REV. 9/08. Mounting Structure PRE-INSTALLATION GUIDE for Ergon Skybooms

Similar documents
SERIES SURGICAL LIGHTS

SERIES SURGICAL LIGHTS

PRE-INSTALLATION GUIDE

Installation Manual. Retractable Ceiling Column Models 8301 & 8301PC

This guide can be used to prepare a specification for incorporating free standing jib cranes into a competitively bid construction project.

INSTALLATION MANUAL RECTANGULAR COLUMN

INSTALLATION INSTRUCTIONS

This guide can be used to prepare a specification for incorporating free standing jib cranes into a competitively bid construction project.

SECTION STRUCTURAL METAL FRAMING

SECTION (16071) - VIBRATION AND SEISMIC CONTROLS FOR ELECTRICAL SYSTEMS

MODEL SPECIFICATIONS: Hydraulically Driven Steel Underpinning Pile Specifications

SECTION HANGERS AND SUPPORTS FOR ELECTRICAL SYSTEMS

SECTION VIBRATION AND SEISMIC CONTROLS FOR ELECTRICAL SYSTEMS

SECTION SEISMIC CONTROL OSHPD

This guide can be used to prepare a specification for incorporating wall bracket jib cranes into a competitively bid construction project.

SECTION VIBRATION CONTROLS FOR PLUMBING PIPING AND EQUIPMENT

WALL MOUNTED ARTICULATING JIB CRANE

Gallery Wall Rough-In Assembly P6918

SEISMIC PROTECTION SYSTEMS Bracing Products for Racks & Cabinets

Student Services & Classroom Addition SECTION VIBRATION AND SEISMIC CONTROLS FOR ELECTRICAL SYSTEMS

Installation and Maintenance Manual for EMH Freestanding Workstation Jib Cranes

SECTION VIBRATION AND SEISMIC CONTROLS FOR ELECTRICAL SYSTEMS

SECTION GROUNDING AND BONDING FOR ELECTRONIC SAFETY AND SECURITY

SECTION VIBRATION CONTROLS FOR HVAC PIPING AND EQUIPMENT

Stainless Steel Aeration Equipment Specification

FIXED CEILING COLUMN. Installation Manual FIXED CEILING COLUMN

Foundation Specifications for 3.6-/3.7-Meter Earth Station Antennas

Foundation Specifications

SEISMIC RESTRAINT SYSTEM

CEILING MOUNTED ARTICULATING JIB CRANE

NORTHWESTERN UNIVERSITY PROJECT NAME JOB # ISSUED: 03/29/2017

Compression Superstructure TELETOM 500/700 Series Equipment Management Booms TELETOM 400/600 Series Anesthesia Booms

Fig Beam Clamp Retaining Strap (B-Line B3367)

Cantilevered Rail System

MICHIGAN DEPARTMENT OF TRANSPORTATION SPECIAL PROVISION FOR LOW-TENSION CABLE BARRIER. OPR:CT 1 of 5 APPR:JAR:DBP: FHWA:APPR:

JAMISON SOUND DOOR SPECIFICATIONS VERTICAL SLIDING SOUND REDUCTION DOOR - STC 53 POWERED OPERATION

SECTION ENCLOSED BUS ASSEMBLIES

SECTION CRL70 HEAVY GLASS TOP HUNG SLIDING DOOR SYSTEMS

SECTION VIBRATION CONTROLS FOR HVAC PIPING AND EQUIPMENT

ELECTRICAL CONTRACTORS THE UNIVERSITY OF TENNESSEE KNOXVILLE, TENNESSEE SECTION PAGE 1 SECTION SUPPORTING DEVICES

L SERIES NEBS COMPLIANT STANDBY BATTERY RACKS ASSEMBLY INSTRUCTIONS

SECTION ENCLOSED BUS ASSEMBLIES

SECTION EZ-SLIDE 80 AND 180 HEAVY GLASS TOP HUNG SLIDING DOOR SYSTEMS

STRUCTURAL STEEL FRAMING

SECTION VIBRATION AND SEISMIC CONTROLS FOR HVAC PIPING AND EQUIPMENT

Midwest Roadside Safety Facility

NORTHWESTERN UNIVERSITY PROJECT NAME JOB #

Table of Contents... 1, 2. (EF) Equipment Floor... VISR EF 102 Water Heater VISR EF 104 Chiller VISR EF 105 End Suction Pump...

SECTION CRL50 AND CRL51 HEAVY GLASS TOP HUNG SLIDING DOOR SYSTEMS

1. Hangers and supports for electrical equipment and systems.

C. OSHPD: Office of Statewide Health Planning and Development for the State of California.

SECTION LAGUNA SERIES TOP HUNG SLIDING DOOR SYSTEM

DARTMOUTH COLLEGE DESIGN September 15, 2004 &CONSTRUCTION GUIDELINES

Diffused Aeration Piping Equipment

CANTILEVER RACK SPECIFICATIONS PART 1 GENERAL 1.1 SCOPE 1.2 APPROVED MANUFACTURER 1.3 REGULATORY ORGANIZATIONS AND GROUPS 1.4 QUALITY ASSURANCE

ELECTRICAL SECTION 2

NORTH HARRIS COUNTY REGIONAL WATER AUTHORITY PIPE HANGERS, Section PIPE HANGERS, SUPPORTS, AND RESTRAINTS

1.02 RELATED SECTIONS: Water supply piping, water use piping, electrical, irrigation, foundations, site plans.

1.02 RELATED SECTIONS: Water supply piping, water use piping, electrical, irrigation, foundations, site plans.

MODELS: SM & SMSS

UNIVERSITY SERVICES ANNEX James Madison University Harrisonburg, Virginia State Project Code: Architect s Project Number:

Essex County College - West Essex Campus Addition And Renovations dlb # / SECTION CABLE TRAYS FOR COMMUNICATIONS SYSTEMS

Privacy Plus Partitions

Item 442 Metal for Structures

WALL CANTILEVER WORK STATION JIB CRANE

NORTHWESTERN UNIVERSITY PROJECT NAME JOB # ISSUED: 03/29/2017

Large cable sizes stronger than pipe/strut

EVALUATION REPORT. PFC-5551 Reissued April 1, Filing Category: DESIGN Steel (038)

SECTION FACILITY LIGHTNING PROTECTION

Water & Industrial Products Catalog

SPECIAL SPECIFICATION 1924 Neighborhood Friendly High Mast Illumination System

1. Hangers and supports for electrical equipment and systems. 2. Construction requirements for concrete bases.

INSTRUCTIONS FOR ASSEMBLING NEBS FLOODED RACKS

05510: TAPERED STEEL TROLLEY POLES AND ACCESSORIES

Concrete Inserts. B Adjustable Metal Deck Ceiling Bolt

SECTION GROUNDING

1. Temperature Change: 120 deg F (67 deg C), ambient; 180 deg F (100 deg C), material surfaces.

STANDARDIZED STRUCTURES

SECTION MODULAR GALVANIZED RF SHIELDING SYSTEM SPECIFICATIONS

SECTION ALUMINUM GATES

RHE3 Dock Leveler Installation Manual

1.2 ACTION AND INFORMATIONAL SUBMITTALS.1 Provide submittals in accordance with Section Submittal Procedures.

Section 7: Specifications

SECTION LOW-VOLTAGE TRANSFORMERS

CITY OF LOS ANGELES CALIFORNIA

Student Services & Classroom Addition

TYPICAL INSTALLATION INSTRUCTIONS

Seismic Restraints. Cent-R-Rail Supplement* SYSTEMS THAT MAKE SENSE

Title: YALE OFFICE OF FACILITIES PROCEDURE MANUAL Chapter: 01 - Yale Design Standard Division: Electrical Standards

Test Chain Instruction Manual

CONSOLIDATED EDISON COMPANY OF NEW YORK, INC. 4 IRVING PLACE NEW YORK, NY DISTRIBUTION ENGINEERING DISTRIBUTION EQUIPMENT

HIGH MAST CAMERA POLE ASSEMBLY GENERAL

SECTION VIBRATION ISOLATION FOR PLUMBING PIPING AND EQUIPMENT

SECTION MISCELLANEOUS METAL FABRICATIONS

SECTION CHAIN-LINK FENCES AND GATES 02821/1

Operation & Service Manual

DIVISION 15-MECHANICAL SECTION SUPPORTS AND ANCHORS PN PART 1 - GENERAL 1.01 SECTION INCLUDES. A. Pipe and equipment support systems.

SECTION CLIMBING WALLS

SECTION RADIO FREQUENCY SHIELDING ENCLOSURE RF Modular Pan System

A. Section includes busway and fittings including plug-in units.

Transcription:

REV. 9/08 Mounting Structure PRE-INSTALLATION GUIDE for Ergon Skybooms

SKYTRON MOUNTING STRUCTURES for Ceiling Mounted Equipment Foreword Skytron s objective is to provide a program and guideline to assist individuals (hospital owners, architects, structural engineers) responsible for the design of mounting structures. The mounting structure of a Skytron fixture should always be considered the most important detail of any project prior to installation. Skytron s ceiling mounted systems depend upon properly designed and installed mounting structures to deliver years of dependable, service-free performance. We realize that individual mounting structures vary in design due to unique structural requirements, physical obstructions, structural member availability and room layout limitations, to name a few. The mounting structure design in this Mounting Structure Guideline Booklet is a successful, proven design. The design consists of a welded, flanged tube assembly combined with (4) angled sway braces ( kickers ). A structural steel tube combined with welded steel plates on each end facilitate its attachment to the structural ceiling and provide an attachment point for the Skytron fixture. This is a simple effective design that is forgiving to restricting structural conditions that are usually encountered during fabrication. This design also eliminates guess work by the steel fabricator when compared to structures that are fabricated on site with angle iron. In many cases the flanged tube assembly can be pre-fabricated reducing the on-site construction time and eliminating the need to correct structural problems associated with angle iron structures. The process of building a proper mounting structure can be complemented with Skytron s continued support on the project. Once the structures are fabricated, Skytron will perform a visual inspection and on-site consultation followed by an actual performance test. The performance test of the structure essentially involves hanging a test jig from the structure, and then measuring the amount of rotation that occurs at the mounting plate using an inclinometer (digital level). The test jig is similar in size, shape, and weight to a Skyboom. In other words, once the test jig is installed and the proper weights have been added to simulate the product s moment load, a reading is taken on the digital level to verify that there is no more that two-tenths of a degree of rotational movement at the mounting plate. Rotation of the mounting plate causes the radial arm(s) to become out of level and drift. For a structure to meet Skytron specifications, we require that the mounting plate does not rotate more that two-tenths of a degree while loaded with the specified weight and moment. The testing process should occur in the early stages of construction to provide optimal time if additional reinforcement is needed on the structure. This should be performed prior to the completion of the finished ceiling. Skytron s program is available to help you and provide you with the necessary counseling. However, the final responsibility to insure that the structure is adequate and meets specification lies with the structural engineer and the contractor for the project. Our support services are geared to minimize the effort on all parties involved and to insure a successful product installation. This Mounting Structure is applicable for all SKYTRON Skyboom models. We recognize that not all situations will permit the use of this Tube Structure. Please contact SKYTRON with your special needs so that we may be able to guide you to other alternatives. Page 1

INTRODUCTION This publication is meant to provide the information needed to design and build a mounting structure that will provide the proper support for a SKYTRON Skyboom fixture as well as the guidance for supplying the necessary services. The responsibilities and items supplied by both SKYTRON and the Customer for pre-installation of the fixture will vary depending on the equipment and services that will be provided on the Skyboom fixture. The mounting structure requirements however, are the same for all models. The structure design is based on the heaviest Skyboom model with the greatest moment load. By using this criteria, one structure design can be utilized to minimize design costs and provide adequate support for future modifications. If applicable, specific Seismic Calculations can be provided from SKYTRON. The Mounting Structure design should always be approved by a Structural Engineer. An inadequate Mounting Structure may allow drifting and other performance problems for the fixture. Equipment warranty or service charges due to an inadequate Mounting Structure are at the customer s expense. SKYTRON does not warrant or certify superstructure designs. This publication is intended to address various areas and responsibilities of multiple individuals. It is important that this information is shared with: - Architects and Structural Engineer - Electrical Engineer - Medical Gas/Piping Engineer - Communication/Video/Data Engineer - Project Manager(s) SKYTRON Supplied Components SKYTRON supplies the following items to assist in the design and construction of the mounting structure: Customer approved room layout and Skyboom configuration drawings. Skyboom mounting plate Medical gas riser plate with appropriate medical gas connectors Electrical junction box Fixture mounting hardware Pump Enclosure (if applicable) CUSTOMER Supplied Components and Responsibilities Mounting Structure - Architect and Structural Engineer The Customer is responsible for the design and fabrication of the structure which connects the SKYTRON mounting plate to the structural ceiling. The mounting plate must be perfectly level (±0.1 ). The design and fabrication must include the necessary bracing to prevent twisting or lateral motion and to allow no more than two-tenths of a degree (0.2 ) of rotation at the mounting plate when the specified load is applied. Refer to Page 5 for recommended specifications. An inadequate mounting structure can result in damage to the equipment. Equipment warranty charges related to an inadequate structure design or installation are at the customers expense. The mounting structure design specifications included in this publication depict a successful, proven design. The actual building design, additional loading required by local building codes and various obstacles may exist that do not allow this design system to be followed. It is the customer s responsibility to provide the appropriate alternate mounting structure. SKYTRON recommends that a Structural Engineer be consulted whenever a design variation is needed. Page 2

Electrical Provisions - Electrical Engineer Each Skyboom fixture is fabricated in accordance to the specifications required by the customer. The customer is responsible for the appropriate electrical supply to the fixture. The Configuration drawings supplied by SKYTRON will indicate the type and quantity of circuits required. The customer is responsible for connections from facility supply to fixture, wall control and wiring installation for surgical lights and pump motor connection if applicable. Refer to Page 7 for Electrical Requirements. Medical Gas Provisions - Medical Gas/Piping Engineer Each Skyboom fixture is fabricated in accordance to the specifications required by the customer (to include the type of faceplate connectors to be provided on the fixture). The customer is responsible for the appropriate medical gas supply and connection to the fixture. The Configuration drawings supplied by SKYTRON will indicate the type and quantity of gas supply lines required. Refer to Page 7 for Medical Gas Requirements. Communications Provisions - Communication/Video/Data Engineer Each Skyboom fixture is fabricated in accordance to the specifications required by the customer. The customer is responsible for the appropriate communication cable routing to the fixture. The Configuration drawings supplied by SKYTRON will indicate the type and quantity of cables required. Refer to Page 7 for Communications Requirements. Testing NOTE A visual review of the mounting structure construction should be requested before the finished ceiling is installed to detect possible interference potentials, and arrangements can be made to conduct a load simulation using a Test Jig. This process is designed to simulate the fixture load and is an effective means to determine where the structure may require improvement. It is not a certification of the structure. The customer is responsible for the testing and any necessary certification requirements for all system utilities prior to the installation of the Skyboom fixture. The testing and certification must be in accordance with all state, federal, local, NFPA and NEC codes and must include the following items: Mounting Structure (Testing performed by Skytron) Electrical Wiring Medical Gas Supply INSTALLATION KIT SKYTRON provides an installation kit which is typically shipped in advance and contains all of the Skytron supplied components required for pre-installation and the fabrication of the Skyboom Mounting Structure. The Installation Kit includes the following: -Customer approved room layout and Skyboom configuration drawings. -Mounting Plate -Medical gas riser plate(s) with appropriate D.I.S.S. medical gas connectors -Electrical junction box -Mounting hardware for riser plate and junction box Page 3

17.50 1.00 4.25 9.00 4.25 9.50 4.25 0.25 1.00 Ø10.63 9.50 Ø1.31 (6 places) Ø6.00 +0.15-0.00 9.75 9.00 1.50 Ø1.00 (4 places) 9.75 1.00 4.25 1.50 R0.50 27.5 0.25 030504.02b NOTE: This illustration depicts a typical mounting structure design and its components. Always consult specific structural criteria defined by a structural engineer. NOTE: Do not cover or block any holes with sway bracing, gussets, weld, weld slag or etc. STRUCTURAL CEILING STRUCTURAL CONCRETE SLAB NORMAL WEIGHT fc = 3000 PSI (20,684 kpa) STRUCTURAL CEILING 11 HIGH CLEAR SPACE ABOVE MOUNTING PLATE REQUIRED ON 3 SIDES FOR GAS AND ELECTRICAL COMPONENT INSTALLATION SWAY BRACING 3"x 3"x 1/4" (4) MOUNTING PLATE (SKYTRON SUPPLIED) FIXTURE HUB 45-60 TYP 1/4" PJP TYP 1/4" TOP & BOTTOM GUSSET 6" DIA SUPPORT TUBE STRUCTURAL CEILING PLATE 17" DIA x 1" 12" MIN / 60" MAX * 10" (-0"/+1") EXPANSION ANCHORS: HILTI KB-II (ICBO 4627) CARBON STEEL ACCESS DOOR 24" X 24" (REQUIRED) 20" DIAMETER OPENING REQUIRED (CENTERED UNDER MOUNT) FINISHED CEILING * CRITICAL DIMENSION 030504.03mj_2 Page 4 NOTE: Typical dimensions shown. Refer to specific structural drawings and/or Seismic drawings for each application.

STRUCTURAL REQUIREMENTS - Architect and Structural Engineer MOUNTING STRUCTURE COMPONENTS The fabrication of each mounting structure may be slightly different but they each require the same basic components to ensure stability. Refer to the figure on page 4 for the relationship of these components. Sway Bracing (by others) Sway bracing is designed to rigidly affix the Mounting Plate to the structural ceiling. The primary purpose of Sway Bracing is to eliminate sway, or lateral twisting and flexing of the mounting structure as it reacts to dynamic load changes caused by moving the fixture radial arms. The sway bracing should be welded to the Mounting Plate and extend away from the center of the mount. A minimum of four sway braces placed 90 apart and positioned at a 45 to 60 angle is recommended. Minimum recommended material for sway bracing is 3 x 3 x 1/4 angle iron. It is recommended that in all applications that the sway bracing be fastened to the structural ceiling. Structural Ceiling Plate (by others) The Structural Ceiling Plate rigidly attaches the mount to the Structural Ceiling using structural anchors appropriate for the ceiling construction. The Structural Ceiling Plate should be a minimum of, 1 ASTM A36 steel plate,17 diameter with (6) 5/8 diameter holes for structural anchors and is fabricated by others. Expansion Anchors Test 50% of the anchors at 2000 pounds (907 kg) tension, or 50 ft lb (68 N m) torque per CBC 1925A.3.5. Installed anchors must meet the following criteria: 1. Hydraulic Ram Method: The anchor should have no observable movement at the applicable test load. For wedge and sleeve type anchors, a practical way to determine observable movement is that the washer under the nut becomes loose. 2. Torque Wrench Method (Wedge or Sleeve Type): The applicable test torque must be reached within one-half (1/2) turn of the nut. Testing should occur no sooner than 24 hours after installation of anchors. If any anchor fails testing, test all anchors until 20 consecutive anchors pass, then resume the initial testing frequency. Test equipment is to be calibrated by an approved testing laboratory in accordance with standard recognized procedures. Support Tube (by others) The Support Tube required to attach the Mounting Plate to the Structural Ceiling Plate is ASTM 500 Grade B, 6 diameter tube. Support Tube is to be welded to Structural Ceiling Plate and Mounting Plate. A minimum of 6 gussets placed 60 apart should be welded to the support tube at the structural ceiling plate and the Mounting Plate. Mounting Plate (Skytron supplied) The 17.5 x 17.5 x 1 ASTM A36 steel Mounting Plate is a Skytron supplied item. The Support Tube and Sway Bracing are welded to the Mounting Plate. The mounting plate contains the corresponding bolt pattern for attaching the fixture and provides the mounting areas for the junction box and gas riser plates. MOUNTING STRUCTURE DESIGN Seismic structural applications may differ. Please contact your local SKYTRON distributor for specific calculations. The Mounting Structure must be designed and fabricated to position the bottom of the SKYTRON Mounting Plate 10 (-0 +1 ) above the finished ceiling. This is a critical dimension in order to accommodate proper clearance required for ceiling cover function. The mounting plate must be perfectly level (±0.1 ) and allow no more than two-tenths of a degree (0.2 ) of rotation at the mounting plate when the specified load is applied. The Mounting Structure must be tested for strength and stiffness prior to installation of the fixture. A Test Jig is available from SKYTRON that includes all components and documentation required for performing an approved Mounting Structure test. Refer to page 8 for testing procedures. Please consult your SKYTRON representative during early stages of construction to facilitate this process. The testing process is a required, documented function prior to closing of the finished ceiling. Ceiling Requirements A 24 x 24 access door must be mounted adjacent to the mounting structure for entry by service personnel for service access. SKYTRON provides a 24 diameter ceiling cover designed to cover a 20 diameter ceiling hole cutout. Page 5

PUMP ENCLOSURE 18 L x 6 W x 12 H 1-1/4" THREADED ROD (6) BRACKET WELDED MOUNT ASSEMBLY (SUPPLIED BY OTHERS) JUNCTION BOX RISER PLATE INLET TUBE 1-1/4" HEX NUT (24) SKYBOOM FIXTURE 030504.04rr NOTE: This illustration depicts a typical mounting structure design and its components. Always consult specific structural criteria defined by a structural engineer. Mounting bolts and nuts shipped with fixture. Page 6

Electrical Requirements - Electrical Engineer Each Skyboom fixture is fabricated in accordance to the specifications required by the customer. The Configuration drawings supplied by SKYTRON will indicate the type and quantity of circuits required. SKYTRON provides all wiring and electrical materials for connection from fixture to junction box or pump enclosure (if applicable). SKYTRON supplies either an electrical junction box to facilitate field wiring for up to six circuits that is mounted on the mounting plate in the correct position OR if applicable, a hydraulic pump enclosure/junction box (18 L x 6 W x 12 H) that is to be remote mounted within 24 of the mounting structure (by contractor). The pump enclosure can be shipped with the Installation Kit upon request. Typical wire type is 12AWG, 600V, XHHW-2. Each circuit requires a separate, properly circuit protected, 120 VAC, 60 Hz power supply line enclosed in rigid metal conduit. All electrical materials for connection to SKYTRON supplied junction box or pump enclosure and installation labor for such materials to be provided by customer. All wiring and materials to be in accordance with federal, state and local codes. It is the customers responsibility to meet conformity to NFPA and NEC standards with respect to the number of receptacles provided in a patient care area. (Page 4 - Drawing Package) Medical Gas Requirements - Medical Gas/Piping Engineer NOTES: Medical gas riser plates comply with NFPA 99-2005 Edition Cleanliness of gas outlets must be maintained through installation. Purge gas lines and test at least 24 hrs. prior to installation. Test results must be made available to SKYTRON for verification & comparison. Each Skyboom fixture is fabricated in accordance to the specifications required by the customer. The Configuration drawings supplied by SKYTRON will indicate the type and quantity of gas supply lines required. D.I.S.S. connection medical grade hoses connect the fixture to the Riser Plate connectors. The customer is responsible to deliver the appropriate medical gas from the facility supply to the Riser Plate connectors and for the design of a medical gas system with adequate flow capacity capable of compensating for the accumulative flow restrictions associated with conventional construction methods i.e. flex gas hoses. SKYTRON provides medical gas riser plate(s) and the appropriate connectors for attachment to the Mounting Plate. The Riser Plate attaches to the Mounting Plate and will accommodate up to 12 gas connectors. The connectors are D.I.S.S. type medical gas connectors with single check valves and provide a 1/2 copper tube for attachment to facility supply lines. Caps are provided with Vacuum & WAGD gas connectors to facilitate testing procedures. All connection and testing of Medical Gas Piping to be performed in accordance with NFPA 99-2005 Edition-5.1.6 requirements. - Evacuation lines and fittings rated at 70-75 l/min (2.47-2.64scfm) at user connection point. - Vacuum lines (WAGD) and fittings rated at 85 l/min (3 scfm), based on 115 l/min facility supply - Nitrogen system requires Nitrogen supplied directly from facility supply at 185 psi minimum to 200 psi maximum (if applicable). Avoid designs which feature multiple in-line Nitrogen control systems in order to avoid loss of flow capability. (Page 3 - Drawing Package) Recomended Gas Riser Placement - When mount is in center of room, Riser Plate should face head end of room, when mount is near walls, Riser Plate should be adjacent to wall. If more than one Riser Plate is required they should be placed next to each other. Communications Requirements - Communication/Video/Data Engineer Each Skyboom fixture is fabricated in accordance to the specifications required by the customer. The Configuration drawings supplied by SKYTRON will indicate the type and quantity of cables required. The customer is responsible for the appropriate communication cable routing to the fixture. Special arrangements can be coordinated for custom cable sets to be installed at the time of installation. Contact your SKYTRON representative. (Page 5 - Drawing Package) ADDITIONAL SKYTRON SUPPLIED ITEMS In addition to the installation kit as outlined on page 3, SKYTRON provides the following items: (6) 1-1/4 x 10 Threaded Rods, (24) 1-1/4 Hex Nuts, Pump Enclosure (if applicable) Page 7

TESTING - MOUNTING STRUCTURE Each mounting structure must be tested by a SKYTRON representative to verify that there is no more than two-tenths of a degree of rotational movement at the mounting plate prior to installing a fixture. A Mounting Structure Test Jig is available from SKYTRON to facilitate this process. The Mounting Structure Test Jig is a fixture which simulates the weights and moment loads created by the SKYTRON fixture. It consists of a Hub Assembly, an Upper and Lower Radial Arm, a Chain Fall, Weight Support Tube and six 100lbs. weights (refer to illustration on page 9). The Drawing Package and the Test Jig Instructions are required to perform an acurate test. The Test Jig Instructions include the Load Simulator Weight Chart, the Test Jig Report Form and complete instructions for performing the test. RECOMMENDED TOOLS Ladder(s) 19mm Socket (3/8 Drive) or 19mm wrench Genie Lift / Overhead Hoist 3/8 Drive Ratchet Two 1-7/8 wrenches Two or more people capable of lifting 100 lbs. Appropriate cart Page 8

SKYTRON MOUNTING STRUCTURE TEST JIG 1-1/4 NUTS 1-1/4 THREADED ROD (6) MOUNTING HUB ASSY. UPPER RADIAL ARM 12mm NUT (6) LOWER RADIAL ARM 12mm NUT (4) 12 mm x 30mm BOLT (4) 5/8 NUT (2) CHAIN CHAIN FALL CHAIN 3/8 HITCH PIN WEIGHT SUPPORT TUBE WEIGHTS (6) Page 9

comm/sky/skyboom/mounting/mounting guide/ergon pre-install guide 9-08 5085 Corporate Exchange Blvd.S.E. Grand Rapids, MI 49512 1-800-SKYTRON or 1-616-656-2900 FAX 1-616-656-2906