Characterization of Glycosylation in the Fc Region of Therapeutic Recombinant Monoclonal Antibody

Similar documents
Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Analysis of Monoclonal Antibody (mab) Using Agilent 1290 Infinity LC System Coupled to Agilent 6530 Accurate-Mass Quadrupole Time-of-Flight (Q-TOF)

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

Separation of Native Monoclonal Antibodies and Identification of Charge Variants:

Cell Culture Optimization Using an Agilent Bio-Monolith Protein A Column and LC/MS

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS

Profiling Glycosylation of Monoclonal Antibodies at Three Levels Using the Agilent 6545XT AdvanceBio LC/Q TOF

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Characterize mab Charge Variants by Cation-Exchange Chromatography

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

Application Note. Authors. Abstract. Ravindra Gudihal Agilent Technologies India Pvt. Ltd. Bangalore, India

Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

Quantification of Host Cell Protein Impurities Using the Agilent 1290 Infinity II LC Coupled with the 6495B Triple Quadrupole LC/MS System

Antibody Analysis by ESI-TOF LC/MS

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

A Comprehensive Approach for Monoclonal Antibody N-linked Glycan Analysis from Sample Preparation to Data Analysis

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Peptide Mapping: A Quality by Design (QbD) Approach

Put the PRO in Protein Characterization

ENHANCED PROTEIN AND PEPTIDE CHARACTERIZATION

Fast Analysis of Environmental Phenols with Agilent Poroshell 120 EC-C18 Columns

Application Note. Authors. Abstract. Ravindra Gudihal Agilent Technologies India Pvt. Ltd. Bangalore India

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

AdvanceBio Peptide Mapping

Reproducible LC/MS Peptide Separation Using Agilent AdvanceBio Peptide Plus Columns

LC/MS Based Quantitation of Intact Proteins for Bioanalytical Applications

BioHPLC columns. Tim Rice Biocolumn Technical Specialist

Analysis of Monoclonal Antibody N-glycans by Fluorescence Detection and Robust Mass Selective Detection Using the Agilent LC/MSD XT

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

MAbPac RP Column. High-performance reverse phase chromatography column for monoclonal antibody analysis

Application Note. Authors. Abstract. Introduction. Pharmaceutical

Enabling routine characterization of proteins. Agilent MassHunter BioConfirm software

Analysis of Nucleosides Using an Agilent Infinity II High Speed UHPLC with the 6130 Single Quadrupole Mass Selective Detector

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs)

4/4/2013. BioHPLC columns. Paul Dinsmoor Biocolumn Technical Specialist. April 23-25, Size Exclusion BioHPLC Columns

Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins

Analysis of Monoclonal Antibodies and Their Fragments by Size Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer

Unified Workflow for Monoclonal Antibody Charge Heterogeneity, Purity, and Molecular Weight Analysis

Background - Importance of monoclonal antibody characterization

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

Ensure your Success with Agilent s Biopharma Workflows

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis

Accelerate mab Characterization Using Automated Sample Prep

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

Faster, easier, flexible proteomics solutions

Size Exclusion Chromatography/ Mass Spectrometry Analysis of Antibody Drug Conjugates Using the Agilent 1260 Infinity II Bio-Inert LC

Application Note. Authors. Abstract. Biopharmaceuticals

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

Determination of Asarinin in Xixin (Asari Radix Et Rhizoma)

Biotherapeutic Method Development Guide

Host Cell Protein Analysis Using Agilent AssayMAP Bravo and 6545XT AdvanceBio LC/Q-TOF

Critical Quality Attribute Assessment by Peptide Mapping Using LC/MS with Superficially Porous Columns

Monoclonal Antibody Charge Variants Identification by Fully Automated Capillary Isoelectric Focusing Mass Spectrometry

Biotherapeutic Method Development Guide

An Integrated Workflow for Intact and Subunits of Monoclonal Antibody Accurate Mass Measurements

Faster Separations Using Agilent Weak Cation Exchange Columns

Answers, Simple and Streamlined. Solutions for Biotherapeutic Intact Mass Analysis

mabs and ADCs analysis by RP

NISTmAb characterization with a high-performance RP chromatography column

mab Glycopeptide Profiling with V-Tag Adding reliable glycoprofiling to your peptide mapping workflow with ease and simplicity

Simple, Robust, High Quality Intact Mass Analysis A Biosimilars Case Study

A Highly Accurate Mass Profiling Approach to Protein Biomarker Discovery Using HPLC-Chip/ MS-Enabled ESI-TOF MS

Agilent 1290 Infinity II 2D-LC Solution Biopharmaceutical Polymer Analysis. WCBP Jan 2017 Washington, DC

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Bio-Monolith Protein G Column - More Options for mab Titer Determination

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

CQA Assessment by Peptide Mapping Using LC/MS with an AdvanceBio Peptide Plus Column

Columns for Biomolecules BioLC Column Lines

Clearly better LC/MS solutions

Thermo Scientific Solutions for Intact-Protein Analysis. Better, Faster Decisions for. Biotherapeutic Development

Highly Confident Peptide Mapping of Protein Digests Using Agilent LC/Q TOFs

High-Resolution Separation of Intact Monoclonal Antibody Isoforms

Choosing the Right Calibration for the Agilent Bio SEC-3 Column

Method Translation in Liquid Chromatography

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Heart-cut 2D-LC/MS approach for pharmaceutical impurity identification using an Agilent 6540 Q-TOF LC/MS System

Advanced Characterization of Antibody Drug Conjugates (ADCs) by Liquid Chromatography and Mass Spectrometry (LC/MS) John Gebler, Ph.D.

Mass Spectrometry Analysis of Liquid Chromatography Fractions using Ettan LC MS System

Application Note. Abstract. Author. Biopharmaceutical

Analysis of Monoclonal Antibodies and Their Fragments by Size-Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer

Characterization of a fusion protein under native and denaturing conditions with maxis II

E f f ic i ent ly Com pa ring Bat c h e s o f a n Intac t Mono c lo na l A n t ibo dy using t h e

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

Trastuzumab Glycan Batch-to-Batch Profiling using a UPLC/FLR/Mass Spectrometry Platform

Agilent Biocolumns. Titer Determination. Application Compendium

Development of Analysis Methods for Therapeutic Monoclonal Antibodies Using Innovative Superficially Porous Particle Biocolumns

Molecular Characterization of Biotherapeutics The Agilent 1260 Infi nity Multi-Detector Bio-SEC Solution with Advanced Light Scattering Detection

Transcription:

Characterization of Glycosylation in the Fc Region of Therapeutic Recombinant Monoclonal Antibody Application Note BioPharma Authors James Martosella, Phu Duong, and Alex Zhu Agilent Technologies, Inc. Abstract Biotherapeutic drugs, such as monoclonal antibodies, constitute the most rapidly growing drug class, and have become a major clinical success in human therapeutics over the past decade. However, the process of discovery and development of biotherapeutics poses throughput challenges that are different from those set by traditional small molecule drugs. They require more complicated manufacturing and characterization processes that need faster data turnaround times, yet they must be well characterized with regard to potency, identity, quality, purity, and stability. Thus the need for rapid and accurate profiling techniques is necessary. In this note we describe a fast and efficient liquid chromatography-tof mass spectrometry approach to characterize an IgG derived from CHO cells. An Agilent Poroshell 3SB-C3, 5 µm superficially porous IgG1 column was used to obtain accurate glycoform masses of an intact IgG1, with a subsequent analysis of papain-digested IgG1 to obtain site-specific glycosylation profile information of the Fc region for further characterization of IgG1 heterogeneity. Additionally, a subsequent mab-glyco chip LC/MS analysis was used to build a glycan accurate mass database for additional validation of the glycans identified in the TOF mass analysis.

Introduction Monoclonal antibodies (mabs) are used increasingly as therapeutic agents to target various diseases, and form a large fraction of therapeutic proteins commercially approved or currently under development. With an increasing number of mabs moving to market, the need for rapid analytical methods for their detailed characterization has dramatically increased [1]. For example, during mab development it is important to find the recombinant production system that offers optimal productivity with rapid turnaround times to meet process development demands. Assessing the glycosylation profiles in shortened analysis times is vital to mab development and the need to have rapid and reliable liquid chromotography-mass spectrometry (LC/MS) glycoprofiling methods are growing. This application note describes a rapid online reversed-phase (RP) LC/MS approach to profile glycosylation for a therapeutic mab (IgG1). Separations of intact IgG1 and papain-cleaved IgG1, to yield Fc and Fab fragments, were optimized for rapid analysis using an Agilent Poroshell 3SB- C3, 5 µm superficially porous column. Specifically, intact IgG1 was characterized to obtain glycoform and accurate glycan mass information. The IgG1 was then subjected to papain digestion and the glycosylation containing Fc portion analyzed by time-of-flight (TOF) mass spectrometry to determine glycospecific modifications. The methods were developed to provide essential glycoform profile information using rapid LC/MS procedures for achieving quick data turnaround times. For further validation of the Fc glycan-specific assignments from the TOF analysis, the intact mab was then subjected to a fully automated rapid glycan analysis. This was achieved using an mab-glyco chip kit to establish a glycan accuratemass database. The methods described herein are useful for cell line and clone selection processes, and cell culture process optimizations where absolute quantitation is not necessary, but where rapid, reliable and efficient characterizations are desired. Experimental Humanized monoclonal antibody derived from a Chinese hamster ovary (CHO) cell line was purchased from Creative Biolabs, Shirley, NY. Triflouroacetic acid was purchased from Sigma-Aldrich Corp., St. Louis, MO, and isopropanol, n-propanol and acetonitrile were supplied from Honeywell Burdick & Jackson, Muskegon, MI. The 1-propanol was purchased from VWR ( p/n BJ322-4). Dialysis cassettes had a 3,5-Da protein MW cut-off (MWCO) and were purchased from Thermo Scientific ( p/n 6633). Papain digestion of mab To prepare the Fc and Fab fragments, a Pierce Fab Micro Preparation Kit (Thermo Scientific p/n 44685) was used. Final Fab and Fc cleanup used microcentrifugation with spin columns to separate the immobilized papain from the fragments. The final concentration was 2 µg/µl. Intact IgG1 chromatographic conditions Column: Agilent Poroshell 3 SB-C3, 2.1 75 mm, 5 µm (p/n 6675-99) Eluent: Injection volume: 2 µl Flow rate: Gradient: A: water (5% AcOH, 1.% FA,.5% TFA) B: 7/2/1 IPA/ACN/water (5% AcOH, 1.% FA,.5% TFA) 1. ml/min Segmented Time (min) Temperature: 8 C Detector: % B 2 4 45 8 45 9 9 1 2 UV, 28 nm Fc fragment chromatographic conditions Same as intact IgG1 separation, except: Injection volume: 5 µl (2 µg/µl) Gradient: 2 to 5% B, 7 min, 5 to 9% B, 4 min, 9 to 2% B, 1 min (re-equilibration) Detector: Instrumentation DAD, 225 nm LC/MS Agilent 12 Infinity Series with auto injector (High Performance Autosampler), binary pump, thermostatted oven (TCC) and diode array detector (DAD) coupled to an Agilent 6224 TOF LC/MS. Glycan library Agilent 12 Infinity HPLC-Chip/ MS system (Agilent mab- Glyco Chip Kit) coupled to an Agilent 653 Accurate-Mass Q-TOF LC/MS. 2

6224 TOF instrument parameters Spectra were recorded in positive ion and centroid modes. Gas temperature: 35 ºC Drying gas: Nebulizer: Vcap: Octupole 1 RF: MS: Mass range: 1 L/min 45 psi 5,5 V (intact), 5, V (Fc) 75 V 4 Hz 1 to 7, m/z Ref. mass: 149.2332, 922.9798 Acquisition mode: Range mode: Extended dynamic range 2 GHz MS data analysis Data was processed using Agilent MassHunter Qualitative Analysis and MassHunter BioConfirm software. Agilent mab-glyco Chip Kit The kit incorporates the mab-glyco chip (p/n G424-6421), reagent pack, mab-glyco chip content disk, and a quick-start guide. The reagent pack provides all chemicals needed for ready chip operation, that is, system conditioning reagent for flow path deactivation and carry over minimization, glycan standards for chromatographic checkout and method development, antibody standard for functional checkout and troubleshooting, and deglycosylation buffer for dilution of standards and samples, and for loading mab samples onto the chip s enzyme reactor. Conditions for the glyco chip were; reactor (31 nl PNGase F), separation column (43 mm PGC, 5 µm), and enrichment column (16 nl PGC, 5 µm). Results and Discussion Analysis of intact IgG1 Reversed-phase separation of intact IgG1 can be challenging due to the large molecular size, diffusion limitations and increased hydrophobicity. For this reason, obtaining high speed separations with narrow and symmetrical peak shapes has become more challenging, primarily due to the diffusion paths of these molecules within the particle pore structure. More recently, superficially porous materials have gained attention to circumvent the problems associated with faster, large-molecule separations afforded by the thin layer of porous packing around a solid core particle. Decreasing the diffusion time and limiting the diffusion path will result in narrower peaks and more efficient separations for macromolecules such as mabs. Here, the migration path of the protein (mab) is shortened compared to that of a fully porous particle. Figure 1 shows how this migration can improve mab separation during rapid run times. Figure 1. Biomolecule migration paths in totally porous and superficially porous particles. 3

For the separation of IgG1, a Poroshell 3 SB-C3 superficially porous column was used at elevated temperature, with carefully optimized conditions for highest MS analytical sensitivity, to enable a rapid and efficient separation (Figure 2A). Positive mode electrospray ionization (ESI) MS showed multiple peaks from 2,8 to 4,8 m/z corresponding to isoforms at different charge states (Figure 2B). These multiple charged envelopes were deconvoluted by the Agilent MassHunter software to obtain the molecular weight values of the different intact IgG1 glycoforms and to identify the accurate glycan masses derived from the IgG1 sequence data shown in Table 1. The MS spectrum peaks are 14 to 15 Da apart. These data indicate that other sources of IgG1 heterogeneity are present and that the difference in mass observed is a sum effect of these different modifications, along with the various glycosylation forms. mau 1 3 5 1.8 1.6 3,8.6876 4 3 1.871 Counts 1.4 1.2 1. 2.154.8.6 3,85.718 1.193 1 2 3 Figure 2A. LC/MS analysis of intact IgG1, with rapid reversed-phase separation using an Agilent Poroshell 3 SB-C3 LC column..4.2 2, 2,4 2,8 3,2 3,6 4, 4,4 4,8 5,2 5,6 Mass-to-charge (m/z) Figure 2B. ESI mass spectrum generated by averaging the MS scans across the main peak of the UV chromatogram in Figure 2A. 4

Table 1. Three major IgG1 glycoforms identified using MassHunter maximum entropy deconvolution from Figure 2B. Intact mab glycoform mass Gylcan assignment Structure Accurate glycan mass 148253.5366 G1/G2 (+384.8581) G1 (1) G1 (1) or G1/G2F (+3231.19) G1 (1) G2F (18) 1484.7426 G1F/G2 (+3231.19) G1F (19) G2 (9) 148549.614 G1F/G2F (+3377.1458 G1F (19) G2F (18) Papain-digested IgG1 and Fc fragment analysis To further profile the nature of the heterogeneity from the intact IgG1 analysis, the IgG1 was cleaved into its Fc and Fab fragments by papain digestion. Analysis of the Fc region bears a highly conserved N-glycosylation site. Glycosylation of the Fc fragment is essential for Fc receptor-mediated activity, and the N-glycans attached to this site are predominantly core-fucosylated biantennary structures of the complex type [2]. IgG1 cleavage with papain was used to clip below and above the hinge region to generate Fc and Fab fragments. Figure 3 shows the separation, and the conditions after cleavage in the reversed-phase chromatogram. In this separation, performed in under 3 minutes, two primary peaks at 2.4 and 2.7 minutes are shown, representing the Fc and Fab fragments, respectively. Additionally, there are smaller sets of partially resolved peaks at 2.1 to 2.5 minutes among the Fc region, representing variants of the Fc fragment. mau 25 2 mau 15 14 1 12 5 Fc 2.144 2.25 2.295 2.41 Fab 2.59 2.656 2.73 2.93 Expanded view Fc Glycoform Resolution (arrows) 1 1.5 1.75 2 2.25 2.5 2.75 3 3.25 3.5 3.75 min 8 Fab region 6 4 Fc region 2.5 1 1.5 2 2.5 3 3.5 4 min Figure 3. Reversed-phase separation of IgG1 after papain digestion showing two primary peaks of the Fc and Fab fragments. The inset details partially resolved peaks representing variants of the Fc and Fab fragments (arrows). 5

Figures 4A and 4B show the average mass spectrum and deconvoluted spectra of the primary Fc peak at 2.4 minutes. Fc with different glycans on it differ by a mass of 162 Da. Table 2 lists the glycosylation forms identified in the Fc fragmented IgG1. Fractionation of the intact IgG1 into Fab and Fc by papain digestion and the optimized rapid chromatographic reversed-phase separation of Fc have thus enabled a more efficient and rapid analysis of the glycoform distribution, demonstrating the value of the method for gaining relative estimations that can be used to assess glycosylation profiles. Counts 1 3 5 4.5 4 3.5 3 2.5 2 1.5 1.5 1,5.51 1,593.7447 1,699.9275 1,821.2788 1 3 5 1,821.2788 1,832.8552 4 Counts 3 2 1 1,961.371 2,124.668 1,844.4489 1,795.258 1,86.7526 1,853.5892 1,865.238 1,78 1,79 1,8 1,81 1,82 1,83 1,84 1,85 1,86 1,87 1,88 1,89 Mass-to-charge (m/z) 2,317.756 2,565.5739 2,85.5228 1,2 1,4 1,6 1,8 2, 2,2 2,4 2,6 2,8 3, 3,2 Mass-to-charge (m/z) 3,186.4395 Figure 4A. Fc peak average mass spectrum showing charge distribution profile from 1,4 to 3,2 m/z. The inset is a zoomed spectrum of the +14 charge state. The main peaks correspond to different glycoforms. 1 5 1.2 25,484.5 1 Hexose -.5 Galactose -.5 Mannose -.5 25,646.8 Counts.8.6 D -162 Da.4.2 25,612. 25,28.89 25,536.12 25,442.79 25,699.77 25,337.65 25,88.1 25,774.37 25,25 25,3 25,35 25,4 25,45 25,5 25,55 25,6 25,65 25,7 25,75 25,8 25,85 25,9 Deconvoluted mass (amu) Figure 4B. Deconvoluted ESI mass spectra of an IgG1 Fc fragment generated from the chromatographic peak at 2.4 minutes in the chromatogram shown in Figure 3, representing glycoforms regularly spaced by hexose units (D m/z = 162). 6

Table 2. TOF LC/MS Fc fragment glycan assignments. Glycan assignment Accurate glycan mass 21 A G (G1) (+1,461.3574) 21 A G (G2) (NA2) (+1,623.57) 211 A G (G1F) (NA2) (+1,67.512) 211 A G (G1F) (NA2) (+1,769.6445) 23 A G (+1,785.6439) 11 A G (+1,258.1613) 1 A G (+1,96.181) To validate the glycan assignments obtained from the Fc fragment analysis, we performed rapid chip-based online glycan analysis that enriches the released glycans cleaved from the intact mab and then separates them prior to nanospray ionization-tof analysis. A glycan accurate mass database was then established (Table 3) allowing quick assignment and validation of the glycans identified in the TOF mass analysis performed on the Fc region of the papain digested mab (Table 2). This additional investigation provided a fast complementary structure analysis while giving a higher confidence in the glycan characterizations. Table 3. Intact glyco-chip glycan results. Although this method provides only relative estimation and not absolute quantitation, it gives sufficient information and is useful for assessing the glycosylation profile of a monoclonal antibody. The rapid separation and data turnaround time afforded by the LC/MS TOF analysis can facilitate biopharma product quality analyses, such as glycosylation profiling, during process optimization studies, and additionally provides an alternative to peptide mapping validation that can slow productivity times. References 1. A. Lim, A. Reed-Bogan, B. J. Harmon. Anal. Biochem. 375, 163 (28). 2. H. S. Gadgil, P. V. Bondarenko, G. D. Pipes, T. M. Dillon, D. Banks, J. Abel, G. R. Kleemann, M. J. Treuheit. Anal. Biochem. 355, 165 (26). For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. Glycan assignment Accurate glycan mass 21 A G (G1) 1,461.3574 211 A G (G1F) 1,67.512 22 A G (G2) 1,623.57 212 A G (G2F) 1,769.6445 23 A G 1,785.6439 1 A G 1,96.181 11 A G 1,258.1613 2 A G 1,299.2141 Conclusions A rapid LC/MS-based approach was developed for the glycosylation profiling of a recombinant therapeutic monoclonal antibody. Optimized high resolution fast separation by an Agilent 5 µm Poroshell 3SB-C3 column enabled sensitive characterization of an intact mab (IgG1) and its Fc fragment after papain digestion. Coupled with TOF MS analysis, this application note demonstrates fast and effective characterization of mab glycosylation. Additionally, we have validated the TOF Fc region glycan analysis using mab-glyco chip LC/MS analysis on the intact mab through glycan cleavage. 7

www.agilent.com/chem For Research Use Only. Not for use in diagnostic procedures. Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 213, 217 November 2, 217 5991-2323EN