URGENT FIELD SAFETY NOTICE. AQUIOS CL Flow Cytometry System (PN B30166)

Similar documents
LONDON HEALTH SCIENCES CENTRE (LHSC) WORKFLOW COMPARISON STUDY WITH THE AQUIOS CL FLOW CYTOMETER

New Client Training. Paychex Time and Labor Online Paychex, Inc. All rights reserved. TLO

Eclipse Palm RDC. Release (Eterm)

Electronic Invoicing

Requisition line View New Change Cancel... 60

Copyright Basware Corporation. All rights reserved.. Vendor Portal User Guide Basware P2P 17.3

Sage ERP Accpac Online 5.6

Application Information Bulletin: DuraClone IM Tubes Compensation setup for high content DuraClone reagents

FileBound AP Demo Overview

Lesson 7 Vendors and Expenses Part 1

FundTracker 21 Updates

LowesLink Spend Management (SM)

Completing the performance picture.

RUO Addendum. UniCel DxH Series with System Manager Software. Coulter Cellular Analysis System. B26673AC July 2015

Robinson Manufacturing Company, Inc. College Concepts, LLC. Vendor Shipment Management System (VSMS)

UPS WorldShip TM 2010

URGENT MEDICAL DEVICE RECALL. Alere San Diego, Inc. is initiating a voluntary recall of certain lots of the following products:

Nanyang Technological University(NTU)

UNITY CRM INTEGRATION AGILE CRM

Oracle Hospitality Suites Management User Guide. Release 3.7

Sage Accpac ERP Integration with SageCRM 6.1

Or, from the View and Manage RFx(s) page, click the Create RFx button. Create New RFx - Use this button to create an RFQ from scratch.

Oracle Warehouse Management Cloud. Getting Started Guide Release 8.0 Part No. E

THE UNIVERSITY O K L A H O M A

Microsoft Dynamics GP. Inventory Inflation Adjustment

Oracle Hospitality RES 3700 Enterprise Management. Installation Guide

IBM Maximo Mobile Suite Version 7 Release 5.1. User Guide

Cipher Lab 8200 Installation & Usage Guide

Invoice Manager Admin Guide Basware P2P 17.3

Refer to this guide for instructions about using Quicken s online account

AQUIOS CL LOAD & GO AND MAKE FLOW CYTOMETRY SIMPLICITY ITSELF

Version 4 Reference Guide. Mobile Inspections. January 2013

User s Manual. esims Web. Synthes Information Management Suite.

Configuring IBM Cognos Controller 8 to use Access Manager Authentication

TRAINING GUIDE. Lucity Mobile Warehouse

Viewing Test Material Reports and Tracking Information

Managing the Holding Tank: Resolve Inventory Problems

Business Portal for Microsoft Dynamics GP. Requisition Management Administrator s Guide Release 10.0

Chapter 20. Label Maker. FrontDesk v Label Maker Range Tab Select Tab

Payroll Process Guide

Sage HRMS 2014 Sage Employee Self Service Installation and Setup Guide. October 2013

Table of Contents. Instructions for Completing Activities Associated with My Performance Plan... 3 Create My Performance Plan... 3

Contents OVERVIEW... 3

Performing a Payroll Export to Payroll City

Table of contents. Reports...15 Printing reports Resources...30 Accessing help...30 Technical support numbers...31

SAFETYLINK USER GUIDE -EMPLOYEE TRAINING

Items and Warehouses - Solutions

SunRISE. Chemical Manager Training

PUBLIC. How to Work with the SBO Mailer Signature. All Countries. Solutions from SAP. SAP Business One 2007 B. June English

Oracle FLEXCUBE Core Banking

Sage 100. Sage Payroll Services Getting Started Guide

Navigating in ADP Workforce Now for Practitioners

MBS ARC (Barcodes) MANUAL

Proteus MMX Computerized Maintenance Management Software. Proteus MMX Mobile V User /Training Manual

MACH VBS. External Haulier User Guide. VBS Invoicing User Manual. Version 2.0 Issue Date: 1 st January 2016

FAQs CTI Integration with SAP Hybris Cloud for Customer SAP Hybris Cloud for Customer PUBLIC

ReCPro TM User Manual Version 1.15

Module Simple Barcode Inventory User Manual Version 1.0

TrackITSystem. Facility Manager Documentation Installation and User Guide

Enhanced Time & Attendance

POS Support Manual 01/10/2013

Microsoft Dynamics GP. Field Service Returns Management

User Manual NSD ERP SYSTEM Customers Relationship Management (CRM)

Oracle Hospitality Inventory Management Mobile Solutions. Installation and Configuration Guide

Variant Support in Microsoft Dynamics AX R3 - CU8

PCH Loyalty Reward Perks Reference Manual Finestra POS Version: 6.8.5

Lesson 12. Doing Payroll with QuickBooks

SUPPLEMENTS MANAGEMENT PROGRAM v2.1 Administrator User Guide. Table of contents

Deltek Touch for Maconomy. Touch 2.2 User Guide

Sage HRMS 2014 Sage Employee Self Service. Installation and Setup Guide. October has been retired. This version of the software

OMNI AR/Billing: Create Contract (Base + Variable/Rate) Detailed Business Process Guides ACONTRACTS1B

iphone/ipad User Guide

BlackBerry User Guide

Revision Control Date By Action Pages T Mistry Document created M Walker Modified 31

Published by ICON Time Systems A subsidiary of EPM Digital Systems, Inc. Portland, Oregon All rights reserved 1-1

PeopleSoft Financials epro Process Guide

Business Portal for Microsoft Dynamics GP. Human Resources Management Self Service Suite Administrator s Guide Release 10.0

Microsoft Dynamics GP Business Portal. Project Time and Expense User s Guide Release 3.0

Version No: 3.0, Build No: 3. Portal Release Notes V3B3

1. Click the IT HelpDesk icon on your desktop or open Internet Explorer and point your browser to

SunRISE. Chemical Manager Training

Bed Management Solution (BMS)

Asset Management for Inventory Scanners. Core-CT 9.2 Finance Implementation

Deltek Touch Time & Expense for Vision. User Guide

After logging into PrintGizmo create your profile and complete the information requested by navigating the menu to Configuration > Profile.

Release Notes Version 6.4

3. Click Exception List on the bottom of the Generate Resupply List screen.

Contents OVERVIEW... 3

ADP Vantage HCM: Transfer an Employee s Time to a Different Labor Account

Aldelo Liquor Controller Server User Manual

Oracle. SCM Cloud Using Receiving. Release 13 (update 17D)

iphone/ipad User Guide

Version /2/2017. Offline User Guide

Procurement Card Exercise Workbook. PeopleSoft Procurement Card

Oracle. SCM Cloud Getting Started with Your Manufacturing and Supply Chain Materials Management Implementation. Release 13 (update 18B)

Aldelo Gift Card Server User Manual

FedEx Ship Manager. Software. Version Quick Guide

Asset Management for Asset Processors. Core-CT 9.2 Finance Implementation

RAMCO AVIATION SOLUTION VERSION 5.8 USER GUIDE SMART OPERATIONS

Product Maintenance. Product Codes

Transcription:

URGENT FIELD SAFETY NOTICE September 28, 2017 Attention Beckman Coulter (BEC) Customer, AQUIOS CL Flow Cytometry System (PN B30166) Beckman Coulter is initiating a field action for the product listed above. This letter contains important information that needs your immediate attention. ISSUE: IMPACT: ACTION: When AQUIOS CL systems are connected to a Laboratory Information System (LIS), the system may duplicate sample requests leading to sample misidentification. For this to happen, the following conditions must be met: The AQUIOS CL Flow Cytometer is connected to an LIS and The AQUIOS CL system has the DEFAULT TEST enabled and Host Query is enabled in LIS setup screen on system and There are multiple cassettes in the Autoloader and The LIS host query response is received in a brief time window before LIS response timeout occurs. When all of these conditions are met, a sample will follow two (2) paths of test request creation (LIS and DEFAULT TEST). All software versions (2.0, 2.0.1 and 2.1) in existence are impacted by this issue and all applications used on the AQUIOS CL are impacted by this issue. When this issue occurs: It will result in sample mis-identification and erroneous results will be generated. The system will not always identify or flag the erroneous results however, some erroneous results may be flagged with Sample ID reused. The frequency will not be predictable from the limited data set that BEC has analyzed, this random mis-identification occurred 24 times out of 92,000 runs. Once the duplicate test is created, the following events will occur: The software will continue to create multiple tests as long as there are tubes available (in the cassette in the Autoloader). The system may associate an incorrect sample ID with the run data. Immediately perform an adjustment to the software options on your AQUIOS CL System by turning off the DEFAULT TEST. This will eliminate the issue and result in one sample producing one set of results (see Attachment 1). After doing this, samples with no LIS test requests will become entries in the incomplete tab (see Attachment 2 for instructions on how to run samples after the DEFAULT TEST is turned OFF). BEC will contact your laboratory in the coming weeks and assist in determining if any samples previously run in your laboratory were impacted. RESOLUTION: Beckman Coulter is working on a permanent resolution to correct and eliminate this issue. P.O. Box 169015 Internet: www.beckmancoulter.com

Page 2 of 5 Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: Electronically, if you received this communication via email. Manually, complete and return the enclosed Response Form. If you have any questions regarding this notification, please contact: From our website: http://www.beckmancoulter.com By phone: call 800-369-0333 in the United States and Canada. By email: LScustomerLetter@Beckman.com Outside the United States and Canada, contact your local Beckman Coulter representative. We apologize for the inconvenience that this has caused your laboratory. Sincerely, Marwan Fathallah Vice President, Quality Assurance and Regulatory Affairs Enclosure: Response Form Beckman Coulter, the stylized logo, and the Beckman Coulter product and service marks mentioned herein are trademarks or registered trademarks of Beckman Coulter, Inc., in the United States and other countries P.O. Box 169015 Internet: www.beckmancoulter.com

Page 3 of 5 Attachment 1 Instructions for Disabling the Default Test Feature Please refer to the AQUIOS CL Instructions for Use, PN B21896, Chapter 8 Setup for additional information. The Admin User is the user type which must perform this action. 1. Select from the right side of the screen, the software will open in the area. 2. Toggle the Default Test option to off. 3. Select Save on this screen prior to moving to any other areas of the software. When the Default test is disabled, there are 4 potential ways in which the system can recognize test requests for samples, utilizing the internal barcode scanning process: 1. Tube barcode is scanned 1 st time and there is an order in LIS AQUIOS software creates the test request. 2. Tube barcode is scanned 2 nd time and LIS is configured to return order again AQUIOS software creates a test request. 3. Tube is scanned and there is no order in LIS the sample information is placed in the Incomplete tab, and a manual test request needs to be created. Follow the instructions indicated below for manual test creation. 4. Tube is scanned 2 nd time and LIS not configured to return order again - the sample information is placed in the Incomplete tab, and a manual test request needs to be created. Follow the instructions indicated below for manual test creation. P.O. Box 169015 Internet: www.beckmancoulter.com

Page 4 of 5 Attachment 2 Instructions for Manually Creating Test Requests on the AQUIOS CL Please refer to the AQUIOS CL Instructions for Use, PN B21896, Chapter 5 Sample Processing Add Test Request for Patient Sample for additional information. The test request can be created before the tubes are ready to be loaded into the system. Once the test request is created, the request appears on the Request Details List. 1. Select from the Main Screen. 2. The following screen appears. 3. Select the from the top right side of the screen. 4. The Add Request screen is obtained. P.O. Box 169015 Internet: www.beckmancoulter.com

Page 5 of 5 5. Ensure that the Patient radial is selected. 6. Enter the Sample ID using one of the following methods: If the sample has a barcode, select the icon to turn on the external barcode reader, and scan the barcode. The barcode information will appear in the Sample ID field of the Add Request screen Manually type the Sample ID into the field. NOTE Only samples with a barcode can be run via the autoloader. 7. Use the Test dropdown menu to select the desired test. 8. Enter any additional information to the Add Request screen, and select OK. 9. The sample can now be loaded into a cassette and placed on the autoloader, if a barcode is present. Alternatively, it can also be run via the single tube loader. 10. Refer to the AQUIOS IFU, PN B21896, Chapter 5 Sample Processing for additional information on the Sample Processing options available on the AQUIOS CL System. P.O. Box 169015 Internet: www.beckmancoulter.com