STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation

Similar documents
REUSABLE MEDICAL DEVICE VALIDATION

IN THIS SECTION MICROBIOLOGY TESTING EXPERT SOLUTIONS FOR PRODUCT DEVELOPMENT. Bacterial Endotoxin (LAL) Testing

DUAL SPECIES BIOLOGICAL INDICATOR SPORE STRIPS For Monitoring Ethylene Oxide (EO), Steam and Chemiclave Processes

sterility assurance Prove the power and your processes

Guidance for Industry

applying a unique combination of disciplines consulting, preclinical, toxicology,

2018 North America Lab Services Fee Schedule

Sterilization - validation, qualification requirements. Sterilization - Overview

CERTIFICATE OF ACCREDITATION

CERTIFICATE OF ACCREDITATION

How is Sterilization Validation Performed by the Medical Device Manufacturer?

Non-Heat Terminal Sterilization of Controlled Release Materials

Date: 17/12/2012 Reviewed: 21/12/2012 Issue: 2 SI No: P3149 Prepared by: G.Ostacchini Reviewed by: A.Green Approved by: D. Johnson

NO 2 Gas Sterilization and Decontamination: Keeping Pace, from Start to Finish

Biological Indicator Reduced Incubation Time (RIT) Presented by: Andrew Melloy

Raven Labs. Sterility Assurance for Industry. Sterility Assurance Products Since A division of Mesa Laboratories, Inc.

Moist Heat Terminal Sterilization for Controlled Release Materials

Quality Control EQUIPMENT CONTROL. Monitor critical performance characteristics of sterilization, decontamination or cleaning equipment

Annex A2. Guidance on Process Validation Scheme for Aseptically Processed Products

RADIATION IS COOL! Martell Winters. Director of Science

Secure. Tested. Certified.

5.1.1 & Changes started in 2008 and were mainly driven under the Chairmanship of Prof Hans van Doorne 11/10/2017

TEST CODE DEFINITION STERRAD H202. Test Code Test Name Testing Description. Cycle. Efficacy. SE-200 STERRAD 200 Efficacy. Cycle. STD Cycle Efficacy

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar

Validating reusable medical and dental devices for over 30 years

SterilAmp II Geobacillus stearothermophilus

Medline Gemini Sterilization Wrap. Test Studies

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية

ISO INTERNATIONAL STANDARD

PRODUCT D-VALUE STUDIES: A CRITICAL TOOL WHEN DEVELOPING A STERILIZATION PROCESS

THE TERMS VALIDATION AND VERIFICATION ARE FREQUENTLY

Biological Monitoring

Low Temperature Processing. An Update on Guidelines, Standards and Requirements

Instructions for Validation of Autoclaves

ISO INTERNATIONAL STANDARD

Validation of cleaning and sterilization of reusable medical devices

What Dentistry Can Learn from AAMI ST79. Best Practices for device and instrument reprocessing

Written By Dr. Robert Ramaley (Univ. NE Med CTR), Mia Ware, Dan Pearson, and Christopher R. Dwyer* (Raven Biological Laboratories)

Sterilization & Chemical Resistance of Healthcare Polymers

PACKAGE TESTING & VALIDATION IN THIS SECTION EXPERT SOLUTIONS FOR PRODUCT DEVELOPMENT

Secure. Tested. Certified.

Biological Indicators and Their Application in Clinical Sterilization Spores Don t Lie 1

Quality Control LOAD CONTROL. VERIFY Load Control Products. Use to monitor and release instrumentation with confidence RECORD KEEPING & LABELING

Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy!

THE BASICS OF STERILIZATION

Monitoring Flexible Endoscope Channels to Assure Cleaning

TECHNICAL DATASHEET. in compliance with: USP, ISO and all appropriate subsections FQ 106 v3.0 August, 2015

Lets Talk Indicators Challenging the process Sterilization. ICAN 5 th November Peter Newson

Services for Medical Devices. Expert solutions for product development from concept to commercialization

Your Sterile Processing Partner. Duraholder IPS

ISO INTERNATIONAL STANDARD

White Paper. Population C, The Standard Distribution of Resistances: Analysis and Commentary 5/30/17

Loaner Program Guidance

Technology Overview. Ethylene Oxide

PACKAGE TESTING AND VALIDATION SERVICES

Your Sterile Processing Partner. Duraholder IPS

BIOTECH-GERMANDE ENDOSCOPE REPROCESSING WITH BRUSHLESS CLEANING PROCESS CONDITIONS. Lionel PINEAU PhD

EU Guidance / Q&A. Greg McGurk, GMP Manager (acting) GMP Conference. 7 February 2017 Dublin

Doc No.: MHD 12(0333)/ISO :2006 TITLE: Sterilization of health care products Biological indicators Part 1: General requirements

ASTM D10-F02 Workshop FDA Regulatory Perspective. Patrick Weixel FDA Center for Devices and Radiological Health

Objectives. The Perioperative Nurse s Role 6/24/2016. Instrument Processing & Sterilization Chesapeake Perioperative Consortium 2016

HANDLING OF WRIGHT DISPOSABLE PROPHECY ANKLE INSTRUMENTS

SUBJECT: Detection of Blood (hemoglobin) residue inside the biopsy channel of an endoendoscope

Safe to Handle? Comparing Manually and Machine-Washed Medical Devices

Tissue cleaning and sterilization provide an additional level of safety

GENERAL SURGERY DEVICES

Chemical Indicators. Biological Indicators. Physical Monitors. Welcome! How do I get a CE Certificate? Learning Objectives

SUBJECT: Detection of residual organic soils inside various channels of lumened items.

CAN STERILE REALLY BE STERILE?: EFFECTIVE AUTOCLAVING METHODS

Single-use bioprocess

Effect of Storage Conditions upon the Resistance (D-value) of Spore Strip Biological Indicators (BIs) used in EtO Sterilization Monitoring

Laboratory Testing Service Guide

Innovative. Smart. Trusted.

MICROBIOLOGICAL VALIDATION PROGRAMME FOR SRCL CLINICAL WASTE TREATMENT AUTOCLAVE PROCESSING METALS (SINGLE USE DEVICES) AT LEEDS

Microbiology Research Associates

CSA Z314 Series Standards Understanding the Standards

Parametric Release An Industry Perspective. October 22, Mike Sadowski Director Sterile Manufacture Support Baxter Healthcare Corporation

Why Do I Test, What Do I Test & When Do I Test It? Ross Caputo, PhD Chief Technical Officer Eagle Analytical Services

SUBJECT: Detection of Blood residue inside the biopsy channel of a scope

Guidance. Media Fills for Validation of Aseptic Preparations for Positron Emission Tomography (PET) Drugs DRAFT GUIDANCE

Loaner Programme Guidance

Loaner Programme Guidance

Enhancement will increase speed and accuracy of data collection during sample preparation. Additional space is needed for HCR scope testing

Overview and Introduction Annex 1 Revision

Post Sterilization Lab Contamination

Cleaning Efficiency of The MicroOne Wipe and Mop Cleaning System

3/11/ M Sterile U Network 3M Sterile U Web Meeting March 11, Welcome! Topic: An Overview of ANSI/AAMI ST58: 2013

MICROBIOLOGICAL CULTURE MEDIA

International and European Standardization of Sterilization Processes and Equipment

Main Sterilization Technologies

GUIDELINE C-17. Non-Incineration Technologies for Treatment of Biomedical Waste (Procedures for Microbiological Testing)

Risk based approaches for a sterility assurance application review: a microbiologist s perspective Dr. Yeissa Chabrier Roselló

Consistent. Reliable. Innovative.

L E X A M E D. Knowledgedriven. Compliancedriven Educationdriven Analysisdriven Clientdriven

Critical Environment Products and Services

Guide on the use of low energy electron beams for microbiological decontamination of surfaces

Specialty Testing of Steam Sterilizers

The Significance of Validation to US CSSD Managers

Biological Indicators

USP Chapter 823 USP 32 (old) vs. USP 35 (new)

Transcription:

STERILIZATION VALIDATION All sterilization processes require validation of the efficacy and reproducibility of the process. Depending on the type of sterilization, this may be accomplished by partial, sublethal, or repetitive processing, using representative product and/or biological challenges. WuXi AppTec offers a full range of services in this area from testing alone to full management of the validation. REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES The FDA expects manufacturers to validate all instructions for reusable devices, including cleaning, disinfection, sterilization parameters and dry times, if applicable. WuXi AppTec offers a comprehensive program for evaluation of cleaning and sterilization processes for reusable medical devices. Your WuXi AppTec Account Manager can provide you with initial information regarding any of these testing programs. Also working closely with you will be a highly trained and knowledgeable technical expert. IN THIS SECTION Radiation Sterilization Validation Method 1, Method 2, and VDmax + Dose Audits Service Options for Radiation Sterilization Validation Programs Service Options for Dose Audit Programs Dose Audit Reminder Program EO Sterilization Validation Reprocessing Validations for Reusable Medical Devices Overview Cleaning Efficacy Studies Inoculated Product Tests Sterilization Efficacy Studies VALIDATIONS WuXi AppTec F-1

RADIATION STERILIZATION VALIDATION For validation of radiation (gamma, electron beam or x-ray) sterilization, certain steps must be followed as outlined in ISO and AAMI standards. As part of the performance qualification, a dose-setting or dose substantiation study must be performed to demonstrate the adequacy of the minimum dose to achieve the desired sterility assurance level (SAL). Several methods are available for validation of the minimum SAL dose, and the choice of method is dependent on a number of variables. Complete radiation validation studies can be designed for a particular product and process, and all aspects of the studies follow the requirements of ISO and AAMI standards. ANSI / AAMI / ISO 11137 Method 1 Validation Total Bioburden Panel 10 products from three batches Page C-5 Bioburden Recovery Efficiency Test 3-5 products from any batch Page C-5 Sterility Method Suitability Test (B/F) One Medium 3 products from any batch Page C-15 Product Test of Sterility 100* samples from one batch Page C-16 NOTE: Single batch validations are also available. Dose Audit Aerobe or Total Bioburden Panel 10 products from one batch Page C-5 Product Test of Sterility 100* samples from one batch Page C-16 ANSI / AAMI / ISO 11137 Method 2 Validation The requirements of Method 2 are not outlined here because of the complexity of the sampling scheme. Please contact the laboratory for more information. Dose Audit Aerobe or Total Bioburden Panel 10 products from one batch Page C-5 Product Test of Sterility 100* samples from one batch Page C-16 * Reduced number of samples may apply, based on specific criteria. Contact Atlanta facility. F-2 WuXi AppTec VALIDATIONS

RADIATION STERILIZATION VALIDATION ANSI / AAMI / ISO 11137 VDmax 15 kgy or 25 kgy Validation and AAMI TIR 33 Total Bioburden Panel 10 products from three batches Page C-5 Bioburden Recovery Efficiency Test 3-5 products from one batch Page C-5 Sterility Method Suitability Test (B/F) One Medium 3 products from any batch Page C-15 Product Test of Sterility 10 samples from one batch Page C-16 NOTE: Single batch validations are also available. Quarterly Dose Audit Aerobic or Total Bioburden Panel 10 products from one batch Page C-5 Product Test of Sterility 10 samples from one batch Page C-16 VALIDATIONS WuXi AppTec F-3

ETHYLENE OXIDE (EO) STERILIZATION VALIDATION For validation of ethylene oxide (EO) sterilization, certain steps must be followed as outlined in ISO and AAMI standards. As part of the performance qualification, a microbiological challenge must be performed to demonstrate the adequacy of the process to achieve the desired sterility assurance level (SAL). One of the most utilized methods is the half-cycle (overkill) method, which uses a biological indicator (BI) challenge, typically 10 6 spores of Bacillus atrophaeus. Complete EO validation studies can be designed for a particular product and process, and all aspects of the studies follow the requirements of ISO and AAMI standards. SUB-LETHAL CYCLE STUDIES Total Bioburden Panel 10 products Page C-5 Bioburden Recovery Efficiency Test 3-5 products from any batch Page C-5 Inoculated Biological Indicators Dependent on load size Page C-11 Sterility Method Suitability Test (B/F) Two Media 6 products from any batch Page C-15 Product Test of Sterility Dependent on load size Page C-16 TERMINAL STERILIZATION STUDIES Biological Indicators Per client specifications Page C-14 EO Residual Panel Varies Page B-6 F-4 WuXi AppTec VALIDATIONS

VALIDATION PROGRAMS SERVICE OPTIONS SERVICE OPTIONS FOR VALIDATION PROGRAMS To help clients get their products through the validation process, WuXi AppTec offers different levels of service options for validation studies with varying degrees of involvement. Your choice of service levels would depend on how much of your company s time, manpower and expertise you want to commit to your validation program. For example, WuXi AppTec can provide only the testing services while you schedule and manage all aspects of the sterilization services and develop all the documentation to present an organized study. Or, in addition to the testing, we can handle all the sterilization services for you, and you would be responsible only for producing the final documentation. Or we can take care of everything. You simply choose the validation method, give us your product samples, and 5 to 7 weeks later we give you a completed validation and finished manual. SERVICE OPTIONS FOR DOSE AUDIT PROGRAMS Most products that are validated for radiation sterilization require periodic dose audits. As with validations, WuXi AppTec offers varying levels of service options for dose audits. DOSE AUDIT REMINDER PROGRAMS Dose audits are typically performed every three (3) months. However, the dose audit requirements for Method 1, Method 2 and VDmax validations may allow for a reduction in the frequency and in the case of Method 1 and 2 a reduction in the number of verification samples required, based on certain criteria. WuXi AppTec offers a dose audit reminder program that informs clients of their options, and issues reminders when dose audits are due for each product or family of products. Contact your Account Manager for more information about these programs. VALIDATIONS WuXi AppTec F-5

REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES When reusable medical devices are cleaned and sterilized in a health care facility, manufacturers are responsible for providing their customers with complete and comprehensive written instructions for handling, cleaning, disinfection and sterilization. The FDA expects manufacturers to validate all reuse instructions including cleaning and disinfection procedures, sterilization parameters, and dry times, if applicable. WuXi AppTec offers a comprehensive program for evaluation of cleaning and sterilization processes for reusable medical devices. Testing follows the guidelines outlined in AAMI TIR No. 12 and AAMI TIR 30 ( Designing, Testing & Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: A Guide for Device Manufacturers ). This program assists the manufacturer in meeting the requirements of the FDA Reviewer Guidance: Labeling Reusable Devices for Reprocessing in a Health Care Facility. PROTOCOL DEVELOPMENT A custom protocol is written for each study, tailored specifically to the device and the manufacturer s instructions for reuse. WuXi AppTec s scientific staff assists clients in assessing cleaning processes and developing protocols. CLEANING EFFICACY STUDIES Manufacturers must verify the efficacy of their recommended cleaning processes. Following the manufacturer s cleaning instructions, this study tests those processes using simulated soil inoculated with an appropriate marker. Those markers can be microbial (Gram positive, Gram negative, spore-forming) and/or physical markers (such as protein, carbohydrate, hemoglobin and TOC), which are representative of typical contamination. STERILIZATION EFFICACY STUDIES Manufacturers must provide health care facilities with detailed sterilization instructions for their particular medical device. Sterilization parameters are tested to determine capability of producing a sterility assurance level of at least 10-6. Studies are available for evaluating the following sterilization processes: Liquid chemical sterilants Ethylene Oxide (EO) Pre-vacuum, steam Gravity, steam DRY TIME VALIDATION STUDIES Manufacturers must provide heath care facilities with an effective dry time to be used in conjunction with the required steam sterilization cycle for reusable devices that are manufactured and supplied non-sterile to the user. Dry time validations will be based on the standard, full cycle steam sterilization cycle recommended by the manufacturer for the specified device in healthcare facilities as previously validated. At the conclusion of the sterilization cycle the devices will be exposed to the specified dry time for the evaluation of the effectiveness of that time to remove all residual moisture from the device as well as the sterilization wrap. SUPPORT FOR FUNCTIONALITY STUDIES These studies involve exposure to multiple cleaning and/or sterilization cycles as part of the functionality studies required to determine the useful life of a device. For more information on reprocessing validations, contact Technical Services at the WuXi AppTec Atlanta facility. F-6 WuXi AppTec VALIDATIONS

REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES STERILIZATION EFFICACY STEAM GRAVITY OR PRE-VAC CYCLES Product Inoculation 4 samples (3 test samples, 1 positive control) Page F-10 Autoclave Cycles 3 cycles, minimum Page F-12 Biological Indicators Varies per test article Page F-10 Inoculated Product Sterility Varies per test article Page F-10 STERILIZATION EFFICACY EO CYCLES [Hospital Cycles] Product Inoculation 4 samples (3 test samples, 1 positive control) Page F-10 EO Cycles 3 cycles, minimum Page F-12 Biological Indicators Varies per test article Page F-10 Inoculated Product Sterility Varies per test article Page F-10 CLEANING EFFICACY MICROBIAL / PHYSICAL CHALLENGE [Manual or Automated] Soiling Page F-8 Cleaning Page F-8 Spore Count for Inoculated Product Residual Protein Cleaning Efficacy Residual Carbohydrates Cleaning Efficacy 5 samples (3 test samples, 1 positive control, 1 negative control) Page F-8 Page F-9 Page F-9 Residual Total Organic Carbon (TOC) Cleaning Efficacy Page F-9 Residual Hemoglobin Cleaning Efficacy Page F-9 DRY TIME VALIDATION STEAM GRAVITY OR PRE-VAC CYCLES Autoclave Cycles 3 cycles, minimum Page F-12 Dry Time Evaluation 3 cycles, minimum Page F-12 VALIDATIONS WuXi AppTec F-7

REPROCESSING VALIDATIONS CLEANING EFFICACY STUDIES Manufacturers must verify the efficacy of their recommended cleaning processes. Following the manufacturer s cleaning instructions, this study tests those processes using simulated soil inoculated with an appropriate marker. Those markers can be microbial (Gram positive, Gram negative, spore-forming) and/or physical markers (such as protein, carbohydrate, hemoglobin and TOC), which are representative of typical contamination. 180115 Soiling Simulated soil is used to inoculate the devices prior to cleaning the devices per manufacturer s instructions. Various soils are available and the intended use of a device will determine which soil is most appropriate. 4 product samples (1 to be used as positive control) 180116 Manual 180117 Automated The manufacturer s instructions for cleaning are followed to determine the effectiveness of the process after devices are soiled. Cleaning may be performed manually or using an automated washer/disinfector. Cleaning 4 product samples (1 to be used as positive control) 160210 Spore Count for Inoculated Product After devices are cleaned per manufacturer s instruction, a bioburden test is performed to evaluate the effectiveness of the cleaning. Results are reported as a log reduction. 4 product samples (1 to be used as positive control) 400443 Residual Protein 400442 Residual Carbohydrates 400363 Residual Total Organic Carbon (TOC) 400444 Residual Hemoglobin After devices are cleaned per manufacturer s instruction, these tests are used to determine the residual concentration of various physical markers to evaluate the effectiveness of the cleaning. 10 to 50 ml extract from cleaned product samples Tests for Physical Markers See next page for descriptions of these individual tests. F-8 WuXi AppTec VALIDATIONS

REPROCESSING VALIDATIONS CLEANING EFFICACY STUDIES This test is used to determine the residual concentration of Total Organic Carbon (TOC) on reprocessed devices. 400363 50 ml extract Residual Total Organic Carbon (TOC) Cleaning Efficacy This test is used to determine the concentration of carbohydrates on reprocessed devices. A colorimetric method is used. 400442 E 10 ml extract Residual Carbohydrates Cleaning Efficacy This test is used to determine the concentration of protein on reprocessed devices. The Modified Lowry Protein Assay, a colorimetric method, is used. 400443 10 ml extract Residual Protein Cleaning Efficacy This test is used to determine the concentration of hemoglobin on reprocessed devices. 400444 10 ml extract Residual Hemoglobin Cleaning Efficacy VALIDATIONS WuXi AppTec F-9

REPROCESSING VALIDATIONS INOCULATED PRODUCT TESTS Inoculated product consists of actual devices or materials that have been inoculated with a specified level of a liquid biological indicator (BI) suspension. Inoculated products are used to validate and / or monitor certain sterilization processes. Testing is performed by product immersion using either the BI manufacturer s parameters or those found in USP, ISO or AAMI standards. Biological indicators (BIs) are carriers, such as a paper strip, that are inoculated with a specified level of a particular organism (typically bacillus species). BIs are used to validate and/or monitor certain sterilization processes. 1902000 B. atrophaeus 1902100 G. stearothermophilus Product Inoculation Devices are inoculated (usually in a location determined as most difficult to sterilize) with an indicator organism appropriate to the sterilization system in use. No minimum. Indicate required population and sterilization method. 1204100 Laminar Flow Hood Only Individual spore strips are transferred from their primary package to SCDM and incubated for recovery of the indicator organism. Biological Indicator Direct Transfer SHIPPING Spore strips (Client-provided positive control recommended.) Overnight air. Protect from temperature extremes. 1205100 Laminar Flow Hood Only Biological Indicator Within Product Spore strips that have been placed within a product or its package are retrieved from the product or package and transferred to SCDM for recovery of the indicator organism. SHIPPING Spore strips (Client-provided positive control recommended.) Overnight air. Protect from temperature extremes. F-10 WuXi AppTec VALIDATIONS

REPROCESSING VALIDATIONS INOCULATED PRODUCT TESTS Product that has been inoculated with a liquid spore solution and exposed to a sterilization process is tested in SCDM to detect surviving organisms. Inoculated Product Sterility Dependent on method and sterilizer volume. 1221010 Extra Small [ 100 ml of media] 1221000 Small [100 & 200 ml of media] 1221210 Medium [300 & 400 ml of media] 1221410 Large [500 & 600 ml of media] 1221610 Extra Large [800 & 1000 ml of media] 1221620 Jumbo [1200 & 1500 ml of media] 1221630 Extra Jumbo [2000 ml of media] Liquid samples, including spore suspensions and inoculated liquids, are enumerated to confirm spore population. 190501 Dependent upon expected population. Liquid Sample Population Confirmation Before using a new lot of BIs for sterilization load monitoring, the average population per unit should be independently confirmed per USP regulations. 190300 3 Sample Composite 120200 Single Sample Dependent on selected test. Biological Indicator Total Viable Spore Count VALIDATIONS WuXi AppTec F-11

REPROCESSING VALIDATIONS STERILIZATION EFFICACY STUDIES Manufacturers must provide health care facilities with detailed sterilization instructions for their particular medical device. Sterilization parameters are tested to determine capability of producing a sterility assurance level of at least 10-6. 800445 Autoclave Cycle Steam sterilization cycle run per client specifications. cycles are available Dependent on study protocol Gravity and pre-vacuum 180112 Data Logger Use for Validation Cycle Data loggers utilized during a steam sterilization cycle run per client specifications. Gravity and pre-vacuum cycles are available Dependent on study protocol 180111 Dry Time Evaluation Steam sterilization cycle run per client specifications. cycles are available Dependent on study protocol Gravity and pre-vacuum 800446 Ethylene oxide sterilization cycle run per client specifications. EO (Hospital) Cycle Dependent on study protocol 180110 Samples processed per client specified disinfection instructions. Disinfection Cycles Cycle Dependent on study protocol 1801500 Decontamination of Test Samples Samples are processed between validation cycles. The method of decontamination varies per client specifications. Dependent on study protocol F-12 WuXi AppTec VALIDATIONS