Capillary GC Troubleshooting: a Practical Approach

Similar documents
GC Troubleshooting. Frank Michel, Katherine K. Stenerson T HCE

Ghost Peaks in Gas Chromatography Part 2: The Injection Port

Gas Filtration Catalogue

GC Consumables Product Guide C184-E034

Super Clean Gas Filters Simply the best

Super Clean Gas Filters Simply the best

Improving GC-MS Method Robustness and Cycle Times Using Capillary Flow Technology and Backflushing

Installation, Care and Maintenance of Capillary Gas Chromatography Columns. Mark Sinnott Application Engineer January 2015 Happy New Year!

Analysis of Semivolatiles by GC/FID using the Ultra Inert Inlet Liners with Wool

2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial waste waters.

Gas Chromatography Assignment Chem 543/443

Please enclose a item wise compliance report along with the technical bid.

Content Page OPTIC Liners OPTIC 2 OPTIC 3 OPTIC-4 LINEX PHASER

MARKET FOCUSED - environmental ORGANOCHLORINE PESTICIDES

Varian Supplies/Parts. GC Accessories. Septum Nut. Merlin Microseal Kit

Tips on Making your GC System and Analysis more Robust. Agilent Restricted

GC Method Development

A Direct Column-Performance Comparison for Rapid Contract Laboratory Program (CLP) Pesticide Analysis. Application. Authors. Abstract.

Dräger Gas Filtration SUPER CLEAN GAS FILTERS. Dräger: Technology for Life

Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D7504 on the Agilent 8890 GC System

Agilent Consumables Improve Performance with Shimadzu GC/MS

PerkinElmer Supplies/Parts. GC Accessories. Autosampler Syringes

Super Clean TM Gas Filters

GC Method Developement. Simon Jones GC Application Engineer GC Columns and Consumables

Contract Laboratory Program (CLP) Pesticide Analysis with 0.18 mm ID High-Efficiency GC Columns Utilizing Helium Carrier Gas. Application.

Gas Chromatography SSI-GC-003

Olefin Analyzer Solutions for Polymer Grade Ethylene and Propylene

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations

Liner Style* Function Advantages Disadvantages Recommended For Straight Low surface area for less activity Least expensive Low activity

PAL SPME Fibers Optimized for Automation

PAL SPME Fibers Optimized for Automation

Prefractionator for Reliable Analysis of the Light Ends of Crude Oil and other Petroleum Fractions

When optimization is discussed throughout this talk, it s in terms of performing more samples, in less time, while minimizing investment in capital

Improved Gas Chromatograph Method for the Analysis of Trace Hydrocarbon Impurities in 1, 3-Butadiene

Direct Injection of Fish Oil for the GC-ECD Analysis of PCBs: Results Using a Deans Switch With Backflushing. Application. Author.

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

Chromatography Applications and Technology

Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP1 and DB-CLP2

GC Method Developement

Installation, and Maintenance of Capillary Gas Chromatography Columns

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D*

Gas Chromatography (GC) Operation, Calibration, Troubleshooting and Practice Maintenance

APPLICATIONS BENEFITS INTELLIGENT

Technical Article. Innovative Chromatography Solutions. The Rtx -CLPesticides Column Set Lets You Make the Choice

Liners and consumables for OPTIC Inlet. Liner Selection Guide

Volatile Organic Carbon Contaminated Site Assessment

Corn Syrup Analysis E-1a- 1. ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application

CORESTA Recommended Method No. 57

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

TRACE 1300 and TRACE 1310 Gas Chromatographs Dual Detector Microfluidics Installation Guide

Fast Refinery Gas Analysis Using the Agilent 490 Micro GC QUAD

Agilent Ultimate Plus Fused Silica Tubing

Alumina BOND/MAPD PLOT Columns!

High temperature Simdist analysis using the new MXT-1 HT Simdist phase. Jaap de Zeeuw and Gary Stidsen Restek corporation, Bellefonte, PA, USA

Residual Solvents in medical cannabis analysis using the SRI 8610C FID GC June 2014

Pesticide Analysis in a Complex Matrix using a 2D GC/MS System

STANDARD OPERATING PROCEDURES

Large Volume Injection of Organochlorine Pesticide Extracts Using a Programmable Temperature Vaporizer in the Solvent Split Mode

The Use Of Automated Backflush on the 7890A/5975A GC-MS System. Application. Setup Example Using Biomarkers. Authors. Abstract.

Reduced carryover and increased reproducibility in the analysis of cold pressed pink grapefruit essential oil

Low-level Detection of Ethanol, 1,4-Dioxane, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES

Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730

Gas Generators. Parker Balston

The Suite for Environmental GC Analysis

Miniature Helium Purifier and Nitrogen Purifier Instruction Manual

DETECTORS Dry Electrolytic Conductivity Detector - DELCD. Overview

3-IN-1 TECHNOLOGY Highest Inertness Lowest Bleed Exceptional Reproducibility

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE.

SCION GC Series. The Gas Chromatographers Choice for Separations

Operating the SRI RGD ( Reduction Gas Detector )

Sampling Techniques Handbook. Agilent Automatic Liquid Samplers

7890A An Essential Partner for the Trace

Hydrocarbon Processing Industry Gas Chromatographs. Standard GC Analyzer Solutions

SSENTIALS RESTEK. Must-Have Solutions for Your Lab Leverage 3-in-1 Technology for Outstanding Inertness Low Bleed High Reproducibility

Hydrogen Generators for Fuel Gas

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

Analyze more samples. faster. while maintaining resolution. NEW Agilent J&W High Efficiency Capillary GC Columns

GC Columns GC COLUMN INSTALLATION GUIDE.

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC

FINAL REPORT. Participants in the reported tasks 1. Ravi Subramanian 2. Manoranjan Misra 3. Kent Hoekman

GAS CHROMATOGRAPHY CLARUS 400/480 GC

PROCESS ANALYZERS FOR PURIFIED BULK GAS QC/QA: ENTERING THE PPT RANGE

TRACE ANALYTICAL. ta7000 Series UHP Gas Purity Monitors At the critical link in your manufacturing process. Features /Benefits

Productivity and accuracy enhancements

Chemical Analysis. DHA Analyzer Family. Optimized Solutions for Detailed Hydrocarbon Analysis. Gas Chromatography. think forward

Standard Operation Procedure (SOP) for the Extraction and Instrument Analysis of Siloxanes from Cartridges

ACETALDEHYDE. The procedure is applicable to corn syrups including those containing fructose, and to most other starch hydrolyzates (Note 1).

Improving Productivity in Your Environmental and Food Testing Labs with Agilent 7890A/5975C GC/MS System

Capillary Flow Technology: Backflush REDUCE RUN TIME AND INCREASE LABORATORY THROUGHPUT

GC-C and GC- Pyrolysis IRMS Short Course: Putting the Gas In Gas Chromatography. Paul Eby

NEW OUTSIDE DIAMETER FUSED SILICA FOR AGILENT J&W CP AND VF GC COLUMNS

Markes International Thermal Desorption G8125A, G8126A, G8127A, G8128A, G8131A, G8132A, G8136A, G8133A, MKI-MicroC Site Preparation Checklist

Electron Capture Detector for GC-2010 Plus

SPECIFICATION. The System shall be easy to use and rugged enough to handle consistently high through performance of the activities.

TVOC Heated FID VE 7

Transcription:

Capillary GC Troubleshooting: a Practical Approach Katherine Stenerson Gas Separations R&D Supelco, Supelco Park, Bellefonte, PA 16823 1999 Sigma-Aldrich Co.

Outline Basic Troubleshooting Strategy Preventing Problems Identifying Common Problems Recommended Reading Discussion

Troubleshooting Strategy Have appropriate equipment and supplies on hand. Establish a systematic approach. Know what to look for first. Record what you did to correct the problem.

Suggested Equipment to Have on Hand for Troubleshooting: Electronic Leak Detector Flow Meter Test Column Replacement Accessories (Syringes, Ferrules, Septa) Replacement Purifiers

Five Major Sources of Chromatographic Problems: ❶ Operator Error ❷ The Sample ❸ The Column ❹ The GC Electrical System ❺ The Gas Flow System (both internal and external to the GC)

Approaching the Problem Check first to see if a fix for the problem is already known. Check the Supelco Capillary GC Troubleshooting Guide (Bulletin 853.) Check back in the instrument maintenance record. Talk to others in your lab.

Isolate the Source of the Problem Run Ref. Std. OK Check operating parameters Not OK Correct the parameter OK Not OK Problem was sample related Install Test Column OK Problem was Column related Not OK Cap off Detector OK Not OK Problem is in the Inlet or with the carrier gas Problem is in the Detector

When reviewing operating method parameters consider the following:? Is my starting temp. low enough to allow sufficient sample focusing?? For splitless injections, is my splitter opening at the appropriate time?? Is my column flow set to give me maximum efficiency at the most critical point?? Are heated zones (injectors, detectors, interfaces) set appropriately?

The Best Way to Solve Problems is to Prevent Them! Install and maintain proper purification for all gases in the GC system. Maintain the injector by periodically inspecting and changing the liner, septa, and seal (H/P.) Use the proper injection technique-this includes using the right liner for the job. When necessary, use a guard column to protect the analytical column.

Gas Purification Carrier Gas At minimum, remove hydrocarbons, water, and oxygen. Hydrogen (FID) At minimum, remove hydrocarbons. Air (FID) At minimum, remove water and hydrocarbons. Nitrogen make-up (FID, ECD) At minimum, remove hydrocarbons. P-5 make-up (ECD) At minimum remove hydrocarbons (especially halogencontaining), oxygen.

Acceptable Purity Levels for Chromatography Grade Gases Impurity / Maximum Concentration Gas O2 H2O CO2 CO Total Hydrocarbons Helium <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm Nitrogen <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm Air 20-22% <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm Hydrogen <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm Argon/ Methane <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm <1.0 ppm

P-5 Suggested purifiers Hydrocarbons Water Oxygen Carrier Supelcarb HC Supelpure HC Mole Sieve 5A OMI -2 H 2 Air Mole Sieve 5A N 2 makeup P-5 OMI -2 OMI -2

What are some signs that my purifiers need to be changed? Hydrocarbon Traps Noise in the baseline (FID) Increase in background peaks on tune (MSD) Higher than normal baseline reading on FID Extra peaks visible in run Molecular Sieve 5A Increase in column bleed Water visible in MS background Poor peak shapes for gaseous VOCs (purge and trap) Extra peaks visible in run OMI -2 color change

Injector Maintenance Change (as needed): 1. Liner and O-ring * 2. Seal and washer * * H/P GCs Inspect the inlet periodically -Look for contamination in the liner -Look for residue on the seal

Using the right liner and injection technique can also prevent problems: Split Injection used for concentrated samples high flow of carrier gas through liner during injection should use liner designed for split injection Splitless Injection used for trace analysis low flow of carrier gas through liner during injection inertness and internal volume of liner used are critical

Some liners used for split injection Cup (unpacked) Cup (wool packed) Split/splitless- wool packed

Some liners used for splitless injection 2 mm ID, straight Dual-tapered Single-tapered

If you must clean a liner. Handle liners with gloves or forceps. Use clean compressed gas and/or a fine brush to remove particles. Rinse liner in an appropriate solvent and dry with clean compressed gas. Use mineral acid and/or detergent only if absolutely necessary.

Using a Guard Column Choose a guard column which has been deactivated. Usually, the ID of the guard matches the analytical column. A 5-10 meter length is normally used. Connect with either a GlasSeal or butt connector.

Common Problems 1 Poor Peak Shapes (either tailing, fronting, or just generally ugly.) 2 Nonlinearity 3 Baseline Noise and /or Drift 4 Ghost Peaks 5 Missing Peaks / Poor Response 6 Insufficient Resolution

1. Poor Peak Shapes Fronting can indicate column overload. Tailing can indicate activity in the system or improper column installation.

1. Poor Peak Shapes (cont.) Generally ugly peaks, such as a,adimethylphenethylamine, can be caused by a variety of problems.

2. Nonlinearity The most common causes are: Column overload Detector overload Standards preparation Poor peak shape resulting in improper integration

An Example of Overload: Peaks have broad bases, fronting on last few visible in run.

Preventing column overload for splitless injections: Inject a smaller amount / use a 1 ul syringe. Use a thicker film column. Use a column with a wider ID. Decrease upper limit of calibration range. Use a column of slightly different polarity.

An example of poor peak shape affecting linearity: Benzoic acid is typically of poor shape when doing splitless injections.

3. Common causes of baseline noise / drift. Column bleed Dirty detector Contaminants in carrier gas / carrier gas purity

Effect of carrier gas purity on baseline noise: 12000 11500 11000 10500 10000 9500 9000 8500 8000 7500 7000 6500 6000 5500 5000 4500 4000 3500 3000 2500 2000 1500 1000 500 0 1 H2 carrier from tank H2 carrier from a generator

Column bleed results from the normal degradation of the stationary phase. All columns bleed to some extent. Bleed increases with temperature. The amount of bleed will increase in the presence of oxygen.

A Typical Bleed Profile: 2 Bleed measured as the difference between 1 and 2. 1 4

Column Bleed on an MSD Visible as baseline rise in the TIC. Check spectra for key bleed ions: PTE -5: 207, 281 SPB -1: 73, 207, 281 SPB -624: 207, 269 Make sure interface temp. is < column max. temp.

Ion 207 corresponds to a fragment known as D3: CH 3 CH 3 CH 3 O Si Si O O + Si CH 3 CH 3

MS spectra of bleed from a PTE -5 Column Abundance 6000 5500 Scan 2895 (27.487 min): 1118015.D 207 207 5000 4500 4000 3500 3000 2500 281 2000 1500 73 96 133 281 1000 191 253 500 50 115 162 223 327 355 415 0 40 60 80 100120140160180200220240260280300320340360380400 m/z-->

MS spectra of bleed from an SPB -1 Column Abundance 11500 11000 10500 10000 9500 9000 8500 8000 7500 7000 73 73 Scan 1941 (28.793 min): 0122002.D 6500 6000 5500 5000 44 207 4500 4000 3500 3000 207 281 2500 2000 1500 147 281 341 1000 96 119 179 500 253 429 313 401 0 40 60 80 100120140160180200220240260280300320340360380400420 m/z-->

MS spectra of bleed from an SPB -624 Column Abundance 24000 Scan 8581 (33.704 min): 1022006.D 207 22000 20000 207 18000 16000 14000 12000 269 10000 269 8000 6000 44 253 4000 133 191 73 96 147 177 2000 119 283 322 343 59 163 223 239 298 0 40 60 80 100 120 140 160 180 200 220 240 260 280 300 320 340 m/z-->

MS Spectra of Septa Bleed Abundance 44000 42000 40000 73 Scan 604 (7.460 min): 1201001.D 38000 36000 34000 73 32000 30000 28000 26000 24000 147 22000 20000 18000 16000 14000 147 281 281 12000 10000 8000 m/z--> 6000 4000 2000 0 45 40 60 80 100120140160180200220240260280300320340360380400420 327 207 399 415 131 95 191 115 163 223 251 297 343 383

Prevent column bleed! Sufficiently purge column with carrier before ramping it up in temperature. Make sure carrier gas is filtered for water and oxygen. Check integrity of all fittings leading to the column. Do not heat the column above its maximum temp. Precondition the column prior to use.

4. Ghost Peaks Residue in the inlet liner and at the head of the column Contaminated syringe / and or wash solutions on an autosampler. Sample carryover Contaminated carrier gas

If pieces of septa get into an inlet liner... 2200000 2000000 1800000 1600000 1400000 Response test mix, before 1200000 1000000 800000 600000 400000 200000 0

even a simple analysis can be ruined. 3500000 3000000 2500000 2000000 Response test mix, after 1500000 1000000 500000 0 Time-->

5. Missing Peaks / Poor Response Sample decomposition Activity in the inlet or column Injection port temperature too high Sample not stable enough for GC Standards not stable Coelution Insufficient run time / final temperature Sample not volatile enough for GC Improper column installation

2-methyl-3,5-dinitrophenol Nasty samples can damage a column by creating active sites: 450000 Before Sample Injection 400000 350000 300000 250000 200000 2,4-dinitrophenol 4-nitrophenol pentachlorophenol 150000 100000 50000 0

Pentachlorophenol should be here Here, the responses of some acids were affected: Abundance After sample injection 900000 800000 700000 600000 500000 400000 300000 200000 2,4-DNP & 4-NP should be here 2-methyl-3,5-dinitrophenol 100000 0 Time-->

Response can also be affected by the position of the column in the inlet: 2800000 2600000 2400000 2200000 2000000 1800000 8 mm above top of ferrule 1600000 1400000 1200000 1000000 800000 600000 400000 200000 0

Here, the column was not inserted far enough: 2200000 2000000 1800000 1600000 1400000 5 mm above top of ferrule 1200000 1000000 800000 600000 400000 200000 0

Here, the column was inserted too far: 1800000 1600000 1400000 1200000 20 mm above top of ferrule 1000000 800000 600000 400000 200000 0

6. Insufficient Resolution Wrong column Longer columns increase resolution. Smaller ID columns increase resolution. A different phase altogether may be needed. Wrong Conditions Carrier gas flow too fast or slow. Oven ramp rate too fast.

Recommended Reading Supelco Bulletins 1. 741: The Supelco Guide to Leak-Free Connections 2. 783: Cleaning Flame Ionization Detectors 3. 853: Capillary Troubleshooting Guide 4. 875: Supelco Capillary GC Selection Guide 5. 895: Installation and Maintenance Instructions for.25 mm and.32 mm ID Fused Silica Capillary Columns 6. 897: Installation and Maintenance Instructions for.53 mm ID Fused Silica Capillary Columns 7. 898: Gas Management Systems for GC 8. 899: Capillary GC Inlet Sleeve Selection Guide 9. 916: Purge and Trap System Guide 10. 918: Selecting Purifiers for Gas Chromatography

Help is just a phone call or mouse click away! Supelco Supelco Technical Service phone: 1-800-359-3041 email: techservice@supelco.sial.com Supelco Supelco Customer Service phone: 1-800-247-6628 email: supelco@sial.com Sigma-Aldrich Website www.sigma-aldrich.com