Spectro-photometric determinations of Mn, Fe and Cu in aluminum master alloys

Similar documents
Analysis of Trace Metals in Tobacco and Tobacco Ash using Atomic Emission Spectroscopy

National Food Safety Standard

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron with 1,10-Phenanthroline

SPECTROPHOTOMETRIC DETERMINATION OF IRON

Background Statement for SEMI Draft Document 4795 (Yellow Ballot) NEW STANDARD: SPECIFICATIONS FOR NITRIC ACID, USED IN PHOTOVOLTAIC APPLICATIONS

Determination of iron content in iron oxide used for fireworks and firecrackers based on Energy Dispersive X-ray Fluorescence Spectrometry (EDXRF

Background The document will be revised for its five year review and updated per the committee agreed upon changes.

SEMI C SPECIFICATIONS AND GUIDELINES FOR AMMONIUM HYDROXIDE

ICP-AES Determination of 15 Kind of impurity elements in the Vanadium-aluminum alloy

2017 International Conference on Materials Science and Biological Engineering (ICMSBE 2017)

ELEMENTS by ICP 7301 (Aqua Regia Ashing)

Analysis of 14 Trace Elements in Stainless Steel

Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric)

DETERMINATION OF SULPHUR CONTENT OF UN-ALLOYED AND LOW-ALLOYED STEEL BY ICP-AES SPECTROMETRY USING WET CHEMICAL SAMPLE PREPARATION

Soil Sampling, Preparation and Analysis

Analysis of steel and its alloys using the GB/T standard and an Agilent 5100 ICP-OES in dual view mode

Rubber and plastics gloves for food services Limits for extractable substances

Experiment 18 Determination of Iron by Visible Spectrophotometry

ABSTRACT. This study investigates the use of inductively coupled plasma-optical emission spectrometry

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB/T Translated English of Chinese Standard: GB/T223.

Measuring Manganese Concentration Using Spectrophotometry

Determination of sodium, calcium, copper, iron and manganese in pulp and paper by atomic absorption spectroscopy (Five-year review of T 266 om-11)

Determination of Iron Content in Different Hemoglobin Samples from Some Patients by UV-Visible Spectrophotometer

IRON (Colorimetric) 2. Muffle Furnace: Equipped with pyrometer and capable of operating at controlled temperatures up to 600 C

SEMI C SPECIFICATIONS AND GUIDELINES FOR HYDROGEN PEROXIDE

ELEMENTS by ICP 7301 (Aqua Regia Ashing)

Standard Test Methods for Rubber from Natural Sources Chemical Analysis 1

Method 5.9 Syrup: calcium and magnesium by EDTA titration

Method for Estimation of Na 2 O and K 2 O in Ores, Fluxes, Coal and Coke Ash by Inductively Coupled Plasma-Atomic Emission Spectroscopy

Experiment 8. Determination of Iron in an Ore by Potentiometric Titration. Iron ores are often completely decomposed in hot concentrated HCl.

Improved Antimony Recoveries in Soil Matrices by 3050B/6010B. DataChem Laboratories, Inc. Robert P. Di Rienzo Jeffery S. Ward John T.

CERTIFICATE OF ANALYSIS

The image part with relationship ID rid4 was not found in the file. Welcome

Analysis of wear metals and contaminants in engine oils using the 4100 MP-AES

Corrosion Behavior of Carbon Steel in Concrete Material Composed of Tin Slag Waste in Aqueous Chloride Solution

The effect of immersion time to low carbon steel hardness and microstructure with hot dip galvanizing coating method

NATIONAL STANDARD OF THE PEOPLE S REPUBLIC OF CHINA (Zhenghua Renmin Gongheguo Guojia Biaozhun)

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

Some Basic Concepts of Chemistry

Disclosure to Promote the Right To Information

Ultra-fast ICP-OES determinations of base metals in geochemical samples using next generation sample introduction technology

Background Statement for SEMI Draft Document 5134 Revision to SEMI C , Specifications for Hydrofluoric Acid

Determination of Elemental Impurities in Graphite-based Anodes using the Agilent 5110 ICP-OES

METHOD 7550 OSMIUM (ATOMIC ABSORPTION, DIRECT ASPIRATION)

X-Ray Fluorescence Measurements of Molten Aluminum Elemental Composition

ISO INTERNATIONAL STANDARD. Iron ores Determination of various elements by X-ray fluorescence spectrometry Part 1: Comprehensive procedure

The Determination of Trace Elements in Stainless Steel by Forked Platform GFAAS

Assaying and Analysis Choices in Practice

ISO INTERNATIONAL STANDARD

1. Introduction. 2. Scope and application. 3. Objectives

ISO INTERNATIONAL STANDARD

Optical spectroscopy of sodium silicate glasses prepared with nano- and micro-sized iron oxide particles

EXPERIMENT III. Determination of Iron in Iron Oxide, (Fe 2 O 3 ), Using Dichromate Method. Chemical Overview

ANALYTICAL PERFORMANCE OF A HANDHELD EDXRF SPECTROMETER WITH MINIATURE X-RAY TUBE EXCITATION

ISO INTERNATIONAL STANDARD

Background Statement for SEMI Draft Document 4546A REVISION OF SEMI C , with title change to SPECIFICATIONS FOR HYDROGEN PEROXIDE

Thin-Layer Chromatography of Aluminium: Quantitative Densitometric Determination of Fe 2+, Ni 2+, Cu 2+, and Si 4+

Determination of the Mass Percentage of Copper in a Penny

Fields of Application / Industry:

IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT:

Analysis of Copper and Copper Base Alloys, Using Shimadzu PDA-7000

Analysis of Cast Iron Using Shimadzu PDA-7000

Metals Analyzer. OES 6000 Optical Emission Spectrometer. fast and accurate metal analysis

Method 108C - Determination of Arsenic Content in Ore Samples From Nonferrous Smelters (Molybdenum Blue Photometric Procedure)

MICROSTRUCTURAL CHARACTERIZATION OF MODIFIED COMMERCIAL 2219 ALUMINUM ALLOY

The following are the completed but unbalanced equations. Each equation is numbered to match each step of the cycle:

Multi-Element Analysis of Trace Metals in Animal Feed using ICP-OES

SPECTRAL INTERFERENCE IN X-RAY FLUORESCENCE ANALYSIS OF COMMON MATERIALS

R = kc Eqn 5.1. R = k C Eqn 5.2. R = kc + B Eqn 5.3

GB/T Translated English of Chinese Standard: GB/T

DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, & USING THE AGILENT 5100 ICP-OES

Chapter 8. Gravimetric Analysis

ELEMENTS by ICP 7300 (Nitric/Perchloric Acid Ashing)

SurTec 712 Alkaline Cyanide-Free Zinc/Iron Process

ANALYTICAL REFERENCE MATERIALS INTERNATIONAL Part of LGC Standards 3 Perimeter Road, Unit 9 Manchester, NH USA

Institute for Reference Materials and Measurements Community Reference Laboratory for Feed Additives

Increased Sample Throughput for ICP-OES Applied to U.S. EPA Method 200.7

ElvaX Light for Petrochemical analysis

Determination of Ferrous Iron in EASTMAN Color Films, KUL Bleach

Method 3.8 Juices: phosphates (P 2 O 5 )

Certificate of Analysis

DETERMINATION OF SILVER, GOLD AND COBALT IN SULFIDE ORES AND PRODUCTS OF THEIR PROCESSING BY ICP-AES

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Surface formation in direct chill (DC) casting of 6082 aluminium alloys

ROLE OF SOLUTE AND TRANSITION METALS IN GRAIN REFINEMENT OF ALUMINUM ALLOYS UNDER ULTRASONIC MELT TREATMENT

SurTec 717 Alkaline Zinc/Nickel Electroplating Process (Electrolyte based on Sodium)

Researches on the production of self-reducing briquettes from waste containing iron and carbon

The Determination of Alloying Elements in High-Carbon Steel using Radially Viewed Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES)

Sample preparation for X-ray fluorescence analysis V. Fusion bead method part 2: practical applications

Standard Solutions (Traceable to NIST)

THE MOLECULAR WEIGHT OF A SOLUTE*

Standard Solutions (Traceable to NIST)

ElvaX ProSpector in Exploration & Mining

Metallurgy, Alloys, and Applications p. 1 Introduction and Overview p. 3 Major Groups of Copper and Copper Alloys p. 3 Properties of Importance p.

O 3, SIO 2. O, CAO, Al 2 AND MG IN ANT-HILL SOIL SAMPLES WITHIN ABRAKA TOWN IN NIGERIA

Ginger extract as green corrosion inhibitor of mild steel in hydrochloric acid solution

Effects of Gamma Radiation on Iron (III) Complex with Sodium Salicylate in Aqueous Solutions

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Manufacturing and STA-investigation of witness-samples for the temperature monitoring of structural steels under irradiation

Rev Experiment 10

Transcription:

IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Spectro-photometric determinations of Mn, Fe and Cu in aluminum master alloys To cite this article: Rehan, A. Naveed et al 2016 IOP Conf. Ser.: Mater. Sci. Eng. 146 012048 View the article online for updates and enhancements. This content was downloaded from IP address 37.44.199.180 on 08/01/2018 at 16:52

Spectro-photometric determinations of Mn, Fe and Cu in aluminum master alloys A. Rehan, A. Naveed, A. Shan 1, M. Afzal, J. Saleem, and M. A. Noshad Soan Enterprises, P. O. Box 2723, Islamabad, Pakistan 1 Department of Physics, University of Engineering and Technology, Lahore, Pakistan E-mail: rehanchatta@gmail.com, chathashan@gmail.com Abstract. Highly reliable, fast and cost effective Spectro-photometric methods have been developed for the determination of Mn, Fe & Cu in aluminum master alloys, based on the development of calibration curves being prepared via laboratory standards. The calibration curves are designed so as to induce maximum sensitivity and minimum instrumental error (Mn 0.1mg/100ml-2mg/100ml, Fe 0.01mg/100ml-0.2mg/100ml and Cu 2mg/100ml-10mg/ 100ml). The developed Spectro-photometric methods produce accurate results while analyzing Mn, Fe and Cu in certified reference materials. Particularly, these methods are suitable for all types of Al-Mn, Al-Fe and Al-Cu master alloys (5%, 10%, 50% etc. master alloys).moreover, the sampling practices suggested herein include a reasonable amount of analytical sample, which truly represent the whole lot of a particular master alloy. Successive dilution technique was utilized to meet the calibration curve range. Furthermore, the workout methods were also found suitable for the analysis of said elements in ordinary aluminum alloys. However, it was observed that Cush owed a considerable interference with Fe, the later one may not be accurately measured in the presence of Cu greater than 0.01 %. 1. Introduction Aluminum master alloy is the aluminum combined with one or two other elements (e.g., Al-Cu 50 % master alloy) and is used as a CHARGE for alloying during the melting of wrought aluminum alloys. Normally aluminum master alloys are produced for such elements that are the alloying elements of Aluminum wrought alloys. Commonly used Aluminum master alloys are Al-Mn, Al- Si, Al-Ni, Al-Cu, Al-Mg, Al-V, Al-Zn, Al-Zr, Al-Ti, Al-Cr and Al-Fe and are available in 5 or 7 Kg ingots. The use of Aluminum master alloys offers distinct advantages over the addition of individual element during alloying of aluminum wrought alloys. Three major advantages are: (a) these can be calculated to become part of the initial cold charge i.e. melted with the metal at low temperature. This rapid dissolution results in a significant reduction in dross formation, minimal gas pick up and extended furnace life (b) significantly reduces the amount of energy required and shortens the length of melt time (c) gives higher accuracy to meet specified composition limits [1-2]. The above mentioned benefits can only be attained by knowing the true composition of a particular master alloy [3]. Instantly, master alloy manufacturer provides a broad range of chemical composition for a single lot of particular master alloy, for example, in Al-Cu 50 % master alloy, Cuelement may varies from 47 to 52% or else as affirmed by the manufacturer. On the other hand for charge calculation one should know the nearest possible true composition of a particular master alloy so that melt could be adjusted as quickly as possible to save time, energy, and material. Also, chemical Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. Published under licence by Ltd 1

composition analysis of master alloys is indispensable for the strict quality control of master alloys production. Spark optical emission spectroscopy (OES) is a quick and reliable tool for analysis of aluminum alloys but it does not work for master alloys because of high percentage of alloy elements (e.g., Al-Cu 50% master alloy)[6]. Other techniques such as: Gravimetric technique, Atomic Absorption Spectrometry (AAS), Atomic Emission Spectrometry (AES), and X-ray fluorescence spectroscopic (XRF) technique are also not valid for such a high concentrations [4-8]. Previously Naveed et al [3] reported a method in which Aluminum master alloys were re-melted along with Aluminum ingots of 99.9% purity in an electrical crucible furnace and subsequently analyzed by spark optical emission spectrometer (OES). However, the technique was costly. Also, the obtained analysis results were significantly dependent upon recovery of a certain element in a particular melting process. Spectro-photometry is a technique that uses the absorbance of light by an analyte (the substance to be analyzed) at a certain wavelength to determine the analyte concentration. Spectro-photometric analysis is based on development of calibration curve: several standards containing exactly known concentrations of the analyte are introduced into instrument, and the instrumental response is recorded. Ordinarily this response is corrected for instrument output obtained with a blank. Ideally, the blank contains all of the components of the original sample except for the analyte. The resulting data are then plotted to give a graph of corrected instrument response versus analyte concentration, and this graph in turn can be used to find the concentration of an unknown [9-10]. In this contribution we presented a cost effective and reliable spectro-photometric method for the composition analysis of aluminum master alloys. Analysis of Al-Mn, Al-Fe, & Al-Cu master alloys using the calibration curve of Mn, Fe, & Cu are reported. Furthermore, the designed methods are verified by analyzing certified reference materials () containing Mn, Fe, & Cu against their calibration curves. 2. Experimental methods 2.1. Development of calibration curve of Mn A calibration curve ranging from 0.1mg/100ml to 2mg/100mlMn was obtainedaccording to ASTM standard [11].This was done by transferring 1.000 g of pure Mn metal powder(weighed accurately to 3 rd decimal) to a 500-ml beaker and dissolved in 100 ml of H₂SO₄ (1+9 v/v). The solution was then boiled for 5 minutes and transferred to 1 liter volumetric flask (when cool down), diluted to the volumeand mixed. This 100mg/100ml solution was named as Mn standard solution-a. From the Mn standard solution-a transferred 100 ml in 1 liter volumetric flask, diluted to the volume and mixed. This 10mg/100ml solution was named as Mn standard solution-b. Calibration solutions were prepared by transferring0 (reagent blank), 1.0, 2.0, 5.0, 10.0, 15.0, and 20.0 ml of Mn Standard Solution B (1 ml = 0.1mg Mn (10mg/100ml)) to 100-mlvolumetric flasks. To each 15 ml of H₂SO₄ (1+1 v/v), 10 ml of HNO₃, and 5 ml of H₃PO₄ were added and Diluted to about 70 ml. For color development, the solutions were heated to boiling in a water bath and 20 ml of KIO 4 (25 g/l) solution was added in each. The boiling process was continued for an hour until the characteristic color developed. The solution was then cooled to room temperature and diluted to volume with potassium iodated water (water free from reducing agents). Photometric measurements were made at approximately 525 nm against the reagent blank. 2.2. Development of calibration curve of Fe A calibration curve ranging from 0.02mg/100ml to 0.2mg/100mlfor Fe was obtained according to ASTM standard [11]. This was done by transferring 0.1000 g of pure Fe metal powder (weighed accurately to 3 rd decimal) to a 500-ml beaker and dissolved in 50 ml of H₂O, 25 ml HCL, 1ml HNO 3. The solution was then heated to dissolve Fe and transferred to 1 liter volumetric flask (when cool down), diluted to the volume and mixed. This 10mg/100ml solution was named as Fe standard solution-a. 2

From the Fe standard solution-a transferred 100 ml in 500 ml volumetric flask, added 10 ml HCL, diluted to the volume and mixed. This 2mg/100ml solution was named as Fe standard solution-b. Calibration solutions were prepared by transferring 0 (reagent Blank), 1.0, 2.0, 4.0, 7.0, 10.0ml of Fe Solution B (1 ml = 0.02 mg Fe) to five 100-ml volumetric flasks. Diluted to 60 ml and for color development : added 8mlNH₂OH HCl (100g/L) solution, 10ml Sodium acetate-acid (272g/L) buffer solution and 10ml of1,10-phenanthroline(2g/l) solution, diluted to volume, mixed and allowed to stand for 5 min. Photometric measurements were made at approximately 510 nm against the reagent blank. 2.3. Development of calibration curve of Cu 0.1g of pure Cu metal powder (weighed accurately to 3rd decimal) was dissolved in a 5ml of nitric acid and diluted to 100 ml in a volumetric flask (100mg/100mlstock Cu solution). 2ml, 4ml, 6ml, 8ml, and 10mlof the stock Cu solution were transferred in five different volumetric flasks of100ml. After that the ammonia solution was gradually added in each volumetric flask, resulting in a change of color(the color turns to dark blue). Adding ammonia solution was continued until a stabilized color was achieved (i.e., no further color change occurs by adding more ammonia solution). Additional 3mlof ammonia solution was added in each and diluted to the volume. Photometric measurements were made at approximately 610 nm against distilled water as a reagent blank. 2.4. Determination of Mn, Fe, and Cu in aluminum master alloys After the development of calibration curves, some test runs were conducted to characterize the calibration curves. This was done by testing Mn, Fe and Cu in certified reference materials ( s). Sample preparation was started with the chipsformation of s using lathe machine. Similarly, master alloy samples were collected in chips form by drilling several master alloy ingots of a single lot (e.g., Al-Mn 10 % master alloy lot) and mixed them.the test and master alloy solutions were prepared in the same way as that of calibration solutions but using successive dilutions to meet with the ranges of Mn, Fe and Cu calibration curves. 3. Results and discussion Calibration standards for Mn, Fe and Cu were prepared and their response in the visible region of electromagnetic spectrum was recorded. Figures 1-3 represent the corresponding calibration curves. Figure 1.Calibration curve of Mn. Photometric measurements were made at 525 nm Figure 2.Calibration curve of Fe. Photometric measurements were made at 510 nm 3

Figure 3. Calibration curve of Cu. Photometric measurements were made at 610 nm At the outset the test run analysis is carried out in order to characterize the calibration curves. This is done by testing Mn (wt. %) in Certified Reference Materials ( s) against the calibration curve prepared for Mn. The photometric calculations and analysis results of Mn are shown in Table 1. It may be seen from Table 1 that the results of photometric analysis of Mn in s (49712b-4 & A333.O AE) are in well agreement with certified values. Also the spectro-photometric results of Mn matches good with spark OES results. These results standardize the characteristic calibration curve of Mn. Table-1 also enlists the spectro-photometric calculations and analysis results of 5%, 10% and 20% Al- Mn master alloys. The Mn % is within the prescribed range for all types of master alloys. Sample Identity Alcan International Ltd. ALC-A 333.0-AE Bureau of Analysed Samples (BAS) Ltd. NCS HS 49712b-4 Table 1.Spectro-photometric data for analysis of Mn Known Value, Mn Wt.% Sample weight/dilution for photometry Concentration found by photometry (mg/100ml) Result, Mn Wt.% 0.32 0.1g/100ml 0.333 0.333 1.93 0.05g/100ml 0.967 1.934 AA 2024 0.601 0.1g/100ml 0.609 0.609 Mn 10% MA 9-11 0.1g/20x100ml 0.507 10.14 Mn 20% MA 18-22 0.1g/20x100ml 1.024 20.48 Where, =certified reference material, MA=master alloy, x=multiplication The photometric calculations and analysis results of Fe are represented in Table 2. It may be seen from Table 2 that the results of photometric analysis of Fe in (49712b-4) are close tothe certified value. However for (A333.O AE) there is a large deviation in spectro-photometric analysis results as compared to certified value. This happens because (A333.O AE) contains high % of 4

Cu (3.5%) that interferes with Fe [11]. Also the spectro-photometric results of Fe for AA1050 are closed to OES value because of low % of Cu. These results standardize the characteristic calibration curve of Fe and express that the Fe can only be determined accurately in the absence of Cu. The spectro-photometric calculations and analysis results for 10% Al-Fe master alloy are also shown in Table 2. The Fe % is within the prescribed range. Table 2.Spectro-photometric data for analysis of Fe Sample Identity known value, Fe Wt.% Sample weight /Dilution for photometry Concentration found by photometry (mg/100ml) Result, Fe Wt.% Bureau of Analysed Samples (BAS) Ltd. NCS HS 49712b-4 Alcan International Ltd. ALC-A 333.0-AE 0.242 0.78 0.1g/2x100ml 0.111 0.222(Cu=0.057) 0.1g/5x100ml 0.051 0.255* AA 1050 0.37 0.1g/10x100ml 0.033 0.33 (Cu= 0.02) Fe 10% MA 9-11 0.1g/100x100ml 0.101 10.1 Where, =certified reference material, MA=master alloy, *low reading due to Cu (3.53%) interferes, x=multiplication First the Cu (wt. %) is determined in against the calibration curve prepared for Cu. The photometric calculations and analysis results of Cu are shown in Table 3. It may be seen from Table 3 that the results of photometric analysis of Cu in (A333.O AE) are just close to the certified value. Also the spectro-photometric results of Cu for AA2219 do not match well with OES results.this variation as compared to the Certified or OES value is due to the reason that the absorption coefficient for copper tetra amine complex is very small (Figure 3. calibration curve of Cu). The spectro-photometric calculations and analysis results for 10% and 50% Al-Cu master alloys are also shown in Table 3. The Cu % is within the prescribed range but lies away from central value. Again this is because the absorption coefficient for copper tetra amine complex is very small which probably increases the instrumental error. Table 3.Spectro-photometric data for analysis of Cu Sample Identity Known value, Cu Wt.% Sample weight /Dilution for photometry Concentration found by photometry (mg/100ml) Result, Cu Wt.% Alcan International Ltd. ALC-A 333.0-AE 3.53 0.5g/5x100ml 3.3 3.3 AA 2219 6 0.5g/5x100ml 6.23 6.23 Cu 10% MA 9-11 0.5g/5x100ml 9.5 9.5 Cu 50% MA 47.5-52.5 0.1g/5x100ml 9.42 47.1 Where, =certified reference material, MA=master alloy, x=multiplication 5

4. Conclusions Chemical compositional analysis of Al-Mn, Al-Fe, and Al-Cu master alloys are necessary to adjust the melt chemistry quickly. Also, it is indispensable for the strict quality control of master alloys production. The presented spectro-photometric methods are reliable and cost effective. Particularly, these spectro-photometric methods cover all the ranges for Mn, Fe and Cu determinations in aluminum master alloys. Also, the sampling practices (chips formation and mixing)suggested here easily include a large number of master alloy ingots, whose average results are the true representatives of the whole lot. 5. Acknowledgement The authors are deeply indebted to Mr. Muhammad Waseem and Mr. Muhammad Awais for their assistance during the preparation of manuscript and proof reading. 6. References [1] Totten G E and Mackenzie D S 2003 Handbook of Aluminum: Alloy production and materials manufacturing(marcel Dekker Inc. New York)2 [2] Moldovan P, Popescu G, Popescu C A and Apostolescu I2007Advanced Materials Res.23 295. [3] Naveed A 2010 Pak. J. Anal. Environ. Chem. 11 2 51-55 [4] ASTM E60.98 1998 Standard practice for analysis of metals, Ores and related materials by molecular absorption spectrometry [5] ASTM E1452.92 1996 Standard practice for preparation of calibration solutions forspectrophotometricanalysis [6] ASTM E1282.98 1998 Standard guide for specifying the chemical compositions and selecting sampling practices and quantitative analysis methods for metals [7] Ingle J D and Crouch S R 1988 Spectro chemical Analysis (Prentice Hall International, Inc. New Jersey) 257 [8] Jenkins R, Gould R W and Gedcke D1995Quantitative X-Ray Spectrometry (Marcel DekkerInc. New York) [9] Grieken R E V and Markowicz A A 2002Handbook of X-Ray Spectrometry(Marcel Dekker Inc. New York) [10] ASTM E169.99 1999 General Techniques of Ultraviolet-Visible Quantitative Analysis. [11] Jeffery G H, Bassett J, Mendham J, Denney R C 1989 quantitative chemical analysis(longman Group UK Limited) [12] ASTME34.94 2002 Standard test methods for chemical analysis of aluminum and aluminumbase alloys 6