High performance radio frequency generator technology for the Thermo Scientific icap 7000 Plus Series ICP-OES

Similar documents
icap 7000 Plus Series ICP-OES Streamlined performance and ultra-low ICP-OES detection limits for routine laboratories

Trace Elements in Petrochemical Samples: Achieving Excellent Analytical Results

High matrix tolerance of the Thermo Scientific icap 7000 Plus Series ICP-OES with the radial Ceramic D-Torch and the sheath gas adaptor

Benefits of running organic matrices using the Agilent 5100 ICP-OES fast, robust, high performance analysis

Eco Plasma Innovative Inductively Coupled Plasma Technology

TECHNIQUES OF INDUCTIVELY COUPLED PLASMA OPTICAL EMISSION SPECTROMETER S. KARTHIKEYAN SSA CENTRAL POLLUTION CONTROL BOARD SOUTH ZONE OFFICE BANGALORE

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

Analysis of steel and its alloys using the GB/T standard and an Agilent 5100 ICP-OES in dual view mode

Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

Analysis of wear metals and contaminants in engine oils using the 4100 MP-AES

Analysis of Mineral and Heavy Metal Content in Beverages Using the Teledyne Leeman Labs Prodigy Plus ICP-OES

The Determination of Alloying Elements in High-Carbon Steel using Radially Viewed Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES)

Increase productivity for environmental sample analysis using the SVS 2+ Switching Valve System for Agilent 5100 SVDV ICP-OES

Optima 8x00 Series. ICP Optical Emission Spectrometers

Optima 8x00 Series. ICP Optical Emission Spectrometers

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

EXPAND YOUR RANGE EXTEND YOUR RESOURCES. Avio 500 ICP Optical Emission Spectrometer

Determination of Elemental Impurities in Graphite-based Anodes using the Agilent 5110 ICP-OES

Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

Analysis of inorganic impurities in lubricating oils by ICP-MS

Agilent 5100 ICP-OES DUAL VIEW ICP-OES MINUS THE WAIT

Metals Analyzer. OES 6000 Optical Emission Spectrometer. fast and accurate metal analysis

Analysis of Trace Metals in Tobacco and Tobacco Ash using Atomic Emission Spectroscopy

AMAZINGLY CAPABLE REMARKABLY AFFORDABLE. Avio 200 ICP Optical Emission Spectrometer. For research use only. Not for use in diagnostic procedures.

Analysis of elemental impurities in drug products using the Thermo Scientific icap 7600 ICP-OES Duo

Multi-Element Analysis of Trace Metals in Animal Feed using ICP-OES

Avio 200 ICP-OES Flat Plate

Direct Analysis of Photoresist by ICP-MS. Featuring the Agilent Technologies 7500s ICP-MS

EU water analysis using the Thermo Scientific icap 7400 ICP-OES Duo

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo

Tender Document. National Centre of Excellence in Analytical. Chemistry, University of Sindh, Jamshoro. Inductive Coupled Plasma (ICP -OES) ISSUED BY:

Analytically Monitoring the Effect of Fracturing Activity

Utilizing Glow Discharge in Optical Emission Spectroscopy

Abstract No. 21. Excitation Mechanisms in Microwave-induced Plasma Excited with Okamoto-cavity and the Spectrometric Application to Steel Analysis

Thermo Scientific icap 6000 Series ICP Emission Spectrometer

Thermo Scientific icap RQ ICP-MS. Simplicity, productivity and robustness for routine labs

Analysis of elemental impurities in drug products using the Thermo Scientific icap 7400 ICP-OES Duo

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

Automating EPA 6020 Compliant Analysis with the Agilent 7900 ICP-MS and ESI prepfast Autodilution System

GDMS. High Resolution GDMS.

DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, & USING THE AGILENT 5100 ICP-OES

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

ICP-OES. Varian Vista-MPX CCD Simultaneous ICP-OES

Thermo Scientific icap Q ICP-MS

Ultra-fast ICP-OES determinations of base metals in geochemical samples using next generation sample introduction technology

Radial, Axial or Dual View ICP: Which Do You Choose? Manny Almeida Teledyne Leeman Labs, Inc. Hudson, NH

EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo

Water Analysis Using ICP-OES with an Ultrasonic Nebulizer

Thin Films: Sputtering Systems (Jaeger Ch 6 & Ruska Ch 7,) Can deposit any material on any substrate (in principal) Start with pumping down to high

ICP-AES Determination of 15 Kind of impurity elements in the Vanadium-aluminum alloy

Atomic Spectroscopy Solutions for Environmental Applications

Korkealämpötilaprosessit

Analysis of Samples Related to Nuclear Submarine Operation

Linear Plasma Sources for Surface Modification and Deposition for Large Area Coating

Inductively coupled plasma (ICP) spectroscopy for online measurements of trace metal emissions at the 250 kw PACT furnace

Thin Films: Sputtering Systems (Jaeger Ch 6 & Ruska Ch 7,) Sputtering: gas plasma transfers atoms from target to substrate Can deposit any material

Thermo Scientific icap 7000 Series ICP-OES Accessories Guide

Multi-element determination in pharmaceutical preparations using the Thermo Scientific icap Q ICP-MS

Most semiconductor devices contain at least one junction between p-type and n-type material. These p-n junctions are fundamental to the performance

Plasma Inductive Coupling Plasma-Atomic Emission Spectrometer (ICP-AES) >>>>>>>>

Evaluation of Effects of Stirring in a Melting Furnace for Aluminum

ICP Expert software. Technical Overview. Introduction

Agilent Atomic Spectroscopy Solutions for Energy & Chemicals AA, MP-AES, ICP-OES, ICP-MS.

Introducing the new Agilent 7800 Solution-Ready ICP- MS. Agilent Technologies 2015

Chapter 5 LOWERING THE DETECTION LIMIT FOR ARSENIC: OUTLOOK FOR A NEW PRACTICAL QUANTITATION LIMIT

Simple, fast and reliable analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

NO INTERFERENCE BETWEEN YOU AND BETTER THROUGHPUT. NexION 1000 ICP Mass Spectrometer. For research use only. Not for use in diagnostic procedures.

Thermal Evaporation. Theory

DETERMINATION OF SULPHUR CONTENT OF UN-ALLOYED AND LOW-ALLOYED STEEL BY ICP-AES SPECTROMETRY USING WET CHEMICAL SAMPLE PREPARATION

Increased Sample Throughput for ICP-OES Applied to U.S. EPA Method 200.7

The Determination of Trace Elements in Waters using the US EPA Method and the Thermo Scientific icap 6500 ICP-OES (Duo)

Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines

WM 05 Conference, February 27 March 3, 2005, Tucson, AZ THE ARCHIMEDES FILTER PLANT: A NEW CONCEPT FOR REDUCING THE AMOUNT OF HLW AT HANFORD

Table I: MCL and MRL Concentrations for Contaminants Monitored Under the Safe Drinking Water Act National Primary Drinking Water Regulations

Analysis of Heavy Metals in Iron-based Fertilizers by HR ICP-OES

Accurate. Robust. Reliable. AGILENT 710 SERIES ICP-OES

Accurate. Robust. Reliable. AGILENT 710 SERIES ICP-OES

Image Area YOUR PRESCRIPTION FOR USP 232/233 COMPLIANCE. USP 232 / 233 Readiness Solutions and Services Overview

Partial Oxidation of Methane to Form Synthesis Gas in a Tubular AC Plasma Reactor

Dr. Jianfeng Cui, Thermo Fisher Scientific, Bremen, Germany APPLICATION NOTE AN43384

Lab 6 Measurement of Ozone

Metal Aerosol Emissions from Biomass Combustion

STUDY OF ELEMENT FINGERPRINTING IN GOLD DORÉ BY GLOW DISCHARGE MASS SPECTROMETRY

Optimization of Ion and Electron Properties in IC Packaging Applications

SMS 100 MERCURY ANALYZER SMS 100. Mercury analysis directly on solid samples from the undisputed leader in Atomic Spectroscopy

Ultima Expert Ultimate ICP-OES Spectrometer Best Performance & Ease-of-Use

Characterization of Surface Metals on Silicon Wafers by SME-ICP-MS. Featuring the Agilent Technologies 7500s ICP-MS

USABILITY BY DESIGN Every component of the Agilent 7900 ICP-MS is engineered for performance and reliability

Ultima Expert Ultimate ICP-OES Spectrometer Best Performance & Ease-of-Use

Analysis of Trace Metals in Surface and Bottled Water with the Optima 7300 DV ICP-OES

Product Descriptions and Offerings

Plasma Quest Limited

Featuring Analyst software under Windows NT for enhanced performance and ease of use. API 150EX. LC/MS System

// XPLORER-NS. Total Nitrogen and Sulfur Analyzer

Analysis of Environmental Samples with the Agilent 730-ES Following US EPA Guidelines

Spectroscopy Performance Note

Confident Quantitation

Transcription:

TECHNICAL NOTE 43334 High performance radio frequency generator technology for the Thermo Scientific icap 7000 Plus Series ICP-OES Keywords Free-running, Plasma, RF generator, Solid-state Using inductively coupled plasmas for optical emission spectrometry An inductively coupled plasma (henceforth referred to as a plasma ) is formed by seeding ions in a stream of argon which is flowing through Radio Frequency (RF) and magnetic fields in the region of the induction coil (Figure 1). The ions are formed by subjecting the argon gas to a high voltage spark and this causes small amounts of ionization of the argon. These electrons and ions are then accelerated by the magnetic field and collide with other argon atoms, causing further argon ionization and the subsequent formation of a plasma. The plasma is sustained by continual ionization of the argon via collisions induced by the magnetic field. This transfer of energy is known as inductive coupling. The energy that is transferred to the electrons and ions originates from the RF generator. The plasma is contained within a quartz or ceramic tube known as a torch and can reach temperatures of up to 10,000 C. To prevent melting or damage to the torch the argon is forced to flow tangentially at a high velocity, this cools the inner surface of the torch, forming a compact and stable plasma (Figure 1).

Temperature 6000 C 6500 C 7000 C 8000 C 100000 C Induction zone Normal analytical zone of plasma Induction coils Magnetic field Common challenges for plasma generation The design challenges for plasma generation in new Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES) instruments are most noticeable in the older style generators. If the challenges are not solved then the effectiveness of the ICP-OES instrumentation is reduced. These challenges may include considerations such as size, robustness, efficiency, reliability and ease of service: RF generators are commonly large, heavy units, thereby forcing the ICP-OES instrument to be even larger. Sample flow Coolant gas Figure 1. Plasma diagram, heating and viewing zones. Aerosols generated from liquid or solid samples are fed into the plasma from various types of sample introduction systems or external accessories through a centrally located tube inside the torch. The plasma is at its most energetic and hottest where the sample is injected and the liquid evaporates and completely dissociates into its constituent atoms and ions within milliseconds (Figure 2). An efficient and powerful RF generator will develop and continually sustain a plasma that is capable of completely dissociating almost any sample matrix, thereby reducing oxide formation and other chemical interferences to a minimum. Traditional generators use power tubes and may not be able to cope with rapid transitions between heavy matrices, such as high metal or salt content and organic solutions. ICP-OES instrumentation with power tube generators have always been inefficient. Much of the power that is used with these instruments must be extracted as heat, this is wasteful and measures must be taken by laboratory to counteract the heat generated. Generators utilizing power tubes will need to have the power tubes replaced approximately every two years, depending on usage, and breakdowns due to old power tubes can occur at any time. Other generators may use many different parts, increasing the likelihood of failure. Traditional generators are complex items which require time-consuming procedures to repair if a failure occurs. Simplified generators can be serviced in a very short time with minimum downtime. Figure 2. Sample processes in the plasma. The constituent atoms and ions of the sample will now be able to absorb the energy from the surrounding plasma and will emit light at their own characteristic wavelengths. The emissions result from electrons absorbing energy in specific amounts, entering an excited state and subsequently releasing that energy as a photon of characteristic wavelength after returning to their original state. Due to the large energies present in the plasma, intense emission wavelengths are obtained making the plasma a very good tool to enable sensitive measurements of trace-level impurities with simultaneous measurement of high concentration sample constituents. The stability of an RF generator greatly depends on its ability to adjust for changing conditions in the plasma due to differing samples or sample matrices by changing the power conditions to suit the variations this is known as matching. Historically, there have always been two main approaches to the control and matching of RF generators, crystal controlled or free-running. Crystalcontrolled generators match and lock the frequency of the RF generator to the oscillation of a reference crystal (Figure 3). Free-running generators match the power generated to the power required by the plasma and allow the frequency to vary slightly (Figure 4).

Figure 3. Simplified crystal-controlled RF generator design. Rapid changes in plasma composition, such as changes in sample type or matrix, require an RF generator that will quickly respond to changes and achieve stable plasma as rapidly as possible. For instance, the plasma may be extinguished on older, slower reacting RF generators by introducing an organic solvent into the plasma directly after an aqueous sample. Figure 4. Simplified free-running RF generator design. Thermo Scientific icap 7000 Plus Series ICP-OES solid-state RF generator design The Thermo Scientific icap 7000 Plus Series ICP-OES solid-state RF generator is an advanced, free-running RF generator, with solid-state chips and circuits replacing the vacuum power valves and other outdated components of traditional designs. Historic designs tended to be slow to match the plasma to different matrices whereas the icap 7000 Plus Series ICP-OES RF generator is solid-state and responds rapidly to changes within the plasma. In addition, the icap 7000 Plus Series ICP- OES RF generator is a simple design which is compact, robust, power efficient and reliable. The efficient design with fewer components enables easier, rapid servicing and helps to reduce instrument downtime since the components can be easily replaced if required. The icap 7000 Plus Series ICP-OES RF generator is a free-running, solid-state, 27.12 MHz RF generator capable of transferring 1600 W of RF power into the plasma with power regulation better than 0.1%. The solid-state design provides stability, reliability and highpower in a small footprint. This innovative RF generator utilizes solid-state transistors, a 3kW solid state DC power supply, printed circuit boards and field-effect transistors (FET) in an integrated design. The design incorporates only three major parts plus the induction coil (Figure 6), all solidstate, providing a robust and reliable design. Figure 5. Pictures of a traditionally designed crystal-controlled RF generator. The power supply (top) and match box (bottom) give a total mass of both is approximately 90 100 kg. The total volume and weight of this generator is greater than that of a complete icap 7000 Plus Series ICP-OES instrument. Induction coil DC power supply Power supply controller RF generator board Figure 6. The schematic shows the DC Power Supply Unit (PSU), feedback Power Supply Controller, main RF Generator Board and Induction Coil.

The diagram in Figure 4 shows the simplicity of the icap 7000 Plus Series ICP-OES RF generator design. In comparison to the crystal-controlled RF generator (Figure 3), the new RF generator board performs the same function as the match box and the new DC power supply equates to the large, heavy power cube. The entire mass of the icap 7000 Plus Series ICP-OES RF generator is approximately 20 kg with the casing forming the majority of the mass and less than one quarter of the volume of the crystal-controlled generator in Figure 5. Thermo Scientific icap 7000 Plus Series ICP-OES RF generator capabilities and benefits The icap 7000 Plus Series ICP-OES highly advanced solid-state RF generator has many advantages over the original power tube designs and less efficient solid-state designs. Efficiency The icap 7000 Plus Series ICP-OES RF generator provides high efficiency of greater than 78% DC to RF power even under the most difficult conditions. This means that the RF generator supplies a minimum of 78% of the incoming power to the plasma, enabling the icap 7000 Plus Series ICP-OES instrument to analyze challenging organic solvents and salt solutions containing high concentrations of metals, even when using low power settings. Impurities in metal plating solutions are a common, challenging analysis due to the high metal content of the matrix (typically containing 50-100 g L -1 of metal or more). Figure 7 shows a calibration graph of silver standards (using the Ag 328.068 nm wavelength) in 50 g L -1 nickel solutions at various power levels. All calibrations obtained are remarkably linear even when low RF power is used and this in the presence of high concentrations of nickel in solution. If a less efficient RF generator was used, there is the possibility that excessive curvature of the calibration may occur at lower RF powers or the plasma may even be extinguished. Methanol is a volatile water-miscible organic solvent. When analyzed by ICP-OES, the solvent may generate too much organic aerosol and extinguish the plasma. Commonly, these types of volatile solvents are analyzed at high power settings and the spray chamber is cooled to reduce the amount of organic vapor produced, sometimes as low as -15 C. Efficient transfer of energy into the plasma enables the icap 7000 Plus Series ICP-OES RF generator to keep the plasma stable, even at ambient temperatures or lower RF powers. A 50% methanol solution was analyzed for copper (Cu 327.396 nm) using a standard organic sample introduction kit at ambient temperature. The calibrations shown in Figure 8 demonstrate that the RF generator within the instrument is capable of enabling excellent linear calibrations in the most difficult matrices. Figure 8. Calibration graphs of copper in 50% methanol at various power levels. Although some RF generators cannot generate stable plasmas with high concentrations of metal and organic solvents, especially at low powers, the icap 7000 Plus Series ICP-OES solid-state RF generator removes the restrictions on power for even the most difficult matrices. The instrument solid-state RF generator, with its efficient power transfer, producing excellent calibrations at high or low RF powers, enables the user to determine the optimum power settings for the element suite to be analyzed. The icap 7000 Plus Series ICP-OES RF generator also demonstrates efficiency by the minimal amount of heat energy it wastes. Even at the highest RF power settings, the amount of heat generated from the entire instrument is minimal (less than 700 W) with the main RF generator contributing less than 300 W. Figure 7. Calibration graphs of silver in 50 g L -1 nickel solutions at various power levels. The efficiency of the icap 7000 Plus Series ICP-OES RF generator enables reduced power consumption, lower waste heat generation coming from the generator and a greater capacity to be able to reduce power while still achieving the same performance, with or without challenging matrices.

Stability Stability of the icap 7000 Plus Series ICP-OES RF generator is enhanced because of its efficient power usage better cooling and less susceptibility to intensity drift due to environmental change. In addition, matrix composition does not affect the RF generator performance as much as in older designs due to the rapid response and power efficient design. The icap 7000 Plus Series ICP-OES solid-state RF design allows the plasma to reach a stable state after changing matrices much more rapidly than older designs. This produces more consistent, stable intensities when changing samples, organic solvents or even from water to high salts solutions. Figure 9 shows the excellent stability (with <2% RSD over the 8 hour period) achievable by the icap 7000 Plus Series ICP-OES RF generator. The solution used was a 50 g L -1 nickel solution spiked with various concentrations of elements. The solution was continuously aspirated and the intensities plotted as a percentage over the full 8 hour analysis. Such stability enables fewer calibrations, fewer QC failures and, hence, increased sample throughput and reduced costs compared to older style, legacy ICP-OES instruments RF generators. Robustness The icap 7000 Plus Series ICP-OES RF generator is especially robust with a rapid response time to changes in the plasma loading. In contrast to older designs, solidstate RF generators are known for their rapid reaction times and robustness when dealing with changing matrices. This rapid response enables the intensity to stabilize rapidly, giving a flat plateau immediately after the wash-in period. This enables the analyst to shorten the stabilization time since the RF generator requires less time to reach its optimum and stable power level. The benefit of the greater robustness of the icap 7000 Plus Series ICP-OES is improved stability when changing sample matrices in the same method and reduced stabilization time, reducing the analysis time as a whole. Conclusion The development of a small, efficient, rapid response solid-state RF generator enables the Thermo Scientific icap 7000 Plus Series ICP-OES instrument to have a very small footprint and robust sample handling capabilities. Even at low RF powers, the RF generator robustness keeps the plasma stable and enables analyses in challenging matrices that most other RF generators need much higher power settings to perform. Greater stability from the solid-state RF generator compared to traditional RF generators also enables the analyst to utilize longer, more efficient analyses without the need for recalibration. The Thermo Scientific icap 7000 Plus Series ICP-OES RF generator s innovative solid-state design has resulted in an efficient, ICP-OES solution which provides the lowest cost of analysis per sample. Figure 9. Stability analysis (8 hours) in a 50 g L -1 nickel solution. Find out more at thermofisher.com/icp-oes 2017 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. TN43334-EN 1117