Second stage cooling from a Cryomech PT415 cooler at second stage temperatures up to 300 K with cooling on the first-stage from 0 to 250 W

Similar documents
Connecting Coolers to Superconducting Magnets with a Thermal-Siphon Cooling Loop

The Role of Quench-back in the Passive Quench Protection of Uncoupled Solenoids in Series with and without Coil Sub-division

A cryogenic tensile testing apparatus for microsamples cooled by miniature pulse tube cryocooler

Study on a High-Power Stirling Cryocooler

Operation of the Levitated Dipole Experiment Floating Coil

A Superconducting Magnet System for Whole-Body Metabolism Imaging

Measured and Calculated Performance of a High Frequency, 4 K Stage, He-3 Regenerator

Measurement Probe Brochures

Cryogenic sample environment on TOSCA

Cryogenic Workstations for Nanotechnology Experiments

Ball Aerospace Next Generation Two-Stage 35 K Coolers: The SB235 and SB235E

Energy efficiency and saving potential analysis of the high intensity proton accelerator HIPA at PSI

Characteristics of a VM Type Thermal Compressor for Driving a Pulse Tube Cooler

Effect of the Charged Pressure on GM Cryocooler Performance

Experimental Investigation of Piston-Free Expander Accompanied by Cold Linear Compressor

Thermo-Fluid Modeling of a Full Scale Liquid Hydrogen Storage System with Integrated Refrigeration

A simulation study for the virtual commissioning of the CERN central helium liquefier

Performance analysis of a solar still coupled with evacuated heat pipes

Energy Efficiency and Recovery at Large Scale Cryogenic Plants: A Survey

Practical experience of using ultrasound flowmeters at the measurement associated petroleum gas

THE CRYOGENIC SYSTEM FOR THE LHC TEST STRING 2: DESIGN, COMMISSIONING AND OPERATION

Specialised Hydrocarbon-based Grease for Cryogenic Applications

Available online at ScienceDirect. Procedia Engineering 152 (2016 )

Cryogen Free Superconducting Magnet Systems

R. C. Hon, C. S. Kirkconnell, and J. A. Shrago. Raytheon Space and Airborne Systems El Segundo, California, 90245, USA

Preliminary Results on a Large Acoustic-Stirling ("Pulse Tube") Cooler Designed for 800W at 70K

The DAΦNE Cryogenic System

The Measurement of RRR and Resistance for Aluminum Alloy Based on a Cryocooler

Special features of ITER Cryogenic system

Y, XY, XYZ and Sample Manipulators. MultiCentre Sample Handling Systems. MultiCentre XYZ Stages

Photoetched Regenerator for Use in a High Frequency Pulse Tube

Available online at ScienceDirect. Physics Procedia 67 (2015 ) K Liquid Neon Energy Storage Unit

CERN experience and strategy for the maintenance of cryogenic plants and distribution systems

Development of a 15 W Coaxial Pulse Tube Cooler

Development of a Diaphragm Stirling Cryocooler

Vertically aligned 14 Tesla split-pair

Experimental and numerical investigation of cooling performance of a cold storage in a pharmaceutical industry

AN ADJUSTABLE THICKNESS Li/Be TARGET FOR FRAGMENTATION OF 3-kW HEAVY ION BEAMS

Development of portable power unit with catalytic micro-combustor

Testing of a 1 kw-class Cryogenic Turboalternator

Construction of a Low Temperature Heat Exchanger Cell

Earthquake Analysis for Cryogenic Components of NPDG Liquid Hydrogen Target

The change of transformation temperature on NiTi shape memory alloy by pressure and thermal ageing

Design, construction and evaluation of a system of forced solar water heating.

Systems for Mӧssbauer Spectroscopy

A millikelvin cryocooler for space

Earlier Lecture. In the earlier lecture, we have seen the importance of instrumentation in Cryogenic Engineering.

Thermally operated mobile air-conditioning system

CRYOGENICS. MSN 506/Phys 580

Identification and reduction of vibration and noise of a glass tempering system

Effects of shaft supporting structure on performance test of axial flow fan

Second-Law Analysis of a Hybrid Reverse Brayton Stirling Cryocooler

Vibration-Free, Hybrid Cryocooler for 4 K Space Applications

Elastic heat exchanger in Stirling cycle machines

Series Supply of Cryogenic Venturi Flowmeters for the ITER Project

Resistance Spot Welding of AA5052 Sheet Metal of Dissimilar Thickness

Research on a Magnetic Refrigeration Cycle for Hydrogen Liquefaction

Establishment of Low Energy Building materials and Equipment Database Based on Property Information

1. Introduction. Piyush Dnyaneshwer Raut

Detector Solenoid and Iron Structure

Estimation of natural circulation flow based on temperature in the FASSIP-02 large-scale test loop facility

Developments of 1-4 K Class Space Mechanical Coolers for New Generation Satellite Missions in JAXA

Decline Curve Analysis for Production Forecast and Optimization of Liquid-Dominated Geothermal Reservoir

Analysis of mismatch and shading effects in a photovoltaic array using different technologies

> HGW WATER - WATER AND BRINE - WATER HEAT PUMPS FOR INDOOR INSTALLATION. Available range. Unit description. Options. Accessories

Water Cooling System of PETRA III. MAC - Meeting at DESY 11/08/2004 Jensen / Wanning - MKK - 1

ANSI/ASHRAE STANDARD , METHODS OF TESTING CHILLED BEAMS

LOSS OF COOLANT ACCIDENT AND LOSS OF FLOW ACCIDENT ANALYSIS OF THE ARIES-AT DESIGN

Study of a 35 K Regenerator Performance Operating at High Frequency

CRYOGENICS OPTIONS FOR THE MMA. L. D'Addario, 97/09/20 REQUIREMENTS. We begin by assuming that the required cooling at each antenna

2. Inclined solar panel basin solar still in passive and active. mode

Vibration influence on structure and density of aluminum alloys

A Pulse Tube Cryocooler with 300 W Refrigeration at 80 K and an Operating Efficiency of 19% Carnot

Technical Guide to Solar Thermal. Better World Energy. How to harness the sun s energy and create hot water with a Solar Thermal System

ITER Organization. Procurement Process Industrial Strategy Business Opportunities

Influence of nickel addition on magnetic and electro-mechanical behaviour of permalloys

Thermal Performance of Thermoelectric Cooler (TEC) Integrated Heat Sink and Optimizing Structure for Low Acoustic Noise / Power Consumption

Recent activity toward high-q at KEK. 2016/11/3 TTC High-Q meeting KEK Kensei Umemori

LABORATORI NAZIONALI DI FRASCATI SIS-Pubblicazioni

CRYOGENICS AND LHC BEAM COMMISSIONING

EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH Laboratory for Particle Physics RECENT DEVELOPMENTS AND QUALIFICATION OF CRYOGENIC HELIUM FLOW METERS

Experience with Cryogenic Operations During the Fermilab Collider Program

Certification, Testing, Approval and R&D work on the Absorber Safety

Thermoacoustics for cold production

PROGRESS ON THE SECTOR MAGNETS FOR THE RIKEN SRC

Temperature Control of Industrial Water Cooler using Hot-gas Bypass

ENGINEERING INFORMATION Hot water and steam service

Mini-project report. Production of Optimised Metal Foams for Stirling Engine Regenerators. Erardo Elizondo

Compression Characteristics of Honeycomb Sandwich Panels to Improve Their Impact Resistances

Power quality analysis of DC arc furnace operation using the Bowman model for electric arc

Brazed aluminium heat exchangers (BAHXs), also referred to

LAB 13: HEAT ENGINES AND THE FIRST LAW OF THERMODYNAMICS

Optimization control of LNG regasification plant using Model Predictive Control

Influences of the current density on the performances of the chromeplated layer in deterministic electroplating repair

Cavitation measurements on a pump-turbine model

Realization of ITS-90 from K through K: One Company s Approach

Chem 355 Jasperse DISTILLATION

ScienceDirect. Cryogenics at the European Spallation Source

Master of Technology (M. Tech.)

A High Frequency Thermoacoustically Driven Thermoacoustic-Stirling Cryocooler

Transcription:

IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Second stage cooling from a Cryomech PT415 cooler at second stage temperatures up to 3 K with cooling on the first-stage from to 25 W To cite this article: M A Green et al 215 IOP Conf. Ser.: Mater. Sci. Eng. 11 122 Related content - Performance improvement of a large capacity GM cryocooler C Wang, A Olesh and J Cosco - Progress on the superconducting magnets for the MICE cooling channel M A Green, S P Virostek, D Li et al. - Acknowledgments View the article online for updates and enhancements. This content was downloaded from IP address 37.44.196.122 on 2/1/218 at 21:42

Second stage cooling from a Cryomech PT415 cooler at second stage temperatures up to 3 K with cooling on the first-stage from to 25 W M A Green 1,2, S S Chouhan 1, C Wang 3, and A F Zeller 1 1 FRIB, Michigan State University, East Lansing MI 48824, USA 2 Lawrence Berkeley Laboratory, M/S161, Berkeley CA 9472, USA 3 Cryomech Inc. 113 Falso Drive, Syracuse NY 13211, USA green@frib.msu.edu and magreen@lbl.gov Abstract. The amount of cooling delivered to the second stage of a two stage cooler is dependent on the second stage temperature and the amount of refrigeration provided by the cooler first stage. The second stage cooling as a function of temperature for a Cryomech PT-415 cooler (1.5 W at 4.2 K with 42 W on the first stage) has been estimated by scaling similar data that was measured for a Cryomech PT-41 cooler (1. W at 4.2 K with 28 to 3 W on the first stage). In order to accurately calculate the cool-down time for a superconducting magnet using PT-415 K coolers one must know how much cooling can be delivered by the cooler second-stage as a function of the second-stage temperature and the added cooling delivered to the cooler first-stage. There are applications where PT415 coolers are used in the temperature range from 15 to 25 K to liquefy hydrogen or cool magnets fabricated from MgB 2. This report describes the method for estimating the cooler performance for a PT415 cooler as well as the results of the measurements on several PT415 coolers. 1. Introduction The Cryomech PT41 and PT415 pulse tube coolers are used for cooling superconducting magnets and other applications [1]. These coolers are also the central element in the small helium liquefiers produced by the company for universities and other institutions that require liquid helium for research and product development [2]. Most of the uses for these coolers involve the re-condensation of liquid helium using a thermal-siphon cooling-loop or in a liquid helium tank [3], [4]. When a cooler is used for helium re-condensation, the cooler second stage performance 4.2 K is often all that is needed. When something goes wrong with a magnet cryostat or magnet cooling system, more knowledge about the cooler performance is needed to determine what went wrong [5], [6]. Using a cooler to cooldown a cryostat requires knowledge of the cooler performance over a wide temperature range [7]. Cooler operating data over a range of temperatures on both stages is critical for understanding how a device is cooled-down using these coolers. Two-stage coolers may also be used for cooling at higher temperatures (say up to 28 K) to cool MgB 2 magnets [8]. These coolers can also be used for the hydrogen liquefaction and hydrogen re-condensation after liquefaction [9], [1]. For these reasons, it is important to understand how these coolers behave as a function of 1 st and 2 nd stage temperatures. Content from this work may be used under the terms of the Creative Commons Attribution 3. licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. Published under licence by Ltd 1

2. Estimating PT415 Performance based on PT41 Performance Measurements In order to understand how a thermal siphon cooling loop behaves during a magnet cool-down one must have cooler performance curves that apply to the coolers that are being used to cool-down a magnet or other device. This statement is also true for cryogen free systems as well. Michigan State University (MSU) looked at cooling down two systems using a thermal-siphon cooling loop. These systems are; 1) an ECR Ion source magnet [11] and 2) the cyclotron gas-stopper magnet [7], [12]. In both cases, MSU used the measured extended PT41 cooler parameters to estimate the performance of the PT415 cooler and the PT415-RM cooler (the remote valve motor version of the PT415 cooler that is used in certain circumstances where there are high magnetic fields [13] or high voltages [14]). Figures 1 and 2 show the measured performance of the PT41 cooler taken in 24 [1]. Figure 1. The operating parameters of a PT41 cooler in the nominal liquid helium range on the cooler second-stage. Figure 2. The extended operating parameters for a PT41 cooler from zero to 3 K on both the first and second stages of the cooler. From the data in Figure 1 and Figure 2, MSU estimated what the output of the PT415-RM coolers might be. The heat loads on the second-stage and first-stage was multiplied by 1.35 because a remote motor PT415 cooler has an output of 1.35 W at 4.2 K and about 36 W at 4 K as compared to 1.5 W at 4.2 K and 4 W at 4 K for a standard PT415. This assumption was used to calculate the refrigeration on the cooler second-stage at temperatures above 4.2 K. This cooling model was used to calculate the cooling needed to cool-down 124 kg of cold mass in each of the cyclotron gas stopper magnet coils and cryostats plus the mass of the second stage cold heads, the condenser heat exchangers, and the pipes that connect the coolers to the magnet cold mass. The mass cooled-down was supposed to average about 425 kg per cooler. Figure 3 shows the refrigeration R on the second stage is plotted as a function of the second stage temperature T of a PT415-RM cooler from the model based on the PT41 cooler with no heat on the 1 st -stage. From this graph a curve fit was made. The equation developed from the fit is as follows; R = 5.259 +1.4738T 8.8986x1 3 T 2 + 5.229x1 5 T 3 1.319x1 7 T 4. (1) Figure 4 shows the inverse case where the temperature of the 2 nd -stage cold head T is plotted as a function of the heat load on the 2 nd -stage cold head Q with heat on the 1 st -stage. From this graph a curve fit was made. The fitted equation was used in the model of the cool-down of the MSU cyclotron 2

gas-stopper magnet coils using three PT415-RM coolers to cool-down each of the magnet coils. The equation developed from the fit is as follows; T = 2.415 +1.1719Q 2.639x1 2 Q 2 + 6.7884x1 4 Q 3 6.189x1 6 Q 4 +1.953x1 8 Q 5. (2) The fit of equations (1) and (2) is pretty good with an R 2 = 1.. The equations don t fit the data very well at very low temperatures or at temperatures above 17 K 16 18 14 16 12 14 Second Stage Refrigeration R (W) 1 8 6 4 Cooler 2nd Stage Temperature T (K) 12 1 8 6 4 2 2 2 4 6 8 1 12 14 16 Second Stage Temperature T2 (K) 18 2 2 4 6 8 1 12 Cooler Second Stage Heat Load Q (W) 14 Figure 3. A model of the 2 nd stage refrigeration as a function of the 2 nd stage temperature for a PT415-RM cooler based on a PT41 cooler. Figure 4. A model of the 2 nd stage temperature as a function of the 2 nd stage heat load for a PT415-RM cooler based on a PT41 cooler. The model based on a PT41 cooler worked reasonably well for modeling the performance of a thermal-siphon cooling loop as the cyclotron gas-stopper magnet was built. We know from Cryomech measurements extended performance that this model doesn t work very well at high heat loads on the cooler. The model worked reasonably well for cooler heat loads between 1 W and 6 W. For cooler heat loads above 75 W, the cooler performance deviates substantially from measured performance of a PT415 cooler. The fit falls apart above R =12 W. The performance curves in Figures 3 and 4 don t fit very well at low heat loads as well. These difficulties didn t affect the modeling of the cool-down of the cyclotron gas-stopper magnet cool-down very much [15]. It was clear that more temperature data was needed for second stage heat loads between 3 W and about 8 W. 3. Methodology for measuring Cooler Performance All of the cooler performance data taken used same methodology. The cooler must be in a good vacuum during the test. The part of the cooler below the first stage of the cooler must be in a shield that is cooled by the cooler first stage. The shield should have a low emissivity surface and there should be at least five layers of MLI between the shield and the second stage cooler parts. There must be at least ten layers of MLI between the shield and the first stage parts and the vacuum vessel. 3

CEC 215 Figure 5 shows the test set up for a PT415 cooler that was tested in June 215. Figures 6a and 6b show a smaller cooler at Cryomech that had already been tested. The photos in Figures 6a shows the copper shield attached to the first stage cold head after the outer MLI was removed. The photo in Figure 6b shows the MLI that covers the second stage pulse tube, regenerator tube and the second stage cold head to prevent radiation shine between the first stage and shield and the second stage. Figure 5. The PT415 cooler test setup at Cryomech. The compressor is on the left. The cooler cold head and vacuum vessel are to the right. Twenty meters of line connecting the compressor on the left to the cooler cold head on the right is coiled up to the left of the PT415 compressor. Figure 6. a) The cooler shield connected to the 1st stage and b) the MLI inside of the shield 4

For tests done at Cryomech in June of 215, there were heaters capable of generating up to 1 W were attached to both stages of the coolers. During the cooler extended characteristic test done at Cryomech in 26, there were heaters on both stages that could generate up to 25 W. The heaters must be firmly attached to the cooler stage cold heads. The temperature sensors must be firmly mounted to the cold heads in a location that is well away from the heaters. The temperature sensors connect to the cooler colds must be fully calibrated over a temperature range from 2 K to 4 K. Cryomech uses Scientific Instruments fully calibrated silicon diode sensors. Since there is no magnetic field, this sensor is a good choice. Cernox sensors can be used as well as long as they are fully calibrated over a temperature range similar to the silicon diode sensors. The temperature sensor wires must be thermally connect to the cold heads. The heaters are controlled by a controller that keeps the heater power constant to within.1 W. The temperatures of both cold heads are measured every second. The pulse tubes in the PT415 cooler pulse at the rate of 1.2 Hz. This pulse rate is independent of the line frequency. Unlike most coolers, the refrigeration on both cold heads is independent of the line frequency. The temperature on the cold heads pulses at the rate of 1.2 Hz. The temperature pulsation is quite evident at low temperatures. The cold head temperatures should be time averaged over several minutes. At low temperatures on the 2 nd -stage cold head temperature peak-to-peak amplitude can be as high as.6 K. Figure 7 shows the temperature oscillations on the cooler second stage. Figure 8 shows the measured peak-to-peak amplitude as a function of 2 nd -stage heat load Q2 as two different 1 st -stage heat loads Q1. To first order, the 2 nd -stage temperature T2 is proportional to the 2 nd -stage heat load Q2 at Q2 less than 4 W..6.5 Q1 = W Q1 = 4 W Temperature Spread due to Pulsing (K).4.3.2.1 Figure 7. The transition between a 15 W heat load and a 1 W heat load on the cooler 2 nd -stage. In 3 minutes, the cooler pulses 216 times. The average cold head T is at the center of the band.. 1 2 3 Heat Load on Second Stage (W) Figure 8. A plot of the amplitude (peak-topeak) versus the second-stage heat load for 1 st -stage heat loads Q1 of and 4 W. 4 The temperature amplitude of the pulse increases in amplitude from 4 K to about 1 K. The amplitude of the temperature pulsation is larger at 2.7 K than it is at 4.2 K. As the temperature increases above 1 K the amplitude of the pulse decreases rapidly. This is a manifestation of the increased specific heat of the cold head with respect to that of the helium gas. The increased pulsation amplitude between 4.2 K and 2.7 K may reflect an increase in the specific heat of the helium gas. The authors do not have an explanation for the temperature oscillation amplitude behavior as a function of 1 st -stage heat load Q1. 5

4. PT415 Cooler Performance Graphs over Various Temperature Ranges This section presents the operating parameters for the PT415 cooler. The operating parameter data is for three different coolers. The performance of any type of cooler is not exactly the same from cooler to cooler. This is true for coolers from all of the small cooler manufacturers. Some components in small coolers are hand made. Each cooler is packed with regenerator material by hand. Variations of the helium pressure in the hoses between the cold head and the compressor affects the cooler operating performance on both stages. Each cooler has its own personality and its own sweet spot where the cooler works the best. Figure 9 shows the temperature map for a PT415 cooler measured by Florida State University [16]. Similar measurements were made by Cryomech [17] in 26, but these measurements didn t cover as much of the operating space as the FSU measurements. Figure 9 shows the temperature map for firststage heat loads from to 84 W combined with second stage heat loads from to 3 W. Having the FSU temperature map proved invaluable in determining what went wrong with the MICE spectrometer solenoids [18]. 5.5 Second Stage Temperature (K) 5. 4.5 4. 3.5 3. Q2 = 3. W Q2 = 2.5 W Q2 = 2. W Q2 = 1.5 W Q2 = 1. W Q2 =.5 W Q2 =. W Q1 = W Q1 = 21W Q1 = 42W Q1 = 63W Q1 = 84W 2.5 3 4 5 6 First Stage Temperature (K) 7 Figure 9. PT415 temperature map as a function of Q1 from to 84 W and Q2 from to 3W [16]. 36 Second Stage Temperature T2 (K) 3 24 18 12 6 3 Q2 = 4 W Q2 = 35 W Q2 = 3 W Q2 = 25 W Q2 = 2 W Q2 = 15 W Q2 = 1 W 35 4 45 Q2 = W 5 Q2 = 5 W Q2 = 1.5 W 55 6 Q1 = W Q1 = 2 W Q1 = 4 W Q1 = 5 W First Stage Temperature T1 (K) Figure 1. PT415 temperature map as a function of Q1 from to 5 W and Q2 from to 4 W. 6

Figure 1 shows an extended temperature map measured in June 215 at Cryomech. This map covers temperatures from to 36 K on the second-stage cold head and 3 to 6 K on the first-stage cold head. The temperature map in Figure 1 was measured with a heat load Q1 on the first-stage from to 5 W while a heat load Q2 was applied on the second-stage from to 4 W. The parameter space was extended to Q2 up to 8 W for a Q1 between and 4 W, but that is not shown in figure 1. The temperature map shown in Figure 1 is very useful for applications in the liquid hydrogen range (from a triple point temperature at 13.8 K and a critical temperature of 32.2 K for para-hydrogen) such as are found in the MICE hydrogen absorbers [9] or in a liquid hydrogen-cooled magnet [8]. Figure 11. PT415 temperature map as a function of Q1 from to 25 W and Q2 from to 125 W. Figure 11 above includes the temperature range from absolute zero to room temperature on both of the PT415 cooler stages. The cooler map shown in Figure 11 is useful for determining the cool-down behavior of a PT415 cooler. The data from Figure 11 and Figure 1 would be useful for improving the cool-down model for the cyclotron gas-stopper magnet at MSU [7], [15]. 5. Concluding Comments The data that is in this paper provides a more complete picture of the operating parameters of the PT415 cooler and its remote motor cousin the PT415-RM cooler, which has ninety percent of the cooling of the PT415 cooler on both cold heads. Having more than one temperature map of a cooler such as the PT415 cooler is important in understanding the behavior of a piece of apparatus that has one or more coolers installed within it. The caveat is that one must have fully calibrated temperature sensors firmly attached to the cooler cold heads, in a helium bath (if there is one) and in key places on the apparatus being cooled-down and being kept cold using the coolers. The temperature sensors must be accurate under the operating conditions of the device being cooled. If there is liquid helium in the apparatus, a measurement of the vapor pressure of the helium is a useful check. Acknowledgments The authors thank Joe Costco of Cryomech for helping the lead author do the tests at their plant. This work was supported in part by an NSF grant PHY-9-58726 and PHY-11-2511 and by the Office of Science, United States Department of Energy, under DOE contract DESC661. 7

References [1] Data Sheets from Cryomech Inc. at 113 Falso Drive, Syracuse NY 13211, USA. For data sheets and drawings for the PT41 and PT415 coolers go to www.cryomech.com. [2] Helium liquefier data sheets from Cryomech Inc. at 113 Falso Drive, Syracuse NY 13211, USA. For He liquefier data sheets go to the company web address www.cryomech.com. [3] Green M A and Wang S T, 21, Tests of four PT415 coolers installed in the drop-in mode, Proceedings of ICEC-22, Seoul Korea 21-25 July 29, pp 15-11 [4] Green M A, and Wang S T, 212, Tests of copper and HTS leads with a two-stage pulse tube cooler, Advances in Cryogenic Engineering 57, pp 581-588, AIP Press, Melville NY [5] Green M A, Pan H, and Preece R M, 214, Changes made on a 2.7-m long superconducting solenoid cryogenic system that allowed the magnet to be kept cold using 4 K pulse tube coolers, Advances in Cryogenic Engineering 59B, AIP Conference Series 1573B, pp 1551-1558, AIP Press, Melville NY [6] Green M A, 214, Lesson learned concerning the use of 4 K coolers to cool LTS Magnets, Proceedings of ICC-18, Syracuse NY, USA 9 to 12 June 214, ISSN 1549-1757, pp 567-576 [7] Green M A, Chouhan S S, and Zeller A F, 214, Design limitations on a thermal-siphon 4 K helium loop for cooling-down the cyclotron gas stopper magnet coils, Advances in Cryogenic Engineering 59B, AIP Conference Series 1573B, pp 1543-155, AIP Press, Melville NY [8] Green M A, 214, Cooling and cooling-down a MgB2 and HTS magnets using a hydrogen thermal siphon cooling loop and coolers running from 15 K to 28 K, IEEE Trans. Appl. Supercon. 24, No. 3, paper 51134 [9] Green M A, et al, 26, Cooling the MICE liquid absorber with a small cooler, Advances in Cryogenic Engineering 51, pp 176-183, AIP Press, Melville NY [1] Green M A, 21, Re-condensation and liquefaction of helium and hydrogen using coolers, Advances in Cryogenic Engineering 55, pp 73-71, AIP Press, Melville NY [11] Green, M A, et al, The cool-down and cooling of an ion source magnet using a thermal-siphon between the coolers and the magnet, presented at the 24th International Cryogenic Engineering Conference, Fuluoka Japan, 14 to 18 May 212 (This paper was not published, but it is an MSU FRIB note FRIB-P215-CA-19-R1 issued 21 May 213.) [12] Chouhan S S, et al, 213, The superferric cyclotron gas-stopper magnet design and fabrication, IEEE Trans. Appl. Supercon. 23, No. 3, p 41185 [13] Green M A and Witte H, 28, The use of small coolers in a magnetic field, Advances in Cryogenic Engineering 53, pp 1299-136, AIP Press, Melville NY [14] Green M A, et al, 215 Progress on the cryogenic and current tests of the MSUcyclotron gasstopper superconducting magnet, Proceedings of IPAC-15, Richmond Virginia USA [15] Green M A, et al, 215, Lessons learned from the cool-down of a superconducting magnet using a thermal-siphon cooling-loop, Advances in Cryogenic Engineering 61, IOP Conference Series (this publication) [16] Choi Y S, et al, 26 Helium-liqufaction by cryocoolers for high field magnet cooling, Proceedings of the International Cryocooler Conference ICC-14 [17] Private communication with C. Wang of Cryomech Inc. 113 Falso Dr, Syracuse NY 13211, USA (215) [18] Green M A, What happened with spectrometer magnet 2B, MICE Note 292, (27 May 21), http://www.mice.iit.edu. 8