Software Regulation: The Transfusion Medicine Experience

Similar documents
Introduction to the FDA. Historical Context. Changes over Time FDA - BE-200 9/23/2004. Copyright C. S. Tritt, Ph.D. 1

Draft Guidance for Industry, Clinical Laboratories, and FDA Staff. In Vitro Diagnostic Multivariate Index Assays

The FDA/CDRH Perspective in the Regulation of Surgical Mesh

Paths to Market & FDA Product Lifecycle Regulation. Patricia Kaeding Design of Medical Devices University of Minnesota April 14, 2005

Oversight of Laboratory Developed Tests

MOBILE MEDICAL APPLICATIONS REVIEWING FDA GUIDANCE FOR INDUSTRY USE

First Global Forum on Medical Devices Bangkok-Thailand, September Regulatory Control of Medical Devices in Tanzania

fix Menu FDA Pharm 411 Medical Devices Overview of Device Regulations

Device research sponsors, whether companies or investigators, are held responsible for meeting the same regulations.

3.1. Overall Principal Investigator (PI), who holds the IDE and/or is the Sponsor

Title: Review of Medical Devices Page: 1 of 5 Written by:

The Institutional Review Board (IRB) Manual

Role of the WHO IVD Prequalification Programme in Light of National Regulatory Authority Approval

Disclosures. Laboratory Stakeholders. IVD vs. LDT. FDA Regulation of Laboratory Developed Tests 10/2/2015. FDA Regulation of LDTs

Navigating the FDA, and understanding Medical Device Preemption. Special Counsel Foulston Siefkin

Medical Device Accessories: Defining Accessories and Classification Pathway for New Accessory Types. Draft Guidance for Industry and

Reducing Unnecessary Medical Imaging Exposure

Carequality Governance Charter

INNOVATION ACTION PLAN

MEDICAL DEVICE CLINICAL INVESTIGATIONS AND ISO 14155

Niccolo Machiavelli (1523)

MCW Office of Research Standard Operating Procedure

Molecular Diagnostics Regulation Shifting from a Biomarker-Based Approach to an Algorithm-Based Approach

The Children s Hospital of Philadelphia Committees for the Protection of Human Subjects Policies and Procedures Determination of IND/IDE Requirement

Stephanie Moravec, MPH October 25, 2017

The In Vitro Diagnostic CRO

INVESTIGATIONAL DEVICES & DRUGS. What is an IRB to do?

OMBU. Ombu Enterprises, LLC Ombu The Operational Excellence Company Ph: Fax: ENTERPRISES, LLC

FDA Regulation of Diagnostic Tests An Evolving World

ANDREW BALO SENIOR VICE PRESIDENT, REGULATORY, BIOMETRICS, & CLINICAL AFFAIRS DEXCOM, INC 6340 SEQUENCE DRIVE SAN DIEGO, CA 92121

Medical Device Classification

Perspective: Convergence of CLIA and FDA Requirements A Rational Shift in the Regulatory Paradigm

FDA s Proposed Regulatory Changes to Laboratory-Developed Tests (LDTs) By Roz Sweeney, Ph.D., Nerac Analyst

The Convergence of Software in the Medical Device Industry. Joseph Azary

Deciding When to Submit a 510(k) for a Change to an Existing Device Guidance for Industry and Food and Drug Administration Staff

International Standards and EU regulation of medical device software an update

Medical Device Regulatory Framework 9 SEPTEMBER 2015 FUNDISA CONFERENCE JANE ROGERS

U.S. FDA CENTER FOR DEVICES AND RADIOLOGICAL HEALTH UPDATE

Research: Ethics, Informed Consent, FDA, Off Label Use

FDA identifies this generic type of device as: Absorbable perirectal spacer.

Regulatory Overview of Proposed LDT Framework. FDA Concerns. Background. FDA Proposed Regulatory Approach. By Ben Berg, Meaghan Bailey, RAC

Medical Devices; Immunology and Microbiology Devices; Classification of the Automated

UDI: The Ins and Outs of Global Compliance and Value

Molecular Diagnostics: The Shift to Complexity. Molecular Diagnostics Regulation: Where do we go from here? David W Feigal, Jr.

Medical Devices; Obstetrical and Gynecological Devices; Classification of the Software

Due diligence in the European medical devices industry

FDA Regulatory Environment for Wearables, Implantables, and Other Digital Health Devices

Copyright. Jeremiah J. Kelly (2015). All rights reserved. Further dissemination without express written consent strictly prohibited.

Is FMT A Drug? Lance Shea, M.S., J.D. Washington Square, Suite Connecticut Ave., NW Washington, DC, D

The GS1 System and services (GTIN, GLN and GDSN) can be used to comply with this rule.

Tuesday, 21 October Jang Yong Choi

Devices 3/26/2014. Investigational Devices. What is considered a device?

FDA PLANS TO REGULATE TESTS AS DEVICES. August 2010 SPECIAL REPRINT. By Ellen Flannery and Scott Danzis

Office for Human Subject Protection. University of Rochester

TERMS AND DEFINITIONS

[[Page 63034]] ======================================================================= DEPARTMENT OF HEALTH AND HUMAN SERVICES

Niccolo Machiavelli (1523)

FDA Regulation of Testing

Re: Docket No. FDA-2017-D ; Draft Guidance on Recommendations for Dual 510(k) and CLIA Waiver by Application Studies

Convergence of Clinical Engineering and Information Technology: Trends, Opportunities & Challenges

June 26, Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane, Room 1061 Rockville, MD 20852

Medical Devices; Immunology and Microbiology Devices; Classification of the Nucleic Acid-

The Medical Device Coordination Group: a new Authority Under EU Device Regulations

Medical Devices; Immunology and Microbiology Devices; Classification of the Brain Trauma

The Future of Pharmaceutical Quality

Combination Products at US FDA

Update on the IVDR. Sue Spencer

U.S. Food and Drug Administration (FDA) Regulatory Pathways for Medical Devices

Docket No. FDA-2011-D-0215 Draft Guidance for Industry and FDA Staff on In Vitro Companion Diagnostic Devices

Medidée Services SA. Nano-Tera.ch. 05 February 2015 part 8. PMA, 510k, IDE. Pierre-Alain Sommer

RE: Docket No. FDA 2017 N 0041: Agency Information Collection Activities; Proposed Collection; Comment Request; Safety Assurance Case

External Defibrillator Improvement Initiative

Points to Consider for Marketing of Computerized Surgical Systems in the U.S.

Testing: The Critical Success Factor in the Transition to ICD-10

Distinguishing Medical Device Recalls from Medical Device Enhancements

October 26, Unique Device Identification System (FDA-2011-N-0090) Dear Commissioner Hamburg:

FDA Regulation of In Vitro Diagnostics: Current Perspectives and Initiatives

Drug Safety and FDA Revitalization Legislation

Table of Contents. Role of Standards in the Assessment of Medical Devices Study Group 1 Final Document GHTF/SG1/N044:2008

The Intersection of Genomics Research and the IDE Regulation

March 19, Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane Room 1061 Rockville, MD 20852

Medical Devices; Neurological Devices; Classification of the Non-Electroencephalogram

So You Want to Develop a Test? Navigating the Regulatory Waters of Laboratory Developed Testing

FDA's Initiative to Regulate Lab- Developed Tests (LDT) will Harm Patients and Academic Pathology

Tracking blood products in hospitals using radio frequency identification: Lessons from a pilot implementation

The Future of Drug Safety

Data Transparency and Quality Metrics

Molecular Diagnostics

State Control of Medicines and Medical Devices in Russian Federation

Digital Health A PRAGMATIC REGULATORY APPROACH

Impact of the Regulatory. Development and Innovation

Overview of the FDA Approval Process for TB Diagnostics

An Introduction to the Worldwide Regulatory Framework for Medical Devices. Elizabeth Malo M.S., Director, Regulatory Affairs

RE: FDA-2011-N-0090: Food and Drug Administration; Unique Device Identification System

Medical Devices; Gastroenterology-Urology Devices; Classification of the Endoscopic

The power of the Converge platform lies in the ability to share data across all aspects of risk management over a secure workspace.

FDA-Regulated Research

Implement an Effective Change Management System throughout GMP and Validation Environments. Eileen Cortes February 23, 2017

Draft Guidance for Industry Development and Use of Risk Minimization Action Plans

Medical Devices; Obstetrical and Gynecological Devices; Classification of the Closed Loop

Transcription:

AMIA Invitational Policy Meeting September 9-10, 2009 Software Regulation: The Transfusion Medicine Experience Rodeina Davis Vice President & CIO Milwaukee, WI

Founded 60 Years Ago, BloodCenter of Wisconsin seeks to advance patient care by delivering life-saving solutions grounded in unparalleled medical and scientific expertise. Blood Research Institute

Blood Services Experts from Bench to Bed site Diagnostic Lab Services Medical Services Blood Research Institute

t1 Relevant Background 20 years experience as Chief Information Officer at 3 large blood centers Member of IT Executive Council of International Society for Blood Transfusion (ISBT) Participated in creating and adoption of an international consensus standard for blood banking data definition and data structure Worked with regulatory agencies and industry to develop & improve IT use in transfusion medicine Chaired 2 task forces to develop international industry guidelines for security and RFID Member of Biovigilance Working Party to develop national surveillance system for transfusion medicine Developed systems with 510(k) premarket approval from FDA

Slide 4 t1 test, 9/8/2009

Transfusion Medicine Transfusion medicine is a segment of healthcare comprised of blood centers and hospital transfusion services. Blood centers collect, process, and distribute blood products to hospitals Transfusion services cross match, issue, and transfuse blood products to patients Both function under direction of transfusion medicine physicians.

Software as a Medical Device Software used in both blood centers and transfusion services is highly regulated The software is considered a Class II medical device requiring 510(k) premarket notification No other software used in healthcare or pharma requires 510(k)

Regulation of Medical Devices Is it a medical device? An instrument, apparatus, implement, machine, contrivance, implant, in vitro reagent, or other similar or related article including any component, part, or accessory which is intended for use in the diagnosis of disease or other conditions, or in the cure, mitigation, treatment, or prevention of disease * Is it in interstate commerce/commercial distribution? Is the data transmitted or accessed across state lines? *Section 201(h) of the FD&C Act

Medical Device Classification (by risk) Class I - General controls alone are sufficient to provide reasonable assurance of safety and effectiveness Class II General controls alone are insufficient to provide reasonable assurance of safety and effectiveness and there is sufficient information to establish special controls 510(k) applies Class III Insufficient information that general or special controls will provide reasonable assurance of safety and effectiveness, i.e., there is no predicate

Questions to Ask Why and how did our software become highly regulated? What are the benefits? What are the unintended consequences? Are the benefits of 510(k) worth the risk of unintended consequences? Are there alternative ways to ensure both the confidence of the regulators and the needs of our industry?

Why and how? Why and how did our software become highly regulated? Major recall & several major incidents in early 1990s involved poorly designed and unvalidated software In response, FDA issued memo ascribing responsibility to transfusion medicine establishments Our industry had little understanding of or investment in IT, depended on software vendors, and was unable to enforce vendor software quality Industry leadership requested that the FDA regulate blood banking software as a medical device March 31, 1994 FDA issued memo that software used in blood establishments is a medical device and will require 510(k) premarket notification

What are the benefits? Gives FDA authority to intervene to correct problems Moves system error responsibility from users to vendors Early identification of defects in software products Substandard software/vendors no longer in market Safety of software on the market improved (80% survey response from industry members July 2008)

Unintended Consequences Introduced market barriers Slowed down time to market for improvements and upgrades Inhibited introduction of new technology Created misconception of intent Increased costs for the vendor and industry Placed boundaries on vendors and products Made integration of systems difficult or almost impossible

The Worth Question Are the benefits of 510(k) worth the risk of unintended consequences? State of our industry has changed since regulation was requested 14 years ago Quality systems in place Clear understanding of verification & validation IT leadership is part of best practices Automation has improved but complicates what we do Business decisions and patient care rely on shared information from multiple systems

The Worth Question Are the benefits of 510(k) worth the risk of unintended consequences? This question was examined at a conference July 10-11, 2008 with FDA, software vendors, blood centers, hospital transfusion services, etc. Benefits in first 10 years of regulation outweighed unintended consequences Lack of integrated solutions in the last 4 years have promoted adoption of suboptimal alternative solutions that are: more error prone less cost-effective Our industry has not seen the benefits of widely adopted technologies in other industries in the last 4 years

Alternatives Are there alternative ways to ensure both the confidence of the regulators and the needs of our industry? Refine definitions of portions of software that must be regulated Streamline 510(k) submission process Standardize interfaces between systems so 510(k) amendment not required Establish discussion forums: FDA, vendors, and industry Reconsider classification of software as Medical Device Class I, not requiring 510(k)

Conclusion While our industry benefitted from regulation for a period of time, the same regulation is a deterrent to adoption of new technology and innovation At this point our industry would benefit more from regulation oversight without the 510(k) process: No adverse impact on safety and effectiveness Reduced risk of work-arounds Improved availability of software solutions and technology No impact on patient safety Be careful what you ask for; it is almost impossible to remain agile in a regulated environment

Links Deciding When to Submit a 510(k) for a Change to an Existing Device http://www.fda.gov/cdrh/ode/510kmod.html Premarket Notification (510(k) http://www.fda.gov/cdrh/devadvice/314.html 510(k) Blood Establishment Computer Software (list) http://www.fda.gov/cber/products/510ksoft.htm Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices http://www.fda.gov/cdrh/ode/guidance/337.html