE TOTAL/FREE CHLORINE METER

ppm Dissolved Oxygen Measurement

Biochemical Oxygen Demand

Chlorine Dioxide Amperometric 4-20mA Sensors

z /zr/r/ lnitials: tv4'wl z"z{-t L 3 /, /tt Operation and Calibration of the Dissolved Oxygen Meter Revision 5 Approval: Laboratory Man ager/lqao/rso

Tracer TM Total Chlorine PockeTester Code 1740

User Guide. Conductivity Cell

ALKALINE Ti Owner s Manual

Chlorine Dioxide Amperometric 4-20mA Sensors

MASTER. Water Conditioning Corp. OPERATION MANUAL. PURO PRO 1200 SCP Series Modular Reverse Osmosis System. January 2012 Version

L A B D I S S O L V E D O X Y G E N I N S T R U M E N T S

Owner s Manual Calcite Cartridge System

AM-FCL Operation Manual AquaMetrix AM-FCL Free Chlorine Sensor

Global Water Instrumentation, Inc.

PTC1. Paint Test Cell Kit

HD Firewall Mount 850 all, S/V , C ipd #

Universal Tactical Light

Instruction Manual. HI HI Portable ph/ec/tds/ C Meters

Instruction Manual. ph20 ph21. Basic ph meters

Global Water Instrumentation, Inc.

Global Water Instrumentation, Inc.

V 2.0 ASSEMBLY GUIDE. TS Platform Scale Series 2

Instruction Manual HI Portable ph/ec/tds/ C Meters

Pallet Jack 4,400 Lbs Capacity Model #31001

Vernier Software & Technology

6 Closet light. Model Due to continuing improvements, actual product may differ slightly from the product described herein.

HyProTec. Hexavalent free RoHS, ELV and WEEE compliant. HyPro System

PS-31 (SUS304, 300 mm) PS-31 (SUS304, 1,000 mm)

ODYSSEY FLOOR MODEL. Owner s Manual. O-6400-F shown rev 118

Installation and User Manual ispring 3-Stage Ultra-Filtration Water Purifier

MASTER. Water Conditioning Corp. OPERATION MANUAL. PURO PRO 1200 SC Series Modular Reverse Osmosis System. January 2012 Version

Measuring ORP : Tips, Cautions and Limitations

Model T2642E Wall Mount. Television Wall Mount with Tilt Option

Drag Chain Conveyor Installation and Operation Manual

PORTABLE ph METER PRICE CODE NUMBER MODEL 100

Typical Experiments. How it Works

L505 VISUAL LEVEL INDICATOR. Dip-Tape Visual Level Indicator INSTALLATION AND OPERATIONS MANUAL

MultiLab BROCHURE MULTIPARAMETER LABORATORY INSTRUMENTS W61-02

Operating Manual for Digital ph-meter GPHR 1400

We greatly appreciate your business and promise to work hard to provide you with the highest level of service at all times.

Mechanical Post Lift. Owner s Manual

Free-standing rail systems, adjustable. Item no:

Lumina. Bright, leveled deposits in less time Excellent ductility Good chromium reception Very low additive consumption rate. Nickel Plating System

Adjustable Fused Power Supply

Cole Parmer Conductivity Meter

B4 Servo Cover Update S , V ipd #

Features Reliable, DOW FILMTEC Membrane 4 - Stage Reverse Osmosis System - Part Number:

OPERATOR S MANUAL Model 56A Prefeed / Dereeler

Introduction The Van London Co. Cadmium Ion Electrode is used to measure Cadmium ions in aqueous solutions.

Scotch-Weld Metal Bonder Acrylic Adhesive DP8407NS Gray

WARNING! Read all of the information in this booklet, and any relevant Application Notes, before using the Eco Sensors DO3.

Technical Data Sheet February 2014

499AOZ. Features and Applications. Dissolved Ozone Sensor. Product Data Sheet LIQ-PDS-499-AOZ March Rugged construction.

Global Water Instrumentation, Inc.

HI 4105 CARBON DIOXIDE. Instruction Manual HI Carbon Dioxide Ion Selective Electrode

8 DIAL CALIPER. Model Diagrams within this manual may not be drawn proportionally.

Miniglaze/2. Owner & Operator's Manual PORCELAIN STAINING AND GLAZING FURNACE. Model No: Voltage: V 50/60Hz V 50/60Hz

AQUA UNO TUB DOOR INSTALLATION INSTRUCTIONS

Global Water Instrumentation, Inc.

Owners Manual Models: 200-RODI4-XXX. US Water Systems 4 Stage RO/DI System Installation and Maintenance Manual.

Dissolved Oxygen. LabQuest INTRODUCTION

INSTALLATION, OPERATION AND MAINTENANCE MANUAL

Advanced Life Support Basic Life Support Trainer

MiT. MOVING image TECHNOLOGIES INSTRUCTIONS FOR INSTALLATION, OPERATION, AND MAINTENANCE. MPL Digital Pedestal. Manual Version 0.2

Dynatronics Parallel Bars

Lionel Truss Bridge with Flasher and Piers Owner s Manual

RapidChek SELECT Salmonella Test Kit

GC-X360S-W GC-X360S-B

OPERATOR S MANUAL Model 27G Rotary Collet Magnet Wire Stripper

INSTALLATION, OPERATION AND MAINTENANCE MANUAL

Product Name. Here. FLUORAway. Operator s Manual. Histology Innovation for a NEW Generation. Catalog # FA-S. A Division of General Data Healthcare

Model WHARSS5. How to install, operate and maintain your In-Line Refrigerator Water Filtration System. Do not return unit to store

Model 292A. Surface Resistance Test Kit. Instruction Manual

117-Bin Wire Shelf Storage Rack

VWR Zero Air Generator Models , Installation, Operation, and Maintenance Manual. Bulletin TI-ZAG3500-L1466

USE AND CARE/INSTALLATION INSTRUCTIONS Shower Filter model Eco50K and model Eco50KM

Please read through the usage and service instructions outlined in this owner s manual before operating the unit. Warnings Diagrams...

Installation Guide Document No Transaction Windows Copyright 2017 Terra Universal Inc. All rights reserved. Revised October 2017

Low Odour Acrylic Adhesives DP8805NS Green DP8810NS Green DP8825NS Green. Preliminary Technical Data Sheet May 2017

To Contact PS DOORS:

66-Bin Wire Shelf Storage Rack

Mobile Drum Carrier. Owner s Manual

Contact. (984) Distribution Drive, #E Morrisville, NC 27560

DGF-P 2001 CONVEYORS

DESCRIPTION SAFETY: WARNINGS ORDERING INFORMATION T01-B INSTALLATION INSTRUCTIONS DROP-IN AND FREESTANDING BATH K-11343T / K-11344T

Printed in Switzerland EI2354 INSTRUCTIONS AND GUARANTEE CARD AUTOMATIC CHRONOGRAPH CALIBRE 18

FPF-P 2045 FLAT TOP CHAIN CONVEYOR

VISE MOUNTED. Model 97359

Q45WQ. Water Quality Monitoring System. 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park

Agilent 16034E Test Fixture

Transcription:

YSI 5905, 5010, 200-BOD, and Pro-BOD Probe Instruction Manual

TABLE OF CONTENTS INTRODUCTION... 1 OPERATING PRINCIPLES... 2 USING THE PROBE... 3 Changing the Membrane Cap...3 Warm Up...5 Electrode Cleaning...5 OPERATING PROCEDURES AND PRECAUTIONS... 7 Air Calibration...7 Checking the Probe Zero...7 Membrane Life...7 Interferences...8 Acids...8 Erroneous Readings...8 Storage...8 PROBLEM SOLVING... 9 Trouble Shooting Guide...9 SERVICING... 11 WARRANTY AND REPAIR... 13 Authorized Service Centers...14 Cleaning Instructions...14 Packing Instructions...15 PROBE SPECIFICATIONS... 17 APPENDIX A... 18 APPENDIX B... 19 i

INTRODUCTION The YSI Model 5905/5010/200-BOD/Pro-BOD (Biochemical Oxygen Demand) probe is used for measuring dissolved oxygen in all popular size BOD bottles. These probes feature self stirring, easily replaceable membrane caps, and a refurbishable electrode system. The 5905 probe is designed for use with all YSI grey-box-style dissolved oxygen meters, except for the Model 55 or Model 550A. It must be used with an adapter (5011) in order to be used on the Model 5000 and 5100 bench top instruments. The 5010 probe is specifically designed for use with the Model 5000 and 5100 only. The 200-BOD probe is specifically designed for use with the Model DO200 only. The Pro-BOD probe is specifically designed for use with the Pro20 and Professional Plus instruments. This manual also applies to probes 5905-J, 5905-K, 5905-L, 5905- W (Appendix A), and probes 5010-J, 5010-K, 5010-L, and 5010- W (Appendix A) ---IMPORTANT--- Intended Use The 5905/5010/200-BOD/Pro-BOD probe is intended for laboratory use, not field use. It is not waterproof and should never be immersed past the taper on the stem of the probe. 1

OPERATING PRINCIPLES The 5905/5010/200-BOD/Pro-BOD probe is a polarographic dissolved oxygen sensor. An oxygen permeable membrane covers an electrolytic cell consisting of a cathode and an anode. This membrane acts as a diffusion barrier and an isolation barrier preventing the fouling of the cathode surface by impurities in the environment. The cathode is the gold electrode, and the anode is the silver electrode which completes the electrolytic cell and acts as a reference electrode. Upon entering the cell through the membrane, oxygen is reduced at an applied potential of -0.8 V referenced to the silver electrode. The reduction current at the cathode is directly proportional to the partial pressure of oxygen in liquid (expressed as %-air saturation) which is also proportional to the concentration of dissolved oxygen (in mg/l) at a particular temperature. Thus the same partial pressure of oxygen (%-air saturation) in liquid gives different concentrations of dissolved oxygen (mg/l) at different temperatures because of the different solubilities of oxygen at different temperatures. 2

USING THE PROBE The 5905/200-BOD/Pro-BOD must be plugged into a YSI meter and the stirrer power plug must be plugged into its power adapter. The 5010 must be plugged into a YSI meter only, as it gets its stirring power from the 5000/5100 instrument. Changing the Membrane Cap ---IMPORTANT--- New Probe Use New probes are shipped with a dry membrane cap to protect the electrodes. A new membrane cap must be installed before first use. For the 5905/5010, use a Teflon membrane from a 5906 Membrane Kit. For the 200-BOD/Pro-BOD, use a PE membrane from a 5908 Membrane Kit. Follow these instructions to change membrane caps (Figure 1): 1. Remove the stir paddle from the probe by pulling it straight out. 2. Unscrew the old membrane cap from the probe. Before installing a new membrane, clean the probe tip with deionized water in order to remove any contaminants. 3. Hold the membrane cap and fill it at least ½ full with the electrolyte solution provided. Screw the membrane cap onto the probe moderately tight. A small amount of electrolyte should overflow. 4. Rinse off excess electrolyte from the probe with deionized water. 5. Reinstall the stir paddle. 3

Remove Stir Paddle Fill Membrane Cap with Solution Unscrew Cap Replace Stir Paddle Screw Cap on moderately tight Figure 1: Membrane Installation on the 5905/5010/200-BOD/Pro- BOD. 4

Warm Up Every time the instrument is turned on, the 5905/5010/200- BOD/Pro-BOD Probe must be allowed to warm up for about 10-20 minutes before taking the first reading. This allows the probe to use up all the oxygen retained in the cap. ---IMPORTANT--- New Probe Break In (motor off) For best results, please allow the probe more than the usual 20 minutes to warm up after the first membrane is installed. The process may be faster if the probe is allowed to warm up in a BOD bottle filled with clean water. If possible, it s a good idea to let new probes warm up overnight (with a new membrane and KCl) prior to first use. Electrode Cleaning The 5905/5010/200-BOD/Pro-BOD should be cleaned only when erratic readings occur or after about every 500 hours of use (two months). Each cleaning removes material and reduces the life of the probe, so excessive cleaning should be avoided. Gold Cathode For correct probe operation, the gold cathode must be textured properly (for its location, see figure 2). It can become tarnished or plated with silver after extended use. The gold cathode can be cleaned by using the sanding disc provided in the 5906 and 5908 Membrane Kits. Place the disc in the palm of your hand, then sand the gold with a twisting motion about 3 times or until all silver deposits are removed and the gold appears to have a matte finish. If the cathode remains tarnished, return the probe for service. 5

Silver Anode It is normal for a dark layer of silver chloride to cover the silver anode (for its location, see figure 2). After prolonged use it may become necessary to clean the anode. Soak the probe in a 14% ammonium hydroxide solution for 2 to 3 minutes or overnight in a 3% ammonium hydroxide solution. Rinse with deionized water, recharge with electrolyte, and install a new membrane. The anode may also need a very light sanding with the supplied sanding discs. Sand lightly in a helical fashion (down and around), then rinse well with deionized water. Temperature Sensor Anode (Silver) Cathode (Gold) Figure 2: 5905/5010/200-BOD/Pro-BOD Sensor 6

OPERATING PROCEDURES AND PRECAUTIONS ---IMPORTANT--- Intended Use The 5905/5010/200-BOD/Pro-BOD probe is intended for laboratory use, not field use. It is not waterproof and should never be immersed past the taper on the stem of the probe. Air Calibration When calibrating in air, make sure there are no water droplets on the membrane. Water droplets will cause a low calibration reading. Excess water may be removed by shaking the probe downward. If droplets persist it may be necessary to carefully remove them using a clean cloth or paper towel. Checking the Probe Zero The probe zero is checked by immersing the probe in a sodium sulfite solution (0.08M or 3g Na 2 SO 3 /300mL), or in water which has an inert gas bubbling through it (e.g. nitrogen, argon). The meter should read less than 1% dissolved oxygen in either of these environments. If it does not, change the membrane or clean the probe. Membrane Life Membrane life depends on use. Membranes will last longer if installed properly and treated with care during use. Erratic readings will result from loose, wrinkled, or fouled membranes or from bubbles larger than 1/8 in the electrolyte solution. If erratic readings or other evidence of membrane damage occurs, replace the membrane and KCl solution. 7

The average replacement interval is two to four weeks. Probes in constant or heavy use may require more frequent membrane changes. Interferences Hydrogen sulfide, sulfur dioxide, halogens, carbon monoxide, chlorine, nitric oxide, and nitrous oxide can cause the probe to give erroneous readings. If you suspect erroneous readings, it may be necessary to determine if these are the cause. Acids Avoid any environment that contains substances such as concentrated acids, caustics, and strong solvents that may attack the probe. Probe materials that may be damaged by these substances include FEP Teflon or Polyethylene, EPR rubber, ABS plastic, and stainless steel. Erroneous Readings Erroneous readings may occur if the membrane is coated with oxygen consuming bacteria or oxygen evolving algae. Heavy residue may coat the membrane causing incorrect readings. Frequent membrane changes will eliminate this problem. Storage When the probe is not in use, store the probe in a BOD bottle containing at least 1 inch of water. For long term storage, remove the membrane cap, rinse the probe tip with deionized water, and install a dry membrane cap (without electrolyte solution). 8

PROBLEM SOLVING Trouble Shooting Guide Use this quick reference table to correct symptoms you might be experiencing. Symptoms Unable to calibrate Actions Allow approximately 20 min. and up to 30 min. for probe to polarize when first turned on Change membrane Clean probe (see instructions under Electrode Cleaning) Does not return to calibration value Unstable readings Allow sufficient time for temperature and oxygen stabilization Check for water droplets on membrane Change membrane Clean probe (see instructions under Electrode Cleaning) Allow sufficient time for temperature and oxygen stabilization Allow 60 sec for reading to stabilize Change membrane Clean probe (see instructions under Electrode Cleaning) 9

Symptoms Unstable readings Actions Allow sufficient time for temperature and oxygen stabilization Allow 60 sec for reading to stabilize Change membrane Clean probe (see instructions under Electrode Cleaning) Motor not working Check power supply Manually turn the stir paddle to help start the motor Replace motor Loud motor Check if stir paddle is pushed all the way in Replace motor Membrane cap is tight Use pliers to loosen the cap (turn counter clockwise while facing sensor tip) 10

SERVICING The 5905/5010/200-BOD/Pro-BOD has four user replaceable parts. The motor assembly can be replaced if it becomes noisy, unbalanced, or malfunctions. The probe assembly, housing assembly, and stir paddle can also be replaced if they become damaged. To disassemble the probe, perform the following steps while referring to figure 3: 1. Remove stir paddle by pulling it straight out. 2. Remove the two screws (A) on the outside of the housing using a 3/32 Allen wrench. 3. Loosen the top nut of the cable strain relief (B) from the housing. This will allow the cable to feed through the housing assembly. 4. Pull the probe and motor assembly out of the housing and disconnect the two plugs (C). 5. To remove the motor assembly from the probe assembly, remove the two screws (D) using the 3/32 Allen wrench. 11

(B) (A) Housing Assembly (A) (C) (D) (C) (D) Motor Assembly Probe Assembly Stir Paddle Figure 3: Exploded View of 5905/5010/200-BOD/Pro-BOD Probe 12

WARRANTY AND REPAIR YSI Models 5905/5010/200-BOD/Pro-BOD Dissolved Oxygen Probes are warranted for one year from date of purchase by the end user against defects in materials and workmanship. Electrode cleaning or maintenance is not covered under warranty. Within the warranty period, YSI will repair or replace, at its sole discretion, free of charge, any product that YSI determines to be covered by this warranty. To exercise this warranty, write or call your local YSI representative, or contact YSI Customer Service in Yellow Springs, Ohio. Send the product and proof of purchase, transportation prepaid, to the Authorized Service Center selected by YSI. Repair or replacement will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days from date of repair or replacement. Limitation of Warranty This warranty does not apply to any YSI product damage or failure caused by (i) failure to install, operate or use the product in accordance with YSI s written instructions, (ii) abuse or misuse of the product, (iii) failure to maintain the product in accordance with YSI s written instructions or standard industry procedure, (iv) any improper repairs to the product, (v) use by you of defective or improper components or parts in servicing or repairing the product, or (vi) modification of the product in any way not expressly authorized by YSI. THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. YSI s LIABILITY UNDER THIS WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF THE PRODUCT, AND THIS SHALL BE YOUR SOLE AND EXCLUSIVE REMEDY FOR ANY DEFECTIVE PRODUCT COVERED BY THIS WARRANTY. IN NO EVENT SHALL YSI BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING FROM ANY DEFECTIVE PRODUCT COVERED BY THIS WARRANTY. 13

Authorized Service Centers YSI has several Authorized Service Centers in the U.S. and around the world. Please refer to the YSI web site (www.ysi.com) for your nearest Authorized Service Center. Cleaning Instructions NOTE: Before they can be serviced, equipment exposed to biological, radioactive, or toxic materials must be cleaned and disinfected. Biological contamination is presumed for any instrument, probe, or other device that has been used with body fluids or tissues, or with wastewater. Radioactive contamination is presumed for any instrument, probe or other device that has been used near any radioactive source. If an instrument, probe, or other part is returned or presented for service without a Cleaning Certificate, and if in our opinion it represents a potential biological or radioactive hazard, our service personnel reserve the right to withhold service until appropriate cleaning, decontamination, and certification has been completed. We will contact the sender for instructions as to the disposition of the equipment. Disposition costs will be the responsibility of the sender. When service is required, either at the user's facility or at YSI, the following steps must be taken to insure the safety of our service personnel. 1. In a manner appropriate to each device, decontaminate all exposed surfaces, including any containers. 70% isopropyl alcohol or a solution of 1/4 cup bleach to 1 gallon tap water is suitable for most disinfecting. Instruments used with waste water may be disinfected with.5% Lysol if this is more convenient to the user. 2. The user shall take normal precautions to prevent radioactive contamination and must use appropriate decontamination procedures should exposure occur. 3. If exposure has occurred, the customer must certify that 14

decontamination has been accomplished and that no radioactivity is detectable by survey equipment. 4. Any product being returned to the YSI Repair Center should be packed securely to prevent damage. 5. Cleaning must be completed and certified on any product before returning it to YSI. Packing Instructions 1. Clean and decontaminate items to insure the safety of the handler. 2. Complete and include the Cleaning Certificate. 3. Place the product in a plastic bag to keep out dirt and packing material. 4. Use a large carton, preferably the original, and surround the product completely with packing material. Insure for the replacement value of the product. 15

Alternatively, if you are planning to send your probe directly to YSI for servicing, please visit www.ysi.com, download and fill out the Product Return Form. 16

Probe Specifications Cathode: Anode: Gold Silver Membrane: FEP Teflon (for 5905/5010) Polyethylene (for 200-BOD/Pro-BOD) Dissolved Oxygen Accuracy: ± 0.1 mg/l or ± 2% of reading, whichever is greater Operating Temperature Range: 0 to 45 C Temperature Accuracy: ± 0.2 C Polarizing Voltage: 0.8 volts Response Time: Typical response to a step change in dissolved oxygen at 20 C is 90% of the change in 30 seconds. 17

APPENDIX A The 5905 and 5010 probes come in several versions, designed to fit different bottle sizes. 5905 5010 200-BOD Pro-BOD 5905 - J 5010 - J 5 taper per side to fit U.S. Standard BOD bottles. Includes an adapter to fit 5 taper per side smaller bottles with built - in overflow funnels. 5905 - K 5010 - K Includes a 19 funnel. 5905 - L 5010 - L Includes a 24 funnel. 5905 - W 5010 - W 6 long probe for use with wine bottles. The 5010 probe will only work with the Model 5000 and 5100 Instruments. A 5905 probe will also work on these instruments by using a 5011 plug adapter. A 200-BOD will only work with a DO200 Instrument. The Pro-BOD will only work with the Pro20 and ProPlus Instruments. 18

19 APPENDIX B Description 5905-K 5905-L 5905-W 5010 5010-J 5010-K 5010-L 5010-W 200-BOD Pro-BOD 5906 Membrane Kit 059880 059880 059880 059880 059880 059880 059880 059880 N/A 5908 Membrane Kit 605306 605306 Stir Paddle 059902 059902 059950 059895 059902 059902 059902 059950 059895 059895 Probe Assembly 115904 115904 110593 110590 115904 115904 115904 110593 609204 115776 Motor Assembly 110594 110594 110592 110592 110592 110594 110594 110592 110592 110592 Housing Assembly 110591 110591 110591 115011 115011 115011 115011 115011 609210 115777 Funnel / Adapter 059933 059911 059929 059933 059911 N/A 115 V Power Supply See dealer 073035 073035 073035 073035 230 V Power Supply See dealer 073047 073047 073047 N/A 5905 059880 059895 110590 110592 110591 073035 073047 5905-J 059880 059902 115904 110592 110591 059929 073035 073047 Accessories and Replacement Parts by Item Number

1700/1725 Brannum Lane Yellow Springs, Ohio 45387 USA 059882 +1 937 767-7241 800 897-4151 (US) A59882L Fax +1 937 767-9353 May 09 environmental@ysi.com www.ysi.com 1997 YSI Incorporated 20