Improved Peptide Maps Using Core-Shell Media: Explaining Better Biopharmaceutical Applications With Kinetex C18 Columns

Similar documents
Biotherapeutic Method Development Guide

Biotherapeutic Method Development Guide

Novel Wide-Pore Superficially Porous Particles for Biomacromolecular Separations

Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

Discovery BIO Wide Pore

Sean M. McCarthy and Martin Gilar Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL RESULTS AND DISCUSSION

BioHPLC columns. Tim Rice Biocolumn Technical Specialist

Bivalirudin Purification:

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Peptide Mapping: A Quality by Design (QbD) Approach

AdvanceBio Peptide Mapping

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

Chromatography column for therapeutic protein analysis

4/4/2013. BioHPLC columns. Paul Dinsmoor Biocolumn Technical Specialist. April 23-25, Size Exclusion BioHPLC Columns

Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS

Application Note. Authors. Abstract. Biopharmaceuticals

Improving Sensitivity for an Immunocapture LC-MS Assay of Infliximab in Rat Plasma Using Trap-and-Elute MicroLC-MS

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

ProSwift and PepSwift Monolith Columns for Biomolecule Analysis

Method Development Considerations for Reversed-Phase Protein Separations

Reverse-Phase Separation of Proteins, Peptide and Other Biomolecules. Bill Long Applications Engineer October 8, 2008

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Method Optimisation in Bottom-Up Analysis of Proteins

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

Application Note. Abstract. Author. Biopharmaceutical

NISTmAb characterization with a high-performance RP chromatography column

Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

HALO BioClass Gives You the Right Column for Every Situation

Peptide Mapping. Hardware and Column Optimization

Reproducible LC/MS Peptide Separation Using Agilent AdvanceBio Peptide Plus Columns

2015 CATALOG. Discover the Advantages of HALO and HALO BioClass Fused-Core Columns

A Unique LC-MS Assay for Host Cell Proteins(HCPs) ) in Biologics

columns P r o P a c H I C C o l u m n S o l u t i o n s f o r P r o t e i n A n a l y s i s

Charged Surface Hybrid C18 for High Resolution LC and LC/MS Peptide Separations

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

Promix TM. Enter a New Era in Biomolecule Analysis with. Columns. Unsurpassed Selectivity and Peak Capacity for Peptides and Proteins

Profiling Glycosylation of Monoclonal Antibodies at Three Levels Using the Agilent 6545XT AdvanceBio LC/Q TOF

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

columns PepSwift and ProSwift Capillary Monolithic Reversed-Phase Columns

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns

Developing Quantitative UPLC Assays with UV

Optimization of the Effective Separations for Peptides and Proteins Using High Durable Packing Materials for HPLC. YMC CO., LTD.

Characterize mab Charge Variants by Cation-Exchange Chromatography

Zwitterion Chromatography ZIC

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

MAbPac RP Column. High-performance reverse phase chromatography column for monoclonal antibody analysis

mabs and ADCs analysis by RP

Reversed Phase Solutions for the Analysis of Proteins, Peptides, and Oligonucleotides

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

Size Exclusion BioHPLC columns Ion Exchange BioHPLC columns

Faster Separations Using Agilent Weak Cation Exchange Columns

Development of Analysis Methods for Therapeutic Monoclonal Antibodies Using Innovative Superficially Porous Particle Biocolumns

Critical Quality Attribute Assessment by Peptide Mapping Using LC/MS with Superficially Porous Columns

Characterization of a fusion protein under native and denaturing conditions with maxis II

Comprehensive Single-Shot Proteomics with FAIMS on a Hybrid Orbitrap Mass Spectrometer

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

New Core-Shell Technology

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis

Fast Agilent HPLC for Large Biomolecules

Barry Boyes 1,2, Tim Langlois 1, Brian Wagner 1, Stephanie Schuster 1 and Joe DeStefano 1

OptiFlow Quant Solution for High-Sensitivity Quantitation

Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins

Agilent OpenLAB CDS MatchCompare for the Comparison of Peptide Mapping Samples

The Agilent 1260 Infinity BioInert Quaternary Pump. Scope of a low-pressure mixing UHPLC pump with Bio-Inert Capabilities

for Acclaim Carbamate Column

TSK-GEL BioAssist Series Ion Exchange Columns

Analysis and Purification of Polypeptides by Reversed-Phase HPLC

Peptide enrichment and fractionation

Nano LC at 20 nl/min Made Easy: A Splitless Pump Combined with Fingertight UHPLC Nano Column to Boost LC-MS Sensitivity in Proteomics

Application Note. Authors. Abstract. Ravindra Gudihal Agilent Technologies India Pvt. Ltd. Bangalore, India

See. Beyond HALO. and Sigma-Aldrich Biotechnology. Advanced Materials Technology. Ascentis. Express

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs)

Method Translation in Liquid Chromatography

Agilent Biocolumns. Peptide Mapping. Application Compendium

CQA Assessment by Peptide Mapping Using LC/MS with an AdvanceBio Peptide Plus Column

Universal Solution for Monoclonal Antibody Quantification in Biological Fluids Using Trap-Elute MicroLC-MS Method

Analysis of Monoclonal Antibodies and Their Fragments by Size Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI

Mobile Phase Optimization in SEC Method Development

A Brief Overview of HPLC & UHPLC Method Development and Optimization. Dr. Chris Message UHPLC/HPLC Product Specialist Phenomenex

Intact & Subunit Purity

Improve Your Analyses of Large Proteins, Antibodies and Antibody-Drug Conjugates Using Reversed-Phase Liquid Chromatography by Maximizing Pore Access

Rapid Extraction of Therapeutic Oligonucleotides from Primary Tissues for LC/ MS Analysis Using Clarity OTX, an Oligonucleotide Extraction Cartridge

Biotherapeutic Peptide Mapping Information Dependent Acquisition (IDA) Method

Biotherapeutic Non-Reduced Peptide Mapping

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Higher Order mab Aggregate Analysis using New Innovative SEC Technology

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Fast protein separations with ion exchange columns

Superficially porous particles with carbon core and nanodiamond polymer shell for protein separations. Szabolcs FEKETE, Davy GUILLARME

Transcription:

Improved Peptide Maps Using Core-Shell Media: Explaining Better Biopharmaceutical Applications With Kinetex C18 Columns Michael McGinley, Jeff Layne, and Vita Knudson Phenomenex, Inc., 411 Madrid Ave.,Torrance, CA 90501 USA PO72760110_L_5

Introduction Other posters presented at this meeting show many performance examples of how the Kinetex core-shell media delivers sub-2 µm efficiency, better resolution, and reduced run times versus fully porous silica media, all while operating at backpressures amenable for operation using standard HPLC systems. While reduced run time and increased throughput are big motivators for using next generation LC columns in small molecule separations, bioanalytical separations require the increased resolution and sensitivity ultra-high performance LC columns offer for the analysis of closely related peptides in low concentrations. Peptide mapping is the principal bioanalytical application where resolution is important; proteins are complex biopolymers that often have post translational modifications (PTM s) that can affect the activity and immunogenicity of a protein therapeutic. As such modifications are only a slight chemical change in the overall protein; proteins must be enzymatically fragmented into dozens or hundreds of peptides that are then separated by reversed phase HPLC and positively identified by LC/MS. Good resolution of closely related peptides is critical for identifying PTM s and next generation HPLC columns with higher efficiency have allowed for more accurate information from complex protein therapeutics like Ig-G proteins which represent a majority of candidates currently in clinical trials. However, high efficiency is not the only important parameter in peptide mapping of complex proteins. HPLC medias must have high selectivity of minute chemical differences, have acceptable loading of analytical amounts of digested protein so that low level PTM s can be detected, and media must maintain good performance in MS-compatible buffers used for such analyses. Kinetex, a recently introduced core-shell media, was evaluated against popular fully porous media as well as other superficially porous media in regards to peptide mapping performance, peptide loading, performance in volatile buffers and with Ig-G peptide separations to demonstrate the advantages of such a media over existing chromatographic solutions

Materials and Methods Recombinant and natural proteins and peptides, dithiothreitol (DTT) and iodoacetic acid (IAA) were purchased from Sigma Chemicals (St. Louis, MO, USA). Recombinant human protein Interferon-alpha was obtained from Abcam, Inc. (Cambridge, MA, USA). Solvents were purchased from Fisher Scientific (Fairlawn, NJ, USA). Mobile phase modifiers and Asp-N endoproteinase were obtained from EMD (San Diego, CA). Lys-C endoproteinase was obtained from Wako Chemicals (San Diego, CA). Sequencing grade trypsin was purchased from Roche Applied Science (Indianapolis, IA, USA). Human Ig-G2 was obtained from Dako (Carpentaria, CA). Kinetex 2.6 µm C18 columns (100 or 150 x 4.6 mm) and a wide-pore 3 µm C18 were used for HPLC analysis (Phenomenex, Torrance, CA). HALO C18 columns were obtained from Mac Mod (Chadds Ford, PA) and XBridge C18 columns were purchased from Waters (Milford, MA).

Experimental For BSA digests, protein was digested with sequencing grade trypsin 1:50 E/S in 0.1 M NH 4HCO 3 ph 8.0 at 37 C for 18 hours then quenched with TFA or formic acid before injection. Different gradients of TFA/water or formic acid/ water from 5 to 45 % B (Acetonitrile) over 30 minutes were used based on the application shown. For peptide mixtures, fast gradients from 5 to 50 % in 5 minutes were used. For Human Ig-G, protein was digested with Lys-C for 18 hours at 37 C (E/S 1:30) in 4 M guanidine/ 0.1 M ammonium bicarbonate. Sample was diluted to 1 M guanidine then digested with Asp-N for 12 hours (E/S 1:50). The reaction was quenched with TFA and 20 µg aliquots were injected on HPLC. HPLC analyses were performed on an Agilent 1100 HPLC equipped with an autosampler, column oven, and MWD using ChemStation software (Agilent, Santa Clara, CA). The mobile phases used were: A=0.1% TFA in water/ 2 % acetonitrile and B=0.085 % TFA in acetonitrile with a gradient from 2 to 45 % B in 30 minutes. A flow rate of 1 ml/min was used and peptide elution was monitored at 214 nm.

Experimental A direct comparison of a BSA tryptic digest run on the Kinetex 2.6 µm C18 column versus a fully porous XBridge 3.5 µm C18 column using identical running conditions on a standard HPLC system is shown in Figure 1. The improved efficiency of the Kinetex core-shell column is readily apparent with significantly narrower peak widths and increased peak height versus the fully porous media. Upon closer inspection one can also see the better selectivity that the Kinetex media delivers over the gradient with over double the peak count (63 vs. 30) that the fully porous media delivers. Such results demonstrate the utility of the Kinetex column for peptide mapping. Improved diffusion is one of the reasons why Kinetex media far outperforms fully porous media of similar size; another relevant performance advantage that improved porosity of Kinetex offers relates to the performance of large peptide/ small proteins on coreshell media as is shown in Figure 2. Small proteins above 10 Kilodaltons tend to perform poorly with badly tailing peaks on fully porous media. When proteins as large as 16 Kilodaltons are analyzed on Kinetex, good peak shape is maintained probably due to the improved porosity that core-shell media delivers. This is especially useful for peptide mapping of large proteins where some glycosylated peptides can be especially large. While core-shell media demonstrates better porosity than fully porous media, one area of concern relates to potential reduced loading capacity of such media due to their lower surface area. When different analytical loading levels of peptides on Kinetex are evaluated (Figure 3) against high surface area fully porous media, Kinetex core-shell media still delivers narrower peak widths than fully porous media for peptide in excess of a 10 µg loads on a 4.6 mm ID column. Such high analytical loading demonstrates the superior performance of Kinetex coreshell for peptide mapping applications where loading might be a concern. Another relevant concern for any high performance separation relates to the inertness of a media and its effect on the peak shape of basic peptides. Inertness of a media can be a significant concern when LC/MS compatible volatile (non-ion pairing) buffers are used. Figure 4 demonstrates the inertness of Kinetex over other core-shell porous media (HALO ) when peptide maps of apomyoglobin are compared. Note that Kinetex chromatograms maintain good peak shape in formic acid LC/MS buffer; HALO demonstrates dramatic loss of selectivity and peak shape when volatile buffers are used. While phase bonding plays a likely role in media inertness, particle morphology may also be a factor. Figure 5 shows different electron micrographs of Kinetex and HALO particles.

Figure 1. 0 20 40 60 XBridge APP ID 19079 0 10 20 30 min 0 20 40 60 Kinetex APP ID 19078 0 10 20 30 min Figure 1: BSA tryptic digests run on a 2.6 µm core-shell C18 column versus a fully porous 3.5 µm column. Note the dramatic improvement in efficiency and peak height for the Kinetex core-shell column as well as a doubling in peak count versus the fully porous media.

Figure 2. 200 175 150 125 100 75 Met-Enkephalin (547) Angiotensin I (1,100) Insulin (5,700) 50 25 0 Cytochrome c (12,000) Myoglobin (16,700) 1 1.25 1.5 1.75 2 2.25 2.5 2.75 3 min Figure 2: Large peptides/ small proteins were run on a core-shell Kinetex 2.6 µm C18 column. Note the reasonable peak width for myoglobin compared to smaller peptides. The good peak shape for myoglobin highlights an additional benefit of improved mass transfer; better performance for large peptides is much greater than expected for a fully porous 90 Å media.

Figure 3. 1200 10 0.03 Kinetex 2.6 µm C18 HALO 2.7 µm C18 100 6 0.025 Gemini -NX 3 µm C18 Poly. (Kinetex 2.6 µm C18 ) 800 600 400 200 1 2 4 Peak Width 0.02 0.015 Peak Width Poly. (HALO 2.7 µm C18) Expon. (Gemini-NX 3 µm C18) 0.01 0 2.4 2.45 2.5 2.55 2.6 2.65 2.7 min 0.005 0.1 1 10 Mass of Met-Enkephalin on column (µg) Figure 3: Peptide loading studies on a core-shell Kinetex 2.6 µm C18 column. The chromatogram shows increasing loads of met-enkephalin from 1 to 10 µg and the peak width at different loading is graphed against a high surface area fully porous 3 µm media and a competitor fused-core media. Note that even at high loads (>10 µg) the peak width is narrower than the fully porous media suggesting that analytical loading is not a concern with core-shell media like Kinetex.

Figure 4. 140 120 Kinetex HALO 140 120 100 100 80 60 80 60 40 40 23.111 20 20 2.050 0 0 2.5 5 7.5 10 12.5 15 17.5 20 min 2.5 5 7.5 10 12.5 15 17.5 20 min Figure 4: A comparison of peptide maps of apo-myoglobin with different core-shell media using LC/MS friendly buffers. Note that the Kinetex column maintains good peak shape and resolution when using volatile buffers. This data suggests that the Kinetex C18 media is more inert than other core-shell media and better suited for LC/MS applications.

Figure 5. HALO Kinetex Figure 5: Electron micrographs comparing Kinetex core-shell media to HALO Fused-Core media. Kinetex media is shown on the right. Note the spherical shape and improved surface smoothness of the Kinetex media.

Figure 6. 600 3 µm C18 Column 400 Kinetex 2.6 µm C18 200 5 10 15 20 25 30 min Figure 6: Ig-G Lys-C/ Asp-N peptide map on a 3 µm wide-pore C18 (red) and Kinetex 2.6 µm C18 (blue). Note the large number of peaks generated from the 150 KDa Ig-G protein and the incomplete resolution of several peptide on the 3 µm C18 column (red). For the Kinetex 2.6 µm C18 column note the dramatically improved resolution of many Ig-G peptide peaks compared to the 3 µm column. Kinetex provides ultra-high resolution of complicated peptide maps run on a standard HPLC system.

Conclusion Other presentations on core-shell Kinetex media have demonstrated its utility for small molecule applications; however have not addressed its capabilities for peptide mapping and other bioanalytical applications. Results obtained in these studies clearly show how the better efficiency, selectivity, loading and inertness make core-shell Kinetex a superior solution for peptide mapping and other bioanalytical applications. A final example of this is shown in Figure 6 where the improved performance of Kinetex over a fully porous media for an Ig-G peptide map is readily apparent, even on a standard HPLC system. Trademarks Kinetex and Gemini are registered trademarks of Phenomenex, Inc. XBridge is a trademark of Waters Corporation. HALO and Fused-Core are registered trademarks of Advanced Materials Technology, Inc. Disclaimer Comparative separations may not be representative of all applications. Columns used for comparison were purchased directly from the original manufacturer. Phenomenex is not affiliated with Waters Corp. or Advanced Materials Technology. 2010 Phenomenex, Inc. All rights reserved.