Specimen configuration

Similar documents
MODEL PicoMill TEM specimen preparation system. Achieve ultimate specimen quality free from amorphous and implanted layers

MODEL NanoMill TEM Specimen Preparation System. Ultra-low-energy, inert-gas ion source. Concentrated ion beam with scanning capabilities

MODEL 1051 TEM Mill ION MILLING. Ion milling is used on physical science. specimens to reduce thickness to electron

MODEL TEM Mill. Two independently adjustable TrueFocus ion sources

Model TEM Mill. Tabletop precision preparation for producing high-quality TEM specimens from a wide variety of materials EXCELLENCE MAGNIFIED

MODEL Plasma Cleaner. Simultaneously cleans specimen and specimen holder. Cleans highly contaminated specimens in 2 minutes or less

MODEL SEM Mill. Two independently adjustable TrueFocus ion sources

METHOD FOR IMPROVING FIB PREPARED TEM SAMPLES BY VERY LOW ENERGY Ar + ION MILL POLISHING

MODEL 1061 SEM Mill ION MILLING. Ion milling is used in the physical. sciences to enhance the sample s surface. characteristics. Inert gas, typically

Specimen Preparation Technique for a Microstructure Analysis Using the Focused Ion Beam Process

Transzmissziós és pásztázó elektronmikroszkópos minták készítése ionsugaras vékonyítással

Sample Repositioning Solutions for

Silver Diffusion Bonding and Layer Transfer of Lithium Niobate to Silicon

Cross sectional TEM analysis of duplex HIPIMS and DC magnetron sputtered Mo and W doped carbon coatings

Application Note. Sample Protection prior to FIB Processing. related instruments: Leica EM ACE600. Material Research. Life Science Research

Cross sectional TEM analysis of duplex HIPIMS and DC magnetron sputtered Mo and W doped carbon coatings

Microstructural Characterization of Materials

SUPPLEMENTARY INFORMATION

Model Plasma Cleaner. Effectively cleans specimens for electron microscopy EXCELLENCE MAGNIFIED

A New Development to Eliminate Artifacts during TEM Sample Preparation in the FIB

Transmission Electron Microscopy (TEM) Prof.Dr.Figen KAYA

Focused Ion Beam CENTRE INTERDISCIPLINAIRE DE MICROSCOPIE ELECTRONIQUE. Marco Cantoni, EPFL-CIME CIME ASSEMBLEE GENERALE 2007

Microscopy AND Microanalysis MICROSCOPY SOCIETY OF AMERICA 2007

Crystallographic Orientation Relationship between Discontinuous Precipitates and Matrix in Commercial AZ91 Mg Alloy

The principles and practice of electron microscopy

Electronic Supplementary Information

Kinematical theory of contrast

Supporting information. In-situ TEM observation of phase transition of nanoscopic patterns on. baroplastic block copolymer film during nanoindentation

STUDY & ANALYSIS OF ALUMINIUM FOIL AND ANATASE TITANIUM OXIDE (TiO2) USING TRANSMISSION ELECTRON MICROSCOPY

3D Nano-analysis Technology for Preparing and Observing Highly Integrated and Scaled-down Devices in QTAT

Supplementary Materials for

Supporting Information

Supplementary Information

IBS/e Ion Beam Sputter Deposition and Etching System. IBS/e with KDC-10 Ion Beam Sputter Deposition and Etching System with Kaufman Ion Source

Carnegie Mellon MRSEC

In Situ Observation of Dislocation Nucleation and Escape in a Submicron Al Single Crystal

Scanning thermal microscopy probe capable of simultaneous electrical imaging and the addition of a diamond tip

Scanning Electron Microscope & Surface Analysis. Wageningen EM Centre Marcel Giesbers

zyvex TEM Sample Lift-out Using the Zyvex Nanoprober System By Kimberly Tuck, Zyvex Corporation

Applications of Energy-Assistance to the formation of novel surface coatings

Supplementary Information. for

Practical 2P8 Transmission Electron Microscopy

Practical 2P8 Transmission Electron Microscopy

Evaluation of Top, Angle, and Side Cleaned FIB Samples for TEM Analysis

Surface Layer Characterization of Atomized Magnesium for use in Powder Metallurgy Products Paul Burke and Georges J. Kipouros

Crystallographic Textures Measurement

Supplementary Material

Dislocations Linear Defects

FePd (216 Å) grown on (001) MgO. 2θ(deg)

Microscopy and imaging science: practical approaches to applied research and education (A. Méndez-Vilas, Ed.)

AFRL-RX-WP-TP

Various Techniques for Reliability Estimation and Failure Analysis of Electronic Products and Components

Electron microscopy II

SUPPLEMENTARY INFORMATION

Sr and Pb additions. L. Affleck, C. Leach *

Transmission Kikuchi Diffraction in the Scanning Electron Microscope

In situ TEM Characterization of Shear Stress-Induced Interlayer. Sliding in the Cross Section View of Molybdenum Disulfide

Facet-Selective Epitaxy of Compound Semiconductors on

AP 5301/8301 LABORATORY MANUAL

Formation of High-quality Aluminum Oxide under Ion Beam Irradiation

XTEM characterization of modulated ion implantation through self-organized anodic aluminium oxide (AAO) membranes

AIP ADVANCES 6, (2016)

INA-X System for SNMS and SIMS

Defect and chemical analysis in the TEM

TED PELLA, INC. Microscopy Products for Science and Industry

Observation in the GB (Gentle Beam) Capabilities

Simple method for formation of nanometer scale holes in membranes. E. O. Lawrence Berkeley National Laboratory, Berkeley, CA 94720

Microstructure Analysis by Means of the Orthogonallyarranged

Supporting Information

Practical Technique for Restraining Differential Charging in X-ray Photoelectron Spectroscopy: Os Coating


Laser Micromachining for Industrial Applications and R&D. 3D-Micromac AG. Symposium on Smart Integrated Systems in Chemnitz. 3D-Micromac AG

General Introduction to Microstructure Technology p. 1 What is Microstructure Technology? p. 1 From Microstructure Technology to Microsystems

Supporting Information for Surface-energy induced formation of single crystalline. bismuth nanowires over vanadium thin film at room temperature

Microstructural study of titanium carbide coating on cemented carbide

Q150R Series Rotary Pumped Coaters

Ion channeling effects on the focused ion beam milling of Cu

From Eye to Insight. Leica EM ACE. Coater Family

Supplementary Figure 1: Geometry of the in situ tensile substrate. The dotted rectangle indicates the location where the TEM sample was placed.

Nanoscale Imaging, Material Removal and Deposition for Fabrication of Cutting-edge Semiconductor Devices

Growth Of TiO 2 Films By RF Magnetron Sputtering Studies On The Structural And Optical Properties

TEM and Electron Diffraction Keith Leonard, PhD (1999) U. Cincinnati

In-situ Electron Microscopy Mechanical Testing for Steels

Influence of Alloy Microstructure on Oxide Growth in HCM12A in Supercritical Water

SPECIMEN PREPARATION FOR ELECTRON MICROSCOPY

Supplementary. N. Akhtar a,b, M.Y. Emran a, M. A. Shenashen a,, T. Osaka b, A. Faheem c, T. Homma b, H. Kawarada. , S. A.

ANALYTICAL SERVICES X-RAY SCATTERING ATOM PROBE TOMOGRAPHY RAMAN SPECTROSCOPY TIME-OF- FLIGHT SECONDARY ION MASS SPECTROMETRY FOCUSED ION BEAM (FIB)

Ultramicroscopy 111 (2011) Contents lists available at ScienceDirect. Ultramicroscopy. journal homepage:

Electrical Characterization of Tungsten Nanowires Deposited by Focused Ion Beam (FIB) *

Electron Microscopy Studies of Niobium Thin Films on Copper

PHASE SEPARATION BY INTERNAL OXIDATION AND REDUCTION IN A Cu-5at%Ni-ALLOY

Growth of monocrystalline In 2 O 3 nanowires by seed orientation dependent vapour-solid-solid mechanism

Previous Lecture. Vacuum & Plasma systems for. Dry etching

MATERIALS. Sample Preparation & Grids for SEM & TEM... L 17 L 01. MATERIALS / Inorganics & thin films guide

Evidence of nanopores in sol gel based TiO 2 and TiN ultrafiltration membranes

Microstructure of as-coated thermal barrier coatings with varying lifetimes

Supplementary Figure S1. High-resolution XPS spectra in the Cu 2p region and Cu LMM spectra are shown in (A) and (B) respectively for CdSe NCs

JSM-7800F Field Emission Scanning Electron Microscope

EUV Transmission Lens Design and Manufacturing Method

ELECTRON MICROSCOPY MODERN SURFACE, LAYER AND DEFECT ANALYSIS USING REM, FIB, EDX, STEM

Transcription:

APPLICATIONNOTE Model 1040 NanoMill TEM specimen preparation system Specimen configuration Preparing focused ion beam (FIB) milled specimens for submission to Fischione Instruments. The Model 1040 NanoMill TEM specimen preparation system is ideal for specimen processing following FIB milling. The NanoMill system s concentrated argon ion beam, typically in the energy range of 50 to 2000 ev, excels at targeted milling and specimen surface damage removal. Ioninduced secondary electron imaging is used to locate the FIB-produced lamella and then to target the region that will receive low energy milling in either raster or spot modes. Only the lamella is targeted during ion milling. Ion bombardment of the supporting grid is eliminated, which helps prevent redeposition. The NanoMill system can also remove damage from electropolished or broadbeam ion milled specimens. This document explains how to configure specimens to obtain optimal results from the NanoMilling SM process. Before you start, it is important to understand the following terms: Specimen geometry. Refers to the type of specimen, i.e., FIB in situ lift out, H bar, dimpled, wedge polished, etc. Specimen configuration. Refers to the position in which the lamella is mounted onto a support grid. Preparing in situ specimens For FIB-produced specimens, the preferred configuration is an in situ lift-out lamella with a thickness that is dictated by the material type. If the material: is in a bulk state or has a simple interface, the optimal specimen thickness is 50 to 80 nm. has an existing microstructure that is a 2- or 3-dimensional microelectronic pattern, (e.g., integrated circuits and FinFETs), the optimal specimen thickness is 50 nm or less. Figure 1. Light optical microscope image of the lamella attached onto the support grid (left) and an enlarged schematic of the lamella (right). Figure 2. Specimen dimensions. 1

Figure 3. The secondary electron image shows the support grid and the lamella. Attach the lamella to the side of the central finger (B pillar) of an Omniprobe copper* grid (Figure 1). The three-position ( ABC ) grid is preferred because the mounted lamella will be near the eucentric position in both the NanoMill system and the transmission electron microscope (TEM). The minimum lamella size is 10 µm long x 5 µm wide (Figure 2) as determined by the minimum diameter of the argon ion beam (~1 µm). Larger specimens are preferred for imaging and targeting. It is very important that the shorter side of the lamella be attached: To the side and very close to (but not on) the top of the grid s B pillar (Figure 1). This positioning allows for a low-angle tilt during milling. Tilting a grid as it is imaged in a FIB or scanning electron microscope (SEM) can establish the attachment point required for reaching a desired milling angle. The typical range of milling angles is 7 to 10 (the angles are measured with respect to the plane of the grid), both in the negative and positive directions of tilt. Attaching the lamella in this position also protects the lamella during shipping and handling. Coplanar with the grid surface so that it may be tilted symmetrically. On both top and bottom lamella surfaces at the attached end to create a rigid and firm bond. With a metal such as platinum, rather than carbon, because the lamella is subsequently plasma cleaned. A cap layer is needed if the lamella has any of the following characteristics: is organic 2

Figure 4. Specimen configuration. has a coating is multilayer is a low atomic number material The preferred cap layer material is platinum. Carbon is not recommended, because material may be lost during initial plasma cleaning. Do not use tungsten as a cap layer either; its lower sputtering rate results in longer NanoMill system cycle times. It is important to use the FIB in situ lift-out geometry and to attach the lamella to the top left- or right-side of the B pillar because other specimen configurations may result in sputtering and subsequent redeposition of grid material onto the specimen surface. The described specimen configuration ensures that the argon ion beam bombards only the lamella, even during low-angle milling. Keep the following guidelines in mind when FIB processing the lamella: Do not attach the lamella in the V notch above the A or C pillars in an Omniprobe grid because the short distance between the top edge of the lamella and the grid may result in redeposition of the grid material onto the specimen surface. Do not attach more than one lamella onto a grid, even if they are placed on opposites sides of the B pillar. Redeposition of material from one lamella to another may occur. Ensure that the protective platinum cap on the top face of the lamella is 0.3 to 0.5 µm in thickness after FIB. Electron beam deposition is recommended because the small grains will not cause shadowing of the ion beam in the NanoMill system. This protective layer is sacrificial and is most often sputtered away during the NanoMilling process. The deposition of gold prior to platinum should be avoided because it may be sputtered preferentially due to its very low sputtering threshold (~ 14 ev). Ensure that the lamella is polished in the FIB at 5 kv or lower, it is as free from curtaining, and it is as parallel-sided as possible. No trimming artifacts should be present at the bottom edge of the lamella (opposite the cap). Trimming artifacts reduce the ability to mill from either the top or bottom edge of the lamella (see Figure 4). Preparing H-bar specimens The NanoMill system can enhance FIB-milled H-bar specimens. Prevention of sputtering and subsequent redeposition of the specimen substrate onto the lamella depends on the proper construction of the FIB box. A single FIB box at the center of an H bar is preferred (minimum Figure 5. An unsupported (left) and a grid-mounted (right) H bar. 3

The grid should be electrically conductive. The adhesive for the H-bar specimen should be electrically conductive to ensure that the grid and the specimen material have good electrical contact. Leave sufficient area surrounding the specimen so that the grid may be clamped securely. Figure 6. Model 1050 TEM Mill specimen holder with a grid bridging the gap. dimensions of 20 x 20 µm). Ensure that the lamella receives ion beam polishing, it is free from curtaining, and is as parallel-sided as possible. The optimal final lamella thickness is dictated by the material type. If the material: is in a bulk state or has a simple interface, the optimal specimen thickness is 50 to 80 nm. has an existing microstructure that is a 2- or 3-dimensional microelectronic pattern, (e.g., integrated circuits and FinFETs), the optimal specimen thickness is 50 nm or less. The H bar can be either free standing or attached to a ½ grid with a minimal amount of conductive adhesive. The edge of the H bar containing the FIB box should be positioned near the center line of the 3 mm diameter support grid (Figure 5). The overall dimensions of the specimen geometry must allow specimen loading into a side-entry TEM holder. Keep the following guidelines in mind when FIB processing the grid-mounted H bar: Submitting specimens to Fischione When submitting a specimen for preparation with the NanoMill system: Provide images (TEM/SEM micrographs and/ or sketches) to show the areas from which you would like to see FIB-induced damage removed or thinning enhanced. Indicate the microscopy that you intend to use to analyze the specimen (e.g., TEM, SEM, high-angle annular dark-field [HAADF], high-resolution electron microscopy [HREM], electron energy loss spectroscopy [EELS], energy-filtered transmission electron microscopy [EFTEM], energy-dispersive X-ray spectroscopy [EDS], etc.). Indicate the initial thickness and desired final thickness (following the NanoMilling process) of the lamella. The specimen geometry should span the 2 mm gap in the specimen holder (see Figure 6) and fit within the 3 mm diameter of the holder. Ship specimens in containers that are standard practice for FIB users and follow standard shipping procedures. Standard membrane boxes used for storage of conventional TEM specimens are suitable; do not insert filter or lens paper into the membrane box. Contact sales@fischione.com or applications@fischione.com with questions before FIB (or submission) of samples. 4

Specimen preparation summary When configuring FIB-processed specimens for the NanoMill system, follow the suggested practices: Parameter Specimen geometry Lamella dimensions Lamella position Attachment Cap layer FIB Suggested practice In situ lift-out or H bar The optimal final lamella thickness is dictated by the material type. If the material: is in a bulk state or has a simple interface, the optimal specimen thickness is 50 to 80 nm. has an existing microstructure that is a 2- or 3-dimensional microelectronic pattern, (e.g., integrated circuits and FinFETs), the optimal specimen thickness is 50 nm or less. Side of B pillar (finger) of a copper Omniprobe grid Platinum, top and bottom lamella surfaces, to the grid ledge Platinum, electron beam deposited, 0.3 to 0.5 μm thick after FIB; area of interest no more than 1 µm away from cap On all surfaces and edges of lamella: 30 kv maximum for cutting 5 kv or less for final polishing Note *Use a copper, not molybdenum, Omniprobe grid. Molybdenum grids are too short for these purposes. E.A. Fischione Instruments, Inc. 9003 Corporate Circle Export, PA 15632 USA Tel: +1 724.325.5444 Fax: +1 724.325.5443 info@fischione.com www.fischione.com 2014 E.A. Fischione Instruments, Inc. All rights reserved. The Model 1040 NanoMill TEM specimen preparation system is the subject of United States Patent Nos. 7,132,673 and 7,504,623. NanoMill is a registered trademark of E.A. Fischione Instruments, Inc. NanoMilling is a service mark of E.A. Fischione Instruments, Inc. All other trademarks are the property of their respective owners. Document Number AN006 Revision 03 01/2014