DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, & USING THE AGILENT 5100 ICP-OES

Similar documents
The Agilent Atomic Spectroscopy portfolio for Environmental applications AA, MP-AES, ICP-OES, ICP-MS & ICP-QQQ

Analysis of steel and its alloys using the GB/T standard and an Agilent 5100 ICP-OES in dual view mode

Multi-Element Analysis of Trace Metals in Animal Feed using ICP-OES

Table I: MCL and MRL Concentrations for Contaminants Monitored Under the Safe Drinking Water Act National Primary Drinking Water Regulations

Determination of Elemental Impurities in Graphite-based Anodes using the Agilent 5110 ICP-OES

Ultra-fast ICP-OES determinations of base metals in geochemical samples using next generation sample introduction technology

Increased Sample Throughput for ICP-OES Applied to U.S. EPA Method 200.7

Benefits of running organic matrices using the Agilent 5100 ICP-OES fast, robust, high performance analysis

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

Agilent 5100 ICP-OES DUAL VIEW ICP-OES MINUS THE WAIT

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo

EU water analysis using the Thermo Scientific icap 7400 ICP-OES Duo

Increase productivity for environmental sample analysis using the SVS 2+ Switching Valve System for Agilent 5100 SVDV ICP-OES

Analysis of Mineral and Heavy Metal Content in Beverages Using the Teledyne Leeman Labs Prodigy Plus ICP-OES

Analysis of Environmental Samples with the Agilent 730-ES Following US EPA Guidelines

Composition Analysis of Animal Feed by HR ICP-OES

Analysis of wear metals and contaminants in engine oils using the 4100 MP-AES

Amplifying ICPMS Productivity Using Discrete Sampling Technology

The Determination of Trace Elements in Waters using the US EPA Method and the Thermo Scientific icap 6500 ICP-OES (Duo)

Agilent Atomic Spectroscopy Solutions for Energy & Chemicals AA, MP-AES, ICP-OES, ICP-MS.

Determination of 22 Elements Following US EPA Guidelines with a New Megapixel CCD ICP-OES

Accurate. Robust. Reliable. AGILENT 710 SERIES ICP-OES

Accurate. Robust. Reliable. AGILENT 710 SERIES ICP-OES

Improved Antimony Recoveries in Soil Matrices by 3050B/6010B. DataChem Laboratories, Inc. Robert P. Di Rienzo Jeffery S. Ward John T.

Flame Atomic Absorption Spectrometer (FAAS) Brand : PERKINALMER Model : PinAAcle 900T Custodian : PRADUP MESAWAT SOOKRUTHAI WONGSUBIN

The Determination of Toxic Metals in Waters and Wastes by Furnace Atomic Absorption

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

Automating EPA 6020 Compliant Analysis with the Agilent 7900 ICP-MS and ESI prepfast Autodilution System

Analysis of Toxic Trace Elements in Coffee Bean Products by HR ICP-OES

PINAACLE 500/900 AA AND SUPPLIES CONSUMABLES. Consumables and Supplies Reference Guide

Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

EPA Primary. (mg/l as CaCO3) (mg/l as CaCO3)

PinAAcle 900 AA Flame AA

The Determination of Alloying Elements in High-Carbon Steel using Radially Viewed Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES)

Analysis of inorganic impurities in lubricating oils by ICP-MS

VHG Labs. Standards and Supplies for Spectrochemical Analysis QUALITY. TRACEABILITY. DEPENDABILITY. WORLDWIDE

The Determination of Trace Elements in Stainless Steel by Forked Platform GFAAS

Standard Solutions (Traceable to NIST)

High-Purity Acids & Reagents

Application Note. Abstract. Authors. Environmental

Water Analysis Using ICP-OES with an Ultrasonic Nebulizer

Standard Solutions (Traceable to NIST)

Analysis of Metals in Water, Stream Sediments and Floodplain Soils Collected March 21-23, 2005 from the Bayou Creek System. David J.

ESI SC-FAST Sample Introduction Solutions for PerkinElmer ICP-OES and ICP-MS Systems

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo

Radial, Axial or Dual View ICP: Which Do You Choose? Manny Almeida Teledyne Leeman Labs, Inc. Hudson, NH

Atomic Spectroscopy Solutions for Environmental Applications

Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

US EPA Method using the XSERIES 2 ICP-MS

ELEMENTS by ICP 7301 (Aqua Regia Ashing)

Analysis of Heavy Metals in Iron-based Fertilizers by HR ICP-OES

Metal Aerosol Emissions from Biomass Combustion

High matrix tolerance of the Thermo Scientific icap 7000 Plus Series ICP-OES with the radial Ceramic D-Torch and the sheath gas adaptor

Collaborative Efforts for Creating a Robust Trace Metals Analytical Procedure for Flue Gas Desulfurization Wastewaters by ICP-MS

Functionality. Normal Operation

Atomic Absorption Spectrometry. The Analyzer 4 You novaa 800

Direct Analysis of Photoresist by ICP-MS. Featuring the Agilent Technologies 7500s ICP-MS

TITLE: SABS PROFICIENCY TESTING SCHEME, RSA

Analytically Monitoring the Effect of Fracturing Activity

Analysis of Trace Metals in Surface and Bottled Water with the Optima 7300 DV ICP-OES

CERTIFICATE OF ANALYSIS 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION

Characterization of Surface Metals on Silicon Wafers by SME-ICP-MS. Featuring the Agilent Technologies 7500s ICP-MS

Leading the way in atomic spectroscopy innovation. AA, MP-AES, ICP-OES, ICP-MS, ICP-QQQ Product Positioning

Agilent 5100 ICP-OES DUAL VIEW ICP-OES MINUS THE WAIT

Achieving Optimum Throughput in ICP-MS Analysis of Environmental Samples with the Agilent 7500ce ICP-MS Application

SUB-Programs - Calibration range Fe Base for "PMI-MASTER Pro" Spark - mode Fe 000

CERTIFICATE OF ANALYSIS

Analysis of elemental impurities in drug products using the Thermo Scientific icap 7400 ICP-OES Duo

Common Challenges facing Laboratories doing Elemental Analysis Today

ELEMENTS by ICP 7301 (Aqua Regia Ashing)

Agilent 200 Series AA Systems AGILENT ATOMIC ABSORPTION SPECTROMETERS

Review on Agriculture and Rural Development vol. 2. (1) ISSN

METHOD 3040A DISSOLUTION PROCEDURE FOR OILS, GREASES, OR WAXES

Trace Metals in Waters by GFAAS, in Accordance with U.S. EPA and Health Canada Requirements

Agilent Resources for Workflow Compliance MEASURING ELEMENTAL IMPURITIES IN PHARMACEUTICAL MATERIALS

High-Purity Acids & Reagents

icap 7000 Plus Series ICP-OES Streamlined performance and ultra-low ICP-OES detection limits for routine laboratories

Background Statement for SEMI Draft Document 4795 (Yellow Ballot) NEW STANDARD: SPECIFICATIONS FOR NITRIC ACID, USED IN PHOTOVOLTAIC APPLICATIONS

CERTIFICATE OF ANALYSIS

CERTIFICATE OF ANALYSIS 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION

Analytical CHEMISTRY

Bottom Ash Data Week 45

Runs On Air. AGILENT 4100 MICROWAVE PLASMA ATOMIC EMISSION SPECTROMETER

Inductively coupled plasma (ICP) spectroscopy for online measurements of trace metal emissions at the 250 kw PACT furnace

Bottom Ash Data Week 52

CERTIFICATE OF ANALYSIS REVISED REPORT

Appendix C: Water Quality Data

Bottom Ash Data Week 47

EPA Methods Update. What s new and what s coming in EPA ICP-MS methods?

Application Note AN1501

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance

Guidelines for Troubleshooting and Maintenance of AA Systems

Productive. Precise. Reliable. AGILENT ATOMIC ABSORPTION SPECTROMETERS

Comparing an instrumental technique based on its

EPA Methods Update 2. What s new and what s coming in EPA ICP-MS methods? 200.8

This method is applicable to the following parameters: Nitric Hydrochloric acid digestion, Instrumental analysis.

Trace Elements in Petrochemical Samples: Achieving Excellent Analytical Results

Multi-element determination in pharmaceutical preparations using the Thermo Scientific icap Q ICP-MS

ELEMENTS by ICP 7300 (Nitric/Perchloric Acid Ashing)

Transcription:

ENVIRONMENTAL ANALYSIS DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, 14543 & 13428 USING THE AGILENT 5100 ICP-OES Solutions for Your Analytical Business Markets and Applications Programs Authors Gaurav Kapadnis Vinay Jain Samir Vyas Agilent Technologies India Centre of Excellence (CoE), Manesar, India ABSTRACT This solution note highlights a total solution for the quantitation of trace and toxic elements in drinking water, packaged drinking water and natural mineral water using the Agilent 5100 ICP-OES, following the Bureau of Indian Standards (BIS) Drinking Water Specifications IS 10500, IS 14543 and IS 13428 respectively. In order to meet the required detection limits for As, Se, Sb and Hg, a specific sample introduction system, the Multimode Sample Introduction System (MSIS) was used. Water standard reference material SRM 1643d was analyzed and showed excellent recoveries. INTRODUCTION Due to an increased awareness in food and drinking water safety, regulatory authorities have stringent regulations in place. There are a wide range of elements that have the potential to cause adverse effects on human health if present in drinking water. Heavy metals have become of particular interest in recent years within the framework of environmental investigation. The Bureau of Indian Standards (BIS) sets regulatory limits for the amounts of contaminants in drinking water (IS 10500), packaged drinking water (IS 14543) and natural mineral water (IS 13428). Currently in India, laboratories are performing elemental analysis of drinking water from low µg/l to high mg/l concentrations using a range of analytical techniques, such as Flame (FAAS) or Graphite Furnace Atomic Absorption Spectrometry (GFAAS) for the lower concentration trace elements such as Cd and Pb, while Hydride Generation and Cold Vapor AAS are used for elements such as As, Se, Sb and Hg in the µg/l range. FAAS can require different burners (Air-Acetylene / Nitrous Oxide-Acetylene) for the different elements. Some labs are even performing water pre-concentration to achieve the desired detection limits. There is now a need for a more simplified routine approach with a high confidence in the results. Agilent s 5100 ICP-OES offers the complete package allowing high throughput analysis, with excellent repeatability and precision, that meets the BIS requirements, without sample pre-concentration, enabling laboratories to analyze more samples per day. A list of elements with their maximum allowable concentrations as per the different BIS regulations is shown in Table 1.

ANALYTICAL TECHNIQUE Instrumentation: Aluminium Table 1. Bureau of Indian Standards Regulations Aluminium IS 13428 IS 14543 IS 10500 mg/l Aluminium Al NA 0.03 0.03 Antimony Sb 0.005 0.005 NA Arsenic As 0.05 0.05 0.01 Barium Ba 1.0 1.0 0.7 Boron B 5.0 5.0 0.5 Cadmium Cd 0.003 0.01 0.003 Calcium Ca 100 75 75 Chromium Cr 0.05 0.05 0.05 Copper Cu 1.0 0.05 0.05 Iron Fe NA 0.1 0.3 Lead Pb 0.01 0.01 0.01 Magnesium Mg 50 30 30 Manganese Mn 2.0 0.1 0.1 Mercury Hg 0.001 0.001 0.001 Molybdenum Mo NA NA 0.07 Nickel Ni 0.02 0.02 0.02 Selenium Se 0.05 0.01 0.01 Silver Ag 0.01 0.01 0.1 Sodium Na 150 200 NA Zinc Zn 5.0 5.0 5.0 NA : Not applicable as per BIS Analysis was performed using an Agilent 5100 Synchronous Vertical Dual View (SVDV) ICP-OES. The SVDV configuration has the capability to run in axial, radial, vertical dual view and synchronous vertical dual view modes for full flexibility in the determination of major, minor and trace elements. Unique Dichroic Spectral Combiner (DSC) technology delivers the fastest analyses and the lowest gas consumption per sample. The sample introduction system used was a single-pass glass cyclonic spray chamber, a seaspray nebulizer, white-white peristaltic pump tubing and a standard 1.8 mm injector torch. A Multimode Sample Introduction System (MSIS) was used for the analysis of hydride forming elements (As, Se, Hg and Sb). An SPS 4 autosampler was used in both setups. The instrument uses a solid state RF (SSRF) system operating at 27 MHz to deliver robust plasma capable of excellent long term analytical stability. The Agilent 5100 ICP-OES uses the VistaChip II detector. This is a high speed (1MHz) CCD detector that enables fast warmup, high throughput, high sensitivity, has a large dynamic range of up to 8 orders of magnitude and provides full wavelength coverage from 167 785 nm from a single entrance slit. The instrument operating conditions used are listed in Table 2. 2

Table 2. Agilent 5100 SVDV ICP-OES operating parameters Parameter Read time (s) 5 Replicates 3 Sample uptake delay (s) 25 Stabilization time (s) 10 Rinse time (s) 15 Pump Speed (rpm) 12 Fast pump (rpm) 80 RF power (kw) 1.20 Aux flow (L/min) 1.0 Plasma flow (L/min) 12.0 Nebulizer flow (L/min) 0.7 Viewing height (mm) 8 Background Correction Fitted Setting Value Sample and Standard Preparation: The Calibration standards for As, Hg, Se and Sb were prepared using single element NIST traceable standards. The samples were prepared with conc. HCl and reduction was carried out using 0.6% NaBH4 and 0.5% NaOH. They were analyzed on the Agilent 5100 ICP-OES using the Multimode Sample Introduction System (MSIS). MSIS is a simple plug and play sample introduction system that replaces the existing spray chamber and has the ability to nebulize liquid sample and create volatile hydrides simultaneously. It can be used in one of three modes, as shown in Figure 1. For this work, the four hydride elements were analyzed simultaneously using the MSIS but separately from the other elements. Peristaltic pump tubing for sample and reagent, Black-Black (Agilent catalog # 3710027200), peristaltic pump tubing for drain, Black-White (Agilent catalog # 3710068900). A. Liquid Nebulization Only B. Hydride Generation Only C. Simultaneous Liquid Nebulization and Hydride Generation Figure 1. The Multimode Sample Introduction System (MSIS) operation 3

For other elements, standards were prepared from multi element NIST traceable standards and diluted with 5% HNO3. They were introduced onto the system using the single-pass glass cyclonic spray chamber fitted with the seaspray nebulizer. The accuracy and precision of the method was assessed by analyzing the trace elements in water Standard Reference Material 1643d (Traceable to NIST). The water analysis workflow is depicted in Figure 2. Calibration curves for As, Cd, Pb and Hg are shown in Figure 3. Figure 2. Schematic of water analysis workflow using the Agilent 5100 ICP-OES Figure 3. Calibration Curves for As, Cd, Pb and Hg 4

RESULTS AND DISCUSSIONS Method Detection limits (MDL): The Method Detection Limits (MDL) for each element are shown in Table 3. Elements were measured simultaneously in the axial and radial views by the DSC and excellent detection limits were obtained. Method detection limits (MDLs) were established by analyzing eight replicate injections of the calibration blank and multiplying the obtained standard deviation by three. Table 3. Method Detection limits Element Wavelength (nm) MDL Aluminum 396.152 nm 1.2 Antimony* 206.834 nm 1.7 Arsenic* 188.980 nm 0.5 Barium 455.403 nm 0.1 Boron 249.772 nm 3.0 Cadmium 228.802 nm 0.4 Calcium 422.673 nm 4.0 Chromium 267.716 nm 0.7 Copper 327.395 nm 0.4 Iron 238.204 nm 0.4 Lead 220.353 nm 3.0 Magnesium 285.213 nm 5.1 Manganese 257.610 nm 0.1 Mercury* 194.164 nm 0.2 Molybdenum 202.032 nm 1.2 Nickel 231.604 nm 3.1 Selenium* 196.026 nm 0.1 Silver 328.068 nm 6.0 Sodium 589.592 nm 10.1 Zinc 213.857 nm 1.1 * Using MSIS Standard Reference Material Recoveries The results for the analysis of the standard reference material SRM 1643d are given in table 4. Excellent recoveries for all elements were obtained, demonstrating the method repeatability and capability using the Agilent 5100 SVDV ICP-OES to analyze trace as well as major elements in drinking water. Table 4. Recoveries for various elements in water SRM 1643d Element Certified Value Measured n=3 SD Recovery (%) Al 127.6 129.0 0.89 101.1 Sb 54.10 54.49 9.84 100.7 As 56.02 56.19 1.73 100.3 Ba 506.5 512.3 2.70 101.1 B 144.8 144.0 1.70 99.4 Cd 6.47 6.42 0.44 99.2 Ca 31040 29650 326 95.5 Cr 18.53 19.29 0.72 104.1 Cu 20.50 21.50 0.99 104.9 Fe 91.20 89.1 1.10 97.7 Pb 18.15 17.76 1.86 97.9 Mg 7989 8420 97.2 105.4 Mn 37.66 37.74 0.54 100.2 Mo 112.9 113.3 1.10 100.4 Ni 58.10 57.12 2.75 98.3 Se 11.43 11.22 2.72 98.2 Ag 1.27 1.19 2.79 93.7 Na 22070 23500 243 106.5 Zn 72.48 75.37 0.75 104.0 5

CONCLUSIONS The Agilent 5100 SVDV ICP-OES with DSC combines the sensitivity benefits of axial plasma with the robust qualities of radial plasma into a single platform so that all wavelengths can be detected simultaneously in a single measurement. This gives greater precision, faster analysis times and reduced argon gas consumption. For drinking water analysis, the Agilent 5100 SVDV ICP-OES in combination with the MSIS, offers a total solution that replaces multiple analytical approaches with a simpler, faster, more precise and cost effective approach. It is capable of measuring all required elements in drinking water, whilst meeting the required BIS detection limits of all elements. Sample preparation times are reduced and less sample is required for each analysis. Excellent method detection limits in the μg/l range were obtained for all elements and good recovery results for all elements in SRM 1643d were achieved. REFERENCES 1. Bureau of Indian Standards, IS 10500 (2012): Drinking Water [FAD 25: Drinking Water], with latest amendments. 2. Bureau of Indian Standards, IS 14543 (2004): Packaged Drinking Water (Other than Packaged Natural Mineral Water) [FAD 14: Drinks and Carbonated Beverages], with latest amendments. 3. Bureau of Indian Standards, IS 13428 (2005): Packaged Natural Mineral Water [FAD 14: Drinks and Carbonated Beverages], with latest amendments. 4. John Cauduro, Andrew Ryan, Agilent Technologies. Ultra-fast determination of trace elements in water, conforming to US EPA 200.7 using the Agilent 5100 Synchronous Vertical Dual View ICP-OES, 5991-4821EN 5. Neli Drvodelic, Agilent Technologies. High throughput, low cost analysis of environmental samples according to US EPA 6010C using the Agilent 5100 SVDV ICP-OES, 5991-5921EN Solutions for your analytical business Markets & Applications Programs www.solutions-to-win.com This information is subject to change without notice. Agilent Technologies, Inc. 2016 Printed in EU, 22 January, 2016 5991-6631EN