LONDON HEALTH SCIENCES CENTRE (LHSC) WORKFLOW COMPARISON STUDY WITH THE AQUIOS CL FLOW CYTOMETER

Similar documents
AQUIOS CL LOAD & GO AND MAKE FLOW CYTOMETRY SIMPLICITY ITSELF

Navios EX FLOW CYTOMETER POWERFUL, DEPENDABLE CLINICAL FLOW CYTOMETRY

Navios EX FLOW CYTOMETER POWERFUL, DEPENDABLE FLOW CYTOMETRY

Navios Now Because Every Event Matters

Application Information Bulletin: Lymphocyte Subset Analysis

URGENT FIELD SAFETY NOTICE. AQUIOS CL Flow Cytometry System (PN B30166)

a Beckman Coulter Life Sciences: White Paper

Application Information Bulletin: DuraClone IM Tubes Compensation setup for high content DuraClone reagents

a Beckman Coulter Life Sciences: White Paper

A previous ICCS Module entitled Instrument optimization - Adjusting PMT voltages and compensation 1 should be read as a prerequisite to this module.

CLEARLLAB LS LYMPHOID SCREEN REAGENT

Automation In Life Sciences and Diagnostics Applications

WORKFLOW WITHOUT THE WORK

Kaluza Analysis Software SOPHISTICATED FLOW CYTOMETRY, SIMPLIFIED.

2. SUMMARY AND EXPLANATION

Completing the performance picture.

Tailored automation solutions that fit. Automation system solutions component overview

Tailored automation solutions that fit. Automation system solutions component overview

ENUMERATION OF LYMPHOCYTES SUBSETS (IMMUNOPHENOTYPING-IPT) (CD3, CD4, CD8, CD19 & CD16/56)

DURACLONE IM ACCELERATE YOUR PACE IN IMMUNE SYSTEM RESEARCH. For Reseach Use Only - Not for use in Diagnostic procedures

The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee

WORKFLOW WITHOUT THE WORK. Power Express Clinical Automation Solution

Atherosclerosis Risk in Communities Carotid MRI Study Manual 6 Flow Cytometry Laboratory

Beadless Absolute Counting

Improving the Delivery of Transfusion Medicine Lab Services using ImmuLINK

Biomek Automated Genomic Sample Prep Accelerates Research

Simplify your pathogen testing. Amplify your confidence.

Automation of Lexogen s QuantSeq 3 mrna-seq Library Prep Kits on the Biomek FX p NGS Workstation

MANAGEMENT BRIEF. How Automation is Revolutionizing White Cell Differential Analysis.

BR-16994A BIOMEK 4000 LABORATORY AUTOMATION WORKSTATION INTUITIVE SOFTWARE. MAXIMUM VERSATILITY.

The Emergence of Flow Cytometry as a Routine Diagnostic Testing Technology Right Here in Our Own Back Yard! Mike Reed February 2013

Molecular Diagnostics INTERNATIONAL PRODUCT CATALOG

Inoculation efficiency is the first step in the bacteriology testing process and is the first opportunity for automation.

NovoCyte Flow Cytometer

Biomek Automated Genomic Sample Prep Accelerates Research

SPRIworks Systems. Push button. Walk away. Fully automated library construction systems with built-in size selection and cleanup BR-15981B

Essential Time & Attendance Supervisor Basics

FEDERAL PUBLIC SERVICE, HEALTH, FOOD CHAIN SECURITY AND ENVIRONMENT CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS GLOBAL REPORT

XS-1000i New Sysmex 5-part diff haematology analyser with fluorescence technology

Advanced Immunoassay Testing

SCIENTIFIC INSTITUTE OF PUBLIC HEALTH QUALITY OF MEDICAL LABORATORIES CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS GLOBAL REPORT

High-throughput automation with the Attune NxT Autosampler: consistent results across all wells and across plates

determine optimum instrument settings for their own instruments and establish their own daily values.

Optimizing Transfusion and Blood Banking Services by Using Lean and Smart Automation

SYSTEMS. MassARRAY System. Uncompromised Molecular Testing. For Research Use Only. Not for use in diagnostic procedures.

Time Clock World. NOVAtime 2000 Sample Reports. Time Clock World. presents

Business Case for Haematology Analyser Replacement Recommendation

FULLY AUTOMATED ENZYME IMMUNOASSAY ANALYZER AUTOMATED IMMUNOASSAY ANALYZER. Advanced Immunoassay Testing TOSOH BIOSCIENCE

THE POWER OF ONE SMART SOLUTION. Imagine one partner for all your automation needs. Meet the leaders in automation.

THE ESSENTIAL ELEMENT of a

TAKE YOUR NEXT STEP WITH CONFIDENCE

Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement

Connected. The Integrated Lab. The fully integrated autoimmune laboratory.

Directors: Dr. Shivram G. Iyer, MD Dr. Rohit Gupta, MD, DNB, MNAMS. Phone numbers:

Essential Time & Attendance Supervisor Timecard Basics

Quality Control in polychromatic Flow Cytometry - EuroFlow interlaboratory standardization

Application Note. Assay Portability on the BD FACSVerse System. Summary. Maria Jaimes, Yibing Wang, Catherine McIntyre, and Dev Mittar

Automated Liquid Handling for Cell Staining and Imaging

Document Title Versalyse Whole Blood Staining Method Department. Type Status CA

Staffing to Workload in Central Receiving

MICROBIOLOGY AUTOMATION EXPANDS WITH THE COPAN WASP DT: WALK-AWAY SPECIMEN PROCESSOR

PowERFUL. INTUITIvE. customizable. ThE NEw STAR of. BENchToP FLow cytometry

Novel Developments in Cytometry. Martin Adelmann Marketing Manager Cellular Analysis & Life Science Beckman Coulter EMEAI

Fully Automated C100HT Biologics Analyzer Sample Preparation on a Biomek i5 Multichannel Workstation

High Throughput Suspension Cell and Bead Analysis. Intellicyt ique Screener PLUS

Analysis of Monoclonal Antibody Charge Variants by Capillary Zone Electrophoresis

SYSTEMS. MassARRAY System. Uncompromised Molecular Testing. For Research Use Only. Not for use in diagnostic procedures.

ACTG Laboratory Technologist Committee Revised Version 2.0 ACTG Lab Man CD38 Quantitation 07 May 2004

BD FACSMelody. Cell Sorter. The simple solution for consistent, quality results

FirstDose TM Medication Management System. Capsa Solutions. Patent Pending

Illumina TruSeq RNA Access Library Prep Kit Automated on the Biomek FX P Dual-Hybrid Liquid Handler

Designing and Implementing a High-Level Multicolor Flow Cytometry Assay. Brent Wood MD PhD Department of Laboratory Medicine University of Washington

Kit Contents FL-10 FL-12 FL-07 FL-09. Important: Before You Begin. New for this Mailing. Reporting Code Changes. Critical Reporting Information

Exempt Employee Time & Absence Reporting

Fast, accurate sequencing starts here. Accelerate your NGS success with automated sample prep

Your Research, Revolutionized

Sage 300 Direct Deposit. Getting Started Guide

Essential Time & Attendance

Contamination Control in the Hydraulic Industry

Pharmacokinetic Analysis of the mab Adalimumab by ELISA and Hybrid LBA/LC/MS: A Comparison Study

TO RESULT SOFTWARE DATA MANAGEMENT MASS SPECTROMETRY FROM SAMPLE COLLECTION MSMS WORKSTATION SOFTWARE

Automated Optimization of IgG Production in CHO Cells

Nonexempt Employee Time & Absence Reporting

CLINICAL MASS SPECTROMETRY DATA MANAGEMENT SOFTWARE FROM SAMPLE COLLECTION TO RESULT MSMS WORKSTATION SOFTWARE

Embla Enterprise. Sleep Business Software

Minimum Information about a Flow Cytometry Experiment (MIFlowCyt) Annotation

ADDITIONAL SPECIMEN TRANSPORT GUIDELINES

Isolation of total nucleic acids from FFPE tissues using FormaPure DNA

Three Systems, One Touch

What you need to know before designing a panel

The Life of a Specimen Received by Mayo Medical Laboratories. Disclosures. Relevant Financial Relationship(s): Nothing to Disclose

NovoCyte Quanteon F. Flow Cytometer. When Exceptional Performance Meets Simplicity

JOB DESCRIPTION. Date: May FLSA Status: Non-Exempt

Application Note. Introduction. EMD Millipore is a division of Merck KGaA, Darmstadt, Germany

IBM TRIRIGA Application Platform Version Introduction to the Gantt Scheduler

BD FACSCalibur. The flow cytometer for your routine cell analysis needs

PERFECT-COUNT MICROSPHERES

Scaling Up NGS Sample Preparation: Automation and Lab Preparation. Alicia Burt Sr. Product Manager

Immunophenotyping of Peripheral Blood and Bone Marrow Cells by Flow Cytometry *Akanni EO and # Palini A.

Page 1 of 7 ADDENDUM NO. 1. March 16, Histology Specimen Processing and Preparation Services RFP # H General

Transcription:

CASE WHITE STUDY PAPER LONDON HEALTH SCIENCES CENTRE (LHSC) WORKFLOW COMPARISON STUDY WITH THE AQUIOS CL FLOW CYTOMETER Laboratory Profile London Health Sciences Centre (LHSC), Flow Cytometry Laboratory, London, Ontario, Canada Operates Monday to Friday from 8 a.m. to 5 p.m. with an on call service for weekends and holidays Employs 5 medical technologists, 1 technical and medical supervisor, and 2 PhD research scientists Performs 6,000 tests per year The flow cytometry laboratory at Victoria Hospital is situated within a tertiary care academic institution, London Health Sciences Centre. This lab serves more than 1,000,000 residents within London and the greater South Western Ontario region. The flow cytometry laboratory in London Health Sciences Centre performs over 6,000 clinical flow cytometry tests per year in addition to numerous research projects. The lab is staffed by five full-time, registered medical technologists, in addition to a technical and medical supervisor. Two PhD students also work in the laboratory in translational research and development projects. Overview of Traditional Flow Cytometry Workflow Many traditional flow cytometry systems, such as the current Beckman Coulter FC500 with TQ Prep, require either manual or semi-automated preparation of samples, manual creation of worklists, manual data review, and manual tabulation of numerical data. This workflow can result in longer run times, more hands-on time, and require more experienced operators. See Figure 1 for the flow cytometry workflow at London Health Sciences Centre. Figure 1. London Health Sciences Centre (LHSC) Flow Cytometry Workflow Example Apply labels to daughter tubes Pipette antibody Pipette blood Load daughter tubes into carousel Transfer carousel TQ Prep Sample Prep System Flow Count LIS (manual transcription) User Reviews data Results FC500 worklist FC 500 Flow Cytometer

Overview of the AQUIOS CL Flow Cytometer Workflow The AQUIOS CL system is a fully-automated system that performs all of the preparatory steps leading to a test result. See Figure 2 for the AQUIOS CL Flow Cytometer workflow. Figure 2. AQUIOS CL Flow Cytometer Workflow Load specimen tubes into cassette AQUIOS CL Flow Cytometer User Reviews data Results LIS Requests Comparison Protocol The purpose of this case study is to evaluate the AQUIOS CL Flow Cytometer design and its resultant workflow through direct comparison with the Beckman Coulter FC500 and TQ Prep system. This case study evaluates the following workflow parameters: Total elapsed time from startup to shut down The time duration that begins with the first step in starting up the instrument(s) and ends at the last moment of shutting down the instrument(s) when all instrument and software activity has ceased. Time to first result from startup For the AQUIOS CL system and alternate systems, the time duration that begins the moment the system is turned on and ends with the first test result for the same sample displayed on the instrument computer screen, for a batch of 10, 25, or 50 samples. Turnaround time For alternate systems, the time duration begins with the first sample preparation step and ends with completed test results for a batch of 10, 25, or 50 samples, as demonstrated by the last sample test result displayed on the instrument computer screen. For the AQUIOS CL system, time duration begins with the first cassette placed in the autoloader and ends with the completed test result for the last sample being displayed on the instrument computer screen, for a batch of 10, 25, or 50 samples. The calculation is total time divided by the number of samples. The time does not include QC. Number of procedural steps The number of instructions or actions required to accomplish the task. This includes all the distinct steps in the process of generating a test result. To avoid individual interpretation of the meaning of a step, the protocol adheres to the steps as defined by the manufacturer in the applicable Instructions for Use manual. For example, if the Instructions for Use manual states to pipette 100 μl of blood, this is one step. Sub-steps such as insert a pipette tip on the pipette, push down the plunger on the pipette, insert the pipette in the blood sample are not considered steps unless the manufacturer specifies them as such. It is important to note that the AQUIOS CL Flow Cytometer and alternate systems do not employ identical procedural steps. Therefore, the steps are categorized by function even if it is not identical system-to-system. It should also be noted that, for the AQUIOS CL Flow Cytometer the procedural steps for Tetra-1 and Tetra-2+ tests are identical.

Operator hands-on time The time required by the user to physically perform the steps listed in the provided test procedure. Note: The workflow parameters measured by this case study does not affect safety or effectiveness Study Controls To ensure standardization between both comparison platforms, the following controls were used: The same samples are run on both systems. Both systems utilize the manufacturer s 4-color, lymphocyte subset panels using CYTO STAT tetrachrome CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 for the FC500 Flow Cytometer and AQUIOS Tetra-1 for the AQUIOS CL Flow Cytometer Beckman Coulter s IMMUNO-TROL and IMMUNO-TROL Low Controls are run on the FC500 Flow Cytometer and AQUIOS IMMUNO TROL and AQUIOS IMMUNO-TROL Low controls are run on the AQUIOS CL Flow Cytometer for each of the sample numbers studied. The AQUIOS CL Flow Cytometer and the FC500/TQ Prep systems are run per the Instructions for Use manuals for each system. In all instances, the AQUIOS CL Flow Cytometer utilizes the autoloader system. In all instances, one individual performed all steps on both systems. The individual was not permitted to multitask between the two systems. Control test requests were created prior to initiating the study following the system Instructions for Use manual. Comparing the FC500/TQ Prep System to the AQUIOS CL Flow Cytometer Total Elapsed Time from Startup to Shutdown This test case consisted of 10, 25, and 50 sample scenarios. The AQUIOS CL autoloader system, running AQUIOS Tetra-1 was compared directly to the FC500 and TQ Prep systems, running CYTO-STAT tetrachrome CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5. Note that the 50 sample scenario required 3 batches on the FC500 and TQ Prep systems. Refer to Table 1 for a tabular description of the total elapsed time results. Table 1. Hands-On/Hands-Off Time and Total Elapsed Time Note: The time format is displayed as (h:mm:ss). For example, 1:01:01 equals one hour, one minute, and one second. NUMBER OF SAMPLES TIME PER STEP DATA FC500 AQUIOS CL FC500 AQUIOS CL FC500 AQUIOS CL TOTAL Hands-On Time 0:55:51 0:04:23 1:15:47 0:08:18 1:37:51 0:12:15 TOTAL Hands-Off Time 2:17:41 1:15:18 2:54:39 1:51:58 4:14:33 2:52:24 TOTAL Elapsed Time 3:13:32 1:19:41 4:10:26 2:00:16 5:52:24 3:04:39

Refer to Figure 3 for a graphic depiction of the total elapsed time results. Note that combining the Total Hands-On Time (blue bar) with the Total Hands-Off Time (gray bar), depicts total elapsed time. Figure 3. Hands-On/Hands-Off Time and Total Elapsed Time 7:12:00 AQUIOS CL vs FC500 with TQ Prep: Hands-On and Hands-Off Time 6:00:00 4:48:00 3:36:00 2:24:00 1:12:00 0:00:00 FC500-10 1 Aquios-10 FC500-25 Aquios-25 FC500-50 Aquios-50 TOTAL Hands-On Time TOTAL Hands-Off Time Legend Key: The number after the cytometer name represents the number of samples run. In the 10 sample scenario, the AQUIOS CL Flow Cytometer is 2.4X faster than the FC500 and TQ Prep system. In the 25 sample scenario, the AQUIOS CL Flow Cytometer is 2.1X faster than the FC500 and TQ Prep system. In the 50 sample scenario, the AQUIOS CL Flow Cytometer is 1.9X faster than the FC500 and TQ Prep system. Assuming 50 samples per workday over the period of one year, the AQUIOS CL Flow Cytometer saves a lab hundreds of hours of valuable time each year. With such a quick turn-around time, the AQUIOS CL Flow Cytometer is capable of more work in less time and has the capacity to do the same work in less than half the time of a traditional Flow Cytometer combination like the FC500 and TQ Prep system. Amount of Hands-On Time Required In Figure 3, the Total Hands-On Time is depicted by the blue bar and the total Hands-Off Time is depicted by the gray bar. In all test case scenarios, the AQUIOS CL Flow Cytometer required significantly less hands-on time than that of the FC500/TQ Prep system combination. This can be attributed to the integrated sample preparation features in the AQUIOS CL Flow Cytometer. The all-in-one approach of the AQUIOS CL Flow Cytometer is intended to assist in time savings and to help increase the workflow efficiency. Number of Procedural Steps Required This test case consisted of 10, 25, and 50 sample scenarios. The AQUIOS CL autoloader, running AQUIOS Tetra-1 was compared directly to the FC500 and TQ Prep systems running CYTO STAT tetrachrome CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5. See Table 2 for a breakdown of the results. The Instructions for Use manuals were used by the investigator to determine the number of steps for each category. Note: LHSC reported number of steps by multiplying by the number of specimens. The table below shows the number of steps for each specimen.

Table 2. Number of Procedural Steps NUMBER OF STEPS DATA, N NUMBER OF SAMPLES FC500 AQUIOS CL FC500 AQUIOS CL FC500 AQUIOS CL Cytometer Startup and Cleaning 14 10 14 10 14 10 Sample Prep Device Startup and Cleaning 25 0 25 0 25 0 Cytometer Setup, Settings, and Compensation 30 0 30 0 30 0 Sample Prep Device Startup 0 0 0 0 0 0 Manual Sample Prep 13 0 15 0 15 0 Quality Control 26 13 26 13 29 13 Run Samples 13 5 13 5 13 5 Data Review and Analysis 2 2 2 2 2 2 Clean and Shutdown 21 8 21 8 21 8 Total Steps 144 38 146 38 149 38 Steps related to sample preparation, quality control and running samples contributed to the distinct difference in the total number of steps between both systems. Regardless of the number of samples in each test case, the AQUIOS CL system consistently reported steps in significantly lower quantity than the FC500/TQ Prep system setup. In addition, the number of steps remained stable across all sample scenarios due to the fully automated design of the AQUIOS CL Flow Cytometer. The reduced number of steps on the AQUIOS CL Flow Cytometer offers a probable correlation to the reduced hands-on time reported in the section above. Turnaround Time This test case consisted of 10, 25, and 50 sample scenarios. The AQUIOS CL autoloader, running AQUIOS Tetra-1 was compared directly to the FC500 and TQ Prep systems running CYTO STAT tetrachrome CD3/CD8/CD45/CD4 Three results were measured: time to first result from startup, turn-around time, and total elapsed time. Refer to Table 3 for the results. Table 3. Turnaround Time Results NUMBER OF SAMPLES FC500 AQUIOS CL FC500 AQUIOS CL FC500 AQUIOS CL Time to first result from startup 2:38:43 0:53:10 2:53:18 0:53:42 2:50:57 0:53:45 Turn-around time 1:04:37 0:33:27 1:58:25 1:08:29 2:32:59 2:09:10 Total elapsed time 3:13:32 1:19:41 4:10:26 2:00:16 5:52:24 3:04:39 Figure 4 depicts time to first result from startup for both the AQUIOS CL system and the FC500/TQ Prep system. The time duration in this instance begins when the system is turned on and ends with the first test result displayed on the workstation. The AQUIOS CL system delivered the first result approximately three times faster than the FC500/TQ Prep system for all three sample scenarios.

Figure 4. Time to First Result from Startup Aquios CL vs FC500 with TQ Prep: Time to First Result from Startup 3:36:00 Time 2:24:00 1:12:00 0:00:00 10 25 Number of Samples 50 FC 500 AQUIOS Figure 5 depicts turn-around time for both the AQUIOS CL system and the FC500/TQ Prep system. The time duration in this instance began with the first sample preparation step (FC500/TQ Prep system) or the first cassette being placed on the system (AQUIOS CL system) and ended with the last test result displayed on the workstation. The AQUIOS CL system delivered faster turn-around times in all three sample scenarios. The 10 sample scenario turn around time on the AQUIOS CL system was 1.9 times faster than that of the FC500/TQ Prep system. The 25 sample scenario on the AQUIOS CL system was 1.7 times faster, and the 50 sample scenario on the AQUIOS CL system was 1.2 times faster than that of the FC500/TQ Prep system. Figure 5. Turn-Around Time Aquios CL vs FC500 with TQ Prep: Turn Around Time 3:36:00 Time 2:24:00 1:12:00 0:00:00 10 25 Number of Samples 50 FC 500 AQUIOS Figure 6 depicts total elapsed time, turn-around time, and time to first result for both the AQUIOS CL system and the FC500/TQ Prep system. In all three instances including total elapsed time, turn-around time, and time to first result from startup, the AQUIOS CL Flow Cytometer consistently delivered faster results in this case study.

Figure 6. Total Elapsed Time, Turn-Around Time, and Time to First Result Aquios CL vs FC500 with TQ Prep: Total elapsed me, Turn-around me and Time to first result Aquios-50 Number of Samples FC500-50 Aquios-25 FC500-25 Aquios-10 FC500-10 0:00:00 1:12:00 2:24:00 3:36:00 4:48:00 6:00:00 7:12:00 Time (h:mm:ss) Total elapsed me Turn-around me Time to first result from startup Conclusions The study indicates that the AQUIOS CL Flow Cytometer performs well compared to a traditional flow cytometry system by providing faster times to first result from startup, faster turnaround times and faster total elapsed times than the comparator, with fewer total steps and less hands-on time, when processing 10, 25 and 50 samples in the given test case. There are two important factors to note regarding this study. The first factor to note is that the time savings exhibited in this study are conservative estimates. A principal advantage of the AQUIOS CL system, freedom from batching, is not measured as a test case. Most current flow cytometry laboratories including FC500/TQ Prep system users have workflows that involve waiting until a critical mass of samples arrive in the lab before starting sample preparations (batching). An AQUIOS CL system user, however, can start running as soon as the first sample arrives. This amounts to improved workflow results in greater time savings than the study indicates. Additionally, the lab can load new samples at any time during the hours of operation with the AQUIOS CL Flow Cytometer. Due to the workflow, traditional flow cytometers require samples to be held until the next morning at a designated time before closing. The second factor to note is that with most traditional flow cytometers, the operator tends to stay at or near the instrument to monitor progress. If there is not a lot of time in between steps, for example five minutes or ten minutes, the user tends to remain close to the instrument and monitor the data acquisition in real- time rather than walk away to accomplish another task. The time that a user is monitoring the data acquisition is not accounted for in this study. The Load & Go nature of the AQUIOS CL system actually accounts for greater time savings and work efficiency than actually indicated by this study since a user can accomplish more in less time. Aquios CL is a Class 1 laser product. Trademarks Beckman Coulter, the stylized logo, and the Beckman Coulter product and service marks mentioned herein are trademarks or registered trademarks of Beckman Coulter, Inc. in the United States and other countries. 2015 Beckman Coulter Life Sciences. All rights reserved. For Beckman Coulter s worldwide office locations and phone numbers, please visit Contact Us at beckmancoulter.com FLOW-1314WP12.15-A