Lifetime stability of size exclusion chromatography columns for protein aggregate analysis

Similar documents
A universal chromatography method for aggregate analysis of monoclonal antibodies

ph gradient analysis of IgG1 therapeutic monoclonal antibodies using a 5 µm WCX column

NISTmAb characterization with a high-performance RP chromatography column

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

Simple charge variant profile comparison of an innovator monoclonal antibody and a biosimilar candidate

Application Note. Authors. Abstract. Biopharmaceuticals

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants

High-Resolution Charge Variant Analysis for Top-Selling Monoclonal Antibody Therapeutics Using a Linear ph Gradient Separation Platform

Thermo Scientific MAbPac HIC Columns. Novel Hydrophobic Interaction HPLC Columns. Designed for Monoclonal Antibody Analysis

Mobile Phase Optimization in SEC Method Development

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

Chromatography column for therapeutic protein analysis

Using the NISTmAb reference standard to demonstrate a simple approach to charge variant analysis

Molecular Characterization of Biotherapeutics The Agilent 1260 Infi nity Multi-Detector Bio-SEC Solution with Advanced Light Scattering Detection

Analysis of Monoclonal Antibody and Related Substances using a New Hydrophobic Interaction Chromatography (HIC) Column

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

Chromatographic Workflows for Biopharmaceutical Characterization

High-Resolution Separation of Intact Monoclonal Antibody Isoforms

Higher Order mab Aggregate Analysis using New Innovative SEC Technology

Columns for Biomolecules BioLC Column Lines

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Size Exclusion BioHPLC columns Ion Exchange BioHPLC columns

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

Separation Science Built for Biopharma

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

ProPac Elite WCX, 5 μm Particle, for Fast, High Resolution Protein and mab Analysis. July 2018

Thermo Scientific. MAbPac HIC-10. Product Manual. P/N: June Part of Thermo Fisher Scientific

High Throughput Sub-4 Minute Separation of Antibodies using Size Exclusion Chromatography

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Simple, Robust, High Quality Intact Mass Analysis A Biosimilars Case Study

Multiple Detector Approaches to Protein Aggregation by SEC

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

High-speed, High-resolution Oligonucleotide Separations Using Small Particle Anion-Exchangers

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column

Understanding the Effects of Common Mobile Phase Additives on the Performance of Size Exclusion Chromatography

Optimizing Protein Separations with Agilent Weak Cation-Exchange Columns

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

CX-1 ph Gradient Buffer

2012 Waters Corporation 1

Size Exclusion BioHPLC Columns

ProSEC 300S. Protein Characterization columns

Sepax Technologies, Inc.

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis

Use of alternative chromatographic phases and LC-MS for characterization of N-glycans from NISTmAb RM 8671

MAbPac RP Column. High-performance reverse phase chromatography column for monoclonal antibody analysis

A Novel ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge-Variant Analysis

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

CHARACTERIZATION & PURIFICATION

SEC-MS with Volatile Buffers for Characterization of Biopharmaceuticals

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

Characterization of IgG monomers & their aggregates

Analysis of biomolecules by SEC and Ion-Exchange UPLC

MAbPac SCX-10 Columns

IdeS-cleaved mab subunit analysis with LC-HRAM-MS: a quick and accurate comparison of biosimilar and originator biotherapeutics

ION EXCHANGE KIT FOR MAB SEPARATIONS

Evaluation of chromatographic phases for separation of differentially labeled glycans from erythropoietin and trastuzumab

Biotherapeutic Method Development Guide

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

TSK-GEL BioAssist Series Ion Exchange Columns

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

Fast Online Desalting of mabs Using a Reversed-Phase Desalting Cartridge for LC-MS Analysis

BioHPLC columns. Tim Rice Biocolumn Technical Specialist

Biotherapeutic Method Development Guide

Phenyl Membrane Adsorber for Bioprocessing

4/4/2013. BioHPLC columns. Paul Dinsmoor Biocolumn Technical Specialist. April 23-25, Size Exclusion BioHPLC Columns

Polishing of monoclonal antibodies using Capto S ImpAct

mab and ADC Analysis Shanhua Lin, Ph.D. The world leader in serving science

Preparative HPLC is still the

Faster Separations Using Agilent Weak Cation Exchange Columns

Size exclusion chromatography for Biomolecule analysis: A "How to" Guide

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs)

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

Protein Separation with ph Gradients Using Composite Buffer Systems Calculated by the Agilent Buffer Advisor Software

Superdex 200 Increase columns

The New Standard in High Resolution Size Exclusion

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

How to Create Your Own Affinity Titer Column. Kelly J. Flook*, Yury Agroskin and Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

Validation of a Dual Wavelength Size Exclusion HPLC Method with Improved Sensitivity to Detect Aggregates of a Monoclonal Antibody Biotherapeutic

Analysis of Monoclonal Antibodies and Their Fragments by Size Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer

Choosing the Right Calibration for the Agilent Bio SEC-3 Column

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

Growth media effects on post-translational modifications investigated through peptide mapping LC-MS/MS analysis of anti-il8 monoclonal antibody

Data File. HiLoad Superdex 30 prep grade HiLoad Superdex 75 prep grade HiLoad Superdex 200 prep grade. Pre-packed gel filtration columns

EAG.COM MATERIALS SCIENCES APPLICATION NOTE. By J. Tompkins 1, T. Spurgeon 1, R. Tobias 1, J. Anders 1, E. Butler-Roberts 2, and M.

Superdex 200 Increase columns

Agilent 1260 Infinity Bio-inert Quaternary LC System. Infinitely better for bio-molecule analysis

SMART Digest Kit User Manual. Version 2

Transcription:

APPLICATION NOTE 72362 Lifetime stability of size exclusion chromatography columns for protein aggregate analysis Authors Amy Farrell 1, Craig Jakes 1, Alexander Ley 2, Mauro De Pra 2, Frank Steiner 2, and Jonathan Bones 1 1 The National Institute for Bioprocessing Research and Training (NIBRT), Dublin, Ireland 2 Thermo Fisher Scientific, Germering, Germany Keywords MAbPac SEC-1, Monoclonal Antibody, mab, Protein Aggregation Analysis, Size Exclusion Chromatography, Biotherapeutics, Critical Quality Attributes, Vanquish Flex UHPLC Application benefits Columns for analysis of mab aggregates with exceptionally long lifetime A biocompatible UHPLC system capable of operating with high salt content and potentially corrosive mobile phases A UHPLC solution for mab aggregate analysis that operates continuously for several weeks Goal To demonstrate the long-term stability of the Thermo Scientific MAbPac SEC-1 column for monoclonal antibody aggregate analysis using the Thermo Scientific Vanquish Flex Quaternary UHPLC system. Introduction Monoclonal antibodies (mabs) are currently the dominant class of protein therapeutics in the biopharmaceutical industry due to their high specificity to target antigens, long serum half-life in humans, and capabilities for use in the treatment of a wide range of ailments such as inflammatory diseases and cancer. During product expression and purification from host

cells, formulation, and storage, mabs may undergo various degradation processes, which may alter the safety, efficacy, and quality profile of the drug product. Consequently, a number of critical quality attributes (CQAs) must be monitored throughout drug development and production to ensure biotherapeutic substances are suitable for clinical use. Aggregation is a common degradation process for therapeutic proteins that can result from partial unfolding or other types of conformational changes in protein structure to form dimers, trimers, and other higher order structures. Protein aggregation is considered a CQA as aggregates may reduce product efficiency by lowering the effective concentration of the product and have been found to result in immunogenic effects in patients. Hence, it is a regulatory requirement to monitor the aggregation profile of therapeutic proteins. Size exclusion chromatography (SEC) is the most commonly applied method for protein aggregate analysis. In SEC, sample molecules are passed through a column containing porous polymer or silica beads to facilitate separation of species based on their size. For protein therapeutics, a pore size of 300 Å is used, which allows smaller species to penetrate into the porous beads (e.g. fragments, monomers) while larger molecules (e.g. dimers, trimers) are more excluded from the pores and therefore elute more quickly from the column. Most frequently, SEC columns with widths of 7.8 mm are used for aggregate analysis in industry. However, use of low dispersion, biocompatible UHPLC systems and narrow internal diameter pre-column tubing has been shown to improve peak shape in chromatograms generated using columns of reduced widths. 1 Narrow columns have advantages such as small sample volume requirements and reduced mobile phase consumption. Due to high costs associated with drug development and production and the potential for mab-based biosimilars resulting from patent expiry for some of the top-selling mab therapeutics, reliable, long-use consumables and equipment for CQA evaluation is needed. Suppliers of SEC columns have illustrated column stability lifetimes of approximately 550 injections (without a column guard) and up to 902 injections (with a column guard). In this study, a MAbPac SEC-1 column (4 300 mm, 5 µm, 300 Å) and a Vanquish Flex Quaternary UHPLC system were applied to assess the long-term stability of the column based on protein aggregation monitoring of the commercial drug product bevacizumab, a humanized monoclonal IgG1 antibody produced from a Chinese hamster ovary mammalian cell expression system. Use of pre-column tubing with internal diameter of 75 µm ensured that excellent peak shape was achieved using a SEC column with dimensions of 4 300 mm. However, as SEC columns with widths of 7.8 mm are most commonly utilized in the biopharmaceutical industry, the column lifetime stability of an MAbPac SEC-1 column (7.8 300 mm, 5 µm, 300 Å) was also evaluated. Experimental Recommended consumables Deionized water, 18.2 MΩ cm resistivity Thermo Scientific MAbPac SEC-1, 5 µm, 300 Å, 4.0 300 mm (P/N 074696) Thermo Scientific MAbPac SEC-1, 5 µm, 300 Å, 7.8 300 mm (P/N 0088460) Fisher Scientific sodium phosphate dibasic anhydrous (P/N 010440481) Fisher Scientific sodium phosphate monobasic anhydrous (P/N 010751135) Fisher Scientific sodium chloride (P/N 011964051) Thermo Scientific Virtuoso Vial Identification System (P/N 060180-VT100) Virtuoso 9 mm Wide Opening SureStop Screw Thread Vial Convenience Kit (P/N 060180-VT405) Sample preparation Bevacizumab (25 mg/ml) was diluted 1:10 in 100 mm sodium phosphate, ph 6.8, 300 mm NaCl (for MAbPac SEC-1 4.0 300 mm column) or 50 mm sodium phosphate, ph 6.8, 300 mm NaCl (for MAbPac SEC-1 7.8 300 mm column). Diluted samples were aliquoted and stored at -20 C. On each day of analysis, a fresh aliquot of diluted bevacizumab sample was removed from the freezer for analysis. 2

Buffer preparation MAbPac SEC-1 4.0 300 mm column: 100 mm sodium phosphate, ph 6.8 in 300 mm NaCl was used as mobile phase. Buffers were filtered through a 0.2 µm filter membrane before use. MAbPac SEC-1 7.8 300 mm column: 50 mm sodium phosphate, ph 6.8 in 300 mm NaCl was used as mobile phase. Buffers were filtered through a 0.2 µm filter membrane before use. Separation conditions Instrumentation Vanquish Flex UHPLC system, including: Quaternary Pump (P/N VF-P20-A) Column Compartment H (P/N VH-C10-A) Split Sampler FT (P/N VF-A10-A) with 25 µl Sample Loop Diode Array Detector HL (P/N VH-D10-A) with LightPipe 10 mm Standard Flow Cell (P/N 6083.0100) System Base Vanquish Flex (P/N VF-S01-A) Flow rate MAbPac SEC-1 4.0 300 mm column: 0.250 ml/min MAbPac SEC-1 7.8 300 mm column: 0.8 ml/min Column temperature 30 C Injection details MAbPac SEC-1 4.0 300 mm column: 2.0 µl of 2.5 µg/µl bevacizumab, diluted in mobile phase. MAbPac SEC-1 7.8 300 mm column: 0.5 µl of 2.5 µg/µl bevacizumab, diluted in mobile phase. Detector settings MAbPac SEC-1 4.0 300 mm column: Data was collected at 280 nm and 214 nm on a DAD detector. Data processing The Thermo Scientific Chromeleon 7.2 SR4 Chromatography Data System was used for data acquisition and analysis. Results and discussion To evaluate the stability of a MAbPac SEC-1 analytical column for aggregate analysis of therapeutic monoclonal antibodies, repeated injections of bevacizumab drug substance were performed using a Vanquish Flex UHPLC system with Chromeleon software version 7.2. The MAbPac SEC-1 column is a silica-based SEC column covalently modified with a proprietary diol hydrophilic layer to prevent secondary interactions of analytes to the stationary phase. A new and unused column was employed for the study and was conditioned by performing ten injections of mab until a constant peak area and peak height were observed as per the manufacturer s guidelines. Each day of analysis and for each new buffer preparation, blank injections were performed prior to mab determination to ensure column equilibration. The industry-standard SEC column dimension for protein aggregate analysis is 7.8 mm internal diameter (i.d.) using flow rates of 1 ml/min. It has previously been shown that pre-column dispersion often associated with reduced i.d. SEC columns (4 mm) and low flow rates (<0.3 ml/min) may be overcome using a low-dispersion, biocompatible UHPLC, such as the Vanquish Flex UHPLC, with smaller injection volumes and fitted with narrow i.d. pre-column transfer tubing (75 µm i.d. tubing). 1 Using 75 µm pre-column transfer tubing, a 4 mm i.d. MAbPac SEC-1 column was evaluated. Over the course of the study excellent resolution of aggregates and fragment peaks from the main mab monomer peak was observed as shown in Figure 1. MAbPac SEC-1 7.8 300 mm column: Data was collected at 280 nm on a DAD detector. 3

Figure 1. Aggregate analysis of bevacizumab using a MAbPac SEC-1 column. As Table 1 shows, consistent retention times were observed for the monomer peak over the course of more than 1950 injections, with a retention time difference of just 0.043 minutes between injection numbers 25 and 1953; this illustrates the absence of chemical or electrostatic interactions with the column stationary phase following extensive use. Similarly, excellent peak symmetry was detected throughout the column lifetime stability study (asymmetry range 0.88 to 0.93), further demonstrating a lack of secondary interactions with the column packing material and hardware across the lifetime of the column. Column efficiency, based on European Pharmacopoeia theoretical plate count, was found to be greater than 85% of initial efficiency following 1953 injections on the column (equivalent to 1866 injections of mab). After 1866 injections of bevacizumab, a loss in column performance was observed as shown in Figure 2. The loss in column performance (<85% theoretical plate count) did not appear to result in a change to the aggregation profile of bevacizumab. Table 1. Monomer peak information acquired following aggregate analysis of bevacizumab using a MAbPac SEC-1 analytical column. On Column Injection # Retention Time Monomer Relative Peak Area (%) Monomer Peak Width @ 50% Height Asymmetry Theoretical Plates 25 9.554 96.96 0.237 0.92 9032 404 9.558 97.21 0.240 0.92 8797 665 9.558 96.96 0.241 0.93 8736 1213 9.567 97.27 0.243 0.89 8604 1551 9.544 96.99 0.244 0.88 8460 1953 9.524 96.30 0.255 0.89 7737 4

10,000 Monomer Theoretical Plate Count 8000 N 6000 4000 2000 0 0 250 500 750 1000 1250 1500 1750 2000 Bevacizumale Injection Number Figure 2. Theoretical plate count for the mab monomer in each injection using the MAbPac SEC-1, 4.0 300 mm column. Greater than 85% of initial column efficiency was preserved following 1866 injections of bevacizumab (corresponding to 1953 on column injections). A loss of column performance was observed after 1866 injections of bevacizumab. Because SEC columns with internal diameters of 7.8 mm are most often applied for aggregate analysis in the biopharmaceutical industry, the column lifetime stability of a MAbPac SEC-1, 5 µm, 300 Å, 7.8 300 mm column was also evaluated. In this case, a new and unused column was also employed for the study and was conditioned prior to use following recommended guidelines. During analysis the column was injected ten times with bevacizumab, followed by two injections of a protein check standard. This injection cycle was repeated until column degradation was observed. Table 2 shows highlighted chromatography data for bevacizumab analyzed using a MAbPac SEC-1, 7.8 300 mm column. The final line on the table displays information for the final injection that was above the column efficiency specification for monoclonal antibodies, namely injection number 1292 (>6300 theoretical plates for the monomer peak). Like the MAbPac SEC-1, 4.0 300 mm column, the MAbPac SEC-1, 7.8 300 mm column displayed relatively consistent retention time and peak symmetry and excellent column efficiency over the lifetime of the column (Figure 3). Despite exhibiting reduced column lifetime stability in comparison to the MAbPac SEC-1, 4.0 300 mm column, the MAbPac SEC-1, 7.8 300 mm column showed extended column lifetime stability when compared to the lifetime stability previously demonstrated for alternative SEC columns. Greater than 55% of initial column efficiency was preserved following 1077 injections of bevacizumab (corresponding to 1292 on column injections). A loss of column performance was observed after 1077 injections of bevacizumab. The efficiency drops observed between 300 and 500 injections were not attributed to column degradation. It was observed that sample aging negatively affected peak width. When sample was replaced, high efficiency values were restored. The root cause for this behaviour could not be found. When the column degraded, the replacement of aged sample with fresh did not have any beneficial effect on peak width. Table 2. Monomer peak information acquired following aggregate analysis of bevacizumab using a MAbPac SEC-1, 7.8 300 mm analytical column. On Column Injection # Retention Time Monomer Relative Peak Area (%) Monomer Peak Width @ 50% Height Asymmetry Theoretical Plates 30 11.192 97.56 0.247 1.08 11349 236 11.183 97.30 0.252 1.11 10870 555 11.192 97.74 0.252 1.11 10918 769 11.200 97.95 0.253 1.06 10876 975 11.225 97.92 0.262 0.93 10158 1292 11.225 97.40 0.333 1.08 6301 5

12,000 Monomer Theoretical Plate Count 10,000 8000 N 6000 4000 2000 0 0 150 300 450 600 750 Injection Number 900 1050 1200 1350 1500 Figure 3. Theoretical plate count for the mab monomer in each injection using the MAbPac SEC-1, 7.8 300 mm column. Conclusion The MAbPac SEC-1 column coupled to a Vanquish Flex Quaternary UHPLC system is a robust platform for aggregate analysis of mabs. Reference 1. Thermo Scientific Application Note AN21602: The importance of correct UHPLC instrument set-up for protein aggregate analysis by size-exclusion chromatography, 2016. Consistent retention time, excellent peak symmetry, and exceptional column efficiency were observed over the course of 1953 on-column injections for the MAbPac SEC-1, 4.0 300 mm column and for 1292 on-column injections on the MAbPac SEC-1, 7.8 300 mm column. To our knowledge, the MAbPac SEC-1 column lifetime stability determined far exceeds other commercially available SEC columns, without requirements for guard columns. Find out more at thermofisher.com/proteinaggregates 2017 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representatives for details. AN72362-EN 0717S