PolyLab System. Torque Rheometer: RheoDrive

Similar documents
Ultimate Flexibility enhance your material development. Thermo Scientific HAAKE PolyLab Modular torque rheometer platform

The influence of carbon black types on the processability of Rubber Compounds in green tires

powder coating pharmaceuticals. The parallel, segmented compounder

Laboratory-Mixers & Extruders

Brabender Measuring Mixers. for Material Research and Quality Control. where quality is measured.

MiniLab-Compounder and Reactor

Thermo Scientific Twin-Screw Extruders Comprehensive Compounding and Extrusion Solutions for Materials Science

Regards from the Team at Rheology Solutions.

zahnrad_neu_gb :46 Uhr Seite 1 Gear Pumps

Thermo Scientific Twin-Screw Extruders. A Better Mix of Science and Technology

Unit 156: Polymer Manufacturing Processes

Industrial aspects of polymer processing

1/31/ Plastics Processing. Thermoplastics (Tampere)

LCM CC & TCC. A member of the Possehl Group

Gelimat Technology. Ultrahigh-Speed Thermokinetic Mixing, Compounding & Fluxing. Processing Excellence in Motion DUSATEC DUSATEC, INC.

Options. The following is a list of our most popular options. If you don t find what you need, please call us. HARDWARE

High Speed Container Mixers

AGC Chemicals Americas, Inc. Your Global Source for Fluorochemicals. Technical Capabilities Overview

PVC. PVC / Suspension. PVC / Emulsion

Polymers and plastics

Mixing conditions of polymer and ceramic powder determined by ultrasound

Compounding and Processing Additives for Woodfiber-Plastic Composites

Extending the Processing Window of Ca/Zn-stabilised PVC Window Profiles by the Use of Metallocene Waxes

Silicone masterbatch additives for improved plastics processing

AN INTRODUCTION TO SPHERONIZATION (MARUMERIZATION * )

New generation of aluminium hydroxide flame retardant filler for the wire and cable industry

Ingtec Technik AG Company Presentation EnMat EN Feb 2016

greenprofile wpc Wood plastic composite (WPC) profile extrusion

BASICS ELASTOMER TESTING FOR MORE THAN 45 YEARS! Elastomer testing VISCOSITY, RHEO-KINETICS, CAPILLARY RHEOMETRY, VULCAMETRY

Rubber Low Temperature Brittleness Tester TRBT-A10

Title Freeze-Drying cycle monitoring Problem Statement Approach Result Impact

LCR 7000 Series Capillary Rheometers. Single Bore LCR 7000 Dual Bore LCR 7002

In the world market for 25 years...

EFFECTS OF PROCESSING METHOD AND FIBRE CHARACTERISTICS ON MICROSTRUCTURE AND PROPERTIES OF WOOD-PLASTIC COMPOSITES

Rubber Testing Solutions

MELT FLOW INDEXER. Robotic. Automatic. Manual

EXTRUSION 15/03/17. Extrusion. Polymer Types. Extrusion. The extruder is the most important piece of machinery in the polymer processing industry.

CO-ROTATING TSE: FLEXIBILITY IN PLASTIC RECYCLING

Processing Guide CONTENTS

Co-Rotating Twin-Screw Extruder Herausgegeben von Klemens Kohlgrüber

BLOW MOULDING. Blow moulding is a process used to produce hollow objects from thermoplastic.

Dynisco ViscoIndicator Online Rheometer. Simplified Rheology for the Masses

Melt Compounding of Rigid PVC Formulations with Hydrotalcites

EXPLORING POWDER INJECTION MOLDING OF NIOBIUM

Andreas Limper. Mixing of. Rubber Compounds HANSER. Hanser Publishers, Munich. Hanser Publications, Cincinnati

Improved Processing of Thermoplastics and Thermoplastic Elastomers

TECHNICAL SPECIFICATION

Rubber Technology and Turnkey Projects

Additive and Colour Preparations. for Extruded Polystyrene Foams

Advances in Twin Screw Compounding Technology

D-LFT PROCESS: IN-LINE COMPOUNDING & COMPRESSION MOLDING OF LONG-GLASS-FIBER REINFORCED POLYMERS

Role of on-line gauging systems in providing insights into the Web Coating process

ATYPICAL TWIN SCREW EXTRUSION COMPOUNDING SYSTEMS FOR POLYOLEFINS

Environmental Impact of Using Suprmix Versus Liquid

! "! "" # 1

PARALOID K-130D Improved Lower-Dusting Processing Aid for Vinyl

Effect of Modifying Agent (ACR) on Cell Structure and Mechanical Properties of Extrusion-Foamed PVC Sheet

CEAST SmartRHEO Series. Capillary Rheometer Systems

FLEXPACKCON SPE Flexible Packaging Division. Strategies to Avoid Interfacial Instability in Multilayer Films October 31, 2018 Phoenix, Arizona

CUSTOM FORMULATED TIRE-TOLL COMPOUNDS

Course Description EMME 1

A Review of PVC Applications using 4 Flight Kneader Systems

Measuring and understanding. of powders with the FT4 Powder Rheometer. freemantechnology

Micophotometric Control of Particles and Inhomogeneities in Flowing Polymer Melts during Extrusion Processing

Gear Pumps. Precise, Pulseless, Repeatable Performance. In Polymer Applications.

Table of Content. Page No

Developments in Small Scale Twin Screw Extruders for Color Development. Charlie Martin, Leistritz Extruson. Abstract. Background

SolutionProvider. Donglin Chem

Continuous Processing. MODCOS for Continuous Oral Solid Dosage Production Textmasterformat in Mastervorlage eingeben

THE EXTUSION FOAMING OF TPEV/CLAY NANOCOMPOSITE

By Dr. Robert Marsh, Bohlin Instruments, 2003

DIN Viscometer ViscoTester 550

Underwater pelletizing systems. Page 1

Melt Pumps and Screenchangers for Blown Film Extrusion

IMPACT MODIFIERS AND THEIR INFLUENCE IN TOUGHNESS AND PLATE OUT IN HIGHLY FILLED RIGID PVC FORMULATIONS

Chemical Foaming Alternatives to Azodicarbonamide

An Introduction to Single Screw Extrusion

Composite Acrylic Impact Modifier: The Latest Innovation in PVC Window Profiles Impact Strengthening

STEER GENERATION NEXT CO-ROTATING TWIN-SCREW EXTRUDERS

DA910 for Masterbatchers. June 2012

Many processors around the globe are improving their productivity by using Supernova for colour and material changeovers.

Highly efficient flue gas analyser with Bluetooth

TERPOLYMER ETHYLENE VINYL ACETATE MALEIC ANHYDRIDE

Improved Processing and Batch Time Reduction Through Powder Liquid Dispersions

Precision Honing Machines

Dynamar TM Polymer Processing Additives (PPA s)

James L. White / Helmut Potente (Editors) Screw Extrusion. Science and Technology

Highly efficient flue gas analyzer

IN-line. thermoforming system. More than just fast: Reliable

Pressure Measurement 1

IN-LINE THERMOFORMING MACHINES. Complete turn key IN-LINE installations. from material handling through to boxing units of finished products

IN-LINE THERMOFORMING MACHINES. Complete turn key IN-LINE installations. from material handling through to boxing units of finished products

Surfactant Influence on Silica Filled Tread

COURSE TITLE : POLYMER PROCESSING METHODS & MACHINERY COURSE CODE : 3091 COURSE CATEGORY : B PERIODS/WEEK : 6 PERIODS/SEMESTER : 108 CREDITS : 5

Product Information. Aflow extrusion plastometer. Optimum test sequence - time-saving and operator-friendly

Calcium Carbonate in Blown HDPE Film

Improving the abrasion resistance of green tyre compounds

MECHANOCHEMICAL RECYCLING AND PROCESSING OF WASTE CROSSLINKED POLYMERS: WASTE TIRE RUBBER AND WASTE XLPE FROM CABLE SCRAPS

APP Seminar China

Non-Radiometric Measurement. Innovative concept of on-line measurement

Transcription:

PolyLab System Torque Rheometer: RheoDrive 32

Rubber compound formulation Sulphur 2% Rubber 55% Activators 1% Oil 5% Accelerators 1% Carbon black 35% Antioxidants 1% 33

Torque Rheometer System Why use Torque Rheometers? Polymers are processed in a liquid state : (Extrusion, injection moulding) The flowabillity (viscosity) will influence : Processabillity Quality of the final product Torque Rheometer and its' sensors are: Down scaled production machines (mixers, extruders, dies, postex) Test procedure similar to production 34

Torque Rheometer System Who uses Torque Rheometers? Application: Material In R+D: for new polymers small amount of polymers In R+D: for new compound developments polymer alloys In R+D for new additives stabilisers, pigments, fillers, lubricants, etc. In pilot plants to study formulation and processing characteristics Application: Process In R+D for new machine design to evaluate production process In pilot plants to determine optimum process conditions In Q.C to control incoming and outgoing materials ISO 9000 35

BASE UNITS faster, higher torque RheoDrive 36

RheoDrive + Mixer/Extruder Sensor = PolyLab OS RheoDrive Mixer sensors Single screw extruder sensors Twin screw extruder sensor additional analyzing sensors Rotors Feeding systems Screws Dies Postex 37

PolyLab OS System Components HAAKE RheoDrive 4 & 16 kw Drive Unit HAAKE PolySoft Control & Application Software HAAKE Rheomix Mixer HAAKE Rheomex Extruder HAAKE PTW 16/24 Parallel Twin Screw Extruder HAAKE CTW Conical Twin Screw Extruder HAAKE Dies HAAKE Postex HAAKE Ancillaries HAAKE Analyzing Sensors (viscosity, visco-elasticity, spectroscopy) 38

PolyLab OS torque speed energy volume weight temperature frictional heat RheoDrive Feeder Mixer-Extruder Die Postex Grinder dyn.viscosity MFI / MVI conductivity % additive dispersity Xtens viscosity dispersity film quality film strength 39

PolyLab OS New Design 40

PolyLab OS New Design 41

PolyLab OS New Design 42

PolyLab OS - Benefits OS = Open Solutions - integration of new sensors and systems with CAN bus (viscosity, spectroscopy, optical, ) - open solutions for all the applications (expand system with additional sensors) - open for the future (standardized interfaces, serviceability) - new sensors for combined measurements or methods (viscosity, optic, analytical tests) - needs only one instrument for polymer testing plug and measure - system with user compatibility of components, reliability, ease of use - connect all our offerings at site, no factory installation required - connect to PC, network or PDA with TCP/IP (cable-wireless) or USB 43

PolyLab OS - Benefits time saving software - software suite for running the complete PolyLab OS platform - time saving software with job stream option (automation of tests, ease of use, unsupervised running) serviceability - modular in design for service - remote diagnostics, exchange of components, minimum downtime - remote supervision from the office with browser/pda monitoring alarms/events/data traceability of experimental results - traceability of experimental results and FDA compliance (21CFR11) - prepared for FDA regulations in design and materials used 44

New Features with PolyLab Hardware System: Modular system for mixer, single screw extruder, or twin screw extruder to be connected to a single drive unit. Open System to connect and integrate different components RheoDrive: New drive concept with intelligent 4 kw & 16 kw motors. Very smooth and silent running drive system specifically selected for improved mixer test results. Docking station: New mechanical mechanism to look the docked systems in place; easy to use and safe. Only one central socket/connector to eliminate cables and wires hanging all over. Rigid and stable chassis on rolls to travel to different locations in a pilot plant. 45

New Features with PolyLab Hardware CAN Open Bus: Intelligent CAN bus integrates all peripherals with self-identification and transfer of characteristic data like calibration data table. Adaptation of all standard CAN open components (thermocouple, pressure sensor, additional sensors) without software modification. Integration of downstream equipment with standardized interface. Connectivity: Simple plug & play computer connection with USB or TCP/IP (option) Remote controller (cable based) to operate the drive manually Remote diagnose and status info with Web browser or PDA 46

PolyLab OS Central Coupling - Central coupling for docking Mixers and Extruders - Connecting the communication and the power supply 47

PolyLab OS Remote Control Control PolyLab OS where it is suitable for you Optimise the process on site Never loose visual control on your process 48

PolyLab OS Torque Measurement 3 torque measurement options: 1) Via motor power 2) Standard torque sensor 3) Multi range torque sensor Torque Measurement Accuracy Resolution no sensor (motor power) +/- 5 % (FSD) 1 Nm CANopen torque sensor +/- 0,15 % (FSD) 0.01 Nm 49

Torque Sensor The PolyLab torque sensor in-between motor shaft and coupling 50

Multi Range Torque Sensor 400Nm / 120Nm high accuracy: 0.15 % FSD high reproducibility: 0.02 % Strain gauge principle Dynamical Resolution Enhancement 60 Nm : 0.01 Nm > 60 Nm : 0.1 Nm This is two torque Sensors Part No.: 567-0101 The PolyLab Torque Sensors has non contact (no brushes, no collector ring) digital data transmission from shaft to casing. This means no failure of transmission and maintenance free. 51

Strain gauges operation principle a) Change of electrical resistance c) Array of 4 single strain gauges affixed with 90 misalignment (Wheatstone Bridge: The changes of strain resulting from the resistance change are detected. These are proportional to the applied loads) b) Principle of strain gauge (horizontal stress) 52

PolyLab OS Software Concept Two software layers: 1) PolyLab Monitor OS: - Basic software to control and monitor the PolyLab OS - Visualization of the attached equipment and sensors - Online data export in Excel (XP) - No data evaluation 2) PolySoft Modules: - Application dedicated software modules (Mixer Test / Capillary-Rheology / Filter Test / Advanced Extrusion) - Job stream structure, incl. data evaluation (similar to Haake RheoWin) - The user is guided through the measurement - Different user levels 53

PolyLab Monitor OS Plug & Measure easy Data monitoring Scaleable windows for measured values Visualized process values online export data to MS Excel (XP) Online sensor recognition ( hot plugging ) 54

PolyLab Monitor OS Graphic monitoring of instrument and test values Scaleable windows for measured values 55

PolyLab Monitor OS Heating/Cooling visualization Monitor the regulating values (in %) of each control zone Get important information for extruder screwconfiguration 56

PolyLab Monitor OS On-Screen Remote Control to keep control of the PolyLab even when using other Windowsε based software 57

PolyLab OS PolySoft Mixer PolySoft Mixer SW: - Job stream structure analog Haake RheoWin - The user is guided through the measurement - Project handling: several tests can be saved in a project 58

PolyLab OS PolySoft Mixer Protocol Element: - Documentation of sample information 59

PolyLab OS PolySoft Mixer Device Setup: - Automatic recognition of mixer setup - Rotor selection 60

PolyLab OS PolySoft Mixer Measurement Element: - Setting of the testing parameters - Calculation of the sample-weight - Programming of sophisticated test steps 61

PolyLab OS PolySoft Mixer Evaluation Element : - Selection of evaluation routine for the determination of the curve maxima, minima and gradients - Modification of the evaluation routine 62

PolyLab OS PolySoft Mixer Post Element : - Generation of file name - Definition of printouts 63

PolyLab OS PolySoft Mixer Export Element : - Definition of data export 64

PolyLab OS PolySoft Mixer Statistic function : - Statistical evaluation of measurement data - Comparison of sample behavior over a longer time period 65

PolyLab OS PolySoft Capillary SW Step choose : - Speed-steps from low to high, as well from high to low possible - Speed-steps from low to high and from high to low possible in one measurement possible 66

PolyLab OS PolySoft Capillary SW Step settings : - Adjustments for each step possible - Possibility for longer step-time at low shear-rates, to get more accurate data - Possibility for short step-time at high shear-rates, to save time & sample 67

PolyLab OS PolySoft Job Controller 68

PolyLab OS PolySoft Regression 69

PolyLab OS PolySoft Regression 70

Tests with the PolyLab OS PolyLab OS relative Measurements Other CAN Sensors absolute Measurements Mixer tests Extruder tests rheological Measurements Single-screw extruders Twin-screw extruders Rod Capillary tests Slit Capillary tests 71

PolyLab System Laboratory Mixers : 72

Laboratory Mixers - measuring principle Shearing of a test sample in a heated mixing chamber with counter rotating rotors 3 2 Test results: Ł Torque Ł Melt temperature 73

Mixer - Rheogram Temperature Torque Energy 74

Electrically heated Rheomix mixer (principle sketch) 1 4 5 1 Back Section 2 Centre Bowl 2 3 Frontplate 7 6 4 Rotor Shafts 3 5 Bushings 9 6 Rotors 8 7 Melt Thermocouple 8 Control Thermocouple 7 9 Ram 75

Rotors for Rheomix mixers 76

Rotors and Applications Roller Rotors Mixing of thermoplastics, e.g. Polyolefines, PVC, engineering plastics Banbury Rotors Mixing of elastomers, but also for compounding of powder into thermoplastic material. Very common in rubber technology 77

Rotors and Applications Delta Rotors Mixing and crosslinking of Thermosets, used with conical bowl of 540 type mixer Cam Rotors Thermoplastics, less axial distribution, ceramic compounds, food (sticky, higher torque) Sigma Rotors Food application and Plastisols limited in torque 78

Electrically heated Mixers: Rheomix 600 / 3000 electrically heated 3 heating zones : backplate / centre bowl / frontplate temperature range up to 450 C possibility to run temperature programs removable rotor blades: - Roller Rotors - Cam Rotors - Banbury Rotors - Sigma Rotors loading ram for manual or pneumatic operation 79

Liquid heated Mixers: Rheomix 610 / 3010 liquid heated for lower temperatures (i.e. mixertests with rubber) 3 heating zones : backplate / centre bowl / frontplate temperature range below 200 C removable rotor blades: - Roller Rotors - Cam Rotors - Banbury Rotors - Sigma Rotors loading ram for manual or pneumatic operation 80

Advantage of round feed section Fr Fr: resulting force, spreading old style mixer bowls Fm Fm Fm: force created by mixing of the material 81

Mixer Test Procedure: 82

Mixer Cleaning: 83

PolySoft Routine evaluation - Curve Type 1 Torque [Nm] M L 0 1 2 3 4 5 6 7 8 9 10 Runtime Materials: Polyolefins: (Polyethylene PE Polypropene PP ) Engineering Plastics: (PS, PA, PC, PEEK, LCP ) Point detection: L = Loading point M = Minimum point 84

PolySoft Routine evaluation - Curve Type 1 Loading Point Detects complete mixer loading. Point is only used as time base for further calculations Torque value depends on the loading conditions Minimum point Detects lowest melt viscosity 85

Influence of Stabilizer on Polyamide Torque [Nm] 80 70 60 50 40 30 20 10 0 Rheomix600, Roller Rotors Temp.: 280 C, m: 52g, n: 60 rpm PA6 with Stabilizer PA6 without Stabilizer 0 5 10 15 20 25 Time [min] 86

SAN with Carbon Black 30 Torque [Nm] 25 20 15 10 Rheomix600, Roller Rotors Temp.: 230 C, m: 58g, n: 40 rpm SAN & 30% Carbon Black 2 5 0 SAN & 30% Carbon Black 1 0 2 4 6 8 10 12 14 Time [min] 87

Exfoliation of Monmorillonite Torque [Nm] L 0 1 2 3 4 5 6 7 8 9 10 Runtime t [min] 5% organoclay [1] M Materials: Polyolefins: PE, PP Engineering / high performance Plastics: (PS, PA, PC, PEEK, LCP ) Compounds of a.m. polymers with organo clays, nano tubes Point detection: L = Loading point M = Minimum point (steady state torque) Observe the melting behaviour and flowability of a compound We have noted that a simple indicator of the extent of clay exfoliation within polyethylene is the steady-state torque recorded during melt compounding 1. [1]Ref.: Influence of clay exfoliation on the physical properties of montmorillonite/polyethylene composites T.G. Gopakumar, J.A. Lee, M. Kontopoulou*, J.S. Parent Polymer 43 (2002) 5483 5491 88

PIM: Critical (powder) loading of feedstock The erratic reading in the steps for increased powder load show the limiting powder [%] in the mixture. Add powder Contact free and fast, high resolution torque sensors have a big advantage over old style dynamometer where the reaction of a heavy motor is measured, the erratic reading, here key feature is damped or smoothed Ref. LR45-e, Joseph A. Krudys, ThermoHaake 2002 89

PolySoft Routine evaluation - Curve Type 2 Torque [Nm] Materials: PVC (rigid) Point detection: L = Loading point V = Valley point (starting point of melting process) F = Fusion point ( melting process finished) M = Minimum point L V F M 0 1 2 3 4 5 6 7 8 9 10 Runtime [minutes] 90

PolySoft Routine evaluation - Curve Type 2 Loading Peak: Shows when mixer is completely filled and closed. Point is used as time base for calculation only. Torque value depends on kind of loading. It can be manually influenced by loading speed and loading force. Valley Point: Shows when PVC Dryblend starts to melt. Point is significant for compound formulation. Fusion Peak: Shows when melting process of PVC is finished. Point is significant for compound formulation. Beyond this point torque starts decreasing caused by friction/heat. Minimum Point: Shows when material has reached the lowest viscosity. 91

Influence of the sample weight on PVC fusion Rheomix600, Roller-Rotors, Temp.: 160 C, n: 40 rpm m: 66 g m: 64 g m: 60 g 92

Influence of the sample weight on PVC fusion m: 60 g 93

Influence of the sample weight on PVC fusion m: 66 g 94

PolySoft Routine evaluation - Curve Type 3 Torque [Nm] Stable time L V F S M O D 0 3 6 9 12 15 18 21 24 27 30 Runtime [minutes] Materials: PVC (rigid) Stability test Point detection: L = Loading point V = Valley point (starting point of melting process) F = Fusion point ( melting process finished) M = Minimum point O = Onset of degradation D = Degradation 95

PolySoft Routine evaluation - Curve Type 3 Loading Peak Shows when mixer is completely filled and closed. Point is used as time base for calculation only. Torque value depends on method of loading. Valley Point Shows when PVC Dryblend starts to melt. Point is significant for compound formulation. Fusion Peak Shows when melting process of PVC has finished. Point is significant for compound formulation. Beyond this point torque decreases caused by frictional heat. Minimum Point Shows when material has reached the lowest viscosity. 96

PolySoft Routine evaluation - Curve Type 3 Onset of Degradation Shows when material starts to degrade. Point is significant for processing stability. It can be set as percentage value of torque above minimum torque. Stable torque point Shows when torque reaches stable torque. Stable torque is set with onset of degradation point. Stable time Time between "S" (stable torque point) and "O" (onset of degradation point). Value gives information about processing time and stability of compound. Degradation Peak Shows when degradation has reached highest value. Torque value is not significant and only used for the calculating of the degradation rate. 97

Gelling of PVC 1st. Phase Ł Intact globular structure of PVC particles 98

Gelling of PVC 2nd. Phase Ł Globular structure destroyed Ł Flow of primary particles Ł Free volume between particles is reduced 99

Gelling of PVC 3rd. Phase Ł Destruction of primary particles Ł Flow of subparticles Ł Crosslinking moleculare Network Ł Additives (Stabilisers) visible in network 100

Application: PVC Stability test 50 45 40 Rheomix600, Roller Rotors Temp: 170 C, n: 60rpm, m: 65g Torque [Nm] 35 30 25 20 15 10 5 a b a b a: PVC Dry-Blend with 1,9% Stabilizer b: PVC Dry-Blend with 2,0% Stabilizer 0 0 5 10 15 20 Time [min] 101

Correlation Mixertest / Extrusion Example: PVC-Stability Torque Time 102

PolySoft Routine evaluation - Curve Type 4 & 5 Materials: Torque PVC granulate (Stability test) X linking materials (PE, Rubber, Thermoset ) Point detection: Stable time L = Loading point S = Stable Torque L S M O D Runtime M = Minimum point O = Onset of Degradation D = Degradation 103

PolySoft Routine evaluation - Curve Type 4 & 5 Loading Peak Shows when mixer is completely filled and closed. Point is used as time base for calculation only. Value depends on kind of loading. It can be influenced by loading speed and loading force. Minimum Point Shows when material has reached the lowest viscosity. Onset of degradation or cross linking point Shows when material starts to degrade or cross link. Point is significant for process. It can be set as percentage value of torque above minimum torque. 104

PolySoft Routine evaluation - Curve Type 4 & 5 Stable torque Point Shows when torque reaches stable value. Stable torque is set with onset of degradation point. Stable time Time between "S" (stable torque point) and "O" (onset of degradation/cross linking point). Value gives information about processing time and stability of testing material. Ł Degradation Point Shows when material has been degraded. Torque value is not significant. It is used to calculate degradation/cross linking rate. 105

Rubber-Compounds with the same Mooney Viscosity 30 Torque [Nm] 25 20 15 10 5 0 Batch 1 0 1 2 3 4 5 6 7 8 Time [min] Batch 2 Rheomix600, Banbury Rotors Temp.: 130 C, n: 60 rpm, m: 63 g 106

Thermoset Mixer Rheomix 540 (principle sketch) Backwall Bushings with O-rings Conical Rotors.... Conical Centre bowl Frontplate with melt thermocouple 107

Crosslinking of PES-Resins 12 Torque [Nm] 10 8 6 4 Rheomix540, Delta Rotors Temp: 120 C, n: 50rpm, m: 63g Sample 1 Sample 2 2 0 0 1 2 3 4 5 Time [min] 108

Mixer Application: 109

Mixer Application: Batchdifferensation with speed-program 2 Rubber-Compounds Torque [Nm] Rheomix600, Banbury-Rotoren Temp.: 130 C, m: 80g n = 70 rpm Compound 1 Compound 2 n = 5 rpm Time [min] 110

Mixer Application: Production-Improvment with Lab-Mixers Production of Hot-Melts: procedure 1: A + B + C Torque [Nm] procedure 2: A + C + B Torque 1 Energy 1 Energy 2 Energy [kj] Torque 2 Time [min] 111

Gas Flow Sensor for Rheomix Teflonblock Screen Gas-Outlet Application: - foaming Compounds - degradation tests 112

Measuring results Gasflowsensor: 113

RheoMix3010 Food-Applikation 114

Planetary mixer for PolyLab system Application: Determination of plasticizer absorption in vinyl chloride (VC) polymers (PVC- Dryblends) Mixing of solid powders with liquids Preparation of a PVC paste for testing purposes under controlled conditions Circulator Standards: DIN 54 800 / DIN 54802 / ISO 4612 115

Planetary mixer for PolyLab system Technical Specifications mixer bowl volume: app.. 2.500 cm 3 temperature control: liquid (water or oil) temperature range:» Water: Tmax = 95 C» Oil (Synth 210): Tmax = 150 C n max. (PolyLab) = 200 1/min» gear ratio for planetary rotor: i = 1:4» gear ratio for skimmer: i = 3:4 Tq max. = 50 Nm 116

Planetary Mixer curve - Plasticizer absorption Speed 100 rpm 400g PVC Powder 200g DOP Plasticizer Temperature variation 117

Electrical Conductivity Measurement A new approach by combining traditional methods with new technologies 1 3 3 2 1 Ram 2 Control Thermocouple 3 Rotors 4 Conductivity Sensor 4 118

Electrical Conductivity Measurement The electrical conductivity of rubber compounds as a measure for the dispersion of carbon black is most important. Studies on the final compounds by use of classical laboratory instruments have been reported (e.g. Carbon Co., Fort Worth). Getting those information during the mixing process can be of great importance for quality control and waste reduction. Thermo Fisher Scientific has now successfully integrated a electrical conductivity sensor in a laboratory scaled mixer. Integrated Conductivity Sensor 119

Electrical Conductivity Measurement for Rubber Compound Test 1: Test material: SBR 1500 with 32 vol.% Carbon black N220 Carbon Black characteristics: relatively fine with high specific surface Test 2: Reproducibility of Test 1 Test 3 Test material: SBR 1500 with 32 vol.% Carbon black N330 Carbon Black characteristics: relative to N220 the N330 has bigger particle sizes and thus a lower specific surface 120

Electrical Conductivity Measurement for Rubber Compound 1 --- Torque --- Conductivity Test material: SBR 1500 with 32 vol.% Carbon black N220 121

Electrical Conductivity Measurement for Rubber Compound 2 --- Torque --- Conductivity --- R-Torque --- R-Conductivity Reproduction : SBR 1500 with 32 vol.% Carbon black N220 122

Electrical Conductivity Measurement for Rubber Compound 3 --- Torque --- Conductivity Test material: SBR 1500 with 32 vol.% Carbon black N330 123

Change from Mixer to Extruder 124