Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell

Similar documents
Table 163 Instrument-specific information. Instrument Pour-thru Kit Cell orientation Adapter. DR 6000 LQV Arrow faces right

1-(2-Pyridylazo)-2-Naphthol PAN Method 1 Method to mg/l Mn (LR) Powder Pillows

USEPA 1,2 Bicinchoninate Method 3 Method 8506 (CuVer 1) and Method 8026 (CuVer 2) 0.04 to 5.00 mg/l Cu Powder Pillows or AccuVac Ampuls

USEPA 1 Periodate Oxidation Method 2 Method to 20.0 mg/l Mn (HR) Powder Pillows

Phosphorus, Reactive (Orthophosphate)

Scope and application: For water, wastewater, seawater, brine solutions, produced waters and hydraulic fracturing waters.

Hardness, Calcium/Magnesium

Sulfate. USEPA 1 SulfaVer 4 Method 2 Method to 70 mg/l SO 4. Powder Pillows. Test preparation. Before starting.

Method 8017 (0.7 to 80.0 µg/l) Powder Pillows. Scope and Application: For water and wastewater; digestion is required to determine total cadmium.

Method to 500 µg/l CH 2 O Powder Pillows

Scope and application: For water and wastewater. Turbid samples are filtered during the analysis.

Color, True and Apparent, LR

Instrument Adapters Light shield DR 6000, DR 5000 DR 3900 LZV849 DR 3800, DR 2800 LZV646 DR or (A)

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Oxygen Demand, Chemical

Method (0.1 to 8.0 mg/l Cu) TNTplus 860

Sulfate DOC SulfaVer 4 Method 1. Method (0 70, and mg/l) Powder Pillows. Test procedure

Direct ISE Method Method to 4.00 mg/l NO 3 N Nitrate ISE

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Nitrogen, Simplified TKN (s-tkn )

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement, 0 to 40 mg/l COD

Direct ISE Method Method to 10.0 mg/l F Fluoride ISE

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement

Dimethylphenol Method Method LR (0.23 to mg/l NO 3 N or 1.00 to mg/l NO 3 ) TNTplus 835

Oxygen Demand, Chemical

Phosphorus, Reactive (Orthophosphate)

Phosphorus, Reactive (Orthophosphate)

DR/4000 PROCEDURE. IRON, Ferrous. Using Powder Pillows

Amperometric Titrator assembly. Digital Titrator. Beaker, low-form, 250-mL. Stir bar, octagonal, Teflon-coated, 50.8 x 7.9 mm

Oxygen Demand, Chemical

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

Oxygen Demand, Chemical

Nitrogen, Simplified TKN (s-tkn ) DOC

STP (off-line) Water Technologies & Solutions analytical procedure. colorimetric method DCA reagent 1 (5/5/1) 2 to 15 mg/l as STP

3. Add 0.4 ml of. 7. Use a TenSette

3. Add 0.4 ml of. 7. Use a TenSette

DR/4000 PROCEDURE OXYGEN SCAVENGERS. 4. Fill a second sample cell (the blank) with 25 ml of deionized water.

PHOSPHORUS, REACTIVE (0.00 to 5.00 mg/l PO 4

FerroVer Method (Powder Pillows or AccuVac Ampuls) USEPA approved for reporting wastewater analysis (digestion is required; see Section 2 * )

1,5-Diphenylcarbohydrazide Method* (Powder Pillows or AccuVac Ampuls) USEPA accepted for wastewater analyses** Using Powder Pillows

Coliforms, Total and E. coli

USEPA Membrane Filtration Method Method m-tec. Scope and Application: For potable water, nonpotable water, recreation water and wastewater.

Cadmium Testing and On Site Calibration for Water Testing Detection Range: ppm

Professional Boiler Water Test Kit

Zinc Testing and On Site Calibration for Water Testing Detection Range: 1 15ppm

Uranium Testing and Site Calibration for Water Testing Detection Range: 2 60ppb

TPTZ Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows

Oxygen Demand, Biochemical

Analysis of dissolved nitrite (NO 2- )

Methods and Materials Manual Of Ion Chromatography Using DX-100 IC. Collecting Samples. LDPE bottle with cap (pre-rinsed with DI water)

30th International Chemistry Olympiad

Rev Experiment 10

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

Heterotrophic Bacteria

Standard Methods for the Examination of Water and Wastewater

Gravimetric Analysis: Determination of % Sulfur in Fertilizer

TESTING THE WATERS HOW GOOD IS THAT BOTTLED WATER AND HOW EFFECTIVE IS YOUR WATER FILTER

PaSEC Colorimeter Instructions DR850

Water Analysis Kit for Home Brewers

An Oxidation-Reduction Titration: The Reaction of Fe 2+ and Ce 4+

Demonstration of osmosis

P-006 Subsampling Water Samples in the Field. Revision 1.0

The determination of copper in brass

Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric)

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Laboratory & Field Instruments: Sample Cells. Sample Cells. See Inside!

SECTION D.3 CHLOROPHYLL AND PHEOPHYTIN

Density (d) is a property of a substance equal to the ratio of its mass (m) to its volume (V): d = m V

1,10 Phenanthroline Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows

Human Myostatin, ELISA Kit (MSTN)

Write your own lab Spectrophotometry/Colorimetry

Human Angiotensin 2 (Ang2) ELISA

MAROOCHYDORE SHS. Student booklet. Including. ame: Teacher: Name. er: Maroochydore SHS Chemistry Department

Uranium Testing and On Site Calibration for Water Testing Detection Range: 2 60ppb

Polymer Block Cleaning Kit

SurTec 650 chromital TCP

CONSERVATION OF MATTER AND CHEMICAL PROPERTIES

Human Selenoprotein P1, Plasma (SEPP1) ELISA

SOP-C-102 Determination of Chemical Oxygen Demand

Application Note AN1501

Ti-Pure TITANIUM DIOXIDE. CONCENTRATED ACID SOLUBILITY OF TiO 2 METHOD: T WP

Analysis of Calcium Carbonate Tablets

Calcium-EDTA Tutorial for the MicroCal ITC200

BONDERITE NT-1 Phosphate-free, Regulated Heavy Metal-Free Pre-treatment

DREL/2010 SPECTROPHOTOMETER Soil and Irrigation Water Portable Laboratory Manual

IL-6 Ultrasensitive ELISA

Evaluation copy. Chloride and Salinity. Computer INTRODUCTION

BIOSCAN* ATP method for biocide (toxicant) evaluations

Technical Process Bulletin

Archer G11 Partner: Judy Aug Gravimetric Analysis of a Metal Carbonate

Name Honors Chemistry / /

Method 6.1 Boiling house products: Brix, pol and purity

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron with 1,10-Phenanthroline

scd134/ox40 (Human) ELISA

Total Residual Chlorine Testing Methods

Experiment 3: Determination of an Empirical Formula

HotBlock. ENVIRONMENTAL EXPRESS Call or Metals Digestions: HotBlock TM Digestion Systems

HiPer Gel Extraction Teaching Kit (Column Based)

Transcription:

Hardness, Total DOC316.53.01045 Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method 8374 4 to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell Scope and application: For boiler and ultrapure water. Test preparation Instrument-specific information Table 1 shows all of the instruments that have the program for this test. The table also shows sample cell and orientation requirements. To use the table, select an instrument, then read across to find the applicable information for this test. Table 1 Instrument-specific information Instrument Sample cell orientation Pour-Thru Kit Adapter DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 The flow path is to the right. LZV899 1 5940400 5940400 5940400 LZV899 LZV585 (B) LZV585 (B) LZV585 (B) DR 5000 The flow path is toward the user. LZV479 DR 3900 LZV899 Before starting Clean the Pour-Thru Cell and all labware before the analysis, refer to Prepare analysis labware on page 4. For the most accurate magnesium test results, keep the sample temperature between 21 29 C (70 84 F). To test for calcium or magnesium contamination in the labware, do the test until the results are consistent. Use dedicated plasticware for this analysis. To protect the Pour-Thru Cell from contamination when not in use, invert a small beaker over the top of the glass funnel. If the test result is more than 750 µg/l, make a 1:1 dilution of the sample for best accuracy. Use ultra-pure (aldehyde-free) water for the dilution. Do the analysis again on the diluted sample and multiply the result by two. Use "Alternate Forms" ("Chemical Forms" on some instruments) only when the sample is known to contain only Mg or Ca. This method does not distinguish between the two forms. Refer to the instrument documentation for Pour-Thru cell and module assembly and installation. Make sure to install the Pour-Thru cell correctly. Total hardness in mg/l equals mg/l Ca as CaCO 3 plus mg/l Mg as CaCO 3. In bright light conditions (e.g., direct sunlight), close the cell compartment, if applicable, with the protective cover during measurements. Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective equipment. Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory agencies for further disposal information. 1 A cell cover (LZQ105) and a tube set (LZQ102) is also necessary. 1

Items to collect Description Quantity Chlorophosphonazo Indicator Solution 2 ml CDTA Reagent for Ultra Low Range Hardness 2 drops Water, Ultra-pure, Aldehyde-free varies Cylinder, graduated, 50 ml, polypropylene 1 Dispenser, adjustable volume, 1.0 5.0 ml 1 Flask, Erlenmeyer, PMP w/cap, 125-mL 1 Pour-Thru Cell Module (Refer to Instrument-specific information on page 1) 1 Refer to Consumables and replacement items on page 5 for order information. Sample collection Test procedure Collect samples in clean plastic bottles with tight-fitting caps. Do not use glass bottles, which will contaminate the sample. Rinse the sample bottle several times with the sample to be collected. Seal the bottle to prevent contamination during the transport. Analyze the samples as soon as possible for best results. Start 1. Start program 227 Hardness Tot RL ULR. For information about sample cells, adapters or light shields, refer to Instrumentspecific information on page 1. 2. Flush the Pour-Thru Cell with 50 ml of ultra-pure water. 3. Fill a clean 125 ml plastic Erlenmeyer flask to the top with sample. Collect the sample directly in the flask if possible. 4. Rinse a clean 50 ml plastic graduated cylinder three times with sample. 5. Fill the cylinder to the 50 ml mark with sample from the flask. 6. Discard the remaining contents of the flask. 7. Pour the contents of the 50-mL cylinder back into the original flask. 8. Use the bottle-top dispenser to add 2.0 ml of Chlorophosphonazo Reagent to the sample. 2 Hardness, Chlorophosphonazo RL Method (1000 µg/l)

Zero 9. Swirl to mix. 10. Use a clean, dry, plastic 25 ml graduated cylinder to pour approximately half (25 ml) of the sample into the Pour Thru Cell. 11. When the flow stops, push ZERO. The display shows 0 µg/l CaCO 3. 12. Add 2 drops of CDTA Reagent for Ultra Low Range Hardness to the remaining sample in the flask. Complete the next 3 steps within 2 minutes. Read 13. Swirl to mix. 14. Pour the remaining sample into the Pour Thru Cell. 15. Push READ. Results show in µg/l CaCO 3. 16. Rinse the Pour-Thru Cell with at least 75 ml of ultra-pure water immediately after use. 17. Rinse the flask with ultra-pure water. Put a cap on the flask when finished. Interferences Table 2 shows the results of interference studies that used various hardness standard solutions between 0 and 500 µg/l as CaCO 3. Various cations and anions were added to the hardness standard solutions at levels in the range applicable for ultra-pure water Hardness, Chlorophosphonazo RL Method (1000 µg/l) 3

applications. An ion was considered an interference when the measured hardness concentration changed by ±10%. Table 2 Interfering substances Interfering substance Aluminum Ammonium Copper Formaldehyde Nitrate Potassium Silicon Sodium Interference level Negative interference above 150 µg/l No interference at or below 1000 µg/l Positive interference above 250 µg/l No interference at or below 47,000 µg/l Positive interference above 250 µg/l No interference at or below 1000 µg/l Positive interference above 1000 µg/l Negative interference above 79,000 µg/l Prepare analysis labware Fully clean all containers used in this test to remove possible traces of calcium or magnesium. Use plastic containers for all analysis and storage. Before analysis, clean the Pour-Thru cell and all labware as follows: Accuracy check 1. Clean all containers by normal means, then rinse with ultra-pure water. 2. Fill and soak all containers for 10 minutes with a 1:25 dilution of Chlorophosphonazo Reagent in ultra-pure water. Note: If containers are rinsed and capped after each use, only occasional soaking is necessary. 3. Rinse well with ultra-pure water. 4. Keep containers tightly closed and dedicate them for ULR Hardness only. 5. Fill the Pour-Thru cell with this same mixture of chlorophosphonazo and water. Let the mixture soak in the cell for several minutes. 6. Rinse the Pour-Thru Cell with ultra-pure water. 7. Use a wash bottle to fully rinse the inside of the bottle-top dispenser cap and inlet tubing with a large volume of ultra-pure water. 8. Put the inlet tubing of the bottle-top dispenser into a beaker of ultra-pure water and depress the plunger 10-15 times to rinse the dispenser. Note: For best results, pour a small amount of reagent into the beaker of rinse water. 9. Remove the dispenser from the water and depress the plunger until all of the water has been expelled. Shake off any excess water on the dispenser. 10. Install the dispenser on the Chlorophosphonazo Reagent bottle and tighten the cap. Standard additions method (sample spike) Use the standard additions method (for applicable instruments) to validate the test procedure, reagents and instrument and to find if there is an interference in the sample. Items to collect: 50,000-µg/L (50-mg/L) as CaCO 3 Calcium Chloride Standard Solution Flask, volumetric, 50-mL plastic 20-mL volumetric pipet, Class A and pipet filler safety bulb Pipet, TenSette, 0.1 1.0 ml and tips Ultrapure water 4 Hardness, Chlorophosphonazo RL Method (1000 µg/l)

1. Prepare a 20,000-µg/L (20-mg/L) hardness standard solution as follows: a. Use a pipet to add 20 ml of a 50-mg/L Calcium Chloride Standard solution into a 50-mL plastic volumetric flask. b. Dilute to the mark with ultrapure water. Mix well. Prepare this solution daily. 2. Use the test procedure to measure the concentration of the sample, then keep the (unspiked) sample in the instrument. 3. Go to the Standard Additions option in the instrument menu. 4. Select the values for standard concentration, sample volume and spike volumes. 5. Prepare three spiked samples: use the TenSette pipet to add 0.2 ml, 0.4 ml and 0.6 ml of the prepared standard solution, respectively, to three 50-mL portions of fresh sample. Mix well. 6. Use the test procedure to measure the concentration of each of the spiked samples. Start with the smallest sample spike. Measure each of the spiked samples in the instrument. 7. Select Graph to compare the expected results to the actual results. Note: If the actual results are significantly different from the expected results, make sure that the sample volumes and sample spikes are measured accurately. The sample volumes and sample spikes that are used should agree with the selections in the standard additions menu. If the results are not within acceptable limits, the sample may contain an interference. Standard solution method Use the standard solution method to validate the test procedure, the reagents and the instrument. Items to collect: 0.50-mg/L (500-µg/L) as CaCO 3 Calcium Chloride Standard Solution 1. Use the test procedure to measure the concentration of the standard solution. 2. Compare the expected result to the actual result. Note: The factory calibration can be adjusted slightly with the standard adjust option so that the instrument shows the expected value of the standard solution. The adjusted calibration is then used for all test results. This adjustment can increase the test accuracy when there are small variations in the reagents or instruments. Method performance The method performance data that follows was derived from laboratory tests that were measured on a spectrophotometer during ideal test conditions. Users can get different results under different test conditions. Program Standard Precision (95% confidence interval) Sensitivity Concentration change per 0.010 Abs change 227 500 µg/l CaCO 3 495 505 µg/l 8 µg/l Ca Summary of Method Calcium and magnesium mix equally with the Chlorophosphonazo Indicator to give a complex with color which absorbs light very strongly at 669 nm. One drop of the CDTA reagent breaks up this complex. The decrease in color is proportional to the amount of calcium and magnesium in the sample (as CaCO 3 ). The measurement wavelength is 669 nm. Consumables and replacement items Required reagents Description Quantity/test Unit Item no. Chlorophosphonazo Indicator Solution 2 ml 500 ml 2589549 CDTA Reagent for Ultra Low Range Hardness 2 drops 30 ml SCDB 2589723 Hardness, Chlorophosphonazo RL Method (1000 µg/l) 5

Required apparatus Description Quantity/test Unit Item no. Cylinder, graduated, 50 ml, polypropylene 1 each 108141 Dispenser, adjustable volume, 1.0 5.0 ml 2 each 2563137 Flask, Erlenmeyer, Polymethylpentene, screw cap, 125 ml 2 each 2089843 Recommended standards Description Unit Item no. Calcium Standard Solution, 50 mg/l as CaCO 3 946 ml 2127716 Calcium Standard Solution, 0.50 mg/l as CaCO 3 946 ml 2058016 Water, Ultra-pure, Aldehyde-free 500 ml 2594649 Optional reagents and apparatus Description Unit Item no. Pipet, TenSette, 0.1 1.0 ml each 1970001 Pipet tips for TenSette Pipet, 0.1 1.0 ml 50/pkg 2185696 Pipet, volumetric Class A, 20 ml each 1451520 Pipet filler, safety bulb each 1465100 Flask, volumetric, 50 ml, polypropylene each 1406041 FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932 Hach Company/Hach Lange GmbH, 2007, 2010, 2012, 2014, 2017. All rights reserved. 11/2017, Edition 9